N9030A KEYSIGHT | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 28

Technical content

PXA X-Series Signal Analyzer N9030A 3 Hz to 3.6, 8.4, 13.6, 26.5, 43, 44, or 50 GHz Data Sheet

2 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Table of Contents Drive your evolution The Keysight Technologies, Inc. future-ready PXA signal analyzer is the evolutionary replacement for your current high-performance analyzer. It helps you sustain past achievements, enhance current designs and accelerate future in - novations. Its performance, flexibility, ca - pability and compatibility enable you to address demanding ap - plications in aerospace, defense, commercial communications and more. – Reveal new levels of signal detail with outstanding RF performance – Increase test throughput and protect your system investments – Refresh legacy systems with a highly compatible replacement This data sheet is a summary of the specifications and conditions for PXA signal analyzers. For the complete specifications guide, visit: www.keysight.com/find/pxa_specifications

3 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Definitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to temperature ranges 0 to 55 °C, unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2σ) of performance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the statistical observations of a sample of instruments, these values include the effects of the uncertainties of external calibration references. These values are not warranted. These values are updated occasionally if a significant change in the statistically observed behavior of production instruments is observed. Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specifications that 80 percent of the units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C. Typical performance does not include measure - ment uncertainty. Nominal values indicate expected performance, or describe product perfor - mance that is useful in the application of the product, but is not covered by the product warranty. The analyzer will meet its specifications when: – The analyzer is within its calibration cycle. – Under auto couple control, except that Auto Sweep Time Rules = Accy. – For signal frequencies < 10 MHz, DC coupling applied. – The analyzer has been stored at an ambient temperature within the allowed operating range for at least two hours before being turned on, if it had previously been stored at a temperature range inside the allowed storage range but outside the allowed operating range. – The analyzer has been turned on at least 30 minutes with Auto Align set to normal, or if Auto Align is set to off or partial, alignments must have been run recently enough to prevent an Alert message. If the Alert condi - tion is changed from Time and Temperature to one of the disabled duration choices, the analyzer may fail to meet specifications without informing the user. For the complete specifications guide, visit: www.keysight.com/find/pxa_specifications

4 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Frequency range DC coupled AC coupled Option 503 3 Hz to 3.6 GHz 10 MHz to 3.6 GHz Option 508 3 Hz to 8.4 GHz 10 MHz to 8.4 GHz Option 513 3 Hz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 3 Hz to 26.5 GHz 10 MHz to 26.5 GHz Option 543 3 Hz to 43 GHz NA Option 544 3 Hz to 44 GHz NA Option 550 3 Hz to 50 GHz NA Band LO multiple (N) 0 1 3 Hz to 3.6 GHz 1 1 3.5 to 8.4 GHz 2 2 8.3 to 13.6 GHz 3 2 13.5 to 17.1 GHz 4 4 17 to 26.5 GHz 5 4 26.4 to 34.5 GHz 6 8 34.4 to 50 GHz Precision frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate ± 1 x 10–7 / year ± 1.5 x 10–7 / 2 years Temperature stability 20 to 30 °C Full temperature range ± 1.5 x 10 ± 5 x 10–8 Achievable initial calibration accuracy ± 4 x 10–8 Example frequency reference accuracy 1 year after last adjustment 20 to 30 °C = ± (1 x 1 x 10 = ± 1.55 x 10–7 Residual FM Center frequency = 1 GHz

10 Hz RBW, 10 Hz VBW

≤ (0.25 Hz x N) p-p in 20 ms nominal See band table above for N (LO multiple) Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.10% x span + 5% x RBW + 2 Hz + 0.5 x horizontal resolution Marker frequency counter Accuracy ± (marker frequency x frequency reference accuracy + 0.100 Hz) Delta counter accuracy ± (delta frequency x frequency reference accuracy + 0.141 Hz) Counter resolution 0.001 Hz Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept FFT ± (0.1% x span + horizontal resolution) ± (0.1% x span + horizontal resolution) 1. Horizontal resolution is span/(sweep points –1). Frequency and Time Specifications

5 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Sweep time and triggering Range Span = 0 Hz Span ≥ 10 Hz 1 μs to 6000 s 1 ms to 4000 s Accuracy Span ≥ 10 Hz, swept Span ≥ 10 Hz, FFT Span = 0 Hz ± 0.01% nominal ± 40% nominal ± 0.01% nominal Sweep trigger Free run, line, video, external 1, external 2, RF burst, periodic timer Trigger Delay Span = 0 Hz or FFT Span ≥ 10 Hz, swept Resolution –150 to +500 ms 0 to 500 ms 0.1 μs Time gating Gate methods Gate length range (except method = FFT) Gate delay range Gate delay jitter Gated LO; gated video; gated FFT 1 μs to 5.0 s 0 to 100.0 s 33.3 ns p-p nominal Sweep (trace) point range All spans 1 to 40001 Resolution bandwidth (RBW) Range (–3.01 dB bandwidth) 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz Bandwidth accuracy (power) RBW range

1 Hz to 100 kHz

110 kHz to 1.0 MHz (< 3.6 GHz CF) 1.1 to 2 MHz (< 3.6 GHz CF) 2.2 to 3 MHz (< 3.6 GHz CF) 4 to 8 MHz (< 3.6 GHz CF) ± 0.07 dB nominal ± 0.10 dB nominal ± 0.20 dB nominal Bandwidth accuracy (–3.01 dB) RBW range 1 Hz to 1.3 MHz ± 2% nominal Selectivity (–60 dB/–3 dB) 4.1:1 nominal EMI bandwidth (CISPR compliant) 200 Hz, 9 kHz, 120 kHz, 1 MHz (Option EMC required) EMI bandwidth (MIL STD 461E compliant) 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz (Option EMC required) Analysis bandwidth Maximum bandwidth Standard Option B25 Option B40 Option B85 Option B1X

10 MHz

25 MHz

40 MHz

85 MHz

160 MHz

Video bandwidth (VBW) Range 1 Hz to 3 MHz (10% steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz) Accuracy ± 6% nominal (in swept mode and zero span) Measurement speed 2 Standard Local measurement and display update rate 10 ms (100/s) nominal Remote measurement and LAN transfer rate 10 ms (100/s) nominal Marker peak search 2.5 ms nominal Center frequency tune and transfer (RF) 43 ms nominal Center frequency tune and transfer (μW) 69 ms nominal Measurement/mode switching 40 ms nominal 1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for fur - ther analysis or processing in the time, frequency, or modulation domain. 2. Sweep points = 101.

6 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Amplitude Accuracy and Range Specifications Amplitude range Measurement range Preamp Off Preamp On RF (Opt 503) Microwave (Opt 508, 513, 526) Millimeter-wave (Opt 543, 544, 550) Displayed average noise level (DANL) to +30 dBm Displayed average noise level (DANL) to +30 dBm Displayed average noise level (DANL) to +24 dBm Displayed average noise level (DANL) to +20 dBm Input attenuator range (3 Hz to 50 GHz) 0 to 70 dB in 2 dB steps Electronic attenuator (Option EA3) Frequency range 3 Hz to 3.6 GHz Attenuation range Electronic attenuator range Full attenuation range (mechanical + electronic) 0 to 24 dB, 1 dB steps 0 to 94 dB, 1 dB steps Maximum safe input level Average total power (with and without preamp) +30 dBm (1 W) Peak pulse power < 10 μs pulse width, < 1% duty cycle +50 dBm (100 W) and input attenuation ≥ 30 dB DC volts DC coupled AC coupled ± 0.2 Vdc ± 100 Vdc (For frequency Option 503, 508, 513, or 526) Display range Log scale 0.1 to 1 dB/division in 0.1 dB steps 1 to 20 dB/division in 1 dB steps (10 display divisions) Linear scale 10 divisions Scale units dBm, dBmV, dBμV, dBmA, dBμA, V, W, A Frequency response Specification 95th percentile (≈ 2 σ) (10 dB input attenuation, 20 to 30 °C, preselector centering applied above 3.6 GHz) RF/MW (Option 503, 508, 513, 526)

3 Hz to 10 MHz

20 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 22.0 GHz 22.0 to 26.5 GHz ± 0.46 dB ± 0.35 dB ± 0.35 dB ± 1.5 dB ± 2.0 dB ± 2.0 dB ± 2.5 dB ± 0.16 dB ± 0.39 dB ± 0.45 dB ± 0.62 dB ± 0.82 dB Millimeter-Wave (Option 543, 544, 550)

3 Hz to 20 MHz

50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz ± 0.46 dB ± 0.35 dB ± 0.35 dB ± 1.7 dB ± 1.5 dB ± 2.0 dB ± 2.0 dB ± 2.0 dB ± 2.5 dB ± 2.5 dB ± 3.2 dB ± 0.19 dB ± 0.15 dB ± 0.70 dB ± 0.57 dB ± 0.54 dB ± 0.64 dB ± 0.72 dB ± 0.71 dB ± 0.93 dB ± 1.24 dB Preamp on (0 dB attenuation) (Option P03, P08, P13, P26, P43, P44, P50) RF/MW (Option 503, 508, 513, 526) 9 to 100 kHz 100 kHz to 50 MHz 50 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz ± 0.68 dB ± 0.55 dB ± 2.0 dB ± 2.3 dB ± 2.5 dB ± 3.0 dB ± 3.5 dB ± 0.36 dB ± 0.26 dB ± 0.28 dB ± 0.64 dB ± 0.76 dB ± 0.95 dB ± 1.41 dB ± 1.61 dB

7 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Millimeter-Wave (Option 543, 544, 550) 9 to 100 kHz 100 kHz to 50 MHz 50 MHz to 3.6 GHz 3.5 to 5.2 GHz 5.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22.0 GHz 22.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz ± 0.68 dB ± 0.60 dB ± 2.0 dB ± 2.0 dB ± 2.3 dB ± 2.5 dB ± 3.0 dB ± 3.5 dB ± 3.0 dB ± 4.1 dB ± 0.40 dB ± 0.34 dB ± 0.31 dB ± 0.81 dB ± 0.70 dB ± 0.79 dB ± 0.88 dB ± 1.07 dB ± 1.03 dB ± 1.35 dB ± 1.69 dB Input attenuation switching uncertainty Specifications Additional information Relative to 10 dB and preamp off At 50 MHz (reference frequency) attenuation 12 to 40 dB attenuation 2 to 8 dB attenuation 0 dB ± 0.14 dB ± 0.18 dB ± 0.03 dB typical ± 0.05 dB typical ± 0.05 dB nominal attenuation > 2 dB 3 Hz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 26.5 GHz 26.4 to 50 GHz ± 0.3 dB nominal ± 0.5 dB nominal ± 0.7 dB nominal ± 0.7 dB nominal ± 1.0 dB nominal Total absolute amplitude accuracy (10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any reference level, any scale, σ = nominal standard deviation) At 50 MHz At all frequencies 10 Hz to 3.6 GHz ± 0.24 dB ± (0.24 dB + frequency response) ± 0.19 dB (95th Percentile approx. 2σ) Preamp on (Option P03, P08, P13, P26, P43, P44 and P50) At all frequencies ± (0.36 dB + frequency response) Input voltage standing wave ratio (VSWR) Freq Opt 503, 508, 513, 526 Freq Opt 543, 544, 550 (10 dB input attenuation) 50 MHz 10 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz 1.07:1 nominal 1.139 (95th percentile) 1.290 (95th percentile) 1.388 (95th percentile) 1.403 (95th percentile) 1.475 (95th percentile) NA NA 1.025:1 nominal 1.134 (95th percentile) 1.152 (95th percentile) 1.178 (95th percentile) 1.204 (95th percentile) 1.331 (95th percentile) 1.321 (95th percentile) 1.378 (95th percentile) Preamp on (0 dB input attenuation) (Option P03. P08, P13, P26, P43, P44, and P50) 10 MHz to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 26.5 GHz 26.4 to 34.5 GHz 34.4 to 50 GHz 1.45 (95th percentile) 1.54 (95th percentile) 1.57 (95th percentile) 1.48 (95th percentile) 1.54 (95th percentile) NA NA 1.393 nominal 1.50 (95th percentile) 1.310 (95th percentile) 1.330 (95th percentile) 1.339 (95th percentile) 1.41 (95th percentile) 1.42 (95th percentile)

8 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 1.5 MHz RBW ± 0.03 dB 1.6 MHz to 2.7 MHz RBW ± 0.05 dB 3 MHz RBW ± 0.10 dB 4, 5, 6, 8 MHz RBW ± 0.30 dB Reference level Range Log scale Linear scale –170 to +30 dBm in 0.01 dB steps 707 pV to 7.07 V with 0.11% (0.01 dB) resolution Accuracy 0 dB Display scale switching uncertainty Switching between linear and log 0 dB Log scale/div switching 0 dB Display scale fidelity Between –10 dBm and –80 dBm input mixer level ± 0.10 dB total ± 0.04 dB typical Below –18 dBm input mixer level ± 0.07 dB ± 0.02 dB typical Trace detectors Normal, peak, sample, negative peak, log power average, RMS average, and voltage average Preamplifier Frequency range

1 Option P03

9 kHz to 3.6 GHz 9 kHz to 8.4 GHz 9 kHz to 13.6 GHz 9 kHz to 26.5 GHz 9 kHz to 43 GHz 9 kHz to 44 GHz 9 kHz to 50 GHz Gain 9 kHz to 3.6 GHz 3.6 to 26.5 GHz 26.5 to 50 GHz +20 dB nominal +35 dB nominal +40 dB nominal 1. Below 100 kHz, only 95th percentile (approx. 2σ) value for frequency response is provided.

9 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Dynamic Range Specifications 1 dB gain compression (two-tone) Maximum power at input mixer (At 1 kHz RBW with 100 kHz tone spacing, 20 to 30 °C) 20 to 40 MHz 40 to 200 MHz 200 MHz to 3.6 GHz 3.6 to 16 GHz 16 to 26.5 GHz 26.5 to 50 GHz –3 dBm +1 dBm +3 dBm +1 dBm –1 dBm 0 dBm typical +3 dBm typical +5 dBm typical +4 dBm typical +2 dBm typical 0 dBm nominal Preamp on (Option P03, P08, P13, P26, P43, P44, and P50) 10 MHz to 3.6 GHz 3.6 to 26.5 GHz Tone spacing 100 kHz to 20 MHz Tone spacing > 70 MHz Freq Option ≤ 526 Freq Option > 526 26.5 to 50 GHz –14 dBm nominal –28 dBm nominal –10 dBm nominal –20 dBm nominal –30 dBm nominal Displayed average noise level (DANL) Specification Typical (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 1 Hz RBW, 20 to 30 °C) RF/MW (Option 503, 508, 513, 526) Normal 1/LNP enabled 2 Normal 1/LNP enabled 2 Preamp off 3 Hz to 9 kHz 9 to 100 kHz 100 kHz to1 MHz 1 to 10 MHz 10 MHz to 1.2 GHz 1.2 to 2.1 GHz 2.1 to 3.0 GHz 3.0 to 3.6 GHz 3.5 to 4.2 GHz 4.2 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 16.9 GHz 16.9 to 20.0 GHz 20.0 to 26.5 GHz –146 dBm/NA –150 dBm/NA –155 dBm/NA –155 dBm/NA –153 dBm/NA –152 dBm/NA –151 dBm/NA –147 dBm/–153 dBm –150 dBm/–155 dBm –149 dBm/–155 dBm –145 dBm/–152 dBm –143 dBm/–151 dBm –137 dBm/–150 dBm –100 dBm/NA typical –152 dBm/NA typical –156 dBm/NA typical –158 dBm/NA typical –157 dBm/NA typical –155 dBm/NA typical –154 dBm/NA typical –153 dBm/NA typical –150 dBm/–156 dBm typical –152 dBm/–157 dBm typical –151 dBm/–157 dBm typical –147 dBm/–155 dBm typical –145 dBm/–153 dBm typical –140 dBm/–152 dBm typical Preamp on Option P03, P08, P13, P26 100 to 200 kHz 200 to 500 kHz 0.5 to 1 MHz –157 dBm/NA –160 dBm/NA –164 dBm/NA –160 dBm/NA typical –163 dBm/NA typical –166 dBm/NA typical Option P03, P08, P13, P26 Option P03, P08, P13, P26 Option P03, P08, P13, P26 Option P08, P13, P26 Option P13, P26 3 Option P26 3 Option P26 3 Option P26 3 1 to 10 MHz 10 MHz to 2.1 GHz 2.1 to 3.6 GHz 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 16.9 GHz 16.9 to 20.0 GHz 20.0 to 26.5 GHz –164 dBm/NA –165 dBm/NA –163 dBm/NA –164 dBm/NA –163 dBm/NA –161 dBm/NA –159 dBm/NA –155 dBm/NA –167 dBm/NA typical –166 dBm/NA typical –164 dBm/NA typical –166 dBm/NA typical –165 dBm/NA typical –162 dBm/NA typical –161 dBm/NA typical –157 dBm/NA typical DANL with Noise Floor Extension (NFE) on

4 Improvement @ 95th percentile

RF/MW (Option 503, 508, 513, 526) Preamp Off Preamp On LNP enabled 2, 3 Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 9 dB 10 dB 10 dB 9 dB 10 dB 10 dB 9 dB 10 dB 10 dB 8 dB NA 10 dB 10 dB 10 dB 10 dB Examples of effective DANL Frequency 20 to 30 °C Preamp Off Preamp On LNP enabled 2, 3 Mid-Band 0 (1.8 GHz) Mid-Band 1 (5.95 GHz) Mid-Band 2 (10.95 GHz) Mid-Band 3 (15.3 GHz) Mid-Band 4 (21.75 GHz) –161 dBm –171 dBm –158 dBm –172 dBm –159 dBm –168 dBm –152 dBm –165 dBm –149 dBm –160 dBm NA –162 dBm –162 dBm –160 dBm –160 dBm 1. With the NFE (Noise Floor Extension) “Off”. 2. LNP (Low Noise Path) requires option LNP. 3. At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp. 4. Beginning January 2015, all PXAs ship standard with the second-generation of NFE (instrument alignment based) installed as N9030A-NF2.

10 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Millimeter-Wave (Option 543, 544, 550) Normal 1/LNP enabled 2 Normal 1/LNP enabled 2 Preamp off 3 Hz to 9 kHz 9 to 100 kHz 100 kHz to 1 MHz 1 to 10 MHz 10 MHz to 1.2 GHz 1.2 to 2.1 GHz 2.1 to 3 GHz 3 to 3.6 GHZ 3.5 to 4.2 GHz 4.2 to 6.6 GHz 6.6 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 14 GHz 14 to 17 GHz 17 to 22.5 GHz 22.5 to 26.5 GHz 26.4 to 34 GHz 33.9 to 37 GHz 37 to 40 GHz 40 to 46 GHz 46 to 49 GHz 49 to 50 GHz –146 dBm/NA –150 dBm/NA –155 dBm/NA –155 dBm/NA –153 dBm/NA –152 dBm/NA –151 dBm/NA –143 dBm/–150 dBm –144 dBm/–152 dBm –147 dBm/–154 dBm –147 dBm/–153 dBm –143 dBm/–150 dBm –145 dBm/–151 dBm –141 dBm/–149 dBm –139 dBm/–146 dBm –138 dBm/–146 dBm –134 dBm/–141 dBm –132 dBm/–140 dBm –130 dBm/–140 dBm –130 dBm/–138 dBm –128 dBm/–138 dBm –100 dBm/NA nominal –152 dBm/NA typical –156 dBm/NA typical –158 dBm/NA typical –157 dBm/NA typical –155 dBm/NA typical –154 dBm/NA typical –153 dBm/NA typical –153 dBm/NA typical –147 dBm/–154 dBm typical –148 dBm/–155 dBm typical –149 dBm/–156 dBm typical –149 dBm/–152 dBm typical –146 dBm/–153 dBm typical –148 dBm/–152 dBm typical –146 dBm/–150 dBm typical –142 dBm/–149 dBm typical –139 dBm/–147 dBm typical –138 dBm/–145 dBm typical –135 dBm/–145 dBm typical –135 dBm/–142 dBm typical –133 dBm/–142 dBm typical Preamp on Option P03, P08, P13, P26, P43, P44, P50 3 100 to 200 kHz 200 to 500 kHz 500 kHz to 1 MHz 1 to 10 MHz 10 MHz to 2.1 GHz 2.1 to 3.6 GHz –157 dBm/NA –160 dBm/NA –162 dBm/NA –164 dBm/NA –164 dBm/NA –163 dBm/NA –160 dBm/NA typical –163 dBm/NA typical –165 dBm/NA typical –167 dBm/NA typical –166 dBm/NA typical –164 dBm/NA typical Option P08, P13, P26, P43, P44, P50 Option P13, P26, P43, P44, P50 3 Option P26, P43, P44, P50 3 3.5 to 8.4 GHz 8.3 to 13.6 GHz 13.5 to 17 GHz 17 to 20 GHz 20 to 26.5 GHz –161 dBm/NA –161 dBm/NA –161 dBm/NA –160 dBm/NA –158 dBm/NA –163 dBm/NA typical –163 dBm/NA typical –163 dBm/NA typical –163 dBm/NA typical –161 dBm/NA typical Option P43, P44, P50 3 26.4 to 30 GHz 30 to 34 GHz 33.9 to 37 GHz 37 to 40 GHz 40 to 43 GHz –157 dBm/NA –155 dBm/NA –153 dBm/NA –152 dBm/NA –149 dBm/NA –159 dBm/NA typical –158 dBm/NA typical –157 dBm/NA typical –156 dBm/NA typical –154 dBm/NA typical Option P44, P50 3 43 to 44 GHz –149 dBm/NA –154 dBm/NA typical Option P50 3 44 to 46 GHz 46 to 50 GHz –149 dBm/NA –146 dBm/NA –154 dBm/NA typical –150 dBm/NA typical 1. With the NFE (Noise Floor Extension) “Off”. 2. LNP (Low Noise Path) requires option LNP. 3. At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp.

11 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet DANL with Noise Floor Extension (NFE) on Improvement @ 95th percentile Millimeter-Wave (Option 543, 544, 550) Preamp Off Preamp On LNP enabled 1, 2 Band 0, f > 20 MHz Band 1 Band 2 Band 3 Band 4 Band 5 Band 6 10 dB 9 dB 9 dB 9 dB 10 dB 11 dB 11 dB 9 dB 9 dB 8 dB 8 dB 9 dB 9 dB 8 dB N/A 10 dB 9 dB 10 dB 11 dB 12 dB 11 dB Example of effective DANL Frequency 20 to 30 ° C Preamp Off Preamp On LNP enabled 1, 2 Mid-Band 0 (1.8 GHz) Mid-Band 1 (5.95 GHz) Mid-Band 2 (10.95 GHz) Mid-Band 3 (15.3 GHz) Mid-Band 4 (21.75 GHz) Mid-Band 5 (30.4 GHz) Mid-Band 6 (42.7 GHz) –160 dBm –154 dBm –155 dBm –154 dBm –152 dBm –148 dBm –143 dBm –172 dBm –164 dBm –167 dBm –167 dBm –165 dBm –160 dBm –156 dBm N/A –157 dBm –157 dBm –157 dBm –157 dBm –157 dBm –150 dBm 1. LNP (Low Noise Path) requires option LNP. 2. At higher frequency bands (beyond 3.6 GHz), Preamp “On” supersedes “LNP enabled”. LNP cannot operate simultaneously with preamp.

12 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Residues, images, and spurious responses Residual responses (Input terminated and 0 dB attenuation) 200 kHz to 8.4 GHz Zero span or FFT or other frequencies –100 dBm –100 dBm nominal Image responses Tuned Freq (f) Excitation Freq Response (Mixer level at –10 dBm) 10 MHz to 26.5 GHz 10 MHz to 3.6 GHz 10 MHz to 3.6 GHz 3.5 to 13.6 GHz 13.5 to 17.1 GHz 17.0 to 22 GHz 22 to 26.5 GHz f+45 MHz f+10,245 MHz f+645 MHz f+645 MHz f+645 MHz f+645 MHz f+645 MHz –80 dBc –118 dBc typical –80 dBc –112 dBc typical –80 dBc –101 dBc typical –78 dBc –87 dBc typical –74 dBc –84 dBc typical –70 dBc –82 dBc typical –68 dBc –79 dBc typical (Mixer level at –30 dBm) 26.5 to 34.5 GHz 34.4 to 44 GHz 44 to 50 GHz f+645 MHz f+645 MHz f+645 MHz –68 dBc –84 dBc typical –57 dBc –79 dBc typical –75 dBc nominal Other spurious responses Mixer level Response Carrier frequency ≤ 26.5 GHz First RF order (f ≥ 10 MHz from carrier) Higher RF order (f ≥ 10 MHz from carrier) Carrier frequency > 26.5 GHz First RF order (f ≥ 10 MHz from carrier) Higher RF order (f ≥ 10 MHz from carrier) –10 dBm –40 dBm –30 dBm –30 dBm –80 dBc + 20log(N 1) Including IF feedthrough, LO harmonic mixing responses –80 dBc + 20log(N 1) Including higher order mixer responses –90 dBc nominal –90 dBc nominal LO-related spurious responses (200 Hz ≤ f < 10 MHz from carrier), Mixer level at –10 dBm –68 dBc 2 + 20log(N 1) Line-related spurious responses –73 dBc 2 + 20log(N 1) (nominal) Second harmonic distortion (SHI) Source frequency Mixer level Distortion 3 SHI 3 RF/MW (Option 503, 508, 513, 526) 10 to 100 MHz 0.1 to 1.8 GHz 1.75 to 2.5 GHz 2.5 to 4 GHz 4 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –57 dBc/NA –60 dBc/NA –77 dBc/–95 dBc –77 dBc/–101 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc +42 dBm/NA +45 dBm/NA +62 dBm/+80 dBm +62 dBm/+86 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm Preamp level Distortion SHI Preamp on (Option P03, P08, P13, P26) 10 MHz to 1.8 GHz 1.8 to 13.25 GHz –45 dBm –50 dBm –78 dBc nominal –60 dBc nominal +33 dBm nominal +10 dBm nominal Millimeter-Wave Mixer level Distortion

