v2.12 / Modification rights reserved EE300Ex-M1www.epluse.com The EE300Ex intrinsically safe sensor reliably measures relative hu- midity (RH) and temperature (T) in explosion hazard areas. It com- plies with the classifications for Europe (ATEX), International (IECEx), China (NEPSI), USA / Canada (FM) and Korea (KCs) for flammable gas and dust applications. The EE300Ex it is also certified for gas applications according Japan (TIIS) certifications. The entire device can be placed in the explosion endangered area. The remote sensing probe allows for classification up to T6. Measurement performance The well proven E+E humidity sensors and competence in calibration allow for highly accurate and long term stable measurement over the full range 0…100 % RH and -40…180 °C (-40...356 °F), with pressure rating up to 20 bar (300 psi). Besides the RH and T measurement, the EE300Ex calculates all humidity related parameters such as dew point temperature (Td), frost point temperature (Tf), absolute humidity (dv) or mixing ratio (r). Moisture in oil measurement The EE300Ex with ATEX, IECEx, NEPSI and KCs approval is suitable also for measuring water content (X) in ppm and water activity (aw) in isolation, lubrication and hydraulic oils. Typical applications include oil purifiers and online monitoring of lubrication and hydraulic oils. Supply and outputs The device can be powered by any intrinsically safe supply unit or via Zener barriers. The measured or calculated data is available on two 4…20 mA, 2-wire outputs and on the LCD display. Robust, functional design The stainless steel enclosure and sensing probe are suitable for harsh environment in challenging industrial applications. The EE300Ex design facilitates the installation as well as the replacement of the measuring section (electronics and probe) without time consuming wiring. Easy Configuration and Adjustment The setup of the analogue outputs as well as the adjustment of the RH and T reading can be easily performed with the optional EE-PCA Product Configuration Adapter and the free EE-PCS Product Configuration Software. Humidity and Temperature Sensor for Intrinsically Safe ApplicationsEE300Ex-M1 EE300Ex-M1 Typical Applications Features Chemical process control Pharmaceutical applications Explosive / hazardous storage rooms Flour mills Oil purifiers Gas and dust in zone 0 / 20 and Div. 1 Stainless steel enclosure and probe Best accuracy up to 180 °C (356 °F) Pressure tight up to 20 bar (300 psi) Inspection certificate according to DIN EN 10204-3.1

v2.12 / Modification rights reservedEE300Ex-M1 www.epluse.com Types Dimensions in mm (inch) Types: T7 Remote probe 20 bar (300 psi) Type: T10 Remote probe 20 bar (300 psi) with sliding fitting for assembly / disassembly under pressure Types: T1 / T7 / T10 Enclosure Type Pressure range Working range Probe Ø mm (inch) T10 Remote probe with sliding fitting for assembly / disassembly under pressure, pressure tight 179 (7) 60 (2.4)151 (5.9) adjustable min. 23 (0.9) / G1/2" ISO or NPT G1/2" ISO or 1/2" NPT 98 (3.9) 53 (2.1) Ø13 (0.51) Ø12 (0.47) L L L code „probe length “ code „probe length “ code „cable length“ code „cable length“ 15 (0.6) 15 (0.6) L - length of filter mm (inch) Stainless steel sintered filter 33 (1.3) PTFE, H2O2 filter 33 (1.3) Stainless steel grid filter 39 (1.5) Oil filter 32 (1.26) Ø12 (0.47)Type: T1 Wall mount Ø 12 (0.47) Protective sensor coating The E+E proprietary sensor coating is a hygroscopic layer applied to the sensing elements, their leads and soldering points. The coating substantially extends sensor life-time and ensures optimal measurement performance in corrosive environments (salts, off-shore applications). Additionally, it improves the long term stability of E+E sensors in dusty, dirty or oily applications by preventing stray impedance caused by deposits on the active sensor surface or on the electrical connections.

