AC/DC Medical Power Supply TPP 100A-J Series, 100 Watt

  • Open frame 100 W power supply with JST connection in 2.0” x 3.0” package
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP
  • Low leakage current <75 µA rated for BF applications
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • Active power factor correction >0.95
  • Protection class l and ll prepared
  • Operating up to 5000 m altitude
  • Ready to meet ErP directive, <0.3 W no load power consumption
  • 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TPP 100A-J series of 100 Watt AC/DC open frame power supplies feature a reinforced double I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below 75 µA which makes the units suitable for BF (body floating) applications. The excellent efficiency of up to 92% allows a high power density for the standard 2.0” x 3.0” packaging format. The full load operating temperature range is –25°C to +55°C while it goes up to 80°C with 50% load derating. The EMC characteristic is dedicated for applications in industrial and medical fields. High reliability is provided by the use of industrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. (adjustable) max. typ. TPP 100-112A-J 100 W 12 VDC (10.8 - 13.2 VDC) 8'340 mA 91 % TPP 100-115A-J 15 VDC (13.5 - 16.5 VDC) 6'670 mA 92 % TPP 100-124A-J 24 VDC (21.6 - 26.4 VDC) 4'170 mA 92 % TPP 100-128A-J 28 VDC (25.2 - 30.8 VDC) 3'580 mA 92 % TPP 100-136A-J 36 VDC (32.4 - 39.6 VDC) 2'780 mA 91 % TPP 100-148A-J 48 VDC (43.2 - 52.8 VDC) 2'090 mA 91 % Options on demand (backorder with MOQ non stocking item) - Optional model with 18 VDC / 5'560 mA www.tracopower.com March 16, 2023 Page 1 / 5

Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 550 mA max. - Full Load & Vin = 115 VAC 1'150 mA max. Power Consumption - No load & Vin = 230 VAC 300 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 300 mW max. Input Inrush Current - At 230 VAC 60 A max. - At 115 VAC 35 A max. Power Factor - At 230 VAC 0.95 min. (Active Power Factor Correction) - At 115 VAC 0.95 min. (Active Power Factor Correction) Input Protection T 3.15 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Output Voltage Adjustment ±10% (By trim potentiometer) Output power must not exceed rated power! Voltage Set Accuracy ±1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.5% max. Ripple and Noise (20 MHz Bandwidth) 12 VDC model: 120 mVp-p typ. (w/ 10 µF X7R) 15 VDC model: 150 mVp-p typ. (w/ 10 µF X7R) 18 VDC model: 160 mVp-p typ. (w/ 10 µF X7R) 24 VDC model: 160 mVp-p typ. (w/ 1 µF X7R) 28 VDC model: 180 mVp-p typ. (w/ 1 µF X7R) 36 VDC model: 190 mVp-p typ. (w/ 1 µF X7R) 48 VDC model: 340 mVp-p typ. (w/ 0.1 µF X7R) Capacitive Load 12 VDC model: 6'950 µF max. 15 VDC model: 4'450 µF max. 18 VDC model: 3'100 µF max. 24 VDC model: 1'750 µF max. 28 VDC model: 1'280 µF max. 36 VDC model: 770 µF max. 48 VDC model: 430 µF max. Minimum Load Not required Temperature Coefficient ±0.02 %/K max. Hold-up Time - At 230 VAC 16 ms min. - At 115 VAC 16 ms min. Start-up Time - At 230 VAC 1'000 ms max. - At 115 VAC 1'000 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 115 - 150% of Iout max. Overvoltage Protection 115 - 135% of Vout nom. Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 500 µs typ. (50% to 75% Load Step) Output Specifications TPP 100A-J Series, 100 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 2 / 5

Safety Standards - IT / Multimedia Equipment EN 60950-1 EN 62368-1 IEC 60950-1 IEC 62368-1 UL 60950-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tpp100a-j Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) FCC Part 15 class B (internal filter) FCC Part 18 class B (internal filter) - Radiated Emissions EN 55011 class A (internal filter) EN 55032 class A (internal filter) FCC Part 15 class A (internal filter) FCC Part 18 class A (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A EN 61000-3-2, class D - Voltage Fluctuations & Flicker EN 61000-3-3 EMS Immunity EN 55024 (IT Equipment) EN 55035 (Multimedia) EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 20 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A L to PE: EN 61000-4-5, ±2 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 20 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 10 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B

115 VAC / 60 Hz: EN 61000-4-11

30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria B EMC Specifications Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –25°C to +85°C - Storage Temperature –40°C to +85°C General Specifications TPP 100A-J Series, 100 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 3 / 5

Power Derating - High Temperature 2 %/K above 55°C - Low Input Voltage 1.33 %/V below 100 VAC See application note: www.tracopower.com/overview/tpp100a-j Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 45 - 75 kHz (PWM QR) Insulation System Reinforced Insulation Working Voltage (rated) 250 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Case or PE, 60 s 1'500 VAC - Output to Case or PE, 60 s 1'500 VAC Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Leakage Current (at 264 VAC) - Touch Current 75 µA max. Reliability - Calculated MTBF 790'000 h (MIL-HDBK-217F, ground benign) Environment - Vibration IEC 60068-2-6 5 g, 3 axis, sine sweep, 5-500 Hz, 1 oct/min - Mechanical Shock IEC 60068-2-27 50 g, 3 axis, half sine, 11 ms Housing Type Open Frame Mounting Type Chassis Mount Connection Type Pin Connector Weight 156 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tpp100a-j TPP 100A-J Series, 100 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com March 16, 2023 Page 4 / 5

TPP 100A-J Series, 100 Watt Dimension in mm, () = inch Tolerances: x.x ±0.50 (±0.02) x.xx ±0.25 (±0.01) 76.2 (3.00) 69.8 (2.748) 44.45 (1.750) CON2 CON1 50.8 (2.00) Voltage adjustment TOP VIEW Ø 3.2 (0.13 1.6 (0.06) 29.5 (1.16) 2.5 (0.10) max. FRONT VIEW Pin connectors Input (CON1) Output (CON2) Pin Function Pin* Function

1 Line 1,2 –Vout

3 Neutral 3,4 +Vout

*Terminal rated for 7 A max. (at higher current connection has to be split) CON1: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-3N CON2: JST series mates with JST crimp terminal: SVH-21T-P1.1 and terminal housing: VHR-4N © Copyright 2023 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. March 16, 2023 Page 5 / 5