MEMORY HiCORDER MR8827
Document overview
- Manufacturer or author: HIOKI
- PDF pages: 16
Technical content
The Memory HiCorder MR8827 achieves isolated input between the main unit and channel or between channels, at a maximum sampling speed of 20 MS/s on all channels. It provides mixed recording that combines 32 analog channels and 32 logic channels, and logic input can be expanded up to 64 channels. Welcome to the next generation of Hioki Memory HiCorders that deliver multi- channel waveform recording of a diverse array of signals to meet complex and demanding applications. *When using 64 logic channels, 28 analog channels are available. MEMORY HiCORDER Recorders MR8827 32 analog channels + 32 logic channels64ch High-speed Isolated testing
MR8827 - Evolving to the Next Stage of High-Speed Waveform Recording
SSD (Solid State Drive) 8x Internal Memory Capacity With 8 times more internal memory capacity from 64 MW to 512 MW, you can now record signals of fast events easily and for extended pe- riods of time. 64MW 512MW 2x Logic Input Channels A maximum of 8 logic probes can be inserted in the main unit. Use of
2 Logic Unit 8973 will add 8 more
connections, supporting 64 channel logic signal input. (This reduces the number of available analog channels to 28.) Use various storage devices and media with more capacity and faster writing speeds than conventional drives or PC cards. The optional internal SSD has 128 GB of capacity so you can store large amounts of data. The high-performance 8826 delivered the most analog channels out of all portable-type Memory HiCorders. The new MEMORY HiCORDER MR8827 inherits that concept and evolves even further. 32ch 64ch1MS/s 20MS/s The sampling speed (for all channels simultaneously) increased by 20 times, while maintaining isolated input. A/D converter integrated in the input amp A/D converter integrated A/D converter integrated 20x Sampling Speed 2x Paper Feeding Speed 3x PC Transfer Speed Transferring speed of stored data from internal memory or SSD to the PC has greatly increased. LCD Resolution Overlapping waveforms are easier to identify now with a new high resolution LCD. 640×480 800×600 10.4 inch TFT 10.4 inch SVGA Easy Setup of Recording Paper No more hassles of feeding recording paper between the rubber roller and the thermal head. Just drop it in to set it up. Use of a high-speed thermal printer gives you 2 times the printing speed. 50mm/sec25mm/sec Use of a high-speed thermal printer gives you 2 times the printing speed. 25mm/sec Logic Unit 8973 Data transfer time → Conventional transfer time 3 times faster transfer speed!
Various Signal Input Support / Optional Units Various amps Transducers Sensors / industrial meters Supply current Inverter Current Motor current, etc. Relay signals AC / DC signals Strain gauge converter Low voltage High voltage Contacts Current Strain Temperature Frequency Thermocouples K, J, E, T, N, R, S, B, W Voltage / non-voltage contacts ANALOG UNIT 8966 DIGITAL VOLTMETER UNIT MR8990 HIGH RESOLUTION UNIT 8968 DC/RMS UNIT 8972 LOGIC UNIT 8973 CURRENT UNIT 8971 STRAIN UNIT 8969 TEMP UNIT 8967 FREQ UNIT 8970 High Resolution LCD Scroll Scroll through the waveform to check all or just part of it. Expand / shrink Not only can you expand or shrink the time axis or verti- cal axis, you can also split the screen to check the expanded waveform of the shrunk wave- form. + − Scroll Expand/Shrink Expanded waveform Shrunk waveform Encoders, etc. Rotating pulse Conventional devices used a 640×480 dot TFT LCD, but the next-generation MR8827 uses an 800×600 dot SVGA high resolution LCD to make it even easier to identify overlapping measured wave- forms. Conventional devices used a 640×480 dot TFT LCD, but the next-generation MR8827 uses an 800×600 dot SVGA high resolution LCD to make it even easier to identify overlapping measured wave- forms. Supply voltage Inverter Primary / Secondary voltage Motor voltage, etc. Dynamic strain / Acceleration / Vibration Pressure / Weight, etc.
A maximum of 16 modules can be connected on the rear side. The main unit also has connectors for connecting 8 logic probes. The MR8827 differentiates itself from typical oscilloscopes by providing complete isolation for the input of each channel, and between each channel and the main frame, enabling you to handle electrical potential differences among multiple signals without any concern. Isolation element Isolated Input for Security Isolation element Sampling Speed and Recording Time Memory functions A4 Size Printer Print in fine detail, with 2 times the paper feeding speed. Get a printout of enlarged waveforms on A4 size paper so you can check them easily on-site. Scanning Scan data at the cursor and the waveform's cross point. Cutout Specify the segment to save as binary or CSV data. Scanning Cutout Recorder functions Time axis range/div Sampling- speed Maximum recording length 32 channels 160,000 div 5 µs 50 ns 0.8 s 10 µs 100 ns 1.6 s 20 µs 200 ns 3.2 s 50 µs 500 ns 8 s 100 µs 1 µs 16 s 200 µs 2 µs 32 s 500 µs 5 µs 1 min 20 s 1 ms 10 µs 2 min 40 s 2 ms 20 µs 5 min 20 s 5 ms 50 µs 13 min 20 s 10 ms 100 µs 26 min 40 s 20 ms 200 µs 53 min 20 s 50 ms 500 µs 2 h 13 min 20 s 100 ms 1 ms 4 h 26 min 40 s 200 ms 2 ms 8 h 53 min 20 s 500 ms 5 ms 22 h 13 min 20 s 1 s 10 ms 1 d 20 h 26 min 40 s 2 s 20 ms 3 d 16 h 53 min 20 s 5 s 50 ms 9 d 6 h 13 min 20 s 10 s 100 ms 18 d 12 h 26 min 40 s 30 s 300 ms 55 d 13 h 20 min 0 s 50 s 500 ms 92 d 14 h 13 min 20 s 1 min 600 ms 111 d 2 h 40 min 0 s 100 s 1 s 185 d 4 h 26 min 40 s 2 min 1.2 s 222 d 5 h 20 min 0 s 5 min 3 s - abbreviated - Time axis range/div Maximum recording length 80,000 div 10 ms 13 min 20 s 20 ms 26 min 40 s 50 ms 1 h 6 min 40 s 100 ms 2 h 13 min 20 s 200 ms 4 h 26 min 40 s 500 ms 11 h 6 min 40 s 1 s 22 h 13 min 20 s 2 s 1 d 20 h 26 min 40 s 5 s 4 d 15 h 6 min 40 s 10 s 9 d 6 h 13 min 20 s 30 s 27 d 18 h 40 min 0 s 50 s 46 d 7 h 6 min 40 s 1 min 55 d 13 h 20 min 0 s 100 s 92 d 14 h 13 min 20 s 2 min 111 d 2 h 40 min 0 s 5 min 277 d 18 h 40 min 0 s 10 min - abbreviated - 30 min - abbreviated - 1 h - abbreviated - Sampling period: 1 µs, 10 µs, 1 ms, 10 ms, 100 ms *Select within 1/100 of the time axis. Also limited by combination with the time axis setting for memory recording.
Input signals after A/D conversion stored in internal memory can be saved on the optional internal SDD, USB mem- ory, or CF card.Transfer to PC Check and analyze data saved in the internal SSD, USB memory, or CF card, by transferring it to a PC, via LAN or USB. LAN Connection Use the HTTP function to operate MR8827 with a browser on a PC connected via LAN. You can also use the FTP function to retrieve data from internal memory, devices or media connected to the main unit. USB Connection Use a PC to retrieve data saved on devices and media such as internal memory, SSD, or CF card connected to the main unit, via USB. Analyze ■ 9335 Brief Specifications Operating environment Windows 8/7 (32/64-bit), Vista (32-bit), XP Functions - Display functions: Waveform display, X-Y display, Cursor function, etc. - File loading: Readable data formats (.MEM, .REC, .RMS, .POW) / Maximum loadable file size: Maximum file size that can be saved by a given device (file size may be limited depending on the computer con- figuration) - Data conversion: Conversion to CSV format, Batch conversion of mul- tiple files, etc. Print - Print function: Printing image file output (expanded META type, ".EMF") - Print formatting: 1 up, 2-to-16 up, 2-to-16 rows, X-Y 1-to-4 up, preview, hard copy A/D conversion Store in internal memory Save on devices and media SSD CF card USB Memory stick Input Input Transfer LAN USB FTP HTTP WAVE PROCESSOR 9335
- Waveform display, calculations
- Print function (Option) LAN COMMUNICATOR 9333
- Auto-save waveform data to PC
- Remote control via LAN con- nection
- Save in CSV format and trans- fer to spreadsheet programs (Option) LAN USB SSD USB ■ 9333 Brief Specifications Operating environment Windows 8/7 (32/64-bit), Vista (32-bit), XP, (9333 ver.1.09 or later) Functions - Auto-saves waveform data to PC, Remote control of Memory HiCorder (by sending key codes and receiving images on screen), print report, print images from the screen, receive waveform data in same format as waveform files from the Memory HiCorder (binary only) - Waveform data acquisition: Accept auto-saves from the Memory HiCorder, same format as auto-save files of Memory HiCorder (binary only), print automatically with a Memory HiCorder from a PC. The Memory HiCorder’s print key launches printouts on the PC - Waveform viewer: Simple display of waveform files, conversion to CSV format, etc.
iPad App for Memory HiCorder HMR Terminal
- Freely control waveforms using iPad's gesture controls
- Multi-channel support – up to 32 channels (with MR8827, MR8740) of waveform data at your fi ngertips
- Operate the Memory HiCorder via network You can change settings, and monitor waveforms during measurement. *New function on Ver 2.0 Free app (exclusively for iPad) downloadable from the App Store
- Check waveforms with binary data on a PC
- Save data in CSV format and transfer to spreadsheet pro- grams Wave Viewer Wv (Bundled software) A4 size thermal printer 10.4 inch SVGA wave- form monitor CF card slot Rack handle Logic probe connectors (for 8 units) Input module slots (maximum 16 modules) USB 2.0 USB 2.0 (connect to the PC) LAN External control connector Multi-channel support – up to 32 channels (with ■ HMR Terminal Brief specifications (free software) Operating environment iOS on the iPad (Apple Inc. ) Functions - Data acquisition: Send to iPad via FTP using a WiFi router, or load to iPad via iTunes (PC app) - Intuitively operate waveform level searches, maximum / minimum / average values, zero position adjustment, and more at your fingertips - Waveform monitoring - Meter setting * Logic waveforms and computational waveforms are not supported. ■ Wave Viewer (Wv) Brief Specifications Operating environment Windows 8/7 (32/64-bit), Vista (32-bit), XP, 2000 Functions - Simple display of waveform files - Convert binary data files to text format, CSV, etc. - Scroll function, enlarge/reduce display, jump to cursor/trig- ger position, etc.
Transformer Interruption Tests Interchannel isolation allows for safe circuit connections. Simultaneous high-speed sampling can record waveforms before and after the interruption, and allows you to input many control and circuit signals. Inverter / UPS Test Perfect for inverter and UPS evaluation and start-up tests. Record using both logic (control signals) and analog input (primary/secondary voltage or current for a UPS or inverter). Battery Charge/Discharge Tests Input and test the voltage of each battery cell. The MR8827 is built for up to 400 V DC input, protecting it even if high voltage is applied when there is a short-circuit. Power Monitor and Logger By loading the analog output for the effective value (instant power / voltage / current, etc.) calculated by the power analyzer, or by importing the waveform output from the power analyzer to MR8827, you can observe data for long- term tests or irregular waveforms. D/A outputMR8827 Torque / Rotation PW6001 or other PowerPower D/A output MR8827 Power INV Power Power Motor Torque Meter Load D/A output PW6001 or other POWER ANALYZER PPerfect for recording a combina- tion of analog and logic signals that require multiple channels.
The 32 analog channels and 32 logic channels work great for observing input and output signals of an Engine Control Unit. Over 4 hours of recording can be achieved with 1 ms sam- pling. Engine Strain Measurements Use the Strain Unit 8969 to perform strain measurements of up to 32 channels. You can use the numerical value calculation function to automatically calculate the maximum value, minimum value, and P-P value of strain waveforms. Vibration / Endurance Tests Use the long 512MW memory to observe vibration waveforms easily (Memory function). Also, with the recorder function, you can perform long-term observation by capturing the waveform peaks while sampling at high speeds. Injection Molder Evaluation Along with a pneumatic pressure or valve closure, you can record the logic input from control signals. Select from a rich lineup of Hioki input units that support a wide range of sensors and converters. OUTI N I N OUT MR8827 RRecord a diverse array of signals simultaneously
PRINTER UNIT U8350 (Factory-installed option) Features Printer paper one-touch loading, high-speed thermal printing Recording paper 216 mm (8.50 in) × 30 m (98.43 ft), thermal paper roll (use the 9231 paper) Recording witdh: 200 mm (7.87 in) 20 division full scale, 1 div = 10 mm (0.39 in) 80 dots Recording speed Max. 50 mm (1.97 in)/sec Paper feed density 10 lines/mm Display Display 10.4 inch SVGA-TFT color LCD (800 × 600 dots) (Time axis 25 div × Voltage axis 20 div, X-Y 20 div × 20 div) Languages English, Japanese, Korean, Chinese Waveform display zoom/compression Time axis: ×10 to ×2 (zoom at MEMORY function only), ×1, ×1/2 to ×1/20000, Voltage axis: ×100 to ×2, ×1, ×1/2 to ×1/10 Variable display Upper/Lower limit set, display/div set Scaling 10:1 to 1000:1, automatic scaling for various probes Manual scaling (conversion ratio setting, 2-point setting, unit setting) Comment input Alphanumeric input (title, analog and logic channels) Simple input, history input, phrase input Logic waveform Display point move 1 % step, Line width 3 types Display partition Max. eight divisions Monitor function • Input level monitor
- Numerical value (Sampling 10kS/s fixed, refresh rate 0.5s) Other display func- tions
- Waveform inversion (positive/negative)
- Cursor measurement (A, B, 2-cursor, for all channels)
- Vernier function (amplitude fine adjustment)
- Zoom function (horizontal screen division, zoomed waveform shown in lower section)
- 16 selectable colors for waveform display
- Zero position shift in 1% steps for analog waveform
- Global zero adjust for all channels and all ranges MEMORY (high-speed recording) Time axis 5 μs to 5 min/div (100 samples/div) 26 ranges, External sampling (100 samples/div, or free setting), Time axis zoom: ×2 to ×10 in 3 stages, compression: 1/2 to 1/20000 in 13 stages Sampling period 1/100 of time axis range (minimum 50 ns period) Recording length Built-in presets: (at 4, 8, 16ch mode) 25 to 20000 div, (at 4, 8 ch mode) 25 to 500000 div (at 4 ch mode) 25 to 1000000 div Arbitrary presets: setting in 1 div steps, Max. 1280000 div (at 4ch mode), 640000 div (at 8ch mode), 320000 div (at 16ch mode), 160000 div (at 32ch mode) Pre-trigger Record data from before the trigger point at 0 to +100% or -95% of the recording length in 15 stages, or in 1 div step settings Numerical calculation
- Simultaneous calculation for up to 16 selected channels Average value, effective (rms) value, peak to peak value, maximum value, time to maximum value, minimum value, time to minimum value, period, frequency, rise time, fall time, standard deviation, area value, X-Y area value, specified level time, specified time level, pulse width, duty ratio, pulse count, four arithmetic operations, time difference, phase difference, high-level and low-level
- Calculation result evaluation output: GO/NG (with open-collector 5 V out- put)
- Automatic storing of calculation results Waveform processing
- For up to 16 freely selectable channels, the following functions can be performed Four arithmetic operations, absolute value, exponentiation, common logarithm, square root, moving average, differentiation (primary, secondary), integration (primary, secondary), parallel displacement along time axis, trigonometric functions, reverse trigonometric functions, integration time correction for each NPLC setting, auto-saves of calculation results Memory segmentation• Max. 1024 blocks, sequential storage, multi-block storage Other functions
- X-Y waveform synthesis (1 screen, 4 screens)
- Overlay (always overlay when started/overlay only required waveforms)
- Automatic/ Manual/ A-B cursor range printing/ Report printing
- Logging is not available RECORDER (Real-time recording) Time axis 10 ms to 1 hour/div, 19 ranges, time axis resolution 100 points/div Note: Out of data acquired at selected sampling rate, only maximum and minimum value data determined using 100 points/div units are stored Time axis compression selectable in 13 steps, from × 1/2 to × 1/20000 Sampling rate 1/10/100 µs 1/10/100 ms (selectable from 1/100 or less of time axis) Real-time printing Supported * Real-time printing is possible at time axis settings slower than 500 ms/div * Delayed print is performed when recording length is not set to ‘‘Continuous’’ and time axis setting is 10 ms - 200 ms/div * When recording length is set to ‘‘Continuous’’ and time axis setting is 10 ms - 200 ms/div, manual printing can be performed after measurement stop Recording length Built-in presets of 25 - 50000 div, or ‘‘Continuous’’ or arbitrary setting in 1 div steps (max. 80000 div) Waveform memory Store data for most recent 80000 div in memory Auto save Data is automatically saved on CF card, USB memory stick or internal SSD after measurement stops Other functions • Manual/ A-B cursor range printing/ Report printing
