300 WATT MEDICAL POWER SUPPLIES

DESCRIPTION

The TM300 series comprising single and multiple output models for 200 to 300 watts of continuous output power is specially designed for medical applications. They operate at 90 to 264VAC input voltage without the need of a selector strap. All auxiliary outputs are with magnetic amplifier to keep regulation. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover-and-fan assembly can be added during manufacturing. Safety Standard Approvals :

FEATURES

‧ EN61000-3-2 class A and D compliant ‧ Power Factor 0.98 typical ‧ Overvoltage protection ‧ Short-circuit protection ‧ Power Fail Detect (PFD) ‧ 100% burn-in at full rated load ‧ Remote sense on output #1 and output #2 ‧ Remote inhibit – TTL high disables output ‧ Compliant with RoHS requirements OUTPUT SPECIFICATIONS Output voltage/current : See rating chart Total output power : See rating chart Ripple and Noise : 2% peak to peak maximum Overvoltage protection : Provided on output #1 only; set at 115–140% of its nominal output voltage Overcurrent protection : All outputs protected to short circuit conditions Temperature coefficient : All outputs ±0.04%/ºC maximum Transient response : Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500µs after a 25% step load change PFD signal : TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1ms prior to master output dropping 5% below its nominal value. This signal also provides a minimum delay of 100ms after master output is within regulation. Remote inhibit : Requires an external TTL high level signal to inhibit outputs for standard models. GENERAL SPECIFICATIONS Switching frequency : KHz ±10KHz Power factor : 0.98 typical Efficiency : 70% minimum on all models Hold-up time : 12 msec minimum at 110VAC Line regulation : ±0.2% maximum at full load Inrush current : 30 amps @ 115VAC or 60 amps @230VAC, at 25ºC cold start Withstand voltage : 4000VAC from input to output 1500VAC from input to ground 500VAC from output to ground MTBF : 300,000 hours minimum at full load at 25ºC ambient, calculated per MIL-HDBK- 217F EMC Performance (EN60601-1-2: 2001) EN55011: Class B conducted, Class B radiated EN61000-3-2: Harmonic distortion, Class A and D EN61000-3-3: Line flicker EN61000-4-2: ESD, ±8KV air and ±6KV contact EN61000-4-3: Radiated immunity, 3V/m EN61000-4-4: Fast transient/burst, ±2KV EN61000-4-5: Surge, ±1KV diff, ±2KV com. EN61000-4-6: Conducted immunity, 3Vrms EN61000-4-8: Magnetic field immunity, 3A/m EN61000-4-11: Voltage dips, 30% reduction for 500ms, 60% reduction for 100ms and >95% reduction for 10ms UL 60601-1 CSA C22.2 No. 601.1 TÜV EN60601-1 TM300 SERIES INPUT SPECIFICATIONS Input voltage : 90 to 264VAC Input frequency: 47 to 63Hz Input current : 4.7A (rms) for 115VAC 2.3A (rms) for 230VAC Earth leakage current: 100µA max. @115VAC, 60Hz 220µA max. @ 230VAC, 50Hz (LVD) ENVIRONMENTAL SPECIFICATIONS Operating temperature : 0ºC to +70ºC Storage temperature : -40ºC to +85ºC Relative humidity : 5% to 95% non-condensing Derating : Derate from 100% at +50ºC linearly to 50% at +70ºC Cooling : 200/250/300 watts continuous output power at 35 CFM forced air cooling or 100/125/150 watts a t c o n v e n t i o n c o o l i n g RoHS T u m b l e r T e c h n o l o g i e s + T R U M P o w e r 3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com TRUMPower

