CXA X-Series Signal Analyzer, Multi-touch N9000B 9 kHz to 3.0, 7.5, 13.6, or 26.5 GHz Data Sheet
02 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Leading low-cost tool The CXA is today’s leading low-cost tool for essential signal characterization. Its capabilities provide a solid foundation for cost-effective testing in general-purpose and educational applications. Table of Contents This data sheet is a summary of the specifications and conditions for CXA signal analyzers. For the complete specifications guide, visit www.keysight.com/find/cxa_specifications
03 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet For more information This CXA signal analyzer data sheet is a summary of the complete specifications and conditions for N9000B CXA signal analyzers, which are available in the CXA Signal Analyzer Specification Guide. The CXA Signal Analyzer Specification Guide can be obtained on the web at: (www.keysight.com/find/cxa_ specifications Definitions and Conditions Specifications describe the performance of parameters covered by the product warranty and apply to temperature ranges 0 to 55 °C, unless otherwise noted. 95th percentile values indicate the breadth of the population (approx. 2 s) of performance tolerances expected to be met in 95 percent of the cases with a 95 percent confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the statistical observations of a sample of instruments, these values include the effects of the uncertainties of external calibration references. These values are not warranted. These values are updated occasionally if a significant change in the statistically observed behavior of production instruments is observed. Typical describes additional product performance information that is not covered by the product warranty. It is performance beyond specifications that 80 percent of the units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C. Typical performance does not include measurement uncertainty. Nominal values indicate expected performance, or describe product performance that is useful in the application of the product, but are not covered by the product warranty. The analyzer will meet its specifications when: – It is within its calibration cycle – Under auto couple control, except when Auto Sweep Time Rules = Accy – The analyzer has been stored at an ambient temperature within the allowed operating range for at least two hours before being turned on; if it had previously been stored at a temperature range inside the allowed storage range, but outside the allowed operating range – The analyzer has been turned on at least 30 minutes with Auto Align set to Normal, or, if Auto Align is set to Off or Partial, alignments must have been run recently enough to prevent an Alert message. If the Alert condition is changed from “Time and Temperature” to one of the disabled duration choices, the analyzer may fail to meet specifications without informing the user. If Auto Align is set to Light, performance is not warranted, and nominal performance will degrade to become a factor of 1.4 wider for any specification subject to alignment, such as amplitude tolerances. For ordering information, refer to the CXA Signal Analyzer Configuration Guide (5992-1275EN).
04 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency range DC coupled AC coupled Option 503 NA 9 kHz to 3.0 GHz Option 507 NA 9 kHz to 7.5 GHz Option 513 9 kHz to 13.6 GHz 10 MHz to 13.6 GHz Option 526 9 kHz to 26.5 GHz 10 MHz to 26.5 GHz Band LO multiple (N) AC coupled RF (Option 503, 507) 0 1 9 kHz to 3.0 GHz 1 1 2.95 to 3.80 GHz 2 1 3.70 to 4.55 GHz 3 1 4.45 to 5.30 GHz 4 1 5.20 to 6.05 GHz 5 1 5.95 to 6.80 GHz 6 1 6.70 to 7.50 GHz Band LO multiple (N) AC coupled MW (Option 513, 526) 0 1 9 kHz to 3.08 GHz 1 2 2.95 to 7.58 GHz 2 2 7.45 to 9.55 GHz 3 2 9.45 to 12.60 GHz 4 2 12.50 to 13.05 GHz 4 4 12.95 to 13.80 GHz 5 4 13.40 to 15.55 GHz 6 4 15.45 to 19.35 GHz 7 4 19.25 to 21.05 GHz 8 4 20.95 to 22.85 GHz 9 4 22.75 to 24.25 GHz 10 4 24.15 to 26.55 GHz Frequency reference Accuracy ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Aging rate Option PFR ± 1 x 10–7 / year ± 1.5 x 10–7 / 2 years Standard ± 1 x 10–6 / year Temperature stability 20 to 30 °C Full temperature range Option PFR ± 1.5 x 10–8 ± 5 x 10–8 Standard ± 2 x 10–6 ± 2 x 10–6 Achievable initial calibration accuracy Option PFR ± 4 x 10–8 Standard ± 1.4 x 10–6 Example frequency reference accuracy (with Option PFR) 1 year after last adjustment = ± (1 x 1 x 10–7 + 5 x 10–8 + 4 x 10–8) = ± 1.9 x 10–7 Residual FM Option PFR Standard ≤ 0.25 Hz p-p in 20 ms nominal ≤ 10 Hz p-p in 20 ms nominal Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution 1) Marker frequency counter Accuracy ± (marker frequency x frequency reference accuracy + 0.100 Hz) Delta counter accuracy ± (delta frequency x frequency reference accuracy + 0.141 Hz) Counter resolution 0.001 Hz 1. Horizontal resolution is span/(sweep points – 1). Frequency and Time Specifications
05 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency span (FFT and swept mode) Range 0 Hz (zero span), 10 Hz to maximum frequency of instrument Resolution 2 Hz Accuracy Swept ± (0.25 % x span + horizontal resolution) FFT ± (0.10 % x span + horizontal resolution) Sweep time and triggering Range Span = 0 Hz 1 μs to 6000 s Span ≥ 10 Hz 1 ms to 4000 s Accuracy Span ≥ 10 Hz, swept ± 0.01 % nominal Span ≥ 10 Hz, FFT ± 40 % nominal Span = 0 Hz ± 1 % nominal Trigger Free run, line, video, external 1, RF burst, periodic timer Trigger delay Span = 0 Hz or FFT –150 to +500 ms Span ≥ 10 Hz, swept 1 μs to 500 ms Resolution 0.1 μs Time gating Gate methods Gated LO; gated video; gated FFT Gate length range (except method = FFT) 100.0 ns to 5.0 s Gate delay range 0 to 100.0 s Gate delay jitter 33.3 ns p-p nominal Sweep (trace) point range All spans 1 to 40001 Resolution bandwidth (RBW) Range (–3.01 dB bandwidth) 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz Bandwidth accuracy (power) 1 Hz to 750 kHz ± 1.0 % (± 0.044 dB) nominal 820 kHz to 1.2 MHz (< 3 GHz CF) ± 2.0 % (± 0.088 dB) nominal 1.3 to 2.0 MHz (< 3 GHz CF) ± 0.07 dB nominal 2.2 to 3 MHz (< 3 GHz CF) ± 0.15 dB nominal 4 to 8 MHz (< 3 GHz CF) ± 0.25 dB nominal Bandwidth accuracy (–3.01 dB) RBW range 1 Hz to 1.3 MHz ± 2 % nominal Selectivity (–60 dB/–3 dB) 4.1:1 nominal EMI bandwidth (CISPR compliant) 200 Hz, 9 kHz, 120 kHz, 1 MHz (Option EMC required) EMI bandwidth (MIL STD 461E compliant) 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, 1 MHz (Option EMC required) Analysis bandwidth 1 Maximum bandwidth Option B25 25 MHz Standard 10 MHz Video bandwidth (VBW) Range 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz, and wide open (labeled 50 MHz) Accuracy ± 6 % nominal Measurement speed 2 Local measurement and display update rate 11 ms (90/s) nominal Remote measurement and LAN transfer rate 6 ms (167/s) nominal Marker peak search 5 ms nominal Center frequency tune and transfer 22 ms nominal Measurement/mode switching 75 ms nominal 1. Analysis bandwidth is the instantaneous bandwidth available around a center frequency over which the input signal can be digitized for further analysis or processing in the time, frequency, or modulation domain. 2. Sweep points = 101. Frequency and Time Specifications (continued)
