AC/DC Medical Power Supply TMW 24 Series, 24 Watt

  • Fully encapsulated, dust resistant, and waterproof (IP68 certified)
  • Operating temperature range –20° to +80°C
  • Mountable in wall outlet boxes, with flying leads
  • Screw mountable on wood and inside furniture
  • Safety approval IEC/EN 62368-1 and IEC/EN 60335-1
  • 2xMOPP certified according to IEC/EN 60601-1 3rd edition
  • Leakage current < 100 µA for body-floating (BF) applications
  • I/O isolation 4000 VAC (reinforced)
  • Short-circuit protection (SCP), overvoltage protection (OVP) and overtemperature protection (OTP)
  • 5-year product warranty IEC 60601-1 IEC 60335-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TMW 24 is a series of fully encapsulated AC/DC power supplies designed for medical, household, and industrial applications. All models come in com- pact, dust resistant, and waterproof IP68 housings, are IEC/EN 62368-1, IEC/- EN 60335-1 and IEC/EN 60601-1 3rd edition compliant and are certified for 2xMOPP. Featuring a reinforced isolation, the TMW 24 power supplies are prepared for protection class II applications. Their wide operating temperature range from -20°C to +80°C makes them ideal candidates for demanding appli- cations. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. max. typ. TMW 24-105 20.4 W 5.1 VDC 4 A 85 % TMW 24-112 24 W 12 VDC 2 A 88 % TMW 24-124 24 VDC 1 A 89 % www.tracopower.com November 30, 2022 Page 1 / 5

Input Voltage Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) Input Frequency Operational Range: 47 - 63 Hz Certified: 50/60 Hz Power Consumption - No load & Vin = 230 VAC 200 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 200 mW max. Input Inrush Current - At 230 VAC 50 A max. - At 115 VAC 25 A max. Power Factor - At 230 VAC 0.45 min. - At 115 VAC 0.55 min. Input Protection T 2 A (Internal Fuse) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Voltage Set Accuracy ±2% max. Regulation - Input Variation (Vmin - Vmax) 0.5% max. - Load Variation (10 - 90%) 1% max. (12 Vout and 24 Vout model) 2% max. (5 Vout model) Ripple and Noise (20 MHz Bandwidth) 140 mVp-p max. Capacitive Load 10'000 µF max. Minimum Load Not required Temperature Coefficient ±2 %/K max. Hold-up Time - At 230 VAC 40 ms min. - At 115 VAC 7 ms min. Start-up Time - At 230 VAC 500 ms max. - At 115 VAC 500 ms max. Start-up Overshoot Voltage 5% max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 105 - 150% of Iout max. Overvoltage Protection 130% typ. of Vout nom. 120 - 160% of Vout nom. Transient Response - Response Deviation 5% max. (10% to 90% Load Step) - Response Time 2'000 µs max. (10% to 90% Load Step) Output Specifications TMW 24 Series, 24 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 2 / 5

Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Household EN 60335-1 IEC 60335-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 CSA-C22.2, No 60601-1 2 x MOPP (Means Of Patient Protection) - LED Modules Controlgear EN 61347-2-13 - Luminaires for Furniture VDE 0710-14 - Power Transformers EN 61558-1 IEC 61558-1 EN 61558-2-8 IEC 61558-2-8 EN 61558-2-16 IEC 61558-2-16 - Certification Documents www.tracopower.com/overview/tmw24 Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 61000-6-3 (Generic Residential) EN 61204-3 (Low Voltage Power Supplies) - Conducted Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EN 55032 class B (internal filter) - Harmonic Current Emissions EN 61000-3-2, class A EMS Immunity EN 61000-6-2 (Generic Industrial) EN 60601-1-2 edition 4 (Medical Devices) - RF Electromagnetic Field EN 61000-4-3, 10 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 10 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 20%, 250 periods, perf. criteria A 30%, 25 periods, perf. criteria A 60%, 10 periods, perf. criteria A >95%, 5 periods, perf. criteria B >95%, 250 periods, perf. criteria B

115 VAC / 60 Hz: EN 61000-4-11

20%, 250 periods, perf. criteria A 30%, 25 periods, perf. criteria B 60%, 10 periods, perf. criteria B >95%, 5 periods, perf. criteria B >95%, 250 periods, perf. criteria B EMC Specifications Relative Humidity 100% max. Temperature Ranges - Operating Temperature –20°C to +80°C - Storage Temperature –40°C to +90°C General Specifications TMW 24 Series, 24 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 3 / 5

Power Derating - High Temperature 2 %/K above 50°C - Low Input Voltage 2 %/V below 100 VAC Over Temperature Protection Switch Off - Protection Mode 85°C min. / 90°C typ. / 95°C max. (Automatic recovery) - Measurement Point Case Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. (4'000 m max. for IEC 61558-2-8) Switching Frequency 25 - 140 kHz (PWM QR) Insulation System Reinforced Insulation Isolation Test Voltage - Input to Output, 60 s 4'000 VAC - Input to Output, 1 s 4'000 VAC Creepage - Input to Output 8 mm min. Clearance - Input to Output 8 mm min. Isolation Resistance - Input to Output, 500 VDC 100 MΩ min. Isolation Capacitance - Input to Output, 100 kHz, 1 V 1'100 pF typ. / 1'300 pF max. Leakage Current (at 240 VAC / 60 Hz) - Touch Current 100 µA max. Distance Through Isolation 2 mm Reliability - Calculated MTBF 3'400'000 h (see application note) (IEC 61709) Environment - Vibration IEC 60068-2-6 3 g, 3 axis, sine sweep, 10-55 Hz, 1 oct/min - Mechanical Shock IEC 60068-2-27 25 g, 3 axis, half sine, 11 ms Case Ingress Protection IP 68 (acc. IEC 60529) Housing Material Plastic (UL 94 V-1 rated) Potting Material Polyurethane (UL 94 V-2 rated) Housing Type Plastic Case Mounting Type Chassis Mount Connection Type Flying Leads Weight 131 g Power Back Immunity 5.1 VDC model: 6.4 V max. 12 VDC model: 15 V max. 24 VDC model: 30 V max. Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/tmw24 TMW 24 Series, 24 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 4 / 5

TMW 24 Series, 24 Watt 53.0 (2.087) 51.0 (2.008) All dimensions in mm (inch) Tolerances ±0.2 (±0.008) unless stated otherwise 200.0 +20.0/-0 7.874 (+0.787/-0) 200.0 +20.0/-0 7.874 (+0.787/-0) 5.0 +2.0/-0 (0.197 +0.787/-0) 5.0 +2.0/-0 (0.197 +0.787/-0) Tc TOP VIEW BOTTOM VIEW SIDE VIEW 30.0 (1.181) 24.5 (0.965) 12.0 (0.472) 22.2 (0.874) 22.2 (0.874) 13.1 (0.516) 3.0 (0.118) 13.1 (0.516) 1 2 4 3 Ø 3.2 (Ø 0.126) Connection Details Wire Function Color Type 1 AC (N) Blue 20AWG / 0.52mm 2 2 AC (L) Brown 20AWG / 0.52mm 2 3 - Vout Black 20AWG / 0.52mm 2 4 + Vout Red 20AWG / 0.52mm 2 © Copyright 2022 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. November 30, 2022 Page 5 / 5