M9380A KEYSIGHT | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 17
Technical content
1 MHz to 3 GHz or 6 GHz
With high-power levels and accurate measurements, the M9380A CW source provides the Keysight Technologies, Inc. quality and performance in the PXI form factor—a trusted Keysight product with global services and support, fast repair and a wide scope of calibration utilities. Product description The M9380A PXIe CW source is a compact modular solution that provides frequency coverage from 1 MHz to 3 or 6 GHz. A typical M9380A configuration includes three individual PXIe modules—M9310A source output, M9301A synthesizer and the M9300A frequency reference— designed for fast data interfaces and high-speed automated test systems. Instrument control is provided through a soft front panel and programmatic interfaces tuned to your application development environment of choice. Product features – Frequency coverage from 1 MHz to 3 or 6 GHz – Output power of +18 dBm across the frequency range – Output power level of +19 dBm from 1 MHz to 5 GHz – Better than ±0.4 dB absolute amplitude accuracy – License key upgraded frequency range and output power – One day startup assistance – Return to Keysight warranty—3 years – Chassis slot compatibility: PXIe slot Uncompromising values – Keeps costs manageable—purchase what you need today and easily upgrade later using license-key upgrades without returning your modules to Keysight. – Reduces development time and simplifies integration into existing test environments with multiple drivers and programmatic interfaces. – Reduces startup time with Keysight IO libraries easy configuration, one-step software install, and integrated instrument level CW source soft front panel. – Fast repair turnaround time with calibrated core exchange strategy.
Applications
– Aerospace and defense – Interference injection – LO substitution – Wireless component test Figure 1. M9380A PXIe CW source with three
Figure 2. M9380A PXIe CW source with three modules consisting of the M9300A PXIe frequency reference, M9301A PXIe synthesizer and M9310A PXIe source output.
100 MHz PLL
10 MHz OCXO
100 MHz
10 MHz Out10 MHz
100 MHz In
850 MHz
3 GHz to 6 GHz
6 GHz
100 MHz Out
04 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Technical Specifications and Characteristics Definitions for specifications Temperatures referred to in this document are defined as follows: – Full temperature range = Individual module temperature of ≤ 75 °C, as reported by the module, and environment temperature of 0 to 55 °C. – Controlled temperature range = Individual module temperature of ≤ 55 °C, as reported by the module, and environment temperature of 20 to 30 °C. Specifications describe the warranted performance of calibrated instruments. Data represented in this document are specifications unless otherwise noted under the following conditions. – Calibrated instruments have been stored for a minimum of 2 hours within the full temperature range – 45 minute warm-up time – Calibration cycle maintained – When used with Keysight M9300A frequency reference – When used with Keysight interconnection cables Characteristics describe product performance that is useful in the application of the product, but that is not covered by the product warranty. Characteristics are often referred to as Typical or Nominal values and are italicized. – Typical describes characteristic performance, which 80% of instruments will meet when operated within the controlled temperature range. – Nominal describes representative performance that is useful in the application of the product when operated within the controlled temperature range. Recommended best practices in use – Use slot blockers and EMC filler panels in empty module slots to ensure proper operating temperatures. – Keysight chassis and slot blockers optimize module temperature performance and reliability of test. – At environment temperatures above 45 °C, chassis fan should be set to high. Additional information – All graphs contain measured data from one unit and is representative of product performance at the controlled temperature range unless otherwise noted. – The specifications contained in this document are subject to change.
05 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Frequency range Option F03 1 MHz to 3 GHz Option F06 1 MHz to 6 GHz Resolution 0.01 Hz Frequency switching speed 1 ≤ 5 ms, nominal Frequency reference (M9300A PXIe frequency reference module) Reference outputs
100 MHz Out (Out 1 through Out 5)
Amplitude ≥ 10 dBm 13 dBm, typical Connectors 5 SMB snap-on Impedance 50 Ω, nominal
10 MHz Out
Amplitude 9.5 dBm, nominal Connectors 1 SMB snap-on Impedance 50 Ω, nominal OCXO Out Amplitude 11.5 dBm, nominal Connectors 1 SMB snap-on Impedance 50 Ω, nominal 1. Mean time from IVI command to carrier frequency settled within 1 ppm or 1 kHz whichever is greater and amplitude settled within 0.2 dB (at the controlled temperature range) or within 0.5 dB (at the full temperature range). If the ALC is off, the settle limit is 0.5 dB above +10 dBm, (at the controlled temperature range). Simulataneous carrier frequency and amplitude switching. Technical Specifications and Characteristics (continued) Frequency
06 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Technical Specifications and Characteristics (continued) Frequency (continued) Frequency accuracy Same as accuracy of internal time base or external reference input Internal timebase Accuracy ± (time since last adjustment x aging rate) ± temperature effects ± calibration accuracy Frequency stability Aging rate Daily < ±0.5 ppb/day, after 72 hour warm-up Yearly < ±0.1 ppm/year, after 72 hours warm-up Total 10 years < ±0.6 ppm/10yrs, after 72 hours warm-up Achievable initial calibration accuracy (at time of shipment) ±5 x 10 -8 Temperature effects 20 to 30 °C < ±10 ppb Full temperature range < ±50 ppb Warm up 5 minutes over +20 to +30 °C, with respect to 1 hour < ±0.1 ppm 15 minutes over +20 to +30 °C, with respect to 1 hour < ±0.01 ppm External reference input Frequency 1 MHz to 110 MHz, sine wave Lock range ±1 ppm, nominal Amplitude 0 to 10 dBm, nominal Connector 1 SMB snap-on Impedance 50 Ω, nominal
Figure 3. Output power normalized response at –120 dBm.
1 MHz to 5 GHz +10 dBm +19 dBm
- Specifications apply at the controlled temperature range. For temperatures outside this range, absolute level accuracy degrades by ± 0.02 dB/°C.
Figure 4. Amplitude repeatability at various carrier frequencies. Repeatability measures the ability of the instrument to return to a given power setting after a random excursion to any other frequency and power setting. It should not be confused with absolute level accuracy. Figure 5. Relative level accuracy at various carrier frequencies.
850 M H z
1900 M H z
2200 M H z
3500 M H z
5800 M H z
0.85 G H z
1.8 G H z
2 G H z
2.2 G H z
3.5 G H z
5.8 G H z
1 MHz to 6 GHz 1 W, nominal
Figure 6. Measured VSWR from 1 MHz to 6 GHz.
1 GHz -122 dBc/Hz, typical
2 GHz -117 dBc/Hz, typical
3 GHz -112 dBc/Hz, typical
6 GHz -108 dBc/Hz, typical
Figure 7. Single sideband phase noise in normal loop bandwidth from 10 Hz to 10 Figure 8. Single sideband phase noise in narrow loop bandwidth from 10 Hz to 10 MHz,
10 Hz 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz 10 MHz
11 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Technical Specifications and Characteristics (continued) Spectral purity (continued) Analog modulation Pulse parameters Pulse on/off ratio 1 MHz to 400 MHz > 85 dB, typical Pulse on/off ratio > 400 MHz to 6 GHz > 95 dB, typical Pulse rise/fall time < 10 ns, nominal Topic Windows 7 requirements Operating systems Windows 7 (32-bit and 64-bit) Processor speed 1 GHz 32-bit (x86), 1 GHz 64-bit (x64) (no support for Itanium 64) Available memory 4 GB minimum
8 GB or greater recommended
1.5 GB available hard disk space, includes:
– 1 GB available for Microsoft .NET framework 3.5 SP1 4 – 100 MB for Keysight IO libraries suite Video Support for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA graphics is supported) Browser Microsoft Internet Explorer 7.0 or greater Keysight IO libraries Version 16.3.16603.3 or later Broadband noise floor Range
1 MHz to 6 GHz < -140 dBc/Hz, nominal, at +10 dBm output power level
Range ≤ 0 dBm ≤ +10 dBm
1 MHz to < 400 MHz < -43 dBc -46 dBc, typical < -35 dBc -37 dBc, typical
400 MHz to 1.5 GHz < -29 dBc -31 dBc, typical < -27 dBc -29 dBc, typical > 1.5 GHz to 3 GHz < -35 dBc -39 dBc, typical < -30 dBc -33 dBc, typical Nonharmonics Nonharmonic miscellaneous spurious 2 < -70 dBc, nominal Nonharmonic HET band mixing spurs (0 dBm) < -67 dBc, nominal Nonharmonic Frac-N < -66 dBc, nominal Subharmonics
1 MHz to 6 GHz none
