EE360 EPLUSE | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 11

Technical content

High-End Moisture in Oil Sensor EE360

v2.6 / Modification rights reserved | 2 EE360 EE360 is dedicated for reliable monitoring of lubrication, hydraulic and insulation oils as well as diesel fuel. In addition to highly accurate measurement of water activity (aw) and temperature (T), EE360 calculates the absolute water content (x) in ppm. Measurement Performance The EE360 employs high-end E+E humidity sensing elements manufactured in state-of-the-art thin film technology, which are the prerequisite for outstanding measurement accuracy. Process Connection The sensing probe can be employed up to 180 °C (356 °F), 20 bar (290 psi) and is available with either ISO or NPT slide fitting, which allows for variable immersion depth. Using the optional ball valve, the probe can be mounted or removed even without process interruption. Enclosure The EE360 features an IP65 / NEMA 4X polycarbonate or stainless steel enclosure which facilitates installation and maintenance. The enclosure can accommodate a 100 - 240 V AC supply unit or various extension modules. Display and Outputs The measured data is available on two analogue outputs, on the RS485 (Modbus RTU) or Ethernet-PoE (Modbus TCP) interface and on the alarm (relay) outputs. The TFT colour display shows up to four measurands simultaneously and offers extensive setup and diagnosis features. The data logging function saves up to 20 000 measured values for each physical quantity. The logged data can be displayed graphically directly on the device or easily downloaded via the USB interface. Configurable and Adjustable The configuration and adjustment of the EE360 can be performed either using the display and the push buttons or with the free E+E PCS10 Product Configuration Software via the USB interface. High-End Moisture in Oil Sensor EE360 with stainless steel enclosure EE360 with polycarbonate enclosure

v2.6 / Modification rights reserved | 3 Enclosure ▪ IP65 / NEMA 4X protection rating ▪ Polycarbonate or stainless steel ▪ Easy mounting and service ▪ Screws secured in cover Outputs ▪ 2 analogue outputs current / voltage ▪ Error indication according NAMUR ▪ Modbus RTU / Modbus TCP ▪ 2 alarm outputs ▪ Configurable via display or software 3.5" TFT colour display ▪ Shows up to 4 measurands simultaneously ▪ Layout and measurands freely selectable ▪ Data logger for 20 000 values per measurand ▪ Logged data shown graphically ▪ Diagnosis functions ▪ Intuitive device setup with push buttons Probe ▪ Pressure tight up to 20 bar (290 psi) ▪ ISO or NPT process connection ▪ Pluggable probe option USB service interface ▪ Download logged data ▪ Perform configuration, adjustment and firmware update ▪ 4 status LEDs

Features

According to DIN EN 10204-3.1 Ball valve set ▪ Probe mounting and removal without process interruption

v2.6 / Modification rights reserved | 4 The moisture in oil can be expressed in absolute or relative terms. Water activity a w is the relative measure for moisture in oil. It represents the ratio between the actual amount of dissolved water and the maximum possible amount of dissolved water in the oil at a certain temperature. Independently of the oil type, the water activity shows how close to saturation is the oil at any moment in time. ▪ aw = 0 indicates completely dry oil ▪ aw = 1 indicates fully saturated oil EE360 measures directly the water activity. Water content x is an absolute measure equal to the amount of water in the oil. The water content is measured in ppm (parts per million) and is independent from the oil temperature. For assessing how far is the oil from saturation, x must be regarded together with T. EE360 calculates x out of the measured aw and T values. The calculation is oil dependent and requires a set of oil specific parameters. Water activity aw / Water content x Dimensions Enclosure Polycarbonate 54 (2.13) 104 (4.09) 5.1 (0.20) 180 (7.09) 166 (6.54) Values in mm (inch) Enclosure Stainless steel 75 (2.95) 5.1 (0.20) 218 (8.58) 200 (7.87) 186 (7.32) 110 (4.33) 73 (2.87) In certain applications, the oil can become corrosive over time, for example due to continuous contamination of lubricating oils by salt water in the maritime environment. In such demanding applications, the E+E proprietary protection of the sensing element leads can significantly extend the service life of the sensor. Sensor Leads Protection

v2.6 / Modification rights reserved | 5 Dimensions Values in mm (inch) Probe code “probe length”1) code “probe cable length”1) (1.3) Ø13 (0.51) Ø12 (0.47) 15 (0.59) 1) Refer to ordering guide Probe Minimum insertion depth 23 (1) Probe Maximum insertion depth Max. depth acc. to table below Probe length [mm (inch)] Max. insertion depth [mm (inch)] 100 (2.5) 64 (3.9) Ball valve set G 1/2" ISO or NPT 155 (6.1) (0.98) Ø33 (1.30) G 1/2" ISO or 1/2" NPT 63 (2.48) G 1/2" ISO or 1/2" NPT (2.36)

v2.6 / Modification rights reserved | 6 Technical Data Measurands Water Activity (a w) / Water Content (x) Measuring range 0...1 aw 0...100 000 ppm; actual range depends on the oil type, for non-mineral transformer oil, specific Accuracy1) ±(0.013 + 0.3%*mv) a w mv = measured value ±0.023 aw Temperature dependency of electronics , typ. ±0.0001 aw/°C (±5.6 * 10 -5 aw/°F) Response time t 90, typ. @ 20 °C (68 °F) in still oil 10 min. 1) Including hysteresis, non-linearity and repeatability, traceable to intern. standards, administrated by NIST, PTB, BEV... The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). Temperature (T) Accuracy1) -40 -20 0 20 40 60 80 100 120 140 160 180 0.1 0.2 0.3 0.4 0.5 0.6∆T [°C] T [°C] 0.55 Temperature dependency of electronics , typ. ±0.005 °C/°C (±0.016 °F/°F) 1) Traceable to international standards, administrated by NIST, PTB, BEV,... The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). Outputs Analogue Two analogue outputs freely selectable and scalable 0 - 1 / 5 / 10 V -1 mA < IL < 1 mA IL = load current 4 - 20 mA 3-wire RL < 500 Ohm RL= load resistance 0 - 20 mA 3-wire RL < 500 Ohm Digital Digital interface RS485 (EE360 = 1 unit load) Protocol Option J3 Factory settings Supported Baud rates Measured data types Modbus RTU 9 600 Baud, parity even, 1 stop bit, Modbus address 231 300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600 and 76800 FLOAT32 and INT16 Protocol Option J4 Ethernet-PoE Modbus TCP

v2.6 / Modification rights reserved | 7 Technical Data Power supply class III USA & Canada: Class 2 supply necessary, max. voltage 30 V DC 8 - 35 V DC 12 - 30 V AC 100 - 240 V AC, 50/60 Hz with Option AM3 1) Current consumption , typ. @ 24 V DC/AC for 2 voltage outputs for 2 current outputs additional for display additional for Ethernet 15 mA and 40 mA rms 35 mA and 100 mArms 50 mA and 150 mA rms 30 mA and 90 mA rms Electrical connection Screw terminals max. 1.5 mm 2 (AWG 16) Cable glands for polycarbonate enclosure for metal enclosure M16x1.5, for cable Ø3 - 7 mm (0.12 - 0.28") M16x1.5, for cable Ø4.5 - 10 mm (0.18 - 0.39") Temperature range Operation Storage Material Probe Enclosure Stainless steel 1.4404 (AISI 316L) PC (Polycarbonate), UL94-V0 approved Stainless steel 1.4404 (AISI 316L) Protection rating IP65 / NEMA 4X Electromagnetic compatibility EN 61326-1 EN 61326-2-3 Industrial environment FCC Part15 Class A ICES-003 Class A Conformity Two alarm outputs, with option AM2 1) Changeover contact

250 V AC / 6 A 28 V DC / 6 A

Configuration software E+E PCS10 Product Configuration Software Free download from www.epluse.com/pcs10 1) Degree of pollution 2, overvoltage category II, altitude up to 3000 m (9843 ft). General

