C7700 HAMAMATSU | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 4
Technical content
High Dynamic Range Streak Camera C7700 L The relaxation oscillation of a laser diode can be measured with the C7700 under single shot operation. (The data was captured with a C4880-21 readout CCD camera.) Less than 60 ps The C7700 is a newly developed high dynamic range streak camera that can handle a large number of photo- electrons. This feature enables single-shot measurements of ultrafast phenomena with a D-range as high as 10 000 : 1. This system is suitable for high-precision simultaneous measurement of high-intensity and weak intensity pulse light. 10 000 :1
7 Readout System
(HPD-TA)
6 Readout Cameras
Frame Grabber Streak Image Analysis Software Streak Image Analysis Systems for IBM PC/ATDual-Scan Cooled CCD Camera C4880-21 Camera head Control unit
1 Input Optics
3 Main Unit
5 Output Optics ORCA 2
(IBM PC/AT Compatible) DUAL MODE COOLED CCD CAMERA POWER High Dynamic Range Measure
APPLICATIONS
G Research involving laser fusion lasers, free electron lasers and various other types of pulsed lasers G Plasma light emission, radiation, laser ab- lation, combustion and explosion G Picosecond transient absorption measurement (Time dependence of absorption is shown below.) G LIDAR, Thomson scattering, time-of-flight laser ranging G Fluorescence lifetime measurement, time- resolved Raman spectroscopy SYSTEM CONFIGURATION Optical intensity Time Space Trigger signal Slit Lens Incident light Photocathode (light → electrons) Accelerating electrode (where electrons are accelerated) Phosphor screen (electrons → light) Sweep circuit Streak image Time Space
2 Streak Tube
4 Image Intensifier,
(where electrons are swept in the direction from top to bottom) [Operating Principle] The light pulse to be measured is focused onto the photocathode of the streak tube through the slit, where the photons are con- verted into a number of electrons proportional to the intensity of the incident light. These electrons are accelerated and conducted towards the phosphor screen, and a high-speed voltage which is synchronized to the incident light is applied. The electrons are swept at high speed from top to bottom, after which they are bom- barded against the phosphor screen of the streak tube and con- verted to an optical image. When the light intensity of the streak image is very weak, an image intensifier or an image booster amplifies the low light level streak image. ∆OD = 0.1 0 12 34 5 Time (ns) S1 - Sn absorption ∆λ = 482 to 489 nm T1 - Tn absorption ∆λ = 557 to 581 nm L The 266 nm excitation (25 ps, 0.2 mJ, 2 mm focused, single shot) of Chrysene in THF (0.5 mmol/l) D-range of 10 000:1 of ultrafast phenomena under single-shot operation
FEATURES
G High dynamic range of 10 000:1* G Temporal resolution of 5 ps* G Effective photocathode size: 17 mm G Simultaneous measurement of light intensity on temporal and spatial (wavelength) axis G IEEE 488 (GP-IB) control *[About a Dynamic Range] The maximum dynamic range of the streak camera essence is set to 1000:1 by the measurement condition of temporal resolution 5ps, and is set to 10 000:1 by the measurement condition of temporal-resolution 100ps. With a light intensity, the image intensifier tube, such as an image intensifier or an image booster, is combined with a streak camera. In this case, the image intensifier tube may restrict the dynamic range of the measurement system. Furthermore, a read-out camera may also restrict the dynamic range of the measurement system. Image Intensifier Tube C4880-21 ORCA 2 Readout Camera Image Intensifier Image Intensifier Image Booster Image Booster Temporal Resolution 5 ps 100 ps 1000:1 1000:1 2000:1 2000:1 10 000:1 3000:1
ment Possible up to 10 000:1 SPECIFICATIONS No. Overall Length Slit Width Reading Precision Image Multiplica- tion Ratio Spectral Transmission Effective F Value Slit Width A1974 400 nm to 900 nm 1.2 1 : 1 0 to 5 mm 5 µm 159 mm A1976-04 200 nm to 1060 nm 3.5 1 : 1 98.2 mm Wavelength (nm) Spectral transmittance of input optics Transmittance (%) 1000 1200 1400 1600 4 Image Intensifier (I.I.), Image Booster (I.B.)
5 Output Optics
(I.I.) S-20 25 mm P-43 3000 V3346U-24 (I.B.) 15 (typ.) Effective Photocathode Size Phosphor Screen Luminous Gain A7702-02 Magnification Readeout Camera Type No. Imaging Device C4880-21 Full frame transfer CCD 512(H) × 512 (V) 24 µm × 24 µm ORCA 2 Progressive-scan interline CCD 1280(H) × 1024(V) 6.7 µm × 6.7 µm Effective No.of Pixels Pixel SizeType No. No. Effective Photocathode Size N8059 200 nm to 850 nm 7 mm × 17 mm N8059-02 300 nm to 1060 nm 18 lp/mm or more centered on photocathode Note) read- out camera ORCA 2
- Phosphor screen characteristic P-43
- Fiber-optic output
- Effective Phos- phor screen size: 25 mm Phosphor Screen Spatial Resolution Spectral Response Characteristic Streak Tube 105 104 103 102 101 200 400 600 Wavelength (nm) Spectral response characteristics Radiant sensitivity (µA/W) 800 1000 1200 C7700-11 C7700-01
- Temporal resolution: better than 5 ps (at the fastest sweep range)
- Sweep time/full screen0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500 ns 1, 2, 5, 10, 20, 50, 100, 200, 500 µs, 1 ms
- Trigger jitter: less than ± 20 ps (at the fastest sweep range)
- Trigger delay: about 25 ns (at the fastest sweep range)
- Maximum sweep repetition frequency
- Trigger input
- Line voltage
- Operating humidity Less than 70 % (with no condensation) UTILITY The HPD-TA (Temporal Analayzer) is a high-perfomance digital data acquisition and control system specifically designed to read- out images from the phosphor screen of the HAMAMATSU streak camera. It enables precise, quantitative acquisition and preanalysis of two dimensional streak data that includes photon counting plus a full range of data correction and calibration possibilities. The HPD-TA allows the remote control of the C7700 via GPIB interface. The entire system is controlled through a powerful and user-friendly application software that runs on a Microsoft Windows platform C4880-21 Effective F-number 1 : 0.7 F = 2.0 Type No. A2098 ORCA 2 : 1 F = 2.5 2
# IBM is a registered trademark of IBM Co. # Windows is a trademark of Microsoft Corporation in the U.S.A. # is a registered trademark of Apple Computer, Inc. Cat. No. SSCS1049E07 APR/2002 HPK Created in Japan (PDF) Delay Unit C1097-01 This passive delay unit provides convenient timing adjustment with zero jitter. Optional remote GPIB control (C1097-04). Variable Delay Range 0 to 31.96 ns Delay Setting Range 30, 60, 120, 250, 500 ps, 1, 2 4, 8, 16 ns Minimum Delay Time Approx. 12 ns Maximum Input Voltage 30 V Power Supply Dimensional Outline 215 (W) × 350 (D) × 102 (H) mm/3.4 kg G G PIN Photodiode C1083-01 Converts low-repetition light pulses to an electronic trigger for streak sweep. Spectral response 400 to 1100 nm Rise time 0.8 ns Power supply +22.5 V (battery) Dimensional Outline Head: 100 (W) × 160 to 235 (H) × 50 (D) mm/400 g Power supply unit: 100 (W) × 83(H) × 100 (D) mm/400 g DIMENSIONAL OUTLINES (Unit: mm) G Main Unit (approx. 24.0 kg) 285 ± 1 (C-mount)160.5 ± 3106.6 ± 1 390 ± 140.5 ± 1 250 ± 1 160 ± 0.5 85 ± 0.5 310 ± 1 HIGH DYNAMIC RANGE STREAK CAMERA C7700 A1976-04 A2098 # Product and software package names noted in this documentation are trademarks or registered trademarks of their respective manufacturers. G Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office. G Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearance are subject to change without notice.. © 2002 Hamamatsu Photonics K.K. Homepage Address http://www.hamamatsu.com HAMAMATSU PHOTONICS K.K., Systems Division 812 Joko-cho, Hamamatsu City, 431-3196, Japan, Telephone: (81)53-431-0124, Fax: (81)53-435-1574, E-mail:export@sys.hpk.co.jp Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658, E-mail: info@hamamatsu.de, United Kingdom:Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire, AL7 1BW, U.K., Telephone: (44) 1707-294888, Fax: (44) 1707-325777, E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 Solna, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01, E-mail: info@hamamatsu.se - Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Mois, 1/E 20020 Arese (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741, E-mail: info@hamamatsu.it ISO 9001 Certificate: 09 105 79045 AC100 V to 240 V