OUTPUT VOLTAGE/CURRENT RATING CHART M a x i m u m Output # 1 Output # 2 Output # 3 Output # 4 Output TM300-S3.3 3.3V 3.0A 60A 3% 200W TM300-S5.1 5.1V 3.0A 60A 2% 300W TM300-S12 12V 1.2A 25A 2% 300W TM300-S15 15V 1.0A 20A 2% 300W TM300-S24 24V 0.6A 12.5A 2% 300W TM300-S30 30V 0.5A 10A 2% 300W TM300-S48 48V 0.5A 6.3A 2% 300W TM300-Q403 3.3V 3.0A 35A 3% 5.1V 2. 0A 22A 2% 12V 0A 4A 4% 12V 0A 4A 4% 250W TM300-Q40 5.1V 2.0A 35A 2% 12V 1.0A 10A 2% 12V 0A 4A 4% 5.1V 0A 4A 4% 300W TM300-Q41 5.1V 2.0A 35A 2% 15V 0.8A 8A 2% 15V 0A 4A 4% 24V 0A 2.5A 4% 300W TM300-Q42 5.1V 2.0A 35A 2% 12V 1.0A 10A 2% 12V 0A 4A 4% 12V 0A 4A 4% 300W TM300-Q45 5.1V 2.0A 35A 2% 12V 1.0A 10A 2% 12V 0A 4A 4% 24V 0A 2.5A 4% 300W TM300-Q46 5.1V 2.0A 35A 2% 12V 1.0A 10A 2% 12V 0A 4A 4% 15V 0A 4A 4% 300W TM300-Q47 5.1V 2.0A 35A 2% 24V 0.5A 5A 2% 12V 0A 4A 4% 12V 0A 4A 4% 300W TM300-Q48 5.1V 2.0A 35A 2% 24V 0.5A 5A 2% 5.1V 0A 4A 4% 15V 0A 4A 4% 300W TM300-Q410 24V 0.5A 6.3A 2% 12V 1.0A 10A 2% 5.1V 0A 4A 4% 12V 0A 4A 4% 300W TM300-Q412 24V 0.5A 6.3A 2% 12V 1.0A 10A 2% 12V 0A 4A 4% 12V 0A 4A 4% 300W TM300-Q413 24V 0.5A 6.3A 2% 24V 0.5A 5A 2% 5.1V 0A 4A 4% 15V 0A 4A 4% 300W TM300-Q414 24V 0.5A 6.3A 2% 24V 0.5A 5A 2% 12V 0A 4A 4% 12V 0A 4A 4% 300W MECHANICAL SPECIFICATIONS Single Output Models NOTES: 1. Dimensions shown in inch [mm] 2. Tolerance 0.02 [0.5] maximum 3. Input connector P1 is Dinkle DT-35-B01W-03 screws are M3, Nickel plated. 4. Connector P4 mates with Molex housing 50-37-5103 and pins 5263. 5. Connectors P2, P3, P5 and P6: M3*0.5 screw connections 7. DC fan on P4 is rated at 12V/0.1A. MEDICAL GRADE TM300 SERIES NOTES: 1. Add suffix "C" for enclosed format with option fan control, e.g. TM300-Q45C. 2. All outputs are floating. It can be connected externally for positive or negative output. 3. Output #1 & #2 can be adjusted within ±5% of their nominal voltage. 4. Output #3 & #4 can be adjusted within ±15% of their nominal voltage. 5. 300 watts for "C" versio n with cover-and-fan assemb ly. 150 watts for U-bracket version without moving air (maximum current of output #1 & #2 derated to 50%), or 300 watts with 35 CFM forced air provided by user. 6. TM300-S5.1 is rated 200 watts with 35 CFM forced air cooling or 100 watts convection cooled. TM300-Q403 is rated 250 watts with 35 CFM forced air cooling (maximum current of output #1 & #2 derated to 50%) or 125 watts convection coole). Single output models may be operated at no-load. At no-load, output voltage tolerance increases to 10%. T u m b l e r T e c h n o l o g i e s + T R U M P o w e r 3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com

MEDICAL GRADE TM300 SERIES MECHANICAL SPECIFICATIONS Multiple Output Models NOTES: 1. Dimensions shown in inch [mm] 2. Tolerance 0.02 [0.5] maximum 3. Input connector P1 is Dinkle DT-35-B01W-03 screws are M3, Nickel plated. 4. Connector P4 mates with Molex housing 50-37-5103 and pins 5263. 5. Connectors P2, P3, P5 and P6: M3*0.5 screw connections 6. Output connector P7 mates with Molex housing 09-50-3041 and Molex 2878 series crimp terminal. 8. DC fan on P4 is rated at 12V/0.1A. PIN CHART P1 (AC) P4 Product No. 1 2 3 P2 P3 1 2 3 4 TM300-S5.1 TM300-S3.3 TM300-S12 TM300-S15 TM300-S24 TM300-S30 TM300-S48 LIVE NEUTRAL GROUND +VO1 +VO1 SIGNAL GROUND ( -VO1) +S (VO1) -S (VO1) PFD \` P4 P7 Product No. 5 6 7 8 9 10 P5 P6 1 2 3 4 TM300-S5.1 TM300-S3.3 TM300-S12 TM300-S15 TM300-S24 TM300-S30 TM300-S48 P1 (AC) P4 Product No. 1 2 3 P2 P3 1 2 3 4 TM300-Q403 TM300-Q40 TM300-Q41 TM300-Q42 TM300-Q45 TM300-Q46 TM300-Q47 TM300-Q48 TM300-Q49 TM300-Q410 TM300-Q411 TM300-Q412 TM300-Q413 TM300-Q414 LIVE NEUTRAL GROUND +VO1 -VO1 SIGNAL GROUND ( -VO1) +S (VO1) -S (VO1) PFD P4 P7 Product No. 5 6 7 8 9 10 P5 P6 1 2 3 4 TM300-Q403 TM300-Q40 TM300-Q41 TM300-Q42 TM300-Q45 TM300-Q46 TM300-Q47 TM300-Q48 TM300-Q49 TM300-Q410 TM300-Q411 TM300-Q412 TM300-Q413 TM300-Q414 INHIBIT N.C. +S (VO2) -S (VO2) 0V (FAN) FAN +VO2 -VO2 +VO3 -VO3 +VO4 -VO4 CONN PIN CONN PIN CONN PIN CONN PIN Single Output Models Multiple Output Models Tumbler Technologies + TRUMPower 3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com