v2.6 / Modification rights reserved | 8 Ordering Guide Feature Description Code Hardware Configuration EE360- Enclosure material PC (Polycarbonate) No code Stainless steel HS2 Filter Stainless steel, for flow <1 m/s (<3.3 ft/s) No code Stainless steel, for flow >1 m/s (>3.3 ft/s) F18 Probe Cable Length 2 m (6.6 ft) No code 5 m (16.4 ft) K5 10 m (32.8 ft) K10 Probe length 100 mm (3.94") L100 200 mm (7.87") No code 400 mm (15.75") L400 600 mm (23.62") L600 800 mm (31.50") L800 1000 mm (39.37") L1000 Process connection G 1/2" ISO - slide fitting, Ø13 mm (0.51") No code 1/2" NPT - slide fitting, Ø13 mm (0.51") PA25 Electrical connection Cable glands No code 1 x plug for power supply and outputs E4 1 x cable gland and 1 plug for Modbus RTU (requires option J3) E5 2 x plugs for power supply + outputs and Modbus RTU (requires option J3) E6 3 x plugs for power supply + outputs and Modbus RTU (requires option J3) 1) E12 Optional features 3.5" TFT display with integrated data logger D2 RS485 module - Modbus RTU J3 Ethernet module - Modbus TCP 1)2) J4 Pluggable probe 1) PC4 Sensing element protection (Senor leads protection) C2 Alarm outputs (Relay module with cable glands) 2) AM2 Integrated power supply (100 - 240 V AC, 50/60 Hz), with connector 2)3) AM3 Setup (Analogue) Outputs Output 1 measurand Water activity a w [ ] No code Water content x [ppm] MA70 Temperature T [°C] MA1 Temperature T [°F] MA2 Output signal 1 4) 0 - 1 V GA1 0 - 5 V GA2 0 - 10 V GA3 0 - 20 mA GA5 4 - 20 mA GA6 Output 1 scaling low 0 No code Value SALValue Output 1 scaling high 1 No code Value SAHValue Output 2 measurand Temperature T [°C] No code Temperature T [°F] MB2 Water activity a w [ ] MB67 Water content x [ppm] MB70 Output signal 2 4) 0 - 1 V GB1 0 - 5 V GB2 0 - 10 V GB3 0 - 20 mA GB5 4 - 20 mA GB6 Output 2 scaling low Value SBLValue Output 2 scaling high Value SBHValue Oil parameterization for water content calculation Mineral transformer oil No code Customer specific oil PPMxxx5) 1) Only with polycarbonate enclosure. 2) No combination of alarm output (AM2), Ethernet module (J4) and integrated power supply (AM3) is possible. 3) Integrated power supply includes 2 plugs for power supply and outputs, other plug options are not possible. 4) Both analogue outputs shall be either voltage or current. 5) Procedure for customer specific oil (see table below).

v2.6 / Modification rights reserved | 9 Option Description Code Oil number is known Replace the xxx by the corresponding number Obtaining new oil parameters via oil analysis Contact and provide E+E HQ the datasheet of the oil before sending us 2 litres of oil. After determination of the oil specific parameters, the corresponding oil number is available, insert this in place of the xxx. Oil-ppmcal Obtaining new oil parameters via saturation curve Contact and provide E+E HQ the datasheet of the oil together with the saturation curve. After calculation of the oil specific parameters, the corresponding oil number is available, insert this in place of the xxx. Oil-calc 5) Procedure for customer specific oil Order Example EE360-D2J3GA3GB3SBL-40SBH180 Feature Code Description Enclosure material No code PC (Polycarbonate) Filter No code Stainless steel, for flow <1 m/s (<3.3 ft/s) Probe cable length No code 2 m (6.6 ft) Probe length No code 200 mm (7.87") Process connection No code G 1/2" ISO - slide fitting, Ø13 mm (0.51") Electrical connection No code Cable glands Optional features D2 3.5" TFT display with integrated data logger RS485 module - Modbus RTU Output 1 measurand No code Water activity a w [ ] Output 1 signal GA3 0 - 10 V Output 1 scaling low No code 0 Output 1 scaling high No code 1 Output 2 measurand No code Temperature T [°C] Output 2 signal GB3 0 - 10 V Output 2 scaling low SBL-40 -40 Output 2 scaling high SBH180 180 Ordering Guide Oil-ppmcal Contact and provide E+E HQ the datasheet of the oil before sending us 2 litres of oil.

v2.6 / Modification rights reserved | 10 Accessories Description Code PCS10 Product Configuration Software (Free download: www.epluse.com/pcs10 ) PCS10 USB-C to USB-A configuration cable HA010327 Bracket for installation onto mounting rails 1) HA010203 Determination of oil specific parameters ppm-cal Humidity calibration kit see data sheet Humidity calibration kit Ball valve set G 1/2" ISO HA050101 Ball valve set 1/2" NPT HA050104 1) For polycarbonate enclosure only. Two pieces are necessary for each EE360. For further information see datasheet Accessories .

Company Headquarters & Production Site E+E Elektronik Ges.m.b.H. Langwiesen 7

4209 Engerwitzdorf | Austria

info@epluse.com www.epluse.com Subsidiaries E+E Sensor Technology (Shanghai) Co., Ltd. T +86 21 6117 6129 info@epluse.cn E+E Elektronik France SARL T +33 4 74 72 35 82 info.fr@epluse.com E+E Elektronik Deutschland GmbH T +49 6171 69411-0 info.de@epluse.com E+E Elektronik India Private Limited T +91 990 440 5400 info.in@epluse.com E+E Elektronik Italia S.R.L. T +39 02 2707 86 36 info.it@epluse.com E+E Elektronik Korea Ltd. T +82 31 732 6050 info.kr@epluse.com E+E Elektronik Corporation T +1 847 490 0520 info.us@epluse.com Version v2.6 | 09-2023 Modification rights reserved