www.eravant.com | 424-757-0168 | support@eravant.com Copyright © 2024 by Eravant Advanced Rev 1.1 Coaxial VNA Frequency Extender TX/RX Module, 5 to

13.75 GHz Input

STO-20355315-C-E1 is a coaxial vector network analyzer (VNA) frequency extender Tx/Rx designed to achieve full 1-port, S-parameter testing at 20 to 55 GHz. It is compatible with modern vector network analyzers such as the Copper Mountain CobaltFx. The VNA needs dual sources to be extended. The frequency extender can achieve a dynamic range up to 120 dB for certain passive products that require high rejection, isolation, and return loss testing such as directional couplers and filters. An AC to DC Power adapter and two Proxi-Flange™ Contactless Flanges (STQ-WG-19010-FB-CF and STQ-WG-19025- FB-CF), are included. The Eravant calibration kit (STQ-TO-VFVM-U3- CKIT1) is highly recommended to complete the coaxial VNA test set. VNA extender is packaged individually in a rugged equipment box with additional hardware and tools. Electrical Specifications: Parameter Minimum Typical Maximum Frequency Range 20 GHz 55 GHz Test Port Output Power (No Attenuation) +15 dBm Test Port Input Power (Damage) +30 dBm Dynamic Range @ 10 Hz BW 100 dB 120 dB Test Port Match 10 dB Directivity 10 dB RF Source Input Frequency 5 GHz 13.75 GHz RF Source Input Power -6 dBm -3 dBm +0 dBm LO Source Input Frequency (RF±IF) 5 GHz 13.75 GHz LO Source Input Power -3 dBm 0 dBm +3 dBm IF Frequency Range 10 MHz 1000 MHz Multiplication Factor 4 Magnitude Stability @ 300 Hz BW ±0.1 dB Phase Stability @ 300 Hz BW ±2.5° Specification Temperature +20°C +30°C Operating Temperature 0°C +50°C ECCN EAR99

FEATURES

  • Dynamic Range of 120 dB
  • AC Power Input: 100 to 240 VAC

APPLICATIONS

  • VNA Frequency Extension
  • S-Parameter Characterization
  • Test Lab Instrumentation Recommended Pairing
  • Cal Kit: STQ-TO-VFVM-U3-CKIT1
  • Waveguide-Glide™ Rail System
  • Cable: SCW-SMSM040-F1-A-PM Recommended Pairing
  • VNA Extender Configuration Guide
  • VNA Extenders and Cal Kits

www.eravant.com | 424-757-0168 | support@eravant.com Copyright © 2024 by Eravant Advanced Rev 1.1 Mechanical Specifications: Included Components: Simplified Block Diagram: Item Specification Test Port 1.85mm (F) RF and LO Source Input Port SMA (F), SMA (F) IF Output Port SMA (F) IF Reference Port SMA (F) DC Power Receptacle LEMO EGG.0B.304.CLL Finish Black Anodized Weight (Per Module) 4.4 lbs. Size (Without Adjustable Feet) 9.00” (L) x 3.00” (W) x 1.90” (H) Outline TO-SC Item Eravant Model Number Quantity Proxi-Flange™ Contactless Flange, 1.0” Long STQ-WG-19010-FB-CF 1 Proxi-Flange™ Contactless Flange, 2.5” Long STQ-WG-19025-FB-CF 1 Waveguide Screws, 3/32 Hex Head SWH-332-SS-10 1 (10 Screws Total) Waveguide Screwdriver, 3/32 Hex Head SWH-332-DS 1 SMA Connector Torque Wrench SCH-08008-S1 1 AC-to-DC Power Adapter STU-110006005-HF 1 Connecting Cables are not included. Eravant coaxial cable, model SCW-SMSM040-F1-A-PM, is highly recommended. Four (4) cables are required to connect this module with VNA

www.eravant.com | 424-757-0168 | support@eravant.com Copyright © 2024 by Eravant Advanced Rev 1.1 Mechanical Outline: (Unless otherwise specified, all dimensions are in inches [millimeters]) NOTE:

  • Only one extender module is included in STO-20355315-C-E1: Tx/Rx module. A pair of extenders is offered under a different model number and can be found on our VNA Frequency Extenders page.
  • Eravant reserves the right to change the information presented without notice. CAUTION:
  • Exceeding absolute maximum ratings shown will damage the extenders.
  • If a waveguide is present, any foreign objects in the waveguide will cause performance degradation and may damage or destroy the unit. SCH-06004-S1 is highly recommended. (0.90 ± 0.02 Nm). Torque wrench model SCH-08008-S1 is highly recommended.

www.eravant.com | 424-757-0168 | support@eravant.com Copyright © 2024 by Eravant Advanced Rev 1.1 Appendix: Case View with Included Components