AC/DC Medical Power Supply TPP 65 Series, 65 Watt
- Enclosed power supply with screw terminal connection
- Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
- Low leakage current <75 µA rated for BF applications
- Risk management process according to ISO 14971 incl. risk management file
- Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
- EMC compliance to IEC 60601-1-2 ed. 4
- Protection class l and ll
- Operating up to 5000 m altitude
- Ready to meet ErP directive, <0.15 W no load power consumption
- 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 65 Series of 65 Watt AC/DC power supplies feature a reinforced dou- ble I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The leakage current is below 75 µA what makes the units suitable for BF (body floating) applications. The excellent efficiency of up to 92% allows a high power density for the stan- dard 2.38” x 3.53” packaging format. The full load operating temperature range is –40°C to +60°C while it goes up to 85°C with 50% load derating (for single output models). The EMC characteristic is dedicated for applications in industrial and domestic fields. High reliability is provided by use of industrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Output 1 Output 2 Output 3 Efficiency Power Vnom Imax Vnom Imax Vnom Imax typ. TPP 65-105 50 W 5 VDC 10'000 mA 90 % TPP 65-112 65 W
12 VDC 5'420 mA 93 %
TPP 65-115 15 VDC 4'340 mA 94 % TPP 65-124 24 VDC 2'710 mA 94 % TPP 65-221 +12 VDC 5'420 mA +5 VDC 8'000 mA 90 % TPP 65-231 +15 VDC 4'340 mA +5 VDC 8'000 mA 91 % TPP 65-251 +24 VDC 2'710 mA +5 VDC 8'000 mA 89 % TPP 65-321M2 +12 VDC 5'420 mA +5 VDC 8'000 mA –12 VDC 600 mA 89 % TPP 65-331M3 +15 VDC 4'340 mA +5 VDC 8'000 mA –15 VDC 600 mA 90 % TPP 65-3512 +24 VDC 2'710 mA +5 VDC 8'000 mA +12 VDC 600 mA 89 % Note - Total Power must not exceed 65 W. - Other output models are available on request. www.tracopower.com November 30, 2022 Page 1 / 5
Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 950 mA max. - Full Load & Vin = 115 VAC 1'650 mA max. Power Consumption - No load & Vin = 230 VAC 150 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 150 mW max. Input Inrush Current - At 230 VAC 60 A max. - At 115 VAC 35 A max. Input Protection T 3.15 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±10% (only Output 1) (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. (Output 1) ±2% max. (Output 2 and 3) Regulation - Input Variation (Vmin - Vmax) single output models: 0.2% max. dual output models: 0.2% max. triple output models: 0.2% max. - Load Variation (0 - 100%) single output models: 0.7% max. (5 VDC model) 0.5% max. (other output models) dual output models: 0.5% max. (Output 1) 1.5% max. (Output 2) triple output models: 0.5% max. (Output 1) 1.5% max. (Output 2) 0.7% max. (Output 3) - Cross Regulation (25% / 100% asym. load) dual output models: 1.5% max. triple output models: 1.5% max. Ripple and Noise (20 MHz Bandwidth) - single output 5 VDC model: 75 mVp-p typ. (w/ 10 µF X7R) 12 VDC model: 75 mVp-p typ. (w/ 10 µF X7R) 15 VDC model: 75 mVp-p typ. (w/ 10 µF X7R) 24 VDC model: 75 mVp-p typ. (w/ 1 µF X7R) - dual output 12 / 5 VDC model: 120 / 100 mVp-p typ. (w/ 10 µF X7R) 15 / 5 VDC model: 150 / 100 mVp-p typ. (w/ 10 µF X7R) 24 / 5 VDC model: 240 / 100 mVp-p typ. (w/ 10 µF X7R) - triple output 12 / 5 / -12 VDC model: 120 / 100 / 120 mVp-p typ. (w/ 10 µF X7R) 15 / 5 / -15 VDC model: 150 / 100 / 150 mVp-p typ. (w/ 10 µF X7R) 24 / 5 / 12 VDC model: 240 / 100 / 120 mVp-p typ. (w/ 10 µF X7R) Capacitive Load - single output 5 VDC model: 20'000 µF max. 12 VDC model: 4'520 µF max. 15 VDC model: 2'900 µF max. 24 VDC model: 1'130 µF max. - dual output 12 / 5 VDC model: 2'500 / 3'000 µF max. 15 / 5 VDC model: 1'200 / 3'000 µF max. 24 / 5 VDC model: 625 / 3'000 µF max. - triple output 12 / 5 / -12 VDC model: 2'500 / 3'000 / 500 µF max. 15 / 5 / -15 VDC model: 1'200 / 3'000 / 500 µF max. 24 / 5 / 12 VDC model: 625 / 3'000 / 500 µF max. Output Specifications TPP 65 Series, 65 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 2 / 5
(0.5 W for Vout1 and Vout2 if Vout3 = Full Load) Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 115 VAC 16 ms min. Start-up Time - At 230 VAC 1'000 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 120 - 160% of Iout max. 145% typ. of Iout max. (Pout 1 + Pout 2) Overvoltage Protection 125 - 140% of Vout nom. (only Output 1) Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 600 µs typ. (50% to 75% Load Step) (Only Output 1) Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp65 Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 18 class B (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B
115 VAC / 60 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B EMC Specifications TPP 65 Series, 65 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 3 / 5
Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C Power Derating - High Temperature Depending on model - Low Input Voltage 4 %/V below 90 VAC See application note: www.tracopower.com/overview/tpp65 Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 50 - 140 kHz (PWM) (Output 1) 750 kHz typ. (PWM) (Output 2) 510 kHz typ. (PWM) (Output 3) Insulation System Reinforced Insulation Working Voltage (rated) 258 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 2'500 VAC - Output to Case or PE, 60 s 2'500 VAC Creepage - Input to Output 8 mm min. Clearance - Input to Output 8 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 75 µA max. Reliability - Calculated MTBF 1'500'000 h (for single output models) 1'000'000 h (for multi output models) (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 5 g, 3 axis, sine sweep, 5-500 Hz, 1 oct/min - Mechanical Shock IEC 60068-2-27 50 g, 3 axis, half sine, 11 ms Housing Material Aluminum Housing Type Metal Case Mounting Type Chassis Mount Connection Type Screw Terminal Weight - single output 172 g - dual output 221 g - triple output 221 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) General Specifications Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp65 TPP 65 Series, 65 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 4 / 5
TPP 65 Series, 65 Watt Max. corner screw penetration: 2.3 (0.09) Max. center screw penetration: 2.0 (0.08) All dimensions in mm (inch) Tolerance: X.X ±0.5 (X.XX ±0.02) X.XX ±0.25 (X.XXX ±0.010) The screw locked torque: max. 5.0 kgfcm / 0.49 Nm Terminal screw locked torque: max. 4.0 kgfcm / 0.39 Nm 60.5 (2.38) 44.45 (1.75) 8.0 (0.31) 33.3 (1.31) 1.2 (0.047) 10.0 (0.39) CON2 Voltage Adj. 4-M3x0.5 CON1 3 84.24 (3.317) 89.7 (3.53) 40.8 (1.606) Screw Terminal Input (CON1) Output (CON2) Pin Function Pin* Function
1 Line 1,2 –Vout
3 Neutral 3,4 +Vout
*Terminal rated for 10 A max. (at higher current connection has to be split) CON1: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG CON2: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG Max. corner screw penetration: 2.3 (0.09) Max. center screw penetration: 2.0 (0.08) All dimensions in mm (inch) Tolerance: X.X ±0.5 (X.XX ±0.02) X.XX ±0.25 (X.XXX ±0.010) The screw locked torque: max. 5.0 kgfcm / 0.49 Nm Terminal screw locked torque: max. 4.0 kgfcm / 0.39 Nm 60.5 (2.38) 44.45 (1.75) 8.0 (0.31) 34.5 (1.36) 1.2 (0.047) 10.0 (0.39) CON2 Voltage Adj. CON1 3 96.94 (3.817) 102.4 (4.03) 40.8 (1.606) 4-M3x0.5 Screw Terminal Input (CON1) Output (CON2) Pin Function Pin* Dual Triple
1 Line 1 NC Vout 3
3 Neutral 2,3 COM COM
4,5 Vout 2 Vout 2
6 Vout 1 Vout 1
*Terminal rated for 10 A max. (at higher current connection has to be split) CON1: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG CON2: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG © Copyright 2022 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. November 30, 2022 Page 5 / 5