- Logging is not available High-speed sampling Culling save Max. Min. High-speed sampling is performed at the set sampling frequency, culling data other than the maximum and minimum values to create the recording data of a certain time. Sampling is done at the set sampling period. Memory recording method Recorder recording method Sampling period Sampling period Basic specifications (product guaranteed for one year) Measurement func- tions MEMORY (high-speed recording) RECORDER (real-time recording) X-Y RECORDER (X-Y real-time recording) FFT (frequency analysis) Main unit OS μ ITRON (Non-Windows OS) Number of channels (Max.) [16 analog input modules]: 32 analog channels + 32 logic channels (logic probe terminals standard, logic has common GND) [14 analog input modules + 2 logic input modules]: 28 analog channels + 64 logic channels (standard 32 channels + 32 channels in Logic unit 8973 ×2) * Max. up to two modules of the Logic Unit 8973, the Current Unit 8971 up to four modules Maximum sampling rate
20 MS/second (all channels simultaneously)
16MW/ch (total capacity 512MW memory), 16MW/ch (using 32 analog chan- nels), 32MW/ch (using 16 analog channels), 64MW/ch (using 8 analog channels), 128MW/ch (using 4 analog channels) Data storage media CF card slot (standard) ×1 (up to 2GB, FAT, or FAT-32 format), USB port ×2 (USB 2.0) Backup battery life Clock and parameter setting backup: at least 10 years (reference value at 25°C) External control con- nectors External trigger input, Trigger output, External sampling input, GND, Two external outputs (GO/NG output), Three external inputs (start/IN1, stop/ IN2, save/IN3) External interfaces LAN: 100BASE-TX (DHCP, DNS supported, FTP server, HTTP server) USB: USB 2.0 compliant, series A receptacle ×1, series B receptacle ×1, (File transfer SSD/ CF card to PC, or remote control from PC) Environmental condi- tions (No condensation) Operation: 0°C to 40°C (32°F to 104°F), 20% to 80% rh Storage: -10°C to 50°C (14°F to 122°F), 90% rh or less Standards Safety: EN61010 EMC: EN61326, EN61000-3-2, EN61000-3-3 Power supply AC 100 to 240 V, 50/60 Hz Power consumption 220 VA max. (when not using the printer), 350 VA max. (when using the printer) Dimensions and mass401 mm (15.79 in)W × 233 mm (9.17 in)H × 388 mm (15.28 in)D, 12.6 kg (444.4 oz) (main unit only) Supplied accessories Instruction manual ×1, Application disk (Wave Viewer Wv, Communication commands table) ×1, Power cord ×1, Input cord label ×1, USB cable ×1, Printer paper ×1 (when equipped with a printer unit), Roll paper attachment ×2 (when equipped with a printer unit) Main unit Specifications
Designate a range of the wave - form stored in the memory func - tion to perform FFT analysis. It is rendered simultaneously on the screen. How is FFT Analysis Performed? Waveform in the MEM function FFT analysis dB V Convert data measured with few calculation points into data with many points for re-analysis. *Not possible with frequency averaging ON 1000P 10000P Display the spectrum as it changes over time in 3-D. -6.51 dB 93.50 dB Before scaling After scaling Scale by dB. Input the overall value (sum of the power spec - trum) directly as a dB value. X-Y RECORDER (X-Y real-time recording) Sampling period 1/10/100 ms (dot), 10/100 ms (line) Recording length Continuous Screen, Printing Split screen (1 or 4), Manual printing only Number of X-Y 1 to 8 phenomenon X-Y channel setting Any 8 channels out of 16 can be selected for X axis and Y axis respectively X-Y axis resolution 25 dots/div (screen), horizontal 80 dots/div × vertical 80 dots/div (printer) Waveform memory Sampling data for last 16000000 points are stored in memory Pen up/down Simultaneous for all phenomena External pen control Possible via external input connector (simultaneous up/down for all phenomena) FFT Analysis mode Storage waveform, Linear spectrum, RMS spectrum, Power spectrum, Density of power spectrum, Cross power spectrum, Auto-correlation function, Histogram, Transfer function, Cross- correlation function, Impulse response, Coherence function, 1/1 Octave analysis, 1/3 Octave analysis, LPC analysis, Phase spectrum Analysis channels Selectable from all analog input channels Frequency range 133 mHz to 8 MHz, External, (resolution 1/400, 1/800, 1/2000, 1/4000) Number of sampling points 1000, 2000, 5000, 10000 points Window functions Rectangular, Hanning, Hamming, Blackman, Blackman-Harris, Flat-top, Exponential Display format Single, Dual, Nyquist, Running spectrum Averaging function Time axis / frequency axis simple averaging, Exponential averaging, Peak hold (frequency axis), Averaging times (2 to 10000 times) Print functions Same as the MEMORY function (partial print not available) Trigger functions Trigger mode MEMORY (high-speed recording), FFT: Single, Repeat, Auto RECORDER (real-time recording): Single, Repeat Trigger sources CH1 to CH32 (analog), Standard Logic 32ch + Logic Unit (Max. 2 units 32 channels), External (a rise of 2.5V or terminal short circuit), Timer, Manual (either ON or OFF for each source), Logical AND/OR of sources Trigger types
- Level: Triggering occurs when preset voltage level is crossed (upwards or downwards)
- Voltage drop: Triggering occurs when voltage drops below peak voltage setting (for 50/60 Hz AC power lines only)
- Window: Triggering occurs when window defined by upper and lower limit is entered or exited
- Period: Rising edge or falling edge cycle of preset voltage value is monitored and triggering occurs when defined cycle range is exceeded
- Glitch: Triggering occurs when pulse width from rising or falling edge of preset voltage value is under run
- Event setting: Event count is performed for each source, and triggering occurs when a preset count is exceeded
- Logic: 1, 0, or ×, Pattern setting Level setting resolution 0.1% of full scale (full scale = 20 divisions) Trigger filter Selectable 0.1 div to 10.0 div 9 steps, or OFF (at MEMORY function) ON (10 ms fixed) or OFF (at RECORDER function) Trigger output Open collector (5 voltage output, active Low) At Level setting: pulse width (Sampling period × data number after trigger) At Pulse setting: pulse width (2 ms) Other functions Trigger priority (OFF/ON), Pre-trigger function for capturing data from before / after trigger event (at MEMORY function), Level display during trigger standby, Start and stop trigger (At RECORDER function), Trigger search function Other functions Waveform judgment function (In MEMORY or FFT function)
- Area comparison with reference waveform area for time domain waveform, X-Y waveform, or FFT analysis waveform
- Parameter calculated value comparison with reference value
- Output: GO/NG decision, Open-collector 5V, *100 msec/div (1 msec sampling) and thereafter allows for evaluation in almost real-time.
ANALOG UNIT 8966 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for voltage measurement Input connectors Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) Measurement range 5 mV to 20 V/div, 12 ranges, full scale: 20 div, AC voltage for possible measurement/display using the memory function: 280 V rms, Low-pass filter: 5/50/500 Hz, 5 k/50 k/500 kHz Measurement resolution 1/100 of measurement range (using 12-bit A/D conversion) Highest sampling rate 20 MS/s (simultaneous sampling across 2 channels) Measurement accuracy ±0.5 % of full scale (with filter 5 Hz, zero position accuracy included) Frequency characteristics DC to 5 MHz -3 dB, (with AC coupling: 7 Hz to 5 MHz -3dB) Input coupling AC/DC/GND Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage) approx. 250 g (8.8 oz) Accessories: None TEMP UNIT 8967 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for temperature measurement with thermocouple (voltage measurement not available) Input connectors Thermocouple input: plug-in connector, Recommended wire diameter: single-wire, 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16 Input impedance: min. 5 MΩ (with line fault detection ON/OFF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) Temperature measurement range Note: Upper and lower limit values depend on the thermocouple 10°C/div (-100 °C to 200 °C), 50°C/div (-200 °C to 1000 °C), 100°C/div (-200°C to 2000°C), 3 ranges, full scale: 20 div, Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion) Thermocouple range (JIS C 1602-1995) (ASTM E-988-96) K: -200 to 1350°C, J: -200 to 1100°C, E: -200 to 800°C, T: -200 to 400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 400 to 1800°C, W (WRe5-26): 0 to 2000°C, Reference junction compensation: internal/ external (switchable), Line fault detection ON/OFF possible Data refresh rate 3 methods, Fast: 1.2 ms (digital filter OFF), Normal: 100 ms (digital filter 50/60 Hz), Slow: 500 ms (digital filter 10Hz) Measurement accuracy Thermocouple K, J, E, T, N: ±0.1 % of full scale ±1°C (±0.1 % of full scale ±2°C at -200°C to 0°C), Thermocouple R, S, W: ±0.1 % of full scale ±3.5°C (at 0°C to 400°C or less), ±0.1 % of full scale ±3°C (at 400°C or more) Thermocouple B: ±0.1 % of full scale ±3°C (at 400°C or more), Reference junction compensation accuracy: ±1.5°C (added to measurement accuracy with internal reference junction compensation) approx. 240 g (8.5 oz) Accessories: Ferrite clamp × 2 HIGH RESOLUTION UNIT 8968 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for voltage measurement Input connectors Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) Measurement range 5 mV to 20 V/div, 12 ranges, full scale: 20 div, AC voltage for possible measurement/display using the memory function: 280 V rms, Low-pass filter: 5/50/500 Hz, 5k/50k Hz Anti-aliasing filter Integrated filter for suppressing aliasing distortion caused by FFT processing (automatic cutoff frequency setting/OFF) Measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion) Highest sampling rate 1 MS/s (simultaneous sampling across 2 channels) Measurement accuracy±0.3 % of full scale (with filter 5 Hz, zero position accuracy included) Frequency characteristics DC to 100 kHz -3 dB, (with AC coupling: 7 Hz to 100 kHz -3dB) Input coupling AC/DC/GND Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage) approx. 250 g (8.8 oz) Accessories: None STRAIN UNIT 8969 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time and auto-balance; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for distortion measurement (electronic auto- balancing, balance adjustment range within ±10000 με) Input connectors Weidmuller SL 3.5/7/90G (via Conversion Cable 9769, TAJIMI PRC03-12A10- 7M10.5) Max. rated voltage to earth: 33 Vrms or 70 V DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) Suitable transducer Strain gauge converter, Bridge impedance: 120 Ω to 1 kΩ, Bridge voltage: 2 V ±0.05 V, Gauge rate: 2.0 Measurement range 20 με to 1000 με/div, 6 ranges, full scale: 20 div, Low-pass filter: 5/10/100 Hz, 1 kHz Measurement resolution 1/1250 of measurement range (using 16-bit A/D conversion) Highest sampling rate 200 kS/s (simultaneous sampling across 2 channels) Measurement accuracy ±(0.5 % of full scale +4 με) (at 5 Hz filter ON, After auto-balancing) Frequency characteristics DC to 20 kHz +1/-3dB approx. 220 g (7.8 oz) Accessories: Conversion cable 9769 × 2 (cable length 50 cm/1.64 ft) FREQ UNIT 8970 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for voltage input based frequency measurement, rotation, power frequency, integration, pulse duty ratio, pulse width Input connectors Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) Frequency mode Range: Between DC to 100 kHz (minimum pulse width 2 μs), 1 Hz/div to 5 kHz/div (full scale= 20 div), 8 settings Rotation mode Range: Between 0 to 2 million rotations/minute (minimum pulse width 2 μs), 100 (r/min)/div to 100 k (r/min)/div (full scale= 20 div), 7 settings Power frequency mode Range: 50 Hz (40 - 60 Hz), 60 Hz (50 - 70 Hz), 400 Hz (390 - 410 Hz) (full scale= 20 div), 3 settings Accuracy: ±0.03 Hz (exclude 400 Hz range), ±0.1 Hz (400 Hz range) Integration mode Range: 2 k counts/div to 1 M counts/div, 6 settings Accuracy: ±range/2000 Duty ratio mode Range: Between 10 Hz to 100 kHz (minimum pulse width 2 μs), 5%/div (full scale=20 div) Accuracy: ±1% (10 Hz to 10 kHz), ±4% (10 kHz to 100 kHz) Pulse width mode Range: Between 2 μs to 2 sec, 500 μs/div to 100 ms/dv (full scale=20 div) Accuracy: ±0.1% f.s. Measurement resolution 1/2000 of range (Integration mode), 1/500 of range (exclude integration, power frequency mode), 1/100 of range (power frequency mode) Input voltage range and threshold level ±10 V to ±400 V, 6 settings, selectable threshold level at each range Other functions Slope, Level, Hold, Smoothing, Low-pass filter, Switchable DC/AC input coupling, Frequency dividing, Integration over-range keep/return approx. 250 g (8.8 oz) Accessories: None CURRENT UNIT 8971 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, Current measurement with optional current sensor, Note: Maximum 4 units connectable to main unit Input connectors Sensor connector (input impedance 1 MΩ, exclusive connector for current sensor via conversion cable the 9318, common GND with recorder) Compatible current sensors CT6863, CT6862, 9709, 9279, 9278, 9277, 9272 -10 (To connect the 8971 via conversion cable the 9318) Measurement range Using 9272-10 (20 A), 9277: 100 mA to 5 A/div (f.s.=20 div, 6 settings) Using CT6862: 200 mA to 10 A/div (f.s.=20 div, 6 settings) Using 9272-10 (200 A), 9278, CT6863: 1 A to 50 A/div (f.s.=20 div, 6 settings) Using 9279, 9709: 2 A to 100 A/div (f.s.=20 div, 6 settings) Accuracy Using 9278, 9279: ±0.85% f.s. Using other sensor: ±0.65% f.s. RMS amplitude accuracy: ±1% f.s. (DC, 30 Hz to 1 kHz), ±3% f.s. (1 kHz to 10 kHz) RMS response time: 100 ms (rise time from 0 to 90% of full scale), Crest factor: 2 Frequency characteristics: DC to 100 kHz, ±3 dB (with AC coupling: 7 Hz to 100 kHz) Measurement resolution 1/100 of range (using 12-bit A/D conversion) Highest sampling rate 1 MS/s (simultaneous sampling across 2 channels) Other functions Input coupling: AC/DC/GND, Low-pass filter: 5, 50, 500, 5 k, 50 kHz approx. 250 g (8.8 oz) Accessories: CONVERSION CABLE 9318 × 2 (To connect the current sensor to the 8971) ■ Options specifications (sold separately)
DC/RMS UNIT 8972 (Accuracy at 23 ± 5°C/73 ±9°F, 20 to 80 % rh after 30 minutes of warm-up time and zero adjustment; accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for voltage measurement, DC/RMS selectable Input connectors Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) Measurement range 5 mV to 20 V/div, 12 ranges, full scale: 20 div, AC voltage for possible measurement/display using the memory function: 280 V rms, Low-pass filter: 5/50/500 Hz, 5 k/100 kHz Measurement resolution 1/100 of measurement range (using 12-bit A/D conversion) Highest sampling rate 1 MS/s (simultaneous sampling across 2 channels) Measurement accuracy±0.5 % of full scale (with filter 5 Hz, zero position accuracy included) RMS measurement RMS amplitude accuracy: ±1 % of full scale (DC, 30 Hz to 1 kHz), ±3 % of full scale (1 kHz to 100 kHz), Response time: SLOW 5 s (rise time from 0 to 90% of full scale), MID 800 ms (rise time from 0 to 90% of full scale), FAST 100 ms (rise time from 0 to 90% of full scale), Crest factor: 2 Frequency characteristics DC to 400 kHz -3 dB, (with AC coupling: 7 Hz to 400 kHz -3dB) Input coupling AC/DC/GND Max. allowable input 400 V DC (the maximum voltage that can be applied across input pins without damage) mm, approx. 250 g (8.8 oz) Accessories: None DIFFERENTIAL PROBE P9000 (accuracy guaranteed for 1 year) Measurement modes P9000-01: For waveform monitor output, Frequency properties: DC to 100 kHz -3 dB P9000-02: Switches between waveform monitor output/AC effective value output Wave mode frequency properties: DC to 100 kHz -3 dB, RMS mode frequency properties: 30 Hz to 10 kHz, Response time: Rise 300 ms, Fall 600 ms Division ratio Switches between 1000:1, 100:1 Effective value measurement accuracy ±1 % f.s. (30 Hz to less than 1 kHz, sine wave) , ±3 % f.s. (1 kHz to 10 kHz, sine wave) Input resistance/ capacity H-L: 10.5 MΩ, 5 pF or less (at 100 kHz) Maximum input voltage 1000 V AC, DC Maximum rated voltage to ground 1000 V AC, DC (CAT III) Operating temperature range -40°C to 80°C (-40°F to 176°F) Power supply (1) AC adapter Z1008 (100 to 240 V AC, 50/60 Hz), 6 VA (including AC adapter),
0.9 VA (main unit only)
(2) USB bus power (5 V DC, USB-microB terminal), 0.8 VA (3) External power source 2.7 V to 15 V DC, 1 VA Accessories Instruction manual ×1, Alligator clip ×2, Carrying case ×1 LOGIC UNIT 8973 (accuracy guaranteed for 1 year) Measurement functions Number of channels: 16 channels (4 ch/1 probe connector × 4 connectors) Input connectors Mini DIN connector (for HIOKI logic probes only), Compatible logic probes: 9320-01, 9327, MR9321-01 approx. 190 g (6.7 oz) Accessories: None LOGIC PROBE 9320-01/9327 (Accuracy at 23 ± 5°C/73 ±9°F, 35 to 80% rh, accuracy / product guaranteed for 1 year) Function Detection of voltage signal or relay contact signal for High/Low state recording Input 4 channels (common ground between unit and channels), digital/contact input, switchable (contact input can detect open-collector signals) Input resistance: 1 MΩ (with digital input, 0 to +5 V) 500 kΩ or more (with digital input, +5 to +50V) Pull-up resistance: 2 kΩ (contact input: internally pulled up to +5 V) Digital input threshold 1.4 V/ 2.5 V/ 4.0 V Contact input detection resistance 1.4 V: 1.5 kΩ or higher (open) and 500 Ω or lower (short) 2.5 V: 3.5 kΩ or higher (open) and 1.5 kΩ or lower (short)
4.0 V: 25 kΩ or higher (open) and 8 kΩ or lower (short)
Response speed 9320-01: 500 ns or lower, 9327: detectable pulse width 100 ns or higher Max. allowable input 0 to +50 V DC (the maximum voltage that can be applied across input pins without damage) Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 30 cm (0.98 ft), approx. 150 g (5.3 oz) Note: The unit-side plug of the 9320-01 is different from the 9320Cable length and mass: 70 cm (2.30 ft), Output side: 1.5 m (4.92 ft), 170g (6.0 oz) LOGIC PROBE MR9321-01 (Accuracy at 23 ± 5°C/73 ±9°F, 35 to 80% rh, accuracy / product guaranteed for 1 year) Function Detection of AC or DC relay drive signal for High/Low state recording Can also be used for power line interruption detection Input 4 channels (isolated between unit and channels), HIGH/LOW range switching Input resistance: 100 kΩ or higher (HIGH range), 30 kΩ or higher (LOW range) Output (H) detection 170 to 250 V AC, ±DC 70 to 250 V (HIGH range) 60 to 150 V AC, ±DC 20 to 150 V (LOW range) Output (L) detection 0 to 30 V AC, ±DC 0 to 43 V (HIGH range) 0 to 10 V AC, ±DC 0 to 15 V (LOW range) Response time Rising edge 1 ms max., falling edge 3 ms max. (with HIGH range at 200 V DC, LOW range at 100 V DC) Max. allowable input 250 Vrms (HIGH range), 150 Vrms (LOW range) (the maximum voltage that can be applied across input pins without damage) Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 1 m (3.28 ft), approx. 320 g (11.3 oz) Note: The unit-side plug of the MR9321-01 is different from the MR9321 . DIGITAL VOLTMETER UNIT MR8990 (Accuracy at 23 ±5°C, 20 to 80% rh after 30 minutes of warm-up time and calibration, product and accuracy guaranteed for 1 year) Measurement functions Number of channels: 2, for DC voltage measurement Input connectors Banana input connectors (Input resistance: 100 MΩ or higher with 100 mV f.s. to 10 V f.s. range, otherwise 10 MΩ) Max. rated voltage to earth: 300 V AC or DC (with input isolated from the main unit, the maximum voltage that can be applied between input channel and chassis, and between input channels without damage) Measurement ranges100 mV f.s. (5 mV/div) to 1000 V f.s. (50 V/div), 5 ranges, full scale: 20 div Measurement resolution 1/1000000 of measurement range (using 24 bit ΔΣ modulation A/D) Integration time 20 ms ×NPLC (during 50 Hz), 16.67 ms ×NPLC (during 60 Hz) Basic measurement (at range of 1000 mV f.s.) Maximum input voltage
500 V DC
(maximum voltage that can be applied between input connectors without damage) approx. 260g (8.8 oz) Accessories: None ■ Options specifications (sold separately)
Input modules *Input cords are not included. Please purchase them separately. *When using 9709 with Current Unit 8971, a total of 7 current probes can be used. ANALOG UNIT 8966 2 ch, Voltage input, DC to 5 MHz bandwidth TEMP UNIT 8967 2 ch, thermocouple temperature input HIGH RESOLUTION UNIT 8968 2 ch, voltage input, DC to 100 kHz bandwidth STRAIN UNIT 8969 2 ch, strain gauge type converter amp Conversion Cable 9769 For the 8969(MR8847/8827 series), bundled with the 8969 FREQ UNIT 8970 2 ch, for measurement of frequency, rpm, pulse, etc. CURRENT UNIT 8971 2 ch, for measuring current using dedicated current sensors, bundled two Conversion cable 9318 * The Current unit 8971 up to four moduleDC/RMS UNIT 8972 2 ch, voltage/DC to 400 kHz, RMS rectifier, DC and 30 to 100 kHz bandwidth LOGIC UNIT 8973 4 terminals, 16 ch * Max. up to two modules of the Logic unit 8973 DIGITAL VOLTMETER UNIT MR8990 2ch, high-precision DC V input, 0.1μV resolution, high- speed sampling 500 times/s Printer options *PRINTER UNIT is a built-in option that must be specified upon order. PRINTER UNIT U8350 Specified upon order. Printing width 200 mm (7.87 inch). Compatible recording paper: Model 9231 RECORDING PAPER 9231 A4 width 216 mm (8.50 in) × 30 m (98.43 ft), 6 rolls/set Main unit MEMORY HiCORDER MR8827 Note: Main unit MR8827 cannot operate alone. You must install one or more optional input modules in the unit. Logic signal measurement LOGIC PROBE 9327 4-channel type, for voltage/contact signal ON/OFF detection (response pulse width 100 ns or more, minia - ture terminal type) LOGIC PROBE MR9321-01 4 isolated channels, ON/OFF detec - tion of AC/DC voltage (miniature terminal type) LOGIC PROBE 9320-01 4-channel type, for voltage/contact signal ON/OFF detection (response pulse width 500 ns or more, minia - ture terminal type) CONVERSION CABLE 9323 *Used for connecting the 9320/9321/ MR9321 and the 9324 to the Memory HiCorder with small logic terminal models * This cable is not required for the small- terminal types 9327, 9320-01, 9321-01 and MR9321-01. Storage media *SSD is a built-in option that must be specified upon order. *The CF card includes a PC card adapter. SSD UNIT U8330 Specified upon order, built-in type, 128 GB *PC Card Precaution Use only PC Cards sold by HIOKI. Compatibility and performance are not guaranteed for PC cards made by other manufacturers. You may be unable to read from or save data to such cards. PC CARD 2G 9830 (2 GB) PC CARD 1G 9729 (1 GB) PC CARD 512M 9728 (512 MB) Case CARRYING CASE (special order) hard trunk type Inquire with your local Hioki distributor. PC Software WAVE PROCESSOR 9335 Convert data, print and display waveforms LAN COMMUNICATOR 9333
- Waveform data collect function
- Remote control with the PC iPad App for MEMORY HiCORDER HMR Terminal Download from the App Store Free (exclusively for Apple Inc. iPad) LAN CABLE 9642 Straight Ethernet cable, supplied with straight to cross conversion cable, 5 m (16.41 ft) length Recommended System Chart Recommended Recommended RecommendedOrder Code MR8827 Order Code U8350 Order Code 9231 Order Code U8330 Order Code 9830 Order Code 9729 Order Code 9728 Order Code 9335 Order Code 9333 Order Code 9642 Order Code 8966 Order Code 8967 Order Code 8968 Order Code 8969 Order Code 9769 Order Code 8970 Order Code 8971 Order Code 8972 Order Code 8973 Order Code MR8990 Order Code 9327 Order Code MR9321-01 Order Code 9320-01 Order Code 9323
INPUT CORD (B) *Voltage is limited to the specifications of the input modules in use CONNECTION CORD L9198 φ 5.0 mm (0.20 in) dia., cable allowing for up to 300 V input. 1.7 m (5.58 ft) length, small alligator clip CONNECTION CORD L9197 φ 5.0 mm (0.20 in) dia., cable allowing for up to 600 V input. 1.8 m (5.91 ft) length, a detachable large alligator clips are bundled GRABBER CLIP 9243 Attaches to the tip of the banana plug cable, CAT III 1000 V, 196 mm (7.72 in) length INPUT CORD (C) *Voltage is limited to the specifications of the input modules in use 10:1 PROBE 9665 Note: This probe does not expand the maximum rated voltage above ground of an isolated input. Max. rated voltage to earth is same as for input module, max. input voltage 1 kV rms (up to 500 kHz), 1.5 m (4.92 ft) length 100:1 PROBE 9666 Note: This probe does not expand the maximum rated voltage above ground of an isolated input. Max. rated voltage to earth is same as for input module, max. input voltage 5 kV peak (up to 1MHz), 1.5 m (4.92 ft) length Custom cable *For P9000. Inquire with your Hioki distributor. Bus powered USB cable For power supply from a USB-A terminal USB(A)- Micro B cable For power supply via USB Micro B terminal from a USB-A terminal 3-prong cable AC adapter output terminal split into 3 Other options for Input CONNECTION CORD L9217 Cord has insulated BNC connectors at both ends, signal output use, 1.6 m (5.25 ft) length CONVERSION ADAPTER 9199 Receiving side banana, output BNC terminal INPUT CORD (E) *For the MR8990 *Voltage is limited to the specifi - cations of the input modules in use TEST LEAD L2200 Cable length: 70 cm, tips interchangeable with a pin lead or alligator clip, maximum input voltage: CAT IV 600V, CAT III 1000V Up to 30 A (High speed) CLAMP ON PROBE 3273-50 DC to 50 MHz wide band response, mA- class current up to 30 Arms CLAMP ON PROBE 3276 DC to 100 MHz wide band response, mA- class current up to 30 Arms Up to 150 A (High speed) CLAMP ON PROBE 3274 DC to 10 MHz wideband response, mA- class current up to 150 Arms POWER SUPPLY POWER SUPPLY 3272 For the 3270 series, single sensor connectable (2 units pos - sible depending on conditions) POWER SUPPLY 3269 For the 3270 series, connect up to four sensors POWER SUPPLY *Not necessary when using Current Unit 8971 SENSOR UNIT 9555-10 Power supply for the Current Sensor, used alone CONNECTION CORD L9217 Cord has insulated BNC connec - tors at both ends, signal output use, 1.6 m (5.25 ft) length Up to 20A (High precision) AC/DC CURRENT PROBE CT6841
20 A AC/DC rated current, DC to 1 MHz
response, 20 mm (0.79 in) core dia., 3 m (9.84 ft) cord length Up to 50 A (High precision) AC/DC CURRENT SENSOR CT6862 CAT III 1000 V, 50 A AC/DC rated current, DC to 1 MHz response, φ 24 mm (0.94 in) core dia., 3 m (9.84 ft) cord length Up to 200 A (High precision) AC/DC CURRENT SENSOR CT6863 CAT III 1000 V, 200 A AC/DC rated current, DC to 500 kHz response, φ 24 mm (0.94 in) core dia., 3 m (9.84 ft) cord length AC/DC CURRENT PROBE CT6843
200 A AC/DC rated current, DC to 500kHz
response, 20 mm (0.79 in) core dia., 3 m (9.84 ft) cord length CLAMP ON SENSOR 9272-10 CAT III 600 Vrms, 20 A/200 A AC rated current, 1 Hz to 100 kHz response, φ 46 mm (1.81 in) core dia., 3 m (9.84 ft) cord length Up to 500 A (High precision) AC/DC CURRENT SENSOR 9709 CAT III 1000 V, 500 A AC/DC rated current, DC to 100 kHz response, φ 36 mm (1.42 in) core dia., 3 m (9.84 ft) cord length UNIVERSAL CLAMP ON CT 9279-01
600 Vrms insulated wire, 500 A AC/DC
rated current, DC to 20 kHz response, φ 40 mm (1.57 in) core dia., 3 m (9.84 ft) cord length, (CE marked)
500 A to 1000 A *For commercial power lines,
50/60Hz (separate power supply not required) CLAMP ON PROBE 9018-50 Excellent phase characteristics, Input from 10 to 500 A, 40 Hz to 3 kHz for
0.2 V AC output, BNC terminal
Input from 20 to 1000 A, 40 Hz to 1 kHz for 0.2 V AC output, BNC terminal *You can connect up to 4 Current Unit 8971 to the Memory HiCorder main unit, allowing up to 8 current sensors to be used. *There is no limit for connecting current sensors to voltage input analog units. Up to 500 A (High speed) CLAMP ON PROBE 3275 DC to 2 MHz wideband response, mA- class current up to 500 Arms
100 A to 5000 A (Medium speed)
DC to 10kHz (-3dB), 100A, Output 0.1 V/f.s. CLAMP ON AC/DC SENSOR CT9692-90 DC to 20kHz (-3dB), 200A, Output 0.2 V/f.s. CLAMP ON AC/DC SENSOR CT9693-90 DC to 15kHz (-3dB), 2000A, Output 0.2 V/f.s. FLEXIBLE CLAMP ON SENSOR CT9667 10Hz to 20kHz (±3dB), AC 5000A/500A, Output AC 500mV/f.s., φ 254 mm (10.0 in) core dia. LEAK CURRENT *For commercial power lines, 50/60 Hz CLAMP ON LEAK HiTESTER 3283 10mA range / 10μA resolution to 200A range, moniter / analog output 1V f.s. OUTPUT CORD 9094 3.5mm (0.14in) dia. mini plug to banana , 1.5m (4.92ft) length CONVERSION ADAPTER 9199 Receiving side banana, output BNC terminal AC ADAPTER 9445-02 For USA, 100 to 240 V AC, 9 V/ 1 A AC ADAPTER 9445-03 For EU 100 to 240 V AC, 9 V/ 1 A INPUT CORD (A) *Voltage is limited to the specifications of the input modules in use CONNECTION CORD L9790 Flexible φ 4.1 mm (0.16 in) thin dia., cable allowing for up to 600 V input. 1.8 m (5.91 ft) length * The end clip is sold separately. ALLIGATOR CLIP L9790-01 Red/black set attaches to the ends of the cables L9790 GRABBER CLIP 9790-02 Red/black set attaches to the ends of the cables L9790 * When this clip is attached to the end of the L9790, input is limited to CAT II 300 V. Red/black set. CONTACT PIN 9790-03 Red/black set attaches to the ends of the cables L9790 Recommended Recommended INPUT CORD (D) *For measuring high voltage. *Voltage to ground is within this product's specifications, and is not affected by connected input modules. Separate power source is also required. DIFFERENTIAL PROBE P9000-01 (Wave mode only) For the Memory HiCorder series, input up to 1kV AC/DC DIFFERENTIAL PROBE P9000-02 (Select between WAVE/RMS mode) For the Memory HiCorder series, input up to 1kV AC/DC AC ADAPTER Z1008 100 to 240 V AC Recommended Order Code L9790 Order Code L9790-01 Order Code 9790-02 Order Code 9790-03 Order Code L9198 Order Code L9197 Order Code 9243 Order Code 9665 Order Code 9666 Order Code P9000-01 Order Code P9000-02 Order Code Z1008 Order Code L2200 Order Code L9217 Order Code 9199 Order Code CT6841 Order Code CT6862 Order Code CT6863 Order Code CT6843 Order Code 9272-10 Order Code 9709 Order Code 9279-01 Order Code 3273-50 Order Code 3276 Order Code 3274 Order Code 3275 Order Code 9018-50 Order Code 9132-50 Order Code 9555-10 Order Code L9217 Order Code 3272 Order Code 3269 Order Code 3283 Order Code 9094 Order Code 9199 Order Code 9445-02 9445-03 Order Code CT9691-90 Order Code CT9692-90 Order Code CT9693-90 Order Code CT9667 Temperature sensor *For reference only. Please purchase locally. Thermocouple
81 Koizumi, Ueda, Nagano, 386-1192, Japan
TEL +81-268-28-0562 FAX +81-268-28-0568 http://www.hioki.com / E-mail: os-com@hioki.co.jp HIOKI USA CORPORATION TEL +1-609-409-9109 FAX +1-609-409-9108 http://www.hiokiusa.com / E-mail: hioki@hiokiusa.com All information correct as of Jan. 30, 2015. All specifi cations are subject to change without notice. MR8827E1-51E Printed in Japan DISTRIBUTED BYHIOKI (Shanghai) SALES & TRADING CO., LTD. TEL +86-21-63910090 FAX +86-21-63910360 http://www.hioki.cn / E-mail: info@hioki.com.cn HIOKI INDIA PRIVATE LIMITED TEL +91-124-6590210 FAX +91-124-6460113 E-mail: hioki@hioki.in HIOKI SINGAPORE PTE. LTD. TEL +65-6634-7677 FAX +65-6634-7477 E-mail: info-sg@hioki.com.sg HIOKI KOREA CO., LTD. TEL +82-42-936-1281 FAX +82-42-936-1284 E-mail: info-kr@hioki.co.jp Note: Company names and Product names appearing in this catalog are trademarks or registered trademarks of various companies. Evolution Will Never Cease.