Single Output DC Power Supplies

minimize setup time and measurement errors and maximize device reliability. certiications and Keysight quality standards — at an affordable price. manufacturing and educational settings. Figure 1. The U8001A 90 W and U8002A 150 W single output DC power supplies

04 | Keysight | U8000 Series Single Output DC Power Supplies - Data Sheet Keypad Lock/Unlock options to prevent accidental front panel usage Selects the knob control function for voltage and current adjustment Control for enabling and disabling the output of the power supply Features of the U8000 Series Backlight On/Off options enabling brighter data display Highly visible LCD displaying actual values of output voltage and current in meter mode and limit values of voltage and current in limit mode Three storage locations (M1, M2, and M3) for user-deined operating states with save-and-recall capability Fully integrated over voltage and over current protection to prevent damages to the DUT Setting and display of the voltage and current limit values Versatile binding posts offer lexibility to use safety test leads, banana plugs or stripped wires Rotary knob for quick and analog-like control of voltage and current

05 | Keysight | U8000 Series Single Output DC Power Supplies - Data Sheet Performance Speciications Electrical Speciications Parameter U8001A U8002A Output Ratings (at 0 °C to 40 °C) 0 to +30 V 0 to 3 A 0 to +30 V 0 to 5 A Line and Load Regulation CV: <0.01% +2 mV CC: <0.02% +2 mA Ripple and Noise (25 °C ±5 °C) CV: 12 mVp-p, <1 mVrms; CC: 3 mArms Load Transient Response Time (within 15 mV from full load to half load and from half load to full load) <50 µs Programming Accuracy 1 (25 °C ±5 °C) <0.35% +20 mV <0.35% +20 mA Readback Accuracy 1 (25 °C ±5 ° C) <0.35% +20 mV <0.35% +20 mA Meter Resolution Voltage: 10 mV Current: 10 mA Maximum Output Float Voltage ±240 Vdc 1 Speciications are based on one hour warm-up period.

06 | Keysight | U8000 Series Single Output DC Power Supplies - Data Sheet Supplemental Characteristics Supplemental Characteristics Parameter U8001A U8002A Temperature Coefficient (for 12 months) CV: <100 ppm/ °C CC: <380 ppm/ °C CC: <300 ppm/ °C Output Voltage Overshoot (during turn-on or turn-off of AC power state with the output control set to less than 1 V) <1 V Voltage Programming Speed, to within 1% of total excursion Up Full Load No Load 150 ms 100 ms Down Full Load No Load 30 ms 450 ms Last Memory Setting Enabled Yes Over Voltage Protection Response Time <1.5 ms when the trip voltage is equ al to or greater than 3 V and <10 ms when the trip voltage is less than 3 V (average time for output to drop from 90% of output voltage to 1 V after O VP condition occurred) Protection Features Parameter U8001A U8002A Over Voltage Protection Accuracy ± (% of output + offset) <0.5% +0.5 V Over Voltage Protection Programmable Range 1 V to 33 V Over Current Protection Accuracy ± (% of output + offset) <0.5% +0.5 A Over Current Protection Programmable Range 1 A to 3.3 A 1 A to 5.5 A AC Power Input Speciications Parameter U8001A U8002A Input Power Option (selectable) 100 Vac ± 10%, 47 to 63 Hz

115 Vac ± 10%, 47 to 63 Hz

230 Vac ± 10%, 47 to 63 Hz

Maximum Input Power 330 VA 500 VA Fuse External, customer assessable Physical Speciications Parameter U8001A U8002A Dimensions (H x W x L) 88.1 mm x 212.3 mm x 394.3 mm Weight 7.3 kg 8.3 kg

07 | Keysight | U8000 Series Single Output DC Power Supplies - Data Sheet Display Liquid Crystal Display (LCD) Operating Environment – Operating temperature from 0 °C to 40 °C (for full rated output) and 40 °C to 55 °C (for derated output) – Relative humidity up to 95% – Altitude up to 2000 meters Storage Compliance -20 °C to 70 °C Safety and EMC Compliance Certiied with: – IEC 61326:2002/EN61326:1997+ A1:1998+A3:2003 – CISPR 11:1990/EN55011:1990 – Canada: ICES-001:2004 – Australia/New Zealand: AS/NZS CISPR11:2004 – IEC 61010-1:2001/EN61010-1:2001 (2 nd Edition) – Canada: CAN/CSA-C22.2 No. 61010-1-04 – USA: ANSI/UL 61010-1:2004 Memory Operations Up to three operating states Warranty – Three year for U8000 series power supplies – Three months for standard shipped accessories U8001A DC Power Supply, 30 V, 3 A U8002A DC Power Supply, 30 V, 5 A Included Documentation: U8001A-0B0 No documentation package Additional Documentation: – U8001A-AB0 Traditional Chinese User’s and Service Guide – U8001A-AB2 China – Simpliied Chinese printed User’s and Service Guide – U8001A-ABA English User’s and Service Guide – U8001A-ACF Japanese User’s and Service Guide Calibration Document: U8001A-UK6 Commercial calibration with test result data Power Options Opt 0E9 90 to 110 Vac, 47 to 63 Hz Opt 0EM 104 to 126 Vac, 47 to 63 Hz Opt 0E3 207 to 253 Vac, 47 to 63 Hz Other Options – Option U800xA-AUT Custom Power On (available only upon purchase with main model) – E3600A-100 Test lead kit – Option 1CM Rack-mount kit Rack-mount Kits To rack-mount a single instrument: Adapter kit (P/N 5063-9240) To rack-mount two instruments side- by-side: Lock-link kit (P/N 5061-9694) Flange kit (P/N 5063-9212) To rack-mount two instruments in a sliding support shelf: Support shelf (P/N 5063-9255) Slide kit (P/N 1494-0015) General Characteristics

Ordering Information

08 | Keysight | U8000 Series Single Output DC Power Supplies - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2009 - 2014 Published in USA, October 6, 2014 5989-7182EN www.keysight.com For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Paciic Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-09-23-14) myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.axiestandard.org AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Keysight is a founding member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of the PCI Industrial Computer Manufacturers Group. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. www.pxisa.org PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience.