N9067C KEYSIGHT | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 12

Technical content

X-Series Measurement App, Multi-Touch N9067C Technical Overview – Automatically synchronize to pulse modulated signals for radar and electronic warfare (EW) applications up to 50 GHz and bandwidth of 510 MHz – Verify all key pulse signal modulation performance indicators relating to power, droop, overshoot, ripple, time (rise/fall/width/PRI), frequency, phase, and FM modulation using the comprehensive pulse table result metrics – Integrates with popular real-time instrument functionality (UXA, PXA, MXA) such as frequency mask trigger – Select from five X-Series signal analyzers with multi-touch to meet your specific design and test goals – Use multi-touch front panel user interface or SCPI remote control – Extend test assets with transportable licenses between all X-Series signal analyzers with multi-touch UI

.csv for custom reports and post analysis. of performance achieved by the hardware you select. information to actionable insight. Start with a 30-day free trial. Figure 1. The N9067C pulse analysis measurement application provides a user interface designed to make quick

Figure 2. Simultaneous views of pulse signals versus time.

Figure 3. Simultaneous views of pulse magnitude, real-time response and cumulative statistics. measurements has never been easier.

06 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview Measurement Summary Category Measurements Result/Parameter Amplitude Amplitude error time, amplitude meas time, amplitude ref time Phase Phase error time, phase meas time, phase ref time FM FM error spectrum, FM error time, FM meas spectrum, FM meas time, FM ref spectrum, FM ref time Pulse result table Level (in dBm, unless otherwise noted) Top level, base level, top/base ratio (dB), on level, peak level, mean level, peak to average (dB) Droop Droop (%), droop (dB), droop rate (dB/us), droop start (dBm), droop stop (dBm) Overshoot Overshoot (%), overshoot (dB) Ripple Ripple (%), ripple (dB) Time (in seconds, unless otherwise noted) Rise time, rising edge, fall time, falling edge, width, off time, PRI,PRF (Hz), duty cycle (%) Frequency (in Hz, unless otherwise noted) Freq mean, freq pulse-pulse difference, freq pk-pk deviation, freq error time, freq error peak, freq error peak location (sec) Phase (in degrees, unless otherwise noted) Phase mean, phase pulse-pulse difference, phase pk-pk deviation, phase error rms, phase error peak, phase error peak location (sec) Best-fit FM (in Hz, unless otherwise noted) Best-fit FM mean, best-fit FM start, best-fit FM stop, best-fit FM pk-pk deviation, best-fit FM slope (Hz/µsec), best-fit FM INL (%, integral non linearity) Current statistics table Same as above results with Min, Max, RMS, Average, Std Dev, Count Cumulative statistics table Same as above results with Min, Max, RMS, Average, Std Dev, Count, Median, Mode

07 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview Key Specifications Nominal definition Nominal values are designated with the abbreviation “nom.” These values indicate expected performance, or describe product performance that is useful in the application of the product, but is not covered by the product warranty. Note: data is subject to change. Accuracy - Amplitude and timing PXA (N9030B) UXA (N9040B) Description Nominal Nominal Top level 1 ±0.2 dB + Absolute amplitude accuracy (CW) ±0.2 dB + Absolute amplitude accuracy + IF frequency response (chirp) On level 1 ±0.1 dB + Absolute amplitude accuracy (CW) ±0.1 dB + Absolute amplitude accuracy + IF frequency response (chirp) Mean level 1 ±0.1 dB + Absolute amplitude accuracy (CW) ±0.1 dB + Absolute amplitude accuracy + IF frequency response (chirp) Peak level 1 ±0.2 dB + Absolute amplitude accuracy (CW) ±0.2 dB + Absolute amplitude accuracy + IF frequency response (chirp) Width 1 ±1/Sample rate PRI 1 ±1/Sample rate Accuracy - Frequency and phase CW (non-chirp signal) PXA (N9030B) UXA (N9040B) Frequency error RMS 2,3 CF 2 GHz Option B40 ±1.6 kHz ±1.6 kHz Option B85 ±3.5 kHz Option B1X ±8.3 kHz Option B2X ±21 kHz Option B5X ±62 kHz CF 10 GHz Option B40 ±3.5 kHz ±3.5 kHz Option B85 ±5.4 kHz Option B1X ±13 kHz Option B2X ±31 kHz (±18 kHz 4) Option B5X ±87 kHz (±110 kHz 4) CF 20 GHz Option B40 ±5.9 kHz ±5.9 kHz Option B85 ±5.9 kHz Option B1X ±5.9 kHz Option B2X ±52 kHz (±17 kHz Option B5X ±150 kHz (±88 kHz 4) 1. SNR ≥ 30 dB, pulse width ≥ 100/bandwidth 2. ATT = 0 dB. IF gain = low, signal condition with pulse on power = -10 dBm and pulse width ≥ 100/bandwidth, modulation setup with FM filter bandwidth = 10% 3. Frequency/phase analysis setup width = 50% 4. Pulse to pulse analysis setup: reference time = center, offset = 0.0 s, window length = 0.0 s

08 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview PXA (N9030B) UXA (N9040B) Frequency pulse to pulse difference 1,2 CF 2 GHz Option B40 ±3.3 kHz ±3.3 kHz Option B85 ±7.1 kHz Option B1X ±17 kHz Option B2X ±47 kHz Option B5X ±130 kHz CF 10 GHz Option B40 ±7.4 kHz ±7.4 kHz Option B85 ±12 kHz Option B1X ±26 kHz Option B2X ±62 kHz (±22 kHz 2) Option B5X ±180 kHz (±56 kHz 2) CF 20 GHz Option B40 ±13 kHz ±13 kHz Option B85 ±23 kHz Option B1X ±52 kHz Option B2X ±110 kHz (±30 kHz 2) Option B5X ±290 kHz (±68 kHz 2) Phase pulse to pulse difference 1,2 CF 2 GHz Option B40 ±0.3° ±0.3° Option B85 ±0.3° Option B1X ±0.3° Option B2X ±0.3° Option B5X ±0.6° CF 10 GHz Option B40 ±0.7° ±0.7° Option B85 ±0.7° Option B1X ±0.7° Option B2X ±0.45° (±0.3° Option B5X ±0.6° (±0.3° 2 CF 20 GHz Option B40 ±1.3° ±1.3° Option B85 ±1.3° Option B1X ±1.3° Option B2X ±0.75° (±0.55° Option B5X ±1.1° (±0.55° 2) 1. ATT = 0 dB. IF gain = low, signal condition with pulse on power = -10 dBm and pulse width ≥ 100/bandwidth, modulation setup with FM filter bandwidth = 10% 2. Pulse to pulse analysis setup: reference time = center, offset = 0.0 s, window length = 0.0 s

09 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview Accuracy - Frequency and Phase Chirp (Linear Chirp Signal) PXA (N9030B) UXA (N9040B) Frequency Error RMS 1,2 CF 2 GHz Option B40 ±3.5 kHz ±3.5 kHz Option B85 ±6 kHz Option B1X ±15 kHz Option B2X ±21 kHz Option B5X ±87 kHz CF 10 GHz Option B40 ±5 kHz ±5 kHz Option B85 ±12 kHz Option B1X ±25 kHz Option B2X ±31 kHz Option B5X ±87 kHz CF 20 GHz Option B40 ±7.5 kHz ±7.5 kHz Option B85 ±20 kHz Option B1X ±35 kHz Option B2X ±55 kHz Option B5X ±150 kHz Frequency pulse to pulse difference 1,3 CF 2 GHz Option B40 ±3.5 kHz ±3.5 kHz Option B85 ±7. 3 kHz Option B1X ±17 kHz Option B2X ±47 kHz Option B5X ±140 kHz CF 10 GHz Option B40 ±7 kHz ±7 kHz Option B85 ±15 kHz Option B1X ±30 kHz Option B2X ±70 kHz Option B5X ±200 kHz CF 20 GHz Option B40 ±12 kHz ±12 kHz Option B85 ±25kHz Option B1X ±60 kHz Option B2X ±125 kHz Option B5X ±320 kHz 1. ATT = 0 dB. IF gain = low, signal condition with pulse on power = -10 dBm and pulse width ≥ 100/bandwidth, modulation setup with FM filter bandwidth = 10% 2. Frequency/phase analysis setup width = 50% 3. Pulse to pulse analysis setup: reference time = center, offset = 0.0 s, window length = 0.0 s

10 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview Accuracy - frequency and phase chirp (linear chirp signal), continued Phase pulse to pulse difference 1,2 PXA (N9030B) UXA (N9040B) CF 2 GHz Option B40 ±0.35° ±0.35° Option B85 ±0.35° Option B1X ±0.4° Option B2X ±0.3° Option B5X ±0.45° CF 10 GHz Option B40 ±0.7° ±0.7° Option B85 ±0.7° Option B1X ±0.9° Option B2X ±0.5° Option B5X ±0.6° CF 20 GHz Option B40 ±1° ±1° Option B85 ±1.2° Option B1X ±1.5° Option B2X ±1.0° Option B5X ±1.2° For a complete list of specifications, refer to the appropriate specifications guide: UXA: www.keysight.com/find/uxa_specifications PXA: www.keysight.com/find/pxa_specifications MXA: www.keysight.com/find/mxa_specifications EXA: www.keysight.com/find/exa_specifications CXA: www.keysight.com/find/cxa_specifications 1. ATT = 0 dB. IF gain = low, signal condition with pulse on power = -10 dBm and pulse width ≥ 100/bandwidth, modulation setup with FM filter bandwidth = 10% 2. Pulse to pulse analysis setup: reference time = center, offset = 0.0 s, window length = 0.0 s

11 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview Pulse analysis measurement application Model-Option Description, license type N9067C-1FP 1 Pulse analysis measurement application, fixed perpetual license N9067C-1TP 1 Noise figure measurement transportable perpetual 1. N9067C application requires an X-Series signal analyzer with multi-touch user interface. For more information, see hardware configurations.

Ordering Information

Software Licensing and Instrument Configuration Choose from two license types:: – Fixed, perpetual license: This allows you to run the application in the X-Series analyzer in which it is initially installed. – Transportable, perpetual license: This allows you to run the application in the X-Series analyzer in which it is initially installed, plus it may be transferred from one multi-touch X-Series analyzer to another. For more information, please visit www.keysight.com/find/n9067c You Can Upgrade! Options can be added after your initial purchase. All of our X-Series application options are license-key upgradeable. UP GRADE Hardware Configuration For optimizing noise figure measurements with noise figure measurement application, Keysight recommends a minimum level of X-Series multi-touch signal analyzer hardware functionality at each instrument performance point. Supported analyzers include: – UXA N9040B – PXA N9030B – MXA N9020B – EXA N9010B – CXA N9000B Capability Instrument Option Benefit Analysis bandwidth 25 MHz minimum (-B25) or wider Required: Wider analysis bandwidth to support enhanced sample interval, time resolu - tion, rise/fall time and minimum pulse width Microwave pre-selector by-pass -MPB Required : When input frequency is > 3.6 GHz on bandwidths 40 MHz or greater Real-time spectrum analyzer -RT1 or –RT2 Required: Detect signals as short as 3.33 ns with 100% POI Frequency mask trigger -FT1 or –FT2 Required: Focus on the signal of interest in a complex signal environment Pre-amplifier 3.6 GHz (-P03) or higher Recommended: For maximizing the measurement sensitivity Fine resolution step attenuator -FSA on EXA and CXA only Recommended: Useful for maximizing useable dynamic range to see signals

12 | Keysight | N9067C Pulse Analysis X-Series Measurement App, Multi-touch - Technical Overview This information is subject to change without notice. © Keysight Technologies, 2015 - 2016 Published in USA, February 19, 2016 5992-1384EN www.keysight.com www.keysight.com/find/n9067c myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939.

1939 THE FUTURE

For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-02-10-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System