v2.12 / Modification rights reserved EE300Ex-M1www.epluse.com Measurands Relative humidity Measuring range 0...100 % RH Accuracy1) Temperature dependence electronics, typ. 0.03 % RH/°C Response time t90 < 30 s with stainless steel filter at 20 °C (68 °F) Temperature Accuracy ∆T [°C] T [°C] -40 0-10-20-30 10 50403020 60 100908070 140130120110 150 170160 180 ∆T [°C] T [°C] ∆T [°C] T [°C] Temperature dependence of electronics, typ. 0.005 °C/°C Calculated parameters from up to Units Wall mount Remote probe Dew point temperature Td -40 (-40) 60 (140) 100 (212) °C (°F) Frost point temperature Tf -40 (-40) 60 (140) 100 (212) °C (°F) Wet bulb temperature Tw 0 (32) 60 (140) 100 (212) °C (°F) Water vapour pressure e 0 (0) 200 (3) 1 100 (15) mbar (psi) Mixing ratio r 0 (0) 425 (2 900) 999 (9 999) g/kg (gr/lb) Absolute humidity dv 0 (0) 150 (60) 700 (300) g/m³ (gr/ft³) Specific enthalpy h 0 (0) 400 (150 000) 2 800 (999 999) kJ/kg (Btu/lb) Water activity aw 0 - 1 1 Water content X 0 - 100 000 ppm Outputs Freely selectable and scalable outputs 2 x 4-20 mA (2-wire) galvanically isolated RL= (Vcc-9V)/20mA Output 1 must be connected! General Supply voltage Vcc min = (9+RL*0.02) V DC Vcc max = 28 V DC RL = load resistor Current consumption Max. 20 mA per channel Protection class of enclosure IP65 / NEMA 4 Cable gland M16 for cable diameter 5 - 10 mm (0.2" - 0.4") M20 for cable diameter 10 - 14 mm (0.4" - 0.6") Electrical connection Screw terminals max. 1.5 mm² (AWG 16) Working temperature range Probe according measuring range Electromagnetic compatibility EN 61326-1 EN 61326-2-3 ICES-003 ClassB Industrial Environment FCC Part15 ClassB Material Enclosure stainless steel 1.4404 Probe cable PTFE Probe (without filter) stainless steel 1.4404 1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement). Technical Data EE300Ex

v2.12 / Modification rights reservedEE300Ex-M1 www.epluse.com Ex - Classifications Europe (ATEX with order code “EX1”) Certificate: TPS 13 ATEX 38892 003 X by TÜV SÜD Product Service GmbH Safety factors: Ui = 28V; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li ≈ 0mH Ex-Designation: Transmitter without display II 1 G Ex ia IIC T4 Ga / II 1 D Ex ia IIIC T200 80°C Da Transmitter with display II 2 G Ex ia IIC T4 Gb / II 1 G Ex ia IIB T4 Ga International (IECEx with order code “EX2”) Certificate: IECEx FMG 14.0017 X by FM Approvals Safety factors: 6.4 Vdc ≤ Ui ≤ 28Vdc; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH Ex-Designation: Transmitter without display Ex ia IIC T4 Ta = -40°C to 60°C Ga / Ex ia IIIC T131°C Da Transmitter with display Ex ia IIC T4 Ta = -40°C to 60°C Gb / Ex ia IIB T4 Ta = -40°C to 60°C Ga Remote probe Ex ia IIC T6-T1 Ta = -70°C to 200°C Ga / Ex ia IIIC T80°C Da China (NEPSI with order code “EX4”) Certificate: Cert NO. GYJ16.1417X by NEPSI Safety factors: Ui = 28Vdc; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH Ex-Designation: Transmitter without display Ex ia IIC T4 Ga, Ex iaD 20 T131 Transmitter with display Ex ia IIC T4 Gb, Ex ia IIB T4 Ga Remote probe Ex ia IIC T1~T6 Ga, Ex iaD 20 T80 Korea (KCs with order code “EX5”) Certificate gas: Remote probe: 20-AV4BO-0253X Transmitter without display: 20-AV4BO-0254X Transmitter with display: 20-AV4B0-0257X (EPL Ga - Zone 0) 20-AV4BO-0258X (EPL Gb - Zone 1) Certificate dust: Remote probe: 20-AV4BO-0256X Transmitter without display: 20-AV4BO-0255X Safety factors: 6.4 V DC ≤ Ui ≤ 28 V DC; Ii = 100 mA; Pi = 700 mW; Ci = 2.2 nF; Li = 0 mH Ex-Designation: Transmitter (without display): Ex ia IIC T4 -40°C ≤ Tamb ≤ +60°C Ex iaD 20 IP6X T131°C -40°C ≤ Tamb ≤ +60°C Transmitter (with display): Ex ia IIC T4 -40°C ≤ Tamb ≤ +60°C (up to Zone 1) Ex ia IIB T4 -40°C ≤ Tamb ≤ +60°C (up to Zone 0) Remote probe: Ex ia IIC T6-T1 / Ex iaD 20 IP6X T80°C -40°C ≤ Tamb ≤ +60°C Japan (TIIS with order code “EX6”) Certificate: Nr. TC22061 by TIIS Safety factors: Ui = 28Vdc; Ii = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH Ta = -40°C to 60°C Ex-Designation, only for gas: Ex ia IIC T4 Gb

