PXI High Voltage Data Acquisition Module 32-Channel, 250 kS/s, 16-bit, 5 V/100 V Data Sheet

The Keysight Technologies, Inc. M9216A is a multi-channel, large-range vo ltage measurement unit capable of acquiring dynamic positive voltage levels from multiple chann els simultaneously, making it ideal for automated test equipment (ATE) in automotive electronics applications. Product description The M9216A is a two slot, PXI-based, high voltage data acquisition module with a 16 -bit resolution that allows simultaneous measurement of 8 channels of positive voltages r anging from 1 mV to 100 V. Each channel in the module comes with a concurrent 5 V and 100 V measurement rang e instead of having multiple switchable voltage ranges. This allows each channel to ac quire signals that luctuate between very low and very high voltage levels without the need for switching r anges and doing separate measurements. The M9216A module also comes with a built-in 32 to eight multiplexer, enabling 3 2 measurement ports to be connected to the eight acquisition channels expanding it to a full bl own 32 channel acquisition module. Alternatively, these channels can also be routed out v ia auxiliary output connector J101 eliminating the need for external multiplexers and pin matrix cards f or additional measurements. With the support of up to 32 channels with each channel equipped with dual concurre nt measurement ranges, the M9216A module is capable of doing fast parallel voltage level meas urements with guaranteed accuracy. This is ideal for the automotive industry and since m ost automotive applications do not require negative voltage measurements, the M9216A’s 16-bit ADC is ful ly used for positive voltages giving it better resolution hence better accuracy. Application –Automotive

Features

–16-bit resolution – 32 channels – Dual concurrent voltage ranges for each channel: – > 5 V (1 mV to 5 V) – > 100 V (20 mV to 100 V) – Input accuracy at 5 V – > zero offset : 5 V range – 200 uV, 100 V range – 1 mV – > gain (% of reading): 5 V range – 0.05%, 100 V range – 0.05% – > noise @ 3 sigma: 5 V range – 200 uV, 100 V range – 2 mV – Auxiliary outputs for additional measurements – Connector compatibility with cPCI, PXI-H, PXI-1 – Included drivers: IVI-COM, IVI-C, LabVIEW G Customer values – High-channel count for fast and accurate acquisition of multiple voltage levels and doing parallel testing – Concurrent dual range for each channel to support large voltage range without the need for switching ranges – Auxiliary output for additional measurements

format offers high performance in a small, rugged package. converters and switch multiplexers. in Figure 1, making a total of four acquisition banks. capability to include modular instruments. soft front panel directly from Keysight Connection Expert. that the card can be quickly integrated into a test system. Figure 1. Example of a data acquisition bank (Bank 1)

04 | Keysight | PXI High Voltage Data Acquisition Module - Data Sheet Technical Speciications Voltage input input (general) Number of Channels 8 concurrent dual range channels Resolution 16 bit Sampling rate 250 kS/s Analog bandwidth (anti-aliasing filter) 85 kHz Dual range specifications 5 V 100 V 1 mV to 5V 20 mV to 100 V Input coupling DC Input impedance 550 kΩ (typical) Internal sample memory 30,000 samples for 5 V and 100 V range per channel (multiple mode) 240,000 samples for 5 V and 100 V range for one channel (single mode) Trigger Software and digital trigger Pre and post trigger sample acquisition Timed delay Voltage input input accuracy (at 25 + 3° C)

5 V Range

Zero offset ±200 μV Gain (% of reading) ±0.05% Noise 3 sigma 200 μV

100 V Range

Gain (% of reading) 0.05% Noise 3 sigma 2 mV Multiplexer Input channels 32 Output channels 8 channels to ADC 8 channels to primary auxiliary out 8 channels to complementary auxiliary out Maximum input voltage 100 V Maximum input current 0.5 A Maximum common return pin voltage with respect to chassis ground 45 V Power Maximum current consumption from PXI 5 V 3.3 V 0.8 A 0.5 A Maximum input voltage 100 V Maximum input current 0.5 A Maximum common return pin voltage with respect to chassis ground 45 V Warm up time NOTE: A minimum system warm up time of 15 minutes is required. For optimum accuracy, a warm up time of one hour is recommended. 0.5 hour

05 | Keysight | PXI High Voltage Data Acquisition Module - Data Sheet General Speciications Environmental and physical Operating temperature 0 to 55° C Storage temperature –20° C to 70° C Relative humidity 0% to 80% non condensing Dimensions 3U, 2-slot, PXI/cPCI module; 40.30 mm x 129.11 mm × 212.73 mm Weight 0.51 kg (1.12 lb) Certifications and Compliance Altitude : 10,000 ft (Operating)/ 15,000 ft (Non-operating) CE Mark Compliance 2006/95/EC; 2004/108/EC Safety Pollution Degree 2 EMC Immunity EN/IEC 61326-1 Industrial Environment EMC Emissions EN/IEC 61326-1 Class A Deinitions for speciications Specifications describe the warranted performance of calibrated instruments that have been stored for a minimum of 2 hours within the operating temperature range of 0 to 50°C, unless otherwise stated, and after a 45 minute warm-up period. Data represented in this document are specifications unless otherwise noted. Characteristics describe product performance that is useful in the application of the product, but that is not covered by the product warranty. Characteristics are often referred to as Typical or Nominal values. Characteristic are represented in italics. – Typical describes characteristic performance, which 80% of instruments are expected to meet when operated over a 20 to 30 °C temperature range. Typical performance is not warranted. – Nominal describes representative performance that is useful in the application of the product when operated over a 20 to 30 °C temperature range. Nominal performance is not warranted. Note: All graphs contain measured data from several units at room temperature unless otherwise noted

06 | Keysight | PXI High Voltage Data Acquisition Module - Data Sheet Software In addition, the M9216A includes a soft front panel graphical interface. This simple software application can be used to – Enable verification that the modules are properly installed and working in the system. – Assist in learning the module capability and behavior via interactive use. – Assist in acquiring programming knowledge of the instrument. – Assist in the verification and debugging of the system. Coniguration and Ordering Information Model Description M9216A M9216A PXI High voltage data acquisition, 32-channel, 250KS/S, 16-BIT, 100 V input Ordering Information Warranty and Calibration Warranty and calibration are provided at the module level. Advantage Services: Calibration and Warranty Keysight Advantage Services is committed to your success throughout your equipment’s lifetime. Warranty Standard warranty is 1 year R-9MB-001-3C 1 year return-to-Keysight warranty extended to 3 years R-9MB-001-5C 1 year return-to-Keysight warranty extended to 5 years Calibration intervals The M9216A is factory calibrated and shipped with a calibration certificate. Calibration is recommended every year in order to verify product performance. Model Description Keysight IO Libraries Keysight IO Libraries Drivers, soft front panels and programming examples in LabVIEW, LabWindows/CVI, Visual Studio C, C++ and C#, Visual Basic, and MATLAB 1. For the M9216A to work properly, at least one PXI chassis and one PXI controller type must be available. 2. Keysight IO Libraries Suite 16.0 is required. The modular product won’t work with Keysight IO Libraries Suite versions earlier than version 16.0. Accessories M9216A-CD1 Software and product information on CD

www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.axiestandard.org AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Keysight is a founding member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo are registered US trademarks of the PCI Industrial Computer Manufacturers Group. www.pxisa.org PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. www.keysight.com/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/modular www.keysight.com/find/m9216a For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Paciic Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 0800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-07-10-14) 07 | Keysight | PXI High Voltage Data Acquisition Module - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2012 - 2014 Published in USA, August 2, 2014 5990-7582EN www.keysight.com