FA4000 TOKYOKEISO | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 5
Technical content
˙GENERAL The FA4000 series is a local indicating flowmeter with an alarm output for liquid measurement. It covers a wide range of applications from as small as 0.1 to 1 L/min to as large as 13 to 130 L/min in measuring range, from 10 mm to 40 mm in size. ˙FEATURES ❑ COMPACT AND LIGHT DESIGN FA4000 Series is very much suitable for assembling onto equipment and devices due to its compact and light design. ❑ WATERPROOF CONSTRUCTION Suitable for opaque liquid s or hot pressurized water. ❑ WIDE APPLICATION Opaque liquid or pressuriza tion hot water is available. ❑ ALARM SET POINT The alarm set point can be set at the front of the FA4000 Series using the alarm setting pointer in its housing. ❑ QUICK DELIVERY AND COMPETITIVE PRICE The standardized specification allows reasonable quick delivery time. ˙STANDARD SPECIFICATION
- Measuring fluid : Water or Low viscosity liquids
- Measuring range : (Min.) 0.1 to 1 L/min : (Max.) 13 to 130 L/min
- Flow temp. : 0 to 100°C *: It is general data, and the maximum temperature may change by terms of use and environment. (When fluid temperature exceeds 80°C, the material of O-ring is FPM.)
- Fluid press. : Max. 1.0MPa
- Flow direction : Bottom to To p, Left to Right, Right to Left,Top to Bottom
- Connection size : 10mm (3/8") to 40mm (1 1/2")
- Process connection : Rc, NPT, JIS10K FF , JIS10K RF , ANSI150Lb RF , JPI150Lb RF
- Indication accuracy : ±5% F .S.
- Alarm setting accuracy : ±2% F .S.(Against flow calibration)
- Alarm setting range : 10 to 100% (F .S.)
- Alarm contact : 1 point (High or Low)
- Alarm reset span : Less than 15% F .S. (Against flow calibration)
- Alarm switch : Self holding type
- Contact capacity : 100 V DC/10W, 125 V AC / 10VA
- Insulation resistance : 100M Ω or higher at 500 V DC megger
- Withstand voltage : 1500 V AC (1 min.)
- Alarm setting : Set by screw (Freely adjustable from housing front)
- Finishing painted : Metallic silver (Except Stainless steel material)
- Cable entry : G1/2 (Female thread)
- Wiring connection : M3 screw terminal
- Housing construction : Equivalent to IP65 Intrinsically safe (Supplied with safety barrier)
- Installation : Piping support ˙RECOMMENDED APPLICATION ❑ Cooling water line ❑ Monitoring of leakage of sealing liquids ❑ Cooling fluids lines in injection moulding machines ˙BEFORE OPERATING 1) The FA4000 Series has a magn et coupling to indicate flow rate. Provide a magnet strainer which is supplied on request in the upstream of the piping to remove the ferrous materials like iron powders if they ar e contained in the fluid. 2) Add the relay driver to increase the contact capacity if more than the allowable contact capacity of the FA4000 Series is required. 3) Use an intrinsically safe relay if the FA4000 Series with the alarm output is used in the hazardous area. The magnet strainer, relay driver and intrinsically safety relay mentioned above are available on request. See “OPTIONAL UNIT” at the last page for detail specifications. FA4000 Series FLOW MONITOR 5('& "QS +VM 3FWJTFE TUFEJUJPO
FA4000 Series FLOW MONITOR 2 TOKYO KEISO CO., LTD. TG-F966-8E ˙EXTERNAL DIMENSIONS AND MATERIAL L 60 110 L 60 25 110 L 60
110 G1 2
110 G12
❏ DIMENSION Screw connection without alarm FA4 □□ RʢNʣ- □□□ 0 - A Flange connection without alarm FA4 □□ A to D - □□□ 0- A Screw connection with alarm FA4 □□ RʢNʣ- □□□ 1 to 4 - A 'MBOHFDPOOFDUJPOXJUIBMBSNFA4□□ A to D - □□□ 1 to 4 - A ❏ MATERIAL -FGUUP3JHIU 3JHIUUP-FGU #PUUPNUP5PQ 5PQUP#PUUPN ❏ INDICATION DISPLAY DIRECTION (FLOW DIRECTION) ❏ WIRING /P 1BSUOBNF .BUFSJBM 5BQFSFEUVCFSUS316 SUS316 SUS316 'MPBU SUS316 SUS316 SUS316 0SJOH NBRˎ NBR ˎ NBR ˎ $POOFDUJOHpUUJOHSCS14Aˎ SCS14Aˎ SCS14Aˎ )PVTJOH ADC 12 ADC 12 ADC 12 'MBOHF SS 400 SUS304 SUS316 1 2 3 5FSNJOBM ˎ8IFOqVJEUFNQFSBUVSFFYDFFET˃$ UIFNBUFSJBMPG0SJOHJT'1. ˎ5IFNBUFSJBMPGDPOOFDUJOHpUUJOHTNBZCFSFQMBDFECZ464 -PBE ❑ Face to face dimension 'MPXSBUF -NJO $POOFDUJPO TJ[F 4DSFX DPOOFDUJPO ʢ-ʣ 'MBOHF DPOOFDUJPO ʢ-ʣ .BYUP 10 mm (3/8") 15 mm (1/2") 20 mm (3/4") NN NN 25 mm (1") NN .BYUP 20 mm (3/4") NN NN 25 mm (1") NN 32 mm (1-1/4") 40 mm (1-1/2") NN 0UIFSUIBO BCPWF $POTVMUVTGPSEFUBJMT ❑ Approx. pressure loss 'MPXSBUF ʢ-NJOʣ 1SFTTVSFMPTT ʢL1Bʣ $POOFDUJPO TJ[F NN (") NN (") NN (") NN (")
FA4000 Series FLOW MONITOR TOKYO KEISO CO., LTD. 3TG-F966-8E ˙MODEL CODE Measuring range 11 0.1 to 1 L/min (Applicable when flow direction is bottom to top) 2 0.2 to 2 L/min 3 0.5 to 5 L/min 4 1 to 10 L/min 5 1.5 to 15 L/min 6 2 to 20 L/min 7 3 to 30 L/min 8 5 to 50 L/min 9 6 to 60 L/min A 7 to 70 L/min B 8 to 80 L/min C 10 to 100 L/min D 12 to 120 L/min E 13 to 130 L/min Z Consult us for other flow ranges Connection size 1 1 0 m m ( 3 / 8 " ) 2 1 5 m m ( 1 / 2 " ) 3 2 0 m m ( 3 / 4 " ) 4 2 5 m m ( 1 " ) 5 3 2 m m ( 1 - 1 / 4 " ) 6 4 0 m m ( 1 - 1 / 2 " ) Process connection R R c N NPT (Female) A J I S 1 0 K F F B J I S 1 0 K R F C A N S I C l a s s 1 5 0 R F D J P I C l a s s 1 5 0 R F Z O t h e r s Material -1 N M a t e r i a l 1 Standard material O ring: NBR-2 N M a t e r i a l 2 -3 N M a t e r i a l 3 -1 F M a t e r i a l 1 Standard material O ring: FPM-2 F M a t e r i a l 2 -3 F M a t e r i a l 3 - ZZ Consult factory for details. Flow direction
1 B o t t o m t o T o p
6 L e f t t o R i g h t
7 R i g h t t o L e f t
8 T o p t o B o t t o m
0 No alarm contact
1 High alarm (High CLOSE)
2 High alarm (High OPEN)
3 Low alarm (Low CLOSE)
4 Low alarm (Low OPEN)
Option ✽ /DEG Non oil treatment /EXn IS relay (Code "n" indicates the number of contacts) n=1 (for 1 contact), n=2 (for 2 contacts), n=3 (for 3 contacts) /RED Relay driver (RD-1000) /MnR Magnet strainer (Rc) n=1: Size 10 mm (3/8") n=2: Size 15 mm (1/2") n=3: Size 20 mm (3/4") n=4: Size 25 mm (1") n=5: Size 32 mm (1-1/4") n=6: Size 40 mm (1-1/2") /MnA Magnet strainer (JIS10K FF) Special (Blank) Not provided /Z Provided Note) Insert "/" between each code when the plural codes are selected. Code example: Flow range 5 to 50 L/min, size 25 mm, Rating JIS 10K FF , Material 2 (O ring : NBR), Flow direction : Left to Right, Lo w alarm CLOSE, Magnet strainer 25 mm, JIS 10K FF , IS relay for 2 contacts, "FA484A-2N63-A/M4A/EX2"
FA4000 Series FLOW MONITOR 4 TOKYO KEISO CO., LTD. TG-F966-8E ❏ Relay driver (RD-1000) ❏ INTRINSICALLY SAFE RELAY (EB3C) The contact output of FA4000 Series is reed switch contact. In case the contact capacity is not sufficient for the operation, use RD-1000 type RELAY DRIVER for capacity increment. (Separate TECDHNICAL GUIDANCE for RD-1000 RELAY DRIVER is available on request. This is not intrinsically safety relay for intrinsically safe circuit. Intrinsically safe relay is to be inserted into the contact loop of FA4000 Series. We can supply IS relay on request. ❏ RD-1000 Outline drawing ❏ WIRING ❏ EB3C Outline drawing Power supply Ambient Temp. Max. supply voltage Max. supply current Max. contact voltage Max. contact current Max. value of switch capacity Insulation resistance Withstand voltage Power consumption
100 V AC or 200 V AC ±10%, 50 / 60Hz
-10 to 50°C
12 V DC (Terminal 7-8 or 9-10)
3 mA DC (Terminal 7-8 or 9-10)
250 V AC, 125 V DC (Terminal 4-5 or 4-6)
5 A (Terminal 4-5 or 4-6)
- 1100 VA AC (Load resistance)
- 120 W DC (Load resistance) 100MΩ at 500 V DC megger 1500 V AC (1 min.) Less than 2 VA ON OFF ON OFF OFF ON Relay action NO (Terminal 4-5) NC (Terminal 4-6) Switch action (Terminal 9-10) COM NO NC SW
200 V AC
100 V AC
4-ϕ6 (M-5 bolt) 90 52 110 Installation hole dimension 6542 (4) 77.5 MIN.ϕ3 MAX.6 MAX.3 MIN.5.4 Terminal screw size : M3
- Applicable crimping terminal EB3C-R01□ EB3C-R02□, R03□ Type of protection Intrinsically safe (Ex ia ᶘC) Rated operation voltage 12 V DC ±10% Rated operation current 10 m A DC ±20% Protection class IP20 (IEC60529) Non-intrinsically circuit (Relay output) Contact configuration 1a contact Rated insulation voltage 250 V AC / 125 V DC Rated turning on 3A electricity current Contact allowable power 750 VA AC / 72 W DC (Resistance load) Rated load 250 V AC, 3A / 24 V DC, 3 A (Resistance load) General specification 100 V to 240 V AC 24 V DC Acceptable variation rate -15 to +10% ±10% Rated frequency 50/60Hz (Allowable range: 47 to 63Hz) — Inrush current 10 A (100 V AC)
20 A (200 V AC) 10A
Insulation resistance 10MΩ or more (500 V DC megger, between the same poles as the dielectric strength) Model code Description EB3C- Output type No. of contact Power supply R R A D □□ □ Model Relay output 1 point use 2 point use 3 points use
100 V to 240 V AC, 50/60Hz
24 V DC
˙OPTIONAL UNIT L L ❏ MAGNET STRAINER /PNJOBM EJBNFUFS 4DSFX 'MBOHF L L .BHOFUTUSBJOFSEJNFOTJPO NN Fluid press. : Max. 1.5MPa Fluid temp : Max. 200°C Filter : 80 mesh Material (Body) : SCS13 or SCS14, Others (Filter) : SUS304 or SUS316 (Packing) : PTFE Connection rating : Rc Connection rating : JIS10K FF Notes on magnet strainer Above figures and dimensions of magnet strainers are those of JIS 10K FF and Rc threads in connection and SCS13 or SCS14 in body material. The dimensions or appearance may differ if connection or material is different from above.
FA4000 Series FLOW MONITOR TOKYO KEISO CO., LTD. 5TG-F966-8E Head Office : Shiba Toho Building, 1 – 7 – 24 Shibakoen, Minato-ku, Tokyo 105 – 8558 Tel : +81-3 – 3431 – 1625 (KEY) ; Fax : +81-3 – 3433 – 4922 * Specification is subject to change without notice.