DATASHEET SEARCH SITE | WWW.ALLDATASHEET.COM

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 8

Technical content

Single Phase AC Power Analyzer Datasheet The Tektronix PA1000 is a single-phase, single-channel power analysis solution that is optimized for fast, efficient, and accurate power consumption testing to international standards. It's compact size, DMM-like user-interface, graphical display, and powerful software enable users to quickly visualize, analyze, and document the power consumption efficiency of next-generation devices, including standby power measurements and harmonic analysis. Key specifications

1 MHz bandwidth

5 mW standby power measurements Harmonic analysis to the 50th order (standard) +/- 0.04% basic accuracy 20 µA to 20 Arms direct current input

1 V to 600 Vrms (Cat II) voltage input

USB, LAN, and GPIB interfaces (standard) Three-year warranty Essential power measurement tool for the R&D bench Harmonic analysis to IEC/EN 61000-3-2 / 4-7 (pre-compliance testing to the 50th order) Standby power analysis to IEC 62301 / EN 50564 (full compliance testing as low as 5mW) Supports additional testing to CE, EnergyStar, CEC, SPEC Power® 1, CQC-3146, NOM-32-ENER-2013, and more Transient analysis with 1M sample/sec continuous sampling Measure voltage, current, power, VA, WHr, THD, PF, CF, and more Convenient front-panel banana jack inputs, color graphical display, and PWRVIEW software to simplify usage and boost productivity Optional breakout test box simplifies AC line connections between your device under test and the PA1000

Applications

Power, energy, standby power, and harmonics measurements for: Power supply and UPS LED drivers / lighting Wireless charging Consumer electronics Home appliances Computers and IT equipment Inverters and converters Battery chargers

1 Spec Power® is a registered trademark of the Standard Performance Evaluation Corporation (SPEC)

www.tek.com 1

Complete power consumption analysis Most of today's AC-powered electronics and electro-mechanical products have government or consumer efficiency regulations they must meet. The PA1000 simplifies the process of proving that designs meet these requirements by offering a complete bench-top solution for single phase power consumption analysis. Use the standard front-panel input jacks and optional breakout box to simplify connections to the device under test, then analyze and document the results with the free PWRVIEW software. Easily and accurately measure harmonic performance, standby power, and more with the PA1000, optional breakout box, and free PWRVIEW software Visualize signals The color graphics display on the PA1000 provides intuitive readout of measured values, harmonic bar charts, waveforms, energy integration plots, and more. Setup is easy using the menu-driven interface and soft keys. Full color waveform display Harmonic bar chart display mode PA1000 Power Analyzer 2 www.tek.com

The PA1000 features harmonics analysis to the 50th order as a standard feature. Harmonics, THD, and related measurements can all be analyzed simultaneously with other power parameters. The PA1000's free PWRVIEW software enables: Viewing measurement data and system uncertainty in real-time, including waveforms, trend plots, and more. Creating and applying limits you define for simplified pass/fail testing of any parameter, including those based on user-defined math functions. Automating instrument setup, data collection, and report generation for key applications with just a few clicks using wizard-driven interfaces. Communicating with multiple PA1000 instruments for calculation of power efficiency and other parameters. IEC 61000-3-2 / 4-7 Current harmonics, pre-compliance testing. IEC 62301 / EN 50564 Standby power, full-compliance testing. IEC 61000-3-2 current harmonic testing IEC 62301 Standby power test with real time uncertainty and stability measurements Document results PWRVIEW software can automatically generate formatted test reports for IEC 61000-3-2 and 4-7 harmonics (pre-compliance) or IEC 62301 standby power (full compliance). These test reports include pass/fail results, data tables, graphs, and more… everything necessary to prove design performance and ensure a successful result in the compliance test lab. Datasheet www.tek.com 3

All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise. Available measurements Vrms - Volts RMS VTHD - Volts Total Harmonic Distortion VDF - Voltage distortion factor Arms - Amps RMS ATHD - Amps Total Harmonic Distortion ADF - Current distortion factor WATT - True power Z - Impedance VA - Apparent power R - Resistance VAR - Reactive power X - Reactance FREQ - Frequency HR - Integrator time PF - Power factor WHr - Watt Hours VPK+ - Volts peak (positive) VAHr - VA Hours VPK- - Volts peak (negative) VARHr - VAR Hours APK+ - Amps peak (positive) AHr - Amp Hours APK- - Amps peak (negative) Vh - Volts harmonics VDC - Volts DC Ah - Amps harmonics ADC - Amps DC VCF - Voltage crest factor ACF - Current crest factor Voltage and current ranges Voltage ranges 1000 Vpeak, 500 Vpeak, 200 Vpeak, 100 Vpeak, 50 Vpeak, 20 Vpeak, 10 Vpeak Current ranges (20 A shunt) 100 Apeak, 50 Apeak, 20 Apeak, 10 Apeak, 5 Apeak, 2 Apeak, 1 Apeak, 0.5 Apeak, 0.2 Apeak, 0.1 Apeak Measurement accuracy - voltage Voltage accuracy, VRMS (45 Hz to 850 Hz) ± 0.04% of Reading ± 0.04% of Range ± 0.005 V Voltage accuracy, VRMS (10 Hz to 45 Hz,

850 Hz to 1 MHz, typical)

± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± 0.05 V Voltage accuracy, DC (typical) ± 0.1% of Reading ± 0.1% of Range ± 0.05 V Effect of common mode (typical)

100 V, 100 kHz < 500 mV

Measurement accuracy - current Current accuracy, ARMS (45 Hz to 850 Hz) 2 ± 0.04% of Reading ± 0.04% of Range ± (1.8 µV/Zext) Current accuracy, ARMS (10 Hz to 45 Hz, ± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± (50 µV/Zext) Current accuracy, DC (typical) ± 0.1% of Reading ± 0.1% of Range ± (100 µV/Zext) 2 Offset is valid in low bandwidth, with internal shunts, and after a manual zero has been performed. Offset is 10 µV/Zext in high bandwidth, and with external shunt. PA1000 Power Analyzer 4 www.tek.com

accuracy (100 Apeak range) 2% of Range ± 20 mA Effect of common mode (typical)

100 V, 100 kHz, 20 A shunt < 15 mA

100 V, 100 kHz, 1 A shunt < 500 μA

100 V, 100 kHz, external shunt < 40 mV

Frequency (10 Hz to 20 kHz) 0.1% of Reading, with the peak of the signal extending 10% above and 10% below the DC level Frequency (20 kHz to 1 MHz) 0.1% of Reading, with the peak of the signal extending 25% above and 25% below the DC level Measurement accuracy - power Watts accuracy ± 0.075% of Reading ± 0.075% of Range (PF=1, 45 - 850 Hz) VA accuracy (Vrmsacc x Arms) + (Armsacc x Vrms) VAR accuracy (typical) VA ± VAerror 2 - W ± Werror 2 - VA2 - W2 PF Accuracy Cos θ -cos [ θ± ( Vh1 ph.err± Ah1 ph.err)] ± 0.002 Measurement accuracy - harmonic magnitude and phase (typical) Voltage harmonics magnitude (10 Hz to 480 kHz) ± 0.02% of Reading ± 0.1% of Range ± (0.04*F)% of Reading ± 0.05 V Voltage harmonics phase ± 0.04 ± [ 0.01 * (Vrange / Vreading )] ± (0.1 / Vrange) ± (0.005 *F ) Current harmonics magnitude (10 Hz to 480 kHz) ± 0.2% of Reading ± 0.1% of Range ± (0.04*F)% of Reading ± (50 µV / Zext) Current harmonics phase ± 0.04 ± [0.01 * (Arange / Areading)] ± (0.001 / Arange * Zext) ± (0.005 *F) Physical characteristics Dimensions mm in Height 102 4.0 Width 223 8.7 Depth 285 11.2 Weight Kg lb Net (without lead set) 3.2 7.0 Temperature C F Operating 0 °C to +40 °C +32 °F to +102 °F Nonoperating -20 °C to +60 °C -4 °F to +140 °F Notes: All stated accuracies are based upon a minimum of a 30-minute warm-up period. Zext is the external shunt impedance used and must be less than or equal to 10 Ohms. If no frequency is measured, then the signal is considered DC for the purpose of accuracy. F is the frequency measured in kHz. In the case of harmonics, F is the harmonic frequency. Specifications are valid from 1 to 100% of range in low bandwidth and after a manual zero has been performed. Values below 1% are typical. In high bandwidth, specifications are valid when the signal is greater than 10% of the range. Harmonic specifications are always valid when the harmonic is greater than 2% of the range. Measurement conditions during calibration: Instrument default settings unless otherwise stated, sine waves applied to V and I inputs, 30 min warm up, ambient temperature 23 °C ±5 °C. Datasheet www.tek.com 5

Ordering information

PA1000 Single-phase power analyzer Standard accessories Voltage lead set Country-specific power cord USB host-to-device interface cable Documentation CD Includes user manual in English, French, German, Spanish, Japanese, Portuguese, Simplified Chinese, Traditional Chinese, Korean, and Russian languages. Certificate of calibration Documents the traceability to National Metrology Institute(s) and ISO9001 Quality System Registration Three year product warranty Recommended accessories BB1000-NA Breakout box (North America plug configuration) BB1000-EU Breakout box (Europe plug configuration) BB1000-UK Breakout box (United Kingdom plug configuration) BALLAST-CT Specialty current transducer for lamp ballast testing CL200 Current clamp, 1 A - 200 A, for Tektronix Power Analyzers CL1200 Current clamp, 0.1 A - 1200 A, for Tektronix Power Analyzers PA-LEADSET Replacement lead set for Tektronix Power Analyzers (one channel lead set) BB1000-NA breakout box The Tektronix breakout box provides an easy way to make wiring connections between your device under test and the Tektronix power analyzer. Your device power cord plugs directly into the outlet on the breakout box (choose the version that best matches the connector style for your geography). Connection to the power analyzer is then simple, using the standard input lead set with 4 mm safety banana connectors that are provided as a standard accessory with the power analyzer. PA1000 Power Analyzer 6 www.tek.com

Opt. A0 North America power plug (115 V, 60 Hz) Opt. A1 Universal Euro power plug (220 V, 50 Hz) Opt. A2 United Kingdom power plug (240 V, 50 Hz) Opt. A3 Australia power plug (240 V, 50 Hz) Opt. A4 North America power plug (240 V, 50 Hz) Opt. A5 Switzerland power plug (220 V, 50 Hz) Opt. A6 Japan power plug (100 V, 50/60 Hz) Opt. A10 China power plug (50 Hz) Opt. A11 India power plug (50 Hz) Opt. A12 Brazil power plug (60 Hz) Opt. A99 No power cord Service options Opt. C3 Calibration Service 3 Years Opt. C5 Calibration Service 5 Years Opt. D1 Calibration Data Report Opt. D3 Calibration Data Report 3 Years (with Opt. C3) Opt. D5 Calibration Data Report 5 Years (with Opt. C5) Opt. R5 Repair Service 5 Years (including warranty) Opt. R5DW Repair Service Coverage 5 Years (includes product warranty period). 5-year period starts at time of instrument purchase Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar. Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats. Datasheet www.tek.com 7

ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777 Belgium 00800 2255 4835* Brazil +55 (11) 3759 7627 Canada 1 800 833 9200 Central East Europe and the Baltics +41 52 675 3777 Central Europe & Greece +41 52 675 3777 Denmark +45 80 88 1401 Finland +41 52 675 3777 France 00800 2255 4835* Germany 00800 2255 4835* Hong Kong 400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835* Japan 81 (3) 6714 3010 Luxembourg +41 52 675 3777 Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90 Middle East, Asia, and North Africa +41 52 675 3777 The Netherlands 00800 2255 4835* Norway 800 16098 People's Republic of China 400 820 5835 Poland +41 52 675 3777 Portugal 80 08 12370 Republic of Korea +822 6917 5084, 822 6917 5080 Russia & CIS +7 (495) 6647564 South Africa +41 52 675 3777 Spain 00800 2255 4835* Sweden 00800 2255 4835* Switzerland 00800 2255 4835* Taiwan 886 (2) 2656 6688 United Kingdom & Ireland 00800 2255 4835* USA 1 800 833 9200 * European toll-free number. If not accessible, call: +41 52 675 3777 price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.

23 Jan 2016 55W-29535-3

www.tek.com