Humidity and Temperature Sensor for Industrial Applications EE23
v5.4 / Modification rights reserved | 2 EE23 The EE23 is optimized for reliable and cost effective use in industrial applications. In addition to highly accurate measurement of relative humidity (RH) and temperature (T), the sensor also calculates the dew point (Td) and the frost point temperature (Tf). Measurement Performance The EE23 employs high-end E+E humidity sensing elements manufactured in state-of-the-art thin-film technology, which are the prerequisite for outstanding accuracy. Long Term Stability The E+E proprietary coating protects the sensing elements against corrosive and electrically conductive pollution, which leads to outstanding long-term stability even in harsh environment. With the appropriate choice of filter cap, the EE23 tackles even challenging industrial applications. Outputs and Power Supply The measured data is available on two voltage or current outputs as well as on the display. Additional features like alarm (relay) output and integrated supply module 100 - 240 V AC facilitate the use of the EE23 in a wide range of applications. Easy Installation and Service The modular, three parts design of the IP65 / NEMA 4 enclosure, available in polycarbonate or metal, facilitates easy installation, service and replacement. The enclosure consists of the back cover with the terminals for wiring, the pluggable active part with the electronics and the probe, and the front cover. Once installed, the active part of EE23 can be plugged on and off without rewiring. The plastic enclosure is appropriate also for mounting onto DIN rails. Remote Probe and Accessories The remote probe with cable length up to 10 m (32.8 ft) together with a wide choice of accessories such as mounting flanges or brackets, drip water protection or radiation shield allow for easy integration of the EE23 into any measurement task. User Configurable The user can easily perform a two-point humidity and temperature adjustment. The analogue and alarm outputs can be freely configured. Humidity and Temperature Sensor for Industrial Applications Wall mount T1 Duct mount T2 Remote probe T4/T5
v5.4 / Modification rights reserved | 3
Features
▪ Polycarbonate or aluminium die-cast ▪ IP65 / NEMA 4(X) protection rating ▪ Robust and high mechanical stability ▪ Easy mounting and installation Inspection certificate According to DIN EN 10204-3.1 Types and outputs ▪ Types for wall mount, duct mount, remote probe ▪ 2 analogue outputs current / voltage ▪ Optional alarm output Measurement perfomance ▪ High RH / T accuracy ▪ Wide T range from -40 up to +180 °C (- 40...+356 °F) ▪ Calculation of ▪ Dew point temperature (Td) ▪ Frost point temperature (Tf) RH and T sensing head ▪ Very robust ▪ Protected by E+E proprietary coating ▪ Outstanding long term stability ▪ Wide choice of filter caps
v5.4 / Modification rights reserved | 4 The E+E proprietary sensor coating is a protective layer applied to the sensing elements, their leads and soldering points. The coating substantially extends sensor lifetime and ensures optimal measurement performance in corrosive environment (salts, off-shore applications). Additionally, it improves the sensors’ long term stability in dusty, dirty or oily applications by preventing stray impedance caused by deposits on the active sensor surface or on the electrical connections. Protective Sensor Coating Values in mm (inch) Dimensions Enclosure PC (Polycarbonate) Metal (Aluminium die-cast) 135 (5.3) 163 (6.4) 66.5 (2.6) 90 (3.5) 90 (3.5) 32 (1.3) /g13512 (0.5) code “probe length” /g13512 (0.5) 39 (1.54) 50 (2) /g13512 (0.5) (1.3) 15 (0.6) code “probe cable length” code “probe length” 135 (5.3) 163 (6.4) 66.5 (2.6) 90 (3.5) 90 (3.5) 32 (1.3) /g13512 (0.5) code “probe length” /g13512 (0.5) 39 (1.54) 50 (2) /g13512 (0.5) (1.3) 15 (0.6) code “probe cable length” code “probe length” Types T1: Wall mount Probe material: PC (Polycarbonate) T4 / T5: Remote probe Probe material: Stainless steel 135 (5.3) 163 (6.4) 66.5 (2.6) 90 (3.5) 90 (3.5) 32 (1.3) /g13512 (0.5) code “probe length” /g13512 (0.5) 39 (1.54) 50 (2) /g13512 (0.5) (1.3) 15 (0.6) code “probe cable length” code “probe length” 135 (5.3) 163 (6.4) 66.5 (2.6) 90 (3.5) 90 (3.5) 32 (1.3) /g13512 (0.5) code “probe length” /g13512 (0.5) 39 (1.54) 50 (2) /g13512 (0.5) (1.3) 15 (0.6) code “probe cable length” code “probe length” T2: Duct mount Probe material: Stainless steel 135 (5.3) 163 (6.4) 66.5 (2.6) 90 (3.5) 90 (3.5) 32 (1.3) /g13512 (0.5) code “probe length” /g13512 (0.5) 39 (1.54) 50 (2) /g13512 (0.5) (1.3) 15 (0.6) code “probe cable length” code “probe length”
v5.4 / Modification rights reserved | 5 Technical Data Measurands Relative humidity Measuring range 0...100 % RH Accuracy1) ±2.3 %RH ±(1.5 + 1.5 %*mv) %RH mv = measured value Temperature dependency of electronics, typ. ±0.015 %RH / °C Response time t 90 with metal grid filter, @ 20 °C (68 °F) <15 s 1) Including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV... The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). Temperature (T) Working range EE23-T1 EE23-T2 EE23-T4 EE23-T5 Accuracy T [°C] ±∆T [°C] Temperature dependence of electronics, typ. ±0.002°C/°C Calculated Quantities from up to unit EE23-T1 EE23-T2 EE23-T4 EE23-T5 Dew point temperature Td -40 (-40) 60 (140) 80 (176) 100 (212) 100 (212) °C (°F) Frost point temperature Tf -40 (-40) 0 (32) 0 (32) 0 (32) 0 (32) °C (°F) Output Analogue Two freely selectable and scalable outputs 0 - 10 V -1 mA < IL < 1 mA IL = load current 0 - 20 mA / 4 - 20 mA RL < 470 Ω RL = load resistance
v5.4 / Modification rights reserved | 6 Output SPDT-Switch max. 250 V AC/8 A or 28 V DC/8 A Setting range Threshold Hysteresis 10...95 %RH 3...15 %RH Setting accuracy ±3 %RH 1) For types T1, T2, T4 only Alarm Module (optional)1) Technical Data Power supply class III USA & Canada: Class 2 supply necessary, max. voltage 30 V DC 15 - 35 V DC or 15 - 28 V AC 100 - 240 V AC, 50/60 Hz (with option AM3) Current consumption voltage output with DC supply with AC supply ≤ 25 mA (with alarm module ≤ 35 mA) ≤ 45 mArms (with alarm module ≤ 70 mA rms) Current consumption current output with DC supply with AC supply ≤ 55 mA (with alarm module ≤ 65 mA) ≤ 100 mArms (with alarm module ≤ 120 mA rms) Electrical connection Screw terminals max. 1.5 mm 2 (AWG 16) Cable glands M16x1.5, for cable Ø4.5 - 10 mm (0.18 - 0.39") Temperature working range Electronics with display Electromagnetic compatibility EN 61326-1 EN 61326-2-3 Industrial environment FCC Part15 Class A ICES-003 Class A Conformity General
v5.4 / Modification rights reserved | 7 Ordering Guide Features Description Code Hardware Configuration EE23- Type1) Wall mount T1 Duct mount T2 Remote probe up to 120 °C (248 °F) T4 Remote probe up to 180 °C (356 °F) T5 Enclosure PC (Polycarbonate) No code Aluminium die-cast (AlSi9Cu3) HS3 Filter Plastic - metal grid (up to 120 °C / 248 °F) F3 F3 F3 F3 Stainless steel sintered No code No code No code No code PTFE (Polytetrafluoroethylene) F5 F5 F5 F5 Stainless steel grid (up to 180 °C / 356 °F) F9 Probe cable length (incl. probe length) 2 m (6.6 ft) K2 K2 5 m (16.4 ft) K5 K5 10 m (32.8 ft) K10 K10 Probe length 65 mm (2.55”) L65 L65 L65 200 mm (7.84ʺ) No code No code No code 400 mm (15.75”) L400 L400 L400 Electrical connection Standard2) No code 1 plug for power supply and outputs E4 Optional features LC display D1 D1 D1 D24) E+E Sensor-Coating C1 C1 C1 C1 Alarm outputs for RH3) AM2 AM2 AM2 Integrated power supply 100 - 240 V AC, 50/60 Hz3) AM3 AM3 AM3 AM3 Output signal 0 - 10 V GA3 0 - 20 mA GA5 4 - 20 mA GA6 Setup - Analogue outputs 1) Output 1 measurand Relative humidity RH [%] No code Other measurand (xx see measurand code below) MAxx Output 1 scaling low 0 No code Value SALValue Output 1 scaling high 100 No code Value SAHValue Output 2 measurand Temperature T [°C] No code Temperature T [°F] MB2 Other measurand (xx see measurand code below) MBxx Output 2 scaling low Value SBLValue Output 2 scaling high Value SBHValue Display mode Measurand output 1 + 2 alternating DT2 DT2 DT2 Measurand output 1 DT3 DT3 DT3 Measurand output 2 DT4 DT4 DT4 1) For T1, T2 and T4 adjustment changes on the electronics board - see User Manual. For T5 adjustment and configuration changes by E+E PCS Software only - see User Manual. 2) Standard = 2xM16 cable glands, except for AM3 option: 2 plugs for power supply and outputs. 3) With electrical connection standard only (no plug options possible) / combination alarm output and integrated power supply is not possible. 4) Measurand on display can be selected with push buttons. Measurand Code For Output 1 and 2 in the Ordering Guide Measurand Unit Code MAxx / MBxx Relative humidity RH % 10 Temperature T °C Dew Point Td °C Frost point Tf °C
v5.4 / Modification rights reserved | 8 Order Example EE23-T4HS3F3K2D1GA3SBL0SBH50DT2 Feature Code Description Type T4 Remote probe up to 120 °C (248 °F) Enclosure HS3 Aluminium die-cast (AlSi9Cu3) Filter F3 Plastic - metal grid Probe cable length K2 2 m (6.6 ft) Probe length No code 200 mm (7.87ʺ) Electrical connection No code Standard Optional feature D1 LC display Output signal GA3 0 - 10 V Output 1 measurand No code Relative humidity [%] Output 1 scaling low No code 0 Output 1 scaling high No code 100 Output 2 measurand No code Temperature T [°C] Output 2 scaling low SBL0 0 Output 2 scaling high SBH50 50 Display mode DT2 Measurand output 1 + 2 alternating Accessories For further information see datasheet Accessories . Accessories Code Mounting flange HA010201 Bracket for installation onto mounting rails 1) HA010203 Drip water protection HA010503 Radiation shield HA010502 Calibration set (see data sheet „Calibration Kit“) HA0104xx Stainless steel wall mounting clip Ø12 mm (0.5") HA010225 1) For plastic enclosure only
Company Headquarters & Production Site E+E Elektronik Ges.m.b.H. Langwiesen 7
4209 Engerwitzdorf | Austria
info@epluse.com www.epluse.com Subsidiaries E+E Sensor Technology (Shanghai) Co., Ltd. T +86 21 6117 6129 info@epluse.cn E+E Elektronik France SARL T +33 4 74 72 35 82 info.fr@epluse.com E+E Elektronik Deutschland GmbH T +49 6171 69411-0 info.de@epluse.com E+E Elektronik India Private Limited T +91 990 440 5400 info.in@epluse.com E+E Elektronik Italia S.R.L. T +39 02 2707 86 36 info.it@epluse.com E+E Korea Co., Ltd. T +82 31 732 6050 info.kr@epluse.com E+E Elektronik Corporation T +1 847 490 0520 info.us@epluse.com Version v5.4 | 08-2023 Modification rights reserved