TPP40E-J TRACOPOWER | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 5

Technical content

AC/DC Medical Power Supply TPP 40E-J Series, 40 Watt

  • Compact encapsulated 40 Watt Chassis mount module (4.3"x2.2")
  • I/O isolation 4000 VAC rated for 250 VAC working voltage
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • Low leakage current <100 μA rated for BF applications
  • Operating temperature –40°C to 80°C
  • EMC compliance according to IEC 60601-1-2 4th edition and EN55032 class B
  • Operating up to 5000m altitude
  • 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 40E-J is a 40 Watt encapsulated AC/DC modules in a Chassis mount package. It features a reinforced double I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The leakage cur- rent is below 100 μA and makes the module suitable for BF (body floating) ap- plications. The excellent efficiency of up to 93% allows a high power density and compact design (4.3” x 2.2”). The operating temperature range is –40°C to +80°C with derating above 60°C. The EMC characteristic is dedicated for ap- plications in industrial and medical fields. High reliability is provided by use of high quality components and an excellent thermal management making the TPP 40E-J an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. (adjustable) max. typ. TPP 40-105E-J 40 W 5 VDC (4.5 - 5.5 VDC) 8'000 mA 90 % TPP 40-112E-J 12 VDC (10.8 - 13.2 VDC) 3'340 mA 92 % TPP 40-115E-J 15 VDC (13.5 - 16.5 VDC) 2'670 mA 92 % TPP 40-124E-J 24 VDC (21.6 - 26.4 VDC) 1'670 mA 92 % TPP 40-136E-J 36 VDC (32.4 - 39.6 VDC) 1'120 mA 92 % TPP 40-148E-J 48 VDC (43.2 - 52.8 VDC) 840 mA 93 % www.tracopower.com February 13, 2023 Page 1 / 5

Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 500 mA max. - Full Load & Vin = 115 VAC 980 mA max. Power Consumption - No load & Vin = 230 VAC 210 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 210 mW max. Input Inrush Current - At 230 VAC 80 A max. - At 115 VAC 30 A max. Input Protection T 3.15 A / 250 VAC (Internal Fuse L & N) Recommended Input Fuse 3'150 mA (slow blow) (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±10% (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.5% max. Ripple and Noise (20 MHz Bandwidth) 5 VDC model: 100 mVp-p max. (w/ 10 µF) 12 VDC model: 100 mVp-p max. (w/ 10 µF) 15 VDC model: 100 mVp-p max. (w/ 10 µF) 24 VDC model: 120 mVp-p max. (w/ 1 µF) 36 VDC model: 120 mVp-p max. (w/ 1 µF) 48 VDC model: 200 mVp-p max. (w/ 0.1 µF) 5 VDC model: 75 mVp-p typ. (w/ 10 µF) 12 VDC model: 75 mVp-p typ. (w/ 10 µF) 15 VDC model: 75 mVp-p typ. (w/ 10 µF) 24 VDC model: 75 mVp-p typ. (w/ 1 µF) 36 VDC model: 75 mVp-p typ. (w/ 1 µF) 48 VDC model: 150 mVp-p typ. (w/ 0.1 µF) Capacitive Load 5 VDC model: 16'000 µF max. 12 VDC model: 2'785 µF max. 15 VDC model: 1'780 µF max. 24 VDC model: 700 µF max. 36 VDC model: 310 µF max. 48 VDC model: 175 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 230 VAC 120 ms min. - At 115 VAC 25 ms min. Start-up Time - At 230 VAC 1'000 ms max. - At 115 VAC 1'000 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 120 - 180% of Iout max. 145% typ. of Iout max. Overvoltage Protection 130% typ. of Vout nom. 125 - 140% of Vout nom. Transient Response - Response Deviation 3% max. (50% to 75% Load Step at 2.5 A/µs) - Response Time 600 µs typ. (50% to 75% Load Step at 2.5 A/µs) Output Specifications TPP 40E-J Series, 40 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com February 13, 2023 Page 2 / 5

Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp40e-j Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria B - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B

115 VAC / 60 Hz: EN 61000-4-11

30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B EMC Specifications Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C Power Derating - High Temperature Depending on model See application note: www.tracopower.com/overview/tpp40e-j Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 50 - 130 kHz (PWM QR) Insulation System Reinforced Insulation Working Voltage (rated) 250 VAC General Specifications TPP 40E-J Series, 40 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com February 13, 2023 Page 3 / 5

Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 2'500 VAC - Output to Case or PE, 60 s 2'500 VAC Creepage - Input to Output 8 mm min. - Input to Case or PE 4 mm min. - Output to Case or PE 4 mm min. Clearance - Input to Output 4 mm min. - Input to Case or PE 4 mm min. - Output to Case or PE 4 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Isolation Capacitance - Input to Output, 100 kHz, 1 V 1'250 pF typ. / 1'500 pF max. Leakage Current - Touch Current 100 µA max. Reliability - Calculated MTBF 3'010'000 h (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 5 g, 3 axis, sine sweep, 3x30 min, 5-500 Hz - Mechanical Shock IEC 60068-2-27 50 g, 3 axis, 11 ms - Thermal Shock MIL-STD-810F -40 to +85°C, 72 cycles, 30 min each Housing Material Plastic (UL 94 V-0 rated) Potting Material Silicone (UL 94 V-0 rated) Housing Type Plastic Case Mounting Type Chassis Mount Connection Type Pin Connector Weight 295 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp40e-j TPP 40E-J Series, 40 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com February 13, 2023 Page 4 / 5

TPP 40E-J Series, 40 Watt Dimensions in mm (inch) Tolerances: x.x ±0.5 (x.xx ±0.02) x.xx ±0.25 (x.xxx ±0.010) CON2 CON1 TOP VIEW Voltage Adj. Ø 3.5 (Ø 0.138) Output On LED FRONT VIEW 109.2 (4.300) 55.9 (2.200) 44.5 (1.750) 30.5 (1.200) 96.5 (3.800) Pin connectors Input (CON1) Output (CON2) Pin Function Pin Function

1 Line 1, 2 –Vout

2 Neutral 3, 4 +Vout

Input: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-3N Output: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-4N NC: Not connected © Copyright 2023 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. February 13, 2023 Page 5 / 5