TMF05 TRACOPOWER | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 4

Technical content

AC/DC Medical Power Supply TMF 05 Series, 5 Watt

  • Fully encapsulated power supplies in plasic casing for PCB mount
  • Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP
  • Risk management process according to ISO 14971 incl. risk management file
  • Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3
  • Low leakage current <100 μA rated for BF applications
  • Operating temperature range: –25°C to +70°C max.
  • Protection against short-circuit, over load and over voltage
  • Protection class II prepared
  • 5-year product warranty IEC 60601-1ES 60601-1 The TMF 05 Series AC/DC power supply modules are designed and manufac- tured based on workmanship standards and risk management to comply with the requirements for quality, reliability and safety of medical equipment. The units are approved to IEC/EN/ES 60601-1 edition 3.1 for 2 x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk management file. These fully encapsulated modules are for PCB mount. They are designed for protection class II applications (no earth connection) and feature a low leakage current (<100 μA). A compact design and excellent EMC considerations facili- tate the design in. The thermal management enables an operation within a wide temperature range of -25 to +70°C and the isolation system is designed and approved for an altitude of 5000 m (AMSL). This makes the power supplies suitable not only for stationary applications but also for transportable medical equipment. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. max. typ. TMF 05105 5 W

5 VDC 1'000 mA 77 %

TMF 05112 12 VDC 417 mA 82 % TMF 05115 15 VDC 333 mA 82 % TMF 05124 24 VDC 208 mA 82 % www.tracopower.com January 25, 2023 Page 1 / 4

Input Voltage - AC Range Operational Range: 90 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / –DC: N Input Frequency Operational Range: 47 - 440 Hz Certified: 50/60 Hz Input Current - Full Load & Vin = 230 VAC 80 mA max. - Full Load & Vin = 115 VAC 150 mA max. Power Consumption - No load & Vin = 230 VAC 100 mW max. (Ready to meet ErP directive) - No load & Vin = 115 VAC 100 mW max. Input Inrush Current - At 230 VAC 20 A max. - At 115 VAC 10 A max. Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Input Specifications Voltage Set Accuracy ±2% max. Regulation - Input Variation (Vmin - Vmax) 0.5% max. - Load Variation (0 - 100%) 1% max. Ripple and Noise (20 MHz Bandwidth) 5 VDC model: 100 mVp-p max. 12 VDC model: 120 mVp-p max. 15 VDC model: 150 mVp-p max. 24 VDC model: 240 mVp-p max. Capacitive Load 5 VDC model: 8'000 µF max. 12 VDC model: 1'200 µF max. 15 VDC model: 820 µF max. 24 VDC model: 330 µF max. Minimum Load Not required Temperature Coefficient ±0.05 %/K max. Hold-up Time - At 230 VAC 90 ms min. - At 115 VAC 20 ms min. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 160 - 240% of Iout max. Overvoltage Protection 105 - 145% of Vout nom. (By Zener diode) Output Specifications Safety Standards - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 CSA-C22.2, No 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/tmf05 Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class B (internal filter) - Radiated Emissions EN 55011 class B (internal filter) EMC Specifications TMF 05 Series, 5 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com January 25, 2023 Page 2 / 4

EMS Immunity EN 61000-6-2 (Generic Industrial) EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 3 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A L to L: EN 61000-4-5, ±1 kV, perf. criteria A L to PE: EN 61000-4-5, ±2 kV, perf. criteria A - Conducted RF Disturbances EN 61000-4-6, 3 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 30 A/m, perf. criteria A - Voltage Dips & Interruptions 230 VAC / 50 Hz: EN 61000-4-11 30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria A

115 VAC / 60 Hz: EN 61000-4-11

30%, 25 periods, perf. criteria A >95%, 0.5 periods, perf. criteria A >95%, 1 period, perf. criteria A >95%, 250 periods, perf. criteria A Relative Humidity 95% max. (non condensing) Temperature Ranges - Operating Temperature –25°C to +70°C - Storage Temperature –40°C to +85°C Power Derating - High Temperature 4 %/K above 55°C - Low Input Voltage 2 %/V below 100 VAC See application note: www.tracopower.com/overview/tmf05 Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 40 - 140 kHz (PWM) 66 kHz typ. (PWM) Insulation System Reinforced Insulation Working Voltage (rated) 250 VAC Isolation Test Voltage - Input to Output, 60 s 4'000 VAC Leakage Current - Touch Current 100 µA max. Reliability - Calculated MTBF 450'000 h (MIL-HDBK-217F, ground benign) Washing Process Not allowed Housing Material Plastic resin (UL 94 V-0 rated) Potting Material Silicone (UL 94 V-0 rated) (Hermetical sealed structure, dust-proof only non water-proof) Pin Material Brass Pin Surface Plating Tin (120 µm min.), matte Housing Type Plastic Case Mounting Type PCB Mount Connection Type THD (Through-Hole Device) Weight 30 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) General Specifications TMF 05 Series, 5 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com January 25, 2023 Page 3 / 4

Overview Link (for additional Documents) www.tracopower.com/overview/tmf05 Outline Dimensions TMF 05 Series, 5 Watt Dimensions in mm Tolerances ±0.5 mm Pin tolerances ±0.1 mm Ø1.0 19.1 6.0 41.2 top view 12.7 4.1 27.2 19.0 3.4 Pinout Pin Function

1 AC (L)

2 AC (N)

3 –Vout 4 +Vout © Copyright 2023 Traco Electronic AG Specifications can be changed without notice. www.tracopower.com Rev. January 25, 2023 Page 4 / 4