AC/DC Medical Power Supply TPP 30-D Series, 30 Watt

  • High power density power supply (encapsulated)
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
  • Low leakage current <75 µA rated for BF applications
  • EMC compliance to IEC 60601-1-2 4th edition
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • Protection class ll
  • Operating up to 5000 m altitude
  • Ready to meet ErP directive, no load power consumption <60 mW
  • 5-year product warranty IEC 60601-1 IEC 60335-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 30-D AC/DC power supplies feature a reinforced double I/O isolation system according to medical safety standards IEC/EN/ES 60601-1 3rd edition for 2 x MOPP approved for an operating altitude of 5000 m. The earth leakage current is below 75 µA what makes the units suitable for BF (body floating) ap- plications. The excellent efficiency of up to 92% offers a high power density in the packaging format 1.5” x 2.9”. The full load operating temperature range covers –40°C to +60°C while it goes up to 85°C with 50% load derating. The units operate in compliance to the medical EMC emission and immunity levels according to latest standard IEC 60601-1-2 4th edition. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. (adjustable) max. typ. TPP 30-103-D 20 W 3.3 VDC (2.97 - 3.63 VDC) 6'000 mA 84 % TPP 30-105-D 30 W 5 VDC (4.5 - 5.5 VDC) 6'000 mA 87 % TPP 30-109-D 9 VDC (8.1 - 9.9 VDC) 3'340 mA 88 % TPP 30-112-D 12 VDC (10.8 - 13.2 VDC) 2'500 mA 91 % TPP 30-115-D 15 VDC (13.5 - 16.5 VDC) 2'000 mA 91 % TPP 30-124-D 24 VDC (21.6 - 26.4 VDC) 1'250 mA 90 % TPP 30-136-D 36 VDC (32.4 - 39.6 VDC) 840 mA 90 % TPP 30-148-D 48 VDC (43.2 - 52.8 VDC) 630 mA 92 % www.tracopower.com March 16, 2023 Page 1 / 5

Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 400 mA max. - Full Load & Vin = 115 VAC 800 mA max. Power Consumption - No load & Vin = 230 VAC 60 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 60 mW max. Input Inrush Current - At 230 VAC 40 A max. - At 115 VAC 25 A max. Input Protection T 1.6 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment See application note: ±10% (By external trim resistor) www.tracopower.com/overview/tpp30-d Output power must not exceed rated power! Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.7% max. (3.3 and 5 VDC model) 0.5% max. (other output models) Ripple and Noise (20 MHz Bandwidth) 3.3 VDC model: 50 mVp-p typ. (w/ 10 µF X7R) 5 VDC model: 50 mVp-p typ. (w/ 10 µF X7R) 9 VDC model: 50 mVp-p typ. (w/ 10 µF X7R) 12 VDC model: 50 mVp-p typ. (w/ 1 µF X7R) 15 VDC model: 50 mVp-p typ. (w/ 1 µF X7R) 24 VDC model: 50 mVp-p typ. (w/ 1 µF X7R) 36 VDC model: 50 mVp-p typ. (w/ 1 µF X7R) 48 VDC model: 50 mVp-p typ. (w/ 0.1 µF X7R) Capacitive Load 3.3 VDC model: 10'000 µF max. 5 VDC model: 12'000 µF max. 9 VDC model: 3'720 µF max. 12 VDC model: 2'085 µF max. 15 VDC model: 1'350 µF max. 24 VDC model: 520 µF max. 36 VDC model: 235 µF max. 48 VDC model: 130 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 115 VAC 16 ms min. Start-up Time - At 230 VAC 1'500 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 110 - 170% of Iout max. 140% typ. of Iout max. Overvoltage Protection 125 - 140% of Vout nom. Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 500 µs typ. (50% to 75% Load Step) Output Specifications TPP 30-D Series, 30 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 2 / 5

Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Household EN 60335-1 IEC 60335-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Power Transformers IEC 61558-1 IEC 61558-2-16 - Certification Documents www.tracopower.com/overview/tpp30-d Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55014-1 (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EN 55014-1 (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 55024 (IT Equipment) EN 55035 (Multimedia) EN 60601-1-2 edition 4 (Medical Devices) EN 55014-2 (Household Appliances Tools) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A 60%, 1 period, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria A

115 VAC / 60 Hz: EN 61000-4-11

30%, 25 periods, perf. criteria A 60%, 1 period, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria A EMC Specifications Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C General Specifications TPP 30-D Series, 30 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 3 / 5

Power Derating - High Temperature Depending on model - Low Input Voltage 4 %/V below 90 VAC See application note: www.tracopower.com/overview/tpp30-d Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 30 - 60 kHz (PWM) Insulation System Reinforced Insulation Working Voltage (rated) 272 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 1'500 VAC - Output to Case or PE, 60 s 1'500 VAC Creepage - Input to Output 8 mm min. Clearance - Input to Output 8 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 75 µA max. Reliability - Calculated MTBF 3'300'000 h (MIL-HDBK-217F, ground benign) Washing Process According to Cleaning Guideline www.tracopower.com/info/cleaning.pdf Environment - Vibration IEC 60068-2-6 - Mechanical Shock IEC 60068-2-27 Potting Material Silicone (UL 94 V-0 rated) Pin Material Copper Pin Foundation Plating Nickel (2 - 3 µm) Pin Surface Plating Tin (3 - 5 µm), matte Housing Type Plastic Case Mounting Type PCB Mount Connection Type THD (Through-Hole Device) Weight 106 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp30-d TPP 30-D Series, 30 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 4 / 5

TPP 30-D Series, 30 Watt Pin diameter: 1.0 mm (0.04 inch) Dimension in mm, () = inch Tolerances: x.x ±0.50 (±0.02) x.xx ±0.25 (±0.01) Pin pitch tolerance: ±0.25 (±0.010) Pin dimension tolerance: ±0.10 (±0.004) Bottom view 66.14 (2.604) 73.3 (2.89) 38.1 (1.50) 15.1 (0.59) 7.92 (0.312) 10.92 (0.43) 3.96 (0.156) 18.85 (0.742) 25.4 (1.00) 3.8 (0.15) 3.6 (0.14) PCB Pinout Pin Function

1 Neutral

2 Line

4 –Vout

5 Trim

© Copyright 2023 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. March 16, 2023 Page 5 / 5