TPP15-J TRACOPOWER | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 5

Technical content

AC/DC Medical Power Supply TPP 15-J Series, 15 Watt

  • High power density power supply (encapsulated)
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
  • Low leakage current <75 µA rated for BF applications
  • EMC compliance to IEC 60601-1-2 4th edition
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • Protection class ll
  • Operating up to 5000 m altitude
  • Ready to meet ErP directive, no load power consumption <75 mW
  • 5-year product warranty IEC 60601-1 IEC 60335-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 15-J AC/DC power supplies feature a reinforced double I/O isolation system according to medical safety standards IEC/EN/ES 60601-1 3rd edition for 2 x MOPP approved for an operating altitude of 5000 m. The earth leakage current is below 75 µA what makes the units suitable for BF (body floating) ap- plications. The excellent efficiency of up to 88.5% offers a high power density in the packaging format 1.1” x 2.8”. The full load operating temperature range covers –40°C to +70°C while it goes up to 85°C with 50% load derating. The units operate in compliance to the medical EMC emission and immunity levels according to latest standard IEC 60601-1-2 4th edition. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. max. typ. TPP 15-103-J 13.2 W 3.3 VDC 4'000 mA 84 % TPP 15-105-J 15 W

5 VDC 3'000 mA 86 %

TPP 15-109-J 9 VDC 1'670 mA 86 % TPP 15-112-J 12 VDC 1'250 mA 87 % TPP 15-115-J 15 VDC 1'000 mA 87 % TPP 15-124-J 24 VDC 625 mA 88 % TPP 15-136-J 36 VDC 417 mA 88 % TPP 15-148-J 48 VDC 313 mA 89 % www.tracopower.com March 16, 2023 Page 1 / 5

Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 300 mA max. - Full Load & Vin = 115 VAC 450 mA max. Power Consumption - No load & Vin = 230 VAC 75 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 75 mW max. Input Inrush Current - At 230 VAC 40 A max. - At 115 VAC 25 A max. Input Protection T 1.6 A / 250 VAC (Internal Fuse) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.7% max. (3.3 and 5 VDC model) 0.5% max. (other output models) Ripple and Noise (20 MHz Bandwidth) 3.3 VDC model: 40 mVp-p typ. (w/ 10 µF X5R) 5 VDC model: 40 mVp-p typ. (w/ 10 µF X5R) 9 VDC model: 70 mVp-p typ. (w/ 10 µF X5R) 12 VDC model: 70 mVp-p typ. (w/ 10 µF X5R) 15 VDC model: 70 mVp-p typ. (w/ 10 µF X5R) 24 VDC model: 100 mVp-p typ. (w/ 10 µF X5R) 36 VDC model: 100 mVp-p typ. (w/ 10 µF X5R) 48 VDC model: 140 mVp-p typ. (w/ 1 µF X7R) Capacitive Load 3.3 VDC model: 6'000 µF max. 5 VDC model: 4'000 µF max. 9 VDC model: 1'860 µF max. 12 VDC model: 1'200 µF max. 15 VDC model: 820 µF max. 24 VDC model: 470 µF max. 36 VDC model: 220 µF max. 48 VDC model: 150 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 115 VAC 8 ms min. Start-up Time - At 230 VAC 500 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 120 - 200% of Iout max. 145% typ. of Iout max. Overvoltage Protection 125 - 140% of Vout nom. Transient Response - Response Deviation 8% max. (75% to 100% Load Step) - Response Time 500 µs typ. (75% to 100% Load Step) Output Specifications TPP 15-J Series, 15 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 2 / 5

Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Household EN 60335-1 IEC 60335-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Power Transformers IEC 61558-1 IEC 61558-2-16 - Certification Documents www.tracopower.com/overview/tpp15-j Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55014-1 (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EN 55014-1 (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 55024 (IT Equipment) EN 55035 (Multimedia) EN 60601-1-2 edition 4 (Medical Devices) EN 55014-2 (Household Appliances Tools) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A 60%, 1 period, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria A

115 VAC / 60 Hz: EN 61000-4-11

30%, 25 periods, perf. criteria A 60%, 1 period, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria A EMC Specifications Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –40°C to +85°C - Storage Temperature –40°C to +85°C General Specifications TPP 15-J Series, 15 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 3 / 5

Power Derating - High Temperature Depending on model - Low Input Voltage 4 %/V below 90 VAC See application note: www.tracopower.com/overview/tpp15-j Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 75 - 95 kHz (PWM) Insulation System Reinforced Insulation Working Voltage (rated) 250 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 1'500 VAC - Output to Case or PE, 60 s 1'500 VAC Creepage - Input to Output 8 mm min. Clearance - Input to Output 8 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 75 µA max. Reliability - Calculated MTBF 3'100'000 h (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 - Mechanical Shock IEC 60068-2-27 Potting Material Silicone (UL 94 V-0 rated) Housing Type Plastic Case Mounting Type Chassis Mount Connection Type Pin Connector Weight 48 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp15-j TPP 15-J Series, 15 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 4 / 5

TPP 15-J Series, 15 Watt Dimension in mm, () = inch Tolerances: x.x ±0.50 (±0.02) x.xx ±0.25 (±0.01) 71.7 (2.82) 58.00 (2.283) Output 28.9 (1.14) 21.50 (0.846) Input Ø 3.4 (0.13) Front view 20.9 (0.82) Top view Pin connectors Input Output Pin Function Pin Function

1 Line 1 –Vout

3 Neutral 2 +Vout

Input: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-3N Output: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-2N © Copyright 2023 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. March 16, 2023 Page 5 / 5