E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz Data Sheet
02 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Definitions All specifications apply over a 23 °C ± 5 °C range (unless otherwise stated) and 30 minutes after the instrument has been turned on. All specified and supplemental values for RF input signals are ap- plicable to sinusoidal-wave carriers unless otherwise noted. Specification (spec.): Warranted performance. Specifications include guard-bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Following supplemental information is intended to provide information that is helpful for using the instrument but that is not guaranteed by the product warranty. Typical (typ.): Describes performance that will be met by a minimum of 80% of all products. It is not guaranteed by the product warranty. Supplemental performance data (SPD): Represents the value of a parameter that is most likely to occur; the expected mean or average. It is not guaranteed by the product warranty. General characteristics or nominal (nom.): A general, descriptive term that does not imply a level of performance. It is not guaranteed by the product warranty. Description Specification RF IN connector Type-N (female), 50 ohm nominal RF IN frequency range 10 MHz to 7 GHz RF IN measurement level –20 dBm to + 20dBm (> 30 MHz) –15 dBm to +20 dBm (< 30 MHz) Input attenuator 0 to 35 dB (in 5 dB step) Input damage level AC > +23 dBm, DC > 5V Input VSWR @50 ohm 10 MHz to 30 MHz < 1.6 30 MHz to 2 GHz < 1.2 2 GHz to 3 GHz < 1.3 3 GHz to 4 GHz < 1.3 typical 4 GHz to 7 GHz < 1.5 typical Table 1-1. RF IN port RF Input Port
03 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification (E5052B) Specification (E5052B Option 011) RF IN frequency range 10 MHz to 7 GHz Measurement frequency bands 10 MHZ to 41 MHz, 39 MHz to 101 MHz, 99 MHz to 1.5 GHz, 250 MHz to 7 GHz 1 RF frequency tracking range 0.4% of carrier frequency Measurement parameters SSB phase noise [dBc/Hz], Spurious noise [dBc], Integrated rms phase deviation [deg, rad] or time jitter [s], Residual FM [Hz rms] Number of trace 1 data trace and 1 memory trace with ‘data math’ functions Measurement trigger continuous/single/hold source: internal/external/manual/bus Offset frequency range (effective) RF carrier signal > 1 GHz (> 400 MHz for wide capture mode)
1 Hz to 100 MHz
1 Hz to 40 MHz (wide capture mode)
10 Hz to 100 MHz
10 Hz to 40 MHz (wide capture mode)
RF carrier signal < 1 GHz (< 400 MHz for wide capture mode)
1 Hz to 10% of carrier frequency 10 Hz to 10% of carrier frequency
Phase noise uncertainty 2 at effective offset frequencies Offset 1 Hz to 10 Hz ± 4 dB (SPD) N/A Offset 10 Hz to 100 Hz ± 4 dB (SPD) Offset 100 Hz to 1 kHz ± 3 dB Offset 1 kHz to 40 MHz ± 2 dB (± 3 dB for wide capture range mode) Offset 40 MHz to 100 MHz ± 3 dB SSB phase noise sensitivity See Table 1-3, 1-4, 1-5, Figure 1-1, 1-2, 1-3 IF gain setting 0 dB to 50 dB in 10 dB step ( not available in wide capture mode) 0 dB to 30 dB in 10 dB step (not available in wide capture mode) Enhanced sensitivity Cross-correlation method available. Number of correlation = 1 to 10,000 See Table 1-5 and Figure 1-3 N/A Built-in LO phase noise optimization < 150 kHz (optimized for better close-in phase noise measurement) > 150 kHz (optimized for better far-out phase noise measurement) See Figure 1-4. Reference oscillator bandwidth optimization Narrow / Wide See Figure 1-5. Residual spurious response level < –80 dBc (SPD) at > 10 kHz offset frequency with correlation > 120 sec. except for 23.5 MHz ± 1 MHz and 71 MHz ± 3 MHz of carrier frequency < –65 dBc (typical) at 1 kHz to 10 kHz offset frequency Measurement time See Table 1-6 Measurement range Capture mode: Normal or Wide PN mode (Regular) RBW: Auto X-axis: Offset frequency in log scale Segment PN mode 3 RBW: 96 mHz to 25 kHz (stepped), Maximum offset frequency span: 93.2 Hz to 24.4 MHz (stepped) X-axis: Offset frequency in liner scale N/A 1. Wide capture mode is available for 250 MHz to 7 GHz only 2. Phase noise uncertainty: specified at 10 MHz and 1 GHz of carrier frequency with 0 dBm level. PN level > –60 dBc 3. Segment PN mode is available with the firmware revision 3.20 or later. In this datasheet, specification of phase noise sensitivity, residual spurious response level and measurement time are applicable to the regular PN mode. In the segment PN mode, number of these parameters depend on the measurement setting. For more detail, refer to the user’s manual. Table 1-2. Phase noise measurement performance Phase Noise Measurement
