TPP150A-J TRACOPOWER | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 5

Technical content

AC/DC Medical Power Supply TPP 150A-J Series, 150 Watt

  • Open frame 150 W power supply with JST connection in 2.0” x 4.0” package
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
  • Low leakage current <100 µA rated for BF applications
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • Active power factor correction >0.95
  • Protection class l and ll prepared
  • Operating up to 5000 m altitude
  • Ready to meet ErP directive, <0.3 W no load power consumption
  • 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 150A-J series of 150 Watt AC/DC open frame power supplies feature a reinforced double I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below 100 µA which makes the units suitable for BF (body floating) applications. The excellent efficiency of up to 92% allows a high power density for the standard 2.0” x 4.0” packaging format. The full load operating temperature range is –25°C to +55°C while it goes up to 80°C with 50% load derating. The EMC characteristic is dedicated for applications in industrial and medical fields. High reliability is provided by the use of industrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Power Output Voltage Output Current max. Output Current max. Efficiency max. nom. (adjustable) (Forced air cooling) (Natural convection) typ. TPP 150-112A-J 150 W 12 VDC (10.8 - 13.2 VDC) 12'500 mA 8'340 mA 91 % TPP 150-115A-J 15 VDC (13.5 - 16.5 VDC) 10'000 mA 7'340 mA 92 % TPP 150-124A-J 24 VDC (21.6 - 26.4 VDC) 6'250 mA 4'590 mA 92 % TPP 150-128A-J 28 VDC (25.2 - 30.8 VDC) 5'360 mA 3'930 mA 92 % TPP 150-136A-J 36 VDC (32.4 - 39.6 VDC) 4'170 mA 3'060 mA 92 % TPP 150-148A-J 48 VDC (43.2 - 52.8 VDC) 3'130 mA 2'090 mA 92 % www.tracopower.com March 16, 2023 Page 1 / 5

Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 800 mA max. - Full Load & Vin = 115 VAC 1'700 mA max. Power Consumption - No load & Vin = 230 VAC 300 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 300 mW max. Input Inrush Current - At 230 VAC 60 A max. - At 115 VAC 35 A max. Power Factor - At 230 VAC 0.95 min. (Active Power Factor Correction) - At 115 VAC 0.95 min. (Active Power Factor Correction) Input Protection T 3.15 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±10% (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.5% max. Ripple and Noise (20 MHz Bandwidth) 12 VDC model: 120 mVp-p typ. (w/ 1 µF X7R) 15 VDC model: 150 mVp-p typ. (w/ 1 µF X7R) 24 VDC model: 220 mVp-p typ. (w/ 1 µF X7R) 28 VDC model: 220 mVp-p typ. (w/ 1 µF X7R) 36 VDC model: 250 mVp-p typ. (w/ 1 µF X7R) 48 VDC model: 250 mVp-p typ. (w/ 0.1 µF X7R) Capacitive Load 12 VDC model: 10'400 µF max. 15 VDC model: 6'600 µF max. 24 VDC model: 2'600 µF max. 28 VDC model: 1'900 µF max. 36 VDC model: 1'150 µF max. 48 VDC model: 650 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 230 VAC 16 ms min. - At 115 VAC 16 ms min. Start-up Time - At 230 VAC 1'000 ms max. - At 115 VAC 1'000 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 115 - 150% of Iout max. Overvoltage Protection 115 - 135% of Vout nom. Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 500 µs typ. (50% to 75% Load Step) Output Specifications TPP 150A-J Series, 150 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 2 / 5

Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 60950-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp150a-j Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class A (internal filter) EN 55032 class A (internal filter) FCC Part 15 class A (internal filter) FCC Part 18 class A (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A EN 61000-3-2, class D - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 55024 (IT Equipment) EN 55035 (Multimedia) EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A L to PE: EN 61000-4-5, ±2 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 10 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A 60%, 5 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 250 periods, perf. criteria B EMC Specifications Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –25°C to +85°C - Storage Temperature –40°C to +85°C Power Derating - High Temperature Depending on model - Low Input Voltage 1.33 %/V below 100 VAC See application note: www.tracopower.com/overview/tpp150a-j Cooling System - Option 1 Forced air cooling (with external fan, 10 CFM) - Option 2 Natural convection (20 LFM) Fan Power Source - Characteristic Variable fan speed (temperature regulated) - Output Voltage 12 VDC - Output Current 500 mA max. General Specifications TPP 150A-J Series, 150 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 3 / 5

Altitude During Operation 5'000 m max. Switching Frequency 45 - 75 kHz (PWM QR) Insulation System Reinforced Insulation Working Voltage (rated) 250 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 2'000 VAC - Output to Case or PE, 60 s 2'000 VAC Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 100 µA max. Reliability - Calculated MTBF 786'000 h (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 5 g, 3 axis, sine sweep, 10-500 Hz, 1 oct/min - Mechanical Shock IEC 60068-2-27 50 g, 3 axis, half sine, 11 ms Housing Type Open Frame Mounting Type Chassis Mount Connection Type Pin Connector Weight 187 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp150a-j TPP 150A-J Series, 150 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 4 / 5

TPP 150A-J Series, 150 Watt Voltage adjustment Dimension in mm, () = inch Tolerances: x.x ±0.50 (±0.02) x.xx ±0.25 (±0.01) 95.20 (3.748) 2.5 max. (0.10 max.) 101.6 (4.00) 1.6 (0.06) 29.5 (1.16) 50.8 (2.00) 44.45 (1.750) FRONT VIEW TOP VIEW Ø 3.2 (Ø 0.13) CON1 CON2 CON3 Pin connectors Input (CON1) Output (CON2) Fan (CON3) Pin Function Pin* Function Pin Function

1 Line 1-3 –Vout 1 –Fan

3 Neutral 4-6 +Vout 2 +Fan

*Terminal rated for 7 A max. (at higher current connection has to be split) CON1: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-3N CON2: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-6N CON3: Molex series mates with Molex crimp terminals: 2759 and Molex housing: 22-01-1022 © Copyright 2023 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. March 16, 2023 Page 5 / 5