EXA X-Series Signal Analyzer, Multi-touch N9010B 10 Hz to 3.6, 7.0, 13.6, 26.5, 32, or 44 GHz Data Sheet
02 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet This data sheet is a summary of the specifications and conditions for EXA signal analyzers. For the complete specifications guide, visit: www.keysight.com/find/exa_specifications Cost-Effective Millimeter-Wave Signal Analysis Whether you’re focused on time-to- market, time-to-volume, or cost of test, your choice of economy class signal analyzer should help you save both time and money. That’s the idea that drives the Keysight Technologies, Inc. EXA signal analyzer—your first, best choice when you need maximum value in signal analysis up to millimeter- wave frequencies. It helps you find the answer faster, whether you're seeking tighter design margins or shorter test times. Table of Contents
03 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Get More Information This EXA signal analyzer data sheet is a summary of the specifi - cations and conditions for N9010B EXA signal analyzers. A full set of specifications are available in the EXA Signal Analyzer Specification Guide at www.keysight.com/find/ exa_specifications . For ordering information, refer to the N9010B EXA Signal Analyzer Configuration Guide literature number (5992-1253EN). Definitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to the full temperature range of 0 to 55 °C, unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2 s) of perfor- mance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the statistical observations of a sample of instruments, these values include the effects of the uncertainties of external calibration references. These values are not warranted. These values are updated occasionally if a significant change in the statistically observed behavior of production instruments is observed. Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specifications that 80 percent of the units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C. Typical performance does not include measurement uncertainty. Nominal values indicate expected performance, or describe product performance that is useful in the application of the product, but are not covered by the product warranty. The analyzer will meet its specifications when: – It is within its calibration cycle – Under auto couple control, except when Auto Sweep Time Rules = Accy – Signal frequencies < 10 MHz, with DC coupling applied – The analyzer has been stored at an ambient temperature within the allowed operat - ing range for at least two hours before being turned on; if it had previously been stored at a temperature range inside the allowed storage range, but outside the allowed operating range – The analyzer has been turned on at least 30 minutes with Auto Align set to Normal, or if Auto Align is set to Off or Partial, alignments must have been run recently enough to prevent an Alert message. If the Alert condition is changed from “Time and Temperature” to one of the disabled duration choices, the analyzer may fail to meet specifications without informing the user. If Auto Align is set to Light, perfor - mance is not warranted, and nominal performance will degrade to become a factor of 1.4 wider for any specification subject to alignment, such as amplitude tolerances
04 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency and Time Specifications Frequency range DC coupled AC coupled Option 503 10 Hz to 3.6 GHz 10 MHz to 3.6 GHz Option 507 10 Hz to 7 GHz 10 MHz to 7 GHz Option 513 10 Hz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 10 Hz to 26.5 GHz 10 MHz to 26.5 GHz Option 532 10 Hz to 32 GHz NA Option 544 10 Hz to 44 GHz NA Band LO multiple (N) 0 1 10 Hz to 3.6 GHz 1 1 3.5 to 7.0 GHz 1 1 3.5 to 8.4 GHz 2 2 8.4 to 13.6 GHz 3 2 13.5 to 17.1 GHz 4 4 17 to 26.5 GHz 5 4 26.4 to 34.5 GHz 6 8 34.4 to 44 GHz Frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate Option PFR Standard ± 1 x 10–7 / year ± 1 x 10–6 / year ± 1.5 x 10–7 / 2 years Temperature stability Option PFR Standard 20 to 30 °C ± 1.5 x 10–8 ± 2 x 10–6 Full temperature range ± 5 x 10–8 ± 2 x 10–6 Achievable initial calibration accuracy Option PFR Standard ± 4 x 10–8 ± 1.4 x 10–6 Example frequency reference accuracy (with Option PFR) = ± (1 x 10–7 + 5 x 10–8 + 4 x 10–8) 1 year after last adjustment = ± 1.9 x 10–7 Residual FM Option PFR Standard ≤ (0.25 Hz x N) p-p in 20 ms nominal ≤ (10 Hz x N) p-p in 20 ms nominal See band table above for N (LO Multiple) Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution 1) Marker frequency counter Accuracy ± (marker frequency x frequency reference accuracy + 0.100 Hz) Delta counter accuracy ± (delta frequency x frequency reference accuracy + 0.141 Hz) Counter resolution 0.001 Hz Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept FFT ± (0.25 % x span + horizontal resolution) ± (0.10 % x span + horizontal resolution) 1. Horizontal resolution is span/(sweep points – 1).
05 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Sweep time and triggering Range Span = 0 Hz 1 μs to 6000 s Span ≥ 10 Hz 1 ms to 4000 s Accuracy Span ≥ 10 Hz, swept ± 0.01% nominal Span ≥ 10 Hz, FFT ± 40% nominal Span = 0 Hz ± 0.01% nominal Trigger Free run, line, video, external 1, external 2, RF burst, periodic timer Trigger Delay Span = 0 Hz or FFT –150 to +500 ms Span ≥ 10 Hz, swept 0 to 500 ms Resolution 0.1 μs Time gating Gate methods Gated LO; gated video; gated FFT Gate length range (except method = FFT) 100.0 ns to 5.0 s Gate delay range 0 to 100.0 s Gate delay jitter 33.3 ns p-p nominal Sweep (trace) point range All spans 1 to 40001 Resolution bandwidth (RBW) Range (–3.01 dB bandwidth) 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz Bandwidth accuracy (power) 1 Hz to 750 kHz ± 1.0 % (± 0.044 dB) 820 kHz to 1.2 MHz (< 3.6 GHz CF) ± 2.0 % (± 0.088 dB) 1.3 to 2 MHz (< 3.6 GHz CF) ± 0.07 dB nominal 2.2 to 3 MHz (< 3.6 GHz CF) 0 to –0.2 dB nominal 4 to 8 MHz (< 3.6 GHz CF) 0 to –0.4 dB nominal Bandwidth accuracy (–3.01 dB) RBW range 1 Hz to 1.3 MHz ± 2 % nominal Selectivity (–60 dB/–3 dB) 4.1:1 nominal EMI bandwidth (CISPR compliant) 200 Hz, 9 kHz, 120 kHz, 1 MHz (Option EMC required) EMI bandwidth (MIL STD 461E compliant) 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz (Option EMC required) Analysis bandwidth 1 Maximum bandwidth Option B40 40 MHz Standard 25 MHz Video bandwidth (VBW) Range 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz) Accuracy ± 6 % nominal 1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or processing in the time, frequency, or modulation domain.
06 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Amplitude Accuracy and Range Specifications Amplitude range Measurement range Displayed average noise level (DANL) to +23 dBm Input attenuator range (10 Hz to 44 GHz) Standard Option FSA 0 to 60 dB in 10 dB steps 0 to 60 dB in 2 dB steps Electronic attenuator (Option EA3) Frequency range 10 Hz to 3.6 GHz Attenuation range Electronic attenuator range Full attenuation range (mechanical + electronic) 0 to 24 dB, 1 dB steps 0 to 84 dB, 1 dB steps Maximum safe input level Average total power (with and without preamp) +30 dBm (1 W) Peak pulse power < 10 μs pulse width, < 1 % duty cycle +50 dBm (100 W) and input attenuation ≥ 30 dB DC volts DC coupled AC coupled ± 0.2 Vdc ± 100 Vdc Display range Log scale 0.1 to 1 dB/division in 0.1 dB steps 1 to 20 dB/division in 1 dB steps (10 display divisions) Linear scale 10 divisions Scale units dBm, dBmV, dBμV, dBmA, dBμA, V, W, A
07 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency response Specification 95th percentile (≈ 2 s) (10 dB input attenuation, 20 to 30 °C, preselector centering applied, s = nominal standard deviation) RF/MW (Option 503, 507, 513, 526) 9 kHz to 10 MHz ± 0.8 dB ± 0.4 dB 10 MHz 1 to 3.6 GHz ± 0.6 dB ± 0.21 dB 3.5 to 7.0 GHz ± 2.0 dB ± 0.69 dB 7.0 to 13.6 GHz ± 2.5 dB 13.5 to 22.0 GHz ± 3.0 dB 22.0 to 26.5 GHz ± 3.2 dB Millimeter-wave (Option 532, 544) 9 kHz to 10 MHz ± 0.6 dB ± 0.28 dB 10 to 50 MHz ± 0.45 dB ± 0.21 dB 50 MHz to 3.6 GHz ± 0.45 dB ± 0.20 dB 3.5 to 5.2 GHz ± 1.7 dB ± 0.91 dB 5.2 to 8.4 GHz ± 1.5 dB ± 0.61 dB 8.3 to 13.6 GHz ± 2.0 dB ± 0.61 dB 13.5 to 17.1 GHz ± 2.0 dB ± 0.67 dB 17.0 to 22.0 GHz ± 2.0 dB ± 0.78 dB 22.0 to 26.5 GHz ± 2.5 dB ± 0.72 dB 26.4 to 34.5 GHz ± 2.5 dB ± 1.11 dB 34.4 to 44 GHz ± 3.2 dB ± 1.42 dB Preamp on (P03, P07, P13, P26) RF/MW (Option 503, 507, 513, 526) 100 kHz to 3.6 GHz ± 0.28 dB nominal 3.6 to 7.0 GHz ± 0.67 dB nominal 7.0 to 26.5 GHz ± 0.80 dB nominal Preamp on (P03, P07, P32, P44) Millimeter-wave (Option 532, 544) 100 kHz to 3.6 GHz ± 0.28 dB nominal 3.5 to 8.4 GHz ± 0.67 dB nominal 8.4 to 26.5 GHz ± 0.80 dB nominal 26.4 to 44 GHz ± 0.80 dB nominal 1. DC coupling required to meet specifications below 50 MHz. With AC coupling, specifications apply at frequencies of 50 MHz and higher. Statistical observations at 10 MHz with AC coupling show that most instruments meet the DC-coupled specifications, however, a small percentage of instruments are expected to have errors exceeding 0.5 dB at 10 MHz at the temperature extreme. The effect at 20 to 50 MHz is negligible but not warranted.
08 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Input attenuation switching uncertainty Specifications Additional information Attenuation > 2 dB, preamp off 50 MHz (reference frequency) ± 0.20 dB ± 0.08 dB typical Relative to 10 dB (reference setting) 9 kHz to 3.6 GHz ± 0.3 dB nominal 3.5 to 7.0 GHz ± 0.5 dB nominal 6.9 to 13.6 GHz ± 0.7 dB nominal 13.5 to 26.5 GHz ± 0.7 dB nominal > 26.5 GHz ± 1.0 dB nominal Total absolute amplitude accuracy (10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any reference level, any scale, s = nominal standard deviation) At 50 MHz At all frequencies 9 kHz to 3.6 GHz ± 0.40 dB ± (0.40 dB + frequency response) ± 0.27 dB (95th percentile ≈ 2 s) Preamp on 100 kHz to 3.6 GHz ± (0.39 dB + frequency response) Input voltage standing wave ratio (VSWR) (≥ 10 dB input attenuation) Options 503, 507, 513, 526 Options 532, 544 10 MHz to 3.6 GHz < 1.2:1 nominal 1.2:1 nominal 3.6 to 26.5 GHz < 1.9:1 nominal 1.5:1 nominal 26.5 to 44 GHz N/A < 1.8:1 nominal Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 3 MHz RBW ± 0.10 dB 4, 5, 6, 8 MHz RBW ± 1.0 dB Reference level Range Log scale –170 to +23 dBm in 0.01 dB steps Linear scale Same as Log (707 pV to 3.16 V) Accuracy 0 dB Display scale switching uncertainty Switching between linear and log 0 dB Log scale/div switching 0 dB Display scale fidelity Between –10 dBm and –80 dBm input mixer level ± 0.15 dB total Trace detectors Normal, peak, sample, negative peak, log power average, RMS average, and voltage average Preamplifier (Option P03, P07, P13, P26, P32, P44) Frequency range Option P03 100 kHz to 3.6 GHz Option P07 100 kHz to 7 GHz Option P13 100 kHz to 13.6 GHz Option P26 100 kHz to 26.5 GHz Option P32 100 kHz to 32 GHz Option P44 100 kHz to 44 GHz Gain 100 kHz to 3.6 GHz +20 dB nominal 3.6 to 7.0 GHz +35 dB nominal > 7 GHz +40 dB nominal Noise figure 100 kHz to 3.6 GHz 8 to 12 dB nominal (proportional to frequency) 3.6 to 8.4 GHz 9 dB nominal 8.4 to 13.6 GHz 10 dB nominal > 13.6 GHz DANL + 176.24 dB nominal
09 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Dynamic Range Specifications 1 dB gain compression (two-tone) Total power at mixer input RF/MW (Option 503, 507, 513, 526) 20 MHz to 26.5 GHz +9 dBm nominal Total power at mixer input Millimeter-wave (Option 532, 544) 20 MHz to 26.5 GHz +6 dBm nominal 26.5 to 44 GHz 0 dBm nominal Total power at preamp input Preamp on 10 MHz to 3.6 GHz 3.6 to 26.5 GHz Tone spacing: 100 kHz to 20 MHz Tone spacing: > 70 MHz > 26.5 GHz –14 dBm nominal –28 dBm nominal –20 dBm nominal –30 dBm nominal Displayed average noise level (DANL) (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C) Specification Typical RF/MW (Option 503, 507, 513, 526) 1 to 10 MHz –147 dBm –149 dBm 10 MHz to 2.1 GHz –148 dBm –150 dBm 2.1 to 3.6 GHz –147 dBm –149 dBm 3.5 to 7.0 GHz –147 dBm –149 dBm 7.0 to 13.6 GHz –143 dBm –147 dBm 13.5 to 20 GHz –137 dBm –142 dBm 20 to 26.5 GHz –134 dBm –140 dBm Preamp on, RF/MW (Option 503, 507, 513, 526) 10 MHz to 2.1 GHz –161 dBm –163 dBm 2.1 to 3.6 GHz –160 dBm –162 dBm 3.5 to 7.0 GHz –160 dBm –162 dBm 7.0 to 13.6 GHz – 160 dBm –163 dBm 13.5 to 17.1 GHz –157 dBm –160 dBm 17.0 to 20.0 GHz –155 dBm –159 dBm 20.0 to 26.5 GHz –150 dBm –156 dBm Millimeter-wave (Option 532, 544) 1 9 kHz to 1 MHz – –130 dBm 1 MHz to 1.2 GHz –152 dBm –155 dBm 1.2 to 2.1 GHz –151 dBm –154 dBm 2.1 to 3.6 GHz –149 dBm –152 dBm 3.5 to 4.2 GHz –144 dBm –147 dBm 4.2 to 8.4 GHz –145 dBm –150 dBm 8.3 to 13.6 GHz –147 dBm –150 dBm 13.5 to 20 GHz –145 dBm –148 dBm 20 to 26.5 GHz –142 dBm –145 dBm 26.4 to 34 GHz –140 dBm –144 dBm 34.4 to 44 GHz –135 dBm –140 dBm 1. Without Option B40, DP2, or MPB. When any of these options are installed, performance may change. Please refer to the EXA specifications guide for more details.
10 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Displayed average noise level (DANL) (continued) Preamp on, millimeter-wave (Option 532, 544) 1 100 kHz to 1 MHz –145 dBm –148 dBm 1 to 10 MHz –161 dBm –165 dBm 10 MHz to 1.2 GHz –164 dBm –165 dBm 1.2 to 2.1 GHz –163 dBm –164 dBm 2.1 to 3.6 GHz –162 dBm –163 dBm 3.5 to 7 GHz –160 dBm –162 dBm 7 to 20 GHz –160 dBm –162 dBm 20 to 26.5 GHz –158 dBm –160 dBm 26.5 to 32 GHz –156 dBm –159 dBm 32 to 34 GHz –156 dBm –159 dBm 33.9 to 40 GHz –153 dBm –155 dBm 40 to 44 GHz –149 dBm –153 dBm DANL with Noise Floor Extension (Option NFE) on Improvement @ 95th percentile RF/MW (Option 503, 507, 513, 526) Frequency band Preamp Off Preamp On Band 0, f > 20 MHz 9 dB 9 dB Band 1 9 dB 8 dB Band 2 9 dB 9 dB Band 3 11 dB 9 dB Band 4 9 dB 8 dB Example of effective DANL @ 18-30 °C Frequency Preamp Off Preamp On Mid-Band 0 (1.8 GHz) –156 dBm –170 dBm Mid-Band 1 (5.9 GHz) –155 dBm –168 dBm Mid-Band 2 (10.95 GHz) –153 dBm –168 dBm Mid-Band 3 (15.3 GHz) –147 dBm –165 dBm Mid-Band 4 (21.75 GHz) –145 dBm –157 dBm Millimeter-Wave (Option 532, 544) 1 Frequency band Preamp Off Preamp On Band 0, f > 20 MHz 7 dB 9 dB Band 1 8 dB 7 dB Band 2 8 dB 7 dB Band 3 8 dB 7 dB Band 4 8 dB 6 dB Band 5 9 dB 6 dB Band 6 9 dB 5 dB Example of effective DANL @ 18-30 °C Frequency Preamp Off Preamp On Mid-Band 0 (1.8 GHz) –157 dBm –169 dBm Mid-Band 1 (5.9 GHz) –152 dBm –166 dBm Mid-Band 2 (10.95 GHz) –154 dBm –165 dBm Mid-Band 3 (15.3 GHz) –153 dBm –164 dBm Mid-Band 4 (21.75 GHz) –148 dBm –164 dBm Mid-Band 5 (30.4 GHz) –145 dBm –160 dBm Mid-Band 6 (42.7 GHz) –142 dBm –154 dBm 1. Without Option B40, DP2, or MPB. When any of these options are installed, performance may change. Please refer to the EXA specifications guide for more details.
11 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Spurious responses Residual responses (input terminated and 0 dB attenuation) 200 kHz to 8.4 GHz (swept) –100 dBm Zero span or FFT or other frequencies –100 dBm nominal Tuned frequency (f) Mixer level Response Image responses (Excitation freq. = f + 645 MHz) 10 MHz to 3.6 GHz –10 dBm –80 dBc (–107 dBc typical) 3.6 to 13.6 GHz –10 dBm –75 dBc (–87 dBc typical) 13.6 to 17.1 GHz –10 dBm –71 dBc (–85 dBc typical) 17.1 to 22 GHz –10 dBm –68 dBc (–82 dBc typical) 22 to 26.5 GHz –10 dBm –66 dBc (–78 dBc typical) 26.5 to 34.5 GHz –30 dBm –70 dBc (–94 dBc typical) 34.5 to 44 GHz –30 dBm –60 dBc (–79 dBc typical) LO related spurious (f > 600 MHz from carrier, 10 MHz to 3.6 GHz) 10 MHz to 3.6 GHz –90 dBc + 20 logN 1 typical Other spurious response Mixer level Response Carrier frequency ≤ 26.5 GHz First RF order (f ≥ 10 MHz from carrier) –10 dBm –80 dBc + 20log(N 1) Including IF feedthrough, LO harmonic mixing responses Higher RF order (f ≥ 10 MHz from carrier) –40 dBm –80 dBc + 20log(N 1) Including higher order mixer responses Carrier frequency > 26.5 GHz First RF order (f ≥ 10 MHz from carrier) –30 dBm –90 dBc nominal Higher RF order (f ≥ 10 MHz from carrier) –30 dBm –90 dBc nominal 1. N is the LO multiplication factor.
12 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Second harmonic distortion (SHI) Source frequency SHI (nominal) RF/MW (Option 503, 507, 513, 526) 10 MHz to 1.8 GHz +45 dBm 1.75 to 7.0 GHz +65 dBm 7.0 to 11.0 GHz +55 dBm 11.0 to 13.25 GHz +50 dBm Millimeter-wave (Option 532, 544) 10 MHz to 1.8 GHz +45 dBm 1.8 to 6.5 GHz +65 dBm 6.5 to 10 GHz +60 dBm 10 to 13.25 GHz +55 dBm 13.25 to 22 GHz +50 dBm Third-order intermodulation distortion (TOI) (Two –30 dBm tones at input mixer with tone separation > 5 times IF prefilter bandwidth, 20 to 30 °C, see Specifications Guide for IF prefilter bandwidths) TOI TOI (typical) RF/MW (Option 503, 507, 513, 526) 100 to 400 MHz +13 dBm +17 dBm 400 MHz to 3.6 GHz +14 dBm +18 dBm 3.6 to 13.6 GHz +14 dBm +18 dBm 13.6 to 26.5 GHz +12 dBm +16 dBm Preamp on, RF/MW (Option 503, 507, 513, 526) 30 MHz to 3.6 GHz (two –45 dBm tones at preamp) 3.6 to 26.5 GHz (two –50 dBm tones at preamp) 0 dBm nominal –18 dBm nominal Millimeter-wave (Option 532, 544) 10 to 100 MHz +12 dBm +17 dBm 100 MHz to 3.95 GHz +15 dBm +19 dBm 3.95 to 8.4 GHz +15 dBm +18 dBm 8.3 to 13.6 GHz +15 dBm +18 dBm 13.5 to 17.1 GHz +11 dBm +17 dBm 17.0 to 26.5 GHz +10 dBm +17 dBm (nominal) 26.5 to 44 GHz – +13 dBm (nominal) Preamp on, millimeter-wave (Option 532, 544) 30 MHz to 3.6 GHz (two –45 dBm tones at preamp) 0 dBm (nominal) 3.6 to 26.5 GHz (two –50 dBm tones at preamp) –18 dBm (nominal)
100 Hz –87 dBc/Hz –102 dBc/Hz
1 MHz –134 dBc/Hz –136 dBc/Hz
10 MHz – –147 dBc/Hz nominal
Figure 1. Nominal phase noise at different center frequencies.
