FS9168-001 FORTUNE | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 18

Technical content

Datasheet sections

  • 11.1 Power Switch
  • 11.2 Operation Procedure
  • 11.3 Power off
  • 11.4 Buzzer frequency
  • 11.5 Measurement Resolution
  • 11.6 Sensor Type
  • 11.7 Resistance reference value
  • 11.8 Others
  • 13.1 Definition
  • 13.2 Linearity Error Table

REV. 1.1 FS9168-001-DS-11_EN MAY 2014 Datasheet FS9168-001 FS9168 standard code for digital clinical thermometer application FORTUNE' Properties For Reference Only

Fortune Semiconductor Corporation 富晶電子股份有限公司 Danshui Town, Taipei County 251, Taiwan Tel.:886-2-28094742 Fax:886-2-28094874 www.ic-fortune.com This manual contains new product information. Fortune Semiconductor Corporation reserves the rights to modify the product specification without further notice. No liability is assumed by Fortune Semiconductor Corporation as a result of the use of this product. No rights under any patent accompany the sale of the product. FORTUNE' Properties For Reference Only

  1. General Description The FS9168-001 is a digital clinical thermometer IC for body temperature measurement. It provides fever alarm and auto power off functions. LCD driver and voltage doubler are inlcuded for LCD display. The other functions are one-key on/off switch, memory function, buzzer frequency slection and embedded low battery detection. 2. Features  1.5V battery operation; operation current is less than 100 μA, and sleep mode current is less than 1μA.  Centigrade/Fahrenheit measurement selection.  Alarm warning for fever.  Highest temperature hold.  Auto power off after 5 min.  Memory function for last time measurement.  Buzzer output frequency selection.  Embedded low battery detection, 1.35V 0.05 V. 3. Ordering Information Product Number Package Type FS9168-001 Dice form of 36 pin, or 44 pin LQFP FORTUNE' Properties For Reference Only
  1. Pin Configuration VDD 1110 29 28 27 SEG12 SEG6 SE G13 C512 R P RN COM2 SEG2 SEG3 31 30 SEG4 SEG5 PT1.1 CAP VEE PT1.7 PT1.6 RST R S COM3 SEG1 PT1.5 PT1.4 PT1.3 PT1.0 SC 4 87 RF VSS TST FS9168 SEG7 SEG9 SEG10 SEG11 PT1.2 NC NC NC COM1 NC NC33 NC 23 SEG8 NC NC 5. Pin Description Name In/Out Pad NO. Description VDD I 3 Positive input of power supply (1.5V) VSS I 5 Negative input of power supply TST I 6 Test pin for IC RP I/O 7 System oscillator external resistor connection (450k) RN I/O 8 System oscillator external resistor connection (450k) VEE I/O 27 Voltage doubler output (+3.0V) CAP I/O 26 Voltage doubler capacitor positive terminal C512 I/O 25 Voltage doubler capacitor negative terminal COM1~3 O 9~11 LCD common driver SEG1~13 O 12~24 LCD segment driver RF I 1 Reference resistor connection RS I 2 Sensor resistor connection SC I 36 Comparator input P1.0~P1.3 I 32~35 Input Ports P1.4~P1.7 I/O 28~31 In/Out Ports RST I 4 CPU Reset Pin FORTUNE' Properties For Reference Only
  1. Functional Block Diagram System Oscillator RP VDD VSS R/F Conversion RF RS Comparator and 16 bits Counter SC 8 bits FSC CPU 1024Word Program Memory

64 Byte

COM1~3 SEG1~13 TST Timer 4 bits Input Port 4 bits In/Out Port P1.0~P1.3 Buzzer Output Port P1.4~P1.7 Low Battery Detector RN 7. Typical Application Circuit TSTSC RF RS 2200pF 30k VSS VDD 1uF CAP C512 0.01uF RP RN 450k LCD TEST PT1.1 PT1.2 BS ML PT1.3 BUZZER1 SW PT1.0 PT1.4 PT1.5 FEVER C/F BUZZER2 VEE 0.1uF BATTERY  SW (PT1.0): Power on/off switch  TEST (PT1.1): Resistance Reference Test. Close the switch during test  BS (PT1.2): Buzzer frequency select ( Close is 4KHz , Open is 8 KHz, When RP ,RN is 450KΩ)  ML ( PT1.3): Memory function select (Close Select Memory, Open No Memory)  FEVER (PT1.4): Fever warning function select (Close Select FEVER warning function, Open “one Beep” When Measure stable) FORTUNE' Properties For Reference Only

 C/F (PT1.5) : ℃ and ℉ unit select (Default select when battery install)  BUZZER1 and BUZZER2 ( PT1.6 , PT1.7): Buzzer frequency out 8. Absolute Maximum Ratings Parameter Rating Unit Supply Voltage to Ground Potential -0.3 to 1.65 V Applied Input/Output Voltage -0.3 to VDD+0.15 V Ambient Operating Temperature 0 to +70 C Storage Temperature -55 to +150 C Soldering Temperature, Time 260C, 10 Sec 9. Electrical Characteristics Symbol Parameter Test Condition Min Typ Max Units VDD Recommended Operation Voltage 1.2 1.5 1.65 V IDD Supply Current CPU, R2F ADC On 40 60 A IPO Power off Current At Sleep mode 0.1 0.2 1.0 A FOSC System Clock ROSC=350k, VDD=1.5V 25.6 32 38.4 KHz RNE Resistance Non-Linearity Error Rin=22k~38k 0.2 % 10. LCD Pin Assignment LCD PIN 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 SEG1 SEG2 SEG3 SEG4 SEG5 SEG6 SEG7 SEG8 SEG9 SEG10 SEG11 SEG12 SEG13 COM1 COM1 A1 C1 F1 A2 C2 F2 A3 C3 F3 A4 C4 F4 A5 COM2 COM2 B1 D1 G1 B2 D2 G2 B3 D3 G3 B4 D4 G4 B5 COM3 COM3 H1 E1 E2 H2 E3 I4 E4 H4 C5 I4 H4 FORTUNE' Properties For Reference Only

  1. Function Description Power Switch Press to turn on or turn off the power. Operation Procedure Turn on the power: When press the switch, the system will sound one “beep”. a. At first the system will display all LCD signals about 2 seconds. b. Following step a, select unit status ℃ or ℉. c. Following step b, it depends on the memory function. If the memory function exists, then jump to step if no memory function, the system will display 36.50℃ instead of the memory symbol, then jump to step i. d. Following step c, if the memorized temperature is between 32 ℃~43℃, the system will display the last time measured temperature value about 3 seconds and the memory symbol; if it is out of the range (<32℃ or >43℃), the system will display 36.50℃ and not display the memory symbol. e. Test the resistance reference to see whether it is normal or not. If it is normal, the system will display 37.00℃ about 1 second, then jump to step f; if it is abnormal, the system will display Err and recycle in step e. f. Display the highest measured temperature value. In the meantime, ℃ signal will flash with frequency about 1Hz. g. If the temperature is lower than 32.00 ℃, the system displays Lo ℃. ℃, the system displays Hi ℃. i. If the temperature is between 32.00 ℃ ~ 43.00℃, the system will display the highest measured temperature value. j. If the temperature keeps unchanged over 8 seconds, the temperature has reached the desired stable reading. The measurement is over and ℃ signal will stop flashing. k. If fever warning is enabled, jump to step l; if not , the system will beep one sound and jump to step m. l. When the temperature is higher than 37.50 ℃, the beeper will sound “beep-beep-beep---beep-beep-beep---” about 4 seconds. The sound would be as follow: Beep(0.125 seconds ) ―― (0.125 seconds ) Beep(0.125 seconds ) ―― (0.125 seconds ) Beep(0.125 seconds ) ―――― (0.375 seconds ) 4 times When the temperature is lower than or equal to 37.50℃, the beeper will sound “beep-beep-beep-beep-” about 4 seconds. The sound would be as follow: ―― (0.5 second ) Beep(0.5 second )) m. After measurement is over, if the temperature keeps unchanged for 5 minutes, the system will beep one sound and turn off automatically. n. If the temperature changes (larger than or equal to 0.05 ℃) within 5 minutes, ℃ signal will flash again (repeat step f), then start to recount 5 minutes. o. When beeper sounds (within 4 seconds proximately), the system will not measure temperature. Power off The system will enter sleep mode. The current consumption is 1μA. Buzzer frequency The Buzzer frequency can be selected with external pin to be 4kHz or 8kHz. Measurement Resolution The smallest measurement resolution for ℃ is 0.01. Sensor Type The sensor type uses 503ET. FORTUNE' Properties For Reference Only

