PWS2000 TEKTRONIX | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 4
Technical content
Tektronix PWS2000 Series Datasheet Features & Benefits Key Performance Specifications Linear Regulation Up to 72 VOutput Voltage 0.05% Basic Voltage Accuracy 0.2% Basic Current Accuracy 10 mV / 10 mA Programming Resolution Less than 3 mVp-p Ripple and Noise Available Functions and Features Bright Display
20 User-defined Setup Memories
Direct Parameter Entry using Numeric Keypad Vary Voltage or Current in User-selectable Steps 3-year Warranty
Generate the Voltage or Current You Need With wide current and voltage ranges, plus 10 mA and 10 mV resolution, the PWS2000 Power Supply Series satisfies a wide variety of applications. Low-noise linear regulation and 0.05% voltage accuracy provide the clean, accurate poweryou need at a price you can afford. Prevent Damaged Circuits and Lost Test Time The numeric keypad makes it easy to specify a precise current limit before you start your test. You can also specify a maximum output voltage, which will prevent accidental voltage settings above the specified maximum. Au s e r - d efinable password allows you to lock the front panel to prevent unwanted adjustments during critical tests. Designed to Make Your Work Easier The PWS2000 Power Supply Series is designed with the ease-of-use and familiar operation you expect from Tektronix. Intuitive Operation The direct-entry numeric keypad makes it easy to enter precise voltage and current values. Up/Down controls, with variable step-size, allow you vary output parameters in an increment size you choose. Bright Display A bright vacuumfluorescent display with large digits provides excellent readability at a distance, at an angle, or under dim lighting conditions. Setup Memories You can save your instrument settings in one of the 20 internal memory locations by simply pressing the Save button. Each time you want to recall that setting, just push the Recall button and choose the desired preset. Performance You Can Count On In addition to industry-leading service and support, the PWS2000 Power Supply Series comes backed with a three-year standard warranty. Characteristics Electrical Specifications PWS2185 PWS2323 PWS2326 PWS2721 DC Output Rating Voltage 0 to 18 V 0 to 32 V 0 to 32 V 0 to 72 V Current 0t o5A 0t o3A 0t o6A 0t o1 . 5A Load Regulation Current ≤0.1 % +2m At y p i c a l Line Regulation Voltage ≤0.1 % +5m V Current ≤0.1 % +2m At y p i c a l R i p p l ea n dN o i s e( 2 0H zt o7M H z ) Voltage ≤1m VRMS /3m V p-p Current ≤5m ARMS Setting Resolution Voltage 10 mV Current 10 mA Setting Accuracy, (25 °C ± 5 °C) Voltage ≤0.05% + 10 mV Current ≤0.2 % +1 0m A Readback Resolution Voltage 10 mV <20 V: 10 mV ≥20 V: 100 mV Current 10 mA Readback Accuracy, (25 °C ± 5 °C) Voltage ≤0.05% + 15 mV <20 V: ≤0.05% + 15 mV Current ≤0.1 % +1 5m A 2 www.tektronix.com
DC Power Supplies — Tektronix PWS2000 Series Display Vacuum fluorescent display. Memory 20 setup memories. Power Source Characteristic Description
110 V AC Setting 99 VRMS to 132 VRMS
220 V AC Setting 198 VRMS to 264 VRMS
Dimension mm in. With Boots and Handle Height 105 4.15 Width 242 9.52 Depth 384 15.12 Without Boots and Handle Height 91 3.57 Width 217 8.55 Depth 362 14.24 Weight kg lb. PWS2185 5.6 12.3 PWS2323/2326/2721 6.7 14.8 Environmental and Safety Characteristic Description Temperature Operating 0 °C to 40 °C Storage –20 °C to 70 °C Warm up 20 minutes to full uncertainty specifications Relative Humidity (Noncondensing) Operating and storage 5% to 95% relative humidity (% RH) at up to 40 °C, noncondensing Storage 5% to 95% RH at up to 40 °C, noncondensing 5% to 60% RH above 40 °C to 70 °C, noncondensing Altitude Operating 2,000 meters Storage 4,000 meters Floating Voltage Rating Up to 100 V (DC + peak AC) between earth ground and any output terminal Electromagnetic Compatibility European Union: EN 55011, Class A; IEC 61000-3-2; IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-4-8, IEC 61000-4-11 USA: FCC, CFR Title 47, Part 15, Subpart B, Class A Australia: EMC Framework, demonstrated per Emission Standard AS/NZS 2064 (Industrial, Scientific, and Medical Equipment) Safety European Union: Low Voltage directive 2006/95/EC; EN61010-1 2001 USA: Nationally recognized testing laboratory listing UL61010-1-2004 Canada: CAN/CSA C22.2 No. 61010-1 2004 Warranty Three years.
Ordering Information
PWS2185 DC Power Supply, 18 V, 5 A PWS2323 DC Power Supply, 32 V, 3 A PWS2326 DC Power Supply, 32 V, 6 A PWS2721 DC Power Supply, 72 V, 1.5 A PWS2000 Includes: Power supply, line cord, user manual, Traceable Certificate of Calibration, and documentation CD (includes user manuals and technical reference). Note:P l e a se specify power plug and preferred language option when ordering. Optional Accessories Accessory Description RMU2U*1 Rackmount Shelf Kit for 1 or 2 Units 386-7598-00 Rackmount Cosmetic Filler Panel *1 PWS2000 Series power supplies require 3U of height in a rack. The instrument's height, without boots, is 3 mm greater than 2U and 42 mm less than 3U. Filler panel 386-7598-00 may be used tofill the 42 mm gap above power supplies. Power Plug Options Option Description A0 North America A1 Universal Euro A2 United Kingdom A3 Australia A5 Switzerland A10 China A11 India A12 Brazil Language Options Option Description L0 English L1 French L2 Italian L3 German L4 Spanish L6 Portuguese L7 Simplified Chinese L8 Traditional Chinese L9 Korean L10 Russian www.tektronix.com 3
Datasheet Contact Tektronix: ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries+41 52 675 3777 Belgium 00800 2255 4835* Brazil + 5 5( 1 1 )3 7 5 97 6 2 7 Canada 1 800 833 9200 Central East Europe and the Baltics+ 4 15 26 7 53 7 7 7 Central Europe & Greece+41 52 675 3777 Denmark + 4 58 08 81 4 0 1 Finland + 4 15 2675 3777 France 00800 2255 4835* Germany 00800 2255 4835* Hong Kong400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835* Japan 81 (3) 6714 3010 Luxembourg + 4 15 26 7 53 7 7 7 Mexico, Central/South America & Caribbean52 (55) 56 04 50 90 Middle East,Asia, and North Africa+41 52 675 3777 The Netherlands00800 2255 4835* Norway 800 16098 People’s Republic of China400 820 5835 Poland +41 52 675 3777 Portugal 80 08 12370 Republic ofKorea 001 800 8255 2835 Russia & CIS+7 (495) 7484900 South Africa+ 4 15 26 7 53 7 7 7 Spain 00800 2255 4835* Sweden 00800 2255 4835* Switzerland 00800 2255 4835* Taiwan 886 (2) 2722 9622 United Kingdom & Ireland00800 2255 4835* USA 1 800 833 9200 * European toll-free number. If not accessible, call: +41 52 675 3777 Updated 10 February 2011 For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visitwww.tektronix.com Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.
25 Jul 2012 3GW-25620-4
C3 Calibration S e r v i c e3Y e a r s C5 Calibration S e r v i c e5Y e a r s CA1 Provides a single calibration event, or coverage for the designated calibration interval, whichever comesfirst D1 Calibration Data Report D3 Calibration Data Report 3 Years (with Option C3) D5 Calibration Data Report 5 Years (with Option C5) R5 Repair Service 5 Years SILV100 Standard Warranty Extended to 5 Years www.tektronix.com