E5080A ENA Series Network Analyzer - 9 kHz to 4.5/6.5/9 GHz E5092A Configurable Multiport Test Set Data Sheet
02 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Options This document provides technical specifications for the E5080A ENA network analyzer. E5080A-245 2-port test set, 9 kHz to 4.5 GHz, with bias tees E5080A-265 2-port test set, 9 kHz to 6.5 GHz, with bias tees E5080A-295 2-port test set, 9 kHz to 9 GHz, with bias tees E5080A-445 4-port test set, 9 kHz to 4.5 GHz, with bias tees E5080A-465 4-port test set, 9 kHz to 6.5 GHz, with bias tees E5080A-495 4-port test set, 9 kHz to 9 GHz, with bias tees Calibration kits and ECal modules This E5080A data sheet also provides technical specifications for the following calibration kits and ECal modules. For models not listed in this data sheet, please download the free Uncertainty Calculator from ( http://www.keysight.com/find/na_ calculator) to generate the curves for your calibration kit and enable ENA setup. 85032F Calibration kit 85033E Calibration kit 85092C Electronic calibration (ECal) module 85093C Electronic calibration (ECal) module Definitions Specification (spec.): Warranted performance. All specifications apply at 25 ºC (± 5 ºC), unless otherwise stated, and 90 minutes after the instrument has been turned on. Specifications include guard bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Typical (typ.): Expected performance of an average unit which does not include guardbands. It is not covered by the product warranty. General characteristics: A general, descriptive term that does not imply a level of performance.
03 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Corrected System Performance The specifications in this section apply to measurements made with the Keysight E5080A network analyzer under the following conditions: – No averaging applied to data – Environmental temperature of 23 °C (± 3 °C) with less than 1 °C deviation from the calibration temperature – Response and isolation calibration performed System dynamic range1 Specification (10 Hz IFBW) Typical (3 Hz IFBW) 9 k to 100 kHz 100 k to 50 MHz
50 M to 6 GHz
6 G to 8.5 GHz
8.5 G to 9 GHz
System dynamic range (specification and actual measurement data example) 1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effective dynamic range must take measurement uncertainty and interfering signals into account.
04 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Corrected System Performance with Calibration Kit Corrected system performance with type-N device connectors, 85032F calibration kit Network analyzer : E5080A Calibration kit : 85032F (Type-N, 50 Ω) Calibration : full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 3 °C) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification Description 9 k to 300 kHz 300 k to 10 MHz 10 M to 3 GHz 3 G to 6 GHz 6 G to 8.5 GHz 8.5 G to 9 GHz Directivity 49 49 46 40 38 38 Source match 41 41 40 36 35 35 Load match 48 49 46 40 37 36 Transmission uncertainty (Specification)1 Reflection uncertainty (Specification)1 1. Applies to the units with Serial Number Prefix MY552/SG552 and above
05 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Corrected system performance with type-N device connectors, 85092C electronic calibration (ECal) module Network analyzer : E5080A Calibration kit : 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module Calibration : full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 3 °C) with < 1 °C deviation from calibration temperature , isolation calibration is not performed Specification Description 9 k to 300 kHz 300 k to 10 MHz 10 M to 3 GHz 3 G to 6 GHz 6 G to 8.5 GHz Directivity 45 52 49 45 45 Source match 36 44 41 36 36 Load match 37 47 44 38 38 Transmission uncertainty (Specification)1 Reflection uncertainty (Specification)1 1. Applies to the units with Serial Number Prefix MY552/SG552 and above
06 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Corrected system performance with 3.5 mm device connector type, 85033E calibration kit Network analyzer : E5080A Calibration kit : 85033E (3.5 mm, 50 Ω) Calibration : full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 3 °C) with < 1 °C deviation from calibration temperature, isolation calibration performed Transmission uncertainty (Specification)1 Reflection uncertainty (Specification)1 Specification Description 9 k to 300 kHz 300 k to 10 MHz 10 M to 3 GHz 3 G to 6 GHz 6 G to 8.5 GHz 8.5 G to 9 GHz Directivity 46 46 44 38 38 38 Source match 43 43 40 37 36 36 Load match 46 46 44 38 38 38 1. Applies to the units with Serial Number Prefix MY552/SG552 and above
07 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration (ECal) module Network analyzer : E5080A Calibration kit : 85093C (3.5 mm, 50 Ω) Electronic calibration (ECal) module Calibration : full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 3 °C) with < 1 °C deviation from calibration temperature , isolation calibration is not performed Transmission uncertainty (Specification)1 Reflection uncertainty (Specification)1 Specification Description 300 k to 10 MHz 10 M to 3 GHz 3 G to 6 GHz 6 G to 8.5 GHz 8.5 G to 9 GHz Directivity 45 52 50 47 47 Source match 36 44 39 34 34 Load match 37 47 44 40 39 1. Applies to the units with Serial Number Prefix MY552/SG552 and above
08 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Uncorrected System Performance User correction: OFF, system error correction: ON Specification Description 9 k to 50 kHz 50 k to 300 kHz 300 k to3 GHz 3 G to 6 GHz 6 G to 8.5 GHz 8.5 G to 9 GHz Directivity 20 20 25 20 15 15 Source match 20 20 25 20 15 15 Load match 8 12 17 12 10 8 Test Port Output (Source) Test port output frequency Description Specification Typical Frequency range Option 245, 445 Option 265, 465 Option 295, 495 9 k to 4.5 GHz 9 k to 6.5 GHz 9 k to 9 GHz Resolution 1 Hz CW accuracy Standard Option 1E5 ± 7 ppm (23 ± 3 ºC) ± 0.45 ppm ± 3.0 ppm (25 ± 5 ºC) Source stability Standard Option 1E5 ± 7 ppm (0 to 40 ºC) ± 0.05 ppm (0 to 40 ºC) ± 0.4 ppm/year Typical Description 9 k to 50 kHz 50 k to 300 kHz 300 k to3 GHz 3 G to 6 GHz 6 G to 8.5 GHz 8.5 G to 9 GHz Directivity 40 40 40 35 35 35 Source match 40 40 40 35 35 35 Load match 12 18 20 20 16 14 Typical Description 9 k to 100 kHz 100 k to 50 MHz 50 M to 6 GHz 6 G to 8.5 GHz 8.5 G to 9 GHz Crosstalk 132 141 147 140 130
09 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Test port output signal purity Description Specification Typical Harmonics (2nd or 3rd) (at 0 dBm) 9 k to 2 GHz
2 G to 9 GHz
< –25 dBc < –20 dBc Non-harmonic spurious (at 0 dBm) < -30 dBc 1. Source output performance on all ports. 2. Power calibration using an external power meter improves level accuracy of the test port output power. 3. Level linearity given is relative to 0 dBm. Test port output power1 Description Specification Typical Nominal power (preset power) 0 dBm Range 9 k to 100 kHz 100 k to 50 MHz 6 G to 8.5 GHz —90 to 10 dBm —90 to 12 dBm —90 to 15 dBm —90 to 12 dBm —90 to 8 dBm —110 to 12 dBm —110 to 14 dBm —110 to 17 dBm —110 to 14 dBm —110 to 12 dBm Max leveled power 9 k to 100 kHz 100 k to 50 MHz 6 G to 8.5 GHz Resolution 0.01 dB Level accuracy Sweep type: stepped 9 k to 50 kHz 50 k to 9 GHz Sweep type: auto 9 k to 9 GHz ± 2.0 dB ± 1.5 dB ± 2.0 dB Level linearity —20 dBm to maximum power Sweep type: stepped Sweep type: auto —110 dBm to —20 dBm ± 0.75 dB ± 1.0 dB ± 1.0 dB Power sweep range 9 k to 100 kHz 100 k to 50 MHz
50 M to 6GHz
6 G to 8.5 GHz —90 to 10 dBm —90 to 12 dBm —90 to 15 dBm —90 to 12 dBm —90 to 8 dBm
10 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Test Port Input Description Specification Typical Damage Level +27 dBm or ± 35 VDC Test port noise floor 1 9 k to 100 kHz 100 k to 50 MHz 6 G to 8.5 GHz —120 dBm —127 dBm —130 dBm —130 dBm —122 dBm —130 dBm —137 dBm —140 dBm —136 dBm —128 dBm Compression level (@+10dBm input) Magnitude 9 k to 6 GHz
6 G to 9 GHz
0.207 dB 0.207 dB 0.503 deg 0.503 deg 0.04 dB 0.08 dB 0.3 deg 0.6 deg 0.1 dB compression input level 9 k to 100 k 100 k to 50 MHz 6 G to 8.5 GHz
- IFBW=1 Hz, FOM is off.
11 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Trace noise Description Specification Typical (test port input level = maximum power in Specification) Magnitude (Transmission) 9 k to 30 kHz, 1 kHz IFBW 30 k to 10 MHz, 1 kHz IFBW
10 M to 6 GHz, 10 kHz IFBW
6 G to 9 GHz, 10 kHz IFBW
(Reflection) 9 k to 30 kHz, 1 kHz IFBW 30 k to 10 MHz, 1 kHz IFBW 10 M to 1.5 GHz, 10 kHz IFBW
1.5 G to 6 GHz, 10 kHz IFBW
6 GHz to 9 GHz, 10 kHz IFBW
0.003 dBrms 0.0015 dBrms 0.0015 dBrms 0.002 dBrms 0.004 dBrms 0.003 dBrms 0.003 dBrms 0.0015 dBrms 0.002 dBrms 0.001 dBrms 0.0005dBrms 0.0005 dBrms 0.0006 dBrms 0.002 dBrms 0.001 dBrms 0.001 dBrms 0.0005 dBrms 0.0006 dBrms Phase (Transmission) 9 k to 30 kHz, 1 kHz IFBW 30 k to 100 kHz, 1 kHz IFBW 100 k to 10 MHz, 1 kHz IFBW (Reflection) 9 k to 30 kHz, 1 kHz IFBW 30 k to 100 kHz, 1 kHz IFBW 100 k to 10 MHz, 1 kHz IFBW 10 M to 1.5 GHz, 10 kHz IFBW 0.03 degrms 0.02 degrms 0.01 degrms 0.01 degrms 0.02 degrms 0.06 degrms 0.035 degrms 0.02 degrms 0.02 degrms 0.01 degrms 0.02 degrms 0.015 degrms 0.008 degrms 0.003 degrms 0.003 degrms 0.006 degrms 0.02 degrms 0.01 degrms 0.005 degrms 0.005 degrms 0.003 degrms 0.006 degrms Stability Description Specification Typical Magnitude 9 k to 3 GHz
3 G to 6 GHz
± 0.005 dB/°C ± 0.01 dB/°C ± 0.04 dB/°C ± 0.1 deg/°C ± 0.2 deg/°C ± 0.8 deg/°C
12 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Dynamic accuracy1, 2 Description Specification Typical Magnitude 10 dBm 5 dBm 0 dBm —5 dBm —10 dBm (Ref.) —15 dBm —20 dBm —30 dBm —40 dBm —50 dBm —60 dBm —70 dBm —80 dBm —90 dBm —100 dBm ± 0.207 dB ± 0.071 dB ± 0.027 dB ± 0.012 dB ± 0.006 dB ± 0.009 dB ± 0.011 dB ± 0.014 dB ± 0.018 dB ± 0.021 dB ± 0.029 dB ± 0.072 dB ± 0.116 dB ± 0.159 dB ± 0.202 dB Phase 10 dBm 5 dBm 0 dBm —5 dBm —10 dBm (Ref.) —15 dBm —20 dBm —30 dBm —40 dBm —50 dBm —60 dBm —70 dBm —80 dBm —90 dBm —100 dBm ± 5.02 deg ± 0.67 deg ± 0.14 deg ± 0.06 deg ± 0.04 deg ± 0.05 deg ± 0.06 deg ± 0.08 deg ± 0.10 deg ± 0.13 deg ± 0.18 deg ± 0.47 deg ± 0.76 deg ± 1.05 deg ± 1.34 deg 1. Accuracy of the test port input power reading is relative to –10 dBm reference input power level. IFBW is equal to or lower than 10 Hz. 2. Applies to the units with Serial Number Prefix MY552/SG552 and above
13 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Group delay1 Description Specification Typical Aperture (selectable) (frequency span)/(number of points - 1) Maximum aperture 25% of frequency span Minimum delay Limited to measuring no more than 180° of phase change within the minimum aperture. Accuracy See graph below (typical) 1. Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep). The following graph shows group delay accuracy with type-N connectors, full 2-port calibration and a 10 Hz IF bandwidth. – Calibration kit (85032F). – Insertion loss is assumed to be < 2 dB. In general, the following formula can be used to determine the accuracy, in seconds, of a specific group delay measurement: ± phase accuracy (degrees)/[360 x aperture (Hz)]
14 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet General Information Description General characteristic IF band width Range 1 Hz to 1.5 MHz Nominal settings are: 150, 200, 300, 400, 500, 700, 1 k, 1.5 k, 2 k, 3 k, 4 k, 5 k, 7 k, 10 k, 15 k, 20 k, 9 k, 40 k, 50 k, 70 k, 100 k, 150 k, 200 k, 300 k, 400 k, 500 k, 700 k, 1M, 1.5 M Front panel Description Typical General characteristic Test Ports Type –N, female, 50 Ω (nominal) Display Type Resolution 12.1 inch TFT color LCD with touch screen WXGA (1280 x 800) USB host port Universal serial bus jack, Type A configuration; female; provides connection to mouse, key board, printer, ECal module, USB power sensor, or USB/ GPIB interface Rear panel Description Typical General characteristic External trigger input connector Type Input level Pulse width Polarity BNC, female Low threshold voltage: 0.5 V High threshold voltage: 2.1 V Input level range: 0 to + 5 V ≥ 2 μsec Positive or negative External trigger output connector Type Maximum output current Output level Pulse width Polarity BNC, female 50 mA Low level voltage: 0 V High level voltage: 5 V 1 μsec Positive or negative External reference signal input connector Type Input frequency Input level BNC, female
10 MHz ± 10 ppm
—3 to +10 dBm Internal reference signal output connector Type Output frequency Signal Type Output level Output Impedance
10 MHz ± 7 ppm
0 dBm ± 3 dB into 50 Ω BNC, female 50 Ω
15 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Description Specification Typical AUX input connector Type Input range Accuracy 1 % ± 1 mV for 1 V input range 1 % ± 10 mV for 10 V input range BNC, female x 2 ports ± 1 V or ± 10 V selectable Description Typical General characteristic Internal reference signal oven connector (option 1E5) Type Output frequency Output level 10 MHz ± 0.45 ppm 0 dBm minimum BNC, female Bias tee input connector Type Damage level Max. current (no degradation in RF specifications) Over current protection ± 35 V, 500 mA DC ± 300 mA at 100 kHz to 9 GHz 500 mA (with Fuse) BNC, female Video output 15-pin mini D-Sub; female; drives WXGA compatible monitors GPIB 24-pin D-Sub (Type D-24), female; compatible with IEEE-488 USB host port Universal serial bus jack, Type A configuration; female; provides connection to mouse, key board, printer, ECal module, USB power sensor, or USB/ GPIB interface USB (USBTMC 1) interface port Universal serial bus jack, Type B configuration (4 contacts inline); female; provides connection to an external PC; compatible with USBTMC- USB488 and USB 2.0.LA LAN 10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data rates Handler I/O port 36-pin centronics, female; provides connection to handler system Line power 2 Frequency Power consumption 3 Voltage VA max 2-port: 135 W 4-port: 165 W
47 Hz to 63 Hz
90-264 VAC (Vpeak > 120 V) 300 VA max. 1. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards. 2. A third-wire ground is required. 3. At preset.
16 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet EMC, safety, environment and compliance Description General characteristics EMC European Council Directive 2004/108/EC IEC 61326-1:2012 EN 61326-1:2013 CISPR 11:2009 +A1:2010 EN 55011: 2009 +A1:2010 Group 1, Class A IEC 61000-4-2:2008 EN 61000-4-2:2009 4 kV CD / 8 kV AD IEC 61000-4-3:2006 +A1:2007 +A2:2010 EN 61000-4-3:2006 +A1:2008 +A2:2010 3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz /1V/m, 2.0 - 2.7 GHz, 80% AM IEC 61000-4-4:2004 +A1:2010 EN 61000-4-4:2004 +A1:2010 1 kV power lines / 0.5 kV signal lines IEC 61000-4-5:2005 EN 61000-4-5:2006 0.5 kV line-line / 1 kV line-ground IEC 61000-4-6:2008 EN 61000-4-6:2009 3 V, 0.15-80 MHz, 80% AM IEC 61000-4-8:2009 EN 61000-4-8:2010 30A/m, 50/60Hz IEC 61000-4-11:2004 EN 61000-4-11:2004 ICES-001:2006 Group 1, Class A AS/NZS CISPR11:2004 Group 1, Class A KN11, KN61000-6-1 and KN61000-6-2 Group 1, Class A Safety European Council Directive 2006/95/EC IEC 61010-1:2001/EN 61010-1:2001 Measurement Category I Pollution Degree 2 Indoor Use CAN/CSA C22.2 No. 61010-1-12 Measurement Category I Pollution Degree 2 Indoor Use Environment This product complies with the WEEE Directive (2002/96/EC) marking requirements. The af - fixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local Keysight office, or see http://www.keysight.com/environment/product/ for more information. Compliance Class C
17 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Analyzer environmental specifications and dimensions Description General characteristic Operating environment Temperature Error-corrected temperature range Humidity Altitude Vibration 0 °C to +40 °C 23 °C (± 3 °C) with < 1 °C deviation from calibration temperature 20% to 80 % at wet bulb temperature < +29 °C (non-condensation) 0 to 2,000 m (0 to 6561 feet)
0.21 G maximum, 5 Hz to 500 Hz
—10 °C to +60 °C 20 % to 90 % at wet bulb temperature < +40 °C (non-condensation) 0 to 4,572 m (0 to 15,000 feet)
0.5 G maximum, 5 Hz to 500 Hz
Weight (net) Option 245/265/295: 20.3 kg Option 445/465/495: 22.7 kg Dimensions Dimensions (front view, E5080A with option 245/265/295, in millimeters)
18 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Dimensions (front view, E5080A with option 445/465/495, in millimeters) Dimensions (rear view with option 1E5, in millimeters)
19 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Dimensions (rear view with option 1E5, in millimeters) Measurement Throughput Summary Measurement throughput data is supplemental performance data. Common condition for the measurement throughput data: – Analyzer display turned off with: DISPlay:ENABle OFF, DISPlay:VISible OFF – Number of traces = 1 – firmware version: A.11.00 Cycle time for measurement completion Sweep type: Auto System error correction: OFF Sweep type: Stepped System error correction: ON Number of Points 51 201 401 1601 51 201 401 1601 Start 800 MHz, stop 1 GHz, 500 kHz IF bandwidth 4-port cal 4.0 6.2 12 31 4.7 11 18 51 Start 800 MHz, stop 1 GHz, 1 kHz IF bandwidth Uncorrected 51 196 389 1546 51 196 389 1546 2-port cal 101 391 777 3092 101 391 777 3092 4-port cal 201 781 1554 6187 201 781 1554 6187 Start 100 kHz, stop 4.5 GHz, 500 kHz IF bandwidth 2-port cal 5.7 11 15 24 5.8 11 15 39 4-port cal 11 21 31 50 11 21 31 81 Unit: ms
