DAQ6510 TEKTRONIX | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 40

Technical content

The DAQ6510 is a precision data acquisition and logging system that creates a new level of simplicity compared with the often complicated configuration and control found in many stand-alone solutions. A large 5-inch (12.7 cm) multi-touch display will guide users through set-up, data visualization, and analysis, removing the necessity of a PC and custom software for many applications. If you prefer or require a PC, a complement of IVI and Labview drivers and the Keithley KickStart Instrument Control Software are available to help with any application. Using Keithley’s newest 6½-digit multimeter technology, you will find even greater accuracy, functionality, and speed. Two-year accuracy specifications are included, so you can be assured your critical measurements remain accurate and fully traceable. With 12 plug-in switching and control modules available and two module slots, you can build a test system that can measure or control up to 80 devices-under-test (DUTs) in a multiplexing configuration. With two 6×8 matrix modules, a total of 96 crosspoints, the DAQ6510 can test one or multiple ICs. For high volume production testing, select a solid state multiplexer module to maximize throughput with 800 channels/s scanning speed and minimize downtime compared with fixed life, electro-mechanical relays. A number of modules have cold junction compensation, temperature references for thermocouple temperature monitoring of devices during environmental testing, and HALT and HASS accelerated life testing. Even switch signals from wireless devices with RF modules having a maximum switching frequency of 3.5 GHz. Key Features

  • Large, 5-inch (12.7 cm) multi-touch capacitive touchscreen with graphical display
  • 2-year specified, full featured, traceable 6½-digit multimeter with 0.0025% DCV (1 V, 10 V ranges) basic accuracy
  • Standard LAN/LXI and USB-TMC communication interfaces
  • Optional interfaces include GPIB, RS-232, and TSP-Link® Technology
  • 12 different switch, RF, and control modules to connect to as many as 80 DUTs in one test setup
  • Up to 80 2-pole channels of thermocouple, RTD, or thermistor temperature measurements
  • Scanning speeds as high as 800 channels/second with the solid state relay module
  • Front panel jacks for stand-alone DMM operation
  • KickStart coding-free instrument control software with a data acquisition and logging application
  • Three-year warranty Visualize, Execute, and Debug Your Test Setup in Minutes Build a test sequence and define the required measurements from one setup screen. First, select the channels that will be used. Next, assign a measurement DAQ6510 Data Acquisition and Logging, Multimeter System Datasheet

TEK.COM2 function to a channel or a group of channels. Then define both the time interval between measurements and scans if multiple scans are required. That is the minimum set of parameters needed to run a test. There are additional optional parameters to adjust the measurements and the switching to meet a specific application. One option eases the identification of channels by allowing the creation of a name for each channel. Do all this from two menus and see – on one screen – how each channel is configured. The touchscreen, graphic display shows the channel setup and the measurement programmed for each channel. To test channels during setup, manually close channels on the setup screen. The measurement on that channel is displayed next to the closed channel. Thus, channel performance can be verified during set up or debugged if the measurement is out of the expected range before the test is executed. Maximize the Range of Tests with an Extensive Array of Plug-In Switch Modules Build a DAQ6510 test system with a selection of 12 plug-in switch modules with a wide range of functionality. Select from high-density, 40-channel multiplexer modules, a 6×8 matrix module, and multiplexer modules with digital I/O and analog output. Also select a control module and two RF modules with 2 GHz and 3.5 GHz bandwidths. Thus, the DAQ6510 with its 12 switch modules provides an extensive range of flexibility to build a wide range of data acquisition or test systems. Channels can be configured for any DMM measurement. For example, measure the temperature of devices in an environmental chamber with thermocouples using the plug-in switch modules that have automatic cold reference junction compensation for thermocouple measurements. Measure up to 80 devices with two switch modules. Also measure other parameters in addition to temperature. The low frequency, multiple pole switch modules can allow measurement of voltage, 2-wire- and 4-wire resistance, and frequency. In addition, two modules have current measurement channels. Series 7700 Plug-in Switch Modules.

TEK.COM 3 DAQ6510 Data Acquisition and Logging, Multimeter System User-installable communication accessory. Provides connectivity options for GPIB, TSP-Link, or RS-232 and each with 6 programmable digital I/O ports DAQ6510 Touchscreen Display Front Panel Store test results and screen images quickly via the USB

2.0 Host Port

Lower swipe screen speeds up access to often-used features such as function selection, statistics, and plotting 5-inch (12.7 cm) touchscreen allows access to more setup, monitoring, and analysis data One button access for easier navigation Front/rear input selector pushbutton with LED to easily identify position Online HELP is just a button press away DAQ6510 Rear Panel Slots for 2 plug-in switch modules provide flexibility for numerous switching and control configuration LAN / LXI Interface BNC input/output triggers for external control and synchronization USB-TMC Interface Apps key for quick access to customized programs Front panel inputs allow you to use the DAQ6510 as a digital multimeter Line power inlet

TEK.COM4 Test a Wide Range of Devices and Components with 6½-digit DMM High Performance Measurements Make your measurements with instrument-grade quality. The DAQ6510 is a fully-featured 6½-digit DMM. The chart below shows all of the DAQ6510’s 15 measurement functions and ranges. All these measurement functions can be used with the various plug-in switch modules. Furthermore, a channel can be programmed for enhanced measurement features to address requirements for tighter tolerance or higher speed. For maximum accuracy, measurements can be filtered and integrated over a range of time intervals. Low resistance readings can be improved with a thermal offset compensation measurement technique. The DAQ6510 excels in low resistance measurement with 1 µ Ω sensitivity. For maximizing measurement speed, make measurements as short as 0.0005 power line cycles (PLCs) or 8.3 µs for 60 Hz power lines (10 µs for 50 Hz power lines). To analyze transient signals, use the DAQ6510’s digitizing function to sample a voltage or current waveform at up to 1M sample/s. DAQ6510 MEASUREMENT CAPABILITIES DC Voltage AC Voltage DC Current AC Current Resistance (2-wire) Resistance (4-wire) Capacitance Period Frequency Diode Digitize Voltage Digitize Current 100 nV 100 nV Logarithmic Scale Linear Scale 10 pA 100 pA 1 µ/uni03A9 0.1 pF –200°C –80°C –200°C 1820°C 150°C 850°C 3.3 µs 3 Hz 10 µV 10 nA 1010 V

750 VRMS

10.1 A 10.1 A

120 M/uni03A9

10 µ/uni03A9 120 M/uni03A9 120 µF 333 ms 300 kHz 1010 V 10 µV 12 V (Selectable: 10 µA / 100 µA / 1 mA / 10 mA source) 10.1 A 1 f 1 p 1 n 1 µ 1 m 1 1 k 1 M 1 G Thermocouple Thermistor 2-, 3-, 4-Wire RTD –500° 0° 500° 1000° 1500° 2000° DAQ6510’s 15 measurement functions and ranges.

TEK.COM 5 DAQ6510 Data Acquisition and Logging, Multimeter System Monitor Any Temperature Range and Select Optimal Accuracy Using a Wide Range of Temperature Sensors The DAQ6510 can process signals from any of eight thermocouple types to test over an extreme range of temperatures, –200°C to 1820°C. For more accurate temperature readings over a more narrow temperature range, the DAQ6510 can measure temperature from –80°C to 150°C with a 2-year accuracy of 0.08°C using 2.5 kΩ, 5 kΩ, and 10 kΩ thermistors. Alternatively, the DAQ6510 can measure temperature between –200°C and 850°C using resistance temperature detectors (RTDs) with 2-wire, 3-wire, and 4-wire configurations. Using a 4-wire RTD, temperature can be measured with with an excellent 2-year accuracy of 0.06°C. The Series 7700 plug-in switch modules have multiple cold junction compensation (CJC) circuits. The CJC circuits are spaced around the module to ensure maximum accuracy for all channels measuring thermocouples. Save Valuable Time with Continuous Monitoring of Scan Test Status Never lose time with a test in which a problem such as a broken wire, a mis-wired connection, or a defective DUT can invalidate days of data. Use the scan status, swipe screen to continuously monitor channels. The status screen shows the percentage of the test that has been completed, the number of scans that have been completed, measurements on channels selected for monitoring, and whether any measurements are out of range. A quick view of the display can give an indication of the status of the test at any time during the test. In addition to an indication of measurements in an overflow condition, limit settings can be programmed for any channel. The Status screen will display an out-of-limit indication when any channel’s measurement is outside the channel’s limit settings. Maintain an Automated Test while Interacting with the Instrument Need to perform a quick check to ensure there are no problems with the measurements that the DAQ6510 is taking? Want to review the trend of the data? The DAQ6510 can be operated from the front panel while a test is in progress. Data can be viewed in a table or displayed on a plot without interrupting an automated scan sequence. Use a pinch and zoom display feature for detailed study of test data. Graphically view data during or at the end of a test. Display up to 20 channels on one plot. Quickly and easily monitor the status of a test using the DAQ6510 scan status screen.

TEK.COM6 Never Lose Data and Test Time Data can be stored to an external USB memory stick to prevent data loss in the event of interruption of line power. The DAQ6510 also has an option to automatically restart scanning after a power interruption. If main power is lost, the DAQ6510 will re-start the scan when the power is restored. The new data will be stored in a separate data buffer with time-stamped measured data so that the lost time due to the power failure can be noted. Of most importance, there is no lost test time due to the instrument sitting idle following a power outage. Save Test Time in Automated Test Systems To maximize throughput and minimize total test time, use the reed relay or the solid state relay multiplexer module to achieve scanning speeds up to 800 channels/s. Also save test time by reducing communication with the test controller. The DAQ6510 has the built-in intelligence to execute a test script, make decisions, and control other instruments. The DAQ6510 has Keithley’s Test Script Processor (TSP®) measurement and control language that the user can use to either execute existing scripts or write custom scripts to run tests. TSP® scripting allows running powerful test scripts directly on the instrument, without the need for an external PC controller. These test scripts are complete test programs based on an easy-to-use yet highly efficient and compact scripting language, LUA ( www.lua.org). Scripts are a collection of instrument control commands and/or program statements. Program statements control script execution and provide facilities such as variables, functions, branching, and loop control. This allows you to create powerful measurement applications without an integrated development environment (IDE). Test scripts can contain any sequence of routines that are executable by conventional programming languages (including decision-making algorithms), so the instrument can manage every facet of the test without the need to communicate with a PC for decision making. This eliminates delays due to GPIB, Ethernet, or USB traffic congestion and greatly improves test times. DAQ6510 with KTTI-TSP Option Device Under Test

2450 SourceMeter® 2460 SourceMeter®

TSP-Link® I/O Control (Triggering/ Handler) TSP® Script Conditional branching Calculations Pass/Fail test Prober/Handler control Datalogging/ Formatting TSP-Link® Use a TSP script and the TSP-Link interface to execute a test for instrument-to-instrument communication without interaction with a computer. Select how you want to store data in the Scan menu. Choose to automatically restart a test after a power loss.

TEK.COM 7 DAQ6510 Data Acquisition and Logging, Multimeter System TSP® technology also offers mainframe-less channel expansion. The KTTI-TSP is a user installable accessory card offering connectivity to TSP-Link technology. This channel expansion bus allows connecting multiple DMM6510’s, other TSP-enabled instruments, and any LAN-based instruments together to form a tightly-synchronized instrument system. Connection is provided with simple low cost Category 5 Ethernet cabling. The system is organized in a master-subordinate configuration, essentially allowing the connected instruments to act as one. Other Keithley TSP-enabled instruments include the 2450, 2460, and 2461 Graphical SourceMeter ® Source Measure Unit (SMU) Instruments, Series 2600B SourceMeter ® (SMU) Instruments, DMM7510, DMM6500, and Series 3700A Switch/Multimeter systems. TSP- Link technology supports up to 32 instruments, so it’s easy to scale a system to fit the requirements of an application. With the TSP-Link interface option, the KTTI-TSP, the DAQ6510 can control other Keithley TSP instruments or control any instrument with a LAN interface. For example, the DAQ6510 can send commands to a temperature chamber to operate at specific temperatures and execute a scan when the chamber has reached the programmed temperatures. In fact, the DAQ6510 can execute an automated sequence entirely without a PC thus saving test time and reducing the capital cost of the system. Using the TSP language to create a script, the user can display messages for an operator in an automated or a semi-automated test system. Data Logging Software Simplifies PC Control, Acquisition, and Analysis KickStart allows you to configure, test, and collect data from multiple instruments, including DMMs, power supplies, SMU instruments, and dataloggers. You can control up to eight instruments at the same time and retrieve millions of readings from each instrument. This makes KickStart a great solution for your datalogging needs and for capturing lots of data from transient events with a digitizing DMM. Getting insights quickly is important, so KickStart plots your data immediately and dedicates a large portion of the viewing area to the graph, while also allowing you to view and edit the most essential parameters of other instruments in your test setup. Kickstart also includes comparison tools to allow you to plot and overlay data from the run history of each test. KICKSTART DATA LOGGING APPLICATION Use the datalogger application to set up, program, and control the DAQ6510. Within minutes, configure channels, set measurement parameters such as limits and trigger sources, and view results in real time. TSP scripting example. Datalogger setup screen. Datalogger multi-channel plot display screen.

TEK.COM8 Key KickStart features:

  • Automate data collection from up to eight instruments
  • Replicate tests quickly using saved test configurations
  • Use statistical summaries and built-in plotting and comparison tools to quickly discover measurement anomalies and trends
  • Export data in ready-to-use formats for additional analysis or to share test updates with your colleagues Add customized applications to the DAQ6510 to enhance the instrument’s capability. The DAQ6510 includes factory-installed applications that:
  • Allow creation of a modified display to show data from multiple channels.
  • Report plug-in module relay closure counts that enable determination of when a plug-in module needs to be replaced. These examples highlight the unique ability of the DAQ6510 to run specialized applications that customize the user interface. This can significantly change the way information is displayed or even automated while running a test. Ready to Use Instrument Drivers Simplify Programming Prefer to create your own customized application software? Native National Instruments Labview ®, IVI-C, and IVI-COM drivers are available for downloading to simplify the programming process. For the DAQ6510 Labview driver, please Wide Range of PC Interfaces You can choose the interface that is most suitable for your needs. Two PC interfaces, LAN LXI and USB, on the DAQ6510 are standard. In addition, field-installable interface options include a GPIB interface, an RS-232 interface, and the Keithley TSP-Link interface. All these options include six digital I/O ports for direct instrument-instrument synchronization and communication. KTTI-TSP TSP-Link interface with 6 digital I/O ports. KTTI-RS232 RS-232 interface with 6 digital I/O ports. KTTI-GPIB GPIB interface with 6 digital I/O ports.

TEK.COM 9 DAQ6510 Data Acquisition and Logging, Multimeter System With its fron panel connections, the DAQ6510 can be used as a 6½-digit DMM. Use as a DMM for Test and Troubleshooting The DAQ6510 has front panel input terminals that enable use of the DAQ6510 as a 6½-digit DMM. When multiple channel measurements are not needed, the DAQ6510 functions as a DMM. Use the DAQ6510 with its front panel terminals to troubleshoot a test setup, a circuit, or to verify the value of a component. Save bench space and cost with an instrument that performs like a DMM and a data logger. The dual functionality enables maximized utilization of your instrument. Reduce Test System Downtime and Save on Lifetime Costs Unplanned system downtime results in loss of productivity and delayed product shipments or loss of valuable development time during product audit testing, accelerated life testing, and environmental testing. To avoid unplanned downtime, the DAQ6510’s plug-in switch modules have counters for each relay that monitor the number of relay contact closures. Thus, when relay switch closures approach the manufacturers’ relay life specification, the switch card module can be replaced during scheduled maintenance. Many of the plug-in switch modules have multi-pin D-sub connectors to ensure secure, minimal maintenance connections in automated test environments. The D-sub connectors minimize downtime when changing plug-in switch modules during system maintenance or during the setup of a new test system. The DAQ6510 is specified for both 1-year and 2-year accuracy specifications with little difference between the 1-year and 2-year specifications. Thus, for most applications, the calibration cycle for the DAQ6510 can be two years. This will reduce the life cycle costs on your DAQ6510 by half. The DAQ6510 will save on unplanned system downtime, minimize planned downtime, and save on lifetime calibration costs making the DAQ6510 a highly economical asset in any test system.

TEK.COM10 Plug-In Module Selector Guide Module Details Model Number 7700 7701 7702 7703 7705 7706 7707 7708 7709 7710 7711 7712

Description

20 Channel,

32 Ch. Differential Multiplexer Module 40 Ch. Differential Multiplexer Module 32 Ch. High Speed, Differential Multiplexer Module. 40 Ch. Single-pole Control Module All-in-One I/O Module. 32 Ch. Digital I/O Module 40 Ch. Differential Multiplexer Module 6×8 Matrix Module. 20 Ch. Solid-state Differential Multiplexer Module

2 GHz 50 Ω RF

3.5 GHz 50 Ω

No. Analog Inputs 20 32 40 32 40 20 10 40 48 20 8 8 Configuration Multi plexer w/CJC Multi plexer Multi plexer Multi plexer Indepen- dent SPST Multi plexer w/CJC Digital I/O/ Multi plexer Multi plexer w/CJC Matrix Multi plexer w/CJC Multi plexer Multi plexer 1×20 or two 1×10 1×32 or two 1×16 1×40 or two 1×20 1×32 or two 1×16 1×20 or two 1×10 1×10 or two 1×5 1×40 or two 1×20 6×8 1×20 or two 1×10 Dual 1×4 Dual 1×4 Connector Type Oversized screw terminal 50 pin female D-sub and 25 pin female D-sub Oversized screw terminal Two 50 pin female D-sub Two 50 pin female D-sub Screw terminal 50 pin male D-sub and 25 pin female D-sub Oversized screw terminal 50 pin female D-sub and 25 pin female D-sub Quick disconnect screw terminal SMA SMA Max. Voltage 300 V 150 V 300 V 300 V 300 V 300 V 300 V 300 V 300 V 60 V 60 V 42 V Max. Switched Current 1 A 1 A 1 A 500 mA 2 A 1 A 1 A 1 A 1 A 0.1 A 0.5 A 0.5 A Bandwidth 50 MHz 2 MHz 2 MHz 2 MHz 10 MHz 2 MHz 2 MHz 2 MHz 2 MHz 2 MHz 2 GHz 3.5 GHz Contact Life 1 108 108 108 108 108 108 108 108 108 1010 106 106 Switch Speed 3 ms 3 ms 3 ms 1 ms 3 ms 3 ms 3 ms 3 ms 3 ms 0.5 ms 10 ms 10 ms Other Details Max. power = 125 VA. 2 current measure channels Max. power = 125 VA. Max. power = 125 VA. 2 current measure channels Reed relays Max. power = 125 VA 2 analog outputs. 16 digital outputs. Max. power = 125 VA 32 digital I/O. Max. power = 125 VA Max. power = 125 VA Connects to internal DMM. Daisy chain mult. cards for up to 6×16 matrix (DAQ6510). Max. power = 125 VA. Solid state relays, 800 ch./second scan rate Insertion loss <1.0 dB @ 1 GHz. VSWR <1.2 @ 1 GHz. Insertion loss <1.1 dB @ 2.4 GHz. 1. No load contact life. See switch module data sheet for additional specifications. Module Function Capabilities Model Number 7700 7701 7702 7703 7705 7706 7707 7708 7709 7710 7711 7712 DC Current ✓ ✓ Temperature T/C w/Auto- matic CJC ✓ ✓ ✓ ✓ T/C w/ Resistance AC Current ✓ ✓ Event Counter/ Totalizer ✓ Digital Input ✓ Digital Output ✓ ✓ Analog Output ✓ RF Switching ✓ ✓

TEK.COM 11 DAQ6510 Data Acquisition and Logging, Multimeter System Specification Conditions This document contains specifications and supplemental information for the DAQ6510. Specifications are the standards against which the DAQ6510 is tested. Upon leaving the factory, the DAQ6510 meets these specifications. Supplemental and typical values are nonwarranted, apply at 23 °C, and are provided solely as useful information. Measurement accuracies are specified for plug-in module user input terminals and DAQ6510 front input terminals and include conversion error for thermocouple, thermistor, and RTD measurements. Additional switching module errors are included in the Characteristics section under each function. Measurement Conditions Include:

  • After a 30-minute warmup period
  • 1 PLC or 5 PLC measurement rate; for NPLC settings less than 1 PLC, add appropriate noise error from Measurement Noise table under each function
  • Autozero enabled
  • Calibration period: one year (recommended) or two years. Calibration period may vary depending on customer requirements
  • 24-hour accuracy specification is relative to calibrator accuracy
  • The communication accessory slot cover or an optional KTTI interface card is properly installed on the rear of the unit Definitions:
  • TCAL: The temperature at which the instrument was calibrated (23°C for factory calibration)
  • Temperature coefficient: Additional uncertainty added for each °C outside T CAL ±5°C
  • Power Line Cycle (PLC): 16.67 ms at 60 Hz and 20 ms at 50 Hz or 400 Hz line frequency; frequency automatically sensed at power up

TEK.COM12 DC Voltage DC Voltage Accuracy ±(% of reading + % of range) Range Resolution Input Impedance

24 Hours

TCAL ±1°C

90 Days

TCAL ±5°C

1 Year

TCAL ±5°C

2 Years

TCAL ±5°C Temperature Coefficient Measurement Noise Characteristics and Rejection Ratios Measurement Rate in NPLCs Digits DCV RMS Noise Uncertainty (in % of range + fixed base) 2 NMRR 3 CMRR 3 5 4 6.5 0 100 dB 140 dB 5 0 60 dB 140 dB 1 4 0 90 dB 140 dB 1 0 60 dB 140 dB 0.1 4 0.00015 + 1 µV 40 dB 120 dB 0.1 5.5 0.00015 + 4 µV — 120 dB 0.01 0.00030 + 6 µV — 80 dB 0.0005 4.5 0.00500 + 40 µV — 80 dB DC Voltage Characteristics Overrange 20% on 100 mV, 1 V, 10 V, and 100 V ranges. 1% on 1000 V range. ADC Linearity (10 V range) 0.0001% of 10 V range Input Impedance 100 mV to 10 V Ranges: Selectable, >10 GΩ or 10 MΩ ±1%)in parallel with <400 pF.

100 V to 1000 V Ranges: 10 MΩ ±1% in parallel with <400 pF

Input Bias Current <50 pA at 23°C Common Mode Current <600 nA peak-peak at 50 Hz or 60 Hz Earth Isolation 500 Vpeak >10 GΩ and <300 pF any terminal to chassis Common Mode Voltage 500 Vpeak LO terminal to chassis maximum Autozero Off Error Add ±(0.0002% of range + 3 µV) within ±1°C and ≤10 minutes since last autozero. Add ±(0.0010% of range + 10 µV) within ±5°C and ≤60 minutes since last autozero Input Protection Front-panel DAQ input: Input HI 1010 V, Sense HI (SHI) and Sense LO (SLO) 350 V referenced to LO Rear-panel DAQ input through the plug-in modules: 600 V PEAK all ranges (note this level exceeds the maximum input signal for modules: see the table below for module levels) Plug-In Module Additional Uncertainties and Maximum Input Signal Levels Plug-In Module Add the Following Uncertainty Maximum Input Signal Level 7000, 7002, 7008 1 µV 300 V for 7700, 7702, and 7708 7701, 7703, 7707 6 µV

300 V for 7703 and 7707

150 V for 7701

150 V for 7701 3 µV

300 V for 7706 and 7709

60 V for 7710

  1. For each additional volt over ±500 V, add 0.02 mV of uncertainty. 2. Noise values apply to terminals using a low-thermal short for 50 Hz and 60 Hz operation only. Measurements through a card may introduce additional noise. 4. Line sync on.

TEK.COM 13 DAQ6510 Data Acquisition and Logging, Multimeter System Resistance Resistance Accuracy ±(% of reading + % of range) 5 Range Resolution Test Current (±5%) Open Circuit Voltage (±5%) TCAL ±1°C TCAL ±5°C TCAL ±5°C TCAL ±5°C Temperature Coefficient Resistance Measurement Noise Characteristics 8 Measurement Rate in NPLC Digits 2-wire RMS Noise Uncertainty (in % of range + fixed base) 4-wire RMS Noise Uncertainty, Offset Compensation OFF (in % of range + fixed base) 9 4-wire RMS noise uncertainty, offset compensation ON (in % of range + fixed base) 9 6.5 0 0 0 1 0 0 0 Resistance Characteristics Overrange 20% on all ranges Autozero Off Error Add ±(0.0005% of range + 5 mΩ) within ±1°C and ≤10 minutes since last autozero Add ±(0.0020% of range + 10 mΩ) within ±5°C and ≤60 minutes since last autozero Offset Compensation Selectable on 1 Ω, 10 Ω, 100 Ω, 1 k Ω, and 10 kΩ ranges, 4-wire mode only Maximum 4-wire Lead Resistance 5 Ω per lead for 1 Ω range 10% of range per lead for 10 Ω, 100 Ω, 1 kΩ, and 10 kΩ ranges 1 kΩ per lead for 100 kΩ, 1 MΩ, 10 MΩ, and 100 MΩ Open Lead Detector Selectable on all ranges, 4-wire mode only; default is off. Input Protection Front-panel DAQ input: Input HI 1010 V, Sense HI (SHI) and Sense LO (SLO) 350 V referenced to LO Rear-panel DAQ input through the plug-in modules: 600 VPEAK all ranges (note this level exceeds the maximum input signal for modules: see maximum input signal levels for 7700 series plug-in modules listed in DC Voltage Characteristics or separately supplied module specifications) Switching Module Additional Contact Resistance See Plug-in Switching Module Data Sheet Plug-In Switching Module Additional Uncertainties Module Add the following to % of reading accuracy 10 kΩ 100 kΩ 1 MΩ 10 MΩ 100 MΩ 7701, 7703, 7707, 7709 0.001 0.01 0.1 1 10 7706, 7708 0.0005 0.005 0.05 0.5 5 7700, 7702 — — — 0.022 0.22 7710 % of reading error 0.0011 0.011 0.11 1.1 11 Temperature Coefficient Notes 5. Specifications are for 2- and 4-wire resistance. For 2-wire, use relative offset, and add 100 m Ω of additional uncertainty. For 4-wire, turn offset compensation on for ≤ 0 k Ω and off for >10 kΩ. The 1 Ω range is for 4-wire only. 6. Requires a 10-reading digital filter at 1 PLC or 2-reading digital filter at 5 PLC. 7. Specified for < 10% lead-resistance mismatch at HI and LO. 8. Applies for 1 Ω through 1 M Ω ranges. For 100 Ω range, multiple the listed values by five. Noise values apply to terminals using a low-thermal short for 50 Hz and 60 Hz operation only. Measurements through a switching module may introduce additional noise. 9. Open lead detection off. 10. Line sync on.

TEK.COM14 DC Current DC Current Accuracy ±(% of reading + % of range) Range Resolution Burden Voltage TCAL ±1°C TCAL ±5°C TCAL ±5°C TCAL ±5°C Temperature Coefficient DC Current Characteristics Overrange 20% on 10 µA, 100 µA, 1 mA, 10 mA, 100 mA, and 1 A ranges 1% on 3 A and 10 A ranges Front-Panel Input Protection Externally accessible 3 A, 250 V fast-acting fuse, 5 × 20 mm: Keithley replacement part number FU-99-1 Plug-in Module Input Protection Fuse provided in 7700 and 7702 plug-in modules; PCB-mounted 3 A, 250 V, fast-acting fuse Autozero Off Error Add ±0.004% of range within ±1°C and ≤10 minutes since last autozero Add ±0.015% of range within ±5°C and ≤60 minutes since last autozero Nominal Shunt Resistance 14 10 μA 100 μA 1 mA 10 mA 100 mA 1 A 3 A 10 kΩ 1 kΩ 100 Ω 10 Ω 1 Ω 100 mΩ 100 mΩ DC Current Measurement Noise Characteristics 15 Measurement Rate in NPLC Digits 4-wire RMS Noise Uncertainty (in % of range + fixed base) 6.5 1 0 0.1 16 0.0009 + 10.0 pA 0.1 5.5 0.0015 + 3.5 nA 0.01 0.0030 + 3.5 nA 0.0005 4.5 0.0200 + 5.0 nA Notes 11. Specifications apply to front-panel inputs only. 12. 20 mA range for 27xx emulation mode only. 13. Add 1.5 V when using with plug-in modules. 14. Guaranteed by design. 15. Noise values apply to open terminals for 50 Hz and 60 Hz operation only. Measurements through a card may introduce additional noise. 16. Line sync on.

TEK.COM 15 DAQ6510 Data Acquisition and Logging, Multimeter System Temperature Thermocouple Accuracy ±°C 17 Type Resolution Range

2 Year Accuracy

TCAL ±5°C; all uncertainties in °C Temperature Coefficient in°C/°C Simulated or External CJC Internal CJC (on module) Front Terminals, 7700, 7702, 7708 7706, 7709, 7710 7701, 7703, 7707 7700, 7708 7706 7710 J 0.001°C K 0.001°C N 0.001°C T 0.001°C E 0.001°C R 0.010 °C S 0.010 °C B 0.010 °C Resistance Temperature Detector (RTD) Accuracy ±°C Types: 100 Ω platinum PT100, D100, F100, PT385, and PT3916 or user-configurable 0 Ω to 10 kΩ Measurement Method Resolution Range TCAL ±5°C Temperature Coefficient in°C/°C 2-wire 18 0.01°C –200° to 850°C 0.80 0.003 3-wire 19 0.01°C –200° to 600°C 0.35 0.003 >600° to 850°C 0.37 0.003 4–wire 0.01°C –200° to 600°C 0.06 0.003 >600° to 850°C 0.12 0.003 Thermistor Accuracy ±°C Types: 2.2 kΩ, 5 kΩ, and 10 kΩ Measurement Method Resolution Range TCAL ±5°C Temperature Coefficient in°C/°C 2-wire 0.01°C –80° to 150°C 0.08 0.002 For readings >70°C, add this additional uncertainty per Ω of lead, channel, and contact resistance Thermistor Type Common Model Number 70° to 100°C >100° to 150°C 2.2 kΩ 44004 0.22°C per Ω 1.11°C per Ω 5 kΩ 44007 0.10°C per Ω 0.46°C per Ω 10 kΩ 44006 0.04°C per Ω 0.19°C per Ω Notes 17. Accuracy excludes probe errors. 18. Specifications do not include errors that may arise from user’s cable or terminal resistance.

TEK.COM16 Temperature Characteristics Thermocouple Conversion ITS-90 Thermocouple Reference Junction External (CJC on 2001-TCSCAN or user-provided with 2000-SCAN) or simulated (fixed) Open Thermocouple Detection Selectable per channel (open >130 k Ω; default on. Earth Isolation 500 VPEAK > 0 GΩ and <300 pF any terminal to chassis AC Voltage AC Voltage Accuracy ±(% of reading + % of range) 19 Range Resolution Calibration Cycle 3 Hz to 5 Hz 5 Hz to 10 Hz

10 Hz to

1 V 1 µV

10 V 10 µV

100 V 100 µV

750 V 100 µV

AC Voltage Characteristics Overrange (voltages in V RMS) 20% on 100 mV, 1 V, 10 V, and 100 V ranges. 0% for 750 V range. AC Measurement Method AC-coupled digital sampling with anti-alias filter Crest Factor (excludes sine wave) Crest factors of up to 3:1 at full-scale input or 10:1 maximum, whichever is greater Autorange selects optimum range for crest factor up to 10:1 Accuracy specifications apply to all crest factors and are limited to a product of (crest factor) × (fundamental frequency) ≤ 3 kHz Volt*Hertz Product ≤8 × 107 V*Hz 21 Common Mode Rejection Ratio >70 dB, for 1 kΩ unbalance in LO lead Detector Bandwidth Setting of 3 Hz, 30 Hz, or 300 Hz sets maximum measurement aperture of 200 ms, 20 ms, or 2 ms, respectively; only signals with frequency greater than the detector bandwidth are measured. Input Impedance 1.1 MΩ ±2%, in parallel with <100 pF Input Protection Front-panel input: 1100 VPEAK or 400 VDC on all ranges Rear-panel input through the plug-in modules: 600 VPEAK on all ranges (note this level exceeds the maximum input signal for modules; see Maximum Input Signal Levels for 7700 Series plug-in modules listed below. Maximum DCV 400 V on any ACV range ACV Frequency Frequency reading automatically returned in reading buffer when in full buffer mode. Frequency readings are specified as in the frequency and period table. Plug-in Module Maximum Input Signal Levels Module Maximum Input Signal Level 7700, 7702, 7703, 7706, 7707, 7708, 7709 300 VRMS / 425 VPEAK 7701 150 VRMS / 212 VPEAK 7710 42 VRMS / 60 VPEAK Notes 20. Specifications are for sine wave inputs >5% of range. 21. Guaranteed by design.

TEK.COM 17 DAQ6510 Data Acquisition and Logging, Multimeter System AC Current AC Current Accuracy ± (% of reading + % of range) 22 Range Resolution Burden Voltage Frequency TCAL ±1°C TCAL ±5°C TCAL ±5°C TCAL ±5°C Temperature Coefficient 100 µA 23 100 pA <0.14 V 1 mA 1 nA <0.17 V 10 mA 10 nA <0.17 V 100 mA 100 nA <0.20 V 1 A 1 µA <0.75 V 25 3 A 1 µA <1.70 V 25 AC Current Characteristics Overrange 20% on 100 µA, 1 mA, 10 mA, 100 mA, and 1 A ranges 1% on 3 A range AC Measurement Type AC-coupled true RMS; measures the AC component of the input Digital sampling with anti-alias filter. Input Protection See DC Current Characteristics Crest Factor 27 (excludes sine wave) 10:1 maximum crest factor (1.75:1 at full-scale) Autorange selects optimum range for crest factor up to 10:1 Accuracy specifications apply to all crest factors less than 5 and are limited to the product of (crest factor) × (fundamental frequency) ≤ 200 Hz ACI Frequency Frequency readings are automatically returned in the reading buffer when in full buffer mode. Frequency values are typical. Nominal Shunt Resistance 28 100 μA 1 mA 10 mA 100 mA 1 A 3 A 1 kΩ 100 Ω 10 Ω 1 Ω 100 mΩ 100 mΩ Notes 22. Specifications are for sine wave inputs > 5% of range and > 10 µA RMS. 23. The 100 µA range is only specified for front-panel inputs. 24. Typical performance for the indicated frequency ranges. 25. Add 1.5 V for the 1 A and 3 A ranges when used with a plug-in module. 26. For signals of < 5 Hz, add 0.2% of reading uncertainty. 27. 100 µA range is specified only for crest factors < 3. 28. Guaranteed by design.

TEK.COM18 Frequency and Period Frequency and Period Accuracy ± (% of reading) 29 Range Resolution Frequency Period TCAL ±5°C Temperature Coefficient in°C/°C 100 mV to 750 V (For signals >5% of range and >10 mVRMS) 0.0001% ppm of reading 3 Hz to 10 Hz 333 ms to 100 ms 0.100 0.0002 >10 Hz to 100 Hz <100 ms to 10 ms 0.030 0.0002 >100 Hz to 1 kHz <10 ms to 1 ms 0.010 0.0002 >1 kHz to 300 kHz <1 ms to 3.3 µs 0.009 0.0002 Square Wave 30 0.008 0.0002 Frequency and Period Characteristics Measurement Method Reciprocal-counting technique; measurement is AC-coupled using AC measurement functions. Voltage Ranges 100 mVRMS full scale to 750 VRMS; auto or manual ranging. Gate Time User definable from 2 ms to 273 ms (default 200 ms) Continuity Continuity Accuracy 2-Wire ±(% of reading + % of range) 31 Range Resolution Test Current Open Circuit Voltage (±5%) TCAL ±5°C Temperature Coefficient Capacitance Capacitance Accuracy ±(% of reading + % of range) 32 Range Resolution Charge Current (±5%) 33 TCAL ±5°C Temperature Coefficient 10 nF 1 pF 10 µA 0.40 + 0.10 0.05 + 0.01 100 nF 10 pF 100 µA 0.40 + 0.10 0.05 + 0.01 10 µF 1 nF 1 mA 0.40 + 0.10 0.05 + 0.01 100 µF 10 nF 1 mA 0.40 + 0.10 0.05 + 0.01 Capacitance Characteristics Overrange 20% on all ranges. Measurement Method Constant current slope measurement. Maximum Voltage and Voltage Clamp For all devices: Clamped by hardware to <3 V. Notes 29. Specifications apply for sine wave input with detector bandwidth of 3 Hz. For a detector bandwidth of 30 Hz, add 100 mHz uncertainty. For a detector bandwidth of 300 Hz, add 1 Hz uncertainty. 30. Used for square waves with amplitude > 10% of range and 10 Hz to 300 kHz. 31. Does not include user’s lead-resistance or plug-in module contact resistance. 32. Accuracies specified for cable, channel, and other stray connector capacitance properly zeroed with the REL function. 33. Discharge current limited to < 10 mA. 34. Specifications apply to front-panel inputs only.

TEK.COM 19 DAQ6510 Data Acquisition and Logging, Multimeter System Diode Diode Voltage Accuracy ±(% of reading + additional uncertainty) 35 Voltage Measure Range Resolution Maximum Voltage Measurement Test Current (±5%) TCAL ±5°C Temperature Coefficient 12 V 10 µA 0.0045 + 60.0 µV 0.0008 + 10 µV 10 V 100 µA 0.0045 + 80.0 µV 0.0008 + 10 µV 7 V 1 mA 0.0045 + 170.0 µV 0.0010 + 10 µV 7 V 10 mA 0.0045 + 1.1 mV 0.0010 + 10 µV Digitize Digitize DC Voltage Accuracy ±(% of reading + % of range) 36 Range Resolution Input Impedance TCAL ±5°C Temperature Coefficient 100 mV 10 µV >10 GΩ or 10 MΩ ±1% 0.040 + 0.020 0.0025 + 0.0030 1 V 100 µV >10 GΩ or 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010 10 V 1 mV >10 GΩ or 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010 100 V 10 mV 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010 1000 V 100 mV 10 MΩ ±1% 0.030 + 0.010 0.0025 + 0.0010 Digitize DC Current Accuracy ±(% of reading + % of range) 36 Range Resolution Burden Voltage TCAL ±5°C Temperature Coefficient Notes 35. Specifications apply to the front-panel inputs, only. Specifications do not include errors that may arise from user’s cable or connection resistance. 36. DC accuracy specified with 1000 samples per second, 100-reading digital filter. 37. Add 1.5 V for 1 A and 3 A ranges when used with a plug-in module.

TEK.COM20 Typical Digitize Signal Characteristics Typical performance for these conditions: Sample rate 1 MS per s; sine wave input V PEAK = –1 dB full-scale of range Function: Range Spur-free Range SFDR (1 kHz / 10 kHz / 50 kHz) THD + Noise SNDR (1 kHz / 10 kHz / 50 kHz) Bandwidth (–3 dB, 5%) Effective Number of Bits (1 kHz/10 kHz/50 kHz) DCV: 100 mV 75 / 70 / 50 65 / 60 / 50 210 kHz 9 / 9 / 7 DCV: 1 V 95 / 90 / 75 80 / 80 / 75 210 kHz 12 / 12 / 11 DCV: 10 V 95 / 80 / 70 90 / 80 / 70 440 kHz 13 / 12 / 10 DCV: 100 V 50 / 35 / 25 50 / 40 / 30 17 kHz 10 / 8 / 7 DCV: 1000 V 50 / 35 / 25 50 / 40 / 30 17 kHz 13 / 11 / 10 DCI: 100 µA 80 / 65 / 45 70 / 65 / 45 430 kHz 12 / 10 / 8 DCI: 1 mA 80 / 65 / 45 70 / 65 / 45 570 kHz 12 / 10 / 8 DCI: 10 mA 80 / 65 / 45 70 / 65 / 45 230 kHz 12 / 10 / 8 DCI: 100 mA 80 / 65 / 45 70 / 65 / 45 340 kHz 12 / 10 / 8 DCI: 1 A 70 / 50 / 40 65 / 50 / 40 25 kHz 11 / 8 / 7 DCI: 3 A 70 / 50 / 40 65 / 50 / 40 25 kHz 11 / 8 / 7 Digitizing Additional Characteristics Maximum Resolution 16 bits Measurement Input Coupling DC coupled Sampling Rate Programmable 1 kS through 1 MS/s Minimum Record Time 1 µs Maximum Record Length (Volatile) 7 million with standard buffer (includes channel and formatting information) DC Voltage Ratio DC Voltage Ratio Calculation 38 Method Measurement Channel Ratio (through rear input plug-in module) Accuracy = (Accuracy of channel A measure range + Accuracy of channel B measure range) × Channel ratio Channel Average (through rear input plug-in module) Accuracy = Accuracy of channel A measure range + Accuracy of paired channel B measure range DCV Input Ratio (HI-LO/SHI–SLO) 39, 40 Notes 39. Sense terminals on inputs are limited to 10 V range during ratio measurement. Add 0.0015% + 0.0005% per °C temperature coefficient to DCV percent of range accuracy when using the 100 V or 1000 V range on the input terminals. 40. Specified for front inputs only.

TEK.COM 21 DAQ6510 Data Acquisition and Logging, Multimeter System System Specifications Typical Single-Channel Rates, DC Functions 41, 42

60 Hz (50 Hz) Operation

Functions: DCV (10 V) 2-wire Ω (≤10 kΩ), DCI (1 mA) Functions: 4-wire Ω ( ≤1 kΩ) 4-wire and 3-wire RTD Function: Thermistor or Thermocouple Measurements (readings per second) 43 Buffer Computer Buffer Computer Buffer Computer 1 59 (48) 58 (48) 28 (23) 28 (23) 59 (49) 57 (48) 0.1 584 (490) 440 (380) 180 (160) 170 (150) 580 (480) 440 (380) 0.01 4900 (4100) 4800 (4100) 400 (390) 400 (390) 4800 (4100) 4700 (4000) 0.0005 20600 (20600) 19800 (19800) 460 (460) 460 (460) 21000 (21000) 20300 (20300) Typical Single-Channel Reading Rates, AC Functions 41 Function: ACV, ACI Function: Frequency, Period Measurements (readings per second) Detector Bandwidth Aperture Buffer or Computer

3 Hz 200 ms 1

30 Hz 20 ms 10

300 Hz 2 ms 100

Scanning/Multiple Channels (with optional scan cards) 44 Typical Scanning Measurement Rates Measurements Into Buffer/Computer (channels per second) Scanning DCV or 2-wire Ω >90 with 7700 plug-in module >450 with 7703 plug-in module >900 with 7710 plug-in module Scanning Thermocouple, Thermistor, or 2-wire RTD >90 with 7700 plug-in module >450 with 7703 plug-in module >900 with 7710 plug-in module Scanning 4-wire Ω and 3- or 4-wire RTD >80 with 7700 plug-in module >300 with 7703 plug-in module >400 with 7710 plug-in module Scanning ACV 45 >60 with 7700 plug-in module >170 with 7703 plug-in module >220 with 7710 plug-in module Scanning Alternating DCV and 2-wire Ω >90 with 7700 plug-in module >430 with 7703 plug-in module >430 with 7710 plug-in module DIGITAL I/O READ/WRITE, TOTALIZER READ SPEED Read Digital Input 7707 plug-in module > 200 Write Digital Output 7706 plug-in module > 1400 7707 plug-in module > 500 Read Totalizer 7706 plug-in module > 100 Notes 41. Reading speeds for autozero off, fixed range, autodelay off, offset compensation off, and open lead detector off where applicable. 42. Buffer measurements: For <0.1 PLC, multisample, and single buffer transfer binary reading only. 43. Computer measurements: For 5 PLC, 1 PLC, and 0.1 PLC single reading and single transfer to computer (USB). 44. Set-up conditions of the factory default setting with the following exceptions: 3.5 digits (0.0005 PLC), autorange off, autozero off, autodelay off, and open lead detection off. 45. Assume the signal is 10 kHz or above.

TEK.COM22 Typical Function and Range Change Speed Function Function Change Time 46 Range Change Time 47 Autorange Time 46 DCV, DCI, or 2-wire Ω 48 <4 ms <1.3 ms <3.2 ms 4-wire Ω 49 or 3-wire RTD <5.5 ms Thermistor — Frequency or Period (2 ms aperture) <1800 ms <50 ms 50 <50 ms 50 ACV (300 Hz bandwidth) ACI (300 Hz bandwidth) <100 ms <4 ms <5 ms Capacitance <4 ms <3 ms <30 ms Digitize <4 ms <5 ms — Diode <11 ms — — Continuity <11 ms — — Thermocouple <4 ms — — Bus Transfer Speed 51 USB LAN GPIB RS232 (Baud 115200) Average for 1000 readings (binary) 441,000 268,000 201,000 10,000 Average for 1000 readings with relative timestamp (binary) 272,000 150,000 105,000 2,900 Average for 1000 readings with formatted elements 52 46,000 29,000 17 ,000 290 Typical Digitize Voltage or Current 53 Sampling Rate Measurements over USB to Computer (readings per second) 10 kS/s Up to 10,000 50 kS/s Up to 50,000 100 kS/s Up to 100,000

1 MS/s up to 7 s maximum duration At least 90,000

Trigger Sources Front panel trigger key, timer, command interface, LAN/LXI, Trigger In (BNC rear panel), Digital I/O (optional accessory card), and TSP-Link ® (optional accessory card) External Trigger Delay <1 µs when triggering from accessory card or rear BNC input External Trigger Jitter <1 µs when triggering from accessory card or rear BNC input External Trigger In/Trigger Out 0 V to 5 V logic signal input and output, TTL-compatible, programmable edge pulse Minimum pulse width: 1 µ s External Trigger Out, Maximum Rate Up to 90 kHz, measurement dependent External Trigger In, Maximum Rate Up to 150 kHz, measurement dependent Notes 46. 3.5 digits, autozero off, 0.0005 PLC, excludes measurement time. 47. DCV = 10 V; 2-wire or 4-wire = 1 k Ω; DCI = 1 mA; ACI = 1 mA; ACV = 1 V; Capacitance = 10 µ F. 48. 2-wire function for 100 Ω range and up. For the 10 Ω range, add 2.7 ms. 49. 4-wire function for 100 Ω range and up. For the 1 Ω and 10 Ω ranges, add 2.7 ms. 50. When ranging to 10 V and above, add 1.8 s. 51. SCPI programmed using 4-byte binary format. 52. Format elements: Reading, relative timestamp, channel, and unit. 53. SCPI programmed using 4-byte binary format.

TEK.COM 23 DAQ6510 Data Acquisition and Logging, Multimeter System Scanning Scan Count 1 to continuous Scan Interval 0 s to 27.7 hours Channel Delay 0 to 60 s Measure Interval 0 s to 27.7 hours Internal Memory Maximum Reading Memory (volatile) Up to 7 million readings with standard buffer (includes channel and formatting information). Internal (non-volatile) Memory for Saved Scripts and Scan Configurations 6 MB, enables hundreds of scan configurations or TSP scripts to be saved in non-volatile memory. General Specifications Line Power Power Supply 100 V, 120 V, 220 V, and 240 V (±10%) Power Line Frequency 50 Hz to 60 Hz and 400 Hz, automatically sensed at power-up Maximum Power Consumption 65 VA Typical Power Consumption 30 VA Mains Input Fuse 250 V, 1.25 A slow-blow fuse: Keithley replacement part number FU-106-1.25 Environment and Regulatory Operating Environment Specified for 0° to 50°C, ≤80% relative humidity at 35°C, altitude up to 2000 meters Storage Environment –40° to 70°C Vibration MIL-PRF-28800F Class 3, random Warm-up 30 minutes to rated accuracy Safety NRTL listed to UL61010-1 and CSA C22.2 No 61010-1; conforms with European Union Low Voltage Directive EMC Conforms to European Union EMC Directive Mechanical Display 12.7 cm (5 in.) capacitive touch, color TFT WVGA (800 × 480) with LED backlight Shipping Weight 4.54 kg (10.0 lb.) instrument only Input Signal Connections Front plug-in modules Plug-in Module Slot Two slots on the rear panel; see Plug-In Module Accessories Communication Slot One slot on rear panel, see Optional Interfaces and Programmable Digital I/O Cooling Forced air, fixed speed

TEK.COM24 Remote Interface: Standard LAN/LXI Compliance RJ-45 Connector: 10/100BT. IP Configuration: Static or DHCP (manual or automatic). Web Interface: Virtual front panel. LXI Compliance: LXI version 1.4 core 2016. USB Device (rear panel, Type B) 2.0 full speed, USBTMC compliant USB Host (front panel, Type A) USB 2.0, support for flash drives, FAT32. Capability: Import/export instrument configuration files, reading buffers, screen captures, and scripts Language SCPI (default) Default command set, Standard Commands for Programmable Instruments, SCPI-1999 TSP Embedded Test Script Processor (TSP) accessible from any host interface; responds to high-speed test scripts comprised of remote commands and statements (for example, branching, looping, and math); able to execute test scripts stored in memory without host intervention Emulation Modes Keithley Model 2700 and Model 2701 Math Functions REL, Minimum, Maximum, Average, Standard Deviation, peak-peak, dB, Limit Test, Percent, 1/x, and mX+b with user-defined units displayed Miscellaneous Real-time Clock Lithium battery backup, CR2032 coin-type, factory replaceable, (3+ years of battery life); set and read year, month, day, hour, minute, and second. (Note: Seconds are not adjustable.) Timestamp Resolution 15 ns with standard or full buffer style Password Protection 30 characters Alarms Up to six: see Optional Interfaces and Programmable Digital I/O Power Failure Recovery Mode User selectable, resumes scanning once power is re-applied Optional Interfaces and Programmable Digital I/O KTTI-GPIB GPIB IEEE-488.1 compliant; supports IEEE-488.2 common commands and status model topology KTTI-RS232 RS232, 9-pin d-sub female connector; standard baud rates from 300 to 115,200 bps are supported KTTI-TSP RJ-45 (quantity 2); TSP-Link ® expansion interface allows TSP-enabled instruments to trigger and communicate with each other Digital I/O For KTTI-RS232, KTTI-GPIB, and KTTI-TSP Connector: 9 pin d-sub female

5 V Power Supply Pin: Limited to 500 mA > 4 V (solid-state fuse protected)

Lines: Six input / output, user-defined for control, alarms (limits), or triggering Input Signal Levels: 0.7 V (maximum logic low), 3.7 V (minimum logic high) Input Voltage Limits: –0.25 V (absolute minimum), 5.25 V (absolute maximum) Maximum Source Current: 2.0 mA at > 2.7 V (per pin) Maximum Sink Current: –50 mA at 0.7 V (per pin, solid state fused)

TEK.COM 25 DAQ6510 Data Acquisition and Logging, Multimeter System

Ordering Information

DAQ6510 Data Acquisition/Multimeter System with 2 Module Slots DAQ6510/7700 Data Acquisition/Multimeter System with Model 7700 20-Channel Differential Multiplexer Module Supplied Accessories

1757 Pair, general purpose test lead set, 1000 V Cat II

USB-B-1 USB Cable, Type A to Type B, 1 m (3.3 ft) Traceable Calibration Certificate Three-Year Warranty Instruction Manuals/Documentation (available at www.tek.com/DAQ6510) DAQ6510 Quick Start Guide DAQ6510 User’s Manual DAQ6510 Reference Manual Software and Drivers (available at tek.com) IVI/VISA Drivers for Microsoft ® Visual Basic ®, Visual C/C++® National Instruments (NI ®) LabView™, NMI LabWindows™/CVI (available at ni.com ) Keithley Test Script Builder available at https://www.tek.com/keithley-test-scri pt-builder KickStart available at www.tek.com /kickstart Optional Series 7700 Plug-in Switch Modules 7700 20-channel Differential Multiplexer Module with Automatic CJC 7701 32-channel Differential Multiplexer Module 7702 40-channel Differential Multiplexer Module 7703 32-channel, High-Speed, Differential Multiplexer Module 7705 40-channel Single-Pole Control Module

7706 All-in-one I/O Module

7707 32-channel Digital I/O Module 7708 40-channel Differential Multiplexer Module with Automatic CJC 7709 2-pole, 6×8 Matrix Module 7710 20-channel Solid State Differential Multiplexer with Automatic CJC 7711 2 GHz 50 Ω RF Module 7712 3.5 GHz 50 Ω RD Module

TEK.COM26 Optional Interfaces and Programmable Digital I/O KTTI-RS232 RS-232 Communication and Digital I/O Accessory, user-installable KTTI-GPIB GPIB Communication and Digital I/O Accessory, user-installable KTTI-TSP TSP-Link Communication and Digital I/O Accessory, user-installable Available Accessories Test Leads and Probes

1752 Premium Safety Test Lead Kit

1754 2-Wire Universal 10-Piece Test Lead Kit

1756 General Purpose Test Lead Kit

5804 Kelvin (4-Wire) Universal 10-Piece Test Lead Kit

8610 Low Thermal Shorting Plug

PC Communication Interface and Cables KPCI-488LPA IEEE-488 Interface for PCI Bus KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter USB-B-1 USB Cable, Type A to Type B, 1 m (3.3 ft.)

174694600 CAT5 Crossover Cable for TSP-Link/Ethernet

7007 Shielded GPIB Cable

DB9-MM 9-pin, D-sub, male-to-male connector for interface options Triggering and Control

8503 DIN to BNC Trigger Cable

4299-8 Single Fixed Rack Mount Kit 4299-9 Dual Fixed Rack Mount Kit 4299-10 Dual Fixed Rack Mount Kit. Mount One DAQ6510 and One Series 26xxB Instrument 4299-11 Dual Fixed Rack Mount Kit. Mount One DAQ6510 and One Instrument from Series 2400, Series 2000, etc. Replacement Fuses FU-99-1 Current Input Fuse, 3 A, 250 V Fast Acting 5×20 mm FU-106-1.25 Main Input Fuse, 1.25 A, 250 V, Slo-Blo

TEK.COM 27 DAQ6510 Data Acquisition and Logging, Multimeter System Available Services Extended Warranties Instruments DAQ6510-EW 3 year factory warranty extended to 4 years from date of shipment DAQ6510-5Y-EW 3 year factory warranty extended to 5 years from date of shipment Plug-in Modules See Plug-In Module Data Sheets Calibration Contracts Instruments C/DAQ6510-3Y-DATA KeithleyCare 3 Year Calibration w/Data Plan C/DAQ6510-5Y-DATA KeithleyCare 5 Year Calibration w/Data Plan C/DAQ6510-3Y-STD KeithleyCare 3 Year Standard Calibration Plan C/DAQ6510-5Y-STD KeithleyCare 5 Year Standard Calibration Plan

TEK.COM28 The 7700 plug-in module offers 20 channels of 2-pole or 10 channels of 4-pole multiplexer switching that can be configured as two independent banks of multiplexers. There are two additional protected channels for current measurements. Automatic CJC is provided so that no other accessories are required to make thermocouple temperature measurements. In addition, the 7700 contains latching electromechanical relays that enable signal bandwidths of up to 50 MHz. The 7700 is ideal for RTD, thermistor, and thermocouple temperature applications. Key Features

  • 20 channels for generalpurpose measurements, plus two channels to measure current
  • Oversize screw terminal connection blocks are standard for easier connections
  • 50 MHz bandwidth
  • 300 V, 1 A capacity for voltage channels; 60 W, 125 VA
  • 3 A capacity for current channels
  • Low insertion loss of up to 50 MHz
  • Relay closures stored in onboard memory Series 7700 Plug-In Switch Modules Cold Junction Ref ×3 Cold Junction Ref ×3 Channel 1 HI LO Channel 10 HI LO (Channels 2–9) Channel 11 HI LO Channel 20 HI LO (Channels 12–19) Channel 21 HI LO Channel 22 HI LO AMPS HI LO Sense HI LO Input Channel 23 2-Pole (Open) 4-Pole (Closed) (see Note) To Instrument Backplane Channel 25 (see Note) Backplane isolation Channel 24 (see Note) Backplane isolation NOTE: Channels 23–25 in this schematic refer to the designations used for control and not actual available channels. For more information, refer to the ROUTe:MULTiple command section in the Instrument User’s Manual. Channels 24 and 25 can be individually controlled using ROUTe:MULTiple if the module is not to be connected to the internal DMM.Card AMPS LO Card Sense HI LO Card Input HI LO 7700 20-channel, Differential Multiplexer Module with Automatic CJC, Screw Terminals, and up to 50MHz Bandwidth Datasheet Specifications Capabilities Channels 1–20 Multiplex one of twenty 2-pole or one of ten 4-pole signals into DMM. Channels 21–22 Multiplex one of two 2-pole current signals into DMM. Inputs Maximum Signal Level Channels (1–20) 300 V DC or 300 V rms (425 V peak) for AC waveforms, 1 A switched, 60 W, 125 VA maximum. Channels (21–22) 60 V DC or 30 V rms, 3 A switched, 60 W, 125 VA maximum. Contact Life (typ.) >105 operations at max. signal level. >10 8 operations no load 1. 1. Open thermocouple detector on during thermocouple measurements. Minimum signal level 10 mV, 10 μA. Contact Resistance <1 Ω at end of contact life. Contact Potential <±500 nV typical per contact, 1 μV max. <±500 nV typical per contact pair, 1 μV max. Offset Current <100 pA. Connector Type Screw terminal, #20 AWG wire size. Isolation Between Any Two Terminals: >1010 Ω, <100 pF. Isolation Between Any Terminal and Earth: >109 Ω, <200 pF. Insertion Loss (50 Ω Source, 50 Ω Load) w/Internal DMM w/o Internal DMM* <0.1 dB 1 MHz 1 MHz <3 dB 2 MHz 50 MHz Crosstalk (50 Ω Load): w/Internal DMM w/o Internal DMM*

10 MHz <–40 dB <–40 dB

25 MHz ** <–25 dB

300 V or 300 V rms (425 V peak) for AC waveforms between

any terminal and chassis. * Channels 24 and 25 are open. Refer to ROUTe:MULTiple command in 27XX User Manual. ** Not valid. General 20 Channels 20 channels of 2-pole relay input. All channels configurable to 4-pole. 2 Channels 2 channels of current only input. Relay Type Latching electromechanical. Actuation Time <3 ms. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.45 kg (1 lb). Supplied Accessories CC-92-1 Set of 20 Cable Ties TL-23 Screwdriver Available Services 7700-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 7700 20-channel, Differential Multiplexer Module with Automatic CJC and Screw Terminals

TEK.COM 29 DAQ6510 Data Acquisition and Logging, Multimeter System The 7701 plug-in module offers 32 channels of 2-pole or 16 channels of 4-pole multiplexer switching. Its 32 channels can be configured for common-side 4-wire ohms. They can also be configured as two independent banks of multiplexers. It is ideal for RTD or thermistor temperature applications. Key Features

  • Configurable for 32 channels of differential measurements, with up to 16 channels of 4-pole measurements
  • Two female D-shell connectors are standard for secure hook-up and quick teardown
  • 150 V, 1 A capacity for voltage channels; 60 W, 125 VA
  • Relay closures stored in onboard memory
  • Screw terminal jumpers allow user-configurable DMM connections Series 7700 Plug-In Switch Modules Channel 1 HI LO Channel 16 HI LO (Channels 2–15) Channel 17 HI LO Channel 32 HI LO (Channels 18–31) Backplane Isolation HI LO Backplane Isolation HI LO DMM Input Channel 33 2-Pole (Open) 4-Pole (Closed) (see Note) To Interna l DMM NOTE: Channels 33–35 in this schematic refer to the designationsused for control and not actual available channels. For more information, refer to the ROUTE:MULT command section in the Instrument User’s Manual. Channel 35 (see Note) Channel 34 (see Note) Multiplexer 1 HI LO Screw Terminals Screw Terminals The 7701 is rated for low-voltage applications. When connecting the 7701 to the internal DMM via the screw terminals, all other modules in the mainframe must be derated to 150 VDC or 150V rms (212 V peak) for AC waveforms. HI LO HI LO DMM Sense Multiplexer 2 HI LO External Wiring Access User Con/f_igurable Screw Terminals 7701 32-channel Differential Multiplexer Module Datasheet Specifications Capabilities Channels 1–32 Multiplex one of 32 2-pole or one of 16 4-pole signals into DMM. Configuration supports dual 1×16 independent multi- plexers. Inputs Maximum Signal Level Any channel to Any Channel (1–32): 150 V DC or 150 Vrms (212 V peak) for AC waveforms, 1 A switched, 60 W, 125 VA maximum. Contact Life (typ): >105 operations at max. signal level. >108 operations no load 1. 1. Minimum signal level 10 mV, 10 μ A. Contact Resistance <1 Ω any path and additional 1 Ω at end of contact life. Contact Potential <6 μV per contact pair. Offset Current <100 pA. Connector Type 50-pin female D-shell, Channels 1–24. 25-pin female D-shell, Channels 25–32. Supplied with male IDC ribbon cable connectors. Isolation Between Any Two Terminals >109 Ω, <200 pF. Isolation Between Any Terminal and Earth >109 Ω, <400 pF. Cross Talk (1 MHz, 50 Ω Load) <–35 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.35 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage

300 VDC or 300 Vrms (425 V peak) for AC waveforms

between any terminal and chassis. General 32 Channels 32 channels of 2-pole relay input. All channels configur a ble to 4-pole. Relay Type Latching electromechanical. Actuation Time <3 ms. DMM Connections Screw terminals provide internal DMM connections to channels 34 and 35 and connections to external wiring access. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight <0.52 kg (1.16 lb). Supplied Accessories 7709-306A 50-pin D-Sub Male IDC Connector Kit 7709-307A 25-pin D-Sub Male IDC Connector Kit J-15 Jumper Wires, quantity 4 Available Accessories 7789 50-pin male, 25-pin male D-shell solder cup connectors 7790 50-pin male, 50-pin female, 25-pin male D-shell IDC connectors Available Services 7701-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 7701 32-channel, Differential Multiplexer Module

TEK.COM30 The 7702 plug-in module offers 40 channels of 2-pole or 20 channels of 4-pole multiplexer switching that can be configured as two independent banks of multiplexers. The 7702 provides two additional protected channels for current measurements. It is ideal for RTD, thermistor, and thermocouple temperature applications. Key Features

  • 40 channels for generalpurpose measurements, plus 2 channels to measure current
  • Two- or four-wire measurement
  • Oversize screw terminal connection blocks are standard for easier connection
  • 300 V, 1 A capacity for voltage channels; 60 W, 125 VA
  • 3 A capacity for current channels
  • Relay closures stored in onboard memory Series 7700 Plug-In Switch Modules Channel 1 HI LO Channel 20 HI LO (Channels 2–19) Channel 21 HI LO Channel 40 HI LO (Channels 22–39) Channel 41 HI LO Channel 42 HI LO AMPS HI LO Sense HI LO Input Channel 43 2-Pole (Open) 4-Pole (Closed) (see Note) To Instrument Backplane Channel 45 (see Note) Backplane isolation Channel 44 (see Note) Backplane isolation NOTE: Channels 43–45 in this schematic refer to the designations used for control and not actual available channels. For more information, refer to the ROUTe:MULTiple command sectionin the Instrument User’s Manual. Channels 44 and 45 can be individually controlled using ROUTe:MULTiple if the module is not to be connected to the internal DMM. Card Sense HI LO Card Input HI LO 7702 40-channel, Differential Multiplexer Module with Screw Terminals Datasheet Specifications Capabilities Channels 1-40 Multiplex one of 40 2-pole or one of 20 4-pole signals into DMM. Channels 41-42 Multiplex one of 2 2-pole current signals into DMM. Inputs Maximum Signal Level Channels (1-40) 300 V DC or rms, 1 A switched, 60 W, 125 VA maximum. Channels (41-42) 60 V DC or 30 V rms, 3 A switched, 60 W, 125 VA maximum. Contact Life (typ.) >105 operations at max. signal level. >108 operations no load 1. 1. Minimum signal level 10 mV, 10 μ A. Contact Resistance <1 Ω at end of contact life. Contact Potential <±500 nV typical per contact, 1 μ V max. <±500 nV typical per contact pair, 1 μ V max. Offset Current <100pA. Connector Type Screw terminal, #20 AWG wire size. Isolation Between Any Two Terminals >1010 Ω, <100 pF. Isolation Between Any Terminal and Earth >109 Ω, <200 pF. Cross Talk (10 MHz, 50 Ω Load) <–40 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.1 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage 300V between any terminal and chassis. General 40 Channels 40 channels of 2-pole relay input. All channels configurable to 4-pole. 2 Channels 2 channels of current only input. Relay Type Latching electromechanical. Actuation Time <3 ms. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.5 kg (1.1 lb). Supplied Accessories CC-92-1 Set of 20 Cable Ties TL-23 Screwdriver Available Services 7702-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

7702 40-channel Differential Multiplexer Module with Screw Terminals

TEK.COM 31 DAQ6510 Data Acquisition and Logging, Multimeter System The 7703 plug-in module offers 32 channels of 2-pole or 16 channels of 4-pole multiplexer switching that can be configured as two independent banks of multiplexers. The non-latching reed relays provide high speeds and are designed for 300 volt, 500 mA; 10 VA. The relay closures are stored in onboard memory. The 7703 is ideal for RTD and thermistor temperature applications. Key Features

  • 32 channels for general purpose measurements
  • Relay actuation time of less than 1ms for high-speed scanning
  • Two- or four-wire measurement
  • Two 50-pin female D-sub connectors are standard for secure hook-up and quick teardown Specifications Series 7700 Plug-In Switch Modules Channel 1 HI LO Channel 16 HI LO (Channels 2–15) Channel 17 HI LO Channel 32 HI LO (Channels 18–31) Backplane isolation HI LO Sense Backplane isolation HI LO Input Channel 35 2-Pole (Closed) 4-Pole (Open) (see Note) To Instrument Backplane NOTE: Channels 33–35 in this schematicrefer to the designations used for control and not actual available channels. For more information, refer to the ROUTE:MUL T command section in the Instrument User’s Manual. Channel 33 (see Note) Channel 34 (see Note) Card Sense HI LO Card Input HI LO 7703 32-channel, High Speed, Differential Multiplexer Module Datasheet Specifications Capabilities Channels 1–32 Multiplex one of 32 2-pole or one of 16 4-pole signals into DMM. Inputs Maximum Signal Level Channels (1–32) 300 V DC or rms, 0.5 A switched, 10 W maximum. Contact Life (typ.) >5×104 operations at max. signal level. >108 operations cold switching. Contact Resistance <1 Ω at end of contact life. Contact Potential <±3 μV typical per contact, 6 μV max. <±3 μV typical per contact pair, 6 μV max. Offset Current <100 pA. Connector Type 50 pin D-sub ×2. Relay Drive Current 20 mA per channel. Isolation Between Any Two Terminals >109 Ω, <200 pF. Isolation Between Any Terminal and Earth >109 Ω, <400 pF. Cross Talk (1 MHz, 50 Ω Load) <–40 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.35 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage 300 V between any terminal and chassis. General 32 Channels 32 channels of 2-pole relay input. All channels configurable to 4-pole. Relay Type Reed. Actuation Time <1 ms. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 40% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.8 kg (1.75 lbs). Supplied Accessories 7703-306A 50-pin D-Sub Male Shell Connector Kit, quantity 2 Available Accessories 7705-MTC-2 50 Pin Male to Female D-sub Cable, 2 m (6.6 ft). 7788 50-pin D-Sub male connector kit, quantity 2 7790 50-pin male, 50-pin female, 25-pin male D-shell IDC connectors Available Services 7703-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

7703 32-channel, High Speed, Differential Multiplexer Module

TEK.COM32 The 7705 plug-in module offers 40 channels of independent switching. These channels are designed to control power to the DUT and switching loads. They can also directly control light indicators, relays, etc. Key Features

  • 300 V, 2 A capacity
  • Two 50-pin female D-sub connectors are standard for secure hook-up and quick teardown
  • Relay closures stored in onboard memory Series 7700 Plug-In Switch Modules Channel 1 IN OUT Channel 40 IN OUT (Channels 2–39) 7705 40-channel, Single-pole Control Module Datasheet Specifications Inputs Maximum Signal Level

300 VDC or rms, 2 A switched, 60 W (DC, resistive),

125 VA (AC, resistive). Contact Life No Load1: 108 closures. At Maximum Signal Levels: 10 5 closures. 1. Minimum signal level 10 mV, 10 μ A. Channel Resistance (per conductor) <1 Ω. Contact Potential ≤4 μV per contact. Offset Current <100 pA. Actuation Time 3 ms. Isolation Channel to Channel: >109 Ω, <50 pF. Common Mode: >109 Ω, <100 pF. Crosstalk (1 MHz, 50 Ω load) <–35 dB. Insertion Loss (50 Ω source, 50 Ω load) <0.3 dB below 1 MHz, <3 dB below 10 MHz. Common Mode Voltage 300 V between any terminal and chassis. General Relay Switch Configuration 40 independent channels of 1-pole switching. Isolated from internal DMM. Contact Configuration 1 pole Form A. Relay Type Latching electromechanical. Connector Type Two 50-pin female D-sub connectors. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.45 kg (1 lb). Supplied Accessories 7703-306A 50-pin D-Sub Male Shell Connector Kit, quantity 2 Available Accessories 7705-MTC-2 50 Pin Male to Female D-sub Cable, 2 m (6.6 ft). 7788 50-pin D-Sub male connector kit, quantity 2 7790 50-pin male, 50-pin female, 25-pin male D-shell IDC connectors Available Services 7705-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 7705 40-channel, Single-pole Control Module

TEK.COM 33 DAQ6510 Data Acquisition and Logging, Multimeter System The 7706 plug-in module offers 20 channels of 2-pole or 10 channels of 4-pole multiplexer switching with automatic CJC, as well as two analog output channels, 16 digital outputs, and one event counter/totalizer. The event counter/ totalizer can be used to monitor and control system components, such as fixtures, limit switches, pass/fail indicators, external voltage sources, loads, door closures, revolutions, etc., while performing mixed signal measurements. The 7706 is ideal for RTD, thermistor, and thermocouple temperature applications. Key Features

  • 20 channels of analog input (w/automatic CJC) for general- purpose measurements
  • 16 channels of digital output
  • 2 analog outputs (±12 V, 5 mA)
  • 300 V, 1 A capacity; 60 W, 125 VA maximum
  • Configurable as two independent banks of multiplexers
  • Relay closures stored in onboard memoryy Series 7700 Plug-In Switch Modules Channel 22 NOTE: Non-isolated grounds ( ) are referenced to chassis ground. Channel 23 DAC Channel 23 Channel 24 DAC Channel 24 Totalizer +IN —IN +GATE —GATE Channel 25 Cold junction Ref ×2 Channel 1 HI LO Channel 10 HI LO (Channels 2—9) Channel 11 HI LO Channel 20 HI LO (Channels 12—19) HI LO Sense HI LO Input Channel 26 2-Pole (Open) 4-Pole (Closed) (see Note) To Instrument Backplane Channel 28 (see Note) Backplane isolation Channel 27 (see Note) Backplane isolation NOTES: Channels 26—28 in this schematic refer to the designations used for control and not actual available channels. Channels 26, 27, and 28 can be individually controlled using ROUTe:MULTiple if the module is not to be connected to the internal DMM. For more information, refer to theROUTe:MULTiple command section in the Instrument Users Manual. Sense HI LO Input HI LO Bit Channel 21 Digital Output Cold junction Ref ×2

7706 All-in-One I/O Module 20-channel Differential Multiplexer w/Automatic

CJC, 16 Digital Outputs, 2 Analog Outputs, a Counter/Totalizer, and Screw Terminals Datasheet Specifications Capabilities Channels 1–20 Multiplex one of 20 2-pole or one of 10 4-pole signals into DMM. Channels 21–22 16 Digital Outputs. Channels 23–24 Analog Voltage Output (2). Channels 25 Totalize Input. Channels 21–25 are referenced to chassis ground. Inputs (Channels 1–20) Maximum Signal Level (Channels 1–20): 300 V DC or rms, 1 A switched, 60 W, 125 VA maximum. Safety Category CAT 1 Contact Life (typ.) >105 operations at max. signal level; >108 operations no load 1. 1. Minimum signal level 10 mV, 10 μ A. Contact Resistance <1 Ω at end of contact life. Contact Potential <±2 μV typical per contact, 3 μ V max. Offset Current <100 pA. Connector Type Screw terminal, #22 AWG wire size. Isolation Between Any Two Terminals: >109 Ω, <100 pF. Isolation Between Any Terminal and Earth: >109 Ω, <200 pF. Cross Talk (10MHz, 50 Ω Load): <–35 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.1 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage: 300 V between any terminal and chassis. Digital Output (Channels 21 and 22) Vout(L) <0.8 V @ Iout = 400 mA. Vout(H) >2.4 V @ Iout = 1 mA. Vout(H)Max. <42 V with external open drain pull-up. Write Speed 50/s. Analog Voltage Output (Channels 23 and 24) DAC 1, 2 ±12 V @ 1 mA max, non-isolated, ±10 V @5 mA max. Resolution 1 mV. Iout 5 mA max. Settling Time 1 ms to 0.01% of output. Accuracy ±(% of output + mV) 1 year ±5°C: 0.15% + 19 mV; 24 hour ±1°C: 0.04% + 19 mV. Temperature Coefficient: ±(0.015% + 1 mV)/°C. Write Speed 50/s. Totalize Input (Channel 25) Maximum Count 232–1. Totalize Input 100 kHz (max), rising or falling edge, programmable. Signal Level 1 Vp-p (min), 42 Vpk (max). Threshold 0 V or TTL, jumper selectable. Gate Input TTL -Hi, TTL -Lo, or none. Count Reset Manual or Read+Reset. Read Speed 50/s. General 20 Channels 20 channels of 2-pole relay input. All channels configurable to 4-pole. Relay Type Latching electromechanical. Actuation Time <3 ms. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.5 kg (1.1 lbs). Supplied Accessories CC-92-1 Set of 20 Cable Ties TL-23 Screwdriver Available Services 7706-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

TEK.COM34 The 7707 plug-in module offers 10 channels of 2-pole or 5 channels of 4-pole multiplexer switching that can be configured as two independent banks of multiplexers. The 7707 also provides 32 digital input/output channels (four 8-bit ports) for I/O control. Connect the 7707 to industry standard solid-state relays to switch up to 980 VA. Key Features

  • 300 V, 1 A capacity; 60 W, 125 VA maximum (analog)
  • 33 V, 100 mA capacity (digital)
  • Digital outputs are short circuit protected
  • Relay closures stored in onboard memory

7707 32-channel Digital I/O Module with 10-channel Differential Multiplexer Series 7700 Plug-In Switch Modules Channel 12 Channel 14 HI LO Sense HI LO Input To Internal DMM Channel 17 (see Note) Backplane isolation Channel 16 (see Note) Backplane isolation NOTES: Channels 15—17 in this schematic refer to the designations used for control and not actual available channels. For more information, refer to the ROUTe:MULT command section in the Instrument Users Manual. Bit 1—16 Channel 11 DIO Digital I/O Bit 17—32 Channel 13 DIO Digital I/O Channel 1 HI LO Channel 5 HI LO (Channels 2—4) Channel 6 HI LO Channel 10 HI LO (Channels 7—9) Channel 15 2-Pole (Open) 4-Pole (Closed) (see Note) Card Sense HI LO Card Input HI LO 7707 32-channel Digital I/O Module with 10-channel Differential Multiplexer Datasheet Specifications Capabilities Channels 1–10 Multiplex one of 10 2-pole or one of 5 4-pole signals into DMM. Channels 11–14 32 Digital Inputs/Outputs referenced to chassis ground. Thermal Protection Channels 11–14 are thermally protected to 1 A up to 25 V. Inputs (Channels 1–10) Maximum Signal Level: Any Channel to Any Channel (1–10) : 300 VDC or 300 Vrms (425 V peak) for AC waveforms, 1 A switched, 60 W, 125 VA maximum. Contact Life (typ.) >105 operations at max. signal level >108 operations no load 1. 1. Minimum signal level 10 mV, 10 μ A. Contact Resistance <1 Ω any path and additional 1 Ω at end of contact life. Contact Potential <6 μV typical per contact pair and additional 5 μ V with Channels 11–14 at rate VOUT(L). Offset Current <100 pA. Connector Type: 50-pin male D-shell, Channels 11–14. 25-pin female D-shell, Channels 1–10. Supplied with female and male IDC ribbon cable connectors. Isolation Between Any Two Terminals >109 Ω, <100 pF with isolation channels 16 and 17 open. Isolation Between Any Terminal and Earth: >109 Ω, <200 pF. Cross Talk (1 MHz, 50 Ω Load): <–35 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.1 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage: 300 VDC or 300 Vrms (425 V peak) for AC waveforms between any terminal and chassis. Digital Input/Output (Channels 11–14) VIN(L): <0.8 V (TTL). VIN(H): >2 V (TTL). Vout(L): <1.0 V @ Iout = 100 mA. Vout(H): >2.4 V @ Iout = 1 mA. Vout(H)MAX.: <40 V with external open drain pull-up. Read/Write Speed 50/s. General 10 Channels 10 channels of 2-pole relay input. All channels configurable to 4-pole. Relay Type Latching electromechanical. Actuation Time <3 ms. Capacity 2700: (1) 7707 and (1) 77XX, except 7706. 2701: Any combination of 77XX modules. 2750: (4) 7707 and (1) 77XX, except 7706. A 7706 module may be substituted for a 7707 module. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight <0.5 kg (1.16 lb). Supplied Accessories 7707-306A 50-pin D-Sub Female IDC Connector Kit 7709-307A 25-pin D-Sub Male IDC Connector Kits Available Accessories 7790 50-pin Male, 50-pin Female, 25-pin Male D-shell IDC Connectors 7705-MTC-2 50 Pin Male to Female D-sub Cable, 2 m (6.6 ft). 7707-MTC-2 25 Pin Male to Female D-sub Cable, 2 m (6.6 ft).. Available Services 7707-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

TEK.COM 35 DAQ6510 Data Acquisition and Logging, Multimeter System The 7708 plug-in module offers 40 channels of 2-pole or 20 channels of 4-pole multiplexer switching that can be configured as two independent banks of multiplexers. The built-in CJC sensors automatically linearize thermocouples, making the 7708 ideal for RTD, thermistor, and thermocouple temperature applications. It is also well suited for mixed-signal measurement applications that require multi-point monitoring, such as environmental stress screening. Key Features

  • 40 differential channels for general-purpose measurements
  • Two- or four-wire measurements
  • 300 V, 1 A capacity for voltage channels; 60 W, 125 VA
  • Oversize screw terminal connection blocks are standard for easier connection
  • Relay closures stored in onboard memory Series 7700 Plug-In Switch Modules Channel 1 HI LO Channel 20 HI LO (Channels 2–19) Channel 21 HI LO Channel 40 HI LO (Channels 22–39) HI LO Sense HI LO Input Channel 4 12-Pole (Open) 4-Pole (Closed) (see Note) To Instrument Backplane Channel 42 (see Note) Backplane isolation Channel 43 (see Note) Backplane isolation NOTE: Card Sense HI LO Card Input HI LO Cold Junction Ref ×4 Cold Junction Ref ×5 Channels 41 –43 in this schematic refer to the designations used for control and not actual available channels. For more information, refer to the ROUTe:MULTiple command section in the Instrument User’s Manual. Channels 42 and 43 can be individually controlled using ROUTe:MULTiple if the module is not to be connected to the internal DMM. 7708 40-channel, Differential Multiplexer Module with Automatic CJC and Screw Terminals Datasheet Specifications Capabilities Channels 1–40 Multiplex one of 40 2-pole or one of 20 4-pole signals into DMM. Inputs Maximum Signal Level Channels (1–40) 300 V DC or rms, 1 A switched, 60 W, 125 VA maximum. Contact Life (typ.) >105 operations at max. signal level. >108 operations no load 1. 1. Open thermocouple detector on during thermocouple measurements. Minimum signal level 10 mV, 10 μA. Contact Resistance <2 Ω at end of contact life. Contact Potential <±500 nV typical per contact, 1 μ V max. <±500 nV typical per contact pair, 1 μ V max. Offset Current <100 pA. Connector Type Screw terminal, #20 AWG wire size. Isolation Between Any Two Terminals >109 Ω, <200 pF. Isolation Between Any Terminal and Earth >109 Ω, <400 pF. Cross Talk (1 MHz, 50 Ω Load) <–40 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.1 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage 300 V between any terminal and chassis. General 40 Channels 40 channels of 2-pole relay input. All channels configurable to 4-pole. Relay Type Latching electromechanical. Actuation Time <3 ms. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.52 kg (1.16 lb). Supplied Accessories CC-92-1 Set of 20 Cable Ties TL-23 Screwdriver Available Accessories 7708-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

7708 40-channel Differential Multiplexer Module with Automatic CJC and Screw Terminals

TEK.COM36 The 7709 plug-in module is a two-pole, 6×8 matrix module. It can connect any combination of six differential channels of instrumentation to any combination of eight differential device- under-test channels. The instrumentation can be AC and DC sources, internal or external meters, oscilloscopes, etc. This matrix configuration allows wide flexibility for complex test systems. Key Features

  • Automatic two- or four-wire connection to DMM
  • 6 row × 8 column matrix
  • Expandable to larger switch configurations by daisy-chaining or cascading multiple modules
  • Two female D-sub connectors are standard for secure hook-up and quick teardown
  • 300 V, 1 A capacity
  • Relay closures stored in onboard memory

7709 6×8 Matrix Module Series 7700 Plug-In Switch Modules 1 2 3 4 5 Matrix Crosspoint HI LO 6 7 8 HI LO Input HI LO Sense To DMM Backplane Columns Rows 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 1 2 3 4 5 6 7 8 7709 6×8 Matrix Module Datasheet Specifications Capabilities DMM Connection 2-Wire Functions Row 1, channels 1–8, through channel 50. 4-Wire Functions Row 1, channels 1–4 (Source to Input) through channel 50 and Row 2, channels 13–16 (Sense), through channel 49. Close Channel CLOSE command connects channels 1–8 to DMM. For 4-wire, channels 1–4 are automatically paired with channels 13–16. ROUTe:MULTiple allows any combination of rows and columns to be connected at the same time. Inputs Maximum Signal Level Any Channel to Any Channel (1–48): 300 VDC or 300 Vrms (425 V peak) for AC waveforms, 1 A switched, 60 W, 125 VA maximum. Contact Life (typ.) >105 operations at max. signal level. >108 operations no load 1. 1. Minimum signal level 10 mV, 10 μ A. Contact Resistance <1 Ω any path and additional 1 Ω at end of contact life. Contact Potential <3 μV per contact pair. Offset Current <100 pA. Connector Type 50-pin female D-shell for rows and columns. 25-pin female D-shell for “daisy-chain” rows. Supplied with male IDC ribbon cable connectors. Isolation Between Any Two Terminals >109 Ω, <200 F. Isolation Between Any Terminal and Earth >109 Ω, <400 pF. Cross Talk (1 MHz, 50 Ω Load) <–35 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <0.35 dB below 1 MHz. <3 dB below 2 MHz. Common Mode Voltage between any terminal and chassis. General Matrix Configuration 6 rows × 8 columns. Contact Configuration 2 pole Form A. Relay Type Latching electromechanical. Actuation Time <3 ms. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.52 kg (1.16 lb). Supplied Accessories 7709-306A 50-pin D-Sub Male IDC Connector Kit 7709-307A 25-pin D-Sub Male IDC Connector Kit Available Accessories 7789 50/25 Pin Male D-Shell Solder Cup Connectors 7790 50/50/25 Pin Female/Male D-Shell IDC Connectors 7705-MTC-2 50 Pin Male to Female D-sub Cable, 2m (6.6 ft). 7707-MTC-2 25 Pin Male to Female D-sub Cable, 2m (6.6 ft). Available Services 7709-3Y-EW 1-year factory warranty extended to 3 years from date of shipment

TEK.COM 37 DAQ6510 Data Acquisition and Logging, Multimeter System The 7710 plug-in module offers 20 channels of 2-pole or 10 channels of 4-pole relay input that can be configured as two independent banks of multiplexers. The relays are solid state, providing long life and low maintenance. Solid-state relays usually have 100 times longer life than mechanical relays. It is ideal for long-term data logging applications as well as for demanding high-speed applications. Key Features

  • 20 channels for general purpose measurements with scanning speeds up to 800 channels/s
  • High speed production or ATE testing up to 500 channels/s
  • Long lifetime solid state relay
  • Removable screw terminals Series 7700 Plug-In Switch Modules Card Source Card Sense Channel 1 Ch. 21 Ch. 22 Channel 10 Channel 11 Channel 20 HI LO HI LO HI LO HI LO HI LO HI LO HI LO HI LO To Mainframe BackplaneChannel 23 2-Pole (closed) 4-Pole (open) Cold Junction Ref 3x Cold Junction Ref 3x 7710 20-channel Solid-state Differential Multiplexer with Automatic CJC Datasheet Specifications Capabilities Maximum Signal Level Any channel to any channel (1–20) 60 VDC or 42 V rms, 100 mA switched, 6 W, 4.2 VA maximum. Common Mode Voltage

300 VDC or 300 Vrms (425 V peak) maximum between

any terminal and chassis. Contact Life >1010 operations at cold switching or max signal level (guaranteed by design). Relay Drive Current 6 mA per channel continuous, 25 mA during initial pulse. Contact Resistance <10 Ω per channel or <5 Ω per conductor. Refer to the

7710 User’s Guide for measurement considerations

when used on the instrument 1 Ω or 10 Ω ranges. Contact Potential <3 μV per pair. Offset Current <3 nA @ 23°C (per channel); additional 0.13 nA/°C >23°C. Connector Type 3.5 mm removable screw terminals, #20 AWG wire size. Isolation Between Any Two Terminals: >1010 Ω @ 23°C, > 8×109 Ω @ 50°C, <100 pF. Isolation Between Any Terminal and Earth: >1010 Ω, <100 pF. Crosstalk (CH-CH, 500 kHz, 50 Ω Load): <–40 dB. Insertion Loss (50 Ω Source, 50 Ω Load) <1 dB below 500 kHz. Scanning Speeds (see mainframe specifications for details) Multiple Channels, Into Memory 7710 Scanning DCV: >800. 7710 Scanning DCV alternating 2W Ω: >400. General Channels 20 channels of 2-pole relay input. All channels configurable to 4-pole. Relay Type Solid state opto-coupled FET. Actuation Time <0.5 ms (100 mA load). Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.45 kg (1 lb). Supplied Accessories CC-92-1 Set of 20 Cable Ties TL-23 Screwdriver Available Services 7710-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 7710 20-channel Solid-state Differential Multiplexer Module

TEK.COM38 The 7711 plug-in module provides an economical, wideband signal routing solution that complements the DC/low frequency switching and measurement capability of the Integra Series systems. The 7711 offers dual 1×4 configurations and can interface with a wide range of external AC instruments, including oscilloscopes, pulse generators, and signal analysis tools. One channel in each multiplex bank is always closed to the corresponding OUT connector. All connections are easily accessible from the rear panel. Key Features

  • Signal routing performance to 2 GHz
  • Switches up to 60 VDC
  • Rear panel SMA connections
  • Onboard switch closure counter
  • Onboard S parameter storage Series 7700 Plug-In Switch Modules Typical Insertion Loss 0 dB –1 dB –2 dB –3 dB –4 dB 10 100 1000 2000 Frequency (MHz) Typical VSWR 1.8 1.6 1.4 1.2 1.0 10 1001 000 2000 Frequency (MHz) 7711 2 GHz 50 Ω RF Module Datasheet Specifications Inputs (Channels 1–8) Maximum Signal Level Any channel to any channel or chassis (1–8):

30 Vrms (42 V peak for AC waveforms) or 60 VDC,

0.5 A. Maximum Power 20 W per module, 10 W per channel (refer to 7711/7712 Manual PA-818 for meas ure ment considerations). Isolation Multiplexer to Multiplexer: >1 GΩ. Center to Shield: >1 G Ω, <25 pF. Channel to Channel: >100 MΩ. Contact Life 1×106 no load, 1×105 rated load (resistive load). Contact Potential <6 μV. Contact Resistance <0.5 Ω (initial), <1 Ω (end of life). Rise Time <300 ps (guaranteed by design). Signal Delay <3 ns. General Relay Type High frequency electromechanical. Contact Configuration Dual 1×4 multiplexer, single pole four throw, Channels 1 and 5 are normally closed. Note: One channel in each multiplex bank is always closed to the corresponding OUT connector. Close Channel ROUTe:CLOSe allows a single channel in a multiplex bank to be closed. ROUTe:MULTiple:CLOSe allows two channels (one in each bank) to be closed at one time. Open Channel ROUTe:OPEN:ALL closes CH1 and CH5 to OUT A and OUT B respectively. Actuation Time <10 ms. Connector Type Ten external rear panel SMA connectors. Mating Torque 0.9 N·m (8 in-lb). Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.5 kg (1.1 lb). Available Accessories 7051-2 BNC Cable, male to male, 0.6 m (2 ft.) 7051-5 BNC Cable, male to male, 1.5 m (5 ft.) 7051-10 BNC Cable, male to male, 3.0 m (10 ft.) 7711-BNC-SMA Male SMA to female BNC Cables (5), 0.15 m (0.5 ft) 7712-SMA-1 SMA Cable, male to male, 1 m (3.3 ft) 7712-SMA-N Female SMA to Male N-Type Adapter S46-SMA-0.5 SMA Cable, male to male, 0.15 m (0.5 ft.) S46-SMA-1 SMA Cable, male to male, 0.3 m (1 ft.) S46-SMA-1.7 SMA Cable, Male to Male, 0.58 m (1.7 ft) Available Services 7711-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 7711 2 GHz 50 Ω RF Module

TEK.COM 39 DAQ6510 Data Acquisition and Logging, Multimeter System The 7712 plug-in module offers a 50W dual 1×4 multiplexer configuration with rear panel SMA 14 connectors. Multiple multiplexers can be cascaded to build scalable matrix and multiplexer systems for a large number of devices under test and RF source/measurement instruments. One channel in each multiplex bank is always closed to the corresponding OUT connector. The

3.5 GHz RF switching capability of the 7712 makes it ideal for testing

wireless modules operating in the 2.4 GHz and lower industrial, scientific, and medical (ISM) radio bands. Key Features

  • 3.5 GHz bandwidth
  • Dual 1×4 configuration
  • Onboard switch closure counter
  • Onboard S parameter storage Series 7700 Plug-In Switch Modules Typical Insertion Loss 0 dB –1 dB –2 dB –3 dB –4 dB 10 1001 000 4000 Frequency (MHz) Typical VSWR 1.8 1.6 1.4 1.2 1.0 10 1001 000 4000 Frequency (MHz) 7712 3.5 GHz 50 Ω RF Module Datasheet Specifications Inputs (Channels 1–8) Maximum Signal Level Any channel to any channel or chassis (1–8):

30 Vrms (42 V peak for AC waveforms) or 42 VDC,

0.5˛A. Maximum Power 20 W per module, 10 W per channel (refer to 7711/7712 Manual PA-818 for meas ure ment considerations). Contact Life 5×106 no load, 1×105 rated load (resistive load). Contact Potential <12 μV. Contact Resistance <0.5 Ω (initial), <1 Ω (end of life). Rise Time <200 ps (guaranteed by design). Signal Delay <1.5 ns. General Relay Type High frequency electromechanical. Contact Configuration Dual 1×4 multiplexer, single pole four throw, Channels 1 and 5 are normally closed. Note: One channel in each multiplex bank is always closed to the corresponding OUT connector. Close Channel ROUTe:CLOSe allows a single channel in a multiplex bank to be closed. ROUTe:MULTiple:CLOSe allows two channels (one in each bank) to be closed at one time. Open Channel ROUTe:OPEN:ALL closes CH1 and CH5 to OUT A and OUT B respectively. Actuation Time <10 ms. Connector Type Ten external rear panel SMA connectors. Mating Torque 0.9 N·m (8 in-lb). Isolation Multiplexer to Multiplexer: >1 GΩ. Center to Shield: >1 G Ω, <20 pF. Channel to Channel: >100 MΩ. Environmental Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C. Storage Environment: –25° to 65°C. EMC: Conforms to European Union EMC Directive. Safety: Conforms to European Union Low Voltage Directive RoHS: Conforms to European Union RoHS Directive Warranty: 1 year Weight 0.5 kg (1.1 lb). Available Accessories 7712-SMA-1 SMA Cable, male to male, 1 m (3.3 ft) 7712-SMA-N Female SMA to Male N-Type Adapter S46-SMA-0.5 SMA Cable, male to male, 0.15 m (0.5 ft.) S46-SMA-1 SMA Cable, male to male, 0.3 m (1 ft.) S46-SMA-1.7 SMA Cable, Male to Male, 0.58 m (1.7 ft.) Available Services 7712-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 7712 3.5 GHz 50 Ω RF Module

Australia* 1 800 709 465 Austria 00800 2255 4835 Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777 Belgium* 00800 2255 4835 Brazil +55 (11) 3759 7627 Canada 1 800 833 9200 Central East Europe/Baltics +41 52 675 3777 Central Europe/Greece +41 52 675 3777 Denmark +45 80 88 1401 Finland +41 52 675 3777 France* 00800 2255 4835 Germany* 00800 2255 4835 Hong Kong 400 820 5835 India 000 800 650 1835 Indonesia 007 803 601 5249 Italy 00800 2255 4835 Japan 81 (3) 6714 3086 Luxembourg +41 52 675 3777 Malaysia 1 800 22 55835 Mexico, Central/South America and Caribbean 52 (55) 56 04 50 90 Middle East, Asia, and North Africa +41 52 675 3777 The Netherlands* 00800 2255 4835 New Zealand 0800 800 238 Norway 800 16098 People’s Republic of China 400 820 5835 Philippines 1 800 1601 0077 Poland +41 52 675 3777 Portugal 80 08 12370 Republic of Korea +82 2 6917 5000 Russia/CIS +7 (495) 6647564 Singapore 800 6011 473 South Africa +41 52 675 3777 Spain* 00800 2255 4835 Sweden* 00800 2255 4835 Switzerland* 00800 2255 4835 Taiwan 886 (2) 2656 6688 Thailand 1 800 011 931 United Kingdom/Ireland* 00800 2255 4835 USA 1 800 833 9200 Vietnam 12060128 * European toll-free number. If not accessible, call: +41 52 675 3777 Rev. 090617 Find more valuable resources at TEK.COM in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies.

031918 SBG 1KW-61317-0