L8601-01 HAMAMATSU | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 4

Technical content

G Focal Spot Size: 5 µm (at 4 W) Providing 5µm focal spot size, for the sealed-off type X-ray source with built-in power supply. It gives a clear clair- voyance image despite of higher magnification ratio. G Easy Handling Convenient handling (with no high voltage wiring) since the air cooled X-ray source and the high voltage power supply are in one package. G Continuous Operation It allows continuous maximum output operation. G Built-in External Control Function It is automatically operable from external control since it has a built-in RS-232C interface. G X-ray Non-destructive Test

  • Printed Circuits • Electronic Components
  • Semiconductor Devices • Ceramics
  • Soldered Parts • Rubbers
  • Combinated Materials • Plastics
  • New materials • Die casting G Multilayer Printed Circuit Board X-ray Inspection•Drilling Instrument G Industrial On-line Process Control and Measurement G Industrial X-ray Computer Tomography System

NOTE: a In the range of output (= Tube voltage × Tube current) less than 10 W NOTE: a Without moisture condensation. Sealed Type Convection Cooling Beryllium (150 µmt) Tungsten Dispenser Cathode End-window Built-in Type +10 to +40 °C / 0 to +50 °C 85 % RH Max.a Approx. 6 kg Parameter Description G GENERAL IX-RAY TUBE UNIT Function Protection External Control Operation /Storage Temperature Operation/Storage Humidity Power Consumption Input Voltage Installation Method Weight Target Voltage/Current Pre-set, Auto Warm-up Safety Interlock RS-232C 0 to +40 °C 85 % RH Max. a 200 W Max.

24 Vdc

Designed for both bench-top and rack mounting Approx. 5 kg Parameter Description IX-RAY CONTROL UNIT Target Voltage Target Current Maximum Output Power Focal Spot Size Beam Angle Minimum Distance Focus/Object Environmental X-ray Dose Leakage Operation Parameter 20 to 90 0 to 250 a 10 (isowatts) 7 (5 µm at 4 W) 9.5 Continuous Value/Description kV µA W µm degree mm mSv/h Max. Unit

DIMENSIONAL OUTLINE (Unit: mm)

90 KV MICROFOCUS X-RAY SOURCE

135 ± 2 107 ± 2 70 5 36.5 52 30 290.0 ± 0.517 110.0 ± 0.5 16.5 320.5 ± 2.0 9.5 39° MIN. 100 X-RAY UNIT INSTALLATION SCREW HOLES 4-M4 D5 4-M3 L = 4 X-RAY FOCAL SPOT X-RAY MODE VOLTAGE CONTROL kV X-RAY STANDBY AGING X-RAY ON OFF PRE-SET INTER LOCK CURRENT X-RAY90 CONTROL UNIT OPERATE µA CONTROL PRE-SET ON REMOTEOFF OVER ERROR OUTPUT X-RAY REMOTE RS-232C INTER LOCK 2INTER LOCK 1AGING SIGNAL IDENTIFICATION PLATE 426 ± 299 4 0 482 ± 2 300 ± 2 30 100 MIN. 99 ± 212 INTERCONNECTION CABLE CONNECTOR DIA 33 mm MAX. LABEL TLSOA0063EA TLSOA0076EA

MAY. 2001 IP * This product(s) conforms to the EMC directive (89/336/EEC) of the European Union. * To meet FDA standards, the Interlock-1 circuit incorporates a relay that shuts off the X-ray generation power supply circuit if the circuit is open. PRE-CAUTION TO USE Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications are subject to change without notice. No patent rights are granted to any of the circuits described herein. ©2001 Hamamatsu Photonics K.K Subject to local technical requirements and regulations, availability of products included in this promotional material may var y. Please consult with our sales office. # WARRANTY PERIOD

RELATED PRODUCTS

The C7716 series X-ray image intensifier (I. I.) camera unit now offers greatly improved X-ray detection efficiency even at low energy levels. This improve- ment stems primarily from a built-in X-ray image intensifier having an extreme- ly thin aluminum input window. Its thickness is only 0.3mm or less, virtually at the limit of present technology providing excellent X-ray transmit-tance and low scattering. The results are sharp, clear, high-quality images taken at low energy X-ray lev- els down to several keV which penetrate plastic (PET) materials. X-RAY IMAGE INTENSIFIER CAMERA UNIT C7716, C7716-10 The C7876 is an X-ray image intensifier camera unit ideal for non-destructive inspection of light-element materials and radiation imaging at low-energy X-ray levels. The C7876 remarkably improves X-ray transmittance at low-energy X-ray lev- els by using a beryllium window instead of the aluminum window currently used for most X-ray image intensifiers. The result is a sharp and clear, high-contrast image taken in real time even at low energy X-ray levels down to a few keV. X-RAY IMAGE INTENSIFIER CAMERA UNIT C7876, C7876-10 HAMAMATSU PHOTONICS K.K., Electron Tube Center 314-5, Shimokanzo, Toyooka-village, Iwata-gun, Shizuoka-ken, 438-0193, Japan, Telephone: (81)539/62-5248, Fax: (81)539/62-2205 Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658 E-mail: info@hamamatsu.de United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW, United Kingdom, Telephone: 44-(0)1707-294888, Fax: 44(0)1707-325777 E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 SOLNA, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01 E -mail: info@hamamatsu.se Italy: Hamamatsu Photonics Italia: S.R.L.: Strada della Moia, 1/E, 20020 Arese, (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741 E-mail: info@hamamatsu.it HOMEPAGE URL http://www.hamamatsu.com 1. X-ray emitted from this device is harmful for human body. And it should be necessary for the operator to protect himself/her self from it. 2. During an operation, the X-ray tube unit should be installed in the X-ray shielded cabinet in order to avoid any X-ray leakage. Also the safety interlock system in X-ray control unit should be always used in order to avoid any misoperation. This device is guaranteed for one year from data of purchase or 3000 hours of operation, whichever comes first. The warranty extends only to replacement of the products. The warranty does not cover damage due to misuse or natural galamity.