Resistance reference value Resistance reference value is the thermistor resistance at 37.00 ℃. Others ” flag won’t display.  During production in the factories, TEST pin is shorted for adjusting reference resistance value (R F). In this situation, the temperature reading is the temperature of the test environment (such as the temperature tank). The display will jump up or down, and won’t keep the highest value. FORTUNE' Properties For Reference Only

  1. Program Flow Chart Power on (Last power-off status) All LCD turn on 2 sec Measurement Display highest Temp.Display “Lo” Display “Hi” > 43℃< 32℃ Temp. no change within 6 secs No ℃ stop glitter. IF fever alarm is on, 1. <=37.5℃ buzzer: ---Bi–Bi 2. >37.5℃ buzzer: ---BIBIBiBi). IF fever alarm is off, buzzer: -Bi. Temp. change within 5 mins? Yes Power off Temperature check (<43℃ or > 32℃) Yes Display last Temp. 3 secs. Show memory mark. Display “36.5℃”3 secs No show memory mark. No Temp. now > highest temp. Idle more then 5 mins Memory? Yes No SENSOR Calibrate? Display “37.00±0.01℃” 1 sec Yes Display “Err” No Display mode (℃ or ℉) Push button within 3 sec ℃ , ℉ Switch Push button more than 2 secYes No Yes No Battery install Power on (Initial is for ℃ measurement) FORTUNE' Properties For Reference Only
  1. FS9168-001 R2F Converter Linearity Measured Data Definition Reference resistance R f is converted to frequency value CF rf. Thermistor resistance Rs is converted to frequency value CF rs. Temp = ( CF rs / CFrf ) * T a. Rf is the resistance value when the thermistor is at 37 ℃. Ta=37℃ Linearity Error Table  Calculate the linearity error of resistance count readings in temperature range 32 ℃~43℃ Linearity Slope Points: Start 26300Ω; End 36900Ω Reference Resistance RF = 30974Ω Count = (Count(RS) / Count(RF) ) * 3700 Actual Linearity = 10000000/Count Slope δ = (31918 ﹣23159) ÷ (36900 ﹣26300) Ideal Linearity Value = 23159 + δ×【(RS(N)-RS(N-1))】 Linearity Error = Ideal Linearity Value - Actual Linearity Error Percentage = (Linearity Error ÷ Ideal Linearity Value) × 100﹪ Power Source: BATTERY Measured Resistance Value RS Count Actual Linearity Ideal Linearity Value Linearity Error Error Percentage 26300 4318 23159 23159 26400 4302 23245 86 23242 -3 -0.01% 26500 4288 23321 76 23324 3 0.01% 26600 4272 23408 87 23407 -1 0% 26700 4257 23491 83 23490 -1 0% 26800 4242 23574 83 23572 -2 -0.01% 26900 4227 23657 83 23655 -2 -0.01% 27000 4213 23736 79 23737 1 0% 27100 4198 23821 85 23820 -1 0% 27200 4184 23901 80 23903 2 0.01% 27300 4169 23987 86 23985 -2 -0.01% 27400 4154 24073 86 24068 -5 -0.02% 27500 4140 24155 82 24151 -4 -0.02% 27600 4126 24237 82 24233 -4 -0.02% 27700 4112 24319 82 24316 -3 -0.01% 27800 4098 24402 83 24398 -4 -0.02% 27900 4084 24486 84 24481 -5 -0.02% 28000 4070 24570 84 24564 -6 -0.02% 28100 4056 24655 85 24646 -9 -0.04% 28200 4043 24734 79 24729 -5 -0.02% 28300 4029 24820 86 24812 -8 -0.03% 28400 4016 24900 80 24894 -6 -0.02% 28500 4002 24988 88 24977 -11 -0.04% 28600 3989 25069 81 25060 -9 -0.04% 28700 3976 25151 82 25142 -9 -0.04% 28800 3963 25233 82 25225 -8 -0.03% 28900 3950 25316 83 25307 -9 -0.04% 29000 3937 25400 84 25390 -10 -0.04% 29100 3924 25484 84 25473 -11 -0.04% 29200 3912 25562 78 25555 -7 -0.03% 29300 3899 25648 86 25638 -10 -0.04% 29400 3887 25727 79 25721 -6 -0.02% 29500 3875 25806 79 25803 -3 -0.01% 29600 3862 25893 87 25886 -7 -0.03% 29700 3850 25974 81 25968 -6 -0.02% 29800 3837 26062 88 26051 -11 -0.04% 29900 3825 26144 82 26134 -10 -0.04% 30000 3813 26226 82 26216 -10 -0.04% 30100 3801 26309 83 26299 -10 -0.04% FORTUNE' Properties For Reference Only

30200 3789 26392 83 26382 -10 -0.04% 30300 3778 26469 77 26464 -5 -0.02% 30400 3766 26553 84 26547 -6 -0.02% 30500 3753 26645 92 26630 -15 -0.06% 30600 3742 26724 79 26712 -12 -0.04% 30700 3731 26802 78 26795 -7 -0.03% 30800 3719 26889 87 26877 -12 -0.04% 30900 3707 26976 87 26960 -16 -0.06% 31000 3696 27056 80 27043 -13 -0.05% 31100 3685 27137 81 27125 -12 -0.04% 31200 3674 27218 81 27208 -10 -0.04% 31300 3663 27300 82 27291 -9 -0.03% 31400 3651 27390 90 27373 -17 -0.06% 31500 3641 27465 75 27456 -9 -0.03% 31600 3630 27548 83 27539 -9 -0.03% 31700 3619 27632 84 27621 -11 -0.04% 31800 3608 27716 84 27704 -12 -0.04% 31900 3597 27801 85 27786 -15 -0.05% 32000 3587 27878 77 27869 -9 -0.03% 32100 3576 27964 86 27952 -12 -0.04% 32200 3565 28050 86 28034 -16 -0.06% 32300 3555 28129 79 28117 -12 -0.04% 32400 3544 28217 88 28200 -17 -0.06% 32500 3534 28297 80 28282 -15 -0.05% 32600 3524 28377 80 28365 -12 -0.04% 32700 3513 28466 89 28447 -19 -0.07% 32800 3503 28547 81 28530 -17 -0.06% 32900 3493 28629 82 28613 -16 -0.06% 33000 3484 28703 74 28695 -8 -0.03% 33100 3474 28785 82 28778 -7 -0.02% 33200 3463 28877 92 28861 -16 -0.06% 33300 3454 28952 75 28943 -9 -0.03% 33400 3443 29044 92 29026 -18 -0.06% 33500 3434 29121 77 29109 -12 -0.04% 33600 3426 29189 68 29191 2 0.01% 33700 3416 29274 85 29274 0 0% 33800 3406 29360 86 29356 -4 -0.01% 33900 3397 29438 78 29439 1 0% 34000 3387 29525 87 29522 -3 -0.01% 34100 3377 29612 87 29604 -8 -0.03% 34200 3367 29700 88 29687 -13 -0.04% 34300 3358 29780 80 29770 -10 -0.03% 34400 3348 29869 89 29852 -17 -0.06% 34500 3340 29940 71 29935 -5 -0.02% 34600 3331 30021 81 30017 -4 -0.01% 34700 3322 30102 81 30100 -2 -0.01% 34800 3312 30193 91 30183 -10 -0.03% 34900 3303 30276 83 30265 -11 -0.04% 35000 3295 30349 73 30348 -1 0% 35100 3286 30432 83 30431 -1 0% 35200 3277 30516 84 30513 -3 -0.01% 35300 3268 30600 84 30596 -4 -0.01% 35400 3259 30684 84 30679 -5 -0.02% 35500 3251 30760 76 30761 1 0% 35600 3242 30845 85 30844 -1 0% 35700 3234 30921 76 30926 5 0.02% 35800 3225 31008 87 31009 1 0% 35900 3217 31085 77 31092 7 0.02% 36000 3208 31172 87 31174 2 0.01% 36100 3199 31260 88 31257 -3 -0.01% 36200 3191 31338 78 31340 2 0.01% 36300 3182 31427 89 31422 -5 -0.02% 36400 3174 31506 79 31505 -1 0% 36500 3166 31586 80 31587 1 0% FORTUNE' Properties For Reference Only

36600 3158 31666 80 31670 4 0.01% 36700 3149 31756 90 31753 -3 -0.01% 36800 3141 31837 81 31835 -2 -0.01% 36900 3133 31918 81 31918 0 0%  Calculate the linearity error of resistance count readings in temperature range 32 ℃~37℃ Linearity Slope Points: Start 26300Ω; End 30800Ω Reference Resistance RF = 30974Ω Count = (Count(RS) / Count(RF) ) * 3700 Actual Linearity = 10000000/Count Slope δ = (26976 ﹣23159) ÷ (30900 ﹣26300) Ideal Linearity Value = 23159 + δ×【(RS(N)-RS(N-1))】 Linearity Error = Ideal Linearity Value - Actual Linearity Error Percentage = (Linearity Error ÷ Ideal Linearity Value) × 100﹪ Power Source: BATTERY Measured Resistance Value RS Count Actual Linearity Ideal Linearity Value Linearity Error Error Percentage 26300 4318 23159 23159 26400 4302 23245 23242 -3 -0.01% 26500 4288 23321 23325 4 0.02% 26600 4272 23408 23408 0 0% 26700 4257 23491 23491 0 0% 26800 4242 23574 23574 0 0% 26900 4227 23657 23657 0 0% 27000 4213 23736 23740 4 0.02% 27100 4198 23821 23823 2 0.01% 27200 4184 23901 23906 5 0.02% 27300 4169 23987 23989 2 0.01% 27400 4154 24073 24072 -1 0% 27500 4140 24155 24155 0 0% 27600 4126 24237 24238 1 0% 27700 4112 24319 24321 2 0.01% 27800 4098 24402 24404 2 0.01% 27900 4084 24486 24487 1 0% 28000 4070 24570 24570 0 0% 28100 4056 24655 24653 -2 -0.01% 28200 4043 24734 24736 2 0.01% 28300 4029 24820 24819 -1 0% 28400 4016 24900 24902 2 0.01% 28500 4002 24988 24985 -3 -0.01% 28600 3989 25069 25068 -1 0% 28700 3976 25151 25150 -1 0% 28800 3963 25233 25233 0 0% 28900 3950 25316 25316 0 0% 29000 3937 25400 25399 -1 0% 29100 3924 25484 25482 -2 -0.01% 29200 3912 25562 25565 3 0.01% 29300 3899 25648 25648 0 0% 29400 3887 25727 25731 4 0.02% 29500 3875 25806 25814 8 0.03% 29600 3862 25893 25897 4 0.02% 29700 3850 25974 25980 6 0.02% 29800 3837 26062 26063 1 0% 29900 3825 26144 26146 2 0.01% 30000 3813 26226 26229 3 0.01% 30100 3801 26309 26312 3 0.01% 30200 3789 26392 26395 3 0.01% 30300 3778 26469 26478 9 0.03% 30400 3766 26553 26561 8 0.03% 30500 3753 26645 26644 -1 0% FORTUNE' Properties For Reference Only

30600 3742 26724 26727 3 0.01% 30700 3731 26802 26810 8 0.03% 30800 3719 26889 26893 4 0.01%  Calculate the linearity error of resistance count readings in temperature range 37 ℃~43℃ Linearity Slope Points: Start 30900Ω; End 36900Ω Resistance Reference RF = 30974Ω Count = (Count(RS) / Count(RF) ) * 3700 Actual Linearity = 10000000/Count Slope δ = (31918 - 23159) ÷ (36900 ﹣30900) Ideal Linearity Value = 23159 + δ×【(RS(N)-RS(N-1))】 Linearity Error = Ideal Linearity Value - Actual Linearity Error Percentage = (Linearity Error ÷ Ideal Linearity Value) × 100﹪ Power Source: BATTERY Measured Resistance Value RS Count Actual Linearity Ideal Linearity Value Linearity Error Error Percentage 30900 3707 26976 26976 0 0% 31000 3696 27056 27058 2 0.01% 31100 3685 27137 27141 4 0.01% 31200 3674 27218 27223 5 0.02% 31300 3663 27300 27305 5 0.02% 31400 3651 27390 27388 -2 -0.01% 31500 3641 27465 27470 5 0.02% 31600 3630 27548 27553 5 0.02% 31700 3619 27632 27635 3 0.01% 31800 3608 27716 27717 1 0% 31900 3597 27801 27800 -1 0% 32000 3587 27878 27882 4 0.01% 32100 3576 27964 27964 0 0% 32200 3565 28050 28047 -3 -0.01% 32300 3555 28129 28129 0 0% 32400 3544 28217 28212 -5 -0.02% 32500 3534 28297 28294 -3 -0.01% 32600 3524 28377 28376 -1 0% 32700 3513 28466 28459 -7 -0.02% 32800 3503 28547 28541 -6 -0.02% 32900 3493 28629 28623 -6 -0.02% 33000 3484 28703 28706 3 0.01% 33100 3474 28785 28788 3 0.01% 33200 3463 28877 28870 -7 -0.02% 33300 3454 28952 28953 1 0% 33400 3443 29044 29035 -9 -0.03% 33500 3434 29121 29118 -3 -0.01% 33600 3426 29189 29200 11 0.04% 33700 3416 29274 29282 8 0.03% 33800 3406 29360 29365 5 0.02% 33900 3397 29438 29447 9 0.03% 34000 3387 29525 29529 4 0.01% 34100 3377 29612 29612 0 0% 34200 3367 29700 29694 -6 -0.02% 34300 3358 29780 29776 -4 -0.01% 34400 3348 29869 29859 -10 -0.03% 34500 3340 29940 29941 1 0% 34600 3331 30021 30024 3 0.01% 34700 3322 30102 30106 4 0.01% 34800 3312 30193 30188 -5 -0.02% 34900 3303 30276 30271 -5 -0.02% FORTUNE' Properties For Reference Only

35000 3295 30349 30353 4 0.01% 35100 3286 30432 30435 3 0.01% 35200 3277 30516 30518 2 0.01% 35300 3268 30600 30600 0 0% 35400 3259 30684 30683 -1 0% 35500 3251 30760 30765 5 0.02% 35600 3242 30845 30847 2 0.01% 35700 3234 30921 30930 9 0.03% 35800 3225 31008 31012 4 0.01% 35900 3217 31085 31094 9 0.03% 36000 3208 31172 31177 5 0.02% 36100 3199 31260 31259 -1 0% 36200 3191 31338 31341 3 0.01% 36300 3182 31427 31424 -3 -0.01% 36400 3174 31506 31506 0 0% 36500 3166 31586 31589 3 0.01% 36600 3158 31666 31671 5 0.02% 36700 3149 31756 31753 -3 -0.01% 36800 3141 31837 31836 -1 0% 36900 3133 31918 31918 0 0% FORTUNE' Properties For Reference Only

  1. Package Outline FORTUNE' Properties For Reference Only

FORTUNE' Properties For Reference Only

  1. Pad Assignment (0,0) X Y 1 2 3 4 5 6 7 8 192021222368 24252627 Pad opening : 90um Substrate should be connected to VSS 16. Pad Coordinate ※PAD NO: 36 pads Pad No. Name X[um] Y[um] Pad No. Name X[um] Y[um]

1 RF 85 74 19 SEG<8> 1335 1406

2 RS 205 74 20 SEG<9> 1205 1406

3 VDD 329 74 21 SEG<10> 1095 1406

4 RST 449 74 22 SEG<11> 975 1406

5 VSS 569 74 23 SEG<12> 855 1406

6 TST 689 74 24 SEG<13> 738 1406

7 RP 809 74 25 C512 331 1406

8 RN 929 74 26 CAP 200 1406

9 COM<1> 1346 87 27 VEE 74 1406

10 COM<2> 1346 205 28 PT1<7> 74 1262

11 COM<3> 1346 324 29 PT1<6> 74 1142

12 SEG<1> 1346 441 30 PT1<5> 74 1021

13 SEG<2> 1346 561 31 PT1<4> 74 902

14 SEG<3> 1346 681 32 PT1<3> 74 772

15 SEG<4> 1346 802 33 PT1<2> 74 650

16 SEG<5> 1346 924 34 PT1<1> 74 531

17 SEG<6> 1346 1041 35 PT1<0> 74 411

18 SEG<7> 1346 1162 36 SC 74 291

FORTUNE' Properties For Reference Only