3 SHI 3

(Option 543, 544, 550) 10 to 100MHz 100 M to 1.8 GHz 1.8 to 2.5 GHz 2.5 to 3 GHz 3 to 5 GHz 5 to 6.5 GHz 6.5 to 10 GHz 10 to 13.25 GHz 13.25 to 25 GHz –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –15 dBm –57 dBc/NA –60 dBc/NA –72 dBc/–95 dBc –72 dBc/–99 dBc –77 dBc/–99 dBc –77 dBc/–105 dBc –70 dBc/–105 dBc –62 dBc/–105 dBc –65 dBc/–105 dBc (nom.) +42 dBm/NA +45 dBm/NA +57 dBm/+80 dBm +57 dBm/+84 dBm +62 dBm/+84 dBm +62 dBm/+90 dBm +55 dBm/+90 dBm +47 dBm/+90 dBm +50 dBm/+90 dBm (nom.) Preamp on (Option P03, P08, P13, P26, P43, P44, P50) Preamp level Distortion SHI 10 MHz to 1.8 GHz 1.8 to 13.25 GHz 13.25 to 25 GHz –45 dBm –50 dBm –50 dBm –78 dBc (nominal) –60 dBm (nominal) –50 dBm (nominal) +33 dBm (nominal) +10 dBm (nominal) 0 dBm (nominal) 1. N is the LO multiplication factor. Refer to page 4 for the N value verses frequency ranges. 2. Nominally –40 dBc under large magnetic (0.38 Gauss rms) or vibrational (0.21 g rms) environmental stimuli. 3. Normal path/LNP enabled (requires Option LNP).

Figure 1. Nominal TOI performance versus frequency and tone separation

14 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Figure 2a. Third-order dynamic range plots Figure 2b. Third-order dynamic range plots Nominal Dynamic Range at 1 GHz DANL and distortion relative mixer level (dB) Mixer Level (dBm) -60 -70 -80 -90 -100 -110 -120 -130 Nominal Dynamic Range at 13 GHz DANL and distortion relative mixer level (dB) Mixer Level (dBm) -60 -70 -80 -90 -100 -110 -120 -130 Nominal Dynamic Range at 26.5 GHz DANL and distortion relative mixer level (dB) Mixer Level (dBm) -60 -70 -80 -90 -100 -110 -120 -130 Nominal Dynamic Range at 50 GHz DANL and distortion relative mixer level (dB) Mixer Level (dBm) -60 -70 -80 -90 -100 -110 -120 -130 ––– DANL (30 kHz RBW) ––– DANL (1 Hz RBW) – – 2nd Harmonic Distortion ––– 3rd Order Intermodulation

1 MHz

  1. When Option MPB is installed and enabled, some aspects of the analyzer performance change. Please refer to the PXA specification guide for

Figure 3. Nominal PXA phase noise at various center frequencies

16 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet PowerSuite Measurement Specifications Channel power Amplitude accuracy, W-CDMA or IS95 (20 to 30 °C, attenuation = 10 dB) ± 0.61 dB (± 0.19 dB 95th percentile) Occupied bandwidth Frequency accuracy ± [span/1000] nominal Adjacent channel power Accuracy, 3GPP W-CDMA (ACLR) (at specific mixer levels and ACLR ranges) Adjacent Alternate MS (UE) BTS ± 0.09 dB ± 0.18 dB ± 0.16 dB ± 0.31 dB Dynamic range (typical) Without noise correction With noise correction –81.5 dB –82.5 dB –87 dB –88 dB Offset channel pairs measured 1 to 6 Multi-carrier ACP Accuracy, 3GPP W-CDMA (ACPR) (4 carriers, 5 MHz offset, BTS, UUT ACPR range at –42 to –48 dB, optimal mixer level at –21 dBm) ± 0.13 dB Multiple number of carriers measured Up to 12 Power statistics CCDF Histogram resolution 0.01 dB Harmonic distortion Maximum harmonic number 10th Result Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in % Intermod (TOI) Measure the third-order products and intercepts from two tones Burst power Methods Power above threshold, power within burst width Results Single burst output power, average output power, maximum power, minimum power within burst, burst width Spurious emission 3GPP W-CDMA table-driven spurious signals; search across regions Dynamic range (1 to 3.6 GHz) Absolute sensitivity (1 to 3.6 GHz) 97.1 dB –86.4 dBm (101.9 dB typical) (–90.4 dBm typical) Spectrum emission mask (SEM) cdma2000 ® (750 kHz offset) Relative dynamic range Absolute sensitivity Relative accuracy 81.6 dB –101.7 dBm ± 0.08 dB (86.4 dB typical) (–105.7 dBm typical) 3GPP W-CDMA (2.515 MHz offset) Relative dynamic range Absolute sensitivity Relative accuracy 85.4 dB –101.7 dBm ± 0.08 dB (89.8 dB typical) (–105.7 dBm typical)

17 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet General Specifications Temperature range Operating Storage 0 to 55 °C –40 to +70 °C Altitude 4,500 meters (approx 15,000 feet) EMC Complies with European EMC Directive 2004/108/EC – IEC/EN 61326-1 or IEC/EN 61326-2-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11:2002 – ICES/NMB-001 This ISM device complies with Canadian ICES-001 Cet appareil ISM est conforme à la norme NMB-001 du Canada Safety Complies with European Low Voltage Directive 2006/95/EC – IEC/EN 61010-1 3rd Edition – Canada: CSA C22.2 No. 61010-1-12 – USA: UL 61010-1 3rd Edition Acoustic statement (European Machinery Directive 2002/42/EC, 1.7.4.2u) Acoustic noise emission LpA < 70 dB Operator position Normal position Per ISO 7779 Acoustic noise - more information (Values given are per ISO 7779 standard in the “Operator Sitting” position) Ambient temperature < 40 °C Nominally under 55 dBA Sound Pressure. 55 dBA is generally considered suitable for use in quiet office environment ≥ 40 °C Nominally under 65 dBA Sound Pressure. 65 dBA is generally considered suitable for use in noisy office environment Environmental stress Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, tempera - ture, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3. Power requirements Voltage and frequency 100 to 120 V, 50/60/400 Hz 220 to 240 V, 50/60 Hz Power consumption On Stanby

630 W (Maximum)

  1. The N9030A is in full compliance with CISPR 11, Class A emissions and is declared as such. In addition, the N9030A has been type tested and shown to meet CISPR 11, Class B emissions limits. Information regarding the Class B emission performance of the N9030A is provided as a convenience to the user and is not intended to be a regulatory declaration.

18 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Display Resolution Size 1024 x 768, XGA 213 mm (8.4 in.) diagonal (nominal) Data storage Internal Removable solid state drive (80 GB) External Supports USB 2.0 compatible memory devices Weight (without options) Net Shipping 22 kg (48 lbs) nominal 34 kg (75 lbs) nominal Dimensions Height Width Length 177 mm (7.0 in) 426 mm (16.8 in) 556 mm (21.9 in) Warranty The PXA signal analyzer is supplied with a 3-year standard warranty Calibration cycle The recommended calibration cycle is one year. Calibration services are available through Keysight service centers Inputs and Outputs Front panel RF input Connector Standard (Option 503, 508, 513, 526) Option C35 (with Option 526 only) Standard (Option 543, 544, 550) Type-N female, 50 Ω nominal APC 3.5 mm male, 50 Ω nominal 2.4 mm male, 50 Ω nominal Analog baseband IQ inputs (Option BBA) 1 Connectors (I, Q, I-Bar, Q-Bar, and Cal Out) Cal Out Signal Frequency Input impedance (4 connectors: I, Q, I-, Q-) Probes supported Active probe Passive probe Input return loss 50 Ω impedance only selected BNC female AC coupled square wave Selectable between 1 kHz and 250 kHz 50 Ω, 1 MΩ (selectable, nominal) 1130A, 1131A, 1132A, 1134A 1161A –5 dB (0 to 10 MHz, nominal) –0 dB (10 to 40 MHz. nominal) Probe power Voltage/current +15 Vdc, ± 7% at 150 mA max nominal –12.6 Vdc, ± 10% at 150 mA max nominal USB 2.0 ports Master (2 ports) Standard Connector Output current Compatible with USB 2.0 USB Type-A female

0.5 A nominal

Headphone jack Miniature stereo audio jack (3.5 mm, also known as “1/8 inch”) 1. For additional specifications, please refer to Chapter BAA in the PXA Signal Analyzer specification guide 2. For more details, please refer to the Keysight Probe Configuration Guides, literature numbers 5968-7141EN and 5989-6162EN; probe heads are necessary to attach to your device properly and probe connectivity kits such as E2668B, E2669A. or E2675A are required.

19 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet External mixing, Option EXM Connection port Connector Impedance Functions SMA, female 50 Ω nominal Triplexed for mixer bias, IF input and LO output Mixer bias range IF input center frequency Narrowband IF path

40 MHz BW IF path

± 10 mA in 10 uA step

322.5 MHz

250.0 MHz

300 MHz

3.75 to 14.0 GHz Rear panel

10 MHz out

BNC female, 50 Ω nominal ≥ 0 dBm nominal

10 MHz + (10 MHz x frequency reference accuracy)

BNC female, 50 Ω nominal –5 to 10 dBm nominal 1 to 50 MHz nominal (selectable to 1 Hz resolution) ± 2 x 10–6 of specified external reference input frequency Trigger 1 and 2 inputs Connector Impedance Trigger level range BNC female > 10 kΩ nominal –5 to +5 V (TTL) factory preset Trigger 1 and 2 outputs Connector Impedance Level BNC female 50 Ω nominal 0 to 5 V (CMOS) nominal Sync (reserved for future use) Connector BNC female Monitor output Connector Format Resolution VGA compatible, 15-pin mini D-SUB XGA (60 Hz vertical sync rates, non-interlaced) Analog RGB 1024 x 768 Noise source drive +28 V (pulsed) Connector Output voltage BNC female On 28.0 ± 0.1 V (60 mA maximum) Off < 1 V SNS series noise source For use with the Keysight SNS Series noise sources Digital bus Connector MDR-80

20 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Rear panel Analog out Connector BNC female USB 2.0 ports Master (4 ports) Standard Connector Output current Slave (1 port) Standard Connector Output current Compatible with USB 2.0 USB Type-A female Compatible with USB 2.0 USB Type-B female SH1, AH1, T6, SR1, RL1, PP0, DC1, C1, C2, C3, C28, DT1, L4, C0 Controller or device LAN TCP/IP interface Standard Connector 1000Base-T RJ45 Ethertwist IF output Connector Impedance SMA female, shared by Opts CR3, CRP, and ALV 50 Ω nominal 2nd IF output, Option CR3 Center frequency SA mode or I/Q analyzer with IF BW ≤ 25 MHz with Option B40 with Option B85/B1X

250 MHz

Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Low band High band, with preselector High band, with preselector bypassed Up to 160 MHz (nominal) Depends on center frequency Up to 700 MHz (nominal); expandable to 900 MHz with corrections Arbitrary IF output, Option CRP Center frequency Range Resolution 10 to 75 MHz (user selectable)

0.5 MHz

Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Output at 70 MHz Low band or high band with preselector bypassed Preselected band

100 MHz (nominal)

Depends on RF center frequency Lower output frequencies Subject to folding Residual output signals ≤ –88 dBm (nominal) 1. The maximum bandwidth is not centered around the IF output center frequency.

21 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Option YAV Y-Axis output General port specifications Connector Impedance SMA female Shared with other options 50 Ω nominal Fast log video output Output voltage Maximum Slope Open-circuit voltages shown

1.6 V at –10 dBm nominal

25 ± 1 mV/dB nominal Log fidelity Range Accuracy within range 49 dB (nominal) with input frequency at 1 GHz ± 1.0 dB nominal Rise time 15 ns nominal Fall time Bands 1-4 with Option MPB Other cases 40 ns nominal best case, Depends on bandwidth Other Optional Output Option ALV Log video out General port specifications Connector Impedance BNC female Shared with other options 50 Ω nominal Screen video Operating conditions Display scale types Log scales Modes Gating Log or Lin All (0.1 to 20 dB/div) Spectrum analyzer only Gating must be off “Lin” is linear in voltage Output scaling Offset Gain accuracy 0 to 1.0 V open circuit, representing bottom to top of screen ± 1% of full scale nominal ± 1% of output voltage nominal Delay between RF input to analog output 71.7 µs +2.56/RBW + 0.159/VBW nominal Log video (Log envelope) output Amplitude range (terminated with 50 Ω) Maximum 1.0 V nominal for –10 dBm at the mixer Scale factor Bandwidth Operating conditions 1 V per 192.66 dB Set by RBW Select Sweep Type = Swept Linear video (AM Demod) output Amplitude range (terminated with 50 Ω) Maximum Minimum

1.0 V nominal for signal envelope at the reference level

Scale factor If carrier level is set to half the reference level in volts, the scale factor is 200% of carrier level per volt. Regardless of the carrier level, the scale factor is 100% of reference level per volt. Bandwidth Operating conditions Set by RBW Select Sweep Type = Swept

22 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Frequency Frequency span Standard instrument Option B25 Option B40 Option B85 Option B1X

10 Hz to 10 MHz

10 Hz to 25 MHz

10 Hz to 40 MHz

10 Hz to 85 MHz

10 Hz to 160 MHz

Resolution bandwidth (spectrum measurement) Range Overall Span = 1 MHz Span = 10 kHz Span = 100 Hz Window shapes 100 mHz to 3 MHz

50 Hz to 3 MHz

1 Hz to 10 kHz

Flat Top, Uniform, Hanning, Hamming, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB) Analysis bandwidth (waveform measurement) Standard instrument Option B25 Option B40 Option B85 Option B1X IF frequency response (standard 10 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency) Freq (GHz) Analysis BW (MHz) Max error Midwidth error (95th percentile) Slope (dB/MHz) (95th percentile) RMS (nominal) ≤ 3.6 3.6 to 26.5 3.6 to 26.5 26.5 to 50 26.5 to 50 ≤ 10 ≤ 10 preselected ≤ 10 preselector off ≤ 10 preselected ≤ 10 preselected off 1 ± 0.20 dB ± 0.25 dB ± 0.30 dB ± 0.12 dB ± 0.12 dB ± 0.12 dB ± 0.10 dB ± 0.10 dB ± 0.10 dB 0.02 dB 0.23 dB 0.02 dB 0.12 dB 0.024 dB I/Q Analyzer 1. Option MPB is installed and enabled.

23 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet I/Q Analyzer (continued) IF phase linearity Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.02, < 3.6 ≥ 3.6 to ≤ 26.5 ≥ 3.6 ≤ 10 ≤ 10 ≤ 10 NA Off 1 On 0.06° 0.10° 0.11° 0.012° 0.022° 0.024° Dynamic range (standard 10 MHz IF path) Clipping-to-noise dynamic range Excluding residuals and spurious responses Clipping level at mixer IF gain = Low IF gain = High –10 dBm –20 dBm Center frequency ≥ 20 MHz –8 dBm nominal –17.5 dBm nominal Noise density at mixer at center frequency (DANL + IF Gain effect) + 2.25 dB Data acquisition (standard 10 MHz IF path) Time record length Analysis tool IQ analyzer 4,000,000 IQ sample Pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2

29 Sa) 268 MSa (228 Sa) 2 GB total memory

Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 16 bits 1. Option MPB is installed and enabled.

24 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet IF frequency response (B25 IF path) IF frequency response (demodulation and FFT response relative to the center frequency) Freq (GHz) Analysis BW (MHz) Max error Midwidth error (95th percentile) Slope (dB/MHz) (95th percentile) RMS (nominal) 3.6 to 26.5 10 to ≤ 25 preselected 0.50 dB 3.6 to 26.5 10 to ≤ 25 preselector off ± 0.40 dB 0.03 dB 26.5 to 50 10 to ≤ 25 preselected 0.31 dB 26.5 to 50 10 to ≤ 25 preselector off ± 0.40 dB 0.02 dB IF phase linearity Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.02, < 3.6 ≥ 3.6 ≤ 25 ≤ 25 NA Off 0.48° 0.85° 0.12° 0.20° Dynamic range (B25 IF path) Full scale (ADC clipping) Default settings, signal at CF (IF gain = Low) Band 0 Bands 1 through 4 –8 dBm mixer level nominal –7 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Bands 1 through 4 –18 dBm mixer level nominal, subject to gain limitations –17 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal Data acquisition (B25 IF path) Time record length Analysis tool IQ analyzer 4,000,000 IQ sample Pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 16 bits 1. Option MPB is installed and enabled. I/Q Analyzer (continued) Option B25 25 MHz analysis bandwidth (Option B25 is automatically included in Option B40, B85 or B1X)

25 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet IF frequency response (B40 IF path) IF frequency response Relative to center frequency Center freq. (GHz) Span (MHz) Preselector Typical RMS (nominal) ≥ 0.03, < 3.6 ≥ 3.6, ≤ 8.4 > 8.4, ≤ 26.5 ≥ 26.5, < 34.4 ≥ 34.4, < 50 ≤ 40 ≤ 40 ≤ 40 ≤ 40 ≤ 40 NA Off Off 1 Off 1 Off 1 ± 0.4 dB ± 0.4 dB ± 0.7 dB ± 0.8 dB ± 1.0 dB ± 0.25 dB ± 0.16 dB ± 0.20 dB ± 0.25 dB ± 0.35 dB 0.05 dB 0.05 dB 0.05 dB 0.1 dB 0.1 dB IF phase linearity (deviation from mean phase linearity) Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.03, < 3.6 ≥ 3.6 ≤ 40 ≤ 40 NA Off 0.16° 1.5° 0.041° 0.35° EVM (EVM measurement floor for an 802.11g OFDM signal, using 89600B software equalization, channel estimation and data EQ)

2.4 GHz

5.8 GHz with Option MPB

–52.0 dB (0.25%) nominal –49.1 dB (0.35%) nominal Dynamic range (B40 IF path) SFDR (Spurious-free dynamic range) Signal frequency within ± 12 MHz of center –80 dBc nominal Signal frequency anywhere within analysis BW Spurious response within ± 18 MHz of center –79 dBc nominal Response anywhere within analysis BW –77 dBc nominal Full scale (ADC clipping) Default settings, signal at CF (IF gain = Low: IF gain offset = 0 dB) Band 0 Bands 1 through 4 –8 dBm mixer level nominal –7 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Bands 1 through 4 –18 dBm mixer level nominal, subject to gain limitations –17 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal 1. Option MPB is installed and enabled. I/Q Analyzer (continued) Option B40 40 MHz analysis bandwidth (Option B40 is automatically included in Option B85 or B1X)

26 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Data acquisition (B40 IF path) Time record length Analysis tool IQ analyzer 4,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 29 Sa) 268 MSa (228 Sa) 2 GB total memory Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 12 bits I/Q Analyzer (continued) Option B40 40 MHz analysis bandwidth IF frequency response (B85 or B1X IF path) IF frequency response Relative to center frequency Center freq. (GHz) Span (MHz) Preselector Typical RMS (nominal) ≥ 0.1, < 3.6 ≤ 85 ≤ 140 ≤ 160 NA NA NA ± 0.6 dB ± 0.6 dB ± 0.17 dB ± 0.25 dB ± 0.2 dB (nom) 0.05 dB 0.05 dB 0.07 dB ≥ 3.6, ≤ 8.4 ≤ 85 ≤ 140 ≤ 160 Off Off 1 Off 1 ± 0.73 dB ± 0.8 dB ± 0.2 dB ± 0.35 dB ± 0.3 dB (nom) 0.05 dB 0.05 dB 0.07 dB > 8.4, ≤ 26.5 ≤ 85 ≤ 140 ≤ 160 Off Off 1 Off 1 ± 1.10 dB ± 1.30 dB ± 0.50 dB ± 0.75 dB ± 0.5 dB (nom) 0.1 dB 0.1 dB 0.12 dB ≥ 26.5, ≤ 50 ≤ 85 ≤ 140 Off Off 1 ± 1.20 dB ± 1.40 dB ± 0.45 dB ± 0.65 dB 0.12 dB 0.12 dB IF phase linearity (deviation from mean phase linearity) Center freq (GHz) Span (MHz) Preselector Peak-to-peak (nominal) RMS (nominal) ≥ 0.03, < 3.6 ≥ 3.6, ≤ 140 ≤ 160 ≤ 140 ≤ 160 NA NA Off Off 1 0.9° 1.7° 1.6° 2.8° 0.20° 0.42° 0.39° 0.64° EVM (EVM measurement floor) Customized settings required, preselector bypassed (Option MPB) above Band 0 Case 1: 62.5 Msymbol/s, 16QAM signal, RRC filter alpha of 0.2, non-equalized, with approximately 75 MHz occupied bandwidth Band 0, 1.8 GHz Band 1, 5.95 GHz 0.8% nominal 1.1% nominal Case 2: 104.167 Msymbol/s, 16QAM signal, RRC filter alpha of 0.35, non-equalized, with approximately 140 MHz occupied bandwidth Band 1, 5.95 GHz Band 2, 15.3 GHz Band 4, 26 GHz 3.0% nominal, (unequalized) 2.5% nominal, (unequalized) 3.5% nominal, (unequalized) 0.5% nominal, (equalized) 0.6% nominal, (equalized) 1.6% nominal, (equalized) Effect of signal frequency ≠ CF Up to ± 3 dB nominal 1. Option MPB is installed and enabled. Option B85 85 MHz or B1X 160 MHz analysis bandwidth

27 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet Dynamic range (B85 or B1X IF path) SFDR (Spurious-free dynamic range) Signal frequency within ± 12 MHz of center –75 dBc nominal Signal frequency anywhere within analysis BW Spurious response within ± 63 MHz of center –74 dBc nominal Response anywhere within analysis BW –72 dBc nominal Full scale (ADC clipping) Default settings, signal at CF (IF gain = Low: IF gain offset = 0 dB) Band 0 Band 1 through 4 –8 dBm mixer level nominal –7 dBm mixer level nominal High gain setting, signal at CF (IF gain = High) Band 0 Band 1 through 4 –18 dBm mixer level nominal, subject to gain limitations –17 dBm mixer level nominal, subject to gain limitations Effect of signal frequency ≠ CF Up to ± 3 dB nominal Data acquisition (B85 or B1X IF path) Time record length Analysis tool IQ analyzer 4,000,000 IQ sample pairs Waveform measurement Advanced tools Data packing 89600 VSA software or fast capture32-bit 64-bit Length (IQ sample pairs) 536 MSa (2 Length (time units) Samples/Sample rate (IQ pair) Sample rate IQ pairs Span x 1.25 ADC resolution 14 bits Real-time analysis Real-time analysis bandwidth Option RT1 Option RT2 Up to 160 MHz Up to 160 MHz Analysis BW option determines the max real-time bandwidth Analysis BW option determines the max real-time bandwidth Minimum detectable signal duration with > 60 dB StM 2 ratio Option RT1 Option RT2 11.42 ns 5.0 ns Minimum signal duration with 100% probability of intercept (POI) at full ampli - tude accuracy Option RT1 Option RT2 17.3 μs 3.57 μs For Frequency Mask Triggering (FMT) Signal is at mask level Signal is at mask level Minimum acquisition time FFT rate 100 μs 292,969/s 1. For additional RTSA specifications, please refer to Option RT1/RT2 Chapter in the PXA Signal Analyzer specifications guide 2. StM = “Signal-to-Mask” I/Q Analyzer (continued) Option B85 85 MHz or B1X 160 MHz analysis bandwidth Real-time spectrum analyzer (RTSA) 1 Option RT1 or RT2

www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. cdma2000® is a registered certification mark of the Telecommunications Industry Association. www.keysight.com/find/N9030A www.keysight.com/find/PXA www.keysight.com/find/xseries_apps For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 0800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-07-10-14) 28 | Keysight | N9030A PXA X-Series Signal Analyzer - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2013 – 2015 Published in USA, February 13, 2015 5990-3952EN www.keysight.com Related Literature Keysight PXA signal analyzers Brochure 5990-3951EN Configuration guide 5990-3953EN