Data Sheet 902006 Page 1/12 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com www.i RTD Temperature Probe for Devices and Plants Tested/ Inspected According to DIN EN 14597 DIN EN 14597
- For heat tran smission plants according to DIN 4754 For temperature s of up to 700 °C A s single, double, or triple RTD temperature probes For water, oil, or air F o r t ested control and limitation devices The RTD temperature probes specified in this data sheet can be used in connection with tem- perature control and limitation devices tested/inspected according to DIN EN 14597 (DIN 3440) in heat transmission plants according to DIN 4754. RTD temperature probes are the preferred choice for measuring temperatures in liquids and gases. Application areas include air conditioning and refrigeration technology as well as heat- ing, furnace/kiln, and apparatus engineering. Form B terminal heads are suitable for ambient temperatures up to 100 °C. Protection tubes made of various materials protect the measuring insert against chemical influ- ences and mechanical damage. The appropriate protection tube material is selected according to the conditions on site. A Pt100 temperature sensor according to DIN EN 60751, class B is used in the measuring insert in a 2-wire-circuit as standard. Changes to the described versions require new design approval. Important information: state the part no. from price sheet 902006 when ordering!
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 2/12 Dimensions Pos. 1 Po s. 2 Po s. 3 Pos. 4 Po s. 5 Po s. 6
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 3/12 Push-in RTD temperature probe with pipe screw connection and terminal head, according to DIN 43729, form B, M 24 × 1.5 Insertion length EL in mm Nominal length NL in mm Thread G in inches Temperature range in °C Type up to February 28, 2013 1× Pt100 Type up to February 28, 2013 2× Pt100 Type as of March 01, 2013 Order code Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571 65 to 670 710 G 1/2 -170 to +550 90.272-F03 - 902006/54-227-1003-1-15-710-254/000 65 to 670 710 G 1/2 -170 to +550 - 90.272-F02 902006/54-227-2003-1-15-710-254/000 Important information: Please observe the internal lead wire resistance with respect to the operating temperature; see the final page for the resistance values. Operating media: water and oil Operating pressure: max. 27 bar up to 100 °C, 20 bar up to 400 °C, pressureless as of 400 °C Pos. 1 G NL EL Versions only for connecting a JUMO safetyM STB/STW Ex type 701155 Insertion length EL in mm Nominal length NL in mm Thread G in inches Temperature range in °C 1× Pt100 2× Pt100 Type Order code Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571 65 to 670 710 G 1/2 -170 to +550 x - 902006/54-227-1003-1-15-710-254/922 65 to 670 710 G 1/2 -170 to +550 - x 902006/54-227-2003-1-15-710-254/922 Important information: Please observe the internal lead wire resistance with respect to the operating temperature; see the final page for the resistance values. Operating media: water and oil Operating pressure: max. 27 bar up to 100 °C, 20 bar up to 400 °C, pressureless as of 400 °C Pos. 1 G NL EL
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 4/12 Screw-in RTD temperature probe according to DIN 43765, form B and terminal head according to DIN 43729, form B, M24 × 1.5 Insertion length EL in mm Thread G in inches Temperature range in °C Type up to February 28, 2013 1× Pt100 Type up to February 28, 2013 2× Pt100 Type as of March 01, 2013 Order code Steel protection tube St 35.8, material-no. 1.0305 250 G 1/2 -170 to +480 90.239 ● - 902006/10-226-1003-1-9-250-104/000 ●
250 G 1/2 -170 to +480 - 90D239 902006/10-226-2003-1-9-250-104/000
Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571
100 G 1/2 -40 to +400 902006/10-402- 1003 - 902006/10-402-1003-1-9-100-104/000
150 G 1/2 -40 to +400 x - 902006/10-402-1003-1-9-150-104/000
200 G 1/2 -40 to +400 x - 902006/10-402-1003-1-9-200-104/000
100 G 1/2 -40 to +400 - 902006/10-402-2003 902006/10-402-2003-1-9-100-104/000
150 G 1/2 -40 to +400 - x 902006/10-402-2003-1-9-150-104/000
200 G 1/2 -40 to +400 - x 902006/10-402-2003-1-9-200-104/000
Operating media: water and oil Operating pressure: max. 35 bar up to 100 °C, 20 bar up to 400 °C, pressureless as of 400 °C
- Available from stock. Pos. 2 Versions only for connecting a JUMO safetyM STB/STW Ex type 701155 Insertion length EL in mm Thread G in inches Temperature range in °C 1× Pt100 2× Pt100 Type Order code Steel protection tube St 35.8, material-no. 1.0305
250 G 1/2 -170 to +480 x - 902006/10-226-1003-1-9-250-104/922
250 G 1/2 -170 to +480 - x 902006/10-226-2003-1-9-250-104/922
Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571
100 G 1/2 -40 to +400 x - 902006/10-402-1003-1-9-100-104/922
150 G 1/2 -40 to +400 x - 902006/10-402-1003-1-9-150-104/922
200 G 1/2 -40 to +400 x - 902006/10-402-1003-1-9-200-104/922
100 G 1/2 -40 to +400 - x 902006/10-402-2003-1-9-100-104/922
150 G 1/2 -40 to +400 - x 902006/10-402-2003-1-9-150-104/922
200 G 1/2 -40 to +400 - x 902006/10-402-2003-1-9-200-104/922
Operating media: water and oil Operating pressure: max. 35 bar up to 100 °C, 20 bar up to 400 °C, pressureless as of 400 °C Pos. 2
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 5/12 Screw-in RTD temperature probe with weld-in sleeve and terminal head according to DIN 43729, form B, M24 × 1.5 Insertion lengtha EL in mm Length L in mm Length L1 in mm Temperature range in °C Type up to February 28, 2013 1× Pt100 Type up to February 28, 2013 2× Pt100 Type up to February 28, 2013 3× Pt100 Type as of March 01, 2013 Order code Steel welding shoulder 15 Mo 3, material-no. 1.5415 Steel protection tube St 35.8, material-no. 1.0305 100 80 96 -40 to +480 90.239-F01 - - 902006/53-507-1003-1-12-100-815/000 160 140 96 -40 to +480 90.239-F11 - - 902006/53-507-1003-1-12-160-815/000 190 170 146 -40 to +400 90.239-F03 - - 902006/53-505-1003-1-12-190-815/000 220 200 96 -40 to +480 90.239-F21 ● - - 902006/53-507-1003-1-12-220-815/000 ● 100 80 96 -40 to +480 - 90.239-F02 ● - 902006/53-507-2003-1-12-100-815/000 ● 160 140 96 -40 to +480 - 90.239-F12 - 902006/53-507-2003-1-12-160-815/000 190 170 146 -40 to +400 - 90D239-F03 ● - 902006/53-505-2003-1-12-190-815/000 ● 220 200 96 -40 to +480 - 902006/ 53-507-2003 ● - 902006/53-507-2003-1-12-220-815/000 ● 100 80 96 -40 to +400 - - 90.239-F07 902006/53-505-3003-1-12-100-815/000 160 140 96 -40 to +400 - - 90.239-F17 902006/53-505-3003-1-12-160-815/000 220 200 96 -40 to +400 - - 90.239-F27 902006/53-505-3003-1-12-220-815/000 Operating media: water and oil Operating pressure: max. 103 bar up to 100 °C, 30 bar up to 480 °C
- Available from stock. Pos. 3 a Insertion length of the RTD temperature probe Versions only for connecting a JUMO safetyM STB/STW Ex type 701155 Insertion lengtha EL in mm Length L in mm Length L1 in mm Temperature range in °C 1× Pt100 2× Pt100 3× Pt100 Type Order code Steel welding shoulder 15 Mo 3, material-no. 1.5415 Steel protection tube St 35.8, material-no. 1.0305 100 80 96 -40 to +480 x - - 902006/53-507-1003-1-12-100-815/922 160 140 96 -40 to +480 x - - 902006/53-507-1003-1-12-160-815/922 190 170 146 -40 to +400 x - - 902006/53-505-1003-1-12-190-815/922 220 200 96 -40 to +480 x - - 902006/53-507-1003-1-12-220-815/922 100 80 96 -40 to +480 - x - 902006/53-507-2003-1-12-100-815/922 160 140 96 -40 to +480 - x - 902006/53-507-2003-1-12-160-815/922 190 170 146 -40 to +400 - x - 902006/53-505-2003-1-12-190-815/922 220 200 96 -40 to +480 - x - 902006/53-507-2003-1-12-220-815/922 100 80 96 -40 to +400 - - x 902006/53-505-3003-1-12-100-815/922 160 140 96 -40 to +400 - - x 902006/53-505-3003-1-12-160-815/922 220 200 96 -40 to +400 - - x 902006/53-505-3003-1-12-220-815/922 Operating media: water and oil Operating pressure: max. 103 bar up to 100 °C, 30 bar up to 480 °C Pos. 3 a Insertion length of the RTD temperature probe
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 6/12 Screw-in RTD temperature probe with weld-in sleeve and stainless steel braided 0.35 mm2 connecting cable and anti-kink protection Insertion lengtha EL in mm a Insertion length of the RTD temperature probe Length L in mm Temperature range in °C Type up to February 28, 2013 1× Pt100 Type as of March 01, 2013 Order code Steel welding shoulder 15 Mo 3, material-no. 1.5415 Steel protection tube St 35.8, material-no. 1.0305 100 80 -170 to +480 90.280-F32 902006/40-226-1003-1-12-100-815/000 160 140 -170 to +480 90.280-F31 902006/40-226-1003-1-12-160-815/000 220 200 -170 to +480 90.280-F30 902006/40-226-1003-1-12-220-815/000 Operating media: water and oil Operating pressure: max. 103 bar up to 100 °C, 30 bar up to 480 °C Pos. 4 Versions only for connecting a JUMO safetyM STB/STW Ex type 701155 Insertion lengtha EL in mm a Insertion length of the RTD temperature probe Length L in mm Temperature range in °C 1× Pt100 Type Order code Steel welding shoulder 15 Mo 3, material-no. 1.5415 Steel protection tube St 35.8, material-no. 1.0305 100 80 -170 to +480 x 902006/40-226-1003-1-12-100-815/922 160 140 -170 to +480 x 902006/40-226-1003-1-12-160-815/922 220 200 -170 to +480 x 902006/40-226-1003-1-12-220-815/922 Operating media: water and oil Operating pressure: max. 103 bar up to 100 °C, 30 bar up to 480 °C Pos. 4
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 7/12 Push-in RTD temperature probe with displaceable stop flange and terminal head according to DIN 43729, form B, M24 × 1.5 Insertion length EL in mm Nominal length NL in mm Temperature range in °C Type up to February 28, 2013 1× Pt100 Type up to February 28, 2013 2× Pt100 Type as of March 01, 2013 Order code Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571 50 to 460 500 -170 to +700 x - 902006/65-228-1003-1-15-500-668/000 50 to 670 710 -170 to +700 x - 902006/65-228-1003-1-15-710-668/000 50 to 960 1000 -170 to +700 x - 902006/65-228-1003-1-15-1000-668/000 50 to 460 500 -170 to +700 - 90.271-F01 902006/65-228-2003-1-15-500-668/000 ● 50 to 670 710 -170 to +700 - 90.272-F01 902006/65-228-2003-1-15-710-668/000 50 to 960 1000 -170 to +700 - 90.273-F01 902006/65-228-2003-1-15-1000-668/000 Important information: Please observe the internal lead wire resistance with respect to the operating temperature; see the final page for the resistance values. Operating medium: air