- Non-harmonics include mixing spurs for frequencies below 400 MHz, synthesizer spurs, and other miscellaneous chassis and power supply products, for offsets >10 kHz. 2. With a Keysight M9036A embedded controller. 3. Because of the installation procedure, less memory may be required for operation than is required for installation. 4. NET framework runtime components are installed by default with Windows Windows 7. Therefore, you may not need this amount of available disk space. System requirements
12 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Technical Specifications and Characteristics (continued) Environmental and physical specifications Temperature Operating Non-operating (storage) 0 to 55 °C -40 to +70 °C Humidity
1 Type tested at 95%, +40 °C
(non-condensing) Altitude Up to 15,000 feet (4,572 meters) Connectors RF OUT SMA female EMC Complies with European EMC Directive 2004/108/EC – IEC/EN 61326-2-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11 – ICES/NMB-001 This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada. Warm-up time 45 minutes Size M9300A M9301A M9310A
1 PXIe slot
Dimensions Module Length Width Height M9300A 210 mm 22 mm 130 mm M9301A 210 mm 22 mm 130 mm M9310A 210 mm 22 mm 130 mm Weight M9300A M9301A M9310A 0.551 kg (1.215 lbs) 0.535 kg (1.179 lbs) 0.551 kg (1.215 lbs) Power drawn from chassis M9300A M9301A M9310A ≤ 18 W ≤ 25 W ≤ 28 W 1. Samples of this product have been type tested in accordance with the Keysight environmental test manual and verified to be robust against the environmental stresses of storage, transportation and end-use--those stresses include but are not limited to temperature, humidity, shock, vibration, altitude and power-line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3.
13 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Software Instrument connection software Keysight IO library The IO suite offers a single entry point for connection to the most common instruments including AXIe, PXI, GPIB, USB, Ethernet/LAN, RS-232, and VXI test instruments from Keysight and other vendors. It automatically discovers interfaces, chassis, and instruments. The graphical user interface allows you to search for, verify and update IVI instrument and soft front panel drivers for modular and traditional instruments. The IO suite safely installs in side-by-side mode with NI I/O software. Free software download at www.keysight.com/find/iosuite Module setup and usage Keysight soft front panel The PXI module includes a soft front panel (SFP), a software-based graphical user interface (GUI) which enables the instrument’s capabilities from your PC. Included on CD-ROM shipped with module or online Programming Driver Development environments lVI-COM IVI-C LabVIEW MATLAB Visual Studio (VB.NET, C#, C/C++), VEE LabVIEW, LabWindows/CVI, MATLAB Included on CD-ROM shipped with module or online Programming assitance Command expert Assists in finding the right instrument commands and setting correct parameters. A simple interface includes documentation, examples, syntax checking, command execution and debug tools to build sequences for integration in Excel, MATLAB, Visual Studio, LabVIEW, VEE, SystemVue. Free software download at www.keysight.com/find/ commandexpert Programming examples Each module includes programming examples for Visual Studio.net, LabVIEW, MATLAB, LabWindows, and Keysight VEE Pro. Included on CD-ROM shipped with module or online at www.keysight.com/find/m9380a
14 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Assistance One day startup assistance Gain access to a technical expert who will help you get started quickly with the M9380A CW source and its powerful software tools. The flexible instruction format is designed to get you to your first measurements and familiarize you with ways to adapt the equipment to a specific application. Included in base configuration Calibration and traceability Factory calibration The M9380A CW source ships factory calibrated with an ISO-9002, NIST-traceable calibration certificate. Included in base configuration Calibration cycle A one year calibration cycle is recommended. Calibration sites – At Keysight worldwide service centers – On-site by Keysight – By self-maintainers More information visit www.keysight.com/find/infoline N7800A Calibration and adjustment software The M9380A CW source is supported by Keysight’s Calibration and adjustment software. This is the same software used at Keysight service centers to automate calibration. The software offers compliance tests for ISO 17025:2005, ANSI/NCSL Z540.3-2006, and measurement uncertainty per ISO Guide to Expression of Measurement Uncertainty. Licensed software. For more information, visit www.keysight.com/find/calibrationsoftware Keysight calibration status utility The Keysight calibration status utility helps ensure your M9380A is calibrated by managing the calibration interval and providing messages regarding instrument and module calibration status. Included in base configuration Setup and Calibration Services
15 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Support and Warranty Warranty Global warranty Keysight’s warranty service provides standard coverage for the country where product is used. – All parts and labor necessary to return to full specified performance – Recalibration for products supplied originally with a calibration certificate – Return shipment Included Standard Return to Keysight warranty—3 years 15 days typical turnaround repair service Included R-51B-001-5Z Return to Keysight warranty—5 years 15 days typical turnaround repair service Optional R-51B-001-3X Express warranty 3 years The express warranty upgrades the global warranty to provide, for 3 years, a 5 day typical turnaround repair service in the US, Japan, China and many EU countries. Optional R-51B-001-5X Express warranty 5 years The express warranty upgrades the global warranty to provide, for 5 years, a 5 day typical turnaround repair service in the US, Japan, China and many EU countries. Optional Support Core exchange program Keysight’s replacement core exchange program allows fast and easy module repairs. A replacement core assembly is a fully functioning pre-calibrated module replacement that is updated with the defective module serial number, allowing the replacement module to retain the original serial number. For qualified self-maintainers in US only Self-test utility A self-test utility runs a set of internal tests which verifies the health of the modules and reports their status. Included in base configuration
16 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheett Configuration and Ordering Information
Ordering information
M9380A PXIe CW source: 1 MHz to 3 or 6 GHz Includes: M9301A PXIe Synthesizer M9310A PXIe Source Output One day startup assistance Module interconnect cables Software, example programs and product information on CD Return to Keysight Warranty—3 Years Base configuration M9380A-F03 Frequency range: 1 MHz to 3 GHz M9380A-300 Required for warranted specifications PXIe frequency reference: 10 and 100 MHz Adds M9300A PXIe frequency reference: 10 and 100 MHz (M9300A module can support multiple M9380A modular instruments) Configurable options Frequency range M9380A-F03 1 MHz to 3 GHz M9380A-F06 1 MHz to 6 GHz Power M9380A-1EA High output power Calibration M9380A-UK6 Commercial calibration certificate with test data for M9380A (M9301A, M9310A) M9300A-UK6 Commercial calibration certificate with test data for M9300A (module only) Related products in recommended configuration M9037A PXIe embedded controller M9018A 18-slot PXIe chassis Recommended configuration Accessories Model Description Y1212A Slot blocker kit: 5 modules Y1213A PXI EMC filler panel kit: 5 slots Y1214A Air inlet kit: M9018A 18-slot chassis Y1215A Rack mount kit for M9018A 18-slot chassis Software information Supported operating systems Microsoft Windows 7 (32/64-bit) Standard compliant drivers IVI-COM, IVI-C, LabVIEW, MATLAB Supported application development environments (ADE) VisualStudio (VB.NET, C#, C/C++), VEE, LabVIEW, LabWindows/CVI, MATLAB Keysight IO libraries (version 16.3 or newer) Includes: VISA Libraries, Keysight Connection Expert, IO Monitor Keysight Command expert Instrument control for SCPI or IVI-COM drivers
Related products
M9021A PCIe cable interface M9045B PCIe ExpressCard adaptor for laptop connectivity Y1200B PCIe cable for laptop connectivity M9048A PCIe desktop adaptor for desktop connectivity Y1202A PCIe Cable for desktop connectivity M9381A PXIe Vector Signal Generator M9300A PXIe Frequency Reference Advantage services: Calibration and warranty Keysight Advantage Services is committed to your success throughout your equipment’s lifetime R-51B-001-5Z Return to Keysight warranty - 5 years R-51B-001-3X Express warranty - 3 years R-51B-001-5X Express warranty - 5 years N7800A Calibration & adjustment software
For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-09-04-14) 17 | Keysight | M9380A PXIe CW Source 1 MHz to 3 GHz or 6 GHz - Data Sheet myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.pxisa.org PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/m9380a www.keysight.com/find/modular This information is subject to change without notice. © Keysight Technologies, 2012 - 2015 Published in USA, March 25, 2015 5991-0283EN www.keysight.com