M9309A KEYSIGHT | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 10

Technical content

M9309A PXIe Vector Network Analyzer Synthesizer

187.5 MHz to 6 GHz

Figure 1. M9309A PXIe VNA synthesizer with up to 12-ports in one chassis. Figure 2. M9309A PXIe vector network analyzer synthesizer block diagram.

100 MHz

100 MHz In

850 MHz

6 GHz

100 MHz input signal

03 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Definitions for specifications Temperatures referred to in this document are defined as follows: – Full temperature range = Individual module temperature of ≤ 73 °C as reported by the module, and environment temperature of 0 to 55 °C. – Controlled temperature range = Individual module temperature of ≤ 50 °C as reported by the module, and environment temperature of 20 to 30 °C. Specifications describe the warranted performance of calibrated instruments. Data represented in this document are specifications under the following conditions unless otherwise noted. – Calibrated instruments have been stored for a minimum of 2 hours within the full temperature range – 45 minute warm-up time – Calibration cycle maintained – When used with Keysight M9300A frequency reference and Keysight interconnect cables Characteristics describe product performance that is useful in the application of the product. Characteristics are often referred to as Typical or Nominal values and are italicized. – Typical describes characteristic performance, which 80% of instruments will meet when operated within the controlled temperature range. – Nominal describes representative performance that is useful in the application of the product when operated within the controlled temperature range. Technical Specifications and Characteristics Recommended best practices in use – Use slot blockers and EMC filler panels in empty module slots to ensure proper operating temperatures. Keysight chassis and slot blockers optimize module temperature performance and reliability of test. – Set chassis fan to high at environmental temperatures above 45 °C. Additional information – All graphs contain measured data from one unit and are representative of product performance within the controlled temperature range unless otherwise noted. – The specifications contained in this document are subject to change. – The M9309A is only intended for use as part of the M9485A PXIe vector network analyzer. It is not intended to be used as a standalone instrument.

04 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Technical Specifications and Characteristics Frequency Frequency Range 187.5 MHz to 6 GHz Resolution 0.01 Hz Switching speed 1 < 100 µs, nominal Amplitude Amplitude Settable range -10 dBm to +10 dBm Resolution 0.2 dBm, typical Maximum output power ≤ 1 GHz +3 dBm > 1 GHz +5 dBm Power level accuracy 25 °C ± 5 °C ±1.5 dB 0 °C to 55 °C ≤ 3 GHz ±2 dB > 3 GHz ±3 dB Output impedence 50 Ω, nominal 1. Time for hardware to settle to within 1 ppm or 1 kHz, whichever is greater and amplitude to settle to within 0.2 dB.

05 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Spectral purity Phase noise 20 kHz offset

1 GHz -119 dBc/Hz, typical

2 GHz -114 dBc/Hz, typical

3 GHz -109 dBc/Hz, typical

6 GHz -105 dBc/Hz, typical

Range ≤ 0 dBm

300 MHz to 3 GHz < -27 dBc

Technical Specifications and Characteristics Figure 3: Single sideband phase noise from 100 Hz to 10 MHz offsets, at 1 GHz, 2 GHz, 3 GHz and 6 GHz RF frequencies.

06 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Technical Specifications and Characteristics Environmental and physical specifications Operating and storage conditions Temperature Operating 0˚C to 55˚C Non-operating (storage) -40˚C to 70˚C Humidity 1 Type-tested at 95%, +40˚C (non-condensing) Altitude Up to 15,000 feet (4,572 meters) Connectors RF Out, SMA female Regulatory characteristics EMC Complies with the essential requirements of the European EMC Directive as well as current editions of the following standards (dates and editions are cited in the Declaration of Comformity): – IEC/EN 61326-1 – CISPR Pub 11 Group 1, class A – AS/NZS CISPR 11 – ICES/NMB-001 This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada. Other Warm-up time 45 minutes Size 1 PXIe slot Dimensions Length 210 mm Width 22 mm Height 130 mm Weight 0.535 kg (1.179 lbs) Power drawn from chassis ≤ 22 W System requirements Topic Operating systems Windows 7 (32-bit and 64-bit) Processor speed 1 GHz 32-bit (x86), 1 GHz 64-bit (x64). No support for Itanium 64. Available memory 4 GB minimum

8 GB or greater recommended

Available disk space 2 1.5 GB available hard disk space, includes: – 1 GB available for Microsoft .NET framework 3.5 SP1 3 – 100 MB for Keysight IO libraries suite Video Support for DirectX 9 graphics with 128 MB graphics memory recommended (Super VGA graphics is supported) Browser Microsoft Internet Explorer 7.0 or greater Keysight IO libraries Version 16.3.16603.3 or later 1. Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environ - mental stresses of storage, transportation and end-use; those stresses include but are not limited to temperature, humidity, shock, vibration, altitude, and power line conditions. Test methods are aligned with IEC 60068-2 and levels are similar to MIL-PRF-28800F Class 3. 2. Because of the installation procedure, less memory may be required for operation than is required for installation. 3. NET framework runtime components are installed by default with Windows Windows 7. Therefore, you may not need this amount of available disk space.

07 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Software Instrument connection software Keysight IO library The IO library suite offers a single entry point for connection to the most common instruments including AXIe, PXI, GPIB, USB, Ethernet/LAN, RS-232, and VXI test instruments from Keysight and other vendors. It automatically discovers interfaces, chassis, and instruments. The graphical user interface allows you to search for, verify, and update IVI instrument and soft front panel drivers for modular and traditional instruments. The IO suite safely installs in side-by-side mode with NI I/O software. Free software download at: www.keysight.com/find/iosuite Setup and Calibration Services Calibration and traceability Factory calibration The M9309A PXIe VNA synthesizer ships factory calibrated with an ISO-9002, NIST-traceable calibration certificate. Included in base configuration Calibration cycle A one year calibration cycle is recommended. Calibration sites – At Keysight worldwide service centers – On-site by Keysight – By self-maintainers For more information visit www.keysight.com/find/infoline N7800A calibration and adjustment software The M9309A PXIe VNA synthesizer is supported by Keysight’s calibration and adjustment software. This is the same software used at Keysight service cen - ters to automate calibration. The software offers compliance tests for ISO 17025:2005, ANSI/NCSL Z540.3-2006, and measurement uncertainty per ISO Guide to Expression of Measurement Uncertainty. Licensed software. For more information, visit www.keysight.com/find/calibrationsoftware Keysight calibration status utility The Keysight calibration status utility helps ensure your M9309A is calibrated by managing the calibration interval and providing messages regarding instrument and module calibration status. Included in base configuration

08 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Support Support Core exchange program Keysight’s replacement core exchange program allows fast and easy module repairs. A replacement core assembly is a fully functioning pre-calibrated module replacement that is updated with the defective module serial number, allowing the replacement module to retain the original serial number. For qualified self maintainers in US only Self-test utility A self-test utility runs a set of internal tests which verifies the health of the modules and reports their status. Included in base configuration

09 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet Configurations and Ordering Information

Ordering information

M9309A PXIe VNA synthesizer: 187.5 MHz to 6 GHz

Related products

M9485A PXIe vector network analyzer M9018A 18-slot PXIe chassis M9037A PXIe embedded controller M9389A PXIe VNA source M9300A PXIe frequency reference Accessories Model Description Y1212A Slot blocker kit: 5 modules Y1213A PXI EMC filler panel kit: 5 slots Y1214A Air inlet kit: M9018A 18-slot chassis Y1215A Rack mount kit for M9018A 18-slot chassis Calibration Advantage services: Calibration Keysight Advantage Services is committed to your success throughout your equipment’s lifetime N7800A Calibration & adjustment software

10 | Keysight | M9309A PXIe Vector Network Analyzer Synthesizer - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2017 Published in USA, December 1, 2017 5992-0881EN www.keysight.com www.keysight.com/find/m9309a For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-9-7-17) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System Evolving Since 1939 Our unique combination of hardware, software, services, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight. myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. http://www.keysight.com/find/emt_product_registration Register your products to get up-to-date product information and find warranty information. Keysight Services www.keysight.com/find/service Keysight Services can help from acquisition to renewal across your instrument’s lifecycle. Our comprehensive service offerings—one- stop calibration, repair, asset management, technology refresh, consulting, training and more—helps you improve product quality and lower costs. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience.