N9030B KEYSIGHT | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 30
Technical content
PXA X-Series Signal Analyzer, Multi-touch N9030B 3 Hz to 3.6, 8.4, 13.6, 26.5, 44, or 50 GHz Data Sheet
2 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Table of Contents Accelerate signal insight with outstanding all-around signal analysis The PXA is the benchmark for performance that accelerates innovation in demanding applica - tions. With measurement op - tions that range from excellent to exceptional, the PXA puts you in the lead. Analyze the latest signals with up to 510 MHz analysis bandwidth and better than 78 dBc SFDR, and reveal previously hidden signals with Noise Floor Extension (NFE). To see your device’s true behav- ior, get industry-leading phase noise performance by adding the Keysight-proprietary DDS-based LO. Simplify migration from legacy Agilent/HP spectrum analyzers with backward code compatibility and compact 4U form-factor. This data sheet is a summary of the specifications and conditions for PXA signal analyzers. For the complete specifications guide, visit: www.keysight.com/find/pxa_specifications
3 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Definitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to temperature ranges 0 to 55 °C, unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2 s) of performance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the statistical observations of a sample of instruments, these values include the effects of the uncertainties of external calibration references. These values are not warranted. These values are updated occasionally if a significant change in the statistically observed behavior of production instruments is observed. Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specifications that 80 percent of the units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C. Typical performance does not include measure - ment uncertainty. Nominal values indicate expected performance, or describe product perfor - mance that is useful in the application of the product, but is not covered by the product warranty. The analyzer will meet its specifications when: – The analyzer is within its calibration cycle. – Under auto couple control, except that Auto Sweep Time Rules = Accy. – For signal frequencies < 10 MHz, DC coupling applied. – The analyzer has been stored at an ambient temperature within the allowed operating range for at least two hours before being turned on, if it had previously been stored at a temperature range inside the allowed storage range but outside the allowed operating range. – The analyzer has been turned on at least 30 minutes with Auto Align set to Normal, or if Auto Align is set to Off or Partial, alignments must have been run recently enough to prevent an Alert message. If the Alert condition is changed from “Time and Temperature” to one of the disabled duration choices, the analyzer may fail to meet specifications without informing the user. If Auto Align is set to Light, performance is not warranted, and nominal performance will degrade to become a factor of 1.4 wider for any specification subject to alignment, such as amplitude tolerances. The term “mixer level” is used as a condition for many specifications in this document. This term is a conceptual quantity that is defined as follows: Mixer Level (dBm) = RF Input Power Level (dBm) - (Electronic + Mechanical) Attenua - tion (dBm).
4 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency range DC coupled AC coupled Option 503 3 Hz to 3.6 GHz 10 MHz to 3.6 GHz Option 508 3 Hz to 8.4 GHz 10 MHz to 8.4 GHz Option 513 3 Hz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 3 Hz to 26.5 GHz 10 MHz to 26.5 GHz Option 544 3 Hz to 44 GHz NA Option 550 3 Hz to 50 GHz NA Band LO multiple (N) 0 1 3 Hz to 3.6 GHz 1 1 3.5 to 8.4 GHz 2 2 8.3 to 13.6 GHz 3 2 13.5 to 17.1 GHz 4 4 17 to 26.5 GHz 5 4 26.4 to 34.5 GHz 6 8 34.4 to 50 GHz Precision frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate ± 1 x 10–7 / year ± 1.5 x 10–7 / 2 years Temperature stability 20 to 30 °C Full temperature range ± 1.5 x 10–8 ± 5 x 10–8 Achievable initial calibration accuracy ± 4 x 10–8 Example frequency reference accuracy 1 year after last adjustment 20 to 30 °C = ± (1 x 1 x 10–7 + 1.5 x 10–8 + 4 x 10–8) = ± 1.55 x 10–7 Residual FM Center frequency = 1 GHz
10 Hz RBW, 10 Hz VBW
≤ (0.25 Hz x N) p-p in 20 ms nominal See band table above for N (LO multiple) Frequency reference (Option EP0) Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate ± 3 x 10–8 / year Temperature stability Full temperature range ± 4.5 x 10–9 Achievable initial calibration accuracy ± 3.1 x 10–8 Example frequency reference accuracy 1 year after last adjustment = ± 6.6 x 10–8 Residual FM Center frequency = 1 GHz ≤ (0.25 Hz x N) p-p in 20 ms nominal See band table above for N (LO multiple) Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.10% x span + 5% x RBW + 2 Hz + 0.5 x horizontal resolution 1) Marker frequency counter Accuracy ± (marker frequency x frequency reference accuracy + 0.100 Hz) Delta counter accuracy ± (delta frequency x frequency reference accuracy + 0.141 Hz) Counter resolution 0.001 Hz 1. Horizontal resolution is span/(sweep points –1). Frequency and Time Specifications
5 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept FFT ± (0.1% x span + horizontal resolution) ± (0.1% x span + horizontal resolution) Sweep time and triggering Range Span = 0 Hz Span ≥ 10 Hz 1 μs to 6000 s 1 ms to 4000 s Accuracy Span ≥ 10 Hz, swept Span ≥ 10 Hz, FFT Span = 0 Hz ± 0.01% nominal ± 40% nominal ± 0.01% nominal Sweep trigger Free run, line, video, external 1, external 2, RF burst, periodic timer Trigger Delay Span = 0 Hz or FFT Span ≥ 10 Hz, swept Resolution –150 to +500 ms 0 to 500 ms 0.1 μs Time gating Gate methods Gate length range (except method = FFT) Gate delay range Gate delay jitter Gated LO; gated video; gated FFT 1 μs to 5.0 s 0 to 100.0 s 33.3 ns p-p nominal Sweep (trace) point range All spans 1 to 40001 Resolution bandwidth (RBW) Range (–3.01 dB bandwidth) 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz Bandwidth accuracy (power) RBW range
1 Hz to 100 kHz
110 kHz to 1.0 MHz (< 3.6 GHz CF) 1.1 to 2 MHz (< 3.6 GHz CF) 2.2 to 3 MHz (< 3.6 GHz CF) 4 to 8 MHz (< 3.6 GHz CF) ± 0.07 dB nominal 0 to –0.2 dB nominal 0 to –0.4 dB nominal Bandwidth accuracy (–3.01 dB) RBW range 1 Hz to 1.3 MHz ± 2% nominal Selectivity (–60 dB/–3 dB) 4.1:1 nominal EMI bandwidth (CISPR compliant) 200 Hz, 9 kHz, 120 kHz, 1 MHz (Option EMC required) EMI bandwidth (MIL STD 461E compliant) 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz (Option EMC required) Analysis bandwidth 1 Maximum bandwidth Option B25 (standard) Option B40 Option B85 Option B1X Option B2X Option B5X
25 MHz
40 MHz
85 MHz
160 MHz
255 MHz
510 MHz
Video bandwidth (VBW) Range 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz) Accuracy ± 6% nominal (in swept mode and zero span) 1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or processing in the time, frequency, or modulation domain. Frequency and Time Specifications (continued)
6 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Amplitude Accuracy and Range Specifications Amplitude range Measurement range Preamp Off Preamp On RF (Opt 503) Microwave (Opt 508, 513, 526) Millimeter-wave (Opt 544, 550) Displayed average noise level (DANL) to +30 dBm Displayed average noise level (DANL) to +30 dBm Displayed average noise level (DANL) to +24 dBm Displayed average noise level (DANL) to +20 dBm Input mechanical attenuator range (3 Hz to 50 GHz) 0 to 70 dB in 2 dB steps Electronic attenuator (Option EA3) Frequency range 3 Hz to 3.6 GHz Attenuation range Electronic attenuator range Full attenuation range (mechanical + electronic) 0 to 24 dB, 1 dB steps 0 to 94 dB, 1 dB steps Maximum safe input level Average total power (with and without preamp) +30 dBm (1 W) Peak pulse power (< 10 μs pulse width, < 1% duty cycle, input attenuation ≥ 30 dB) +50 dBm (100 W) DC volts DC coupled AC coupled ± 0.2 Vdc ± 100 Vdc (For frequency Option 503, 508, 513, or 526) Display range Log scale 0.1 to 1 dB/division in 0.1 dB steps 1 to 20 dB/division in 1 dB steps (10 display divisions) Linear scale 10 divisions Scale units dBm, dBmV, dBμV, dBmA, dBμA, V, W, A Frequency response Specification 95th percentile (≈ 2 s) (10 dB input attenuation, 20 to 30 °C, preselector centering applied above 3.6 GHz) RF/MW (Option 503, 508, 513, 526)
3 Hz to 10 MHz
20 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 22.0 GHz 22.0 to 26.5 GHz ± 0.46 dB ± 0.35 dB ± 0.35 dB ± 1.5 dB ± 2.0 dB ± 2.0 dB ± 2.5 dB ± 0.16 dB ± 0.39 dB ± 0.56 dB ± 0.81 dB ± 0.82 dB Millimeter-Wave (Option 544, 550)
3 Hz to 20 MHz
50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz ± 0.46 dB ± 0.35 dB ± 0.35 dB ± 1.7 dB ± 1.5 dB ± 2.0 dB ± 2.0 dB ± 2.0 dB ± 2.5 dB ± 2.5 dB ± 3.2 dB ± 0.19 dB ± 0.15 dB ± 0.70 dB ± 0.57 dB ± 0.54 dB ± 0.64 dB ± 0.72 dB ± 0.71 dB ± 0.93 dB ± 1.24 dB Preamp on (0 dB attenuation) (Option P03, P08, P13, P26, P44, P50) RF/MW (Option 503, 508, 513, 526) 9 to 100 kHz 100 kHz to 50 MHz 50 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz ± 0.68 dB ± 0.55 dB ± 2.0 dB ± 2.3 dB ± 2.5 dB ± 3.0 dB ± 3.5 dB ± 0.38 dB ± 0.34 dB ± 0.30 dB ± 0.69 dB ± 0.71 dB ± 0.95 dB ± 1.41 dB ± 1.61 dB
7 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Millimeter-Wave (Option 544, 550) 9 to 100 kHz 100 kHz to 50 MHz 50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz ± 0.68 dB ± 0.60 dB ± 2.0 dB ± 2.0 dB ± 2.3 dB ± 2.5 dB ± 3.0 dB ± 3.5 dB ± 3.0 dB ± 4.1 dB ± 0.40 dB ± 0.34 dB ± 0.31 dB ± 0.81 dB ± 0.70 dB ± 0.79 dB ± 0.88 dB ± 1.07 dB ± 1.03 dB ± 1.35 dB ± 1.69 dB Input attenuation switching uncertainty Specifications Additional information Relative to 10 dB and preamp off At 50 MHz (reference frequency) Attenuation 12 to 40 dB Attenuation 2 to 8 dB Attenuation 0 dB ± 0.14 dB ± 0.18 dB ± 0.04 dB typical ± 0.06 dB typical ± 0.05 dB nominal Attenuation > 2 dB 3 Hz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 26.5 GHz 26.4 to 50 GHz ± 0.3 dB nominal ± 0.5 dB nominal ± 0.7 dB nominal ± 0.7 dB nominal ± 1.0 dB nominal Total absolute amplitude accuracy (10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any reference level, any scale, s = nominal standard deviation) At 50 MHz At all frequencies 10 Hz to 3.6 GHz ± 0.24 dB ± (0.24 dB + frequency response) ± 0.19 dB (95th Percentile approx. 2s) Preamp on (Option P03, P08, P13, P26, P44 and P50) At all frequencies ± (0.36 dB + frequency response) Input voltage standing wave ratio (VSWR) Freq Opt 503, 508, 513, 526 Freq Opt 544, 550 (10 dB input attenuation) 50 MHz 10 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz 1.07 nominal 1.139 (95th percentile) 1.290 (95th percentile) 1.388 (95th percentile) 1.41 (95th percentile) 1.48 (95th percentile) NA NA 1.025 nominal 1.134 (95th percentile) 1.152 (95th percentile) 1.178 (95th percentile) 1.204 (95th percentile) 1.331 (95th percentile) 1.321 (95th percentile) 1.378 (95th percentile) Preamp on (0 dB input attenuation) (Option P03. P08, P13, P26, P44, and P50) 10 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz 1.71 (95th percentile) 1.54 (95th percentile) 1.57 (95th percentile) 1.48 (95th percentile) 1.54 (95th percentile) NA NA 1.393 (95th percentile) 1.50 (95th percentile) 1.310 (95th percentile) 1.330 (95th percentile) 1.339 (95th percentile) 1.41 (95th percentile) 1.42 (95th percentile) Amplitude Accuracy and Range Specifications (continued)
8 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 1.5 MHz RBW ± 0.03 dB 1.6 MHz to 2.7 MHz RBW ± 0.05 dB 3 MHz RBW ± 0.10 dB 4, 5, 6, 8 MHz RBW ± 0.30 dB Reference level Range Log scale Linear scale –170 to +30 dBm in 0.01 dB steps 707 pV to 7.07 V with 0.11% (0.01 dB) resolution Accuracy 0 dB 1 Display scale switching uncertainty Switching between linear and log 0 dB 1 Log scale/div switching 0 dB 1 Display scale fidelity Between –10 dBm and –18 dBm input mixer level ± 0.10 dB total ± 0.04 dB typical Below –18 dBm input mixer level ± 0.07 dB ± 0.02 dB typical Trace detectors Standard Normal, peak, sample, negative peak, log power average, RMS average, and voltage average With Option EMC Add quasi-peak to above Preamplifier Frequency range 2 Option P03 Option P08 Option P13 Option P26 Option P44 Option P50 9 kHz to 3.6 GHz 9 kHz to 8.4 GHz 9 kHz to 13.6 GHz 9 kHz to 26.5 GHz 9 kHz to 44 GHz 9 kHz to 50 GHz Gain 100 kHz to 3.6 GHz 3.6 to 26.5 GHz 26.5 to 50 GHz +20 dB nominal +35 dB nominal +40 dB nominal 1. Only affects the display, not the measurement, so it causes no additional error in measurement results from trace data or markers. 2. Below 100 kHz, only 95th percentile (approx. 2 s) value for frequency response is provided. Amplitude Accuracy and Range Specifications (continued)
9 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications 1 dB gain compression (two-tone) Maximum power at input mixer (At 1 kHz RBW with 100 kHz tone spacing, 20 to 30 °C) 20 to 40 MHz 40 to 200 MHz 200 MHz to 3.6 GHz 3.6 to 16 GHz 16 to 26.5 GHz 26.5 to 50 GHz –3 dBm +1 dBm +3 dBm +1 dBm –1 dBm 0 dBm typical +3 dBm typical +5 dBm typical +4 dBm typical +2 dBm typical 0 dBm nominal Preamp on (Option P03, P08, P13, P26, P44, and P50) 10 MHz to 3.6 GHz 3.6 to 26.5 GHz Tone spacing 100 kHz to 20 MHz Tone spacing > 70 MHz Freq Option ≤ 526 Freq Option > 526 26.5 to 50 GHz –14 dBm nominal –28 dBm nominal –10 dBm nominal –20 dBm nominal –30 dBm nominal Displayed average noise level (DANL) 4 Specification Typical (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 1 Hz RBW, 20 to 30 °C) RF/MW (Option 503, 508, 513, 526) Normal 1/LNP enabled 2 Normal 1/LNP enabled 2 Preamp off 3 Hz to 9 kHz 9 to 100 kHz 100 kHz to1 MHz 1 to 10 MHz 10 MHz to 1.2 GHz 1.2 to 2.1 GHz 2.1 to 3.0 GHz 3.0 to 3.6 GHz 3.5 to 4.2 GHz 4.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 16.9 GHz 16.9 to 20.0 GHz 20.0 to 26.5 GHz –146 dBm/NA –150 dBm/NA –155 dBm/NA –154 dBm/NA –153 dBm/NA –151 dBm/NA –151 dBm/NA –147 dBm/–153 dBm –150 dBm/–155 dBm –149 dBm/–155 dBm –145 dBm/–152 dBm –143 dBm/–151 dBm –137 dBm/–150 dBm –100 dBm/NA typical –151 dBm/NA typical –156 dBm/NA typical –158 dBm/NA typical –155 dBm/NA typical –155 dBm/NA typical –153 dBm/NA typical –153 dBm/NA typical –150 dBm/–156 dBm typical –152 dBm/–157 dBm typical –151 dBm/–157 dBm typical –147 dBm/–155 dBm typical –145 dBm/–153 dBm typical –140 dBm/–152 dBm typical Preamp on 3 100 to 200 kHz 200 to 500 kHz 0.5 to 1 MHz 1 to 10 MHz 10 MHz to 2.1 GHz 2.1 to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 16.9 GHz 16.9 to 20.0 GHz 20.0 to 26.5 GHz –157 dBm/NA –160 dBm/NA –162 dBm/NA –164 dBm/NA –165 dBm/NA –163 dBm/NA –164 dBm/NA –163 dBm/NA –161 dBm/NA –159 dBm/NA –155 dBm/NA –159 dBm/NA typical –161 dBm/NA typical –164 dBm/NA typical –165 dBm/NA typical –166 dBm/NA typical –164 dBm/NA typical –166 dBm/NA typical –164 dBm/NA typical –162 dBm/NA typical –161 dBm/NA typical –157 dBm/NA typical DANL with Noise Floor Extension (Option NF2 ) on 95th percentile DANL improvement Preamp Off Preamp On LNP enabled 2, 3 Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 9 dB 10 dB 10 dB 9 dB 10 dB 10 dB 9 dB 10 dB 9 dB 8 dB NA 10 dB 10 dB 10 dB 10 dB DANL with Noise Floor Extension Preamp Off Preamp On LNP enabled 2, 3 Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 –163 dBm –162 dBm –162 dBm –156 dB –150 dBm –174 dBm –174 dBm –173 dBm –172 dBm –166 dBm NA –166 dBm –167 dBm –164 dBm –162 dBm 1. With Option NF2 (Noise Floor Extension) “Off”. 2. LNP (Low Noise Path) requires option LNP. 3. At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp. 4. With standard LO. Instruments with DDS LO ( Option EP0) may see a few dB degradation in DANL. See specifications guide for details.
10 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Displayed average noise level (DANL) 1 Specification Typical Millimeter-Wave (Option 544, 550) Normal 2/LNP enabled 3 Normal 2/LNP enabled 3 Preamp off 3 Hz to 9 kHz 9 to 100 kHz 100 kHz to 1 MHz 1 to 10 MHz 10 MHz to 1.2 GHz 1.2 to 2.1 GHz 2.1 to 3 GHz 3 to 3.6 GHZ 3.5 to 4.2 GHz 4.2 to 6.6 GHz 6.6 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 14 GHz 14 to 17 GHz 17 to 22.5 GHz 22.5 to 26.5 GHz 26.4 to 34 GHz 33.9 to 37 GHz 37 to 40 GHz 40 to 46 GHz 46 to 49 GHz 49 to 50 GHz –146 dBm/NA –150 dBm/NA –155 dBm/NA –154 dBm/NA –153 dBm/NA –151 dBm/NA –151 dBm/NA –143 dBm/–150 dBm –144 dBm/–152 dBm –147 dBm/–154 dBm –147 dBm/–153 dBm –143 dBm/–150 dBm –145 dBm/–151 dBm –141 dBm/–149 dBm –139 dBm/–146 dBm –138 dBm/–146 dBm –134 dBm/–141 dBm –132 dBm/–140 dBm –130 dBm/–140 dBm –130 dBm/–138 dBm –128 dBm/–138 dBm –100 dBm/NA nominal –151 dBm/NA typical –156 dBm/NA typical –158 dBm/NA typical –155 dBm/NA typical –155 dBm/NA typical –153 dBm/NA typical –153 dBm/NA typical –147 dBm/–154 dBm typical –148 dBm/–155 dBm typical –149 dBm/–156 dBm typical –149 dBm/–156 dBm typical –146 dBm/–152 dBm typical –148 dBm/–153 dBm typical –146 dBm/–152 dBm typical –143 dBm/–150 dBm typical –142 dBm/–149 dBm typical –139 dBm/–147 dBm typical –138 dBm/–145 dBm typical –135 dBm/–145 dBm typical –135 dBm/–142 dBm typical –133 dBm/–142 dBm typical Preamp on 4 100 to 200 kHz 200 to 500 kHz 500 kHz to 1 MHz 1 to 10 MHz 10 MHz to 2.1 GHz 2.1 to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17 GHz 17 to 20 GHz 20 to 26.5 GHz 26.4 to 30 GHz 30 to 34 GHz 33.9 to 37 GHz 37 to 40 GHz 40 to 43 GHz 44 to 46 GHz 46 to 50 GHz –157 dBm –159 dBm –162 dBm –164 dBm –164 dBm –163 dBm –161 dBm –161 dBm –162 dBm –160 dBm –158 dBm –157 dBm –155 dBm –153 dBm –152 dBm –149 dBm –149 dBm –146 dBm –159 dBm typical –161 dBm typical –164 dBm typical –165 dBm typical –166 dBm typical –164 dBm typical –163 dBm typical –163 dBm typical –164 dBm typical –163 dBm typical –161 dBm typical –159 dBm typical –158 dBm typical –157 dBm typical –156 dBm typical –154 dBm typical –154 dBm typical –150 dBm typical 1. With standard LO. Instruments with DDS LO ( Option EP0) may see a few dB degradation in DANL. See specifications guide for details. 2. With Option NF2 (Noise Floor Extension) “Off”. 3. LNP (Low Noise Path) requires option LNP. 4. At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp. Dynamic Range Specifications (continued)
11 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet DANL with Noise Floor Extension (Option NF2) on 95th percentile DANL Improvement Preamp Off Preamp On LNP enabled 1, 2 Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 Band 5 Band 6 10 dB 9 dB 9 dB 9 dB 10 dB 11 dB 11 dB 9 dB 9 dB 8 dB 8 dB 9 dB 8 dB 7 dB N/A 10 dB 9 dB 10 dB 11 dB 12 dB 11 dB DANL with Noise Floor Extension Preamp Off Preamp On LNP enabled 1, 2 Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 Band 5 Band 6 –163 dBm –160 dBm –161 dBm –161 dBm –158 dBm –157 dBm –149 dBm –174 dBm –172 dBm –173 dBm –174 dBm –171 dBm –168 dBm –161 dBm N/A –165 dBm –164 dBm –164 dBm –161 dBm –161 dBm –152 dBm 1. LNP (Low Noise Path) requires option LNP. 2. At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp. Dynamic Range Specifications (continued)
12 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Residuals, images, and spurious responses Residual responses (Input terminated and 0 dB attenuation) 200 kHz to 8.4 GHz Zero span or FFT or other frequencies –100 dBm –100 dBm nominal Image responses 4 Tuned Freq (f) Excitation Freq Response RF/MW (Opt 503, 508, 513, 526) mmW (Opt 544, 550) (Mixer level at –10 dBm) 10 MHz to 26.5 GHz 10 MHz to 3.6 GHz 10 MHz to 3.6 GHz 3.5 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22 GHz 22 to 26.5 GHz f+45 MHz f+10,245 MHz f+645 MHz f+645 MHz f+645 MHz f+645 MHz f+645 MHz –80 dBc –118 dBc typical –80 dBc –112 dBc typical –80 dBc –101 dBc typical –78 dBc –87 dBc typical –74 dBc –84 dBc typical –70 dBc –82 dBc typical –68 dBc –79 dBc typical –80 dBc –118 dBc typical –80 dBc –112 dBc typical –80 dBc –101 dBc typical –80 dBc –102 dBc typical –80 dBc –102 dBc typical –80 dBc –100 dBc typical –80 dBc –97 dBc typical (Mixer level at –30 dBm) 26.5 to 34.5 GHz 34.4 to 44 GHz 44 to 50 GHz f+645 MHz f+645 MHz f+645 MHz –70 dBc – 94 dBc typical –60 dBc – 79 dBc typical – 75 dBc nominal Other spurious responses Mixer level Response Carrier frequency ≤ 26.5 GHz First RF order (f ≥ 10 MHz from carrier) Higher RF order (f ≥ 10 MHz from carrier) Carrier frequency > 26.5 GHz First RF order (f ≥ 10 MHz from carrier) Higher RF order (f ≥ 10 MHz from carrier) –10 dBm –40 dBm –30 dBm –30 dBm –80 dBc + 20log(N 1) Including IF feedthrough, LO harmonic mixing responses –80 dBc + 20log(N 1) Including higher order mixer responses –90 dBc nominal –90 dBc nominal LO-related spurious responses (200 Hz ≤ f < 10 MHz from carrier) –10 dBm –68 dBc 2 + 20log(N 1) Line-related spurious responses –73 dBc 2 + 20log(N 1) (nominal) Second harmonic distortion (SHI) Source frequency Mixer level Distortion 3 (LNP Off/LNP On) SHI 3 (LNP Off/LNP On) RF/MW (Option 503, 508, 513, 526) 10 to 100 MHz 0.1 to 1.8 GHz 1.75 to 2.5 GHz 2.5 to 4 GHz 4 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –57 dBc/NA –60 dBc/NA –77 dBc/–95 dBc –77 dBc/–101 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc +42 dBm/NA +45 dBm/NA +62 dBm/+80 dBm +62 dBm/+86 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm Millimeter-Wave (Option 544, 550) 10 to 100MHz 100 M to 1.8 GHz 1.8 to 2.5 GHz 2.5 to 3 GHz 3 to 5 GHz 5 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz 13.25 to 25 GHz –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –57 dBc/NA –60 dBc/NA –72 dBc/–95 dBc –72 dBc/–99 dBc –77 dBc/–99 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc –65 dBc/–105 dBc (nom.) +42 dBm/NA +45 dBm/NA +57 dBm/+80 dBm +57 dBm/+84 dBm +62 dBm/+84 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm +50 dBm/+90 dBm (nom.) Preamp level Distortion SHI Preamp on (Option P03, P08, P13, P26, P44, P50) 10 MHz to 1.8 GHz 1.8 to 13.25 GHz 13.25 to 25 GHz –45 dBm –50 dBm –50 dBm –78 dBc nominal –60 dBc nominal –50 dBm nominal +33 dBm nominal +10 dBm nominal 0 dBm nominal 1. N is the LO multiplication factor. Refer to page 4 for the N value verses frequency ranges. 2. Nominally –40 dBc under large magnetic (0.38 Gauss rms) or vibrational (0.21 g rms) environmental stimuli. 3. Normal path/LNP enabled (requires Option LNP). 4. With standard LO. Instruments with DDS LO (option EP0) may see a few dB degradation in DANL. See specifications guide for details. Dynamic Range Specifications (continued)
Figure 1. Nominal TOI performance versus frequency and tone separation
1 MHz
10 MHz
- When Option MPB is installed and enabled, some aspects of the analyzer performance change. Please refer to the PXA specification guide for
Figure 3. Nominal PXA phase noise at various center frequencies with Figure 4. Nominal PXA phase noise at various center frequencies with DDS
15 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet PowerSuite Measurement Specifications Channel power Amplitude accuracy, W-CDMA or IS95 (20 to 30 °C, attenuation = 10 dB) ± 0.61 dB (± 0.19 dB 95th percentile) Occupied bandwidth Frequency accuracy ± [span/1000] nominal Adjacent channel power Accuracy, 3GPP W-CDMA (ACLR) (at specific mixer levels and ACLR ranges) Adjacent Alternate MS (UE) BTS ± 0.09 dB ± 0.18 dB ± 0.16 dB ± 0.31 dB Dynamic range (typical) Without noise correction With noise correction –81.5 dB –82.5 dB –87 dB –88 dB Offset channel pairs measured 1 to 6 Multi-carrier ACP Accuracy, 3GPP W-CDMA (ACPR) (4 carriers, 5 MHz offset, BTS, UUT ACPR range at –42 to –48 dB, optimal mixer level at –21 dBm) ± 0.13 dB Multiple number of carriers measured Up to 12 Power statistics CCDF Histogram resolution 0.01 dB Harmonic distortion Maximum harmonic number 10th Result Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in % Intermod (TOI) Measure the third-order products and intercepts from two tones Burst power Methods Power above threshold, power within burst width Results Single burst output power, average output power, maximum power, minimum power within burst, burst width Spurious emission 3GPP W-CDMA table-driven spurious signals; search across regions Dynamic range (1 to 3.6 GHz) Absolute sensitivity (1 to 3.6 GHz) 97.1 dB –86.4 dBm (101.9 dB typical) (–90.4 dBm typical) Spectrum emission mask (SEM) cdma2000 ® (750 kHz offset) Relative dynamic range Absolute sensitivity Relative accuracy 81.6 dB –101.7 dBm ± 0.08 dB (86.4 dB typical) (–105.7 dBm typical) 3GPP W-CDMA (2.515 MHz offset) Relative dynamic range Absolute sensitivity Relative accuracy 85.4 dB –101.7 dBm ± 0.08 dB (89.8 dB typical) (–105.7 dBm typical)
16 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet General Specifications Temperature range Operating Storage 0 to 55 °C –40 to +70 °C Altitude 4,500 meters (approx 15,000 feet) EMC Complies with the essential requirements of the European EMC Directive as well as current editions of the following standards (dates and editions are cited in the Declaration of Conformity): - IEC/EN 61326-1 - CISPR Pub 11 Group 1, class A - AS/NZS CISPR 11 - ICES/NMB-001 This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada South Korean Class A EMC declaration This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home. A 급 기기 (업무용 방송통신기자재)이 기 기는 업무용 (A 급) 전자파적합기기로서 판 매자 또는 사용자는 이 점을 주 의하시기 바라 며, 가 정외의 지역에서 사용하는 것을 목적으 로 합니다. Safety Complies with the essential requirements of the European Low Voltage Directive as well as current editions of the following standards (dates and editions are cited in the Declaration of Conformity): - IEC/EN 61010-1 - Canada: CSA C22.2 No. 61010-1 - USA: UL std no. 61010-1 Acoustic statement (European Machinery Directive) Acoustic noise emission LpA < 70 dB Operator position Normal operation mode per ISO 7779 Acoustic noise - more information (Values given are per ISO 7779 standard in the “Operator Sitting” position) Ambient temperature < 40 °C Nominally under 55 dBA Sound Pressure. 55 dBA is generally considered suitable for use in quiet office environment ≥ 40 °C Nominally under 65 dBA Sound Pressure. 65 dBA is generally considered suitable for use in noisy office environment Environmental stress Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, tempera - ture, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3. Power requirements Voltage and frequency 100 to 120 V, 50/60/400 Hz 220 to 240 V, 50/60 Hz Power consumption On Standby
630 W (Maximum)
- The N9030B is in full compliance with CISPR 11, Class A emissions and is declared as such. In addition, the N9030B has been type tested and shown to meet CISPR 11, Class B emissions limits. Information regarding the Class B emission performance of the N9030B is provided as a convenience to the user and is not intended to be a regulatory declaration.
17 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Display Resolution Size 1280 x 768 269 mm (10.6 in.) diagonal (nominal) capacitive multi-touch screen Data storage Internal Removable solid state drive (≥ 80 GB) and secure digital (SD) memory device External Supports USB 3.0/2.0 compatible memory devices Weight (without options) Net Shipping 22 kg (48 lbs) nominal 34 kg (75 lbs) nominal Dimensions Height Width Length 177 mm (7.0 in) 426 mm (16.8 in) 556 mm (21.9 in) Warranty The PXA signal analyzer is supplied with a 3-year standard warranty Calibration cycle The recommended calibration cycle is one year. Calibration services are available through Keysight service centers Inputs and Outputs Front panel RF input Connector Standard (Option 503, 508, 513, 526) Option C35 (with Option 526 only) Standard (Option 544, 550) Type-N female, 50 Ω nominal APC 3.5 mm male, 50 Ω nominal 2.4 mm male, 50 Ω nominal Analog baseband IQ inputs (Option BBA) 1 Connectors (I, Q, I-Bar, Q-Bar, and Cal Out) Cal Out Signal Frequency Input impedance (4 connectors: I, Q, I-, Q-) Probes supported 2 Active probe Passive probe Input return loss 50 Ω impedance only selected BNC female AC coupled square wave Selectable between 1 kHz and 250 kHz 50 Ω, 1 MΩ (selectable, nominal) 1130A, 1131A, 1132A, 1134A 1161A –5 dB (0 to 10 MHz, nominal) –0 dB (10 to 40 MHz. nominal) Probe power Voltage/current +15 Vdc, ± 7% at 150 mA max nominal –12.6 Vdc, ± 10% at 150 mA max nominal USB ports Host (3 ports) Standard Connector Output current Port marked with lightning bolt Ports not marked with lightning bolt Compatible with USB 2.0 USB Type-A female
1.2 A (nominal)
0.5 A Headphone jack Miniature stereo audio jack (3.5 mm, also known as “1/8 inch”) 1. For additional specifications, please refer to Chapter BBA in the PXA Signal Analyzer specification guide 2. For more details, please refer to the Keysight Probe Configuration Guides, literature numbers 5968-7141EN and 5989-6162EN; probe heads are necessary to attach to your device properly and probe connectivity kits such as E2668B, E2669A. or E2675A are required. General Specifications (continued)
18 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet External mixing, Option EXM Connection port Connector Impedance Functions SMA, female 50 Ω nominal Triplexed for mixer bias, IF input and LO output Mixer bias range IF input center frequency ≤ 25 MHz IF path
40 MHz BW IF path
255 MHz BW IF path
510 MHz BW IF path
± 10 mA in 10 uA step
322.5 MHz
250.0 MHz
300 MHz
750.0 MHz
877 .1484375 MHz 3.75 to 14.0 GHz Rear panel
10 MHz out
BNC female, 50 Ω nominal ≥ 0 dBm nominal
10 MHz + (10 MHz x frequency reference accuracy)
BNC female, 50 Ω nominal –5 to 10 dBm nominal 1 to 50 MHz nominal (selectable to 1 Hz resolution) ± 2 x 10–6 of specified external reference input frequency Trigger 1 and 2 inputs Connector Impedance Trigger level range BNC female > 10 kΩ nominal –5 to +5 V (TTL) factory preset Trigger 1 and 2 outputs Connector Impedance Level BNC female 50 Ω nominal 0 to 5 V (CMOS) nominal Sync (reserved for future use) Connector BNC female Monitor output 1 Connector Format Resolution VGA compatible, 15-pin mini D-SUB XGA (60 Hz vertical sync rates, non-interlaced) Analog RGB 1024 x 768 Monitor output 2 Connector Resolution Mini DisplayPort 1024 x 768 Noise source drive +28 V (pulsed) Connector Output voltage BNC female On 28.0 ± 0.1 V (60 mA maximum) Off < 1 V SNS series noise source For use with the Agilent/Keysight SNS Series noise sources Digital bus Connector MDR-80 Inputs and Outputs (continued)
19 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Rear panel Analog out Connector BNC female USB ports Host, super speed Standard Connector Output current Host Standard Connector Output current Device Standard Connector 2 ports (stacked with each other) Compatible with USB 3.0 USB Type-A female 0.9 A 1 port (stacked with LAN) USB 2.0 USB Type-A female 0.5 A Compatible with USB 3.0 USB Type-B female GPIB interface Connector GPIB codes GPIB mode IEEE-488 bus connector SH1, AH1, T6, SR1, RL1, PP0, DC1, C1, C2, C3, C28, DT1, L4, C0 Controller or device LAN TCP/IP interface Standard Connector 1000Base-T RJ45 Ethertwist IF output Connector Impedance SMA female, shared by Opts CR3, CRP, and ALV 50 Ω nominal 2nd IF output, Option CR3 Center frequency SA mode or I/Q analyzer with IF BW ≤ 25 MHz with Option B40 with Option B85/B1X with Option B2X with Option B5X
250 MHz
750 MHz
877 .1484375 MHz Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Low band IF Path ≤ 160 MHz IF Path 255 MHz IF Path 510 MHz High band, with preselector High band, with preselector bypassed 1 Up to 160 MHz (nominal) Up to 255 MHz (nominal) Up to 510 MHz (nominal) Depends on center frequency Up to 700 MHz (nominal); expandable to 900 MHz with corrections Programmable IF output, Option CRP Center frequency Range Resolution 10 to 75 MHz (user selectable)
0.5 MHz
Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Output at 70 MHz Low band or high band with preselector bypassed Preselected band
100 MHz (nominal)
Depends on RF center frequency Lower output frequencies Subject to folding Residual output signals ≤ –88 dBm (nominal) 1. The maximum bandwidth is not centered around the IF output center frequency. Inputs and Outputs (continued)
20 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Option YAV Y-Axis output General port specifications Connector Impedance SMA female Shared with other options 50 Ω nominal Fast log video output Output voltage Maximum Slope Open-circuit voltages shown
1.6 V at –10 dBm nominal
25 ± 1 mV/dB nominal Log fidelity Range Accuracy within range 49 dB (nominal) with input frequency at 1 GHz ± 1.0 dB nominal Rise time 15 ns nominal Fall time Bands 1-4 with Option MPB Other cases 40 ns nominal best case Depends on bandwidth Other Optional Output Option ALV Log video out General port specifications Connector Impedance BNC female Shared with other options 50 Ω nominal Screen video Operating conditions Display scale types Log scales Modes Gating Log or Lin All (0.1 to 20 dB/div) Spectrum analyzer only Gating must be off “Lin” is linear in voltage Output scaling Offset Gain accuracy 0 to 1.0 V open circuit, representing bottom to top of screen ± 1% of full scale nominal ± 1% of output voltage nominal Delay between RF input to analog output 71.7 µs +2.56/RBW + 0.159/VBW nominal Log video (Log envelope) output Amplitude range (terminated with 50 Ω) Maximum 1.0 V nominal for –10 dBm at the mixer Scale factor Bandwidth Operating conditions 1 V per 192.66 dB Set by RBW Select Sweep Type = Swept Linear video (AM Demod) output Amplitude range (terminated with 50 Ω) Maximum Minimum
1.0 V nominal for signal envelope at the reference level
Scale factor If carrier level is set to half the reference level in volts, the scale factor is 200% of carrier level per volt. Regardless of the carrier level, the scale factor is 100% of reference level per volt. Bandwidth Operating conditions Set by RBW Select Sweep Type = Swept
21 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency Frequency span Option B25 (standard) Option B40 Option B85 Option B1X Option B2X Option B5X
10 Hz to 25 MHz
10 Hz to 40 MHz
10 Hz to 85 MHz
10 Hz to 160 MHz
10 Hz to 255 MHz
10 Hz to 510 MHz
Resolution bandwidth (spectrum measurement) Range Overall Span = 1 MHz Span = 10 kHz Span = 100 Hz Window shapes 100 mHz to 3 MHz
50 Hz to 3 MHz
1 Hz to 10 kHz
Flat Top, Uniform, Hanning, Hamming, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB) Analysis bandwidth (waveform measurement) Option B25 (standard) Option B40 Option B85 Option B1X Option B2X Option B5X IF frequency response (standard 10 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency) Freq (GHz) Analysis BW (MHz) Max error Midwidth error (95th percentile) Slope (dB/MHz) (95th percentile) RMS (nominal) ≤ 3.6 3.6 to 26.5 3.6 to 26.5 26.5 to 50 26.5 to 50 ≤ 10 ≤ 10 preselected ≤ 10 preselector off 1 ≤ 10 preselected ≤ 10 preselected off 1 ± 0.20 dB ± 0.25 dB ± 0.30 dB ± 0.12 dB ± 0.12 dB ± 0.12 dB ± 0.10 dB ± 0.10 dB ± 0.10 dB 0.02 dB 0.23 dB 0.02 dB 0.12 dB 0.024 dB 1. Option MPB is installed and enabled. I/Q Analyzer
22 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer (continued) IF phase linearity Peak-to-peak (nominal) RMS (nominal) Center freq (GHz) Span (MHz) Preselector Std LO DDS LO Std LO DDS LO ≥ 0.02, < 3.6 ≥ 3.6 to ≤ 26.5 ≥ 3.6 ≤ 10 ≤ 10 ≤ 10 NA Off 1 On 0.06° 0.10° 0.11° 0.14° 0.27° 0.93° 0.012° 0.022° 0.024° 0.032° 0.057° 0.22° Dynamic range (standard 10 MHz IF path) Clipping-to-noise dynamic range Excluding residuals and spurious responses Clipping level at mixer IF gain = Low IF gain = High Center frequency ≥ 20 MHz –10 dBm –20 dBm –8 dBm nominal –17.5 dBm nominal Noise density at mixer at center frequency (DANL + IF Gain effect) + 2.25 dB Data acquisition (standard 10 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample Pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 16 bits 1. Option MPB is installed and enabled.
23 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet IF frequency response (standard 25 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency) Freq (GHz) Analysis BW (MHz) Max error Midwidth error (95th percentile) Slope (dB/ MHz) (95th percentile) RMS (nominal) 3.6 to 26.5 10 to ≤ 25 preselected 0.50 dB 3.6 to 26.5 10 to ≤ 25 preselector off 1 ± 0.40 dB 0.03 dB 26.5 to 50 10 to ≤ 25 preselected 0.31 dB 26.5 to 50 10 to ≤ 25 preselector off 1 ± 0.40 dB 0.02 dB IF phase linearity Peak-to-peak (nominal) RMS (nominal) Center freq (GHz) Span (MHz) Preselector Std LO DDS LO Std LO DDS LO ≥ 0.02, < 3.6 ≥ 3.6 ≤ 25 ≤ 25 NA Off 1 0.48° 0.85° 0.41° 1.0° 0.12° 0.20° 0.11° 0.27° Dynamic range (standard 25 MHz IF path) Full scale (ADC clipping) Default settings, signal at CF (IF gain = Low) Band 0 Bands 1 through 4 –8 dBm mixer level nominal –7 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Bands 1 through 4 –18 dBm mixer level nominal, subject to gain limitations –17 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal Data acquisition (standard 25 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 16 bits 1. Option MPB is installed and enabled. I/Q Analyzer (continued)
24 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet IF frequency response (40 MHz IF path) IF frequency response (relative to center frequency) Center freq. (GHz) Span (MHz) Preselector Typical RMS (nominal) ≥ 0.03, < 3.6 ≥ 3.6, ≤ 8.4 > 8.4, ≤ 26.5 ≥ 26.5, < 34.4 ≥ 34.4, < 50 ≤ 40 ≤ 40 ≤ 40 ≤ 40 ≤ 40 NA Off 1 Off 1 Off 1 Off 1 ± 0.4 dB ± 0.4 dB ± 0.7 dB ± 0.8 dB ± 1.0 dB ± 0.25 dB ± 0.16 dB ± 0.20 dB ± 0.25 dB ± 0.35 dB 0.05 dB 0.05 dB 0.05 dB 0.1 dB 0.1 dB IF phase linearity (deviation from mean phase linearity) Peak-to-peak (nominal) RMS (nominal) Center freq (GHz) Span (MHz) Preselector Std LO DDS LO Std LO DDS LO ≥ 0.03, < 3.6 ≥ 3.6 ≤ 40 ≤ 40 NA Off 1 0.16° 1.5° 0.36° 1.0° 0.041° 0.35° 0.083° 0.24° EVM (EVM measurement floor for an 802.11g OFDM signal, using 89600 VSA software equalization, channel estimation and data EQ)
2.4 GHz
5.8 GHz with Option MPB
–52.0 dB (0.25%) nominal –49.1 dB (0.35%) nominal Dynamic range (40 MHz IF path) SFDR (Spurious-free dynamic range) Signal frequency within ± 12 MHz of center –80 dBc nominal Signal frequency anywhere within analysis BW Spurious response within ± 18 MHz of center –79 dBc nominal Response anywhere within analysis BW –77 dBc nominal Full scale (ADC clipping) Std LO/DDS LO Default settings, signal at CF (IF gain = Low: IF gain offset = 0 dB) Band 0 Bands 1 through 4 –8 dBm mixer level nominal –7 dBm/–6 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Bands 1 and 2 Band 3 and 4 –18 dBm/–16 dBm mixer level nominal, subject to gain limitations –17 dBm/–9 dBm mixer level nominal, subject to gain limitations –17 dBm/–6 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal 1. Option MPB is installed and enabled. I/Q Analyzer (continued) Option B40 40 MHz analysis bandwidth (Option B40 is automatically included in Option B85, B1X, B2X, or B5X)
25 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Data acquisition (40 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 12 bits I/Q Analyzer (continued) Option B40 40 MHz analysis bandwidth IF frequency response (85 or 160 MHz IF path) IF frequency response (relative to center frequency) Center freq. (GHz) Span (MHz) Preselector Typical RMS (nominal) ≥ 0.1, < 3.6 ≤ 85 ≤ 140 ≤ 160 NA NA NA ± 0.6 dB ± 0.6 dB ± 0.17 dB ± 0.25 dB ± 0.2 dB (nom) 0.05 dB 0.05 dB 0.07 dB ≥ 3.6, ≤ 8.4 ≤ 85 ≤ 140 ≤ 160 Off 1 Off 1 Off 1 ± 0.73 dB ± 0.8 dB ± 0.2 dB ± 0.35 dB ± 0.3 dB (nom) 0.05 dB 0.05 dB 0.07 dB > 8.4, ≤ 26.5 ≤ 85 ≤ 140 ≤ 160 Off 1 Off 1 Off 1 ± 1.10 dB ± 1.30 dB ± 0.50 dB ± 0.75 dB ± 0.5 dB (nom) 0.1 dB 0.1 dB 0.12 dB ≥ 26.5, ≤ 50 ≤ 85 ≤ 140 Off 1 Off 1 ± 1.20 dB ± 1.40 dB ± 0.45 dB ± 0.65 dB 0.12 dB 0.12 dB IF phase linearity (deviation from mean phase linearity) Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.03, < 3.6 ≥ 3.6, ≤ 140 ≤ 160 ≤ 140 ≤ 160 NA NA Off 1 Off 1 0.9° 1.7° 1.6° 2.8° 0.20° 0.42° 0.39° 0.64° EVM (EVM measurement floor) Customized settings required, preselector bypassed (Option MPB) above Band 0 Case 1: 62.5 Msymbol/s, 16QAM signal, RRC filter alpha of 0.2, non-equalized, with approximately 75 MHz occupied bandwidth Band 0, 1.8 GHz Band 1, 5.95 GHz 0.8% nominal 1.1% nominal Case 2: 104.167 Msymbol/s, 16QAM signal, RRC filter alpha of 0.35, non-equalized, with approximately 140 MHz occupied bandwidth Band 1, 5.95 GHz Band 2, 15.3 GHz Band 4, 26 GHz 3.0% nominal, (unequalized) 2.5% nominal, (unequalized) 3.5% nominal, (unequalized) 0.5% nominal, (equalized) 0.6% nominal, (equalized) 1.6% nominal, (equalized) 1. Option MPB is installed and enabled. Option B85 85 MHz or B1X 160 MHz analysis bandwidth
26 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic range (85 or 160 MHz IF path) SFDR (Spurious-free dynamic range) Signal frequency within ± 12 MHz of center –75 dBc nominal Signal frequency anywhere within analysis BW Spurious response within ± 63 MHz of center –74 dBc nominal Response anywhere within analysis BW –72 dBc nominal Full scale (ADC clipping) Default settings, signal at CF (IF gain = Low: IF gain offset = 0 dB) Band 0 Band 1 through 4 –8 dBm mixer level nominal –7 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Band 1 through 4 –18 dBm mixer level nominal, subject to gain limitations –17 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal Data acquisition (85 or 160 MHz IF path) Time record length Analysis tool IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (IQ sample pairs) 1073 MSa (230 Sa) 536 MSa (229 Sa) 4 GB total memory (Option DP4) Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 14 bits I/Q Analyzer (continued) Option B85 85 MHz or B1X 160 MHz analysis bandwidth
