M9391A PXIe Vector Signal Analyzer Data Sheet
1 MHz to 3 GHz or 6 GHz
Technical Specifications and Characteristics 4 Definitions for specifications 4 Recommended best practices in use 4 Block diagram 5 Frequency 6 Amplitude 8 Dynamic range 11 Spectral purity 14 Data acquisition 16 Measurement speed 17 Format-specific measurement data 17 Environmental and physical specifications 21 S y s t e m r e q u i r e m e n t s 2 2 S o f t w a r e 23 Setup and Calibration Services 24 Support and Warranty 25 Configuration and Ordering Information 26 Table of Contents
3 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Overview Be ready for tomorrow - today RF requirements keep growing while timelines keep shrinking. To help ease the technical and business pressures, the right test solution provides continuity in measurements and longevity in capability. The Keysight Technologies, Inc. M9391A PXIe vector signal analyzer (PXI VSA) is the next logical step in RF signal analysis. The M9391A PXI VSA, combined with the M9381A PXIe vec - tor signal generator provides a complete solution for fast, high quality measurements optimized for RF manufacturing test environments. To help you get proven results even faster, Keysight’s PXI VSA can be used with X-Series measurement applications for modular instruments, 89600 VSA software and System - Vue. These software applications enable you to investigate, validate and test your RF communications designs. From fully modular hardware to software leverage to world - wide support, the PXI VSA is the low-risk way to manage change and be ready for tomorrow—today. Product description The M9391A PXI VSA is a modular vector signal analyzer for frequencies from 1 MHz to 6 GHz and up to 160 MHz analysis bandwidth. The M9391A is comprised of four individual PXI modules - M9350A downconverter, M9214A digitizer, M9301A synthesizer and M9300A frequency reference. A single M9300A frequency reference can be shared between multiple instruments to minimize footprint. The flexible, modular design of the M9391A enables you to efficiently scale to multi-channel signal analysis to test multiple-input, multiple-output (MIMO) devices. Capability can also be scaled with options for memory, frequency range and modulation bandwidth which can be easily upgraded in the field.
Applications
– Power amplifier and front-end-module design validation and manufacturing – Radio transceiver design validation and production test – MIMO and multi-channel device test Figure 1. M9391A PXIe vector signal analyzer with four modules consisting
4 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Definitions for specifications Temperatures referred to in this document are defined as follows: – Full temperature range = Individual module temperature of 25 to 75 °C, as reported by the module, and environment temperature of 0 to 55 °C. – Controlled temperature range = Individual module tempera - ture of 40 to 51 °C, as reported by the module, and environ - ment temperature of 20 to 30 °C. Specifications describe the warranted performance of calibrated instruments. Data represented in this document are specifications under the following conditions unless otherwise noted. – Calibrated instruments have been stored for a minimum of 2 hours within the full temperature range – 45 minute warm-up time – Calibration cycle maintained – When used with Keysight M9300A frequency reference and Keysight interconnect cables Characteristics describe product performance that is useful in the application of the product, but that is not covered by the product warranty. Characteristics are often referred to as Typical or Nominal values and are italicized. – Typical describes characteristic performance, which 80% of instruments will meet when operated within the controlled temperature range. – Nominal describes representative performance that is useful in the application of the product when operated within the controlled temperature range. Recommended best practices in use – Use slot blockers and EMC filler panels in empty module slots to ensure proper operating temperatures. Keysight chassis and slot blockers optimize module temperature performance and reliability of test. – Set chassis fan to high at environmental temperatures above 45 °C – Maintain temperature stability for best multi-channel phase coherence – Set chassis fans to maximum – Maintain stable ambient temperature – Perform warm-up with session open and representative acquisition waveform running Conversion type operating range Conversion types Frequency range Auto 1 MHz to 3 or 6 GHz Image protect 1 MHz to 3 or 6 GHz Single high 400 MHz to 3 or 6 GHz Single low 1.1 GHz to 3 or 6 GHz Additional information – Mixer level offset modifies the receiver gain prior to the first mixer of the receiver. A negative setting improves distortion positive setting improves noise performance at the cost of distortion. – Performance described in this document applies for module temperature within ± 3 degrees of comprehensive alignment, unless otherwise noted. – When used with a Keysight M9018A PXIe chassis, compre - hensive alignment requires chassis FPGA version 1.05 or greater. – When configured for multi-channel, phase-coherent operation (shared synthesizer configuration), instrument level warranted specifications only apply to the M9391A which was previously calibrated with the M9301A synthesizer, showing a valid calibration indicator. For all other M9391A channels, specifications revert to typical performance. If using an external LO distribution unit, such as the V2802A LO distribution network, specifications for all M9391A channels revert to typical performance. – All graphs contain measured data from one unit and is repre - sentative of product performance within the controlled temperature range unless otherwise noted. – The specifications contained in this document are subject to change. Technical Specifications and Characteristics
100 MHz PLL
10 MHz OCXO
100 MHz
10 MHz Out10 MHz
100 MHz In
850 MHz
3 GHz to 6 GHz
2.1 GHz
3.9 GHz
1 MHz to 6 GHz
300 MHz
Figure 3. M9391A PXIe vector signal analyzer block diagram with four modules consisting of the M9301A synthesizer, M9350A downconverter, M9214A digitizer and optional M9300A frequency reference.
6 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Frequency Frequency range and resolution Option F03 1 MHz to 3 GHz Option F06 1 MHz to 6 GHz Tuning resolution 0.001 Hz IF frequency Nominal
15 MHz filter 326 MHz
40 MHz filter 240 MHz
160 MHz filter 300 MHz
Maximum bandwidth Option B04 40 MHz Option B10 100 MHz Option B16 160 MHz Frequency switching speed 2,3 List mode switching speed 4 Sample rate Acquisition bandwidth Standard, nominal Option UNZ, nominal Baseband frequency offset change 5 ≤ 100 MHz ≤ 80 MHz 5 ms 27 µs > 100 MHz to < 180 MHz > 80 MHz to < 144 MHz 5 ms 102 µs ≥ 180 MHz ≥ 144 MHz 5 ms 15 µs Arbitrary frequency change 5 ms 320 µs Non-list mode switching speed 6 Standard, nominal Option UNZ, nominal Baseband frequency offset change 5 5 ms 310 µs Arbitrary frequency change 5 ms 2.3 ms 1. Instantaneous bandwidth (1 dB bandwidth) available around a center frequency over which the input signal can be digitized for further analysis or processing in the time, frequency or modulation domain. 2. When used with the M9018A PXIe chassis (2-link configuration: 1 x 8 [factory default]) and M9036A PXIe embedded controller. 3. Settled to within 1 kHz or 1 ppm, whichever is greater of final value. Does not include data acquisition or processing time. Amplitude settled to within 0.1 dB. Channel filter set to none. Applies for all conversion types. Minimum power acquisition channel filter bandwidth ≥ 4.8 MHz. For lists with first point < 400 MHz or for frequency changes from > 400 MHz to < 400 MHz, add 40 ms. 5. Baseband offset can be adjusted ± from carrier frequency within limits determined by RF analysis bandwidth and IF filter bandwidth. Synthesizer frequency and amplitude are not changing. Baseband offset settled to within 1 kHz. 6. Mean time from IVI command to carrier frequency settled to within 1 kHz or 1 ppm, whichever is greater. Amplitude settled within 0.1 dB. Simultaneous carrier frequency and amplitude switching. For frequency changes from > 400 MHz to < 400 MHz, add 40 ms.
7 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Frequency (continued) Frequency reference (M9300A PXIe frequency reference module) Reference outputs
100 MHz Out (Out 1 through Out 5)
Amplitude ≥ 10 dBm 13 dBm, typical Connectors 5 SMB snap-on Impedance 50 Ω, nominal
10 MHz Out
Amplitude 9.5 dBm, nominal Connectors 1 SMB snap-on Impedance 50 Ω, nominal OCXO Out Amplitude 11.5 dBm, nominal Connectors 1 SMB snap-on Impedance 50 Ω, nominal Frequency accuracy Same as accuracy of internal time base or external reference input Internal timebase Accuracy ± [(time since last adjustment x aging rate) ± temperature effects ± calibration accuracy] Frequency stability Aging rate Daily < ±0.5 ppb/day, after 72 hours of warm-up Yearly < ±0.1 ppm/year, after 72 hours of warm-up Total 10 years < ±0.6 ppm/10yrs, after 72 hours of warm-up Achievable initial calibration accuracy (at time of shipment) ±5 x 10-8 Temperature effects 20 to 30 °C < ±10 ppb Full temperature range < ±50 ppb Warm up 5 minutes over +20 to +30 °C, with respect to 1 hour < ±0.1 ppm 15 minutes over +20 to +30 °C, with respect to 1 hour < ±0.01 ppm External reference input Frequency 1 to 110 MHz, sine wave Lock range ±1 ppm, nominal Amplitude 0 to 10 dBm, nominal Connector 1 SMB snap-on Impedance 50 Ω, nominal
8 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Amplitude Input level Max safe average total power +30 dBm (1 W) Max DC voltage 25 Vdc Max RF input (specified performance) 1 to 2 MHz 0 dBm 2 to 4 MHz +4 dBm 4 to 100 MHz +12 dBm
100 MHz to 6 GHz +30 dBm
Expected input level setting Range Pre-amp ON −170 to 0 dBm Pre-amp OFF −170 to +30 dBm Pre-amp AUTO 7 −170 to +30 dBm Resolution 0.1 dB Absolute amplitude accuracy & total absolute amplitude accuracy Full temperature range Controlled temperature range @ 46 ºC module temp10, typical Conversion type Frequency Total absolute amplitude accuracy 8 Absolute amplitude accuracy 9 Total absolute amplitude accuracy 8 Absolute amplitude accuracy 9 Total absolute amplitude accuracy 8
40 MHz IF filter Module temperature within ± 3 ºC of alignment, pre-amp ON & OFF
160 MHz IF filter Module temperature within ±3 ºC of alignment, Pre-amp OFF 11
160 MHz IF filter Module temperature within ±3 ºC of alignment, Pre-amp ON 12
- At expected input level ≤ −37 dBm, pre-amp is switched on. 8. Total absolute amplitude accuracy is the total of all amplitude measurement errors. This specification includes the sum of the following individual specifica - tions: linearity, expected input level switching uncertainty, IF bandwidth filter switching uncertainty, absolute amplitude accuracy. The wide range of settings confidence are computed for each parameter on a statistically significant number of instruments. The root-sum-square (RSS) of these four independent Gaussian parameters is then taken. To that RSS value, two environmental effects and measurement uncertainty are added. One environmental effect is that of temperature (full and controlled temperature range, as defined above) and the other is the temperature variation of ±3 degrees around a field alignment. Applies over the following subset of settings and conditions: expected input level −50 dBm to +30 dBm; input signals within 60 dB below expected input level; 40 MHz and 160 MHz IF filters; input signal at center frequency over full frequency range. 9. The absolute amplitude accuracy is the amplitude measurement error when only changing frequency. The expected input level, conversion type and IF bandwidth settings remain the same and the error introduced by those parameters are not included. Pre-amp auto/OFF expected input level +10 dBm and −12 dBm. Pre-amp ON expected input level −30 dBm. 10. Typical specifications shown at M9350A downconverter reported module temperature of 46 ºC and a corresponding environment temperature of 25 ºC. 11. When using pre-amp auto mode, applies for signal level within expected input level >−37 dBm. 12. When using pre-amp auto mode, applies for signal level within expected input level ≤−37 dBm.
9 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Amplitude (continued) Amplitude repeatability and linearity Input signal relative to expected input level setting Specification Repeatability <0.05 dB, nominal Linearity 13 >−35 dB ±0.12 dB ±0.03 dB, nominal ≤−35 dB ±0.21 dB ±0.04 dB, nominal IF flatness 14, 15 Analysis bandwidth IF filter Nominal 40 MHz 40 MHz ± 0.08 dB 100 MHz 160 MHz ± 0.09 dB 160 MHz 160 MHz ± 0.10 dB IF phase linearity 15 Analysis bandwidth Conversion type Peak to peak, nominal 40 MHz All 1.0 º 100 MHz Single 0.8 º Image protect 1.7 º 160 MHz Single 1.4 º Image protect 1.8 º 13. Input level 20 dB above the noise floor and dither on, no change in hardware settings, below expected input level. 14. Amplitude deviation from the mean error of the entire bandwidth, all conversion types. 15. Expected input level 0 dBm. Center frequency ≥ 250 MHz.
10 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Amplitude (continued) IF bandwidth filter switching uncertainty 16 Specification Typical Nominal ±0.4 dB ±0.15 dB ±0.09 dB Expected input level switching uncertainty Specification Typical Nominal Pre-amp Auto/OFF Max input to +5 dBm ±0.45 dB ±0.14 dB ±0.10 dB Crossing +5 dBm ±0.63 dB ±0.24 dB ±0.17 dB Pre-amp OFF +5 to −50 dBm ±0.41 dB ±0.16 dB ±0.11 dB Pre-amp ON +0 to −50 dBm ±0.64 dB ±0.27 dB ±0.21 dB Pre-amp AUTO Crossing −37 dBm ±0.95 dB ±0.19 dB ±0.12 dB Amplitude switching speed Arbitrary amplitude change Standard, nominal Option UNZ, nominal List mode switching speed 17 ≤ 5 ms ≤ 136 µs Non-list mode switching speed 18 ≤ 5 ms ≤ 1.5 ms Input voltage standing wave ratio (VSWR) Nominal < 10 MHz 1.7:1 10 MHz to 2.5 GHz 1.4:1 > 2.5 GHz 1.7:1 16. Amplitude error relative to the reference IF bandwidth filter of 40 MHz. 17. Settled to within 0.1 dB of final value. Does not include data acquisition or processing time. When used with the M9018A PXIe chassis (2-link configuration: 1 x 8 [factory default]) and the M9036A PXIe embedded controller. 18. Mean time from IVI command to amplitude settled.
11 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Dynamic range Displayed average noise level (DANL) 19 Conversion type Frequency Specification Nominal Pre−amp OFF Image protect < 100 MHz −145 dBm/Hz 100 to < 700 MHz −137 dBm/Hz −147 dBm/Hz 700 MHz to < 5.75 GHz −140 dBm/Hz −148 dBm/Hz 5.75 to 6 GHz −129 dBm/Hz −146 dBm/Hz Single <1.2 GHz −148 dBm/Hz −154 dBm/Hz 1.2 to 3.1 GHz −143 dBm/Hz −152 dBm/Hz > 3.1 to < 5.4 GHz −138 dBm/Hz −149 dBm/Hz 5.4 to 6 GHz −133 dBm/Hz −148 dBm/Hz Pre−amp ON Image protect < 100 MHz −162 dBm/Hz 100 MHz to < 2.7 GHz −156 dBm/Hz −161 dBm/Hz 2.7 to 4.4 GHz −155 dBm/Hz −160 dBm/Hz > 4.4 to < 5.6 GHz −152 dBm/Hz −157 dBm/Hz 5.6 to 6 GHz −141 dBm/Hz −154 dBm/Hz Single <1.1 GHz −157 dBm/Hz −161 dBm/Hz 1.1 to < 3.6 GHz −154 dBm/Hz −158 dBm/Hz 3.6 to 5 GHz −151 dBm/Hz −156 dBm/Hz > 5 to 6 GHz −146 dBm/Hz −153 dBm/Hz Third order intermodulation distortion (TOI) 20 TOI23 Distortion 24 Conversion type: auto Frequency Specification Typical Specification Pre−amp OFF 21 ≤ 400 MHz +15 dBm +20.5 dBm −52 dBc > 400 MHz to 3 GHz +18 dBm +23 dBm −52 dBc > 3 GHz +20 dBm +23.5 dBm −52 dBc Pre−amp ON 22 ≤ 100 MHz −9.9 dBm −2.5 dBm −56 dBc > 100 to 850 MHz −7.9 dBm +2 dBm −58 dBc > 850 MHz to 2 GHz −4.3 dBm +5 dBm −47 dBc > 2 to 3 GHz −0.9 dBm +7 dBm −41 dBc > 3 to 6 GHz +1 dBm +5 dBm −32 dBc 19. Expected input level of −50 dBm. Mixer level offset +10 dB. 20. Two tone, 100 kHz tone spacing. 21. Expected input level −5 dBm. Mixer level offset +10 dB. 22. Expected input level −25 dBm. Mixer level offset +15 dB. 23. TOI = third order intercept. The TOI is given by the input tone level (in dBm) minus (distortion/2) where distortion is the relative level of the distortion tones in dBc. 24. Expected input level -10 dBm with preamp off and -30 dBm with preamp on.
image protect conversion type. Figure 7. Dynamic range at 5.8 GHz, pre-amp OFF, image protect conversion type.
15 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Spectral purity (continued) Residuals, images & spurious responses Non-input related spurs 29 Conversion type Frequency Nominal Expected input level Pre-amp ON ≤ 0 dBm (measured at −50 dBm) Single All < −120 dBm Image protect All 30 < −120 dBm Pre-amp OFF < +5 dBm (measured at −50 dBm) Single ≤ 3 GHz < −120 dBm > 3 GHz < −116 dBm Image protect All 31 < −105 dBm ≥ +5 dBm (measured at +6 dBm) Single All < −98 dBm Image protect All 32 < −90 dBm LO related spurs 33 Offsets from carrier Frequency Nominal 200 to 10 kHz All − 82 dBc 10 kHz to 10 MHz All − 55 dBc First order RF spurious responses 34 Offsets from carrier Frequency Nominal ≥ 10 MHz ≥ 200 MHz to 6 GHz −60 dBc Higher order RF spurious responses 34 Offsets from carrier Frequency Nominal ≥ 10 MHz ≥ 200 MHz to 6 GHz −60 dBc Image responses 35 Conversion type Frequency Nominal Image protect All < −68 dBc IF rejection 36 IF bandwidth filter Frequency Nominal
15 MHz ≤ 400 MHz < −57 dBc
400 MHz < −105 dBc
40 MHz ≤ 450 MHz < −57 dBc
450 MHz < −98 dBc
160 MHz All < −85 dBc
LO emission 37 Conversion type Frequency Nominal Single ≤ 3 GHz −72 dBm > 3 GHz −62 dBm Image protect All −88 dBm 29. Mixer level offset at 10 dB, input terminated, with 50Ω load. 30. From 4.72 to 4.88 GHz, specification at <-108 dBm, nominal. 31. From 4.72 to 4.88 GHz, specification at <-96 dBm, nominal. 32. From 4.72 to 4.88 GHz, specification at <-80 dBm, nominal. 33. Expected input level 0 dBm. Mixer offset level -10 dB. 34. Conversion type: image protect, pre-amp OFF, expected input level -20 dBm and mixer level offset 0 dB. 35. Excitation frequency: [F=2*Final IF] MHz, expected input level -20 dBm, mixer level offset -30 dB. 36. Suppression of signal at IF frequencies when tuned at least 2 x IF BW away. All input paths, image protect, expected input level -30 dBm. Input signal at -30 dBm and mixer level offset 0 dB. 37. Expected input level -50 dBm. Mixer level offset +10 dB.
16 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Data acquisition Maximum capture memory Non-list mode List mode Option M01 128 MSample (512 MB) 128 MSample (512 MB) Option M05 512 MSample (2 GB) 512 MSample (2 GB) Option M10 1 GSample (4 GB) 38 512 MSample (2 GB) to ~ 1 GSample (3.999 GB) 39 Segments Minimum length 1 sample 40 Maximum length Full capture memory 38 Maximum sample rate Option B04 / 40 MHz 50 MS/s complex, 100 MS/s real Option B10 / 100 MHz 125 MS/s complex, 250 MS/s real Option B16 / 160 MHz 200 MS/s complex, 400 MS/s real List mode Maximum number of segments 3201 Trigger sources External, magnitude Trigger modes Per acquisition, interval timer trigger Triggering Delay range 41 −500 ms to +500 ms, nominal Delay resolution 1 sample, nominal External trigger signal frequency range 10 to 30 MHz for pulse External trigger signal level TTL External trigger signal duty cycle range 20% to 80% External trigger signal waveform Sine, pulse/square, ramp (symmetry 0% to 100%) Channel-to-channel synchronization 42 Timing Phase Skew ≤400 ps, nominal – Jitter 43 ≤50 ps, nominal ≤0.3°, nominal Repeatability 44 ≤80 ps, nominal ≤1.0°, nominal Adjustment resolution 45 50 ps 0.05° Drift over 12 hours 20 ps, nominal 0.5°, nominal 38. The default mode for allocation of capture memory is AgM9391MemoryModeNormal, where the digitizer’s memory is shared by both the default single acquisition (capture ID = 0) and all the other acquisitions with non-zero capture IDs. In particular, the memory for the default single acquisition is allocated from the area unused by the list acquisitions. If the available memory is not sufficient for the single acquisition, the user must release memory allocated for the non-zero capture ID acquisitions manually, thus increasing free space. Total memory usage is limited according to the memory option. Note that the maximum size of acquisition is 2 GB in this mode. To perform the default single acquisition with memory size larger than 2 GB, AgM9391MemoryModeLargeAcquisition must be selected. The non-zero capture ID acquisitions cannot be per formed in this mode. All data acquired with AGM9391MemoryMode Normal will be invalidated. 39. The maximum size for a single list point capture is limited to 512 MSamples (2 GB). However, with option M10, total capture of up to 3.999 GB is available across all list mode captures. 40. 64-bit mode, 2 samples for 32-bit mode. 41. Negative trigger delay limited to capture size. 42. Multi-channel capability only supported with up to 8-channels when configured with a Keysight M9018A PXIe chassis with FPGA version 1.05 or greater. Characteristics measured at 400, 900, 2400, 5800 MHz and apply in Auto Conversion mode at frequencies ≥400 MHz with IF filter = 160 MHz. V2802A LO distribution network used for phase synchronization for more than 4 channels. 43. Jitter indicates measurement-to-measurement variation and applies over short time interval at room temperature without resetting or reinitializing a driver session. 44. Repeatability indicates stability of alignment between channels across power cycles and IVI sessions, with identical cabling and hardware settings (fre - quency, span, sample rate, etc.) 45. Channel time and phase offsets can be adjusted using OffsetDelay and OffsetPhase properties respectively.
17 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Measurement speed46 IQ data capture 47 Nominal Large block (50 MSamples) 1.5 s Transferred in 100 kSa or 1 MSa blocks Small block (100 captures, 100 ksamples each) 292 ms Transferred in 10 kSa blocks Adjust level, freq (10 ksamples) 1.7 ms Transferred in 10 kSa blocks Power measurements 48 Channel power settings & filter bandwidth Acquisition Time Averages Nominal 3.84 MHz 400 µs None 1.8 ms 10 7.6 ms 100 µs None 1.3 ms 10 4.1 ms 50 µs None 1.3 ms 10 3.4 ms 30 kHz 100 µs None 3.9 ms 10 30.4 ms Format specific measurement data GSM 49, 50 Parameters Nominal ORFS dynamic range 200 kHz offset −36 dBc 250 kHz offset −41 dBc 400 kHz offset −69 dBc 600 kHz offset −73 dBc 800 kHz offset −77 dBc 1200 kHz offset −80 dBc 1800 kHz offset −78 dBc EDGE 49, 50 Parameters Nominal ORFS dynamic range 200 kHz offset −37 dBc 250 kHz offset −42 dBc 400 kHz offset −69 dBc 600 kHz offset −73 dBc 800 kHz offset −77 dBc 1200 kHz offset −80 dBc 1800 kHz offset −77 dBc 46. EVM, ACPR and servo loop test times for the RF power amplifier test, reference solution are included in the solution brochure 5991-4104EN. 47. Capture block, transfer to host memory, 160 MHz BW, excludes frequency transitions below 400 MHz, with M9037A embedded controller (2-link configuration: 1 x 8 [factory default]). 48. Transfer to host memory, 160 MHz IF bandwidth filter, excludes frequency transitions below 400 MHz, with M9037A embedded controller (2-link configuration: 1 x 8 [factory default]). 49. Synthesizer PLL mode set to PLL mode best wide offset. 50. Expected input level 0 dBm, input signal (total power) 0 dBm, mixer level offset +10 dB, conversion type: Auto, PeakToAverage set per signal peak to average.
18 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Format specific measurement (continued) W−CDMA 51, 52 Parameters Typical Nominal Residual EVM 2 GHz, 1 DPCH, 1 carrier 0.5% ACLR dynamic range 2 GHz, 1 DPCH, 1 carrier (power mode) Adjacent −68.1 dBc −69.8 dBc Alternate −70.7 dBc −71.7 dBc 802.11g 51, 52, 56 Parameters Nominal EVM 2.4 GHz, 20 MHz BW −52.8 dB 802.11a 51, 52, 56 Parameters Nominal EVM 5.8 GHz, 20 MHz BW −48.1 dB 802.11n 51, 52, 56 Parameters Nominal 1-channel 2-channel 54 3-channel 54 4-channel 54 802.11ac 51, 52 Parameters Nominal 1-channel 2-channel 54 3-channel 54 4-channel 54 8-channel 54 Preamble only Preamble, pilots & data SEM 5.8 GHz, 80 MHz BW see Figure 10 802.11a/g 54, 52 Parameters SEM 2.4 GHz see Figure 11
5.5 GHz see Figure 12
802.11e 54, 52, 57 Parameters OFDMA WiMAX™ EVM 2.5, 3.5, & 5.8 GHz −48.3 dB, nominal 51. Synthesizer PLL mode set to PLL mode best wide offset. 52. Expected input level 0 dBm, input signal (total power) 0 dBm, conversion type: Auto. PeakToAverage set per signal peak to average. 53. Synthesizer PLL mode set to PLL mode normal. 54. Multi-channel performance data applies when each channel is configured with its own independent synthesizer. Sharing a single synthesizer will degrade EVM performance approximately 1 dB. 55. Mixer level offset = +5 dB 56. Mixer level offset = +10 dB 57. Mixer level offset = +15 dB
20 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Format specific measurement (continued) LTE FDD - single channel 58, 59 Parameters 1-channel, nominal
10 MHz BW EVM,
E−TM 3.1 61, 62
10 MHz BW ACLR,
E−TM 1.1 63 (power mode) Adjacent −64.2 dBc Alternate −65.5 dBc LTE FDD - MIMO 58, 59, 60 Parameters 2-channel, nominal 64 4-channel, nominal 64 8-channel, nominal 65 LTE TDD - MIMO 58, 59, 60 Parameters 2-channel, nominal 64 4-channel, nominal 64 8-channel, nominal 65 58. Expected input level 0 dBm, input signal (total power) 0 dBm, conversion type: Auto. PeakToAverage set per signal peak to average. 59. Synthesizer PLL mode set to PLL mode normal. 60. Multi-channel performance data applies when each channel is configured with its own independent synthesizer. Sharing a single synthesizer will degrade EVM performance approximately 1 dB. 61. PDCCH power boost = 1.065 dB 62. Mixer level offset = +10 dB 63. Mixer level offset = +15 dB 64. 10 MHz BW EVM, R9 downlink, 64 QAM, open loop spatial multiplexing 65. 10 MHz BW, DL, TM9 multi-layer, TM4 closed loop spatial multiplexing
21 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics Environmental and physical specifications Temperature Operating Non-operating (storage) Individual module temp 25 to 75 °C as reported by the module and environment temp of 0 to 55 °C Environment temp of −40 to +70 °C Humidity 66 Type tested at 95%, +40 °C (non-condensing) Shock/vibration 66 Operating random vibration Survival random vibration Functional shock Bench handling Type tested at 5 to 500 Hz, 0.21 g rms Type tested at 5 to 500 Hz, 2.09 g rms Type tested at half-sine, 30 g, 11 ms Type tested per MIL-PRF-28800F Altitude Up to 15,000 feet (4,572 meters) 67 Connectors RF In SMA female EMC Complies with European EMC Directive 2004/108/EC – IEC/EN 61326-2-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11 – ICES/NMB-001 This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada. Warm-up time 45 minutes Size M9300A M9301A M9350A M9214A
1 PXIe slot
Dimensions Module Length Width Height M9300A 210 mm 22 mm 130 mm M9301A 210 mm 22 mm 130 mm M9350A 210 mm 22 mm 130 mm M9214A 210 mm 22 mm 130 mm Weight M9300A M9301A M9350A M9214A 0.55 kg (1.21 lbs) 0.54 kg (1.19 lbs) 0.56 kg (1.23 lbs) 0.36 kg (0.79 lbs) Power drawn from chassis M9300A M9301A M9350A M9214A ≤ 18 W ≤ 25 W ≤ 30 W ≤ 35 W 66. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage, transportation and end-use--those stresses include but are not limited to temperature, humidity, shock, vibration, altitude and power-line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3. 67. At 15,000 feet, the maximum environmental temperature is de-rated to 52 °C.
22 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Technical Specifications and Characteristics System requirements Topic Windows 7 requirements Operating systems Windows 7 (32-bit and 64-bit) Processor speed 1 GHz 32-bit (x86), 1 GHz 64-bit (x64) (no support for Itanium 64) Available memory 4 GB minimum
8 GB or greater recommended
1.5 GB available hard disk space, includes:
1 GB available for Microsoft .NET Framework 3.5 SP1 69
100 MB for Keysight IO Libraries Suite
Video Support for DirectX 9 graphics with 128 MB graph - ics memory recommended (Super VGA graphics is supported) Browser Microsoft Internet Explorer 7 or greater M9391A vector signal analyzer instrument drivers Keysight IO libraries Version 16.3.17914 or greater 68. Because of the installation procedure, less disk space may be required for operation than is required for installation. 69. NET Framework Runtime Components are installed by default with Windows 7. Therefore, you may not need this amount of available disk space.
23 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Software Instrument connection software Keysight IO library The IO library suite offers a single entry point for connection to the most common instruments including AXIe, PXI, GPIB, USB, Ethernet/LAN, RS-232, and VXI test instruments from Keysight and other vendors. It automatically discovers interfaces, chas - sis, and instruments. The graphical user interface allows you to search for, verify, and update IVI instrument and soft front panel drivers for modular and traditional instruments. The IO suite safely installs in side-by-side mode with NI I/O software. Free software download at www.keysight.com/find/iosuite Module setup and usage Keysight soft front panel The PXI module includes a soft front panel (SFP), a software- based graphical user interface (GUI) which enables the instru - ment’s capabilities from your PC. Included on CD-ROM shipped with module or online Module management Keysight connection expert Connection expert is the graphical user interface included in the IO libraries suite that allows you to search for, verify and update IVI instrument and soft front panel drivers for modular and traditional instruments Free software download at www.keysight.com/find/iosuite Programming Driver Development environments lVI-COM, IVI-C LabVIEW, MATLAB Visual Studio (VB.NET, C#, C/C++), VEE LabVIEW, LabWindows/CVI, MATLAB Included on CD-ROM shipped with module. Programming assitance Command expert Assists in finding the right instrument commands and setting correct parameters. A simple interface includes documentation, examples, syntax checking, command execution, and debug tools to build sequences for integration in Excel, MATLAB, Visual Studio, LabVIEW, VEE, and SystemVue. Free software download at www.keysight.com/find/commandexpert Programming examples Each module includes programming examples for Visual Studio. net, LabVIEW, MATLAB, LabWindows, and Keysight VEE Pro. Included on CD-ROM shipped with module. Signal analysis software X-Series measurement The X-Series measurement applications transform modular PXI VSAs into standards based RF transmitter testers. Provides conformance measurements for many communications standards including : LTE, WLAN 802.11ac and others. Licensed software. For more information, visit www.keysight.com/find/pxi-x-series_apps
89600 VSA 89600 VSA software sees through the complexity of emerging
and existing industry standards, serving as your window into com - plex signal interactions. Quickly characterize spurs and harmonics with speed-optimized stepped spectrum measurement provided by 89601B-SSA option. Licensed software. For more information, visit www.keysight.com/find/vsa SystemVue SystemVue is a system-level EDA platform for designing com - munications and defense systems. Used with the M9391A, SystemVue enables you to create model-based design validation tests to ensure consistency from design to manufacturing. Licensed software. For more information, visit www.keysight.com/find/systemvue
24 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Setup and Calibration Services Assistance One day startup assistance Gain access to a technical expert who will help you get started quickly with the M9391A PXI VSA and its powerful software tools. The flexible instruction format is designed to get you to your first measurements and familiarize you with ways to adapt the equipment to a specific application. Included in base configuration Calibration and traceability Factory calibration The M9391A PXI VSA ships factory calibrated with an ISO-9002, NIST-traceable calibration certificate. Included in base configuration Calibration cycle A one year calibration cycle is recommended. Calibration sites – At Keysight worldwide service xenters – On-site by Keysight – By self-maintainers For more information visit www.keysight.com/find/infoline N7800A calibration and adjustment software The M9391A PXI VSA is supported by Keysight’s calibration and adjust - ment software. This is the same software used at Keysight service centers to automate calibration. The software offers compliance tests for ISO 17025:2005, ANSI/NCSL Z540.3-2006, and measurement uncertainty per ISO Guide to Expression of Measurement Uncertainty. Licensed software. For more information, visit www.keysight.com/find/ calibrationsoftware Keysight calibration status utility The Keysight calibration status utility helps ensure your M9391A is calibrated by managing the calibration interval and providing messages regarding instrument and module calibration status. Included in base configuration