06 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Amplitude range Measurement range RF (Option 503, 507) Preamp off 100 kHz to 1 MHz Displayed average noise level (DANL) to +20 dBm 1 MHz to 7.5 GHz Displayed average noise level (DANL) to +23 dBm Preamp on 100 kHz to 7.5 GHz Displayed average noise level (DANL) to +15 dBm MW (Option 513/526) Preamp off 100 kHz to 26.5 GHz Displayed average noise level (DANL) to +23 dBm Preamp on 100 kHz to 26.5 GHz Displayed average noise level (DANL) to +23 dBm Input attenuator range RF (Option 503, 507) Standard 0 to 50 dB in 10 dB steps Option FSA 0 to 50 dB in 2 dB steps MW (Option 513, 526) Standard 0 to 70 dB in 10 dB steps Option FSA 0 to 70 dB in 2 dB steps Maximum safe input level Average total power RF (Option 503, 507) +30 dBm (1 W) Input attenuation ≥ 20 dB, preamp off 10 dBm (10 mW) Input attenuation ≥ 20 dB, preamp on MW (Option 513, 526) +30 dBm (1 W) Input attenuation ≥ 10 dB, preamp off +30 dBm (1 W) Input attenuation ≥ 20 dB, preamp on Peak pulse power +50 dBm (100 W) < 10 μs pulse width, < 1 % duty cycle, input attenuation ≥ 30 dB DC volts RF (Option 503, 507) AC coupled ± 50 Vdc MW (Option 513, 526) AC coupled ± 50 Vdc DC coupled ± 0.2 Vdc Display range Log scale 0.1 to 1 dB/division in 0.1 dB steps 1 to 20 dB/division in 1 dB steps (10 display divisions) Linear scale 10 divisions Scale units dBm, dBmV, dBμV, dBmA, dBμA, V, W, A Frequency response Specification 95th percentile (≈ 2 s) (10 dB input attenuation, 20 to 30 °C, s = nominal standard deviation) RF (Option 503, 507) 9 kHz to 10 MHz ± 0.60 dB ± 0.45 dB 10 MHz to 3 GHz ± 0.75 dB ± 0.55 dB 3 to 5.25 GHz ± 1.45 dB ± 1.00 dB 5.25 to 7.5 GHz ± 1.65 dB ± 1.20 dB MW (Option 513, 526) 9 kHz to 10 MHz ± 0.8 dB ± 0.5 dB 10 MHz to 3 GHz ± 0.65 dB ± 0.4 dB 3 to 7.5 GHz ± 1.5 dB ± 0.5 dB 7.5 to 13.6 GHz ± 2.0 dB ± 0.8 dB 13.6 to 19 GHz ± 2.0 dB ± 1.0 dB 19 to 26.5 GHz ± 2.5 dB ± 1.3 dB Preamp on RF (Option 503, 507) (P03, P07) 100 kHz to 3 GHz ± 0.70 dB 3 to 5.25 GHz ± 0.85 dB 5.25 to 7.5 GHz ± 1.35 dB MW (Option 513, 526) (P03, P07, P13, P26) 100 kHz to 3 GHz ± 0.7 dB 3 to 13.6 GHz ± 1.0 dB 13.6 to 19 GHz ± 1.1 dB 19 to 26.5 GHz ± 2.5 dB Amplitude Accuracy and Range Specifications
07 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Input attenuation switching uncertainty Specifications Additional information Attenuation > 2 dB, preamp off 50 MHz (reference frequency) ± 0.32 dB ± 0.15 dB typical Relative to 10 dB (reference setting) 100 kHz to 3.0 GHz ± 0.30 dB nominal 3.0 to 7.5 GHz ± 0.50 dB nominal 7.5 to 26.5 GHz ± 0.70 dB nominal Total absolute amplitude accuracy (10 dB attenuation, 20 to 30 °C, 1 Hz ≤ RBW ≤ 1 MHz, input signal –10 to –50 dBm, all settings auto-coupled except Auto Swp Time = Accy, any reference level, any scale, s = nominal standard deviation) At 50 MHz ± 0.40 dB At all frequencies ± (0.40 dB + frequency response) 100 kHz to 10 MHz ± 0.60 dB (95th Percentile ≈ 2s) 10 MHz to 2.0 GHz ± 0.50 dB (95th Percentile ≈ 2s) 2.0 to 3.0 GHz ± 0.60 dB (95th Percentile ≈ 2s) Preamp on (Option P03/P07/P13/P26) ± (0.39 dB + frequency response) nominal Input voltage standing wave ratio (VSWR) (≥ 10 dB attenuation) Option 503, 507 Option 513, 526 10 MHz to 3 GHz < 1.5 nominal < 1.3 nominal 3 to 7.5 GHz < 2.0 nominal < 1.4 nominal 7.5 to 26.5 GHz N/A < 1.9 nominal Resolution bandwidth switching uncertainty (referenced to 30 kHz RBW) 1 Hz to 3 MHz RBW ± 0.15 dB 4, 5, 6, 8 MHz RBW ± 1.0 dB Reference level Range Log scale –170 to +23 dBm in 0.01 dB steps Linear scale Same as log (707 pV to 3.16 V) Accuracy 0 dB Display scale switching uncertainty Switching between linear and log 0 dB Log scale/div switching 0 dB Display scale fidelity –80 dBm ≤ input mixer level < –15 dBm ± 0.15 dB total –15 dBm ≤ input mixer level < –10 dBm ± 0.30 dB ± 0.15 dB typical Trace detectors Normal, peak, sample, negative peak, log power average, RMS average, and voltage average Preamplifier (Option P03/P07/P13/P26) Frequency range Option P03 100 kHz to 3.0 GHz Option P07 100 kHz to 7.5 GHz Option P13 100 kHz to 13.6 GHz Option P26 100 kHz to 26.5 GHz Gain 100 kHz to 26.5 GHz +20 dB nominal Noise figure 100 kHz to 26.5 GHz DANL + 176.24 dB nominal Amplitude Accuracy and Range Specifications (continued)
08 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet 1 dB gain compression (two-tone) Total power at input mixer RF (Option 503, 507) Preamp off 50 MHz to 7.5 GHz +2 dBm nominal Preamp on 50 MHz to 7.5 GHz –19 dBm nominal MW (Option 513/526) Preamp off 50 MHz to 7.5 GHz +7 dBm noiminal 7.5 to 13.6 GHz +3 dBm noiminal 13.6 to 26.5 GHz +0 dBm noiminal Preamp on 50 MHz to 26.5 GHz –19 dBm nominal Displayed average noise level (DANL) (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, 20 to 30 °C) Parentheses indicate typical performance Preamplifier OFF Preamplifier ON RF (Option 503/507) 9 kHz to 1 MHz (–120) dBm (–139) dBm, 100 kHz to 1 MHz 1 to 10 MHz –130 (–137) dBm –149 (–157) dBm 10 MHz to 1.5 GHz –148 (–150) dBm –161 (–163) dBm 1.5 to 2.2 GHz –144 (–147) dBm –160 (–163)dBm 2.2 to 2.5 GHz –144 (–147) dBm –158 (–161) dBm 2.5 to 2.7GHz –142 (–145) dBm –158 (–161) dBm 2.7 to 3.0 GHz –139 (–143) dBm –158 (–161) dBm 3 to 4.5 GHz –137 (–140) dBm –155 (–159) dBm 4.5 to 6 GHz –133 (–136) dBm –152 (–156) dBm 6 to 7.5 GHz –128 (–131) dBm –148 (–152) dBm MW (Option 513/526) 1 to 10 MHz –143 (–148) dBm –153 (–158) dBm 10 MHz to 1.5 GHz –147 (–150) dBm –160 (–163) dBm 1.5 to 6 GHz –143 (–147) dBm –158 (–161) dBm 6 to 7.5 GHz –141 (–145) dBm –155 (–160) dBm 7.5 to 13.6 GHz –139 (–142) dBm –155 (–160) dBm 13.6 to 20 GHz –134 (–140) dBm –153 (–157) dBm 20 to 24 GHz –132 (–138) dBm –151 (–155) dBm 24 to 26.5 GHz –124 (–129) dBm –142 (–147) dBm Spurious responses RF (Option 503, 507) Residual responses (Input terminated and 0 dB attenuation, 20 to 30 °C) 200 kHz to 7.5 GHz (swept) –90 dBm Zero span or FFT or other frequencies –100 dBm nominal Input related spurious 10 MHz to 7.5 GHz –60 dBc typical MW (Option 513, 526) Tuned frequency (f) Mixer level Response Image responses 10 MHz to 26.5 GHz −10 dBm −60 dBc typical LO-related spurious 10 MHz to 3 GHz −10 dBm −64 dBc typical Other spurious responses First RF order (f ≥ 10 MHz from carrier) −10 dBm −65 dBc High RF order (f ≥ 10 MHz from carrier) −30 dBm −65 dBc Second harmonic distortion (SHI) Source frequency SHI (nominal) RF/MW (Option 503, 507, 513, 526) 10 MHz to 3.75 GHz +42 dBm MW (Option 513, 526) 3.75 to 13.25 GHz +54 dBm Dynamic Range Specifications
400 MHz to 3 GHz +13 (+17) dBm
500 MHz to 2 GHz +12 dBm, (+15) dBm
1 MHz
10 MHz
Figure 1. Nominal phase noise at different center frequencies for CXA
10 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Channel power Amplitude accuracy, W-CDMA or IS95 (20 to 30 °C, attenuation = 10 dB) ± 1.33 dB (± 0.61 dB 95th percentile) Occupied bandwidth Frequency accuracy ± [span/1000] nominal Adjacent channel power Accuracy, W-CDMA (ACLR) (at specific mixer levels and ACLR ranges) Adjacent Alternate MS BTS ± 0.76 dB ± 1.72 dB ± 0.73 dB ± 1.96 dB Dynamic range (typical) RF (Option 503, 507) Without noise correction –63 dB –67 dB With noise correction –73 dB –78 dB MW (Option 513, 526) Without noise correction –66 dB –69 dB With noise correction –73 dB –78 dB Offset channel pairs measured 1 to 6 Power statistics CCDF Histogram resolution 0.01 dB Harmonic distortion Maximum harmonic number 10th Results Fundamental power (dBm), relative harmonics power (dBc), total harmonic distortion in % Intermod (TOI) Measure the third-order products and intercepts from two tones Burst power Methods Power above threshold, power within burst width Results Single burst output power, average output power, maximum power, minimum power within burst, burst width Spurious emission W-CDMA (1 to 2.7 GHz) table-driven spurious signals; search across regions Dynamic range (RBW=1 MHz) 70.7 dB (75.9 dB typical) Absolute sensitivity (RBW=1 MHz) −76.5 dBm (−82.5 dBm typical) Spectrum emission mask (SEM) cdma2000 ® (750 kHz offset) Relative dynamic range (30 kHz RBW) 67.4 dB (72.7 dB typical) Absolute sensitivity −93.7 dBm (−99.7 dBm typical) Relative accuracy ± 0.11 dB 3GPP W-CDMA (2.515 MHz offset) Relative dynamic range (30 kHz RBW) 73.4 dB (80.2 dB typical) Absolute sensitivity −91.7 dBm (−97.7 dBm typical) Relative accuracy ± 0.11 dB PowerSuite Measurement Specifications
11 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Output frequency Frequency range Option T03 1 Option T06 1 9 kHz to 3 GHz 9 kHz to 6 GHz Resolution 1 Hz Output power level Range −50 to 0 dBm Resolution 0.1 dB Absolute accuracy (at 50 MHz, −10 dBm, 20 to 30 °C) ± 0.55 dB Output flatness (referenced to 50 MHz, −10 dBm, 20 to 30 °C) Specification 95th percentile (≈ 2 s) 9 kHz to 100 kHz ± 1.5 dB ± 1.2 dB 100 kHz to 3.0 GHz ± 1.2 dB ± 0.8 dB 3.0 GHz to 6.0 GHz ± 1.5 dB ± 1.2 dB Level accuracy 9 kHz to 100 kHz ± 1.0 dB nominal 100 kHz to 3.0 GHz ± 0.5 dB nominal 3.0 GHz to 6.0 GHz ± 0.8 dB nominal Output power sweep Range −50 to 0 dBm Resolution 0.1 dB Maximum safe reverse level Average total power +30 dBm (1 W) AC coupled ± 50 Vdc Phase noise Noise sidebands (CF = 1 GHz) Offset 10 kHz 100 kHz −102 dBc/Hz nominal −104 dBc/Hz nominal −117 dBc/Hz nominal Spurious outputs (0 dBm output) Harmonic spurs 100 kHz to 3 GHz < −35 dBc
3 GHz to 6 GHz < −30 dBc
9 kHz to 10 MHz < −35 dBc nominal
10 MHz to 6 GHz < −35 dBc
Maximum output power – displayed average noise level 110 dBc nominal Output VSWR 9 kHz to 6 GHz < 1.5:1 nominal 1. Not available on microwave CXA (Option 513 or 526). Tracking Generator Specifications
12 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency range Option C75 1 1 MHz to 1.5 GHz Maximum safe input level Average continuous power or peak pulse power +72.5 dBmV (0.25 W) Input attenuation ≥ 20 dB, preamp off +63 dBmV (25 mW) Input attenuation ≥ 20 dB, preamp on (Option P03/P07) AC coupled ± 50 Vdc Frequency response (10 dB input attenuation) Preamp off 1 MHz to 10 MHz 10 MHz to 1.5 GHz ± 0.6 dB nominal ± 0.75 dB nominal 1 dB gain compression (two-tone) Total power at input mixer Preamp off 50 MHz to 1.5 GHz +57 dBmV nominal Preamp on (Option P03/P07) 50 MHz to 1.5 GHz +35 dBmV nominal Displayed average noise level (DANL) (Input terminated, sample or average detector, averaging type = Log, 0 dB input attenuation, IF Gain = High, nominal) Preamp off 1 to 10 MHz −89 dBmV 10 MHz to 1.5 GHz −97 dBmV Preamp on (Option P03/P07) 1 to 10 MHz −108 dBmV 10 MHz to 1.5 GHz −113 dBmV Second harmonic distortion (SHI) Preamp off (Input level +28.75 dBmV, input attenuation 10 dB) 10 to 750 MHz +95 dBmV nominal Preamp on (Option P03/P07) (Input level +8.75 dBmV, input attenuation 10 dB) 10 to 750 MHz +63 dBmV nominal Third-order intermodulation distortion (TOI) Preamp off (Two +28.75 dBmV tones at input mixer spaced by 100 kHz, 0 dB attenuation) 10 MHz to 1.5 GHz +62 dBmV nominal Preamp on (Option P03/P07) (Two +3.75 dBmV tones at input mixer spaced by 100 kHz, 0 dB attenuation) 10 MHz to 1.5 GHz +40 dBmV nominal Input voltage standing wave ratio (VSWR) Preamp off (10 dB attenuation) 1 MHz to 1.5 GHz < 1.4:1 nominal Preamp on (Option P03/P07) (0 dB attenuation) 1 MHz to 1.5 GHz < 1.4:1 nominal 1. Not available on microwave CXA (Option 513 or 526). 75 Ω Input Specifications
13 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Temperature range Operating 0 to 55 °C Storage –40 to 70 °C EMC Complies with European EMC Directive 2004/108/EC – IEC/EN 61326-1 or IEC/EN 61326-2-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11:2002 – ICES/NMB-001 This ISM device complies with Canadian ICES-001 Cet appareil ISM est conforme à la norme NMB-001 du Canada Safety Complies with European Low Voltage Directive 73/23/EEC, amended by 93/68/EEC – IEC/EN 61010-1 2nd Edition – Canada: CSA C22.2 No. 61010-1 – USA: UL 61010-1 2nd Edition Audio noise Acoustic noise emission Geraeuschemission LpA < 70 dB LpA < 70 dB Operator position Am Arbeitsplatz Normal position Normaler Betrieb Per ISO 7779 Nach DIN 45635 t.19 Environmental stress Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude, and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3. Power requirements Voltage and frequency (nominal) 100 to 120 V, 50/60/400 Hz 220 to 240 V, 50/60 Hz Power consumption On 270 W maximum Standby 20 W Display Resolution 1280 x 768, WXGA Size 269 mm (10.6 in.) diagonal (nominal) Data storage Internal 160 GB nominal (removable solid state drive) External Supports USB 2.0 compatible memory devices Weight (without options) Net 15.4 kg (34.0 lbs) Shipping 27.4 kg (60.4 lbs) Dimensions Height 177 mm (7.0 in) Width 426 mm (16.8 in) Length 368 mm (14.5 in) Warranty The CXA signal analyzer is supplied with a 3-year warranty Calibration cycle The recommended calibration cycle is one year; calibration services are available through Keysight service centers General Specifications
14 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Front panel RF input Connector Type-N female, 50 Ω nominal RF input (Option C75) Connector Type-N female, 75 Ω nominal RF output (Option T03 or T06) Connector Type-N female, 50 Ω nominal Probe power Voltage/current +15 Vdc, ± 7 % at 150 mA max. nominal –12.6 Vdc, ± 10 % at 150 mA max. nominal USB ports Host (3 ports) Standard Connector Output current Port marked with lightning bolt Port not marked with lightning bolt Compatible with USB 2.0 USB Type-A female
1.2 A (nominal)
0.5 A Rear panel
10 MHz out
BNC female, 50 Ω nominal ≥ 0 dBm nominal
10 MHz ± (10 MHz x frequency reference accuracy)
BNC female, 50 Ω nominal –5 to 10 dBm nominal
10 MHz ± nominal
± 5 x 10–6 of specified external reference input frequency Trigger 1 input Connector Impedance Trigger level range BNC female > 10 kΩ nominal –5 to 5 V Trigger 1 output Connector Impedance Level BNC female 50 Ω nominal
5 V TTL nominal
VGA compatible, 15-pin mini D-SUB XGA (60 Hz vertical sync rates, non-interlaced) Analog RGB 1024 x 768 Noise source drive +28 V (pulsed) Connector BNC female SNS Series noise source Analog out Connector BNC female Inputs and Outputs
15 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet USB ports Host, super speed Standard Connector Output current Host Standard Connector Output current Device Standard Connector 2 ports (stacked with each other) Compatible with USB 3.0 USB Type-A female 0.9 A 1 port (stacked with LAN) USB 2.0 USB Type-A female 0.5 A Compatible with USB 3.0 USB Type-B female GPIB interface Connector GPIB codes GPIB mode IEEE-488 bus connector SH1, AH1, T6, SR1, RL1, PP0, DC1, C1, C2, C3, C28, DT1, L4, C0 Controller or device Rear panel (continued) LAN TCP/IP interface Standard Connector 1000Base-T RJ45 Ethertwist Sync (reserved for future use) Connector BNC female IF output Connector Impedance SMA female 50 Ω nominal Wideband IF output, Option CR3 1 Center frequency SA mode or I/Q analyzer 322.5 MHz Conversion gain –4 to +7 dB (nominal) plus RF frequency response Bandwidth Low band High band Up to 120 MHz (nominal) Up to 40 MHz (nominal) 1. Not available on microwave CXA (Option 513 or 526). Inputs and Outputs (continued)
16 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Frequency Frequency span Standard instrument 10 Hz to 10 MHz Option B25 10 Hz to 25 MHz Resolution bandwidth (spectrum measurement) Range Overall 100 mHz to 3 MHz Span = 1 MHz 50 Hz to 1 MHz Span = 10 kHz 1 Hz to 10 kHz Span = 100 Hz 100 mHz to 100 Hz Window shapes Flat top, Uniform, Hanning, Gaussian, Blackman, Blackman-Harris, Kaiser Bessel (K-B 70 dB, K-B 90 dB and K-B 110 dB) Analysis bandwidth Standard instrument 10 Hz to 10 MHz Option B25 10 Hz to 25 MHz IF frequency response (standard 10 MHz IF path) IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C) Center frequency (GHz) Span (MHz) Max. error RMS (nominal) ≤ 3.0 ≤ 10 ± 0.40 dB 0.03 dB 3.0 < f ≤ 26.5 ≤ 10 ± 0.40 dB 0.25 dB IF phase linearity (deviation from mean phase linearity, nominal) Center frequency (GHz) Span (MHz) Peak-to-peak RMS Data acquisition (standard 10 MHz IF path) Time record length 4,000,000 IQ sample pairs Sample rate 30 MSa/s ADC resolution 14 Bits Option B25 25 MHz analysis bandwidth IF frequency response (demodulation and FFT response relative to the center frequency, 20 to 30 °C) Center frequency (GHz) Span (MHz) Max. error RMS (nominal) ≤ 3.0 10 to ≤ 25 ± 0.45 dB 0.03 dB 3.0 < f ≤ 26.5 10 to ≤ 25 ± 0.45 dB 0.65 dB IF phase linearity (deviation from mean phase linearity, nominal) Center frequency (GHz) Span (MHz) Peak-to-peak RMS 0.02 ≤ f ≤ 3.0 10 to ≤ 25 2.7 ° 0.9 ° 3.0 < f ≤ 7.5 10 to ≤ 25 4.7 ° 2.2 ° 7.5 < f ≤ 26.5 10 to ≤ 25 3.5 ° 1.0 ° Data acquisition (B25 IF path) Time record length IQ analyzer 4,000,000 IQ sample pairs Sample rate 90 MSa/s ADC resolution 14 Bits I/Q Analyzer
17 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet Literature Pub number CXA Signal Analyzer N9000B - Configuration Guide 5992-1275EN X-Series Signal Analyzers - Brochure 5992-1316EN For more information or literature resources please visit the web: www.keysight.com/find/cxa Web Product page: www.keysight.com/find/N9000 B X-Series measurement applications: www.keysight.com/find/X-Series_Apps X-Series signal analyzers: www.keysight.com/find/X-Series Related Literature
18 | Keysight | N9000B CXA X-Series Signal Analyzer, Multi-touch - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2016 Published in USA, January 26, 2016 5992-1274EN www.keysight.com From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939.
1939 THE FUTURE
For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-01-01-16) www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with a three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. cdma2000 is a US registered certification mark of the Telecommunications Industry Association.