CO2 Sensor for Railway Applications EE8915
v1.9 / Modification rights reserved | 2 EE8915 EE8915 measures reliably CO 2 concentration in harsh environment and complies with the relevant railway standards. Outstanding Accuracy and Long-Term Stability A multiple point CO 2 and temperature (T) adjustment procedure leads to excellent CO 2 measurement accuracy over the entire T working range -40…+60 °C (-40…+140 °F). The active compensation with on-board sensors leads to best CO 2 measurement accuracy independently of weather conditions, altitude or temperature. The E+E dual wavelength non-dispersive infrared (NDIR) measurement principle compensates automatically for ageing effects and is highly insensitive to pollution. Versatile and Suitable for Demanding Applications EE8915 is available for wall and duct mounting. The innovative design enables the combination of short response time and high protection class. The CO2 measured data is available as voltage and current output signals. Due to the compliance with tough railway standards, the EE8915 stands for excellent performance even under challenging conditions in any process and climate control application. Configurable and Adjustable The free EE-PCS product configuration software and the USB connection enable particularly user-friendly configuration and adjustment. CO2 Sensor for Railway Applications EE8915 wall mount with M12-plug EE8915 duct mount with fix installed cable
v1.9 / Modification rights reserved | 3
Features
▪ Voltage and current output ▪ M12 connector or fix installed cable ▪ User configurable and adjustable ▪ USB configuration interface Enclosure ▪ IP65 protection rating ▪ UL94 V-0 approved material ▪ Easy mounting without opening the device Test report according According to DIN EN 10204-2.2 Measurement performance ▪ E+E dual wavelength NDIR auto calibration ▪ T and p compensation with on-board sensors ▪ CO2 range 0...2 000 / 5 000 / 10 000 ppm ▪ Short response time
v1.9 / Modification rights reserved | 4 Dimensions Values in mm (inch) Wall mount Type:T1 Connection Refer to ordering guide Ø22 (0.87) 98 (3.86) 48 (1.89) 35 (1.38) Ø4.5 (0.18) 26 (1.02) 85 (3.35) Connection Refer to ordering guide 39.8 (1.57) Duct mount Type:T2 Connection Refer to ordering guide 98 (3.86) Ø4.5 (0.18) 85 (3.35) 35 (1.38) 48 (1.89) L Ø12 (0.47) 39.8 (1.57) L in mm (inch) L1 in mm (inch) Connection Refer to ordering guide
v1.9 / Modification rights reserved | 5 Dimensions Values in mm (inch) Duct mount with 90° rotated probe Type:T27 Connection Refer to ordering guide 98 (3.86) Ø4.5 (0.18) 85 (3.35) 35 (1.38) 48 (1.89) 39.8 (1.57) (0.39) 8 (0.31) L L1 Ø12 (0.47) L in mm (inch) L1 in mm (inch) Connection Refer to ordering guide
v1.9 / Modification rights reserved | 6 Technical Data Measurands CO2 Measurement principle Dual wavelength non-dispersive infrared technology (NDIR) Measurement range 0...2 000 / 5 000 / 10 000 ppm Accuracy @ 25 °C (77 °F) and 1 013 mbar (14.7 psi) 0...2 000 ppm 0...5 000 ppm 0...10 000 ppm < ±(50 ppm +2 % of mv) mv = measured value < ±(50 ppm +3 % of mv) < ±(50 ppm +5 % of mv) Temperature dependency, typ., ±(1 + mv / 1 000) ppm/°C mv = measured value Residual pressure dependency 1) related to 1 013 mbar 0.014 % of mv / mbar mv = measured value 0.965 % of mv / psi Response time t 63, typ. Duct mount Wall mount <100 s at 3 m/s (590 ft/min) air speed <160 s Measuring intervall 15 s 1) Pressure dependence of a sensor without pressure correction: 0.14 % mv/mbar. Outputs Analogue CO2 1) 0 - 5 V or 0 - 10 V -1 mA < IL < 1 mA IL = load current and 0 - 20 mA or 4 - 20 mA RL ≤ 500 Ω R L = load resistor 1) Voltage and current output signals are available simultaneously. General Power supply class III USA & Canada: Class 2 supply necessary, max. voltage 30 V DC 10 - 35 V DC