04 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet RF input frequency Offset frequency [Hz] from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M 100 M
10 MHz specification –148 –156 –166 –168 – – –
100 MHz specification –147 –156 –163 –168 –170 – –
1 GHz specification –128 –137 –144 –160 –170 –168 –169
SPD –60 –91 –116 –135 –146 –155 –171 –178 –178 –177
3 GHz specification –118 –127 –133 –149 –163 –164 –165
SPD –50 –81 –106 –127 –135 –142 –161 –175 –177 –177
7 GHz specification –111 –120 –127 –143 –157 –158 –159
SPD –43 –74 –99 –121 –129 –138 –154 –171 –174 –175 Table 1-3. SSB phase noise sensitivity (dBc/Hz) in normal capture range mode (E5052B) LO optimization: < 150 kHz, Ref. BW: narrow, correlation = 1, RF input: +5 dBm, start offset frequency: 1 Hz, measurement time = 12.9 sec RF input frequency Offset frequency [Hz] from the carrier 10 100 1 k 10 k 100 k 1 M 10 M 40 M 100 M
10 MHz specification –135 –147 –160 –160 – – –
100 MHz specification –142 –152 –154 –156 –159 – –
1 GHz specification –125 –134 –141 –157 –160 –160 –160
SPD –86 –111 –132 –143 –152 –168 –170 –170 –170
3 GHz specification –115 –124 –130 –146 –160 –160 –160
SPD –76 –101 –124 –132 –139 –158 –170 –170 –170
7 GHz specification –108 –117 –124 –140 –154 –155 –156
SPD –69 –94 –118 –126 –135 –151 –165 –170 –170 Table 1-4. SSB phase noise sensitivity (dBc/Hz) in normal capture range mode (E5052B Option 011) LO optimization: < 150 kHz, Ref. BW: narrow, correlation = 1, RF input: +5 dBm, start offset frequency: 10 Hz, measurement time = 3.3 sec Number of correlation 10 100 1,000 10,000 Improvement factor 5 dB 10 dB 15 dB 20 dB Table 1-5. SSB phase noise sensitivity improvement by correlation Stop frequency (Hz) Start frequency (Hz) 1 10 100 1 k 100k 8.8 2.2 0.28 0.04 1M 8.8 2.2 0.28 0.04 10M 10 2.5 0.32 0.04 40M 10 2.5 0.32 0.04 100M 12.9 3.3 0.41 0.05 Table 1-6. E5052B Typical measurement time (sec) for phase noise LO optimization: < 150 kHz, Ref. BW: narrow, correlation = 1, RF input: +5 dBm Measurement time (sec) = ( 0.4 (Capture range narrow) or 0.6 (Capture range wide) ) + the above value x number of correlation when applying cross-correlation function (E5052B ONLY). For E5052B Option 011, number of correlation = 1. Phase Noise Measurement – continued RF input frequency Offset frequency (Hz) from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M
1 GHz SPD – – – –108 –128 –144 –155 –160 –160
3 GHz SPD – – – –107 –119 –134 –150 –158 –158
7 GHz SPD – – – –107 –112 –126 –146 –156 –156
Table 1-3-W. SSB phase noise sensitivity (dBc/Hz) in wide capture range mode (E5052B) (SPD) LO optimization: < 150 kHz, Ref. BW: narrow, correlation = 1, RF input: +5 dBm, start offset frequency: 1 Hz, measurement time = 12.9 sec
05 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification RF frequency range 10 MHz to 7 GHz Monitoring span 15 MHz maximum with linear scale RBW 1.53 Hz to 400 kHz Measurement parameters dBm, dBV, watt, volt, dBm/Hz, dBV/Hz, watt/Hz, V/ √Hz‾ Absolute measurement uncertainty ± 2 dB typical @ –10 dBm (att. = 10 dB) Relative measurement uncertainty ± 1.5 dB (–60 dBm to –10 dBm, ratio) Residual noise floor –95 dBm typical @ RBW = 24.4 Hz Measurement trigger continuous/single/hold source: internal/external/manual/bus Table 2-1. Spectrum monitor performance Spectrum Monitor Measurement E5052B SSB-PN Sensitivity Offset frequency [Hz] SSB phase noise [dBc/Hz] –40 –50 –60 –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 –170 –180 –190 –200 1 10 100 1k 10k 100k 1M 10M 100M
10 MHz
100 MHz
1 GHz
Figure 1-1. SSB phase noise sensitivity (E5052B, SPD) (LO < 150 kHz opti - mized, +5 dBm input, start offset frequency = 1 Hz, measurement time = 12.9 sec.) E5052B-011 SSB-PN Sensitivity Offset frequency [Hz] SSB phase noise [dBc/Hz] –40 –50 –60 –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 –170 –180 –190 –200 1 10 100 1k 10k 100k 1M 10M 100M Figure 1-2. SSB phase noise sensitivity (E5052B Option 011, SPD) (LO < 150 kHz optimized, +5 dBm input, start offset frequency = 10 Hz, measurement time = 3.3 sec.) E5052B SSB-PN Sensitivity Improvement by Correlation @1 GHz Offset frequency [Hz] SSB phase noise [dBc/Hz] –40 –50 –60 –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 –170 –180 –190 –200 1 10 100 1k 10k 100k 1M 10M 100M corr. = 1 corr. = 10 corr. = 100 Figure 1-3. SSB phase noise sensitivity improvement by correlation (E5052B, SPD) (carrier 1 GHz, LO < 150 kHz optimized, +5 dBm input, start offset frequency = 1 Hz) E5052B SSB-PN Sensitivity Change by L.O. Phase Optimization @1 GHz Offset frequency [Hz] SSB phase noise [dBc/Hz] –40 –50 –60 –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 –170 –180 –190 –200 1 10 100 1k 10k 100k 1M 10M 100M < 150 kHz opt. > 150 kHz opt. Figure 1-4. SSB phase noise change by LO optimization (SPD) (carrier 1 GHz, +5 dBm input, start offset frequency = 1 Hz, reference oscillator:narrowband) E5052B SSB-PN Sensitivity Change by Reference Oscillator Bandwidth @100 MHz Offset frequency [Hz] 3 phase noise [dBc/Hz] –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 –170 1 10 100 1k 10k 100k 1M 10M 100M Narrow Wide Figure 1-5. SSB phase noise sensitivity change by reference oscillator bandwidth (SPD) (carrier 100 MHz, +5 dBm input, start offset frequency = 1 Hz, LO optimization: < 150 kHz) Phase Noise Measurement – continued
06 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification (E5052B) Specification (E5052B-011) RF frequency range 10 MHz to 7 GHz Measurement frequency bands 10 MHz to 1.5 GHz (low-band), 250 MHz to 7 GHz (high-band) Sweep parameters DC control voltage (Vc) DC supply voltage (Vs) N/A (Vc and Vs: constant) Measurement parameters Full analysis capability available for Frequency [Hz, ∆Hz, %, ppm], Tuning sensitivity (∆f/∆Vc)[Hz/V], frequency pushing (∆f/∆Vs)[Hz/V], RF power level [dBm], DC supply current [A], ‘Meter mode’ is also available. No ‘Analysis mode’. Only ‘Meter mode’ is available. Frequency [Hz], RF power [dBm], DC supply current [A] Frequency resolution 10 Hz, 1 kHz, 64 kHz Frequency uncertainty ± (frequency resolution + time-base uncertainty) RF power measurement range –20 dBm to +20 dBm (carrier 30 MHz to 7 GHz) –15 dBm to +20 dBm (carrier 10 MHz to 30 MHz) RF power resolution 0.01 dB RF power uncertainty (by peak detection) ± 0.5 dB (carrier 30 MHz to 3 GHz, > –10 dBm) ± 1 dB (other than the above) DC (Vs) current measurement range 0 to 80 mA DC (Vs) current resolution 10 μA DC (Vs) current uncertainty ± (0.2% of reading + 160 μA) Swept measurement points 2 to 1,001 N/A DC supply voltage source (Vs) output Setting range 0 to +16 V (sweep) 0 to +16 V (one point) Setting resolution 1 mV Setting uncertainty ± (0.2% of setting + 2 mV) Maximum output current 80 mA Noise level < 10 nVrms/√Hz‾ @ 10 kHz typical Output resistance < 0.3 ohm typical DC control voltage source (Vc) output Setting range –15 V to +35 V (sweep) –15 V to +35 V (one point) Setting resolution 0.1 mV Setting uncertainty ± (0.1% of (setting + 15 V) + 5 mV) (@Vc = –15 V to 0V) ± (0.1% of setting + 2 mV) (@Vc = 0 to +35 V) Maximum output current 20 mA Noise level 1 nVrms/√Hz‾ @ 10 kHz (Vc = 0 to +20V) 1.5 nVrms/√Hz‾ @ 10 kHz (Vc: otherwise) Output resistance < 50 ohm (DC) Output settling time < 20 ms @ 0.1% uncertainty Measurement trigger continuous/single/hold source: internal/external/manual/bus Table 3-1. Frequency and power measurement performance Frequency and RF Power, DC Supply Current Measurements
07 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification Target frequency range 10 MHz to 7 GHz Measurement parameters Narrowband mode Wideband mode Frequency, RF power, phase Frequency Frequency transient bandwidth Wideband Narrowband See Table 4-2. 3.125 kHz/ 25 kHz/ 200 kHz/ 1.6 MHz
25.6 MHz (> carrier 200 MHz)
80 MHz (> carrier 800 MHz)
See Table 4-2. through Table 4-8. ± (resolution + time-base uncertainty) RF power measurement Power level range Resolution Uncertainty –20 dBm to +20 dBm 0.1 dB ± 2 dB typical Phase measurement (when DUT signal is locked to a target fre - quency) Uncertainty Trace noise Stability 0.1 deg + 0.1 deg/GHz typical 0.02 deg + 0.02 deg/GHz (s) typical 10 deg/sec typical Sweep measurement time Time span Time resolution 10 μs to 10 s in 1,2,5 step (in advanced mode: maximum time span = time resolution * 10,000. up to 1000 sec.) See Table 4-2 through 4-8. in details 8 ns to 10 ms, See Table 4-2. to 4-8. in details. *Video filter time-constant: 260 ns to 16.78 ms Measurement trigger Trigger mode Trigger source External trigger polarity Video trigger Pre-trigger delay External trigger delay adjustment External trigger detection jitter continuous/single/hold internal/external/manual/bus/wide-video/narrow-video positive/negative (TTL level) positive/negative/frequency-band in/ frequency-band out –80% of time span to + 1 s 0 to 1 μs < (1 μs + time resolution) 1. Equation is based on simplified model of phase noise characteristic of local oscillator in the E5052B. Table 4-1. Transient measurement performance Transient Measurement 0.2ƒ 2.5 31+11ƒ (Hzrms/GHz),SPD√ ƒ=resolution
08 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Wideband mode Transient time span (X-axis) setting Measurement point 1251 1251 1251 1251 1251 1251 1001 1001 1001 1001 1001 1001 801 801 801 801 801 801 801 Frequency band [GHz] Frequency resolution [Hz] 0.05 to 0.15 28 k 9 k 3 k 1 k 0.1 to 0.3 56 k 19 k 7 k 2 k 0.2 to 0.6 112 k 39 k 14 k 4 k 0.3 to 0.9 168 k 59 k 21 k 7 k 0.4 to 1.2 225 k 79 k 28 k 9 k 0.5 to 1.5 281 k 99 k 35 k 12 k 0.6 to 1.8 337 k 119 k 42 k 14 k 0.8 to 2.4 450 k 159 k 56 k 19 k 1.0 to 3.0 562 k 198 k 70 k 24 k 1.2 to 3.6 675 k 238 k 84 k 29 k 1.4 to 4.2 787 k 278 k 98 k 34 k 1.6 to 4.8 900 k 318 k 112 k 39 k 1.8 to 5.4 1.012 M 357 k 126 k 44 k 2.0 to 6.0 1.125 M 397 k 140 k 49 k 2.2 to 6.6 1.237 M 437 k 154 k 54 k 2.4 to 7.2 1.35 M 477 k 168 k 59 k Table 4-2. Wideband mode frequency resolution vs. time span and frequency band Transient Measurement/Wideband Mode
09 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Measurement point 783 783 783 1564 1564 1564 1564 1251 626 626 626 626 626 626 626 626 Frequency resolution [Hz] 110 39 13.7 4.9 Table 4-5. Narrowband mode (1.6 MHz span)/frequency resolution vs. time span Time span [s] 1 m 2 m 5 m 10 m 20 m1 50 m1 0.11 0.21 0.51 1 2 5 101 Time resolution [s] 1 μ1 2 μ1 5 μ1 10 μ1 20 μ1 51 μ1 128 μ 256 μ 640 μ 1.28 m 2.56 m 6.4 m 12.8 m Measurement point 978 978 978 978 978 978 783 783 783 783 783 783 783 Table 4-6. Narrowband mode (200 kHz span)/frequency resolution vs. time span Time span [s] 10 m 20 m 50 m1 0.11 0.21 0.51 1 2 5 10 Measurement point 1222 1222 1222 1222 1222 1222 978 978 978 978 Frequency resolution [Hz] 0.21 0.08 0.03 0.01 Table 4-7. Narrowband mode (25 kHz span)/frequency resolution vs. time span Time span [s] 0.21 0.51 0.51 1 2 5 10 Time resolution [s] 65 μ1 131 μ1 328 μ1 655 μ1 1.31 m 3.3 m1 8.2 m1 Measurement point 1527 1527 1527 1527 1527 1527 1222 Frequency resolution [Hz] 0.01 3 m 1 m 0.4 m 1. Means approximately Table 4-8. Narrowband mode (3.125 kHz span)/frequency resolution vs. time span Measurement point 1251 1251 1251 1251 1251 1251 1001 1001 1001 1001 1001 1001 801 801 801 801 801 801 801 Frequency resolution [Hz] 7 k 2.5 k 879 Table 4-3. Narrowband mode (80 MHz span)/frequency resolution vs. time span Table 4-4. Narrowband mode (25.6 MHz span)/frequency resolution vs. time span Measurement point 1251 1251 1251 1251 1251 1251 1001 1001 1001 1001 1001 1001 801 801 801 801 801 801 801 Frequency resolution [Hz] 7 k 2.5 k 879 311 Transient Measurement/Narrowband Mode
10 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification RF frequency range 60 MHz to 7 GHz Effective offset frequency range 10 Hz to 40 MHz (@ > carrier 400 MHz)
10 Hz to 10% of carrier frequency
(@ < carrier 400 MHz) AM noise sensitivity See Table 5-2. Measurement uncertainty 1 ± 4 dB (100 Hz to 1 kHz offset) typical ± 2 dB (1 kHz to 1 MHz offset) typical ± 3 dB (1 MHz to 40 MHz offset) typical Spurious level < –65 dBc/Hz (at > 1 kHz offset) typical Measurement trigger continuous/single/hold source: internal/external/manual/bus Table 5-1. AM noise measurement performance AM Noise Measurement AM noise sensitivity Offset frequency (Hz) from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M E5052B start frequency = 1 Hz, measurement time = 13 s specification – – – –127 –138 –147 –150 –154 –155 typical – –103 –117 –131 –142 –151 –154 –158 –159 E5052B-011 (Option 011) start frequency = 10 Hz, measurement time = 3.3 s specification – – – –124 –135 –144 –147 –151 –152 typical – –100 –114 –128 –139 –148 –151 –155 –156 Table 5-2. AM noise sensitivity [dBc/Hz] correlation = 1, RF input: 0 dBm, > 400 MHz 1. AM noise measurement uncertainty: specified at 1 GHz of carrier frequency with 0 dBm level. AM level > –60 dBc
11 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification Baseband input connector BNC, 50 ohm nominal, AC coupled Measurement frequency range 1 Hz to 100 MHz (E5052B)
10 Hz to 100 MHz (E5052B Option 011)
Measurement parameters dBV/Hz, dBm/Hz, V/√Hz Measurement level range < +5 dBm Baseband input damage level > +23 dBm, > 35 V DC Noise floor level See Table 6-2. Measurement uncertainty 1 ± 4 dB (< 1 kHz) SPD ± 2 dB (> 1 kHz) typical Measurement trigger continuous/single/hold source: internal/external/manual/bus Table 6-1. Baseband noise measurement performance Baseband Noise Measurement BB noise floor Baseband frequency [Hz] 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M 100 M E5052B start frequency = 1 Hz, measurement time = 13 s specification – – – –151 –158 –163 –160 –160 –156 –156 typical –119 –132 –145 –155 –162 –167 –164 –164 –160 –160 E5052B Option 011 start frequency = 10 Hz, measurement speed = 3.3 s specification – – – –148 –155 –160 –157 –160 –156 –156 typical – –129 –142 –152 –159 –164 –161 –164 –160 –160 Table 6-2. Baseband noise floor [dBm/Hz] correlation = 1, baseband input: 0 ohm terminated Description Specification Frequency uncertainty ± 5 Hz at 10 MHz (± 0.5 ppm) Frequency temperature coefficient < 0.5 ppb/degC Frequency aging rate < 0.5 ppb/day 24 hours after a cold start for < 30 days continuous operation 1. Baseband measurement uncertainty: specified at > –60 dBm level. Table 7-1. Internal timebase (OCXO) performance Internal Timebase
12 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Supplemental information (nominal) Connectors/terminals RF IN Type-N (female), 50 ohm Baseband IN BNC (female), 50 ohm, AC coupled DC power BNC (female), DC control BNC (female), 50 ohm RF1/RF2, IN/OUT SMA (female), 50 ohm See the simplified block diagram. USB 2 ports (designed for USB2.0) Probe DC power output +15 V, 150 mA maximum –12.6 V, 150 mA maximum Ground terminal 1 Display 10.4 inch TFT color LCD with touch screen 1,024 x 768 resolution 1 Table 8-1. Front panel information Description Supplemental information (nominal) External trigger input port Connector Input signal level Trigger pulse width BNC (female) TTL level, (0 V to +5 V) Threshold Low: 0.5 V, High: 2.1V > 2 μs Trigger polarity positive/negative edge selectable Auxiliary output port Connector Output signal level pulse width BNC (female) TTL level, L: 0 V, H:= +5 V, 50 mA max. 1 μs Reference output port Connector Output frequency Output level Output signal waveform BNC (female), 50 ohm same as timebase 2.5 dBm ± 2 dB typical Sinusoidal wave Reference input ports Connector Input frequency Input signal level (Ref In 1, Ref In 2) BNC (female), 50 ohm
10 MHz ± 10 Hz
24 BIT I/O parallel port
USB (USBTMC 2) port LAN port Video output port 36-pin D-sub (female) connector to a handler system TTL level, 8-bit I/O 16-bit Out 24-pin D-sub (female) connector (compatible with IEEE-488) 4 type-A (compatible with USB 2.0) 1 type-B (compatible with USBTMC-USB488 and USB 2.0) 10/100 base-T ethernet 15-pin mini D-sub (female) connector drives XGA compatible monitors AC power line (a third-wire ground is required) AC frequency AC voltage AC power
47 Hz to 63 Hz
90 to 132 V, or 198 to 264 V (automatically selected)
500 VA maximum
- USB test and measurement class (TMC) interface that communicates over USB, complying with the IEEE-488.1 and IEEE-488.2 standards. Table 8-2. Rear panel information General Information
13 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet General Information – continued Description Supplemental information (nominal) Operating environment Temperature Humidity Altitude Vibration +10 degC to +40 degC RH 20% to 80% at wet bulb temp.< 29 degC (non-condensing) 0 to 2,000 m (0 to 6,561 feet)
0.21 G maximum, 5 Hz to 500 Hz
Non-operating storage environment Temperature Humidity Altitude Vibration Instrument dimensions Weight (NET) –10 degC to +60 degC RH 20% to 90% at wet bulb temp.< 40 degC (non-condensing) 0 to 4,572 m (0 to 15,000 feet)
0.5 G maximum, 5 Hz to 500 Hz
See Figure 8-1, 8-2, 8-3. 24.5 kg Table 8-3. Analyzer environment and dimensions EMC European Council Directive 2004/108/EC IEC 61326-1:2012 EN 61326-1:2013 CISPR 11:2009 +A1:2010 EN 55011: 2009 +A1:2010 Group 1, Class A IEC 61000-4-2:2008 EN 61000-4-2:2009 4 kV CD / 8 kV AD IEC 61000-4-3:2006 +A1:2007 +A2:2010 EN 61000-4-3:2006 +A1:2008 +A2:2010 3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz / 1V/m, 2.0 - 2.7 GHz, 80% AM IEC 61000-4-4:2004 +A1:2010 EN 61000-4-4:2004 +A1:2010 1 kV power lines / 0.5 kV signal lines IEC 61000-4-5:2005 EN 61000-4-5:2006 0.5 kV line-line / 1 kV line-ground IEC 61000-4-6:2008 EN 61000-4-6:2009 3 V, 0.15-80 MHz, 80% AM IEC 61000-4-8:2009 EN 61000-4-8:2010 30A/m, 50/60Hz IEC 61000-4-11:2004 EN 61000-4-11:2004 ICES-001:2006 Group 1, Class A AS/NZS CISPR11:2004 Group 1, Class A Table 8-5. EMC, safety, environment and compliance Table 8-4. LXI compliance LXI Class C (only applies to units that are shipped with firmware revision A.03.10 or later)
14 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet KN11, KN61000-6-1 and KN61000-6-2 Group 1, Class A Safety European Council Directive 73/23/EEC, 93/68/EEC IEC 61010-1:2001 EN 61010-1:2001 IEC60825-1:1994 Measurement category I Pollution degree 2 Indoor use Class 1 LED CAN/CSA C22.2 61010-1-04 Measurement category I Pollution degree 2 Indoor use WEEE European Council Directive 2002/96/EC
16 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description General characteristics Graphical user interface The analyzer employs a graphical user interface based on Windows ® OS. There are three ways to operate the instrument manually; you can use a hard key interface, a touch-screen interface, or a mouse interface. Limit-line test Define the test limit that appears on the display for pass/fail testing. Defined limits may be any combination of horizontal or sloping lines and discrete data points. Data storage Internal removable SSD File sharing Screen hard copy Store and recall instrument states and trace data on internal removable solid state drive. Instrument states include all control settings and memory trace data. Files on user disk drive (F:) can be accessed from an external Windows PC through LAN or USB (USB-TMC) Print-outs of instrument data are directly produced on a printer via USB. Automation Built-in VBA® Controlling via GPIB or USB Controlling via USBTMC LAN Applications can be developed in a built-in VBA (Visual Basic for Applications) language. The GPIB interface operates with IEEE488.2 and SCPI protocols. The instrument can be controlled by a GPIB external controller. The instrument can control external devices using a USB/GPIB interface. The USB interface operates with USBTMC and SCPI protocols. The instrument can be controlled by an external PC using the USB interface with a USB cable. (10/100 base-T) Telnet, SICL-LAN Table 9-2. Data processing capabilities Description General characteristics Measurement functions RJ (random jitter), PJ (periodic jitter) frequency, PJ decomposition with auto-trend correction Measurement parameters RJ: rms, PJ: frequency, rms, p-p, δ-δ, TJ (total jitter): p-p, jitter trend (phase deviation waveform), jitter histogram Jitter spectrum analysis range 1 Hz to 100 MHz (E5052B), 10 Hz to 100 MHz (E5052B Option 011) Table 9-3. E5001A SSA-J precision clock jitter analysis software Description General characteristics Measurement windows Up to 6 windows, and 1 user definable window User definable window 8 data traces and 8 memory traces Trace functions Data traces Trace math Title Auto scale Statistics Display current measurement data and/or memory data Addition, subtraction, multiplication, or division of trace data Add customized title to each measurement window Titles are printed on hard copies of displayed measurements. Automatically selects scale resolution and reference value to vertically center the trace. Calculates and displays mean, standard deviation, and peak-to-peak deviation of the trace. Marker functions Data markers Marker search Marker-to Searching range Tracking 10 independent markers per trace. Reference marker available for “delta marker” operation. Maximum value, minimum value, peak, peak-left, peak-right, target, target-left, target-right, multi-peak and band markers with user-definable bandwidth value. Set, start, stop, center to active marker stimulus value. Set reference to active marker response value. User definable Performs marker search continuously or on-demand. Table 9-1. Display functions (windows and traces) Display Funtions Data Processing Capabilities Optional Application Software
17 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Performance characteristics RF input port Input connector Frequency range Input level Carrier search range APC-3.5 (female), 50 ohm nominal (E5053A input)
10 MHz to 3 GHz (E5052B RF IN port)
3 to 26.5 GHz (E5053A Input port) 3 to 10 GHz frequency band: fundamental mixing 9 to 26.5 GHz frequency band: third harmonics mixing –15 to +20 dBm (10 MHz to 3 GHz, E5052B RF IN port) –30 to +10 dBm (3 to 10 GHz frequency band) –20 to +5 dBm (9 to 26.5 GHz frequency band) –10 to +10 dBm (3 to 10 GHz frequency band) –10 to +5 dBm (9 to 26.5 GHz frequency band) Phase noise measurement 1 SSB phase noise sensitivity Frequency tracking range See Table 10-2, Figure 10-2 and Figure 10-3. 1.8 MHz (< 4.9 GHz carrier in 3 to 10 GHz frequency band) 2.8 MHz (> 4.9 GHz carrier in 3 to 10 GHz frequency band) 1.3 MHz (< 10 GHz carrier in 9 to 26.5 GHz frequency band) 2.6 MHz (> 10 GHz carrier in 9 o 26.5 GHz frequency band) Spectrum monitor measurement Frequency span RBW (resolution bandwidth) Level uncertainty
15 MHz maximum
1.53 Hz to 400 kHz
± 4 dB Frequency & RF power measurement Frequency measurement resolution RF power measurement uncertainty
10 Hz, 1 kHz, or 64 kHz
± 2 dB (10 MHz to 3 GHz, E5052B RF IN port) ± 3 dB (low band: 3 to 10 GHz) ± 4 dB (high band: 9 to 26.5 GHz) Power uncertainty can be improved by applying the ‘user power cal.’ function. Transient measurement Wideband frequency range Narrowband frequency range RF power measurement uncertainty
50 MHz to 3 GHz (E5052B RF IN port)
500 MHz (E5053A Input port)
3.125 kHz, 25 kHz, 200 kHz, 1.6 MHz, 25.6 MHz, or 80 MHz ± 2 dB (10 MHz to 3 GHz, E5052B RF IN port) ± 3 dB (low band: 3 to 10 GHz) ± 4 dB (high band: 9 to 26.5 GHz) Power uncertainty can be improved by applying the ‘user power cal.’ function. 1. Segment PN mode is available with the firmware revision 3.20 or later. In segment PN mode, offset frequency range is limited up to 99.9 MHz when fre - quency range is 9 to 26.5 GHz frequency band. Table 10-1. System performance characteristics E5053A microwave downconverter E5052B signal source analyzer
3 GHz
26.5 GHz
(correlation) LO in IF out IF in LO out LO out IF in IF out LO in RF in RF out RF out RF in RF RF IF amp. IF amp. D-PLL ADC ADC D-PLL FFT FFT Display System Performance with the E5053A Microwave Downconverter The system performance is the combination of the E5052B SSA and the E5053A microwave downconverter. All data is typical performance. Figure 10-1. E5053A with E5052B simplified block diagram
18 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Input frequency Offset frequency (Hz) from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M 100 M
3 GHz –48 –79 –99 –124 –135 –137 –153 –164 –167 –167
10 GHz –38 –72 –91 –116 –124 –128 –147 –156 –160 –160
18 GHz –33 –66 –85 –110 –121 –125 –141 –150 –154 –154
26.5 GHz –30 –63 –82 –107 –118 –122 –138 –147 –151 –151
Table 10-2. System SSB phase noise sensitivity (dBc/Hz) in normal capture range mode (E5053A + E5052B) (SPD) 0 dBm input, start offset frequency = 1 Hz, correlation = 1, LO optimization: < 150 kHz, measurement time = 13 sec System Performance with the E5053A Microwave Downconverter – continued
26.4 GHz
18 GHz
16.2 GHz
Figure 10-3. Measurement samples for the ultra-low noise N5507A LO Input frequency Offset frequency (Hz) from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M Table 10-2-W. System SSB phase noise sensitivity (dBc/Hz) in wide capture range mode (E5053A+E5052B) (SPD) 0 dBm input, start offset frequency = 1 Hz, correlation = 1, LO optimization: < 150 kHz, measurement time = 13 sec AM noise sensitivity Offset frequency (Hz) from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M E5052B start frequency = 1 Hz, measurement time = 13 s 3 to 10 GHz – – 100 –110 –117 –127 –130 –137 –137 –137 10 to 26.5 GHz – –100 –110 –117 –127 –129 –129 –129 –129 E5052B-011 (Option 011) start frequency = 10 Hz, measurement time = 3.3 s 3 to 10 GHz – –97 –107 –114 –124 –127 –134 –134 –134 10 to 26.5 GHz – –97 –107 –114 –124 –126 –126 –126 –126 Table 10-2-A. System AM noise sensitivity (dBc/Hz) (E5053A + E5052B) (SPD) correlation = 1, RF input: 0 dBm, > 400 MHz E5052B + E5053A System SSB-PN Sensitivity (typical) Offset frequency [Hz] SSB phase noise [dBc/Hz] –20 –30 –40 –50 –60 –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 –170 –180 1 10 100 1k 10k 100k 1M 10M 100M
10 GHz
(SPD) Figure 10-2. System phase noise sensitivity (E5053A + E5052B) (SPD)
19 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet System Performance with the E5053A Microwave Downconverter – continued Figure 10-4. System set-up for harmonic mixers (E5053A + E5052B) Mixer model Frequency band N 11970A 26.5 to 40 GHz 8 11970Q 33 to 50 GHz 10 11970U 40 to 60 GHz 10 11970V 50 to 75 GHz 14 11970W 75 to 110 GHz 18 Table 10-2-H. Frequency band example of phase noise measurement with mmW harmonic mixers About “mmW application”: Phase noise measurements above 26.5 GHz can be done by using external harmonic mixers (such as Keysight Technologies, Inc. 11970 series) and a power divider (splitter) with E5053A LO and IF terminals. The E5052B’s mmW application software sets up ap - propriate LO frequencies for the harmonic mixers.
20 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet Description Specification RF Input port Input connector Frequency range Input level Input damage level APC-3.5 (female), 50 ohm nominal 3 GHz to 26.5 GHz
3 GHz to 10 GHz (fundamental mixing)
9 GHz to 26.5 GHz (third harmonics mixing) < +10 dBm ( 3 GHz to 10 GHz band) < +5 dBm (9 GHz to 26.5 GHz band) > +23 dBm LO outputs Output connector Output frequency Frequency resolution Output power LO spurious SMA (female), 50 ohm nominal
3 GHz to 10 GHz
50 MHz
10 dBm to 16 dBm (3 GHz to 6 GHz) 10 dBm to 15 dBm (6 GHz to 10 GHz) < –55 dBc (offset frequency > 300 Hz) typical IF inputs Input connector Frequency range Maximum input level IF gain Noise floor Mixer bias current SMA (female), 50 ohm nominal
250 MHz to 1,250 MHz
0 dB to 35 dB in 5 dB step < –162 dBm/Hz (SPD) –10 mA to +10 mA Table 10-3. E5053A front ports Description Supplemental information (nominal) External reference signal input port Input connector Input frequency Input level BNC (female), 50 ohm nominal
10 MHz ± 10 Hz typical
-6 dBm to 6 dBm typical Internal reference signal output port Output connector Output frequency Output level USB port BNC (female), 50 ohm nominal
10 MHz ± 50 Hz typical
2.5 dBm ± 3 dB typical type-B (female), provides connection to E5052A/B AC power Line (a third -wire ground is required) AC frequency AC voltage AC power
90 V to 132 V, or 198 V to 264 V
(automatically selected)
120 VA maximum
Table 10-4. General information Description Supplemental information (nominal) Operating environment Temperature Humidity Altitude Vibration +10 degC to +40 degC RH 20% to 80% at wet bulb temp. < 29 degC (non-condensing) 0 to 2,000 m (0 to 6,561 feet) Non-operating storage environment Temperature Humidity Altitude Vibration Instrument dimensions Weight (NET) –10 degC to +60 degC RH 20% to 90% at wet bulb temp. < 40 degC (non-condensing) 0 to 4,572 m (0 to 15,000 feet) See Figure 10-6, 10-7, 10-8. 11 kg Table 10-5. Analyzer environmental and dimensions E5053A Microwave Downconverter Specifications and General Information Summary
21 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet RF input frequency Offset frequency [Hz] from the carrier 1 10 100 1 k 10 k 100 k 1 M 10 M 40 M 100 M
3 GHz specification – – – –110 –116 –113 –127 –140 –140 –140
typical –49 –79 –94 –114 –120 –117 –131 –144 –144 –144
6 GHz specification – – – –104 –110 –109 –123 –140 –140 –140
Typical –43 –73 –88 –108 –114 –113 –127 –144 –144 –144
10 GHz specification – – – –100 –103 –102 –119 –140 –140 –140
typical –39 –69 –84 –104 –107 –106 –123 –144 –144 –144 Table 10-6. E5053A LO phase noise performance (dBc/Hz) E5053A Microwave Downconverter Specifications and General Information Summary – continued Web Sources Visit our Signal Source Analyzer Web site for additional product information and literature. http://www.keysight.com/find/ssa Phase noise measurements; http://www.keysight.com/find/phasenoise Jitter measurements; http://www.keysight.com/find/jitter RF and microwave accessories http://www.keysight.com/find/ mta E5053A LO Phase Noise (typical) Offset frequency [Hz] SSB phase noise [dBc/Hz] –20 –30 –40 –50 –60 –70 –80 –90 –100 –110 –120 –130 –140 –150 –160 1 10 100 1k 10k 110k 1M 10M 100M
6 GHz
Figure 10-5. E5053A LO phase noise (typical) Figure 10-6. Front view Figure 10-7. Rear view Figure 10-8. Side view
www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/ssa www.keysight.com/find/phasenoise www.keysight.com/find/jitter www.keysight.com/find/mta From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939.
1939 THE FUTURE
For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-02-10-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System 22 | Keysight | E5052B Signal Source Analyzer 10 MHz to 7 GHz, 26.5 GHz, or 110 GHz - Data Sheet This information is subject to change without notice. © Keysight Technologies, 2013 - 2016 Published in USA, March 23, 2016 5989-6388EN www.keysight.com