UXA X-Series Signal Analyzer, Multi-touch N9040B 3 Hz to 8.4, 13.6, 26.5, 44, or 50 GHz Data Sheet
02 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet This data sheet is a summary of the specifications and conditions for the UXA signal analyzers. For the complete specifications guide, visit: www.keysight.com/find/uxa_specifications Table of Contents
03 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Definitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to temperature ranges 0 to 55 °C, unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2 s) of performance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 °C . In addition to the statistical observations of a sample of instruments, these values include the effects of the uncer - tainties of external calibration references. These values are not warranted. These values are updated occasionally if a significant change in the statistically observed behavior of production instruments is observed. Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specifications that 80 percent of the units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C. Typi - cal performance does not include measurement uncertainty . Nominal values indicate expected performance, or describe product performance that is useful in the application of the product, but is not covered by the product warranty. The analyzer will meet its specifications when: – The analyzer is within its calibration cycle. – Under auto couple control, except that Auto Sweep Time Rules = Accy. – For signal frequencies < 10 MHz, DC coupling applied . – The analyzer has been stored at an ambient temperature within the allowed operating range for at least two hours before being turned on, if it had previously been stored at a temperature range inside the allowed storage range but outside the allowed operating range. – The analyzer has been turned on at least 30 minutes with Auto Align set to Normal, or if Auto Align is set to Off or Partial, alignments must have been run recently enough to prevent an Alert message. If the Alert condition is changed from “Time and Temperature” to one of the disabled duration choices, the analyzer may fail to meet specifications without informing the user. If Auto Align is set to Light, performance is not warranted, and nominal performance will degrade to become a factor of 1.4 wider for any specification subject to alignment, such as amplitude tolerances . – The term “mixer level” is used as a condition for many specifications in this document. This term is a conceptual quantity that is defined as follows: Mixer Level (dBm) = RF Input Power Level (dBm) - (Electronic + Mechanical) Attenuation (dBm) For instruments with option H1G, all standard instrument specifications apply for 0 to 40 °C, except as noted in document N9040-90026. Maximum operating temperature range is 40 °C when using the 1 GHz IF path.
04 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency and Time Specifications Frequency range DC coupled AC coupled Option 508 3 Hz to 8.4 GHz 10 MHz to 8.4 GHz Option 513 3 Hz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 3 Hz to 26.5 GHz 10 MHz to 26.5 GHz Option 544 3 Hz to 44 GHz NA Option 550 3 Hz to 50 GHz NA Frequency band LO multiple (N) Frequency range 0 1 3 Hz to 3.6 GHz 1 1 3.5 to 8.4 GHz 2 2 8.3 to 13.6 GHz 3 2 13.5 to 17.1 GHz 4 4 17 to 26.5 GHz 5 4 26.4 to 34.5 GHz 6 8 34.4 to 50 GHz Frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate ± 3 x 10–8 / year Temperature stability Full temperature range ± 4.5 x 10–9 Achievable initial calibration accuracy ± 3.1 x 10–8 Example frequency reference accuracy 1 year after last adjustment = ± 6.6 x 10–8 Residual FM (Center frequency = 1 GHz
10 Hz RBW, 10 Hz VBW)
≤ (0.25 Hz x N) p-p in 20 ms nominal See band table above for N (LO multiple) Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.10% x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution 1) Marker frequency counter Accuracy ± (marker frequency x frequency reference accuracy + 0.100 Hz) Delta counter accuracy ± (delta frequency x frequency reference accuracy + 0.141 Hz) Counter resolution 0.001 Hz Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept FFT ± (0.1% x span + horizontal resolution) ± (0.1% x span + horizontal resolution) 1 . Horizontal resolution is span/(sweep points –1).
05 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency and Time Specifications (continued) Sweep time and triggering Range Span = 0 Hz Span ≥ 10 Hz 1 μs to 6000 s 1 ms to 4000 s Accuracy Span ≥ 10 Hz, swept Span ≥ 10 Hz, FFT Span = 0 Hz ± 0.01% nominal ± 40% nominal ± 0.01% nominal Sweep trigger Free run, line, video, external 1, external 2, RF burst, periodic timer Trigger Delay Span = 0 Hz or FFT Span ≥ 10 Hz, swept Resolution –150 to +500 ms 0 to 500 ms 0.1 μs Time gating Gate methods Gated LO; gated video; gated FFT Gate length range (except method = FFT) 1 μs to 5.0 s Gate delay range 0 to 100.0 s Gate delay jitter 33.3 ns p-p nominal Sweep (trace) point range All spans 1 to 40,001 Resolution bandwidth (RBW) Range (–3.01 dB bandwidth) 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz Bandwidth accuracy (power) RBW range 1 Hz to 100 kHz 110 kHz to 1.0 MHz (< 3.6 GHz CF) 1.1 to 2 MHz (< 3.6 GHz CF) 2.2 to 3 MHz (< 3.6 GHz CF) 4 to 8 MHz (< 3.6 GHz CF) ± 0.07 dB nominal 0 to -0.2 dB nominal 0 to -0.4 dB nominal Bandwidth accuracy (–3.01 dB) RBW range 1 Hz to 1.3 MHz ± 2% nominal Selectivity (–60 dB/–3 dB) 4.1:1 nominal EMI bandwidth (CISPR compliant) 200 Hz, 9 kHz, 120 kHz, 1 MHz (Option EMC required) EMI bandwidth (MIL STD 461E compliant) 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz (Option EMC required) Analysis bandwidth 1 Maximum bandwidth Option B25 (standard) Option B40 Option B2X Option B5X Option H1G
25 MHz
40 MHz
255 MHz
510 MHz
1 GHz
Video bandwidth (VBW) Range 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz) Accuracy ± 6% nominal (in swept mode and zero span) 1 . Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or processing in the time, frequency, or modulation domain .
06 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Amplitude Accuracy and Range Specifications Amplitude range Measurement range Displayed average noise level (DANL) to +30 dBm (for preamp Off) DANL to +24 dBm (for frequency opts ≤ 526 with preamp On) DANL to +20 dBm (for frequency opts > 526 with preamp On) Input mechanical attenuator range (3 Hz to 50 GHz) 0 to 70 dB in 2 dB steps Electronic attenuator (Option EA3) Frequency range 3 Hz to 3.6 GHz Attenuation range Electronic attenuator range Full attenuation range (mechanical + electronic) 0 to 24 dB, 1 dB steps 0 to 94 dB, 1 dB steps Maximum safe input level Average total power (with and without preamp) +30 dBm (1 W) Peak pulse power (< 10 μs pulse width, < 1% duty cycle, and input attenuation ≥ 30 dB) +50 dBm (100 W) DC volts DC coupled AC coupled ± 0.2 Vdc ± 100 Vdc (For frequency Option 508, 513, or 526) Display range Log scale 0.1 to 1 dB/division in 0.1 dB steps 1 to 20 dB/division in 1 dB steps (10 display divisions) Linear scale 10 divisions Scale units dBm, dBmV, dBμV, dBmA, dBμA, V, W, A Frequency response Specifications 95th percentile (≈ 2 s) (10 dB input attenuation, 20 to 30 °C, preselector centering applied above 3.6 GHz) RF/MW (Option 508, 513, 526)
3 Hz to 10 MHz
50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz
17.0 GHz to 22 GHz
22.0 to 26.5 GHz ± 0.46 dB ± 0.35 dB ± 0.35 dB ± 1.5 dB ± 1.5 dB ± 2.0 dB ± 2.0 dB ± 2.0 dB ± 2.5 dB ± 0.19 dB ± 0.14 dB ± 0.50 dB ± 0.42 dB ± 0.51 dB ± 0.57 dB ± 0.65 dB ± 0.87 dB mmW (Option 544, 550)
3 Hz to 20 MHz
50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 22.0 to 26.5 GHz 26.4 to 34.5 GHz 33.4 to 50 GHz ± 0.46 dB ± 0.35 dB ± 0.35 dB ± 1.7 dB ± 1.5 dB ± 1.5 dB ± 2.0 dB ± 2.0 dB ± 2.0 dB ± 2.5 dB ± 2.5 dB ± 3.2 dB ± 0.20 dB ± 0.16 dB ± 0.69 dB ± 0.42 dB ± 0.42 dB ± 0.39 dB ± 0.54 dB ± 0.62 dB ± 0.59 dB ± 0.93 dB ± 1.28 dB
07 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Amplitude Accuracy and Range Specifications (continued) Frequency response Specifications 95th percentile (≈ 2 s) Preamp on (0 dB attenuation) RF/MW (Option P08, P13, P26) 9 kHz to 1 MHz 1 to 50 MHz 50 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz ± 0.68 dB ± 0.55 dB ± 2.0 dB ± 2.3 dB ± 2.5 dB ± 3.0 dB ± 3.5 dB ± 0.38 dB ± 0.32 dB ± 0.28 dB ± 0.64 dB ± 0.69 dB ± 0.84 dB ± 1.13 dB ± 1.48 dB mmW (Option P44, P50) 9 kHz to 1 MHz 1 to 50 MHz 50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz ± 0.68 dB ± 0.60 dB ± 2.0 dB ± 2.0 dB ± 2.3 dB ± 2.5 dB ± 3.0 dB ± 3.5 dB ± 3.0 dB ± 4.1 dB ± 0.45 dB ± 0.27 dB ± 0.29 dB ± 0.75 dB ± 0.52 dB ± 0.61 dB ± 0.61 dB ± 0.73 dB ± 0.63 dB ± 1.11 dB ± 1.47 dB Input attenuation switching uncertainty Specifications Supplemental information Relative to 10 dB and preamp off At 50 MHz (reference frequency) attenuation 12 to 40 dB attenuation 2 to 8 dB attenuation 0 dB ± 0.14 dB ± 0.18 dB ± 0.04 dB typical ± 0.06 dB typical ± 0.05 dB nominal Attenuation > 2 dB 3 Hz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 26.5 GHz 26.4 to 50 GHz ± 0.3 dB nominal ± 0.5 dB nominal ± 0.7 dB nominal ± 0.7 dB nominal ± 1.0 dB nominal Total absolute amplitude accuracy Specifications (10 dB attenuation, 20 to 30°C, 1 Hz ≤ RBW ≤ 1 MHz, input signal -10 to -50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any reference level, any scale, s = nominal standard deviation) At 50 MHz At all frequencies 0.01 to 3.6 GHz ± 0.24 dB ± (0.24 dB + frequency response) ± 0.16 dB (95th Percentile approx. 2 s) Preamp on (Option P08, P13, P26, P44, P50) At all frequencies ± (0.36 dB + frequency response) Input voltage standing wave ratio (VSWR) 95th percentile (10 dB input attenuation)
50 MHz
10 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz Freq Opt 508, 513, 526 1.07 nominal 1.101 1.278 1.341 1.58 1.60 NA NA Freq Opt 544, 550 1.025 nominal 1.116 1.144 1.158 1.258 1.233 1.363 1.55
08 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Amplitude Accuracy and Range Specifications (continued) Input voltage standing wave ratio (VSWR) 95th percentile Freq Opt 508, 513, 526 Freq Opt 544, 550 Preamp on (Option P08, P13, P26, P44, or P50) (0 dB input attenuation) 10 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz 1.56 1.47 1.57 1.72 1.70 NA NA 1.40 nominal 1.53 1.389 1.316 1.337 1.42 1.62 Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 1.5 MHz RBW 1.6 MHz to 2.7 MHz RBW
3 MHz RBW
4, 5, 6, 8 MHz RBW ± 0.03 dB ± 0.05 dB ± 0.10 dB ± 0.30 dB Reference level Range Log scale Linear scale –170 to +30 dBm in 0.01 dB steps 707 pV to 7.07 V with 0.11% (0.01 dB) resolution Accuracy 0 dB1 Display scale switching uncertainty Switching between linear and log 0 dB1 Log scale/div switching 0 dB1 Display scale fidelity Between –10 dBm and –18 dBm input mixer level ± 0.10 dB total ± 0.04 dB typical Below –18 dBm input mixer level ± 0.07 dB ± 0.02 dB typical Trace detectors Standard Normal, peak, sample, negative peak, log power average, RMS average, and voltage average With Option EMC Add quasi-peak to above Preamplifier Frequency range 2 Option P08 Option P13 Option P26 Option P44 Option P50 9 kHz to 8.4 GHz 9 kHz to 13.6 GHz 9 kHz to 26.5 GHz 9 kHz to 44 GHz 9 kHz to 50 GHz Gain 9 kHz to 3.6 GHz 3.6 to 26.5 GHz 3.6 to 50 GHz +20 dB nominal +35 dB nominal (for freq opts ≤ 526) +40 dB nominal (for freq opts > 526) 1 . Only affects the display, not the measurement, so it causes no additional error in measurement results from trace data or markers . 2 . Below 100 kHz, only 95th percentile (approx. 2s) value for frequency response is provide
09 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications 1 dB gain compression (two-tone) Maximum power at input mixer (At 1 kHz RBW with 100 kHz tone spacing) Preamp Off 20 to 40 MHz
40 MHz to 2 GHz
2 to 26.5 GHz 26.5 to 50 GHz 2 dBm nominal 5 dBm nominal 10 dBm nominal 0 dBm nominal Preamp On 10 MHz to 3.6 GHz 3.6 to 26.5 GHz Tone spacing 100 kHz to 20 MHz Tone spacing > 70 MHz Freq Opt ≤ 526 Freq Opt > 526 26.5 to 50 GHz –14 dBm nominal –28 dBm nominal –10 dBm nominal –20 dBm nominal –30 dBm nominal Displayed average noise level (DANL) 1 Specifications Typical (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 1 Hz RBW, 20 to 30 °C) RF/MW (Option 508, 513, 526) LNP Off/LNP On 2 LNP Off/LNP On 2 Preamp Off 3 to 10 Hz 10 to 100 Hz
100 Hz to 1 kHz
10 MHz to 1.2 GHz 1.2 to 2.1 GHz 2.1 to 3.0 GHz 3.0 to 3.6 GHz 3.5 to 4.2 GHz 4.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 16.9 GHz 16.9 to 20.0 GHz 20.0 to 26.5 GHz –141 dBm/NA –150 dBm/NA –155 dBm/NA –155 dBm/NA –153 dBm/NA –152 dBm/NA –151 dBm/NA –149 dBm/–154 dBm –150 dBm/–155 dBm –149 dBm/–155 dBm –145 dBm/–152 dBm –143 dBm/–151 dBm –136 dBm/–148 dBm –100 dBm/NA nominal –125 dBm/NA nominal –130 dBm/NA nominal –137 dBm/NA nominal –146 dBm/NA typical –155 dBm/NA typical –157 dBm/NA typical –156 dBm/NA typical –155 dBm/NA typical –153 dBm/NA typical –152 dBm/NA typical –152 dBm/–155 dBm typical –152 dBm/–156 dBm typical –151 dBm/–156 dBm typical –147 dBm/–155 dBm typical –146 dBm/–154 dBm typical –139 dBm/–151 dBm typical Preamp On 3 100 to 200 kHz 200 to 500 kHz 0.5 to 1 MHz 1 to 10 MHz 10 MHz to 2.1 GHz 2.1 to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 16.9 GHz 16.9 to 20.0 GHz 20.0 to 26.5 GHz –152 dBm –155 dBm –159 dBm –161 dBm –165 dBm –163 dBm –164 dBm –163 dBm –161 dBm –159 dBm –155 dBm –159 dBm –161 dBm –164 dBm –166 dBm –166 dBm –164 dBm –166 dBm –165 dBm –163 dBm –161 dBm –158 dBm 1 . With Option NF2 (Noise Floor Extension) “Off”. 2 . LNP (Low Noise Path) is standard for the UXA. 3 . At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp.
10 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications (continued) Displayed average noise level (DANL) with Noise Floor Extension (Option NF2) on 95th percentile DANL improvement Preamp Off Preamp On LNP On Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 9 dB 10 dB 10 dB 9 dB 9 dB 10 dB 9 dB 10 dB 9 dB 8 dB NA 10 dB 10 dB 10 dB 9 dB DANL with Noise Floor Extension Preamp Off Preamp On LNP On Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 -163 dBm -162 dBm -162 dBm -159 dBm -148 dBm -174 dBm -174 dBm -174 dBm -172 dBm -166 dBm NA -166 dBm -167 dBm -165 dBm -162 dBm Displayed average noise level (DANL )1 Specifications Typical (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 1 Hz RBW, 20 to 30 °C) mmW (Option 544, 550) LNP Off/LNP On 2 LNP Off/LNP On 2 Preamp Off 3 to 10 Hz 10 to 100 Hz 10 MHz to 1.2 GHz 1.2 to 2.1 GHz 2.1 to 3.0 GHz 3.0 to 3.6 GHz 3.5 to 4.2 GHz 4.2 to 6.6 GHz 6.6 to 13.6 GHz 13.5 to 14 GHz 14 to 17 GHz 17 to 22.5 GHz 22.5 to 26.5 GHz 26.4 to 34 GHz 33.9 to 37 GHz 37 to 40 GHz 40 to 46 GHz 46 to 49 GHz 49 to 50 GHz –141 dBm/NA –150 dBm/NA –154 dBm/NA –153 dBm/NA –151 dBm/NA –150 dBm/NA –149 dBm/NA –145 dBm/–151 dBm –144 dBm/–152 dBm –147 dBm/–153 dBm –144 dBm/–150 dBm –145 dBm/–151 dBm –141 dBm/–149 dBm –139 dBm/–146 dBm –138 dBm/–146 dBm –134 dBm/–142 dBm –132 dBm/–141 dBm –130 dBm/–141 dBm –130 dBm/–139 dBm –128 dBm/–139 dBm –95 dBm/NA nominal –114 dBm/NA nominal –128 dBm/NA nominal –136 dBm/NA nominal –144 dBm/NA typical –154 dBm/NA typical –156 dBm/NA typical –155 dBm/NA typical –153 dBm/NA typical –152 dBm/NA typical –151 dBm/NA typical –148 dBm/–154 dBm typical –148 dBm/–154 dBm typical –149 dBm/–155 dBm typical –148 dBm/–153 dBm typical –148 dBm/–153 dBm typical –146 dBm/–152 dBm typical –143 dBm/–150 dBm typical –143 dBm/–150 dBm typical –140 dBm/–148 dBm typical –139 dBm/–146 dBm typical –137 dBm/–146 dBm typical –137 dBm/–145 dBm typical –135 dBm/–145 dBm typical 1 . With Option NF2 (Noise Floor Extension) “Off”. 2 . LNP (Low Noise Path) is standard for the UXA.
11 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications (continued) Specifications Typical mmW (Option 544, 550) LNP Off/LNP O n2 LNP Off/LNP O n2 LNP Off/LNP O n2 Preamp On 1 100 to 200 kHz 200 to 500 kHz 0.5 to 1 MHz 1 to 2.1 GHz 2.1 to 3.6 GHz 3.5 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 20.0 GHz 20.0 to 26.5 GHz 26.4 to 30 GHz 30 to 34 GHz 33.9 to 37 GHz 37 to 40 GHz 40 to 46 GHz 46 to 50 GHz –157 dBm –159 dBm –162 dBm –164 dBm –162 dBm –161 dBm –161 dBm –160 dBm –158 dBm –157 dBm –155 dBm –153 dBm –152 dBm –149 dBm –146 dBm –159 dBm typical –161 dBm typical –164 dBm typical –165 dBm typical –164 dBm typical –162 dBm typical –164 dBm typical –163 dBm typical –161 dBm typical –160 dBm typical –159 dBm typical –158 dBm typical –156 dBm typical –155 dBm typical –152 dBm typical Displayed average noise level (DANL) with Noise Floor Extension (Option NF2) on 95th percentile DANL improvement Preamp Off Preamp On LNP On Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 Band 5 Band 6 10 dB 8 dB 8 dB 9 dB 10 dB 11 dB 11 dB 9 dB 9 dB 8 dB 8 dB 8 dB 8 dB 7 dB NA 9 dB 9 dB 10 dB 11 dB 11 dB 11 dB DANL with Noise Floor Extension Preamp Off Preamp On LNP On Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 Band 5 Band 6 -160 dBm -154 dBm -154 dBm -154 dBm -146 dBm -140 dBm -172 dBm -165 dBm -167 dBm -167 dBm -164 dBm -160 dBm -154 dBm NA -157 dBm -157 dBm -157 dBm -157 dBm -155 dBm -150 dBm 1 . At higher frequency bands (beyond 3 .6 GHz), Preamp “On” supersedes “LNP enabled ”. LNP cannot operate simultaneously with preamp.
12 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications (continued) Residuals, images, and spurious responses Residual responses (Input terminated and 0 dB attenuation) 200 kHz to 8.4 GHz Zero span or FFT or other frequencies –100 dBm –100 dBm nominal Image responses Tuned Freq (f) Excitation Freq Response RF/MW (Opt 508, 513, 526) mmW (Opt 544, 550) (Mixer level at –10 dBm) 10 MHz to 26.5 GHz 10 MHz to 3.6 GHz 10 MHz to 3.6 GHz 3.5 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22 GHz 22 to 26.5 GHz f+45 MHz f+10,245 MHz f+645 MHz f+645 MHz f+645 MHz f+645 MHz f+645 MHz –80 dBc –80 dBc –80 dBc –78 dBc –74 dBc –70 dBc –66 dBc –105 dBc typical –106 dBc typical –101 dBc typical –86 dBc typical –84 dBc typical –78 dBc typical –75 dBc typical –80 dBc –80 dBc –80 dBc –80 dBc –80 dBc –80 dBc –70 dBc –104 dBc typical –106 dBc typical –101 dBc typical –106 dBc typical –106 dBc typical –101 dBc typical –102 dBc typical (Mixer level at –30 dBm) 26.5 to 50 GHz 26.5 to 34.5 GHz 34.4 to 42 GHz 42 to 50 GHz f+45 MHz f+645 MHz f+645 MHz f+645 MHz –70 dBc –60 dBc –90 dBc nominal –98 dBc typical –84 dBc typical –75 dBc nominal Other spurious responses Mixer level Response Carrier frequency ≤ 26.5 GHz First RF order (f ≥ 10 MHz from carrier) Higher RF order (f ≥ 10 MHz from carrier) –10 dBm –40 dBm –80 dBc + 20log(N 2) Including IF feedthrough, LO harmonic mixing responses –80 dBc + 20log(N 2) Including higher order mixer responses Carrier frequency > 26.5 GHz (f ≥ 10 MHz from carrier) –30 dBm –90 dBc nominal LO-related spurious responses (200 Hz ≤ f < 10 MHz from carrier) Line-related spurious responses –10 dBm –68 dBc 1 + 20log(N 2) –73 dBc1 + 20log (N2) (nominal) Second harmonic distortion (SHI) Source frequency Mixer level Distortion (LNP Off/LNP On) SHI (LNP Off/LNP On) RF/MW (Opt 508, 513, 526) 10 MHz to 1.8 GHz 1.75 to 2.5 GHz 2.5 to 4 GHz 4 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –60 dBc/NA –77 dBc/–95 dBc –77 dBc/–101 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc +45 dBm/NA +62 dBm/+80 dBm +62 dBm/+86 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm mmW (Opt 544, 550) 10 MHz to 1.8 GHz 1.75 to 2.5 GHz 2.5 to 4 GHz 4 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz 13.25 to 25 GHz –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –60 dBc/NA –72 dBc/–95 dBc –72 dBc/–99 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc –65 dBc/–105 dBc (nom) +45 dBm/NA +57 dBm/+80 dBm +57 dBm/+84 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm +50/+90 dBm (nom) Source frequency Preamp level Distortion SHI Preamp On (Option P08, P13, P26, P44, P50) 10 MHz to 1.8 GHz 1.8 to 13.25 GHz 13.25 to 25 GHz –45 dBm –50 dBm –50 dBm –78 dBc nominal –60 dBc nominal –50 dBc nominal +33 dBm nominal +10 dBm nominal 0 dBm nominal 1 . Nominally –40 dBc under large magnetic (0.38 Gauss rms) or vibrational (0.21 g rms) environmental stimuli. 2 . N is the LO multiplication factor. Refer to page 3 for the N value verses frequency ranges.
13 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications (continued) Third-order intermodulation distortion (TOI) (two –16 dBm tones at input mixer with tone separation > 5 times IF prefilter bandwidth, 20 to 30 °C) RF/MW (Opt 508, 513, 526) 10 to 300 MHz 300 to 600 MHz 0.6 to 1.5 GHz 1.5 to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz +13.5 dBm +18 dBm +20 dBm +21 dBm +19 dBm +19 dBm +18 dBm +19 dBm +16 dBm typical +21 dBm typical +22 dBm typical +23 dBm typical +22 dBm typical +23 dBm typical +23 dBm typical +24 dBm typical mmW (Opt 544, 550) 10 to 300 MHz 300 to 600 MHz 0.6 to 1.5 GHz 1.5 to 3.6 GHz 3.5 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.5 to 50 GHz +13.5 dBm +18 dBm +20 dBm +21 dBm +16 dBm +13 dBm +13 dBm +16 dBm typical +21 dBm typical +22 dBm typical +23 dBm typical +23 dBm typical +17 dBm typical +20 dBm typical +13 dBm nominal Preamp On (Option P08, P13, P26, P44, P50) Tones at preamp input (two –45 dBm) (two –45 dBm) (two –50 dBm) 10 to 500 MHz 500 MHz to 3.6 GHz 3.6 to 26.5 GHz +4 dBm nominal +4.5 dBm nominal –15 dBm nominal
1 MHz
10 MHz
1 . For wide reference loop bandwidth. Figure 1. Nominal UXA phase noise at various center frequencies. 50 GHz curve is the predicted phase noise computed from the 25.2 GHz observation.
15 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet General Specifications Temperature range Operating 1 Storage 0 to 55 °C –40 to +70 °C Altitude 4,500 meters (approx. 15,000 feet) EMC Complies with the essential requirements of the European EMC Directive as well as current editions of the following standards (dates and editions are cited in the Declaration of Conformity): – IEC/EN 61326-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11 – ICES/NMB-001 This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada. South Korean Class A EMC declaration This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home. A 급 기기 (업무용 방송통신기자재)이 기 기는 업무용 (A 급) 전자파적합기기로서 판 매자 또는 사용자는 이 점을 주 의하시기 바라 며, 가 정외의 지역에서 사용 하는 것을 목적으 로 합니다. Safety Complies with the essential requirements of the European Low Voltage Directive as well as current editions of the following standards (dates and editions are cited in the Declaration of Conformity): – IEC/EN 61010-1 – Canada: CSA C22.2 No. 61010-1 – ·USA: UL std no. 61010-1 Acoustic statement (European Machinery Directive) Acoustic noise emission LpA < 70 dB Operator position Normal operation mode per ISO 7779 Acoustic noise – more information Values given are per ISO 7779 standard in the “Operator Sitting” position Ambient temperature < 35 °C Nominally under 55 dBA Sound Pressure. 55 dBA is generally considered suitable for use in quiet office environment ≥ 35 °C Nominally under 65 dBA Sound Pressure. 65 dBA is generally considered suitable for use in noisy office environment Environmental stress Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environ - mental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3. Power requirements Voltage and frequency 100 to 120 V, 50/60/400 Hz 220 to 240 V, 50/60 Hz Power consumption On Standby
850 W (Maximum)
1 . Operating temperature range when option H1G is installed is 0 to 40 °C .
16 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet General Specifications (continued) Display Resolution Size 1280 x 800 357 mm (14.1 in.) diagonal (nominal) capacitive multi-touch screen Data storage Internal Removable solid state drive (≥ 80 GB) and secure digital (SD) memory device External Supports USB 3.0/2.0 compatible memory devices Weight (Basic configuration) Net Shipping 30.9 kg (68 lbs) nominal 39.5 kg (87 lbs) nominal Dimensions Height Width Length 280 mm (11 in) 459 mm (18 in) 500 mm (19.8 in) Warranty The UXA signal analyzer is supplied with a 3-year standard warranty Calibration cycle The recommended calibration cycle is one year. Calibration services are available through Keysight service centers
17 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Inputs and Outputs Front panel RF input connector Standard (for Opt 508, 513, 526) Standard (for Opt 544, 550) Option C35 (with Option 526 only) Type-N female, 50 Ω nominal 2.4 mm male, 50 Ω nominal APC 3.5 mm male, 50 Ω nominal Probe power Voltage/current +15 Vdc, ± 7% at 150 mA max nominal –12.6 Vdc, ± 10% at 150 mA max nominal USB ports Master (3 ports) Standard Connector Output current Compatible with USB 2.0 USB Type-A female
0.5 A nominal
Headphone jack Miniature stereo audio jack (3.5 mm, also known as “1/8 inch”) External mixing Connection port Connector Impedance Functions Mixer bias range IF center frequency ≤ 25 MHz IF path
40 MHz BW IF path
255 MHz BW IF path
510 MHz BW IF path
SMA, female 50 Ω nominal Triplexed for mixer bias, IF input and LO output ± 10 mA in 10 uA step
322.5 MHz
250.0 MHz
750.0 MHz
877 .1484375 MHz 3.75 to 14.1 GHz Rear panel
10 MHz out
BNC female, 50 Ω nominal ≥ 0 dBm nominal
10 MHz + (10 MHz x frequency reference accuracy)
BNC female, 50 Ω nominal –5 to 10 dBm nominal 1 to 50 MHz nominal (selectable to 1 Hz resolution) ± 2 x 10–6 of specified external reference input frequency Trigger 1 and 2 inputs Connector Impedance Trigger level range BNC female > 10 kΩ nominal –5 to +5 V (TTL) factory preset Trigger 1 and 2 outputs Connector Impedance Level BNC female 50 Ω nominal 0 to 5 V (CMOS) nominal Sync (reserved for future use) Connector BNC female Monitor output 1 Connector Format Resolution VGA compatible, 15-pin mini D-SUB XGA (60 Hz vertical sync rates, non-interlaced) Analog RGB 1280 x 800 Monitor output 2 Connector Resolution Mini DisplayPort 1280 x 800
18 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Noise source drive +28 V (pulsed) Connector Output voltage BNC female On 28.0 ± 0.1 V (60 mA maximum) Off < 1 V SNS series noise source For use with the Agilent/Keysight SNS Series noise sources Digital bus Connector MDR-80 Analog out Connector BNC female USB ports Master (3 ports) Standard Connector Output current Slave (1 port) Standard Connector Two ports (stacked with each other) are compatible with USB 3.0; one (stacked with LAN port) with USB 2.0 USB Type-A female Compatible with USB 3.0 USB Type-B female GPIB interface Connector GPIB codes GPIB mode IEEE-488 bus connector SH1, AH1, T6, SR1, RL1, PP0, DC1, C1, C2, C3, C28, DT1, L4, C0 Controller or device LAN TCP/IP interface Standard Connector 1000Base-T RJ45 Ethertwist IF output Connector Impedance SMA female, shared by the second IF out (wideband, standard) and Opt CRP, and ALV 50 Ω nominal 2nd IF output Center frequency SA mode or I/Q analyzer with IF BW ≤ 25 MHz with Option B40 with Option B2X with Option B5X Conversion gain Bandwidth Low band IF Path ≤ 40 MHz IF Path 255 MHz IF Path 510 MHz IF Path 1 GHz High band, with preselector bypassed
250 MHz
750 MHz
877 .1484375 MHz 1 dB nominal Up to 160 MHz nominal
255 MHz nominal
510 MHz nominal
1 GHz nominal
Up to 800 MHz (nominal); expandable to 1200 MHz with corrections IF2 output for 1 GHz analog IF Connector Impedance Center frequency SMA female 50 Ω nominal IF2 input for 1 GHz digital section Connector Impedance Center frequency SMA female 50 Ω nominal Trigger 3 input for 1 GHz digitizer Connector Impedance Trigger level range Trigger channel passband BNC female 50 Ω, DC terminated ± 5 V range (minimum amplitude 0.5 V pk-pk) DC to 2 GHz nominal Inputs and Outputs (continued)
19 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Other Optional Outputs Option ALV log video out General port specifications Connector Impedance SMA female Shared with other options 50 Ω nominal Fast log video output Output voltage Maximum Slope Open-circuit voltages shown
1.6 V at –10 dBm nominal
25 ± 1 mV/dB nominal Log fidelity Range Accuracy within range 49 dB (nominal) with input frequency at 1 GHz ± 1.0 dB nominal Rise time 15 ns nominal Fall time Bands 1-4 with Option MPB Other cases 40 ns nominal best case Depends on bandwidth Option CRP programmable IF output General port specifications Connector Impedance SMA female Shared with other options 50 Ω nominal Programmable IF output Center frequency Range Resolution 10 to 75 MHz (user selectable)
0.5 MHz
Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Output at 70 MHz Low band or high band with preselector bypassed Preselected band
100 MHz (nominal)
Depends on RF center frequency Lower output frequencies Subject to folding Residual output signals ≤ –88 dBm (nominal)
20 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Other Optional Outputs (continued) Option YAV Y-axis video output General port specifications Connector Impedance BNC female Shared with other options 50 Ω nominal Screen video Operating conditions Display scale types Log scales Modes Gating Log or Lin All (0.1 to 20 dB/div) Spectrum analyzer only Gating must be off “Lin” is linear in voltage Output scaling Offset Gain accuracy 0 to 1.0 V open circuit, representing bottom to top of screen ± 1% of full scale nominal ± 1% of output voltage nominal Log video (Log envelope) output Amplitude range (terminated with 50 Ω) Maximum Scale factor Bandwidth Operating conditions V nominal for –10 dBm at the mixer 1 V per 192.66 dB Set by RBW Select Sweep Type = Swept Linear video output Amplitude range (terminated with 50 Ω) Maximum Minimum
1.0 V nominal for signal envelope at the reference level
Scale factor If carrier level is set to half the reference level in volts, the scale factor is 200% of carrier level per volt. Regardless of the carrier level, the scale factor is 100% of reference level per volt. Bandwidth Operating conditions Set by RBW Select Sweep Type = Swept
21 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer Frequency Frequency span Option B25 (standard) Option B40 Option B2X Option B5X Option H1G
10 Hz to 25 MHz
10 Hz to 40 MHz
10 Hz to 255 MHz
10 Hz to 510 MHz
10 Hz to 1 GHz
Resolution bandwidth (spectrum measurement) Range Overall Span = 1 MHz Span = 10 kHz Span = 100 Hz Window shapes 100 mHz to 3 MHz
50 Hz to 3 MHz
1 Hz to 10 kHz
Flat Top, Uniform, Hanning, Hamming, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB) Analysis bandwidth (waveform measurement) Option B25 (standard) Option B40 Option B2X Option B5X Option H1G IF frequency response (standard 10 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency) Frequency (GHz) Span (MHz) Preselector Max error Midwidth error (95th percentile) Slope (dB/MHz) (95th percentile) RMS (nominal) ≤ 3.6 3.6 to < 26.5 ≥ 26.5 ≤ 10 ≤ 10 ≤ 10 NA Off12 Off12 ± 0.20 dB ± 0.25 dB ± 0.30 dB ± 0.12 dB ± 0.12 dB ± 0.12 dB ± 0.10 dB ± 0.10 dB ± 0.10 dB 0.02 dB 0.02 dB 0.024 dB IF phase linearity Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.02, < 3.6 ≥ 3.6 ≤ 10 ≤ 10 NA Off 1 0.14° 0.27° 0.032° 0.057° Dynamic range (standard 10 MHz IF path) Clipping-to-noise dynamic range Excluding residuals and spurious responses Clipping level at mixer IF gain = Low IF gain = High –10 dBm –20 dBm Center frequency ≥ 20 MHz –8 dBm nominal –17.5 dBm nominal Noise density at mixer at center frequency (DANL + IF Gain effect) + 2.25 dB 1 . MPB (microwave preselector bypass) is enabled. All UXAs ship with MPB as a standard feature.
22 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer (continued) Data acquisition (standard 10 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tool Data packing With 89600 VSA or fast capture 32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pairs) Sample rate IQ pairs 1.25 x IFBW ADC resolution 16 bits IF frequency response (standard 25 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency) Freq (GHz) Span (MHz) Preselector Max error Midwidth error (95th percentile) Slope (dB/MHz) (95th percentile) RMS (nominal) < 3.6 3.6 to < 26.5 ≥ 26.5 ≤ 25 ≤ 25 ≤ 25 NA Off 1 Off 1 ± 0.30 dB ± 0.40 dB ± 0.40 dB ± 0.12 dB ± 0.12 dB ± 0.1 dB ± 0.1 dB 0.02 dB 0.03 dB 0.02 dB IF phase linearity Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.02, < 3.6 ≥ 3.6 ≤ 25 ≤ 25 NA Off 1 0.41° 1.0° 0.11° 0.27° Dynamic range (standard 25 MHz IF path) Full scale (ADC clipping) Default settings, signal at CF (IF gain = Low) Band 0 Bands 1 through 4 –8 dBm mixer level nominal –7 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Bands 1 through 6 –18 dBm mixer level nominal, subject to gain limitations –17 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal 1 . MPB (microwave preselector bypass) is enabled. All UXAs ship with MPB as a standard feature.
23 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer (continued) Data acquisition (standard 25 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tool Data packing With 89600 VSA or fast capture 32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pairs) Sample rate IQ pairs 1.25 x IF BW ADC resolution 16 bits Option B40 40 MHz analysis bandwidth (Option B40 is automatically included in Option B2X, B5X, or H1G) IF frequency response (40 MHz IF path) IF frequency response (relative to center) Freq (GHz) Span (MHz) Preselector Specification Typical RMS (nominal) ≥ 0.03, < 3.6 ≥ 3.6, ≤ 8.4 > 8.4, ≤ 26.5 > 26.5, ≤ 34.4 > 34.4 ≤ 40 ≤ 40 ≤ 40 ≤ 40 ≤ 40 NA Off 1 Off 1 Off 1 Off 1 ± 0.37 dB ± 0.5 dB ± 0.7 dB ± 0.8 dB ± 1 dB ± 0.22 dB ± 0.13 dB ± 0.14 dB ± 0.25 dB ± 0.35 dB 0.07 dB 0.05 dB 0.05 dB 0.07 dB 0.07 dB IF phase linearity Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.02, < 3.6 ≥ 3.6 ≤ 40 ≤ 40 NA Off 1 0.36° 1.0° 0.083° 0.24° Dynamic range (40 MHz IF path) SFDR (Spurious-free dynamic range) Signal frequency within ± 12 MHz of center Signal frequency anywhere within analysis BW Spurious response within ± 18 MHz of center Response anywhere within analysis BW –80 dBc nominal –79 dBc nominal –77 dBc nominal Full scale (ADC clipping) Mixer level Default settings, signal at CF (IF gain = Low) Band 0 Bands 1 through 4 Bands 5 through 6 RF/MW (Opt 508, 513, 526) –8 dBm nominal –6 dBm nominal mmW (Opt 544, 550) –8 dBm nominal –7 dBm nominal –7 dBm nominal High gain setting, signal at CF (IF gain = High) Band 0 Bands 1 through 2 Bands 3 through 4 Bands 5 through 6 subject to gain limitations –16 dBm nominal –9 dBm nominal –6 dBm nominal –12 dBm nominal –16 dBm nominal –16 dBm nominal –15 dBm nominal Effect of signal frequency ≠ CF Up to ± 4 dB nominal 1 . MPB (microwave preselector bypass) is enabled. All UXAs ship with MPB as a standard feature.
24 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer (continued) Data acquisition (40 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tool Data packing With 89600 VSA software or fast capture 32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pairs) Sample rate IQ pairs 1.25 x IF BW ADC resolution 12 bits Option B2X 255 MHz analysis bandwidth (Option B2X is automatically included with Option B5X or H1G) IF frequency response (255 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Specification Typical RMS (nominal) ≥ 0.4, < 3.6 > 3.6, ≤ 8.4 > 8.4 ≤ 255 ≤ 255 ≤ 255 NA Off 1 Off 1 ± 0.74 dB ± 0.82 dB ± 0.3 dB ± 0.34 dB ± 0.8 dB nominal 0.1 dB 0.1 dB 0.2 dB IF phase linearity (255 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Pk-to-pk (nominal) RMS (nominal) ≥ 0.4, < 3.6 ≥ 3.6, < 26.5 ≥ 26.5 ≤ 255 ≤ 255 ≤ 255 NA Off 1 Off 1 0.6° 0.5° 0.8° Dynamic range (255 MHz IF path) Spurious-free dynamic range (SFDR) Anywhere within the analysis BW –78 dBc nominal Full scale (ADC clipping) Mixer level Default setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 Bands 5 through 6 RF/MW (Opt 508, 513, 526) +2 dBm nominal +4 dBm nominal +4 dBm nominal mmW (Opt 544, 550) +3 dBm nominal +3 dBm nominal +1 dBm nominal +1 dBm nominal High gain setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 Bands 5 through 6 -4 dBm nominal +2 dBm nominal +4 dBm nominal -1 dBm nominal -4 dBm nominal -6 dBm nominal -5 dBm nominal Effect of signal frequency ≠ CF Up to ± 4 dB nominal IF residual responses across the full BW Band 0 Band 1 Preselector off 1 –110 dBFS nominal –108 dBFS nominal Third-order intermodulation distortion (Two tones of equal level, 1 MHz separation, each tone -23 dB relative to full scale (ADC clipping), IF gain = high) Band 0 Bands 1 through 4 Bands 5 through 6 Preselector off 1 Preselector off 1 –85 dBc nominal –85 dBc nominal –80 dBc nominal 1 . MPB (microwave preselector bypass) is enabled. All UXA ship with MPB as a standard feature
25 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Option B5X 510 MHz analysis bandwidth IF frequency response (510 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Specification Typical RMS (nominal) ≥ 0.6, < 3.6 > 3.6, ≤ 8.4 > 8.4, ≤ 26.5 > 26.5 ≤ 510 ≤ 510 ≤ 510 ≤ 510 NA Off 1 Off 1 Off 1 ± 1.0 dB ± 1.25 dB ± 0.41 dB ± 0.42 dB ± 0.8 dB nominal ± 1 dB nominal 0.06 dB 0.3 dB IF phase linearity (510 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Pk-to-pk (nominal) RMS (nominal) ≥ 0.4, < 3.6 ≥ 3.6, < 26.5 ≥ 26.5 ≤ 510 ≤ 510 ≤ 510 NA Off Off 1.4° 1.6° Dynamic range (510 MHz IF path) Spurious-free dynamic range (SFDR) Anywhere within the analysis BW –78 dBc nominal Full scale (ADC clipping) Mixer level Default setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 Bands 5 through 6 RF/MW (Opt 508, 513, 526) +2 dBm nominal +2 dBm nominal +2 dBm nominal mmW (Opt 544, 550) +2.5 dBm nominal +3.5 dBm nominal +1 dBm nominal +1 dBm nominal High gain setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 Bands 5 through 6 –3 dBm nominal 0 dBm nominal +2 dBm nominal –1 dBm nominal –7 dBm nominal –9 dBm nominal –9 dBm nominal Noise density Band Frequency (GHz) 1.80 6.00 10.80 15.15 21.80 30.50 42.25 IF gain = Low –144 dBm/Hz –141 dBm/Hz –140 dBm/Hz –137 dBm/Hz –135 dBm/Hz –130 dBm/Hz –130 dBm/Hz IF gain = High –145 dBm/Hz –142 dBm/Hz –141 dBm/Hz –137 dBm/Hz –135 dBm/Hz –130 dBm/Hz –130 dBm/Hz Data acquisition (255 MHz IF path) Time record length IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing
89600 VSA or fast capture
Length (IQ sample pairs) 1073 MSa (230 Sa) 536 MSa (229 Sa) 4 GB total memory (Option DP4) Maximum IQ capture time (89600 VSA and fast capture) Length of IQ sample pairs/sample rate (IQ pairs) Sample rate (IQ pairs) Minimum of (1.25 x IFBW, 300 Msa/s) ADC resolution 14 bits I/Q Analyzer (continued) 1 . MPB (microwave preselector bypass) is enabled. All UXA ship with MPB as a standard feature
26 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Option H1G 1 GHz analysis bandwidth IF frequency response ( 1 GHz IF path) Center Freq (GHz) Span (MHz) Preselector Max Error (nominal) ≥ 0.7, < 3.6 > 3.6, ≤ 8.4 > 8.4, ≤ 26.5 > 26.5 ≤ 1000 ≤ 1000 ≤ 1000 ≤ 1000 NA Off 1 Off 1 Off 1 0.7 dB 0.7 dB 1.0 dB 1.25 dB IF phase linearity (510 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Pk-to-pk (nominal) RMS (nominal) ≥ 0.7, < 3.6 > 3.6 ≤ 1000 ≤ 1000 NA Off1 1.5° 1.3° Effect of signal frequency ≠ CF Up to ± 4 dB nominal IF residual responses across the full BW Band 0 Band 1 Preselector off 1 –110 dBFS nominal –108 dBFS nominal Third-order intermodulation distortion (Two tones of equal level, 1 MHz separation, each tone -23 dB relative to full scale (ADC clipping), IF gain = high) Band 0 Bands 1 through 4 Bands 5 through 6 Preselector off 1 Preselector off 1 –85 dBc nominal –82 dBc nominal –79 dBc nominal Noise density Band Frequency (GHz) 1.80 6.00 10.80 15.15 21.80 30.50 42.25 IF gain = Low –144 dBm/Hz –140 dBm/Hz –140 dBm/Hz –137 dBm/Hz –135 dBm/Hz –130 dBm/Hz –130 dBm/Hz IF gain = High –144 dBm/Hz –142 dBm/Hz –141 dBm/Hz –137 dBm/Hz –135 dBm/Hz –130 dBm/Hz –130 dBm/Hz Data acquisition (510 MHz IF path) Time record length IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing Length (IQ sample pairs) IFBW ≤255.176 MHz IFBW >255.176 MHz
1073 MSa (230 Sa)
2,147 MSa (230 Sa)
536 MSa (229 Sa)
4 GB total memory
8 GB total memory (Option DP4)
(89600 VSA and fast capture) Length of IQ sample pairs/sample rate (IQ pairs) Sample rate (IQ pairs) Minimum of (1.25 x IFBW, 300 Msa/s) ADC resolution 14 bits I/Q Analyzer (continued) 1 . MPB (microwave preselector bypass) is enabled. All UXA ship with MPB as a standard feature
27 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic range ( 1 GHz IF path) Center frequency Suprious-free dynamic range (SFDR) anywhere within the analysis BW < 3.1 GHz ≥ 3.1 GHz -62 dBc nominal -56 dBc nominal Effect of signal frequency ≠ CF Up to ± 4 dB nominal IF residual responses across the full BW 2 IF gain = Low Band 0 Band 1 Preselector off 1 –67 dBFS nominal –69 dBFS nominal Noise density (preselector off above band 0) Band Frequency (GHz) 1.80 6.00 10.80 15.15 21.80 30.5 42.25 IF gain = High (nominal) –152 dBm/Hz –153 dBm/Hz –151 dBm/Hz –151 dBm/Hz –149 dBm/Hz –147 dBm/Hz –142 dBm/Hz Data acquisition ( 1 GHz IF path) Time record length IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools 32-bit data packing 89600 VSA or fast capture IF bandwidth Length (IQ sample pairs)
1 GHz ≥ IFBW > 500 MHz
500 MHz ≥ IFBW > 250 MHz
250 MHz ≥ IFBW > 125 MHz
125 MHz ≥ IFBW > 62.5 MHz
62.5 MHz ≥ IFBW > 40 MHz
838,859,979 to 419,429,990 838,859,979 to 419,429,990 838,859,979 to 419,429,990 838,859,979 to 419,429,990 838,859,979 to 419,429,990 Maximum IQ capture time (89600 VSA and fast capture) Length of IQ sample pairs/sample rate (IQ pairs) Sample rate (IQ pairs) 1.25 x IFBW ADC resolution 12 bits I/Q Analyzer (continued) 1 . MPB (microwave preselector bypass) is enabled. All UXA ship with MPB as a standard feature. 2 . The residual performance is dominated by a single residual 50 MHz to the left of the center of screen .
28 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet Real-time Spectrum Analyzer (RTSA) Option RT1 real-time spectrum analyzer, basic detection, or RT2 real-time spectrum analyzer, optimal detection Real-time analysis Real-time analysis bandwidth Option RT1 Option RT2 Minimum detectable signal duration with > 60 dB StM 1 ratio Option RT1 Option RT2 Minimum signal duration with 100% probability of intercept (POI) at full amplitude accuracy Option RT1 Option RT2 Minimum acquisition time FFT rate Supported Detectors Number of Traces Number of Markers Supported Markers Supported triggers Up to 509.47 MHz Up to 509.47 MHz 11.42 ns 3.33 ns 17.17 μs 3.51 μs 100 μs 292,969/s Peak, Negative Peak, Sample, Average Normal, Delta, Noise, Band Power Level, Level with Time Qualified (TQT), Line, External, RF burst, Frame, Frequency Mask (FMT), FMT with TQT Analysis BW option determines the max real-time BW (max 255 MHz with B2X or H1G, max 510 MHz with B5X) For Frequency Mask Triggering (FMT) Signal is at mask level Signal is at mask level, span > 85 MHz 1 . “StM” = “Signal-to-Mask” Option RTS Real-time I/Q Data Streaming Real-time streaming 1 Output stream resolution IQ streaming bandwidth Electrical interface 16-bit I + jQ Sample rate varies continuously based on RTSA span setting Max IQ streaming bandwidth and sample rate B1X B2X, B5X, or H1G
160 MHz
200 Msamples/s
300 Msamples/s
Supported data recorder X-COM Systems IQC5255B Capture time Data tagging < 3 hours at 255 MHz bandwidth Event markers, IRIG-B GPS 1 . Use with X-COM Systems IQC5255B data recorder to capture rare events and play back at RF using integrated control software on the UXA. Related Literature UXA Brochure, 5992-0089EN UXA Configuration Guide, 5992-0043EN UXA Specifications Guide, N9040-90002
29 | Keysight | N9040B UXA X-Series Signal Analyzer, Multi-touch - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2014 - 2016 Published in USA, August 10, 2016 5992-0090EN www.keysight.com myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Keysight Infoline www.keysight.com/find/Infoline Keysight’s insight to best in class information management . Free access to your Keysight equipment company reports and e-library . KEYSIGHT SERVICES Keysight Services www.keysight.com/find/services Our deep offering in design, test, and measurement services deploys an industry-leading array of people, processes, and tools . The result? We help you implement new technologies and engineer improved processes that lower costs. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products . Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements . Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/uxa For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt . 2 (FR) Opt . 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-06-08-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc . DEKRA Certified ISO 9001:2015 Quality Management System Evolving Our unique combination of hardware, software, support, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight