E5072A KEYSIGHT | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 23
Technical content
E5072A ENA Series Network Analyzer 30 kHz to 4.5/8.5 GHz Data Sheet
2 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Options This document provides technical specifications for the E5072A ENA network analyzer. E5072A-245 2-port with configurable test set, 30 kHz to 4.5 GHz E5072A-285 2-port with configurable test set, 30 kHz to 8.5 GHz Calibration kits and ECal modules This E5072A data sheet provides technical specifications for the following calibration kits and ECal modules. For models not listed in this data sheet, please download the free Uncertainty Calculator from www.keysight.com/find/na_calculator to generate the curves for your calibration kit. 85032F Calibration kit 85033E Calibration kit 85092C Electronic calibration (ECal) module 85093C Electronic calibration (ECal) module Definitions Specification (spec.): Warranted performance. All specifications apply at 23 ºC (± 5 ºC), unless otherwise stated, and 90 minutes after the instrument has been turned on. Specifications include guard bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Typical (typ.): Describes performance that will be met by a minimum of 80% of all products. It is not guaranteed by the product warranty. Supplemental performance data (SPD): Supplemental performance data represents the value of a parameter that is most likely to occur; the expected mean or average. It is not guaranteed by the product warranty. General characteristics: A general, descriptive term that does not imply a level of performance. Boundary Conditions In this data sheet, boundary conditions are given for the specifications. For example, sys - tem dynamic range is 98 dB with the following boundary conditions. Frequency: 10 MHz, IF bandwidth: 3 kHz If the same boundary conditions fall under more than one category in a table, apply the best value.
3 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Corrected System Performance The specifications in this section apply to measurements made with the Keysight Technologies, Inc. E5072A network analyzer under the following conditions: – No averaging applied to data – Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the calibration temperature – Response and isolation calibration performed Description Specification SPD System dynamic range at test port 1, 2 (IF Bandwidth = 3 kHz) 30 to 300 kHz 65 dB (Typ.) 300 kHz to 10 MHz 82 dB
10 MHz to 6 GHz 98 dB
6 to 8.5 GHz 92 dB (IF Bandwidth = 10 Hz) 30 to 300 kHz 90 dB (Typ.) 100 dB 300 kHz to 10 MHz 107 dB 115 dB
10 MHz to 6 GHz 123 dB 130 dB
6 to 8.5 GHz 117 dB 128 dB Extended dynamic range at direct receiver access input (IF Bandwidth = 10 Hz) 30 to 300 kHz 128 dB 300 kHz to 10 MHz 134 dB
10 MHz to 3 GHz 151 dB
3 GHz to 6 GHz 147 dB
6 to 8.5 GHz 145 dB 1. The test port dynamic range is calculated as the difference between the test port rms noise floor and the source maximum output power. The effec - tive dynamic range must take measurement uncertainty and interfering signals into account. 2. Typical performance might not be met from 60 to 70 kHz. 3. The direct receiver access input system dynamic range is calculated as the difference between the direct receiver access input noise fl oor and the source maximum output power. The effective dynamic range must take measurement uncertainties and interfering signals into account.
Figure 1. System dynamic range (specification and actual measurement data example, IF bandwidth 10 Hz)
Figure 2. Transmission uncertainty (Specification) Figure 3. Re flection uncertainty (Specification) with < 1 °C deviation from calibration temperature, isolation calibration performed.
30 KHz to 300 kHz
300 KHz to 10 MHz
10 MHz to 3 GHz
3 GHz to 6 GHz
with < 1 °C deviation from calibration temperature, isolation calibration is not performed. Figure 4. Transmission uncertainty (Specification) Figure 5. Re flection uncertainty (Specification)
with < 1 °C deviation from calibration temperature, isolation calibration performed. Figure 6. Transmission uncertainty (Specification) Figure 7. Reflection uncertainty (Specification)
9 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Uncorrected System Performance User correction: OFF, System error correction: ON Specification (dB) Description 30 to 300 kHz 300 kHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz Directivity 15 (Typ.) 25 20 15 Source match 15 (Typ.) 25 20 15 Specification (dB) Description 30 to 300 kHz 300 kHz to 100 MHz 100 MHz to 2 GHz 2 to 4.5 GHz 4.5 to 8.5 GHz Load match 14 (Typ.) 22 15 11 8 Test Port Output (Source) Test port output frequency Description Specification Typical Frequency range 1 Option 245 Option 285 30 kHz to 4.5 GHz 30 kHz to 8.5 GHz Resolution 1 Hz Source Stability Standard Option 1E5 ± 7 ppm (5 to 40 ºC) ± 1 ppm (5 to 40 ºC) ± 0.5 ppm/year CW accuracy Standard Option 1E5 ± 7 ppm ± 1 ppm Test port output power 2 Description Specification SPD Nominal Power (Preset Power) 0 dBm Range 3 30 to 300 kHz 300 kHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz –85 to 10 dBm –85 to 16 dBm –85 to 12 dBm –85 to 10 dBm Max Leveled Power 30 to 300 kHz 300 kHz to 1 GHz 1 to 3 GHz 3 to 8.5 GHz 16 dBm 20 dBm 18 dBm Slope from 18 dBm (3 GHz) to 12 dBm (8.5 GHz) 1. Frequency can be set down to 9 kHz 2. Source output performance on port 1 only. Port 2 output performance is typical. 3. Power can be set from –109 to 20 dBm
10 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Description Specification Typical Resolution 0.05 dB Level Accuracy 1 At 50 MHz, 0 dBm, Absolute ± 0.65 dB (Level Flatness) 2 Stepped Sweep Mode 30 to 300 kHz 300 kHz to 8.5 GHz ± 1.0 dB ± 1.0 dB Swept Sweep Mode 30 kHz to 7 GHz 7 to 8.5 GHz ± 2.5 dB ± 3.0 dB Level Linearity (–15 dBm to Maximum Power) Stepped Sweep Mode 30 to 300 kHz 300 kHz to 8.5 GHz ± 0.75 dB ± 0.75 dB Swept Sweep Mode All frequency ± 1.5 dB Power Sweep Range 4 65 dB Test port output signal purity Description Specification Typical Harmonics (2nd or 3rd) (at 5 dBm) 30 kHz to 2 GHz 2 to 8.5 GHz < –25 dBc < –20 dBc Non-harmonic spurious (at 5 dBm) < –30 dBc 1. Power calibration using an external power meter improves level accuracy of the test port output power. 2. Level accuracy of other frequency is taken at 0 dBm, relative to 50 MHz reference unless otherwise stated. 3. Level linearity given is relative to 0 dBm. 4. Stop power may be limited by maximum power.
11 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Test Port Input Description Specification SPD Crosstalk 1 30 to 300 kHz 300 kHz to 10 MHz 6 to 8.5 GHz –90 dB (Typ.) –110 dB –120 dB –110 dB –100 dB –95 dB –120 dB –140 dB –130 dB –120 dB Test Port Noise Floor 30 to 300 kHz 300 kHz to 10 MHz 6 to 8.5 GHz –90 dBm/Hz (Typ.) –101 dBm/Hz –117 dBm/Hz –121 dBm/Hz –117 dBm/Hz –100 dBm/Hz –109 dBm/Hz –124 dBm/Hz –128 dBm/Hz –128 dBm/Hz (at 10Hz IFBW) 30 to 300 kHz 300 kHz to 10 MHz 6 to 8.5 GHz –80 dBm (Typ.) –91 dBm –107 dBm –111 dBm –107 dBm –90 dBm –99 dBm –114 dBm –118 dBm –118 dBm Compression level (at +10dBm input) Magnitude 30 to 300 kHz 300 kHz to 3 GHz 3 to 8.5 GHz 0.3 dB (Typ.) 0.2 dB 0.2 dB 0.15 dB 0.08 dB 0.1 dB Phase 30 to 300 kHz 300 kHz to 3 GHz 3 to 6 GHz 6 to 8.5 GHz 7 deg (Typ.) 5 deg 5 deg 5 deg 0.7 deg 0.3 deg 0.6 deg 1.0 deg 0.1 dB Compression Input Level 30 to 300 kHz 300 kHz to 2 GHz 2 to 6 GHz 6 to 8.5 GHz 6 dBm 16 dBm 14 dBm 10 dBm Damage Level +26 dBm or ± 35 VDC Trace Noise 2 Description Specification SPD (test port input level = maximum power in Specification) Magnitude 30 to 300 kHz, 3 kHz IFBW 300 kHz to 10 MHz, 3 kHz IFBW 10 MHz to 4.38 GHz, 70 kHz IFBW 4.38 to 8.5 GHz, 70 kHz IFBW 0.015 dB rms (Typ.) 0.003 dB rms 0.004 dB rms 0.006 dB rms 0.007 dB rms 0.0005 dB rms 0.001 dB rms 0.0012 dB rms Phase 30 to 300 kHz, 3 kHz IFBW 300 kHz to 10 MHz, 3 kHz IFBW 10 MHz to 4.38 GHz, 70 kHz IFBW 4.38 to 8.5 GHz, 70 kHz IFBW 0.1 deg rms (Typ.) 0.02 deg rms 0.035 deg rms 0.05 deg rms 0.05 deg rms 0.0025 deg rms 0.0075 deg rms 0.015 deg rms 1. Typical performance might not be met from 60 to 70 kHz.
- Accuracy of the test port input power reading is relative to –10 dBm reference input power level.
Figure 10. Dynamic Accuracy
full 2-port calibration and a 10 Hz IF bandwidth. – Insertion loss is assumed to be < 2 dB.
- Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep).
Figure 11. Group delay
14 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Front-Panel Jumpers Measurement Receiver Input (R1, R2, A, B) Description SPD Typical Noise Floor 30 kHz to 10 MHz 10 MHz to 8.5 GHz –128 dBm/Hz –145 dBm/Hz (10Hz IFBW ) 30 kHz to 10 MHz 10 MHz to 8.5 GHz –118 dBm –135 dBm Maximum Input Level (Rcvr R1/R2/A or B IN at 0.1dB typi- cal compression) 30 to 300 kHz 300 kHz to 8.5 GHz –15 dBm –10 dBm Damage Level +15 dBm or ± 16 VDC Coupler arm output Description SPD Typical Damage Level +15 dBm or 0 VDC Coupler through input Description SPD Typical Insertion Loss from Coupler Thru to Test Port Slope from 0.8 dB at 30 kHz to 3.5 dB at 8.5 GHz (0.32 x f [GHz] + 0.8) dB Damage Level +26 dBm or ± 35VDC Source output Description SPD Typical Maximum Leveled Power 30 to 300 kHz 300 kHz to 1 GHz 1 to 3 GHz 3 to 8.5 GHz +18 dBm +20.5 dBm +19 dBm Slope from +19dBm at 3 GHz to +16 dBm at 8.5 GHz Maximum Input Level (at Max Specified Output Power) 30 kHz to 8.5 GHz +20 dBm Damage Level +26 dBm or 0 VDC Reference source output Description SPD Typical Damage Level +15 dBm or 0 VDC
15 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet General Information Measurement Receiver Input (R1, R2, A, B) Description General characteristic System band width Range 10 Hz to 500 kHz Nominal settings are: 10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1 k, 1.5 k, 2 k, 3 k, 4 k, 5 k, 7 k, 10 k, 15 k, 20 k, 30 k, 40 k, 50 k, 70 k, 100 k, 150 k, 200 k, 300 k, 400 k, 500 kHz Front Panel Description Typical General characteristic Test Ports Type –N, female, 50 Ω (nominal) Front Panel Jumpers SMA, female, 50 Ω (nominal) Probe Power Connector Voltage & Maximum current 1 +15 V ± 5% (400 mA) 3 terminal connector x 2 Display Type Resolution 10.4 inch TFT color LCD with touch screen XGA (1024 x 768) USB host port Universal serial bus jack, Type A configuration; female; provides con - nection to mouse, keyboard, printer, ECal module, USB power sensor, or USB/GPIB interface Rear Panel Description Typical General characteristic External trigger input connector Type Input level Pulse width Polarity BNC, female Low threshold voltage: 0.5 V High threshold voltage: 2.1 V Input level range: 0 to +5 V ≥ 2 μsec Positive or negative External trigger output connector Type Maximum output current Output level Pulse width Polarity BNC, female 50 mA Low level voltage: 0 V High level voltage: 5 V 1 μsec Positive or negative External reference signal input connector Type Input frequency Input level
10 MHz ± 10 ppm
–3 to +10 dBm BNC, female 1. Combined load for both probe connections
16 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Rear Panel Description Typical General characteristic Internal reference signal output connector Type Output frequency Signal Type Output level Output Impedance
10 MHz ± 7 ppm
0 dBm ± 3 dB into 50 Ω BNC, female 50 Ω Internal reference signal oven connector (Option 1E5) Type Output frequency Output level
10 MHz ± 1 ppm
BNC, female Bias tee input connector Type Damage level No RF spec degradation level Over current protection ± 35 V, 1A DC 300 mA 1A (Circuit Breaker) BNC, female Video output 15-pin mini D-Sub; female; drives XGA compatible monitors GPIB 24-pin D-Sub (Type D-24), female; compatible with IEEE-488 USB host port Universal serial bus jack, Type A configuration; female; provides connection to mouse, key board, printer, ECal module, USB power sensor, or USB/GPIB interface USB (USBTMC 1) interface port Universal serial bus jack, Type B configuration (4 contacts inline); female; provides connection to an external PC; compatible with USBTMC-USB488 and USB 2.0.LA LAN 10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data rates Handler I/O port 36-pin centronics, female; provides connection to handler system Line Power Frequency Voltage VA max Power consumption
47 Hz to 63 Hz
90-264 VAC (Vpeak ≥ 120V) 350 VA max.
135 W (SPD)
Description Specification General characteristic AUX input connector Type Input Range Accuracy 1% + 1 mV for ± 1 V input 1% + 10 mV for ± 10 V input BNC, female ± 1 V or ± 10 V selectable 1. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards. 2. A third-wire ground is required.
17 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Block Diagram A B Port 1P ort 2 RCVR R1 IN SOURCE OUT CPLR THRU SOURCE OUT RCVR A IN CPLR ARM RCVR B IN CPLR ARM CPLR THRU SOURCE OUT RCVR R2 IN SOURCE OUT Source SPDT Switch Solid-state Attenuator (65 dB) Mechanical Step Attenuator (60 dB, 10 dB step) Bias-Tee Bias-Tee Mechanical Step Attenuator (60 dB, 10 dB step) REF 1 REF 2
18 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet EMC, safety, environment and compliance Description Specification EMC European Council Directive 2004/108/EC IEC 61326-1:2012 EN 61326-1:2013 CISPR 11:2009 +A1:2010 EN 55011: 2009 +A1:2010 Group 1, Class A IEC 61000-4-2:2008 EN 61000-4-2:2009 4 kV CD / 8 kV AD IEC 61000-4-3:2006 +A1:2007 +A2:2010 EN 61000-4-3:2006 +A1:2008 +A2:2010 3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz / 1V/m, 2.0 - 2.7 GHz, 80% AM IEC 61000-4-4:2004 +A1:2010 EN 61000-4-4:2004 +A1:2010 1 kV power lines / 0.5 kV signal lines IEC 61000-4-5:2005 EN 61000-4-5:2006 0.5 kV line-line / 1 kV line-ground IEC 61000-4-6:2008 EN 61000-4-6:2009 3 V, 0.15-80 MHz, 80% AM IEC 61000-4-8:2009 EN 61000-4-8:2010 30A/m, 50/60Hz IEC 61000-4-11:2004 EN 61000-4-11:2004 ICES-001:2006 Group 1, Class A AS/NZS CISPR11:2004 Group 1, Class A KN11, KN61000-6-1 and KN61000-6-2 Group 1, Class A Safety European Council Directive 2006/95/EC IEC 61010-1:2001/EN 61010-1:2001 Measurement Category I Pollution Degree 2 Indoor Use CAN/CSA C22.2 No. 61010-1-04 Measurement Category I Pollution Degree 2 Indoor Use Environment This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation” product. Do not dispose in domestic household waste. To return unwanted products, contact your local Keysight office, or for more information see http://www.keysight.com/environment/product/. Compliance Class C
19 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Analyzer environmental specifications and dimensions Description General characteristics Operating environment Temperature Error-corrected temperature range Humidity Altitude Vibration +5 °C to +40 °C 23 °C (± 5 °C) with < 1 °C deviation from calibration temperature 20% to 80% at wet bulb temperature < +29 °C (non-condensation) 0 to 2,000 m (0 to 6561 feet)
0.21 G maximum, 5 Hz to 500 Hz
–10 °C to +60 °C 20% to 90% at wet bulb temperature < +40 °C (non-condensation) 0 to 4,572 m (0 to 15,000 feet)
0.5 G maximum, 5 Hz to 500 Hz
Weight (net) Option 017: 19.8 kg Option 019: 19.0 kg Dimensions (front view)
20 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Dimensions (rear view) Dimensions (side view)
21 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Measurement Throughput Summary Measurement throughput data is supplemental performance data. Common condition for the measurement throughput data: – Analyzer display turned off with: DISP : ENAB OFF – Number of traces = 1 – Firmware version: A.01.00 Cycle time for measurement completion Sweep mode: Swept System error correction: OFF Sweep mode: Step System error correction: ON Number of Points 51 201 401 1601 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 2 ms 3 ms 4 ms 12 ms 3 ms 6 ms 10 ms 26 ms 2-port cal 2 ms 4 ms 7 ms 23 ms 4 ms 11 ms 18 ms 51 ms Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth Uncorrected 52 ms 199 ms 394 ms 1560 ms 52 ms 199 ms 394 ms 1600 ms 2-port cal 102 ms 396 ms 786 ms 3119 ms 102 ms 396 ms 786 ms 3119 ms Start 30 kHz, stop 4.5 GHz, 500 kHz IF bandwidth Uncorrected 9 ms 11 ms 12 ms 18 ms 5 ms 11 ms 16 ms 45 ms 2-port cal 16 ms 21 ms 23 ms 35 ms 10 ms 20 ms 32 ms 90 ms Start 30 kHz, stop 4.5 GHz, 1 kHz IF bandwidth Uncorrected 54 ms 203 ms 401 ms 1579 ms 54 ms 203 ms 401 ms 1579 ms 2-port cal 107 ms 405 ms 800 ms 3158 ms 107 ms 405 ms 800 ms 3157 ms Start 30 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 12 ms 16 ms 17 ms 20 ms 6 ms 11 ms 17 ms 45 ms 2-port cal 24 ms 31 ms 33 ms 38 ms 10 ms 21 ms 32 ms 90 ms Start 30 kHz, stop 8.5 GHz, 1 kHz IF bandwidth Uncorrected 55 ms 204 ms 401 ms 1579 ms 55 ms 204 ms 401 ms 1579 ms 2-port cal 108 ms 406 ms 801 ms 3157 ms 108 ms 406 ms 801 ms 3157 ms Cycle time vs. number of points Condition: Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Number of Points Sweep mode: Swept System error correction: OFF Sweep mode: Step System error correction: ON 3 1 ms 1 ms 11 2 ms 2 ms 51 2 ms 3 ms 101 2 ms 4 ms 201 3 ms 6 ms 801 7 ms 16 ms 1601 12 ms 27 ms
22 | Keysight | E5072A ENA Series Network Analyze 30 kHz to 4.5/8.5 GHz - Data Sheet Cycle time vs. IF bandwidth Condition: NOP=201, system error correction: OFF, Sweep mode: swep IF BW (Hz) Cycle Time (ms) IF BW (Hz) Cycle Time (ms) IF BW (Hz) Cycle Time (ms) IF BW (Hz) Cycle Time (ms) IF BW (Hz) Cycle Time (ms) 10 19297 100 1931 1000 195 10000 20 100000 3 15 12865 150 1288 1500 130 15000 14 150000 3 20 9649 200 1091 2000 98 20000 11 200000 3 30 6433 300 645 3000 66 30000 8 300000 3 40 4825 400 484 4000 50 40000 6 400000 2 50 3861 500 388 5000 40 50000 5 500000 2 70 2735 700 276 7000 29 70000 4 Data transfer time 1 Number of Points 51 201 401 1601 SCPI over GPIB 64-bit floating point 5 ms 15 ms 28 ms 110 ms 32-bit floating point 3 ms 8 ms 15 ms 56 ms ASCII 21 ms 78 ms 156 ms 618 ms SCPI over 100 Mbps LAN (Socket) 2 REAL 64 1 ms 1 ms 2 ms 3 ms REAL 32 1 ms 1 ms 1 ms 2 ms ASCII 13 ms 47 ms 93 ms 363 ms SCPI over 100 Mbps LAN (SICL-LAN) REAL 64 3 ms 3 ms 4 ms 6 ms REAL 32 3 ms 3 ms 4 ms 4 ms ASCII 3 ms 7 ms 11 ms 37 ms SCPI over USB REAL 64 2 ms 2 ms 2 ms 3 ms REAL 32 2 ms 2 ms 2 ms 3 ms ASCII 3 ms 9 ms 17 ms 66 ms SCPI over GPIB/USB (82357B) REAL 64 8 ms 14 ms 23 ms 77 ms REAL 32 7 ms 10 ms 14 ms 41 ms ASCII 74 ms 285 ms 568 ms 2268 ms COM Variant Type 1 ms 1 ms 1 ms 1 ms 1. Measured using a VEE Pro 7.0 program running on a 3.2 GHz Pentium 4 DELL Precision 370, Transferred complex S11 data, using : CALC{1-36} : DATA : FDAT?. 2. USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1 and IEEE 488.2 standards. 3. Measured using an E5072A VBA macro running inside the analyzer. Transferred complex S11 data
This information is subject to change without notice. © Keysight Technologies, 2013 - 2014, 2016 Published in USA, September 6, 2016 5990-8002EN www.keysight.com 23 | Keysight | E5072A ENA Series Network Analyzer 30 kHz to 4.5 /8.5 GHz – Data Sheet myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Keysight Infoline www.keysight.com/find/Inofline Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. KEYSIGHT SERVICES Keysight Services www.keysight.com/find/service Our deep offering in design, test, and measurement services deploys an industry-leading array of people, processes, and tools. The result? We help you implement new technologies and engineer improved processes that lower costs. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/e5072a For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-06-08-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System Evolving Our unique combination of hardware, software, support, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight