V7.00/EN/00710338/2023-11-01 Page 1/25Data Sheet 706521 70652100T10Z001K000 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc.
6724 Joy Road
East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com JUMO LOGOSCREEN 601 Paperless Recorder with Touchscreen Brief description The JUMO LOGOSCREEN 601 paperless recorder is characterized by an intuitive, icon-based operation and visualization concept that makes it easy to operate. Different versions of the JUMO L OGOSCREEN 601 are available for process data recording. The scalability allows for flexible adaptation to various customer requirements: from the device version without measurement input (24 process values via interface) through to different device versions with up to 6 measurement inputs (universal analog inputs), 2 analog outputs, 12 digital inputs, 12 individually switchabl e digital inputs/outputs. A relay output is available as standard. The version with FDA-compliant data recording fulfills all requirements according to 21 CFR Part 11. In order to display the recorded data, the JUMO LOGOSCREEN 601 features various visualiza- tions. In addition, the user can use the setup program to create up to 6 separate process screens – with up to 100 objects per process screen – according to his individual requirements. For batch- based processes, a special batch recording is available which enables the storage of additional, batch-related information. The extra code "structured text" all ows for the creation of individual measurement and recording applications. Type 706521/... Block diagram Setup programm PCC, PCA3000 PCS, PCAT Software Internal channels 6 math channels (optional) 6 logic channels (optional) 6 counters/integrators 2 high-speed counters ST code (optional) Inputs/outputs Optionen 1 und 2: 3 analog inputs, 6 digital inputs, 1 analog output each 12 digital inputs/outputs, incl. 2 high-speed counting inputs (up to 12.5 kHz) Option 3: Display/operation JUMO LOGOSCREEN 601 Voltage supply Display Operation 14.5 cm (5.7”) TFT color screen 640 x 480 pixel, 65536 colors Touchscreen (resistive) As standard: 24 external analog inputs and 24 external digital inputs and 14 external texts (10 batch texts, 4 event texts) Inputs via interface As standard: 1 relay (changeover contact) Relay output AC 110 to 240 V +10/-15 %, 48 to 63 Hz AC/DC 20 to 30 V, 48 to 63 Hz Interfaces As standard: 1x Ethernet 10/100 MBit/s 1x USB host (flash drive) 1x USB device (setup) 1x RS232/RS485 (Modbus master/slave or barcode scanner) Optional: 1x PROFINET IO Device Internal memory (data transfer via interface or USB flash drive) : 1 GByte Measurem.-data memory Special features
- Intuitive touch operation
- Up to 2 analog outputs
- Up to 6 customer-specific process screens
- PROFINET IO device interface (extra code)
- Integrated web server for online visual- ization as on device
- Recording of a batch report
- Limit value monitoring function (24 chan- nels)
- Flow measurement (2 channels, option)
- 2 counter inputs (max. 12.5 kHz, option)
- User-specific application using structured text (ST code; extra code)
- Automatic data readout via PCA Commu- nication Software PCC
- Data recording compliant with FDA 21 CFR Part 11 (extra code)
- Manipulation detection with digital certifi- cate (extra code)
- PC programs for data evaluation and ac- cess control
- AMS2750/CQI-9 (extra code)
- Wide operating temperature range Approvals and approval marks (see technical data)
V7.00/EN/00710338/2023-11-01 Page 2/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000
Description
Configuration and operation On the device The JUMO operation and visualization con- cept, allows the user to operate the paperless recorder almost intuitively. All operations are performed with an icon-based menu system on the resistive touchscreen. The integrated user management protects the paperless recorder against unauthorized ac- cess. The standard version supports up to five users with varying access rights. The extra code 888 (FDA 21 CFR Part 11) allows for up to 50 users to be managed. With the setup program The paperless recorder can also be config- ured using the setup program; it should be not- ed that some functions are only available in the setup program, such as:
- Editing the operating language
- Assigning user rights
- Creating process screens
- Creating texts (e.g. for batch reports and process screens) The setup program is installed on a PC with a Windows 1 operating system (7/8/10 – 32 or 64 bit) and communicates with the paperless re- corder via USB or Ethernet interface. It is also possible to transfer configuration files to the paperless recorder using a USB flash drive.
1 Windows is a registered trademark of the
Microsoft Corporation. The user can save the configuration data as a file, which can also be printed out for docu- mentation purposes. Operating language Multiple operating languages can be selected on the device. These can be edited and switched using the setup program. The lan- guages German, English, French, Spanish, Czech, Chinese, Russian, and Italian are cur- rently available. Users can also create their own language versions (Unicode compatible). Process screen editor The user can use the setup program to create 6 individual process screens which he can subsequently transfer to the paperless record- er and use to display process data and input text and process values there. Each process screen can consist of up to 100 objects (imag- es, analog channels, digital channels, text, etc.). Interfaces USB The paperless recorder is equipped with two USB interfaces as a standard feature. A USB flash drive can be connected to the host inter- face located on the front. The device interface on the back (Micro-B type) can be used to con- nect the device to a PC (setup program or PCC/PCA3000). The USB host interface has a cover so that the front of the device complies with protection type IP66. Ethernet The paperless recorder is equipped with an Ethernet interface as a standard feature, which supports the following functions:
- Communication with a PC (setup pro- gram, web server, data archiving with PCC/PCA3000)
- Email dispatch via SMTP server
- Time synchronization via SNTP server
- Communication with a Modbus master/ slave The IP address is either configured as a fixed address or received automatically from a DHCP server; DNS is supported. RS232/RS485 This interface available as a standard feature can be configured as RS232 and RS485. It is used for communicating with a Modbus mas- ter or a Modbus slave. It can also be used to connect to a barcode scanner. PROFINET IO device The paperless recorder can also be equipped with a PROFINET interface and integrated into a PROFINET network as an IO device as an optional extra. The interface also supports si- multaneous use of Ethernet standard ser- vices; thus Ethernet interfaces as a standard feature are omitted. A GSD file (GSDML) is available for the pro- gramming system of the IO controller, which describes the features of the paperless re- corder. External inputs via interface The paperless recorder can access 24 exter- nal analog inputs and 24 external digital inputs via the interfaces (Ethernet, RS232/RS485). In addition, 10 texts for batch reports and 4 event texts with a text length of up to 160 char- acters can be transferred. In doing so the Modbus-TCP/Modbus-RTU (master/slave re- spectively) reports are used. These external inputs are also available via the optional PROFINET interface. Inputs and outputs The different device versions of the paperless recorder are available with analog and digital inputs and outputs (options). The analog inputs (max. 6) are universal mea- surement inputs for RTD temperature probes, thermocouples, resistance transmitters, resis- tance/potentiometers, and standard signals (current, voltage). The analog outputs (max. 2) can each be used as voltage output (0 to 10 V) or current output (0/4 to 20 mA). The digital inputs (max. 12) and the individual- ly switchable digital inputs/outputs (max. 12) are operated with a voltage of DC 0/24 V. All device versions feature a relay output with changeover contact. Customer-specific linearization Sensor signals with special characteristic line characteristics can also be used through cus- tomer-specific linearization (e.g. PTC/NTC sensors under consideration of the resistance
V7.00/EN/00710338/2023-11-01 Page 3/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 measuring range). Configuration is carried out in the setup program on the basis of a value ta- ble with up to 40 value pairs or through a for- mula (4th order polynomial). Data recording The measured values are recorded continu- ously with a sampling rate of 125 ms. Report creation and limit value monitoring are per- formed based on these measured values. The measured values are transferred to the work- ing memory of the device depending on the programmable memory cycle and memory value (current value, average value, maximum value, minimum value, or minimum/maximum values). The paperless recorder stores the data according to group; one input can be as- signed to multiple groups (max. 4). Working memory (SRAM) The data stored in the SRAM is copied to the internal memory in 20-kByte blocks at regular intervals. Internal memory (flash) Whenever a memory block in the working memory is full, it is copied to the internal mem- ory. The internal memory has a maximum ca- pacity of 1 GB. Each write operation is monitored to ensure that any data storage er- rors are detected immediately. The device monitors the capacity of the inter- nal memory and, if the remaining capacity falls below the configured minimum, a memory alarm signal is triggered. This can, for exam- ple, control the alarm relay. The data is written to the memory as a ring buffer, which means that when the memory is full, the oldest data is automatically overwrit- ten with new data. To show the history in the paperless recorder, data from the internal memory can be dis- played (history memory: 8 MByte). Data transfer to PC Data can be transferred from the paperless re- corder to a PC via a USB flash drive or via one of the interfaces (USB device, Ethernet). Data security Data is stored in an encrypted format devel- oped by JUMO. This ensures a high level of data security. The following applies if the paperless recorder is disconnected from the voltage supply:
- Measurement data in the working memo- ry and time are buffered by a lithium bat- tery (operating life > 7 years).
- If the lithium battery is discharged, the measurement data in the working memo- ry and the time are lost. For the purposes of a battery change, the data is buffered for approximately 2 minutes by a storage capacitor.
- Measurement and configuration data in the internal memory are not lost. Extra code 887 gives the device reliable ma- nipulation detection. A digital device certificate verifies that the recording data in the device has not been manipulated – which also ap- plies to the transfer into the data archive. Recording time The maximum recording time depends on a number of factors, in particular on the set memory cycle. The values specified in the ta- ble (entries in the event list reduce the maxi- mum recording time) apply when a group with 6 analog channels is activated in standard op- eration and storage of average values (not the minimum/maximum values). Reports For each channel in a group, reports can be maintained over specified time periods (maxi- mum, minimum, and average values). Config- uration takes place for each group. Batch report The paperless recorder allows a batch report to be created for one plant. The measurement data, the beginning, end, and duration of the batch can be displayed together with a batch counter and freely definable texts on the pa- perless recorder and in the PC Evaluation Software PCA3000. A barcode scanner can also be used to start and stop the batch and to load batch texts. GPS data recording GPS data (NMEA 0183 data records) can be received and recorded via the serial interface of the device. The data of the connected GPS receivers (e.g. positioning data) are cyclically Memory cycle Max. recording time 125 ms Approx. 42 days 1 s Approx. 8 months 5 s Approx. 41 months 10 s Approx. 82 months 60 s Approx. 493 months entered into the event list (group-related) and can thus be evaluated in conjunction with fur- ther registration data. Operating modes The operating mode can be selected individu- ally for each group. The memory cycle and memory value can be separately configured for each operating mode. Up to 4 groups can be recorded with one memory cycle of 125 ms. The operating modes have different priorities: Event operation Event operation is activated/deactivated by a control signal (such as a digital input, group, or collective alarm). The device is in event oper- ation for as long as the control signal is active. Event operation has the highest priority. Time operation Time operation is active on a daily basis within a programmable timeframe, providing event operation is not active. Standard operation If the device is not in event or time operation, standard operation is active. Limit value monitoring func- tion Up to 24 analog values can be monitored by the configurable limit value monitoring func- tion. In the event of deviation above or below the limit value, an alarm signal is generated that can be used for individual purposes (such as switching the operating mode from stan- dard to event operation). The alarm delay can be used to hide short- term deviations above or below the limit value so that no alarm signal is issued. It is also pos- sible to suppress the alarm signal by a digital signal. Limit value and switching differential can also be changed as part of parameterization, pro- vided the user has the rights to do this. Counters/integrators Six additional internal channels are available as counters, integrators, operating time count- ers, or to determine the total flow volume. Two high-speed counters (up to 12.5 kHz) can be implemented via the digital inputs/outputs 1 and 2 (option 3). These optional inputs are also required for flow measurement, if the pulses of a flow transmitter are to be evaluat- ed.
V7.00/EN/00710338/2023-11-01 Page 4/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 The counters are controlled via digital signals (counting pulses), whereas the integrators are controlled via analog signals (value integrated according to the selected time base). Operat- ing time counters determine the timeframe during which a digital signal is active. The value of the counter/integrator is dis- played in a separate window of the paperless recorder in numerical format with a maximum of 9 digits (in the event that this is exceeded, the counter restarts with 0). Different recording periods can be set. A minimum and maximum alarm can be configured for each counter/inte- grator. Math and logic module The math and logic module (each with 6 chan- nels) is available as extra code. The math function can be used to link various analog and Boolean input variables using a formula which can be defined freely in accor- dance with mathematical rules (formula with a maximum of 160 ASCII characters). The out- put variables are real values. As an alternative to entering a formula, the following mathemat- ical functions are already available: difference, ratio, humidity, and floating average. The logic function allows various Boolean val- ues to be linked using a logic formula (maxi- mum of 600 ASCII characters). The output variables are Boolean values. The math and logic module can only be config- ured via the setup program. Structured text The user has the option to create his/her own application using the "Structured text" option (extra code). The application with the ST editor, which is part of the setup program, is created in the PLC programming language "Structured text". The finished applicati on is transmitted to the device and continuously processed there. There are online-debugger functions available in the ST editor for testing and troubleshoot- ing. FDA-compliant data record- ing Extra code 888 allows the paperless recorder to fully meet FDA requirements in accordance with 21 CFR Part 11. User management and startup require the PC software package (in- cluding PCS and PCAT). The device supports up to 50 users with spe- cific rights. The user has the option to provide a completed batch or the recording data of a certain time range with their electronic signa- ture. A logged-in user can also provide their signature during logoff – it applies to the entire time period for which the user was logged in. Visualization on the device Various display types are available to visualize the measurement data on the paperless re- corder. The visualization screen after power- on-reset can be selected in the configuration, as can the screen that appears when the home button is pressed. The colors of the individual channels and the background color of the analog curves and the digital traces can be set. Vertical diagram
- Analog curves and digital traces running from top to bottom
- Up to 6 analog and 6 digital channels in one group can be shown on one screen
- Group rotation (max. 4, also with maxi- mum memory cycle)
- Digital traces can be hidden
- Channel information (short description of signal, analog value) can be hidden
- Auxiliary lines can be shown and hidden Horizontal diagram
- Analog curves and digital traces running from right to left
- Digital traces and channel information can be hidden
- Auxiliary lines can be shown and hidden Digital diagram
- Up to 6 digital channels in one group on one screen
- Vertical display (digital traces running from top to bottom)
- Horizontal display ( digital traces running from right to left) Bar graph display
- Up to 6 analog channels in one group as a bar graph on one screen
- Display of scaling and limit values
- Configurable bar color and background color
- Additional display of up to 6 digital chan- nels in one group as a symbol B1 to B6
V7.00/EN/00710338/2023-11-01 Page 5/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Text image
- Numerical display of the measured val- ues from up to 6 analog channels in one group
- Additional display of up to 6 digital chan- nels in one group as a symbol B1 to B6
- Analog channels can be displayed indi- vidually Text image – individual display
- Analog signal also as bar graph with limit values
- Color change in case of an alarm
- Alarm text display Report
- Display of minimum, maximum, and aver- age value of each analog channel in a group
- Various reporting periods
- Separate report for each group
- Display of current and completed reports Batch report
- Logging of a batch recording
- Display of the completed batch as a re- port or curve diagram Counter/integrator
- Display of the current and the completed counter/integrator
- Status of the counter/integrator with start and stop time
- Bar graph display of the current status with limit values Process screen
- Display of process data (analog and digi- tal signals) and texts as well as text and value entry
- Up to 6 process screens each with 100 objects
- Library with pictograms (also possible to import own images)
- Individual configuration using the setup program Web server The paperless recorder is equipped with a web server function as a standard feature. The web server allows the user to display cer- tain settings, process values, and messages using a web browser:
- Parameters of the user level
- Default visualizations
- Individual process screen
- Data of the recording function (including history)
- Alarm and event list The display depends on the web browser used and the PC operating system.
V7.00/EN/00710338/2023-11-01 Page 6/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 PC programs With basic type extension 1, the paperless re- corder will come with a software package con- sisting of the following PC programs: setup, PCC, and PCA3000. With extra code 888 the software package also includes the PC pro- grams PCS and PCAT (see order details). PCA communication software PCC The PCA Communication Software PCC is a PC program for Windows operating systems (7/8/10 – 32/64 bit) for extracting data from the paperless recorder.
- The data can be extracted with a USB flash drive or via an interface (USB de- vice, Ethernet).
- The data can be extracted manually or automatically (for example, every day at 11 pm). PC Evaluation Software PCA3000 The PC Evaluation Software PCA3000 is a PC program for Windows operating systems (7/8/ 10 – 32/64 bit) for managing, archiving, visual- izing, and evaluating the data from the paper- less recorder.
- The data from differently configured de- vices is detected by the PC Evaluation Software and stored in an archive data- base. Management is performed fully au- tomatically. All the user has to do is manually enter an ID (additional descrip- tion).
- The user can access certain data records which are recognizable due to the ID, at any time. The time ranges to be evaluat- ed can also be restricted.
- Any analog and digital channels of a pa- perless recorder (even from different groups) can be subsequently combined in so-called PCA groups in PCA3000.
- Since each group is shown in its own win- dow, several groups can be displayed on the screen in parallel and compared.
- Using the export filter, it is possible to ex- port the stored data in order to process it in other programs, such as Excel.
- The PC Evaluation Software PCA3000 is network compatible, which means that several users can read the data from the same archived file (*.177) in a network di- rectory independently of one another.
- Batch data or even reports can be auto- matically printed or made available in the network as PDF files, using the "automat- ic printout" PCA3000 option in conjunc- tion with the PCC software. The output forms used can be customized. PC Security Manager PCS Software for administration of device user ac- cess control. This software is only accessible to administrators. The PCS software can only be used for devic- es with extra code 888, for managing device users. PC Audit Trail Manager PCAT Software for the documentation of PC opera- tional actions that could lead to alterations in data recording. Print template JUMO PCA3000 Design Automatic printout *.xml*.csv JUMO PCC Running as a Windows service Teleservice Measurement data readout CW1...52.177 *.pdf Content
V7.00/EN/00710338/2023-11-01 Page 7/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Technical data Analog inputs General Thermocouples Number Max. 6 (see connection diagram) A/D converter 24 bit delta-sigma Sampling rate Up to 6 channels: 125 ms Input filter Digital filter, 2nd order; filter constant can be set from 0 to 100.0 s Galvanic isolation See "Galvanic isolation" Designation Type Standard ITS Measuring range Accuracy a Fe-CuNi "L" DIN 43710 (1985) IPTS-68 -200 to +900 °C ≤ 0.1 % Fe-CuNi "J" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -210 to +1200 °C ≤ 0.1 % from -100 °C Cu-CuNi "U" DIN 43710 (1985) IPTS-68 -200 to +600 °C ≤ 0.1 % from -100 °C Cu-CuNi "T" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -270 to +400 °C ≤ 0.1 % from -150 °C NiCr-Ni "K" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -270 to +1300 °C ≤ 0.1 % from -80 °C NiCr-CuNi "E" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -270 to +1000 °C ≤ 0.1 % from -80 °C NiCrSi-NiSi "N" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -270 to +1300 °C ≤ 0.1 % from -80 °C Pt10Rh-Pt "S" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -50 to +1768 °C ≤ 0.15 % from 100 °C Pt13Rh-Pt "R" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 -50 to +1768 °C ≤ 0.15 % from 100 °C Pt30Rh-Pt6Rh "B" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 0 to 1820 °C ≤ 0.15 % from 600 °C W5Re-W26Re "C" DIN EN 60584-1:2014 IEC 60584-1:2013 ITS-90 0 to 2315 °C ≤ 0.1 % from 500 °C W3Re-W25Re "D" ASTM E1751M-15 ITS-90 0 to 2315 °C ≤ 0.1 % from 500 °C W5Re-W20Re "A1" GOST R 8.585-2001 ITS-90 0 to 2500 °C ≤ 0.1 % from 500 °C Chromel®-Copel "L" GOST R 8.585-2001 ITS-90 -200 to +800 °C ≤ 0.1 % from -80°C Chromel®-Alumel® "K" GOST R 8.585-2001 ITS-90 -270 to +1372 °C ≤ 0.1 % from -80 °C PLII (Platinel® II) ASTM E1751M-15 ITS-90 0 to 1395 °C ≤ 0.1 % Ambient temperature influence ≤ 100 ppm/K Cold junction Internal (Pt100) or external (constant) or external terminal with integrated cold junction (accessories) Cold junction accuracy (internal Pt100)) ± 1K Cold junction temperature (exter- nal constant) -30 to +85 °C (adjustable) Cold junction accuracy of the ex- ternal terminal (TC terminal, ac- cessories) b Typically ± 0.4 K at ambient temperature 23 °C ± 0.5 K at ambient temperature 0 to 40 °C Base measuring range -20 to +70 mV a Accuracy refers to the measuring range b The accuracy applies to the measuring range from -100 °C.
V7.00/EN/00710338/2023-11-01 Page 8/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 RTD temperature probe Resistance transmitter and resistance/potentiometer Designation Standard ITS Measuring range Accuracy a Measuring current Pt50 DIN EN 60751:2009 IEC 60751:2008 ITS-90 -200 to +850 °C ≤ 0.1 % 500 μA Pt100 DIN EN 60751:2009 IEC 60751:2008 ITS-90 -200 to +850 °C ≤ 0.1 % 500 μA Pt500 DIN EN 60751:2009 IEC 60751:2008 ITS-90 -200 to +850 °C ≤ 0.1 % 50 μA Pt1000 DIN EN 60751:2009 IEC 60751:2008 ITS-90 -200 to +850 °C ≤ 0.1 % 50 μA Pt100 JIS C 1604:1981 IPTS-68 -200 to +649 °C ≤ 0.1 % 500 μA Pt50 GOST 6651-2009 A.2 ITS-90 -200 to +850 °C ≤ 0.1 % 500 μA Pt100 GOST 6651-2009 A.2 ITS-90 -200 to +850 °C ≤ 0.1 % 500 μA Cu50 GOST 6651-2009 A.3 ITS-90 -180 to +200 °C ≤ 0.4 % 500 μA Cu100 GOST 6651-2009 A.3 ITS-90 -180 to +200 °C ≤ 0.4 % 500 μA Ni100 DIN 43760 (1987) IPTS-68 -60 to +250 °C ≤ 0.2 % 500 μA Ni100 GOST 6651-2009 A.5 ITS-90 -60 to +180 °C ≤ 0.2 % 500 μA Connection type 2/3/4-wire Ambient temperature influence ≤ 50 ppm/K Sensor line resistance Max. 10 Ω per cable for two-wire circuit Max. 30 Ω per cable for three/four-wire circuit a Accuracy refers to the measuring range. Designation Measuring range Accuracy a Measuring current Resistance transmitter 0 to 4000 Ω≤ 0.1 % 50 μA Resistance/potentiometer 0 to 400 Ω≤ 0.1 % 500 μA 0 to 4000Ω≤ 0.1 % 50 μA Ambient temperature influence ≤ 100 ppm/K Connection type Resistance transmitter Three-wire circuit Resistance/potentiometer Two/three/four-wire circuit Smallest measuring span 60 Ω Sensor line resistance Max. 10 Ω per cable for two-wire and three-wire circuits Resistance values Freely programmable within the limits, in incr ements of 0.1 Ω a Accuracy refers to the maximum measuring range. Small measuring spans lead to reduced linearization accuracy.
V7.00/EN/00710338/2023-11-01 Page 9/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Voltage, current (standard signals) Measuring circuit monitoring The device behavior in the event of a malfunction is configurable. Analog outputs Designation Measuring range Accuracy a Input resistance or burden voltage V o l t a g e 0t o7 0m V ≤ 0.1 % > 500 kΩ 0t o1 0V ≤ 0.05 % > 500 kΩ -10 to +10 V ≤ 0.05 % > 500 kΩ -1 to +1 V ≤ 0.08 % > 500 kΩ 0t o1V ≤ 0.08 % > 500 kΩ C u r r e n t 4t o2 0m A ≤ 0.1 % < 2V 0t o2 0m A ≤ 0.1 % < 2V Ambient temperature influence ≤ 100 ppm/K Smallest measuring span Voltage 5 mV Current 0.5 mA Measuring range start/end Voltage Freely programmable within the limits, in increments of 0.01 mV Current Freely programmable within the limits, in increments of 0.01 mA Deviation below/above the measur. range According to NAMUR recom mendation NE 43 (only current input 4 to 20 mA) a Accuracy refers to the maximum measuring range. Small measuring spans lead to reduced linearization accuracy. Measuring probe Probe break Short-circuit Polarity Thermocouple Is detected Is not detected Is detected in certain co nditionsa RTD temperature probe Is detected Is detected Is not detected Resistance transmitter Is detected Is not detected Is not detected Resistance/potentiometer Is detected Is not detected Is not detect ed Voltage 0 to 70 mV Is detected Is not detected Is detected Voltage 0 to 10 V Is not detected Is not detected Is detected Voltage -10 to +10 V Is not detected Is not detected Is not detected Voltage 0 to 1 V Is detected Is not detected Is detected Voltage -1 to +1 V Is detected Is not detected Is not detected Current 0 to 20 mA Is not detected Is not detected Is not detected Current 4 to 20 mA Is detected Is detected Is detected a Dependent on the set characteristic line Number Max. 2 (see connection diagram) Voltage Output signal DC 0 to 10 V Load resistance > 500 Ω Current Output signal DC 0(4) to 20 mA Load resistance < 450 Ω Accuracy 0.5 % Ambient temperature influence 150 ppm/K
V7.00/EN/00710338/2023-11-01 Page 10/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Digital inputs Digital inputs/outputs Relays Number Max. 12 (see connection diagram) Input Level Logic level 0: < 3.5 V; logic level 1: > 10 V Sampling rate 125 ms (max. counting frequency: 4 Hz) Potential-free contact R ON: < 1k Ω; ROFF: > 50 kΩ (use of auxiliary voltage 24 V) Auxiliary voltage supply Voltage DC 24 V +10/-15 % Current Max. 50 mA per slot Number Max. 12 (see connection diagram) Input or output Individually conf igurable as input or output Input Level Logic level 0: < 3.5 V; logic level 1: > 10 V Sampling rate 125 ms (max. counting frequency: 4 Hz) Potential-free contact R ON: < 1k Ω; ROFF: > 50 kΩ (use of auxiliary voltage 24 V) High-speed input Usable inputs 1, 2 (see connection diagram) Function Counts each positive edge of the input signal Max. counting frequency 12.5 kHz Mark-to-space ratio 30 to 70 % (high-pulse ≥ 30 μs, low-pulse ≥ 30 μs) Accuracy for flow measurement 0.5 % of measured value; ambient temperature influence: 50 ppm/K Output Output signal DC 0/24 V +10/-15 %; galvanically isolated Current Max. 40 mA per output, max. 100 mA in total (including a uxiliary voltage supply current) Auxiliary voltage supply Voltage DC 24 V +10/-15 % Current Max. 100 mA (including digital outputs current) Number 1 (see connection diagram) Relay (changeover contact) Switching capacity 3 A at AC 230 V or DC 30 V, resistive load Contact life 30,000 switching operations at rated load
V7.00/EN/00710338/2023-11-01 Page 11/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Interfaces RS232/RS485 Number 1 (can be switched between RS232 and RS485) Connector type SUB-D 9-pin (socket) Baud rate 4800, 9600, 19200, 38400, 115200 Data format 8/1n, 8/1e, 8/1o Protocol Modbus RTU as master or slave, barcode scanner, NMEA 01 83 Application Communication with Modbus master/slave, connection of a barcode scanner or a GPS receiver External inputs Via Modbus master/slave functionality: 24 analog and 24 digital inputs, 10 batch texts, 4 event texts Ethernet Number 1 (alternative to PROFINET interface) Connector type RJ45 (socket) Transfer rate 10 Mbit/s, 100 Mbit/s Protocol IPv4; TCP, UDP; DHCP, DNS, HTTP, SMTP, SNTP , Modbus-TCP Application Communication with PC (setup program, data archiving, web server), email server, SNTP server, and Modbus master/slave External inputs Via Modbus master/slave functionality: 24 analog and 24 digital inputs, 10 batch texts, 4 event texts Max. cable length 100 m PROFINET IO device Number 1 (alternative to Ethernet interface) Connector type 2 x RJ45 (soc ket), integrated switch Transfer rate 100 Mbit/s Conformity class B (CC-B) Netload class III (Netload Class III) Protocol DCP, LLDP , VLAN Priority, PTCP Application Communication with PROFINET IO controller; Ethernet standard services are also supported Max. cable length 100 m USB host Number 1 (on front with cover) Connector type A (socket) Standard USB 2.0 (high speed) Max. load current 100 mA USB device Number 1 (on the back) Connector type Micro-B (socket) Standard USB 2.0 (high speed) Max. cable length 5 m
V7.00/EN/00710338/2023-11-01 Page 12/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Screen Electrical data Type TFT color screen/TFT-touchscreen (resistive) a Size 14.5 cm (5.7") Resolution 640 × 480 pixels (VGA) Number of colors 65536 Frame rate 60 Hz (type) Brightness setting Adjustable on the device Screensaver (switch-off) After wa iting period or due to control signal a TFT color screens may have pixel errors due to technological and/or production-related reasons. Up to four pixel errors are deemed admissible for this paperless recorder; they do not provide the ground for warranty claims. Voltage supply AC 110 to 240 V +10/-15 %, 48 to 63 Hz or AC/DC 20 to 30 V, 48 to 63 Hz (not in conjunction with extra code 970) Electrical safety According to DIN EN 61010-1 Overvoltage category II to 300 V mains voltage, pollution degree 2 Protection rating I with internal isolation from SELV Power consumption AC 110 to 240 V < 45 VA AC/DC 20 to 30 V < 30 VA Data backup Internal flash memory Data buffering Battery (operating life > 7 years); additionally, storage capacitor for buffering during battery change (buffer time approx. 2 minutes) Time Battery-buffered real-time clock Electrical connection On the back via push-in spring-cage terminals Conductor cross section on terminal 5 Wire or stranded wire without ferrule Min. 0.2 mm2, max. 2.5 mm2 Stranded wire with ferrule Min. 0.2 mm 2, max. 2.5 mm2 2 × stranded wire with twin ferrule with plastic collar Min. 0.5 mm2, max. 1.5 mm2 (both stranded wires with identical cross section) Stripping length 10 mm Conductor cross section on terminals 4, 14, and 15 Wire or stranded wire without ferrule Min. 0.2 mm 2, max. 2.5 mm2 (with terminal cover: max. 1.5 mm2) Stripping length 10 mm Conductor cross section on terminals 6 to Wire or stranded wire without ferrule Min. 0.14 mm 2, max. 1.5 mm2 (with terminal cover: max. 0.5 mm2) With plastic collar: min. 0.25 mm2, max. 0.5 mm2 Stripping length 9 mm Conductor cross section on TC terminal (accessories) Wire or stranded wire without ferrule Min. 0,20 mm 2, max. 1,5 mm2 (with terminal cover: max. 0,5 mm2) Stranded wire with ferrule Without plastic collar: min. 0,25 mm 2, max. 1,5 mm2 (with terminal cover: max. 0,5 mm2) With plastic collar: min. 0,25 mm2, max. 0,75 mm2 (with terminal cover: max. 0,5 mm2) Stripping length 10 mm
V7.00/EN/00710338/2023-11-01 Page 13/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Environmental influences Case Approvals and approval marks Ambient temperature range Storage -20 to +60 °C Operation -20 to +50 °C a; in conjunction with extra code 970: 0 to 40 °C Site altitude Max. 2000 m above sea level Climatic environmental influences According to DIN EN 60721-3 with extended temperature range Resistance to climatic conditions ≤ 85 % rel. humidity without condensation Storage According to class 1K2 Operation According to class 3K3 Mechanical environmental influences According to DIN EN 60721-3 Storage According to class 1M2 Transport According to class 2M2 Operation According to class 3M3 Electromagnetic compatibility (EMC) According to DIN EN 61326-1 Interference emission Class A – only for industrial use – Interference immunity Industrial requirements a At temperatures below 0 °C, the build-up of screen contents slows down. Case type Flush-mounted housing according to DIN IEC 61554 made of zinc-plated steel sheet (indoor use) Case front Made of die-cast zinc with decor foil Front frame dimensions 144 mm x 144 mm (front frame depth approx. 8 mm including seal) Mounting depth 120.9 mm (incl. spring-cage terminals) Panel cut-out 138 +1.0 mm × 138+1.0 mm Panel thickness 2 to 8 mm Case fastening In panel using t he four supplied mounting elements Operating position Any, with due c onsideration for the viewing angle of the screen, horizontal ±50°, vertical ±30° Protection type According to D IN EN 60529, front IP66, back IP20; in conjunction with extra code 970: IP20 with open carrying case, IP20D with closed carrying case Weight Max. 1.65 kg (without terminal cover) Approval mark Testing facility Certificate/certification num- ber Inspection basis Valid for c UL us Underwriters Laboratories E201387 UL 61010-1 (3. Ed.), CAN/CSA-22.2 No. 61010-1 (3rd Ed.) All versions of the built-in de- vice; not in conjunction with ex- tra code 970 The device is approved if the approval mark is shown on the device.
V7.00/EN/00710338/2023-11-01 Page 14/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Dimensions Device Panel cut-out Close mounting Device with terminal cover (accessories) 6 120.9 144 144136 138 +1 138 Distance between panel cut-outs Horizontal Vertical Minimum clearance 20 mm 20 mm Recommended distance (easier mounting of mounting elements) 50 mm 50 mm 128.1 147.6 20.2 109
V7.00/EN/00710338/2023-11-01 Page 15/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Universal carrying case, compact (extra code 970) Dimensions 260 297 327 250 200
V7.00/EN/00710338/2023-11-01 Page 16/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Galvanic isolation 1 The voltage specifications correspond to the test voltages (al ternating voltage, rms values) according to EN 61010-1:2011-07 for the type test. 2 Functional galvanic isolation for connection of SELV or PELV electrical circuits. Analog input 1 to Analog input 3 Relay 1/c187 Analog 2ouput/c187/c187 Ethernet/PROFINET interface Analog ouput 1/c187 RS232/RS485 interface 3510 V Digital inputs 1 to 6 USB host- interface USB device- interface /c187/c1873510 V Voltage supply Digital inputs 7 to 12 (1) (2) (2)(2) (2) (2) (2) (1) /c187 (2) /c187/c187 Analog input Analog input to /c187 (2) /c187 (2) /c187Digital inputs/ outputs 1 to 12 (2)
V7.00/EN/00710338/2023-11-01 Page 17/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Connection elements Front USB host interface (without cover) Back connection elements Slot 3 (option 3) Slot 2 (option 2) Slot 1 (option 1) Connection element and assignment Connection element and assignm ent 1. USB device interface 4. Relay 1 (changeover contact) 2.1 2.1, 2.2 Ethernet interface (as a standard feature) or PROFINET interface (including Ethernet; extra code): 2.1 = port 2, 2.2 = port 1 5. Voltage supply 6. - 15. Option inputs and outputs (Slot 1 to Slot 3) 3. RS232/RS485 interface
V7.00/EN/00710338/2023-11-01 Page 18/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Connection diagram The connection diagram in the dat a sheet provides preliminary i nformation about the connection op tions. For the electrical con nection, only use the quick start guide or the oper ating manual. The knowledge and the correct technical execution with the safety information a nd warnings con- tained in these documents are mandatory for mounting, electrical connection, and startup as well as for safety during operation. Analog inputs Measuring probe Terminals and connection symbol Connection elemen t.terminal / assignment Thermocouple Analog/digital option (order code 1): 7.1-5 / Analog input 1 8.1-5 / Analog input 2 9.1-5 / Analog input 3 11.1-5 / Analog input 4 12.1-5 / Analog input 5 13.1-5 / Analog input 6 RTD temperature probe Two-wire circuit RTD temperature probe Three-wire circuit RTD temperature probe Four-wire circuit Resistance transmitter Resistance/potentiometer Two-wire circuit (Continued on the following page) 12345 12345 12345 12345 12345 ESA 12345
V7.00/EN/00710338/2023-11-01 Page 19/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Resistance/potentiometer Three-wire circuit Resistance/potentiometer Four-wire circuit Voltage DC -10(0) to +10 V Voltage DC -1(0) to +1 V Voltage DC 0 to 70 mV Current DC 0(4) to 20 mA Measuring probe Terminals and connection symbol Connection elemen t.terminal / assignment 12345 12345 12345 U +- x 12345 U +- x 12345 U +- x 12345 I + -x
V7.00/EN/00710338/2023-11-01 Page 20/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Analog outputs Version Terminals and connection symbol Connection element.termin al / assignment Analog output DC 0 to 10 V or DC 0(4) to 20 mA (configurable) Analog/digital option (order code 1): 6.9 / Analog output 1 + 6.10 / Analog output 1 - 10.9 / Analog output 2 + 10.10 / Analog output 2 - 1234 9 + - 5678 1 0 x IxU ,
V7.00/EN/00710338/2023-11-01 Page 21/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Digital inputs Version Terminals and connection symbol Connection element.termin al / assignment Digital input DC 0/24 V, auxiliary voltage supply DC 24 V Example: potential-free contact on digital input 1 and +24 V (auxiliary voltage) Analog/digital option (order code 1):: 6.1 / Digital input 1 6.2 / Digital input 2 6.3 / Digital input 3 6.4 / Digital input 4 6.5 / Digital input 5 6.6 / Digital input 6 6.7 / +24 V 6.8 / GND 10.1 / Digital input 7 10.2 / Digital input 8 10.3 / Digital input 9 10.4 / Digital input 10 10.5 / Digital input 11 10.6 / Digital input 12 10.7 / +24 V 10.8 / GND Example: external voltage on digital input 1 and GND 1234 + - 5678 9 10 1234 9 U + - x 5678 1 0
24 V -+
V7.00/EN/00710338/2023-11-01 Page 22/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Digital inputs/outputs Relays RS232/RS485 interface Version Terminals and connection symbol Connection element.termin al / assignment Digital input DC 0/24 V or digital output DC 0/24 V (individually switchable), auxiliary voltage supply DC 24 V Note regarding the digital option: Auxiliary voltage supply and digital outputs to- gether deliver max. 100 mA (at 24 V). Example: potential-free contact on digital input/ output 1 (as input) and +24 V (auxiliary voltage) Digital option (order code 4): 14.1 / Digital input/output 1 14.2 / Digital input/output 2 14.3 / Digital input/output 3 14.4 / Digital input/output 4 14.5 / Digital input/output 5 14.6 / Digital input/output 6 14.7 / +24 V 14.8 / GND 15.1 / Digital input/output 7 15.2 / Digital input/output 8 15.3 / Digital input/output 9 15.4 / Digital input/output 10 15.5 / Digital input/output 11 15.6 / Digital input/output 12 15.7 / +24 V 15.8 / GND Example: external voltage on digital input/out- put 1 (as input) and GND Example: external relay on digital input/output 1 (as output) and GND (max. 40 mA per output, max. 100 mA in total, see note in the "Version" column) 1234 + - 5678 1234 U + - x 5678 + - 5678 Imax = 40 mA Version Terminals and connection symbol Connection element.termin al / assignment Relay (changeover contact) (max. 3 A at AC 230 V, resistive load) Relay 1: 4.1 / Normally open contact (NO) 4.2 / Common contact (C) 4.3 / Normally closed contact (NC) 31 2 Version Connection element.pin / assignment Connection element RS232 9-pin SUB-D socket (switchable to RS485) 3.2 / RxD (received data) 3.3 / TxD (transmission data) 3.5 / GND (ground) RS485 9-pin SUB-D socket (switchable to RS232) 3.3 / TxD+/RxD+ (transmission/received data 3.5 / GND (ground) 3.8 / TxD-/RxD- (transmission/received data -) 6789 2345 1
V7.00/EN/00710338/2023-11-01 Page 23/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Ethernet/PROFINET Voltage supply Version Connection element.pin / assignment Connection element Ethernet 1 x RJ45 (as a standard feature) 2.1.1 / TX+ (transmission data +) 2.1.2 / TX- (transmission data -) 2.1.3 / RX+ (received data +) 2.1.6 / RX- (received data -) PROFINET IO device (incl. Ethernet) 2 x RJ45, integrated switch (as extra code) Port 2: 2.1.1 / TX+ (transmission data +) 2.1.2 / TX- (transmission data -) 2.1.3 / RX+ (received data +) 2.1.6 / RX- (received data -) Port 1: 2.2.1 / TX+ (transmission data +) 2.2.2 / TX- (transmission data -) 2.2.3 / RX+ (received data +) 2.2.6 / RX- (received data -) 8 1 Version Connection element.terminal / assignment Terminals and co nnection symbol A C1 1 0t o2 4 0V + 1 0 / - 1 5% , 4 8t o6 3H z or A C / D C2 0t o3 0V , 4 8t o6 3H z Observe order details! 5.L1 / Line conductor (for DC: positive terminal L+) 5.N / Neutral conductor (for DC: negative termi- nal L-) 5.PE / Protection conductor PEL1 N PEL1 N (L+) (L-)
V7.00/EN/00710338/2023-11-01 Page 24/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Order details (1) Basic type 706521 Paperless recorder with the following interfaces: 1x Ethe rnet, 2x USB (1x host, 1x device), 1x RS232/485 interface, as well as a relay (changeover contact) (2) Basic type extension
0 Without software package
1 With software package (setup program incl. USB cable, PC Evalu ation Software PCA3000, PCA Communi- cation Software PCC; in conjunction with extra code "888" as well with PC Security Manager PCS and PC Audit Trail Manager PCAT software) (3) Language
8 Default setting (German/English)
9 Set according to customer specifications
(4) Option 1 (slot 1) a a Subsequent expansion is only possible in JUMO Central Services.
0 Not used
1 Analog/digital: 3 analog and 6 digital inputs, 1 analog output
(5) Option 2 (slot 2) a (6) Option 3 (slot 3) a
4 Digital: 12 digital inputs/out puts (individually switchable)
(7) Voltage supply 2 3 A C1 1 0t o2 4 0V + 1 0 / - 1 5% , 4 8t o6 3H z 2 5 A C / D C2 0t o3 0V , 4 8t o6 3H z (8) Extra code 1 . Not used
260 Math and logic module (6 channels each)
221 Structured text (ST code)
(9) Extra code 2 . Not used
887 Manipulation detection wi th digital certificate
888 FDA 21 CFR Part 11 with digital certificate
(10) Extra code 3 . Not used 163 PROFINET IO device in terface (incl. Ethernet)
879 AMS2750/CQI-9
b b For the calibration certificate it is necessary to state the channels along with the thermocouple type and the desired measuring points. The device is to be used as a permanently installed fie ld device. Use as a mobile field test device for SAT and TUS testing is not permitted. (11) Extra code, housing . Not used
970 Universal carrying case, compact c
c The extra code is only available in conjunction with voltage supply AC 110 to 240 V. The UL approval does not apply. Use only for personnel with technical qualifications who have been specially trained, and have the relevant knowledge in the field of au- tomation technology! Specifications for ambient temperature and for protection type are to be observed (see technical data)! a Multiple selection at positions 8 and 10 is possible. Specify extra codes one after the other, and separate them with com- mas. (9) (10) a (11) Order code / - - /,,, Order example 706521 / 1 8 - 1 1 4 - 23 / 260 , 887 , 163 , 970
V7.00/EN/00710338/2023-11-01 Page 25/25Data Sheet 706521 JUMO GmbH & Co. KG Delivery address: Mackenrodtstraße 14 Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 Email: mail@jumo.net Internet: www.jumo.net JUMO UK LTD JUMO House Temple Bank, Riverway Harlow, Essex, CM20 2DY, UK Phone: +44 1279 63 55 33 Fax: +44 1279 62 50 29 Email: sales@jumo.co.uk Internet: www.jumo.co.uk JUMO Process Control, Inc. East Syracuse, NY 13057, USA Phone: +1 315 437 5866 Fax: +1 315 437 5860 Email: info.us@jumo.net Internet: www.jumousa.com 70652100T10Z001K000 Stock versions Scope of delivery Accessories Order code Part no. 706521/08-000-23/000 00727734 706521/08-100-23/000 00727735 706521/18-100-23/000 00727736 706521/08-110-23/000 00727737 706521/18-110-23/000 00727738 1 paperless recorder in the ordered version 1 quick start guide (brief instructions) 4 mounting elements Beschreibung Teile-Nr. Setup program 00645110 USB cable, A connector to Micro-B connector, length 3 m 00616250 PC Evaluation Software PCA3000 00431882 PCA communication software PCC 00431879 PC software package consisting of: setup program, PC Evaluation Software PCA3000, PCA Communication Software PCC, PC Security Manager PCS, PC Audit Trail Manager PCAT. Please specify all version num- bers when placing follow-up orders. 00666817 USB flash drive, 2 GBa a The indicated USB flash drive has been tested and is designed for industrial applications. Other brands with larger storage ca- pacity can also be used, but no liability is assumed for this. 00505592 Activation for math and logic module (setup program required) 00 716354 Activation for structured text (ST code; setup program required) 00716357 Activation, automatic printout (PCA3000) 00505548 TP-Link TL-WR802N (Wi-Fi router) 00658592 Sealable terminal cover 00712239 Relay (N/O contact) AC 230 V / 3 A for DIN rail 00515872
3 TC terminals (2-pole) with integrated cold junctionb, usable for slot 1 and slot 2
b The TC terminals with integrated cold junction are used to connect thermocouples for high-precision temperature measurement (usable as of version 323.04.05 of the device software). 30053080