THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. U.S. Government Rights U.S. Government Rights. The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EU LA), a copy of which can be found at http://www.keysight.com/find/sweula. The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that i s not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Table 1. Key Specifications 1 Fast CW mode - no point triggering. 2 For single parameter; two parameters are 250 million points each.
Table 2. Measurement Throughput Summary
Start 2 GHz, Stop 18 GHz, 600 kHz IF bandwidth (with band crossings) RF = MXG, N5183A opt. UNZ, Fast switching LO = MXG, N5183A opt. UNZ, Fast switching 0.580 0.580 -92.5 dBm, 2 – 3 GHz - 96.5 dBm, 3 – 12.5 GHz - 85 dBm, 12.5 – 18 GHz RF = MXG, N5183A opt. UNZ, Fast switching LO = N5264A opt. 1083 0.580 0.580 -85.5 dBm, 2 – 3 GHz - 92.5 dBm, 3 – 12.5 GHz - 83 dBm, 12.5 – 18 GHz RF = UXG, N5193A opt. SS1, 1 µs switching speed LO = UXG, N5193A opt. SS1, 1 µs switching speed 0.045 0.039 -92.5 dBm, 2 – 3 GHz - 96.5 dBm, 3 – 12.5 GHz - 85 dBm, 12.5 – 18 GHz Start 2 GHz, Stop 18 GHz, 10 kHz IF bandwidth (with band crossings) RF = MXG, N5183A opt. UNZ, Fast switching LO = MXG, N5183A opt. UNZ, Fast switching 0.730 0.730 -110.5 dBm, 2 – 3 GHz - 114.5 dBm, 3 –12.5 GHz - 103 dBm, 12.5 – 18 GHz RF = MXG, N5183A opt. UNZ, Fast switching LO = N5264A opt. 1083 0.730 0.730 -103.5 dBm, 2 – 3 GHz - 110.5 dBm, 3 –12.5 GHz - 101 dBm, 12.5 –18 GHz RF = MXG, N5183A opt. UNZ, Fast switching LO = PSG E8267D opt. 520, UNX 9.50 9.50 -110.25 dBm, 2 – 3 GHz - 112.50 dBm, 3 –12.5 GHz - 96.50 dBm, 12.5 – 18 GHz RF = MXG, N5183A opt. UNZ, Fast switching LO = 83623B 7.80 -- -108.5 dBm, 2 – 3 GHz - 113.0 dBm, 3 –12.5 GHz - 96.0 dBm, 12.5 –18 GHz RF = UXG, N5193A opt. SS1, 1 µs switching speed LO = UXG, N5193A opt. SS1, 1 µs switching speed 0.170 0.167 -110.5 dBm, 2 – 3 GHz - 114.5 dBm, 3 –12.5 GHz - 103 dBm, 12.5 – 18 GHz
Start 2 GHz, Stop 18 GHz, 1 kHz IF bandwidth (with band crossings) RF = MXG, N5183A opt. UNZ, Fast switching LO = MXG, N5183A opt. UNZ, Fast switching 1.5 1.5 -120.5 dBm, 2 – 3 GHz - 124.5 dBm, 3 –12.5 GHz - 113 dBm, 12.5 – 18 GHz RF = MXG, N5183A opt. UNZ, Fast switching LO = N5264A opt. 1083 1.5 1.5 -113.5 dBm, 2 – 3 GHz - 120.5 dBm, 3 –12.5 GHz - 111 dBm, 12.5 – 18 GHz RF = UXG, N5193A opt. SS1, 1 µs switching speed LO = UXG, N5193A opt. SS1, 1 µs switching speed 0.970 0.970 -120.5 dBm, 2 – 3 GHz - 124.5 dBm, 3 –12.5 GHz - 113 dBm, 12.5 – 18 GHz Start 2 GHz, Stop 18 GHz, 500 Hz IF bandwidth (with band crossings) RF = UXG, N5193A opt. SS1, 1 µs switching speed LO = UXG, N5193A opt. SS1, 1 µs switching speed 1.85 1.85 -120.5 dBm, 2 – 3 GHz - 124.5 dBm, 3 –12.5 GHz - 113 dBm, 12.5 – 18 GHz Option 118 Fast-CW mode (CW frequency) Number of Points per Second (#pt/Sec) External Trigger C.W, 7.0 GHz, ≥1 MHz IF bandwidth -- 400,000 C.W, 7.0 GHz, 600 KHz IF bandwidth Up to 400,000 240,000 C.W, 7.0 GHz, 10 KHz IF bandwidth Up to 8,200 7,000 C.W, 7.0 GHz, 1 KHz IF bandwidth Up to 1,000 1,000 Time/Point (ms) Description Typical Performance Start 2 GHz, Stop 18 GHz, 801 points (with band crossings), hardware trigger IF Bandwidth 1 MHz 600 kHz 100 kHz 10 kHz 1 kHz 500 Hz RF = UXG, N5193A opt. SS1, 1 µs switching speed LO = UXG, N5193A opt. SS1, 1 µs switching speed
Data Transfer Time (ms) Description Typical Performance Number of Points 201 401 1601 16,001 SCPI over GPIB Program executed on external PC 5 32-bit floating point 5.6 10.5 39.9 400 64-bit floating point 10.5 20.3 79.2 788 ASCII 46 92.5 370 3702 SCPI over SICL/LAN or TCP/IP Socket Program executed in the analyzer 32-bit floating point 0.18 0.21 0.5 3.6 64-bit floating point 0.22 0.28 0.62 5.3 ASCII 6.3 12.3 47.3 470 COM6 Program executed in the analyzer 32-bit floating point <0.15 0.15 0.2 0.7 Variant type 0.75 1.2 4.5 50 DCOM over LAN6 Program executed on external PC 32-bit floating point <1.0 1.2 2.1 13 Variant type 2.7 4.5 15 150 1 Includes sweep time, retrace time and band-crossing time. Analyzer display turned on. Minus 21 ms from total time for display off with DISPLAY:ENABLE OFF. Data for two traces (A & B receiver) per measurement. 2 After first complete sweep. 3 When configuring the N5264A Option 108 as the LO source, you may improve system measurement sensitivity by using a method of AM noise suppression.
4 Performance Characteristics when connected with 85309A and 85320A/B mixers - system noise floor +
conversion gain. 5 Measured when using the SCPI command DISPlay: VISible OFF. 6 Values are for real and imaginary pairs, with the analyzer display off.
Table 3. Rear Panel Information
10 Hz IF BW -143 dBm
The IF Input frequencies are different depending on the DSP Version. With DSP Version 4: RF or Transmitting frequency < 53 MHz: IF = 2.535211 MHz [3 x (60e6 / 71)] RF or Transmitting frequency >= 53 MHz: IF = 7.605634 MHz [9 x (60e6 / 71)] With DSP Version 5, the IF frequency is dependent on the RF or Transmitting frequency AND the current IFBW setting: All RF or Transmitting frequency; IF Bandwidth >= 1MHz IFBW Setting IF Frequency 1 MHz 7.692 MHz 1.5 MHz 7.368 MHz 2 MHz 8.450 MHz 3 MHz 8.163 MHz 5 MHz 6.897 MHz 7 MHz 10.53 MHz 10 MHz 15.38 MHz 15 MHz 22.22 MHZ IF Bandwidth <= 600 kHz: RF or Transmitting frequency < 53 MHz; IF = 2.479339 MHz [(3 x (100e6 / 121)] RF or Transmitting frequency >= 53 MHz; IF = 7.438017 MHz [(9 x (100e6 / 121)] Manually change the IF frequency The IF frequency can be changed to any value between +14.9999 MHz and -14.9999 MHz using SENS:IF:FREQ (SCPI) or IFFrequency (COM) commands. With DSP Version 4 - 34 and above, min and max IF frequencies up to +/- 20.1 MHz are available. With DSP Version 5, min and max IF frequencies up to +/- 38 MHz are available. Performance is degraded drastically above +/- 14.9999 MHz.
External IF Inputs (Cont.) Description Typical Performance Dynamic Accuracy (Cont.) -40 dBm reference, over range set by compression and noise floor @ IF Frequencies -80 dBm 0.041 dB -90 dBm 0.049 dB -100 dBm 0.057 dB -110 dBm 0.072 dB -120 dBm 0.188 dB LO output 2 (Option 108) Description Specification Frequency Stability +/- 0.05 ppm, -10 to 70C, +/- 0.1ppm/yr max Frequency Accuracy +/- 1 ppm Description Typical Performance Frequency Range 10 MHz to 26.5 GHz Frequency Switching Speed3 < 100 microsecond/point Frequency Resolution 1 Hz Power Flatness +/- 1.0 dB Power Output +10 dBm 2nd Harmonics4 20 MHz to 2.0 GHz -23 dBc 2.0 GHz to 5.0 GHz -28 dBc 5.0 GHz to 23.0 GHz -35 dBc 23.0 GHz to 26.5 GHz -27 dBc
LO output 2 (Option 108) Description Typical Performance 3rd Harmonics3 30 MHz to 8.0 GHz -32 dBc 8.0 GHz to 15.0 GHz -38 dBc 15.0 GHz to 26.5.0 GHz -48 dBc Phase Noise
1 KHz
10 KHz Offset 100 KHz
1 MHz
10 MHz to 500 MHz -80 dBc/Hz -85 dBc/Hz -76 dBc/Hz -113 dBc/Hz
500 MHz to 1 GHz -90 dBc/Hz -110 dBc/Hz -106 dBc/Hz -115 dBc/Hz
1 GHz to 2 GHz -85 dBc/Hz -105 dBc/Hz -101 dBc/Hz -110 dBc/Hz
2 GHz to 4 GHz -80 dBc/Hz -100 dBc/Hz -96 dBc/Hz -105 dBc/Hz
4 GHz to 8 GHz -74 dBc/Hz -94 dBc/Hz -90 dBc/Hz -99 dBc/Hz
8 GHz to 16 GHz -68 dBc/Hz -88 dBc/Hz -84 dBc/Hz -93 dBc/Hz
16 GHz to 26.5 GHz -62 dBc/Hz -82 dBc/Hz -78 dBc/Hz -87 dBc/Hz
10 MHz Reference
10 MHz Reference In
Connector BNC, female Input Frequency 10 MHz ± 10 ppm, typical Input Level -15 dBm to +20 dBm, typical Input Impedance 200 Ω, nom.
10 MHz Reference Out
Connector BNC, female Output Frequency 10 MHz ± 1 ppm, typical Signal Type Sine Wave, typical Output Level +10 dBm ± 4 dB into 50 Ω Output Impedance 50 Ω, nominal Harmonics <-40 dBc, typical
External Monitor Information Description Typical Performance VGA Video Output Connector 15-pin mini D-Sub; Drives VGA compatible monitors Devices Supported: Resolutions: Flat Panel (TFT) 1024 X 768, 800 X 600, 640 X 480 Flat Panel (DSTN) 800 X 600, 640 X 480 CRT Monitor 1280 X 1024, 1024 X 768, 800 X 600, 640 X 480 -- Simultaneous operation of the internal and external displays is allowed, but with 640 X 480 resolution only. If you change resolution, you can only view the external display (internal display will "white out"). Test Set IO 25-pin D-Sub connector, available for external test set control. Power IO 9-pin D-Sub, female; analog and digital IO Handler IO 36-pin parallel I/O port; all input/output signals are default set to negative logic; can be reset to positive logic via GPIB command. Trigger Information Description Typical Performance Trigger In/Meas Trigger Nominal Input Impedance 5K Ohms Minimum Pulse Width 1 us DC Damage Level 5.5 volts Drive Voltage TTL (0, +5.0) Volts
Trigger Information (Cont.) Description Typical Performance Trigger out/Meas Trigger Ready Nominal Input Impedance 5K Ohm Pulse Width = Data acquisition Polarity Selectable with sweep or point mode Drive Voltage TTL (0, +5.0) Volts Trigger Inputs/Outputs (Aux. 1 & 2) BNC(f), TTL/CMOS compatible GPIB (two ports - dedicated controller and dedicated talker/listener) 24-pin D-sub (Type D-24), female; compatible with IEEE-488. Parallel Port (LPT1) 25-pin D-Sub miniature connector, female; provides connection to printers or any other parallel port peripherals Serial Port (COM 1) 9-pin D-Sub, male; compatible with RS-232 USB Port Four ports on front panel (all Host) and five ports (four hosts and one Device) on rear panel. Type A configuration (eight hosts) and Type B configuration (one Device), USB 2.0 compatible. LAN 10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data rates Line Power Description Typical Performance Power supply is auto switching Frequency, Voltage 50/60 Hz for 100 240 VAC Max 450 watts 1 Measurement conditions: normalized to -10 dBm, 10 Hz IFBW, averaging factor of 8. 2 Absolute LO frequency is Front Panel set frequency plus 1 IF. 3 No band crossings; IFBW ≥ 100 kHz with 801 measurement points. 4 Listed frequency is the harmonic frequency setting entered with front panel (frequency setting entered with front panel plus {IF frequency} * {harmonic number}) at typical power.
Table 4. Front Panel Information black appearance that will not change. black appearance that will not change.
Table 5. Analyzer Dimensions and Weight and feet. Including rack-mount flanges. connectors, rear panel bumpers, and feet. 1 Feet removed from the N5264A.
This information is subject to change without notice. © Keysight Technologies 2014 - 2016 Print Date: August 4, 2016 www.keysight.com