- When Option MPB is installed and enabled, some aspects of the analyzer performance changes. Please refer to the EXA specification guide for more
phase noise computed from the 25.2 GHz observation. See the Frequency Stability section for the details of phase noise performance versus CF.
14 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet PowerSuite Measurement Specifications Channel power Amplitude accuracy, W-CDMA or IS95 (20 to 30 °C, attenuation = 10 dB) ± 1.04 dB (± 0.27 dB 95th percentile) Occupied bandwidth Frequency accuracy ± [span/1000] nominal Adjacent channel power Adjacent Alternate Accuracy, W-CDMA (ACLR) (at specific mixer levels and ACLR ranges) MS ± 0.17 dB ± 0.22 dB BTS ± 0.70 dB ± 0.57 dB Dynamic range (typical) Without noise correction With noise correction –68 dB –74 dB –73 dB –76 dB Offset channel pairs measured 1 to 6 ACP measurement and transfer time (fast method) 10 ms nominal (s = 0.2 dB) Multiple number of carriers measured Up to 12 Power statistics CCDF Histogram resolution 0.01 dB Harmonic distortion Maximum harmonic number 10th Result Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in % Intermod (TOI) Measure the third-order products and intercepts from two tones Burst power Methods Power above threshold, power within burst width Results Single burst output power, average output power, maximum power, minimum power within burst, burst width Spurious emission W-CDMA (1 to 3.6 GHz) table-driven spurious signals; search across regions Dynamic range 80.4 dB 82.9 dB typical Absolute sensitivity –82.5 dBm –86.5 dBm typical Spectrum emission mask (SEM) cdma2000 ® (750 kHz offset) Relative dynamic range (30 kHz RBW) Absolute sensitivity Relative accuracy 76.2 dB 82.8 dB typical –97.7 dBm –101.7 dBm typical ± 0.12 dB 3GPP W-CDMA (2.515 MHz offset) Relative dynamic range (30 kHz RBW) Absolute sensitivity Relative accuracy 79.3 dB 84.9 dB typical –97.7 dBm –101.7 dBm typical ± 0.15 dB
15 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet General Specifications Temperature range Operating 0 to 55 °C Storage –40 to 70 °C EMC Complies with European EMC Directive 2004/108/EC – IEC/EN 61326-1 or IEC/EN 61326-2-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11:2002 – ICES/NMB-001 This ISM device complies with Canadian ICES-001 Cet appareil ISM est conforme à la norme NMB-001 du Canada Safety Complies with European Low Voltage Directive 2006/95/EC – IEC/EN 61010-1 3rd Edition – Canada: CSA C22.2 No. 61010-1-12 – U.S.A.: UL 61010-1 3rd Edition Acoustic statement (European Machinery Directive 2002/42/EC, 1.7.4.2u) Acoustic noise emission LpA < 70 dB Operator position Normal position Per ISO 7779 Environmental stress Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3. Power requirements Voltage and frequency 100 to 120 V, 50/60/400 Hz 220 to 240 V, 50/60 Hz Power consumption On Standby
350 W maximum
Size 269 mm (10.6 in.) diagonal (nominal) capacitive multi-touch screen Data storage Internal ≥ 80 GB nominal (removable solid-state drive) External Supports USB 2.0 or 3.0 compatible memory devices Weight (without options) Net 16 kg (35 lbs) nominal Shipping 28 kg (62 lbs) nominal Dimensions Height 177 mm (7.0 in) Width 426 mm (16.8 in) Length 368 mm (14.5 in) Warranty The EXA signal analyzer is supplied with a standard 3-year warranty Calibration cycle The recommended calibration cycle is two years; calibration services are available through Keysight service centers
16 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Inputs and Outputs Front panel RF input connector Standard (Option 503, 507, 513, or 526) Standard (Option 532 or 544) Type-N female, 50 Ω nominal 2.4 mm male, 50 Ω nominal Probe power Voltage/current +15 Vdc, ± 7 % at 150 mA max nominal –12.6 Vdc, ± 10 % at 150 mA max nominal USB ports Master (3 ports) Standard Connector Output current Port marked with lightning bolt Ports not marked with lightning bolt Compatible with USB 2.0 USB Type-A female
1.2 A nominal
0.5 A nominal
External mixing, Option EXM (available only with EXA millimeter wave, Option 532 or 544) Connection port Connector Impedance Functions SMA, female 50 Ω nominal Triplexed for mixer bias, IF input and LO output Mixer bias range ± 10 mA in 10 μA step IF input center frequency Narrowband IF path
40 MHz IF path
322.5 MHz
250 MHz
LO output frequency range 3.75 to 14.0 GHz Rear panel
10 MHz out
BNC female, 50 Ω nominal ≥ 0 dBm nominal
10 MHz ± (10 MHz x frequency reference accuracy)
BNC female, 50 Ω nominal –5 to 10 dBm nominal
10 MHz nominal
± 5 x 10–6 of specified external reference input frequency Trigger 1 and 2 inputs Connector Impedance Trigger level range BNC female > 10 kΩ nominal –5 to 5 V Trigger 1 and 2 outputs Connector Impedance Level BNC female 50 Ω nominal
5 V TTL nominal
VGA compatible, 15-pin mini D-SUB XGA (60 Hz vertical sync rates, non-interlaced) analog RGB 1024 x 768
17 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Rear panel Noise source drive +28 V (pulsed) Connector BNC female SNS Series noise source connector For use with Keysight SNS Series noise sources Analog out Connector BNC female (used with N9063A analog demod app and Option YAS) USB ports Master, super speed 2 ports Compatibility USB 3.0 Connector USB Type-A female Output current 0.9 A nominal Master, stacked with LAN 1 port Compatibility USB 2.0 Connector USB Type A female Output current 0.5 A nominal Slave 1 port Standard USB 3.0 Connector USB Type-B female Output current 0.9 A nominal GPIB interface Connector IEEE-488 bus connector GPIB codes SH1, AH1, T6, SR1, RL1, PP0, DC1, C1, C2, C3, C28, DT1, L4, C0 GPIB mode Controller or device LAN TCP/IP interface Standard Connector 1000Base-T RJ45 Ethertwist IF output Connector Impedance SMA female, shared by Option CR3 and CRP 50 Ω nominal Wideband IF output, Option CR3 Center frequency SA mode or I/Q analyzer with IF BW ≤ 25 MHz with Option B40 Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Low band Up to 140 MHz (nominal) High band, with preselector Depends on center frequency High band, with preselector bypassed 1 Up to 410 MHz (nominal) Programmable IF output, Option CRP Center frequency Range Resolution 10 to 75 MHz (user selectable)
0.5 MHz
Conversion gain –1 to +4 dB (nominal) plus RF frequency response Bandwidth Output at 70 MHz center Low band or high band with preselector bypassed 1 100 MHz (nominal) Preselected band Depends on RF center frequency Lower output frequencies Subject to folding Residual output signals ≤ –88 dBm (nominal) 1. Option MPB installed and enabled.
18 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet I/Q Analyzer Frequency Frequency span Standard 10 Hz to 10 MHz Option B25 (standard) 10 Hz to 25 MHz Option B40 10 Hz to 40 MHz Resolution bandwidth (spectrum measurement) Range Overall 100 mHz to 3 MHz Span = 1 MHz 50 Hz to 1 MHz Span = 10 kHz 1 Hz to 10 kHz Span = 100 Hz 100 mHz to 100 Hz Window shapes Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB) Analysis bandwidth Standard 10 Hz to 10 MHz Option B25 (standard) 10 Hz to 25 MHz Option B40 10 Hz to 40 MHz IF frequency response (standard 10 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C) Center frequency (GHz) Span (MHz) Preselector Max. error RMS < 3.6 ≤ 10 n/a ± 0.40 dB 0.04 dB nominal ≥ 3.6 ≤ 10 on 0.25 dB nominal ≥ 3.6 ≤ 10 off 1 ± 0.45 dB 0.04 dB nominal > 26.5 (Option 532 or 544) ≤ 10 on 0.35 dB nominal IF phase linearity (deviation from mean phase linearity, nominal) Center frequency (GHz) Span (MHz) Preselector Peak-to-peak RMS Data acquisition (10 MHz IF path) Time record length IQ analyzer 4,000,000 IQ sample pairs Sample rate at ADC Option DP2, B40 or MPB 100 MSa/s None of the above 90 MSa/s ADC resolution Option DP2, B40 or MPB 16 bits None of the above 14 bits Option B25 (standard) 25 MHz analysis bandwidth IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C) Center frequency (GHz) Span (MHz) Preselector Max. error RMS ≤ 3.6 10 to ≤ 25 n/a ± 0.45 dB 0.051 dB nominal > 3.6 10 to ≤ 25 on 0.45 dB nominal > 3.6 10 to ≤ 25 off 1 ± 0.45 dB 0.071 dB nominal IF phase linearity (deviation from mean phase linearity, nominal) Center frequency (GHz) Span (MHz) Preselector Peak-to-peak RMS 1. Option MPB is installed and enabled.
19 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet Data acquisition (25 MHz IF path) Time record length (IQ pairs) IQ Analyzer 4,000,000 IQ sample pairs 89600 software 32-bit packing 64-bit packing Memory Option DP2, B40 or MPB 536 MSa 268 MSa 2 GB None of the above 4,000,000 IQ sample pairs (independent of data packing) Sample rate at ADC Option DP2, B40 or MPB 100 MSa/s None of the above 90 MSa/s ADC resolution Option DP2, B40 or MPB 16 bits None of the above 14 bits Option B40 40 MHz analysis bandwidth IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C), nominal Center frequency (GHz) Span (MHz) Preselector Max. error RMS 0.03 ≤ f < 3.6 ≤ 40 n/a ± 0.3 dB 0.08 dB 3.6 ≤ f ≤ 26.5 ≤ 40 off 1 ± 0.25 dB 0.08 dB > 26.5 ≤ 40 off 1 ± 0.25 dB 0.12 dB IF phase linearity (deviation from mean phase linearity, nominal) Center frequency (GHz) Span (MHz) Preselector Peak-to-peak RMS ≥ 3.6 40 off 1 5° 1.4° Data acquisition (40 MHz IF path) Time record length (IQ pairs) IQ Analyzer 4,000,000 samples (I/Q pairs)
89600 VSA software 32-bit packing 64-bit packing 2 GB total memory (nominal)
Length (IQ sample pairs) 536 MSa 268 MSa Length (time units) Samples/(span x 1.28) (nominal) Sample rate At ADC 200 MSa/s IQ pairs Span x 1.28 (nominal) ADC resolution 12 bits 1. Option MPB is installed and enabled. Related Literature Publication title Publication number X-Series Signal Analyzers – Brochure 5992-1316EN N9010B EXA X-Series Signal Analyzer, Multi-touch – Configuration Guide 5992-1253EN For more information or literature resources please visit the web: Product page: www.keysight.com/find/N9010B X-Series measurement applications: www.keysight.com/find/X-Series_Apps X-Series signal analyzers: www.keysight.com/find/X-Series
20 | Keysight | N9010B EXA X-Series Signal Analyzer, Multi-touch - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2016 Published in USA, January 21, 2016 5992-1256EN www.keysight.com myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with a three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. cdma2000 is a US registered certification mark of the Telecommunications Industry Association. www.keysight.com/find/exa For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-01-01-16) www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939.