AC/DC Medical Power Supply TPP 40 Series, 40 Watt

  • Enclosed power supply with screw terminal connection
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
  • Low leakage current <75 µA rated for BF applications
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • EMC compliance to IEC 60601-1-2 ed. 4
  • Protection class l and ll prepared
  • Operating up to 5000 m altitude
  • Ready to meet ErP directive, <0.15 W no load power consumption
  • 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 40 Series of 40 Watt AC/DC power supplies feature a reinforced dou- ble I/O isolation system according to latest medical safety standards IEC/EN/- ES 60601-1 3rd edition for 2 x MOPP up to 5000 m altitude. The leakage cur- rent is below 75 µA what makes the units suitable for BF (body floating) appli- cations. The excellent efficiency of up to 92% allows a high power density for the stan- dard 2.38” x 3.53” packaging format. The full load operating temperature range is –40°C to +70°C while it goes up to 85°C with 50% load derating. The EMC characteristic complies to IEC 60601-1-2 ed.4 and is dedicated for appli- cations in industrial and domestic fields. High reliability is provided by use of in- dustrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Output 1 Output 2 Output 3 Efficiency Power Vnom Imax Vnom Imax Vnom Imax typ. TPP 40-105 40 W

5 VDC 8'000 mA 90 %

TPP 40-112 12 VDC 3'340 mA 92 % TPP 40-115 15 VDC 2'670 mA 92 % TPP 40-124 24 VDC 1'670 mA 92 % TPP 40-221 +12 VDC 3'340 mA +5 VDC 6'000 mA 89 % TPP 40-231 +15 VDC 2'670 mA +5 VDC 6'000 mA 89 % TPP 40-251 +24 VDC 1'670 mA +5 VDC 6'000 mA 86 % TPP 40-321M2 +12 VDC 3'340 mA +5 VDC 6'000 mA –12 VDC 500 mA 88 % TPP 40-331M3 +15 VDC 2'670 mA +5 VDC 6'000 mA –15 VDC 500 mA 88 % TPP 40-3512 +24 VDC 1'670 mA +5 VDC 6'000 mA +12 VDC 500 mA 86 % Note - Total output power must not exceed 40 W. - Other output models are available on request. www.tracopower.com November 30, 2022 Page 1 / 5

Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC single output models: 500 mA max. dual output models: 550 mA max. tripleoutput models: 550 mA max. - Full Load & Vin = 115 VAC single output models: 1'000 mA max. dual output models: 1'050 mA max. triple output models: 1'050 mA max. Power Consumption - No load & Vin = 230 VAC 150 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 150 mW max. Input Inrush Current - At 230 VAC 60 A max. - At 115 VAC 35 A max. Input Protection T 3.15 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±10% (only Output 1) (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. (Output 1) ±2% max. (Output 2 and 3) Regulation - Input Variation (Vmin - Vmax) single output models: 0.2% max. dual output models: 0.2% max. triple output models: 0.2% max. - Load Variation (0 - 100%) single output models: 0.7% max. (5 VDC model) 0.5% max. (other output models) dual output models: 0.5% max. (Output 1) 1.5% max. (Output 2) triple output models: 0.5% max. (Output 1) 1.5% max. (Output 2) 0.7% max. (Output 3) - Cross Regulation (25% / 100% asym. load) dual output models: 1.5% max. triple output models: 1.5% max. Ripple and Noise (20 MHz Bandwidth) - single output 5 VDC model: 75 mVp-p typ. (w/ 10 µF X7R) 12 VDC model: 75 mVp-p typ. (w/ 10 µF X7R) 15 VDC model: 75 mVp-p typ. (w/ 10 µF X7R) 24 VDC model: 75 mVp-p typ. (w/ 1 µF X7R) - dual output 12 / 5 VDC model: 120 / 100 mVp-p typ. (w/ 10 µF X7R) 15 / 5 VDC model: 150 / 100 mVp-p typ. (w/ 10 µF X7R) 24 / 5 VDC model: 240 / 100 mVp-p typ. (w/ 10 µF X7R) - triple output 12 / 5 / -12 VDC model: 120 / 100 / 120 mVp-p typ. (w/ 10 µF X7R) 15 / 5 / -15 VDC model: 150 / 100 / 150 mVp-p typ. (w/ 10 µF X7R) 24 / 5 / 12 VDC model: 240 / 100 / 120 mVp-p typ. (w/ 10 µF X7R) Output Specifications TPP 40 Series, 40 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 2 / 5

Capacitive Load - single output 5 VDC model: 16'000 µF max. 12 VDC model: 2'785 µF max. 15 VDC model: 1'780 µF max. 24 VDC model: 700 µF max. - dual output 12 / 5 VDC model: 1'750 / 2'000 µF max. 15 / 5 VDC model: 1'670 / 2'000 µF max. 24 / 5 VDC model: 440 / 2'000 µF max. - triple output 12 / 5 / -12 VDC model: 1'750 / 2'000 / 420 µF max. 15 / 5 / -15 VDC model: 1'670 / 2'000 / 420 µF max. 24 / 5 / 12 VDC model: 440 / 2'000 / 420 µF max. Minimum Load Not required (0.5 W for Vout1 and Vout2 if Vout3 = Full Load) Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 115 VAC 25 ms min. Start-up Time - At 230 VAC 1'000 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 115 - 180% of Iout max. 145% typ. of Iout max. (Pout 1 + Pout 2) Overvoltage Protection 125 - 140% of Vout nom. (only Output 1) Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 600 µs typ. (50% to 75% Load Step) (Only Output 1) Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp40 Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 18 class B (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A - Voltage Fluctuations & Flicker EN 61000-3-3 EMC Specifications TPP 40 Series, 40 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 3 / 5

EMS Immunity EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A L to PE: EN 61000-4-5, ±2 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B

115 VAC / 60 Hz: EN 61000-4-11

30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C Power Derating - High Temperature Depending on model - Low Input Voltage Depending on model See application note: www.tracopower.com/overview/tpp40 Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 50 - 140 kHz (PWM) (Output 1) 750 kHz typ. (PWM) (Output 2) 510 kHz typ. (PWM) (Output 3) Insulation System Reinforced Insulation Working Voltage (rated) 258 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 2'500 VAC - Output to Case or PE, 60 s 2'500 VAC Creepage - Input to Output 8 mm min. Clearance - Input to Output 8 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 75 µA max. Reliability - Calculated MTBF 3'000'000 h (for single output models) 1'700'000 h (for multi output models) (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 5 g, 3 axis, sine sweep, 5-500 Hz, 1 oct/min - Mechanical Shock IEC 60068-2-27 50 g, 3 axis, half sine, 11 ms Housing Material Aluminum Housing Type Metal Case Mounting Type Chassis Mount Connection Type Screw Terminal Weight - single output 169 g - dual output 216 g - triple output 216 g General Specifications TPP 40 Series, 40 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 4 / 5

Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp40 Outline Dimensions Single Output Models Multi Output Models TPP 40 Series, 40 Watt Max. corner screw penetration: 2.3 (0.09) Max. center screw penetration: 2.0 (0.08) All dimensions in mm (inch) Tolerance: X.X ±0.5 (X.XX ±0.02) X.XX ±0.25 (X.XXX ±0.010) The screw locked torque: max. 5.0 kgfcm / 0.49 Nm Terminal screw locked torque: max. 4.0 kgfcm / 0.39 Nm 60.5 (2.38) 44.45 (1.75) 8.0 (0.31) 33.3 (1.31) 1.2 (0.047) 10.0 (0.39) CON2 Voltage Adj. 4-M3x0.5 CON1 3 84.24 (3.317) 89.7 (3.53) 40.8 (1.606) Screw Terminal Input (CON1) Output (CON2) Pin Function Pin* Function

1 Line 1,2 –Vout

3 Neutral 3,4 +Vout

*Terminal rated for 10 A max. (at higher current connection has to be split) CON1: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG CON2: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG Max. corner screw penetration: 2.3 (0.09) Max. center screw penetration: 2.0 (0.08) All dimensions in mm (inch) Tolerance: X.X ±0.5 (X.XX ±0.02) X.XX ±0.25 (X.XXX ±0.010) The screw locked torque: max. 5.0 kgfcm / 0.49 Nm Terminal screw locked torque: max. 4.0 kgfcm / 0.39 Nm 60.5 (2.38) 44.45 (1.75) 8.0 (0.31) 34.5 (1.36) 1.2 (0.047) 10.0 (0.39) CON2 Voltage Adj. CON1 3 96.94 (3.817) 102.4 (4.03) 40.8 (1.606) 4-M3x0.5 Screw Terminal Input (CON1) Output (CON2) Pin Function Pin* Function

1 Line 1 Vout 3

3 Neutral 2,3 COM

4,5 Vout 2

6 Vout 1

*Terminal rated for 10 A max. (at higher current connection has to be split) CON1: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG CON2: Terminal Block mates with Screw locked torque MAX 2Kgf.cm/0.2N.m Wire dimension range: 26 - 16 AWG © Copyright 2022 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. November 30, 2022 Page 5 / 5