20 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Start 100 kHz, stop 4.5 GHz, 1 kHz IF bandwidth Uncorrected 53 199 393 1554 53 199 393 1554 2-port cal 104 396 784 3107 104 396 784 3107 4-port cal 207 792 1569 6217 207 792 1569 6217 Start 100 kHz, stop 9 GHz, 500 kHz IF bandwidth 2-port cal 7.0 12 17 35 7.0 12 17 40 4-port cal 14 23 33 73 14 23 34 83 Start 100 kHz, stop 9 GHz, 1 kHz IF bandwidth Uncorrected 53 199 393 1554 53 199 393 1554 2-port cal 105 397 785 3108 105 397 785 3108 4-port cal 210 794 1571 6219 210 794 1571 6219 Unit: ms Cycle time vs. number of points Condition: Start = 800 MHz, Stop = 1 GHz, 500 kHz IF bandwidth Number of points Sweep type: Auto System error correction: OFF Sweep type: Stepped System error correction: ON 3 1.1 1.0 11 1.1 1.1 51 1.3 1.5 101 1.4 2.0 201 1.8 2.8 401 2.5 4.3 801 3.9 7.0 1601 7.0 13 Unit: ms Cycle time vs. IF bandwidth Condition: Frequency = 4 GHz, NOP = 201, system error correction: OFF IF BW (Hz) Cycle time (ms) IF BW (Hz) Cycle time (ms) IF BW (Hz) Cycle time (ms) IF BW (Hz) Cycle time (ms) IF BW (Hz) Cycle time (ms) IF BW (Hz) Cycle time (ms) 10 19299 100 1932 1000 195 10000 21 100000 3.1 1000000 1.8 15 12867 150 1288 1500 131 15000 14 150000 2.4 1500000 1.8 20 9651 200 967 2000 98 20000 11 200000 2.1 30 6434 300 645 3000 66 30000 7.6 300000 1.9 40 4826 400 484 4000 50 40000 6.0 400000 1.8 50 3861 500 388 5000 40 50000 5.0 500000 1.8 70 2736 700 275 7000 29 70000 3.9 700000 1.8
21 | Keysight | E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet Data transfer time (ms)1 Number of Points 51 201 401 1601 SCPI over GPIB 2 64-bit floating point 2.7 6.8 13 45 32-bit floating point 1.9 4.0 6.8 23 ASCII 6.1 21 37 143 SCPI over 1 Gbps LAN (Socket) 2 REAL 64 1.4 1.4 1.4 2.0 REAL 32 1.4 1.4 1.4 1.6 ASCII 4.3 13 23 86 SCPI over 1 Gbps LAN (SICL-LAN) 2 REAL 64 1.9 1.8 1.8 2.6 REAL 32 2.0 1.9 1.9 2.1 ASCII 2.8 4.3 6.4 20 SCPI over USB (SICL-USB) 2 REAL 64 2.3 2.5 2.5 3.0 REAL 32 2.3 2.3 2.5 2.7 ASCII 2.3 5.2 8.2 28 SCPI over GPIB/USB (82357B) REAL 64 5.9 9.5 15 44 REAL 32 5.3 7.1 9.6 25 ASCII 36 133 265 1061 1. Supplemental performance data. Data transfer time varies depending on the type of PC and control software. 2. Transferred LogMag S11 data, using :CALC{ch}:MEAS{tr}:DATA:FDAT?.
The section provides test input/output performance without calibration by the E5080A. Table 46. Test set input/output performance Table 47. Option E5092A-020 port performance
50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz
2 GHz to 3 GHz
3 GHz to 8 GHz
50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
10 GHz to 20 GHz
1.3 GHz to 4 GHz
- SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
- SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
Table 48. Option E5092A-020 port performance (continued)
50 MHz to 100 MHz
100 MHz to 2 GHz
3 GHz to 4 GHz
4 GHz to 6 GHz
6 GHz to 8 GHz
10 GHz to 14 GHz
14 GHz to 18 GHz
Condition: Enviroment Temperature +23 °C ± 3 °C and Internal DC source: ≤ 100 mA (Sum of 4 channels), no heat source and no wall close to the unit.
50 M to 4 GHz
4 G to 12 GHz
12 G to 20 GHz
50 MHz to 500 MHz
500 MHz to 1 GHz
1 GHz to 2 GHz
2 GHz to 6 GHz
6 GHz to 10 GHz
- SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
- SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)
- This specification is defined when all ports are terminated with a 50 ohm load.
Table 49. Control line Table 50. DC source Table 51. Operating storage environment Table 52. Non-operating storage environment Table 53. Front panel information
- Input voltage will be clipped at about ± 5.2 V when over this range.
- The output voltage can be set in this range.
- The output voltage resolution becomes effective between 0 V to 5.2 V.
28 | Keysight |E5080A ENA Series Network Analyzer, E5092A Configurable Multiport Test Set - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2015, 2016 Published in USA, January 19, 2016 5992-0291EN www.keysight.com For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-01-01-16) www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2008 Quality Management System myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with a three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939.