- Available from stock. Pos. 5 Versions only for connecting a JUMO safetyM STB/STW Ex type 701155 Insertion length EL in mm Nominal length NL in mm Temperature range in °C 1× Pt100 2× Pt100 Type Order code Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571 50 to 460 500 -170 to +700 x - 902006/65-228-1003-1-15-500-668/922 50 to 670 710 -170 to +700 x - 902006/65-228-1003-1-15-710-668/922 50 to 960 1000 -170 to +700 x - 902006/65-228-1003-1-15-1000-668/922 50 to 460 500 -170 to +700 - x 902006/65-228-2003-1-15-500-668/922 50 to 670 710 -170 to +700 - x 902006/65-228-2003-1-15-710-668/922 50 to 960 1000 -170 to +700 - x 902006/65-228-2003-1-15-1000-668/922 Important information: Please observe the internal lead wire resistance with respect to the operating temperature; see the final page for the resistance values. Operating medium: air Pos. 5
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 8/12 Screw-in RTD temperature probe with pipe screw connection and terminal head, according to DIN 43729, form B, M 24 × 1.5 Insertion length EL in mm Thread G in inches Nominal length NL in mm Temperature range in °C Type up to February 28, 2013 1× Pt100 Type up to February 28, 2013 2× Pt100 Type as of March 01, 2013 Order code Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571 50 to 460 G 1/2 500 -170 to +700 x - 902006/55-228-1003-1-15-500-254/000 50 to 670 G 1/2 710 -170 to +700 x - 902006/55-228-1003-1-15-710-254/000 50 to 960 G 1/2 1000 -170 to +700 x - 902006/55-228-1003-1-15-1000-254/000 50 to 460 G 1/2 500 -170 to +700 - x 902006/55-228-2003-1-15-500-254/000 ● 50 to 670 G 1/2 710 -170 to +700 - x 902006/55-228-2003-1-15-710-254/000 50 to 960 G 1/2 1000 -170 to +700 - x 902006/55-228-2003-1-15-1000-254/000 Important information: Please observe the internal lead wire resistance with respect to the operating temperature; see the final page for the resistance values. Operating medium: air
- Available from stock. Pos. 6 Versions only for connecting a JUMO safetyM STB/STW Ex type 701155 Insertion length EL in mm Thread G in inches Nominal length NL in mm Temperature range in °C 1× Pt100 2× Pt100 Type Order code Stainless steel protection tube X 6 CrNiMoTi 17 12 2, material-no. 1.4571 50 to 460 G 1/2 500 -170 to +700 x - 902006/55-228-1003-1-15-500-254/922 50 to 670 G 1/2 710 -170 to +700 x - 902006/55-228-1003-1-15-710-254/922 50 to 960 G 1/2 1000 -170 to +700 x - 902006/55-228-1003-1-15-1000-254/922 50 to 460 G 1/2 500 -170 to +700 - x 902006/55-228-2003-1-15-500-254/922 50 to 670 G 1/2 710 -170 to +700 - x 902006/55-228-2003-1-15-710-254/922 50 to 960 G 1/2 1000 -170 to +700 - x 902006/55-228-2003-1-15-1000-254/922 Important information: Please observe the internal lead wire resistance with respect to the operating temperature; see the final page for the resistance values. Operating medium: air Pos. 6
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 9/12 Allocation list of temperature probes to the respective JUMO devices For connection to... JUMO safetyM STB/STW type 701150 The temperature probes are suitable for use according to DIN EN 14597. Based on the test report from TÜV-Nord no. SAS-110-2006TB-2. Temperature probes for air Temperature probes for water and oil Type as of March 01, 2013 Order code Type up to February 28, 2013 Measuring probe Temperature range in °C Weld-in sleeve 902006/65-228-1003-1-15-500-668/000 - 1× Pt100 -170 to +700 - 902006/65-228-1003-1-15-710-668/000 - 1× Pt100 -170 to +700 - 902006/65-228-1003-1-15-1000-668/000 - 1× Pt100 -170 to +700 - 902006/65-228-2003-1-15-500-668/000 90.271-F01 2× Pt100 -170 to +700 - 902006/65-228-2003-1-15-710-668/000 90.272-F01 2× Pt100 -170 to +700 - 902006/65-228-2003-1-15-1000-668/000 90.273-F01 2× Pt100 -170 to +700 - 902006/55-228-1003-1-15-500-254/000 - 1× Pt100 -170 to +700 - 902006/55-228-1003-1-15-710-254/000 - 1× Pt100 -170 to +700 - 902006/55-228-1003-1-15-1000-254/000 - 1× Pt100 -170 to +700 - 902006/55-228-2003-1-15-500-254/000 - 2× Pt100 -170 to +700 - 902006/55-228-2003-1-15-710-254/000 - 2× Pt100 -170 to +700 - 902006/55-228-2003-1-15-1000-254/000 - 2× Pt100 -170 to +700 - Type as of March 01, 2013 Order code Type up to February 28, 2013 Measuring probe Temperature range in °C Weld-in sleeve 902006/54-227-1003-1-15-710-254/000 90.272-F03 1× Pt100 -170 to +550 - 902006/54-227-2003-1-15-710-254/000 90.272-F02 2× Pt100 -170 to +550 - 902006/10-226-1003-1-9-250-104/000 90.239 1× Pt100 -170 to +480 - 902006/10-226-2003-1-9-250-104/000 90D239 2× Pt100 -170 to +480 - 902006/10-402-1003-1-9-100-104/000 902006/10-402-1003 1× Pt100 -40 to +400 - 902006/10-402-1003-1-9-150-104/000 - 1× Pt100 -40 to +400 - 902006/10-402-1003-1-9-200-104/000 - 1× Pt100 -40 to +400 - 902006/10-402-2003-1-9-100-104/000 902006/10-402-2003 2× Pt100 -40 to +400 - 902006/10-402-2003-1-9-150-104/000 - 2× Pt100 -40 to +400 - 902006/10-402-2003-1-9-200-104/000 - 2× Pt100 -40 to +400 - 902006/53-507-1003-1-12-100-815/000 90.239-F01 1× Pt100 -40 to +480 x 902006/53-507-1003-1-12-160-815/000 90.239-F11 1× Pt100 -40 to +480 x 902006/53-505-1003-1-12-190-815/000 90.239-F03 1× Pt100 -40 to +400 x 902006/53-507-1003-1-12-220-815/000 90.239-F21 1× Pt100 -40 to +480 x 902006/53-507-2003-1-12-100-815/000 90.239-F02 2× Pt100 -40 to +480 x 902006/53-507-2003-1-12-160-815/000 90.239-F12 2× Pt100 -40 to +480 x 902006/53-505-2003-1-12-190-815/000 90D239-F03 2× Pt100 -40 to +400 x 902006/53-507-2003-1-12-220-815/000 902006/53-507-2003 2× Pt100 -40 to +480 x 902006/53-505-3003-1-12-100-815/000 90.239-F07 3× Pt100 -40 to +400 x 902006/53-505-3003-1-12-160-815/000 90.239-F17 3× Pt100 -40 to +400 x 902006/53-505-3003-1-12-220-815/000 90.239-F27 3× Pt100 -40 to +400 x 902006/40-226-1003-1-12-100-815/000 90.280-F32 1× Pt100 -170 to +480 x 902006/40-226-1003-1-12-160-815/000 90.280-F31 1× Pt100 -170 to +480 x 902006/40-226-1003-1-12-220-815/000 90.280-F30 1× Pt100 -170 to +480 x
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 10/12 For connection to... JUMO safetyM TB/TW type 701160 JUMO safetyM TB/TW 08 type 701170 JUMO DICON 500 type 703570 JUMO DICON 501 type 703580 The temperature probes are suitable for use according to DIN EN 14597. Based on the test report from TÜV-Süd no. C-T 1487-00/12. Temperature probes for air Temperature probes for water and oil Type as of March 01, 2013 Order code Type up to February 28, 2013 Measuring probe Temperature range in °C Weld-in sleeve 902006/65-228-1003-1-15-500-668/000 - 1× Pt100 -170 to +700 - 902006/65-228-1003-1-15-710-668/000 - 1× Pt100 -170 to +700 - 902006/65-228-1003-1-15-1000-668/000 - 1× Pt100 -170 to +700 - 902006/65-228-2003-1-15-500-668/000 90.271-F01 2× Pt100 -170 to +700 - 902006/65-228-2003-1-15-710-668/000 90.272-F01 2× Pt100 -170 to +700 - 902006/65-228-2003-1-15-1000-668/000 90.273-F01 2× Pt100 -170 to +700 - 902006/55-228-1003-1-15-500-254/000 - 1× Pt100 -170 to +700 - 902006/55-228-1003-1-15-710-254/000 - 1× Pt100 -170 to +700 - 902006/55-228-1003-1-15-1000-254/000 - 1× Pt100 -170 to +700 - 902006/55-228-2003-1-15-500-254/000 - 2× Pt100 -170 to +700 - 902006/55-228-2003-1-15-710-254/000 - 2× Pt100 -170 to +700 - 902006/55-228-2003-1-15-1000-254/000 - 2× Pt100 -170 to +700 - Type as of March 01, 2013 Order code Type up to February 28, 2013 Measuring probe Temperature range in °C Weld-in sleeve 902006/54-227-1003-1-15-710-254/000 90.272-F03 1× Pt100 -170 to +550 - 902006/54-227-2003-1-15-710-254/000 90.272-F02 2× Pt100 -170 to +700 - 902006/10-226-1003-1-9-250-104/000 90.239 1× Pt100 -170 to +480 - 902006/10-226-2003-1-9-250-104/000 90D239 2× Pt100 -170 to +480 - 902006/10-402-1003-1-9-100-104/000 902006/10-402-1003 1× Pt100 -40 to +400 - 902006/10-402-1003-1-9-150-104/000 - 1× Pt100 -40 to +400 - 902006/10-402-1003-1-9-200-104/000 - 1× Pt100 -40 to +400 - 902006/10-402-2003-1-9-100-104/000 902006/10-402-2003 2× Pt100 -40 to +400 - 902006/10-402-2003-1-9-150-104/000 - 2× Pt100 -40 to +400 - 902006/10-402-2003-1-9-200-104/000 - 2× Pt100 -40 to +400 - 902006/53-507-1003-1-12-100-815/000 90.239-F01 1× Pt100 -40 to +480 x 902006/53-507-1003-1-12-160-815/000 90.239-F11 1× Pt100 -40 to +480 x 902006/53-505-1003-1-12-190-815/000 90.239-F03 1× Pt100 -40 to +400 x 902006/53-507-1003-1-12-220-815/000 90.239-F21 1× Pt100 -40 to +480 x 902006/53-507-2003-1-12-100-815/000 90.239-F02 2× Pt100 -40 to +480 x 902006/53-507-2003-1-12-160-815/000 90.239-F12 2× Pt100 -40 to +480 x 902006/53-505-2003-1-12-190-815/000 90D239-F03 2× Pt100 -40 to +400 x 902006/53-507-2003-1-12-220-815/000 902006/53-507-2003 2× Pt100 -40 to +480 x 902006/53-505-2003-1-12-100-815/000 90.239-F07 3× Pt100 -40 to +400 x 902006/53-505-3003-1-12-160-815/000 90.239-F17 3× Pt100 -40 to +400 x 902006/53-505-3003-1-12-220-815/000 90.239-F27 3× Pt100 -40 to +400 x 902006/40-226-1003-1-12-100-815/000 90.280-F32 1× Pt100 -170 to +480 x 902006/40-226-1003-1-12-160-815/000 90.280-F31 1× Pt100 -170 to +480 x 902006/40-226-1003-1-12-220-815/000 90.280-F30 1× Pt100 -170 to +480 x
JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com 2013-07-30/00311472 Data Sheet 902006 Page 11/12 For connection to... JUMO safetyM STB/STW Ex type 701155 The temperature probes are suitable for use in potentially explosive areas according to DIN EN 14597. Based on the test report from TÜV-Nord no. 12 203 556139. Temperature probes for air Temperature probes for water and oil Type as of March 01, 2013 Order code Type up to February 28, 2013 Measuring probe Temperature range in °C Weld-in sleeve 902006/65-228-1003-1-15-500-668/922 - 1× Pt100 -170 to +700 - 902006/65-228-1003-1-15-710-668/922 - 1× Pt100 -170 to +700 - 902006/65-228-1003-1-15-1000-668/922 - 1× Pt100 -170 to +700 - 902006/65-228-2003-1-15-500-668/922 90.271-F01 2× Pt100 -170 to +700 - 902006/65-228-2003-1-15-710-668/922 90.272-F01 2× Pt100 -170 to +700 - 902006/65-228-2003-1-15-1000-668/922 90.273-F01 2× Pt100 -170 to +700 - 902006/55-228-1003-1-15-500-254/922 - 1× Pt100 -170 to +700 - 902006/55-228-1003-1-15-710-254/922 - 1× Pt100 -170 to +700 - 902006/55-228-1003-1-15-1000-254/922 - 1× Pt100 -170 to +700 - 902006/55-228-2003-1-15-500-254/922 - 2× Pt100 -170 to +700 - 902006/55-228-2003-1-15-710-254/922 - 2× Pt100 -170 to +700 - 902006/55-228-2003-1-15-1000-254/922 - 1× Pt100 -170 to +700 - Type as of March 01, 2013 Order code Type up to February 28, 2013 Measuring probe Temperature range in °C Weld-in sleeve 902006/54-227-1003-1-15-710-254/922 90.272-F03 1× Pt100 -170 to +550 - 902006/54-227-2003-1-15-710-254/922 90.272-F02 2× Pt100 -170 to +700 - 902006/10-226-1003-1-9-250-104/922 90.239 1× Pt100 -170 to +480 - 902006/10-226-2003-1-9-250-104/922 90D239 2× Pt100 -170 to +480 - 902006/10-402-1003-1-9-100-104/922 902006/10-402-1003 1× Pt100 -40 to +400 - 902006/10-402-1003-1-9-150-104/922 - 1× Pt100 -40 to +400 - 902006/10-402-1003-1-9-200-104/922 - 1× Pt100 -40 to +400 - 902006/10-402-2003-1-9-100-104/922 902006/10-402-2003 2× Pt100 -40 to +400 - 902006/10-402-2003-1-9-150-104/922 - 2× Pt100 -40 to +400 - 902006/10-402-2003-1-9-200-104/922 - 2× Pt100 -40 to +400 - 902006/53-507-1003-1-12-100-815/922 90.239-F01 1× Pt100 -40 to +480 x 902006/53-507-1003-1-12-160-815/922 90.239-F11 1× Pt100 -40 to +480 x 902006/53-505-1003-1-12-190-815/922 90.239-F03 1× Pt100 -40 to +400 x 902006/53-507-1003-1-12-220-815/922 90.239-F21 1× Pt100 -40 to +480 x 902006/53-507-2003-1-12-100-815/922 90.239-F02 2× Pt100 -40 to +480 x 902006/53-507-2003-1-12-160-815/922 90.239-F12 2× Pt100 -40 to +480 x 902006/53-505-2003-1-12-190-815/922 90D239-F03 2× Pt100 -40 to +400 x 902006/53-507-2003-1-12-220-815/922 902006/53-507-2003 2× Pt100 -40 to +480 x 902006/53-505-3003-1-12-100-815/922 90.239-F07 3× Pt100 -40 to +400 x 902006/53-505-3003-1-12-160-815/922 90.239-F17 3× Pt100 -40 to +400 x 902006/53-505-3003-1-12-220-815/922 90.239-F27 3× Pt100 -40 to +400 x 902006/40-226-1003-1-12-100-815/922 90.280-F32 1× Pt100 -170 to +480 x 902006/40-226-1003-1-12-160-815/922 90.280-F31 1× Pt100 -170 to +480 x 902006/40-226-1003-1-12-220-815/922 90.280-F30 1× Pt100 -170 to +480 x
Data Sheet 902006 Page 12/12 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 E-mail: mail@jumo.net Internet: www.jumo.net JUMO Instrument Co. Ltd. JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 E-mail: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: 315-437-5866 1-800-554-5866 Fax: 315-437-5860 E-mail: info.us@jumo.net Internet: www.jumousa.com Temperature errors The internal lead wire causes a systematic measure- ment error in the versions 902006/54 and 902006/65. This depends on both the nominal length (NL) as well as the measured temperature. In the table, the error is shown per 100 mm insertion length (EL) depending on the temperature measurement. Lead wire resistance of a 0.5 mm NiCr wire per 100 mm insertion length (EL) (double wire length). The specifications are to be viewed as reference values. Lead wire resistance R/Ω Measurement error Δt/K Measured temperature t/°C t/°C R/Ω Δt/K 0 0.73 1.87 100 0.75 1.98 200 0.77 2.09 300 0.79 2.22 400 0.81 2.35 500 0.83 2.49 600 0.85 2.64 700 0.87 2.80 800 0.89 2.98