TPP450A TRACOPOWER | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 5
Technical content
AC/DC Medical Power Supply TPP 450A Series, 450 Watt
- High power density 3" x 5" open frame medical power supply
- 450 Watt with forced air cooling, up to
320 Watt convection cooled without
derating up to 50°C
- Medical certification to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
- EMC compliance to IEC/EN 60601-1-2 4th edition
- Risk management process according to ISO 14971 incl. risk management file
- Acceptance criteria for electronic assemblies acc. to IPC-A-610 class 3
- Isolation (4000 VAC) and leakage current (<100 µA) rated for BF
applications
- Standard features: 5 V standby output
12 V fan output, Remote On/Off, Power
Good Signal, variable fan speed
- Operating up to 5000 m altitude
- 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 450A Series of 450 Watt AC/DC power supplies feature a reinforced double I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below 100 µA what makes the units suitable for BF (body floating) applications. The excellent efficiency of up to 94% allows a high power density for the standard 3" x 5" packaging format. Natural convection cooled power up to 320 W at +50°C and 150W at +85°C. Thus you can power your medical device in a quiet and hygienic way as you don’t need to run a fan to cool down the power supply. High reliability is pro- vided by use of industrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Power Output Voltage Output Current max. Output Current max. Efficiency max. nom. (adjustable) (Forced air cooling) (Natural convection) typ. TPP 450-112A-M 450 W 12 VDC (11.0 - 13.0 VDC) 37'500 mA 20'800 mA 91 % TPP 450-115A-M 15 VDC (13.8 - 16.2 VDC) 30'000 mA 16'600 mA 92 % TPP 450-124A-M 24 VDC (22.1 - 25.9 VDC) 18'750 mA 13'300 mA 93 % TPP 450-128A-M 28 VDC (25.8 - 30.2 VDC) 16'100 mA 11'400 mA 93 % TPP 450-136A-M 36 VDC (33.1 - 38.9 VDC) 12'500 mA 8'900 mA 93 % TPP 450-148A-M 48 VDC (44.2 - 51.8 VDC) 9'400 mA 6'650 mA 94 % TPP 450-153A-M 53 VDC (48.8 - 57.2 VDC) 8'550 mA 6'050 mA 94 % Options TPP 450-AUX1 - Optional Cable for auxiliary connection (2 x 4 pin): www.tracopower.com/products/tpp450-aux1.pdf TPP 450-AUX2 - Optional Cable for auxiliary connection (2 x 5 pin): www.tracopower.com/products/tpp450-aux2.pdf www.tracopower.com March 16, 2023 Page 1 / 5
Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 2'400 mA max. - Full Load & Vin = 115 VAC 5'800 mA max. Power Consumption - No load & Vin = 230 VAC 650 mW max. - No load & Vin = 115 VAC 900 mW max. Input Inrush Current - At 230 VAC 100 A max. - At 115 VAC 55 A max. Power Factor - At 230 VAC 0.95 min. (Active Power Factor Correction) - At 115 VAC 0.95 min. (Active Power Factor Correction) Input Protection T 6.3 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±8% (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.5% max. Ripple and Noise (20 MHz Bandwidth) 12 VDC model: 250 mVp-p typ. (w/ 1 µF X7R) 15 VDC model: 300 mVp-p typ. (w/ 1 µF X7R) 24 VDC model: 240 mVp-p typ. (w/ 1 µF X7R) 28 VDC model: 280 mVp-p typ. (w/ 1 µF X7R) 36 VDC model: 360 mVp-p typ. (w/ 1 µF X7R) 48 VDC model: 480 mVp-p typ. (w/ 1 µF X7R) 53 VDC model: 530 mVp-p typ. (w/ 0.1 µF X7R) Capacitive Load 12 VDC model: 31'250 µF max. 15 VDC model: 20'000 µF max. 24 VDC model: 7'820 µF max. 28 VDC model: 5'750 µF max. 36 VDC model: 3'500 µF max. 48 VDC model: 1'960 µF max. 53 VDC model: 1'600 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 230 VAC 12 ms min. - At 115 VAC 12 ms min. Start-up Time - At 230 VAC 2'000 ms max. - At 115 VAC 2'000 ms max. Short Circuit Protection Continuous, Automatic recovery (Level 1, nom.) Latch (Level 2, instantaneous high current) Output Current Limitation 115 - 155% of Iout max. Overvoltage Protection 110 - 135% of Vout nom. (Latch off, Standby Power Source always present) Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 600 µs typ. (50% to 75% Load Step) Output Specifications TPP 450A Series, 450 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 2 / 5
Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Railway Fire Protection EN 45545-2 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp450a Protection Class Class I (Prepared): Connection to PE Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class A (internal filter) EN 55032 class A (internal filter) FCC Part 15 class A (internal filter) FCC Part 18 class A (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A EN 61000-3-2, class D - Voltage Fluctuations & Flicker EN 61000-3-3 (For optimal EMI performance the power supply should be mounted to a grounded aluminium plate (480 x 248 x 12 mm) with electrical contact to the four PCB mounting holes. To comply with safety standards, this plate must be grounded.) EMS Immunity EN 55024 (IT Equipment) EN 55035 (Multimedia) EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 3 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A L to PE: EN 61000-4-5, ±2 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B
115 VAC / 60 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B EMC Specifications Relative Humidity 95% max. (non condensing) General Specifications TPP 450A Series, 450 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 3 / 5
Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C Power Derating - High Temperature Depending on model - Low Input Voltage 1.33 %/V below 100 VAC See application note: www.tracopower.com/overview/tpp450a Over Temperature Protection Switch Off - Protection Mode 110°C to 125°C (Latch off) - Measurement Point See application note: www.tracopower.com/overview/tpp450a (Standby Power Source always present) Cooling System - Option 1 Forced air cooling (with external fan, 21 CFM) - Option 2 Natural convection (20 LFM) Fan Power Source - Characteristic Variable fan speed (temperature regulated) - Output Voltage 12 VDC - Output Current 500 mA max. Standby Power Source - Output Voltage 5 VDC - Output Current 2000 mA max. Remote Control - Voltage Controlled Remote On: 3.0 to 12 VDC or open circuit Off: 0 to 1.2 VDC or short circuit Refers to '+Remote' and '-Remote' Pin - Remote Pin Input Current -0.5 to 1.0 mA (Standby power source is always present) Altitude During Operation 5'000 m max. Switching Frequency 55 - 85 kHz (PFM) Insulation System Reinforced Insulation Working Voltage (rated) 312 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 2'500 VAC - Output to Case or PE, 60 s 2'500 VAC Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 100 µA max. Reliability - Calculated MTBF 410'000 h (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 - Mechanical Shock IEC 60068-2-27 Housing Type Open Frame Mounting Type Chassis Mount Connection Type Pin Connector Weight 462 g Power OK Signal Open collector output - Trigger Threshold 12 VDC model: 9.8 - 11 VDC 15 VDC model: 12.3 - 13.8 VDC 24 VDC model: 19.7 - 22.1 VDC 28 VDC model: 23 - 25.8 VDC 36 VDC model: 29.5 - 33.1 VDC 48 VDC model: 39.4 - 44.2 VDC 53 VDC model: 43.5 - 48.8 VDC - Power OK Low level - Power Off High resistance (Refers to 'PG' and '-Vout' Pin) - Pin Specifications 50 VDC / 50 mA / 120 mW max. Sense Function 8% max. of Vout nom. (If sense function is not used, sense pins should be connected to output pins.) TPP 450A Series, 450 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 4 / 5
Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) - Flammability (EN 45545-2) www.tracopower.com/info/en45545-declaration.pdf Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp450a Outline Dimensions TPP 450A Series, 450 Watt 127.0 (5.00) 3.05 (0.12) 70.1 (2.760) 3.05 (0.12) 3.05 (0.12) FRONT VIEW BOTTOM VIEW CON3 SIDE VIEW TOP VIEW Ø3.2 (0.13) HOLE Screw 1 4 M3x0.5 Voltage Adj. 120.90 (4.760) 76.2 (3.00) 40.1 (1.58) 3 1 CON1 CON2 4-Ø3.2 holes Max. screw penetration: 1.3 (0.05) All dimensions in mm (inch) Tolerance: X.X ±0.5 (X.XX ±0.02) X.XX ±0.25 (X.XXX ±0.01) Screw locked torque: max. 5.2 kgfcm / 0.51 Nm CON1: Molex housing: 09-50-8031 Molex crimp terminals: 08500106 (2478), 08520112 (6838), 45570 CON2: Molex housing: 39-01-2105 Molex crimp terminals: 5556,45750 CON3: Molex housing: 90143-0010 Molex crimp terminals: 90119 Input CON1 Pin Function
1 AC (L)
3 AC (N)
6 –Fan 7 –Sense 8 –Remote
9 No Pin
10 –Standby** Output CON2 Pin* Function 1-5 +Vout 6-10 –Vout *Terminal rated for 13 A max. (at higher current connection has to be split) **Internally connected with –Vout © Copyright 2023 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. March 16, 2023 Page 5 / 5