25 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Support and Warranty Warranty Global warranty Keysight’s warranty service provides standard coverage for the country where product is used. – All parts and labor necessary to return to full specified performance – Recalibration for products supplied originally with a calibration certificate – Return shipment Included Standard Return to Keysight warranty—3 years 15 days typical turnaround repair service Included R-51B-001-5Z Return to Keysight warranty—5 years 15 days typical turnaround repair service Optional R-51B-001-3X Express warranty 3 years The express warranty upgrades the global warranty to provide, for 3 years, a 5 day typical turnaround repair service in the US, Japan, China and many EU countries. Optional R-51B-001-5X Express warranty 5 years The express warranty upgrades the global warranty to provide, for 5 years, a 5 day typical turnaround repair service in the US, Japan, China and many EU countries. Optional Support Core exchange program Keysight’s replacement core exchange program allows fast and easy module repairs. A replacement core assembly is a fully functioning pre- calibrated module replacement that is updated with the defective module serial number, allowing the replacement module to retain the original serial number. For qualified self-maintainers in US only Self-test utility A self-test utility runs a set of internal tests which verifies the health of the modules and reports their status. Included in base configuration
26 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Configuration and Ordering Information
Ordering information
M9391A PXIe vector signal analyzer:
1 MHz to 3 or 6 GHz
Includes: M9301A PXIe synthesizer M9350A PXIe downconverter M9214A PXIe IF digitizer One day startup assistance Module interconnect cables Software, example programs and product information on CD Return to Keysight warranty—3 Years Base configuration M9391A-F03 Frequency range: 1 MHz to 3 GHz M9391A-B04 Analysis bandwidth, 40 MHz M9391A-M01 Memory, 128 MSa M9391A-300 Required for warranted specifications PXIe frequency reference: 10 and 100 MHz Adds M9300A PXIe frequency reference: 10 and 100 MHz (M9300A module can sup - port multiple M9391A modular instruments) Configurable options Frequency M9391A-F03 1 MHz to 3 GHz M9391A-F06 1 MHz to 6 GHz Switching speed M9391A-UNZ Fast switching Analysis bandwidth M9391A-B04 40 MHz M9391A-B10 100 MHz M9391A-B16 160 MHz Memory M9391A-M01 128 MSa M9391A-M05 512 MSa M9391A-M10 1024 MSa Other M9391A-012 Phase coherency M9391A-UK6 Commercial calibration certificate with test data for M9391A (M9301A, M9350A, M9214A) M9300A-UK6 Commercial calibration certificate with test data for M9300A (module only) Related products in recommended configuration M9037A PXIe embedded controller M9018A 18-slot PXIe chassis Indicates recommended configuration For configurations of the M9391A PXI VSA, including com - binations with a single or multiple M9381A PXI VSGs, please consult the M9391A & M9381A configuration guide, literature number 5991-0897EN.
27 | Keysight | M9391A PXIe Vector Signal Analyzer - Data Sheet Software information Supported operating systems Microsoft Windows 7 (32/64-bit) Standard compliant drivers IVI-COM, IVI-C, LabVIEW, MATLAB Supported application development environ - ments (ADE) VisualStudio (VB.NET, C#, C/C++), VEE, LabVIEW, LabWindows/CVI, MATLAB Keysight IO libraries (version 16.3 or newer) Includes: VISA libraries, Keysight Connection Expert, IO monitor Keysight Command Expert Instrument control for SCPI or IVI-COM drivers
89600 VSA Software
(version 17.21 or newer; Option SSA added in ver - sion 18.5) 89600B-200 Basic VSA software 89601B-300 Hardware connectivity 89601B-SSA Spectrum analysis 89601B-AYA GP analysis 89601B-B7T cdma2000®/1xEV-DO 89601B-B7U W-CDMA/HSPA+ 89601B-B7R WLAN 802.11a/b/g/j/p 89601B-B7X TD-SCDMA 89601B-BHD LTE FDD 89601B-BHG LTE FDD - Advanced 89601B-BHE LTE TDD 89601B-BHH LTE TDD - Advanced X-Series Measurement Applications for Modular Instruments transportable perpetual license. M9063A Analog demodulation M9064A Vector signal analysis M9071A GSM/EDGE/Evo M9072A cdma2000 ®/cdmaOne M9073A W-CDMA/HSPA+ M9076A 1xEV-DO M9077A WLAN 802.11a/b/g/n/ac M9079A TD-SCDMA/HSDPA M9080B LTE/LTE-A FDD M9081A Bluetooth® M9082B LTE/LTE-A TDD Accessories Model Description Y1212A Slot blocker kit: 5 modules Y1213A PXI EMC filler panel kit: 5 slots Y1299A PXI solutions startup kits Y1243A Cable kit for M9301A LO distribution M9021A PCIe® cable interface M9045B PCIe express card adaptor for laptop connectivity Y1200B PCIe cable for laptop connectivity M9048A PCIe desktop adaptor for desktop connectivity Y1202A PCIe cable for desktop connectivity
Related products
M9381A PXIe vector signal generator M9380A PXIe CW source M9300A PXIe frequency reference M9018A PXIe 18-slot chassis M9037A PXIe embedded controller Advantage services: Calibration and warranty Keysight Advantage Services is committed to your success throughout your equipment’s lifetime R-51B-001-5Z Return to Keysight warranty - 5 years R-51B-001-3X Express warranty - 3 years R-51B-001-5X Express warranty - 5 years N7800A Calibration & adjustment software Configuration and Ordering Information
28 | Keysight | M9391A PXIe Vector Signal Analyzer 1 MHz to 3 GHz or 6 GHz - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2013-2015 Published in USA, July 10, 2015 5991-2603EN www.keysight.com myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.pxisa.org PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. Bluetooth and the Bluetooth logos are trademarks owned by Bluetooth SIG, Inc., U.S.A. and licensed to Keysight Technologies, Inc. “PCIe” and “PCI EXPRESS” are registered trademarks and/or service marks of PCI-SIG. cdma2000 is a registered certification mark of the Telecommunications Industry Association. www.keysight.com/find/m9391a www.keysight.com/find/modular For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-02-20-15)