AC/DC Medical Power Supply TPP 180A-M Series, 180 Watt
- High power-density: 180 Watt in 3”x2” package (open frame)
- I/O isolation 4000 VAC rated for 250 VAC working voltage
- Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
- Risk management process according to ISO 14971 incl. risk management file
- Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
- Peak power operation up to 220 Watt for 5s
- Operating temperature –40°C to 85°C
- Active power factor correction >0.9
- Operating up to 5000m altitude
- 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 180A-M is series of open frame AC/DC power supplies coming in an open frame package. They feature a reinforced double I/O isolation (4000 VAC) system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The leakage current is below 100 μA and makes the units suitable for BF (body floating) applications. The excellent efficiency of up to 94% allows a high power-density and compact design (3” x 2”). The operating temperature range is –40°C to +85°C with derating above 50°C. In natural convection oper- ation these power supplies deliver 150 Watt going up to 180 Watt with forced air cooling. Additionally, they can deliver 220 Watt peak power for 5s. The EMC characteristic is dedicated for applications in industrial and medical fields. High reliability is provided by using high quality components and an excellent ther- mal management making the TPP 180A-M an ideal solution for industrial and medical devices and for demanding safety and space critical applications. Models Order Code Output Power Output Voltage Output Current max. Output Current max. Output Current Efficiency max. nom. (adjustable) (Forced air cooling) (Natural convection) peak typ. TPP 180-112A-M 180 W 12 VDC (11.0 - 13.0 VDC) 15'000 mA 12'500 mA 18'334 mA 92 % TPP 180-115A-M 15 VDC (13.8 - 16.2 VDC) 12'000 mA 10'000 mA 14'667 mA 92 % TPP 180-124A-M 24 VDC (22.1 - 25.9 VDC) 7'500 mA 6'250 mA 9'167 mA 94 % TPP 180-136A-M 36 VDC (33.1 - 38.9 VDC) 5'000 mA 4'170 mA 6'111 mA 93 % TPP 180-148A-M 48 VDC (44.2 - 51.8 VDC) 3'750 mA 3'130 mA 4'583 mA 93 % TPP 180-153A-M 53 VDC (48.8 - 57.2 VDC) 3'400 mA 2'830 mA 4'151 mA 93 % Options on demand (backorder with MOQ non stocking item) - Optional model with 18 VDC / 10'000 mA - Optional model with 28 VDC / 6'430 mA www.tracopower.com November 30, 2022 Page 1 / 5
Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 1'500 mA max. - Full Load & Vin = 115 VAC 2'550 mA max. Power Consumption - No load & Vin = 230 VAC 200 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 200 mW max. Input Inrush Current - At 230 VAC 100 A max. - At 115 VAC 50 A max. Power Factor - At 230 VAC 0.9 min. (Active Power Factor Correction) - At 115 VAC 0.9 min. (Active Power Factor Correction) Input Protection T 4 A / 250 VAC Recommended Input Fuse 4'000 mA (slow blow) (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±8% (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.5% max. Boost Power Output Current peak: See model table Peak power time: 5 s max. Peak power duty cycle: 20% max. Average operation power: 99 W max. (at natural convection) Ripple and Noise (20 MHz Bandwidth) 12 VDC model: 200 mVp-p max. (w/ 1 µF) 15 VDC model: 200 mVp-p max. (w/ 1 µF) 18 VDC model: 220 mVp-p max. (w/ 1 µF) 24 VDC model: 220 mVp-p max. (w/ 1 µF) 28 VDC model: 220 mVp-p max. (w/ 1 µF) 36 VDC model: 220 mVp-p max. (w/ 1 µF) 48 VDC model: 350 mVp-p max. (w/ 0.1 µF) 53 VDC model: 350 mVp-p max. (w/ 0.1 µF) Capacitive Load 12 VDC model: 10'000 µF max. 15 VDC model: 6'800 µF max. 18 VDC model: 4'700 µF max. 24 VDC model: 2'700 µF max. 28 VDC model: 1'800 µF max. 36 VDC model: 1'200 µF max. 48 VDC model: 680 µF max. 53 VDC model: 560 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 230 VAC 10 ms min. - At 115 VAC 10 ms min. Start-up Time - At 230 VAC 1'500 ms max. - At 115 VAC 1'500 ms max. Short Circuit Protection Continuous, Automatic recovery Output Specifications TPP 180A-M Series, 180 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 2 / 5
Output Current Limitation 130 - 175% of Iout max. 150% typ. of Iout max. Overvoltage Protection 115 - 135% of Vout nom. Transient Response - Response Deviation 3% typ. / 10% max. (75% to 100% Load Step) - Response Time 600 µs typ. (75% to 100% Load Step) Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp180a-m Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class A (internal filter) EN 55032 class A (internal filter) FCC Part 15 class A (internal filter) FCC Part 18 class A (internal filter) - Harmonic Current Emissions EN 61000-3-2, class D - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 55024 (IT Equipment) EN 55035 (Multimedia) EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A L to PE: EN 61000-4-5, ±2 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A 1 s: EN 61000-4-8, 1000 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria B >95%, 250 periods, perf. criteria B
115 VAC / 60 Hz: EN 61000-4-11
30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria B >95%, 250 periods, perf. criteria B EMC Specifications Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C General Specifications TPP 180A-M Series, 180 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 3 / 5
Power Derating - High Temperature Depending on model - Low Input Voltage 1.33 %/V below 100 VAC See application note: www.tracopower.com/overview/tpp180a-m Over Temperature Protection Switch Off - Protection Mode 115°C min. / 125°C typ. / 135°C max. (Automatic recovery at 112°C typ.) - Measurement Point See application note: www.tracopower.com/overview/tpp180a-m Cooling System - Option 1 Forced air cooling (with external fan, 10 CFM) - Option 2 Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 135 - 205 kHz (PWM) 170 kHz typ. (PWM) Insulation System Reinforced Insulation Working Voltage (rated) 250 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 2'500 VAC - Output to Case or PE, 60 s 2'500 VAC Creepage - Input to Output 27.5 mm min. - Input to Case or PE 5 mm min. - Output to Case or PE 4.3 mm min. Clearance - Input to Output 9.3 mm min. - Input to Case or PE 5 mm min. - Output to Case or PE 4.3 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Isolation Capacitance - Input to Output, 100 kHz, 1 V 1'250 pF typ. / 1'500 pF max. Leakage Current (264 VAC / 63 Hz) - Touch Current 100 µA max. Distance Through Isolation 0.4 mm Reliability - Calculated MTBF 1'145'000 h (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 5 g, 3 axis, sine sweep, 3x30 min, 5-500 Hz - Mechanical Shock IEC 60068-2-27 50 g, 3 axis, 11 ms - Thermal Shock MIL-STD-810F Housing Type Open Frame Mounting Type Chassis Mount Connection Type Pin Connector Weight 162 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp180a-m TPP 180A-M Series, 180 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 4 / 5
TPP 180A-M Series, 180 Watt 76.2 (3.00) 50.8 (2.00) 44.45 (1.75) 1.6 (0.06) 3.0 (0.12) max. CON1 CON2 69.8 (2.748) Voltage Adj. Ø 3.3 (0.13) 31.6 (1.24) FRONT VIEW All dimensions in mm (inch) Tolerance: X.X ±0.5 (X.XX ±0.02) X.XX ±0.25 (X.XXX ±0.01) CON2 screw locked torque: max. 2.5 kgfcm / 0.25 Nm wires 24 -14 AWG Pin connectors Input (CON1) Output (CON2) Pin Function Pin Function
1 AC (N) / D 1 +Vout
3 AC (L) / DC+ 2 –Vout
CON1: Molex Housing 09-50-8031 Molex Crimp Terminals 08500106 (2478), 08520112 (6838), 45570 © Copyright 2022 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. November 30, 2022 Page 5 / 5