v2.12 / Modification rights reserved EE300Ex-M1www.epluse.com USA (FM with order code “EX3”) Certificate: No. FM17US0302X by FM Approvals Safety factors: 6.4 Vdc ≤ Vmax (or Ui) ≤ 28Vdc; Imax (or Ii) = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH Ex-Designation: Equipment Group I: EE300Ex without display Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C; Entity – M1_139080; IP65 Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C Class I, Zone 0, AEx ia IIC T4 Ta = -40°C to +60°C Ga; Entity – M1_139080; IP65 Zone 20, AEx ia IIIC T131°C Ta = -40°C to +60°C Da; Entity – M1_139080; IP65 Remote Probe: Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65 Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1 Class I, Zone 0, AEx ia IIC T6…T1 Ga; Entity – M1_139080; IP65 Zone 20, AEx ia IIIC T80°C Da; Entity – M1_139080; IP65 Equipment Group II: EE300Ex with display Class I, Division 1, Groups C, and D; T4 Ta = -40°C to +60°C; Entity – M1_139080 Class I, Division 2, Groups A, B, C and D; T4 Ta = -40°C to +60°C; Entity – M1_139080 Class I, Zone 0, AEx ia IIB T4 Ta = -40°C to +60°C Ga; Entity – M1_139080 Class I, Zone 1, AEx ia IIC T4°C Ta = -40°C to +60°C Gb; Entity – M1_139080 Remote Probe: Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65 Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1 Class I, Zone 0, AEx ia IIC T6…T1 Ga; Entity – M1_139080; IP65 Zone 20, AEx ia IIIC T80°C Da; Entity – M1_139080; IP65 CANADA (FM with order code “EX9”) Certificate: No. FM17CA0154X by FM Approvals Safety factors: 6.4 Vdc ≤ Vmax (or Ui) ≤ 28Vdc; Imax (or Ii) = 100mA; Pi = 700mW; Ci = 2.2nF; Li = 0mH Ex-Designation: Equipment Group I: EE300Ex without display Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C; Entity – M1_139080; IP65 Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T4 Ta = -40°C to +60°C Zone 0, Ex ia IIC T4 Ta = -40°C to +60°C Ga; Entity – M1_139080; IP65 Zone 20, Ex ia IIIC T131°C Ta = -40°C to +60°C Da; Entity – M1_139080; IP65 Remote Probe: Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65 Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1 Zone 0, Ex ia IIC T6…T1 Ga; Entity – M1_139080; IP65 Zone 20, Ex ia IIIC T80°C Da; Entity – M1_139080; IP65 Equipment Group II: EE300Ex with display Class I, Division 1, Groups C, and D; T4 Ta = -40°C to +60°C; Entity – M1_139080 Class I, Division 2, Groups A, B, C and D; T4 Ta = -40°C to +60°C; Entity – M1_139080 Zone 0, Ex ia IIB T4 Ta = -40°C to +60°C Ga; Entity – M1_139080 Zone 1, Ex ia IIB T4 Ta = -40°C to +60°C Gb; Entity – M1_139080 Remote Probe: Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T6…T1; Entity – M1_139080; IP65 Class I, II, III, Division 2, Groups A, B, C, D, E, F and G; T6…T1 Zone 0, Ex ia IIC T6…T1 Ga; Entity – M1_139080; IP65 Zone 20, Ex ia IIIC T80°C Da; Entity – M1_139080; IP65 The USA and Canada approvals are valid for air and gas measurement only.

v2.12 / Modification rights reservedEE300Ex-M1 www.epluse.com 1) No display possible for environments with combustible dust, fibers and flyings and in gases with EPL Ga IIC (EX1/EX2) / Gas Groups A, B for Division 1 (EX3/EX9) / Zone 0 IIC (EX5). 2) Allowed only in combination with PA0. 3) May not be used in EPL Ga IIC (EX1/EX2) / Gas Groups A, B for Division 1 (EX3/ EX9) / Zone 0 IIC (EX5). For TIIS (Japan) approval not allowed in models T1, T7 and T10. EE300Ex-M1A6HS2 Hardware Configuration Type Wall mount T1 Remote probe with cut-in fitting, pressure tight, 20 bar (300 psi) T7 Remote probe with sliding fitting, pressure tight, 20 bar (300 psi) T10 Display1) Without display D0 With display D1 Electrical Connection 2 x M16 cable gland E2 1/2" NPT conduit E13 2 x M20 cable gland E15 Probe Cable Length Wall mount K0 1 m (3.3 ft) K1 K1 2 m (6.6 ft) K2 K2 5 m (16.4 ft) K5 K5 10 m (32.8 ft) K10 K10 Probe Length Wall mount, 50 mm (1.97ʺ) L50 65 mm (2.56ʺ)2) L65 100 mm (3.95") L100 200 mm (7.9ʺ) L200 L200 400 mm (15.8") L400 L400 Process Connection (Zone Feedthrough) Without probe fitting PA0 PA0 G1/2" ISO - cut-in fitting, Ø 12 mm (0.47") PA20 1/2" weld cut-in fitting, Ø 12 mm (0.47") PA21 1/2" NPT - cut-in fitting, Ø 12 mm (0.47") PA22 G1/2" ISO - sliding fitting, Ø 13 mm (0.51") PA23 1/2" NPT - sliding fitting, Ø 13 mm (0.51") PA25 Filter Stainless steel sintered F4 F4 F4 PTFE3) F5 F5 F5 Stainless steel grid, stainless steel body, up to 180 °C F9 F9 F9 Catalytic fo H2O2 sterilisation3) F12 F12 F12 Stainless steel with boreholes Ø 3 mm (0.12") F13 F13 Sensing Element Protection Without coating C0 C0 With coating4) C1 C1 C1 Ex Approval ATEX (Europe) EX1 IECEx (International) EX2 FM (USA) EX3 NEPSI (China) EX4 KCs (Korea) EX5 TIIS (Japan)5) EX6 FM (Canada) EX9 4) Not appropriate for moisture in oil measurement, obligatory for all other applications, free of charge. 5) Only gas Ex up to EPL Gb (Zone 1). 6) Assign the most relevant measurand to output 1. 7) For TIIS (Japan) approval, models T1, T7, and T10 have a maximum temperature Setup - analogue outputs Output 16) Measurand (xx see measurand code below)7) MAxx Scaling 1 low Value SALvalue Scaling 1 high Value SAHvalue Output 2 Measurand (xx see measurand code below)7) MBxx Scaling 2 low Value SBLvalue Scaling 2 high Value SBHvalue 1) Not allowed for FM (USA / Canada) and TIIS (Japan) approval. Measurand Code for output 1 and 2 in the ordering guide MAxx / MBxx Relative humidity % 10 Temperature °C 1 °F 2 Dew point Td °C 52 °F 53 Frost point Tf °C 65 °F 66 Mixing ratio r g/kg 60 gr/lb 61 Absolute humidity dv g/m³ 56 gr/ft³ 57 MAxx / MBxx Wet bulb temperature Tw °C 54 °F 55 Water vapour partial pressure e mbar 50 psi 51 Specific enthalpy h kJ/kg 62 BTU/lb 64 Water activity1) aw 67 Water content X in mineral transformer oil1) ppm 70 Water content X in customer specific oil1) ppm 70PPMxxx

v2.12 / Modification rights reserved EE300Ex-M1www.epluse.com Order Example EE300Ex-M1A6HS2T7D1E2K10L200PA20F4C1EX1/ MA1SAL-40SAH180MB10SBL0SBH100 Type: Remote probe up to 20 bar (300 psi) Display: With display Electrical Connection: 2 x M16 cable gland Probe Cable: 10 m (32.8 ft) Probe Length: 200 mm (7.9") Process connection (Zone Feedthrough): G1/2" ISO - cut-in fitting, Ø 12 mm (0.47") Filter: Stainless steel sintered Sensing Element Protection: With coating Ex Approval: ATEX (Europe) Output 1: Temperature [°C] Scaling Output 1: -40...180 °C Output 2: Relative humidity [% RH] Scaling Output 2: 0...100 % RH EE300Ex-M1A6HS2T1D0E2K0L50PA0F9C1EX3/ MA2SAL-40SAH140MB53SBL-40SBH140 Type: Wall mount Display: Without display Electrical Connection: 2 x M16 cable gland Probe Cable: Wall mount Probe Length: Wall mount, 50 mm (1.97ʺ) Process connection: (Zone Feedthrough): Without probe fitting Filter: Stainless steel grid Sensing Element Protection: With coating Ex Approval: FM (USA) Output 1: Temperature [°F] Scaling Output 1: -40...140 °F Output 2: Dew point temperature [°F] Scaling Output 2: -40...140 °F Example 1: Example 2: Accessories Blank cover for enclosure base HA011401 Safety barrier, 1-channel, STAHL 9002/13-280-093-001 HA011410 Intrinsically safe supply unit, 1-channel, STAHL 9160/13-11-11 HA011405 Intrinsically safe supply unit, 2-channel, STAHL 9160/23-11-11 HA011406 Sealing plug for unused M16 cable glands HA011402 Sealing plug for unused M20 cable glands HA011404 Ball valve with 1/2 ISO female thread, ATEX certified HA011403 Product Configuration Software EE-PCS (free download: www.epluse.com/configurator) Adapter Kit for configuration and adjustment (must be ordered together, see datasheet EE-PCA): Pos. 1: Product Configuration Adapter EE-PCA Pos. 2: Connection cable HA011068