24 V DC nominal voltage Un according to EN 50155
Current consumption , typ. Average @ 24 V DC/AC Peak 10 mA + output current 105 mA for 0.3 s Minimum air speed in the duct 1 m/s (196 ft/min) Electrical connection Connector M12x1 or cable with flying leads, max. 2 m (6.56 ft) 0...95 %RH, non-condensing Enclosure Material Protection rating Polycarbonate, UL94 V-0 approved IP65 / NEMA 4X Electromagnetic compatibility Railway standard: EN 50121-3-2:2016 EN 50121-1:2017 EN 61326-1 EN 61326-2-3 Industrial environment FCC Part15 Class B ICES-003 Class B Conformity Configuration and adjustment Software Interface EE-PCS Product Configuration Software (free download: www.epluse.com/configurator ) USB, micro B
v1.9 / Modification rights reserved | 7 Compliance with Railway Standards ▪ EN 50155:2017 Electronic equipment used on rolling stock ▪ EN 50121-1:2017 Electromagnetic compatibility - general ▪ EN 50121-3-2:2016 Electromagnetic compatibility - rolling stock ▪ EN 61373:2010 Rolling stock equipment - shock and vibration tests ▪ EN 50125-1 Environmental conditions for equipment - rolling stuck on - board equipment ▪ EN 45545-2 Fire protection on railway vehicles ▪ EN 50306 Railway rolling stock cables having special fire performance Technical Data Ordering Guide Feature Description Code Hardware Configuration EE8915- Type Wall mount T1 Duct mount T2 Duct mount with 90° rotated probe T27 CO2 measuring range 0...2 000 ppm HV1 0...5 000 ppm HV2 0...10 000 ppm HV3 Electrical connection M12 plug E4 Cable E8 Probe length 50 mm (1.97“) L50 200 mm (7.87”) L200 Cable length (for cable version E8 only) 0.5 m (1.64 ft) KL50 2 m (6.56 ft) KL200 SW Setup Output Output 1: 0 - 10 V Output 2: 4 - 20 mA GA7 Output 1: 0 - 5 V Output 2: 0 - 20 mA GA11 Order Example EE8915-T1HV2E8KL50GA7 Feature Code Description Type T1 Wall mount CO2 measuring range HV2 0...5 000 ppm Electrical connection E8 Cable Cable length KL50 0.5 m (1.64 ft) Output GA7 Output 1: 0 - 10 V Output 2: 4 - 20 mA
v1.9 / Modification rights reserved | 8 Accessories Description Code Plastic mounting flange Ø12mm (0.47") HA010202 M12 cable connector for self assemply, 5 pin HA010708 Connection cable M12x1 Socket 5 Poles / Free Cable Ends 1.5 m 5 m 10 m HA010819 HA010820 HA010821 Protection cap for M12 socket HA010781 Protection cap for M12 plug HA010782 For further information see datasheet Accessories .
Company Headquarters & Production Site E+E Elektronik Ges.m.b.H. Langwiesen 7
4209 Engerwitzdorf | Austria
info@epluse.com www.epluse.com Subsidiaries E+E Sensor Technology (Shanghai) Co., Ltd. T +86 21 6117 6129 info@epluse.cn E+E Elektronik France SARL T +33 4 74 72 35 82 info.fr@epluse.com E+E Elektronik Deutschland GmbH T +49 6171 69411-0 info.de@epluse.com E+E Elektronik India Private Limited T +91 990 440 5400 info.in@epluse.com E+E Elektronik Italia S.R.L. T +39 02 2707 86 36 info.it@epluse.com E+E Elektronik Korea Ltd. T +82 31 732 6050 info.kr@epluse.com E+E Elektronik Corporation T +1 847 490 0520 info.us@epluse.com Version v1.9 | 08-2023 Modification rights reserved