27 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer (continued) Option B2X 255 MHz analysis bandwidth (Option B2X is automatically included with Option B5X) IF frequency response (255 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Specification Typical RMS (nominal) ≥ 0.4, < 3.6 > 3.6, ≤ 8.4 > 8.4 ≤ 255 ≤ 255 ≤ 255 NA Off 1 Off 1 ± 0.75 dB ± 0.85 dB ± 0.3 dB ± 0.34 dB ± 0.6 dB nominal 0.1 dB 0.1 dB 0.2 dB IF phase linearity (255 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Pk-to-pk (nominal) RMS (nominal) ≥ 0.4, < 3.6 ≥ 3.6, < 26.5 ≤ 255 ≤ 255 NA Off 1 0.6° 0.5° Dynamic range (255 MHz IF path) Spurious-free dynamic range (SFDR) Anywhere within the analysis BW –78 dBc nominal Full scale (ADC clipping) Mixer level Default setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 RF/MW (Opt 508, 513, 526) +3 dBm nominal +4 dBm nominal +1 dBm nominal High gain setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 -4 dBm nominal +2.5 dBm nominal +1 dBm nominal Effect of signal frequency ≠ CF Up to ± 4 dB nominal IF residual responses across the full BW Band 0 Band 1 Preselector off 1 –110 dBFS nominal –108 dBFS nominal Third-order intermodulation distortion (Two tones of equal level, 1 MHz separation, each tone -23 dB relative to full scale (ADC clipping), IF gain = high) Band 0 Bands 1 through 4 Preselector off 1 –85 dBc nominal –85 dBc nominal Noise density Band Frequency (GHz) 1.80 6.00 10.80 15.15 21.80 IF gain = Low –144 dBm/Hz –141 dBm/Hz –140 dBm/Hz –137 dBm/Hz –135 dBm/Hz IF gain = High –145 dBm/Hz –141 dBm/Hz –140 dBm/Hz –137 dBm/Hz –135 dBm/Hz Data acquisition (255 MHz IF path) Time record length IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA or fast capture32-bit 64-bit Length (IQ sample pairs) 1073 MSa (230 Sa) 536 MSa (229 Sa) 4 GB total memory (Option DP4) Maximum IQ capture time (89600 VSA and fast capture) Length of IQ sample pairs/sample rate (IQ pairs) Sample rate (IQ pairs) Minimum of (1.25 x IFBW, 300 Msa/s) ADC resolution 14 bits 1. MPB (microwave preselector bypass) is enabled. All UXA ship with MPB as a standard feature
28 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer (continued) Option B5X 510 MHz analysis bandwidth IF frequency response (510 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Specification Typical RMS (nominal) ≥ 0.6, < 3.6 > 3.6, ≤ 8.4 > 8.4, ≤ 26.5 ≤ 500 ≤ 500 ≤ 510 NA Off 1 Off 1 ± 1.0 dB ± 1.25 dB ± 0.41 dB ± 0.42 dB ± 0.8 dB nominal 0.06 dB 0.3 dB IF phase linearity (510 MHz IF path) Center Freq (GHz) Span (MHz) Preselector Pk-to-pk (nominal) RMS (nominal) ≥ 0.4, < 3.6 ≥ 3.6, < 26.5 ≤ 510 ≤ 510 NA Off 1.4° Dynamic range (510 MHz IF path) Spurious-free dynamic range (SFDR) Anywhere within the analysis BW –78 dBc nominal Full scale (ADC clipping) Mixer level Default setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 RF/MW (Opt 508, 513, 526) +2 dBm nominal +3 dBm nominal +1 dBm nominal High gain setting, signal at CF Band 0 Bands 1 through 2 Bands 3 through 4 –3.5 dBm nominal –1 dBm nominal +1 dBm nominal Effect of signal frequency ≠ CF Up to ± 4 dB nominal IF residual responses across the full BW Band 0 Band 1 Preselector off 1 –104 dBFS nominal –103 dBFS nominal Third-order intermodulation distortion (Two tones of equal level, 1 MHz separation, each tone -23 dB relative to full scale (ADC clipping), IF gain = high) Band 0 Bands 1 through 4 Preselector off 1 –85 dBc nominal –82 dBc nominal Noise density Band Frequency (GHz) 1.80 6.00 10.80 15.15 21.80 IF gain = Low –144 dBm/Hz –140 dBm/Hz –140 dBm/Hz –137 dBm/Hz –135 dBm/Hz IF gain = High –144 dBm/Hz –142 dBm/Hz –141 dBm/Hz –137 dBm/Hz –135 dBm/Hz Data acquisition (510 MHz IF path) Time record length IQ analyzer 8,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA or fast capture32-bit 64-bit Length (IQ sample pairs) IFBW ≤255.176 MHz IFBW >255.176 MHz
1073 MSa (230 Sa)
2,147 MSa (230 Sa)
536 MSa (229 Sa)
4 GB total memory
8 GB total memory (Option DP4)
(89600 VSA and fast capture) Length of IQ sample pairs/sample rate (IQ pairs) Sample rate (IQ pairs) Minimum of (1.25 x IFBW, 300 Msa/s) ADC resolution 14 bits 1. MPB (microwave preselector bypass) is enabled. All UXA ship with MPB as a standard feature
29 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet Real-time analysis Real-time analysis bandwidth Option RT1 Option RT2 Up to 509.47 MHz Up to 509.47 MHz Analysis BW option determines the max real-time bandwidth Analysis BW option determines the max real-time bandwidth Minimum detectable signal duration with > 60 dB StM 2 ratio Option B85 Option B1X Option B2X or B5X 11.42 ns 5.0 ns 3.33 ns Minimum signal duration with 100% probability of intercept (POI) at full ampli - tude accuracy Option RT1 Option RT2 17.3 μs 3.57 μs For Frequency Mask Triggering (FMT) Signal is at mask level Signal is at mask level Minimum acquisition time FFT rate 100 μs 292,969/s Real-time I/Q data streaming 3 Output stream resolution IQ streaming bandwidth Electrical interface Sample rate 16-bit I + jQ Up to 255 MHz LVDS Varies continuously based on RTSA span setting Max IQ streaming bandwidth and sample rate B1X B2X or B5X
200 Msamples/s
300 Msamples/s
< 3 hours at 255 MHz bandwidth Event markers, IRIG-B GPS 1. For additional RTSA specifications, please refer to Option RT1/RT2 Chapter in the PXA Signal Analyzer specifications guide 2. StM = “Signal-to-Mask” 3. Use with X-COM Systems IQC5255B data recorder to capture rare events and play back at RF using integrated control software on the PXA. Real-time spectrum analyzer (RTSA) 1 Option RT1 or RT2 Option RTS Related Literature Keysight PXA signal analyzers Brochure 5992-1316EN Configuration guide 5992-1318EN
30 | Keysight | N9030B PXA X-Series Signal Analyzer, Multi-touch - Data Sheet myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Keysight Infoline www.keysight.com/find/Infoline Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. KEYSIGHT SERVICES Keysight Services www.keysight.com/find/services Our deep offering in design, test, and measurement services deploys an industry-leading array of people, processes, and tools. The result? We help you implement new technologies and engineer improved processes that lower costs. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/N9030B www.keysight.com/find/PXA www.keysight.com/find/X-Series For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-06-08-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System This information is subject to change without notice. © Keysight Technologies, 2016 Published in USA, September 14, 2016 5992-1317EN www.keysight.com Evolving Our unique combination of hardware, software, support, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight