Measurements Ranging From X-Ray to Near Infrared With a Temporal Resolution of 2 ps The streak camera is an ultra high-speed detector which captures light emission phenomena occurring in extremely short time periods. Not only can the streak camera measure intensity variations with superb temporal resolution, but it can also be used for simultaneous measurement of the spatial (or spectral) distribution. The C5680 Streak Camera Series is a universal streak camera which incorporates all of the specialized technology and expertise HAMAMATSU has acquired in over 20 years of research. The streak tubes are manufactured on a regular production schedule at Hamamatsu to provide consistency and reliability. Special requests and custom designs are also available.
APPLICATIONS
Measurement of electron bunch for synchrotron and LINAC applications Research involving X-ray lasers, free electron lasers, and various other types of pulsed lasers Plasma light emission, radiation, laser ablasion, combustion and explosions Fluorescence lifetime measurement, transient absorption measurement, time-resolved raman spectroscopy Optical soliton communications, response measurement with quantum devices Lidar Thomson scattering, laser distance measurement Universal Streak Camera C5680 Series
FEATURES
- Temporal resolution of within 2 ps A temporal resolution of 2 ps is achieved for both synchroscan and single shot.
- Ultra-high sensitivity (detection of single photons) Several plug-in module, operating mode. Accommodates a diverse range of experimental setups from single light emitting phenomena to high-speed repeated phenomena in the GHz. Can be used in X-ray to near infrared fields By selecting the appropriate streak tube (light sensor), the C5680 can be used in a wide range of measurement appli- cations, from X-rays to near infrared light. Simultaneous measurement of light intensity on temporal and spatial (wavelength) axes Spectrograph can be placed in front of the streak camera, to convert the spatial axis to a wavelength axis. This enables changes in the light intensity to be measured over various wavelength (time-resolved spectroscopy). The streak tube converts light into electrons which are then multiplied by an electron multiplier. This enables detection of extremely faint light (at the single-photon level). (See photon counting integration principle) IEEE-488 (GP-IB) control Computer control enables remote control and advanced measurements to be performed out using very simple op- eration. Diverse selection of peripheral equipment A full lineup of peripheral devices is available, including spectroscopes, optical trigger heads, and expansion units. OPERATING PRINCIPLE The light pulse to be measured is projected onto the slit and is focused by the lens into an optical image on the photocathode of the streak tube. Changing the temporal and spatial offset slightly each time, four light pulses, each with a different light itensity, are introduced through the slit and conducted to the photocathode. Here, the photons are converted into a number of electrons proportional to the intensity of the incident light. The four light pulses are converted sequentially to electrons which are then accelerated and conducted towards the photocathode. As the group of electrons created from the four light pulses passes between a pair of sweep electrodes, a high voltage is applied (see above), resulting in a high-speed sweep (the elec- trons are swept in the direction from top to bottom). The elec- trons are deflected at different times, and at slightly different angles in the perpendicular direction, and are then conducted to the MCP (micro-channel plate). As the electrons pass the MCP, they are multiplied several thousands of times and are then bombarded against the phos- phorscreen, where they are converted back into light. The fluorescence image corresponding to the first incident light pulse is positioned at the top of the phosphor screen, follo- wedby the others, with images proceeding in descending or- der; inother words, the axis in the perpendicular direction on the phosphor screen serves as the temporal axis. The bright- nesses ofthe various fluorescence images are proportional to theintensities of the corresponding incident light pulses. The positions in the horizontal direction on the phosphor screen correspond to the positions of the incident light in the horizon- tal direction. THE PRINCIPLE OF PHOTON COUNTING INTEGRATION Photoelectrons given off from the photocathode of the streak tube are multiplied at a high integration rate by the MCP, and one photoelectron is counted as one intensity point on the phosphor screen. A threshold value is then used with this pho- toelectron image to clearly separate out noise. The operating principle of the streak camera Positions in the photoelectron image which are above the threshold value are detected and are integrated in the memory, enabling noise to be eliminated completely. This makes it possi- ble to achieve data measurements with a high dynamic range and high S/N. Photon Counting Integration Separation of Photoelectron Image and Noise Optical intensity Time Space Trigger signal Slit Lens Incident light Photocathode (light→ electrons) Accelerating electrode (where electrons are accelerated) Phosphor screen (electrons→ light) MCP (which multiplies electrons) Streak image on phosphor screen Time Space Sweep electrode (where electrons are swept in the direction from top to bottom) Sweep circuit The intensity of the incident light can be read from the brightness of the phosphor screen, and the time and space from the position of the phosphor screen. A/D conversion value Signal output from CCD camera Time (wavelength)Noise Photoelectron image Threshold value 0ps Light source: PLP (λ = 800 nm) Integration time: 1 min. 200ps400ps600ps800ps 1ns 1.2ns 1.4ns 1.6ns1.8ns
Wavelength (nm) Spectral response of the streak tube Radiant sensitivity (µA/W) 800 1000 1200 1400 1600 N5716-01 N5716, N5864 N5716-02 N5716-03 Selection of C5680 main unit Selection of input optics system Selection of streak tube Selection of output format Selection of sweep unit Selection of function expansion unit C5680 Main Unit [Suffix (Model No.)]
2 Lens output type
3 Video output type
. . . .
1 Input Optics System
A1976-01 200 nm to 1600 nm 5.0 1 : 1 98.2 mm A1974 400 nm to 900 nm 1.2 1 : 1 0 to 5 mm 5 µm 159 mm A1974-01 400 nm to 1600 nm 1.2 1 : 1 159 mm A1976-04 200 nm to 1600 nm 3.5 1 : 1 98.2 mm The A1974 and A1974-01 are optional units. Input optics system Output format Streak tube
2 Streak Tube
0.15 × 5.3 mm Lens output type 0.15 × 4.8 mm Video output type N5716 200 nm to 850 nm N5716-02 300 nm to 1600 nm N5716-01 115 nm to 850 nm N5716-03 200 nm to 900 nm N5864 200 nm to 850 nm 25 lp/mm or more centered on photocath- ode Photocathode characteristic P-43 Fiber-optic output Effective photo- cathode size 18 mm 6 × 10 Phosphor Screen Spatial Resolution MCP Gain £Function expansion unit ™Sweep unit 100 0 200 400 600 A1974 A1976-01 A1974-01 800 Wavelength (nm) Spectral transmittance of input optics system Transmittance (%) 1000 1200 1400 1600 FUNCTION CONFIGURATION C5680 Main Unit (with power supply and camera controller) SPECIFICATIONS One of the following suffixes is appended to the model number of the C5680, depending on the type of streak tube and output format used.
1 Accommodates 200 nm to 850 nm, 1 MCP
2 Accommodates 300 nm to 1600 nm, 1 MCP
3 Accommodates 115 nm to 850 nm, 1 MCP
4 Accommodates 200 nm to 900 nm, 1 MCP
5 Accommodates 200 nm to 850 nm, 2 MCPs
X-ray streak cameras designed for use with 10 eV to 10 keV can also be selected. Effective F Value 3 × 10 Effective Photocathode Size Spectral Response Characteristic
3 Output Formats
- Video output...
4 Other 5680 Specifications
- Gate Gating Method 50 ns to continuous 50 ns to continuous
- M5675 Synchroscan Unit Better than 3 ps at 800 nm (N5716-02) Sweep range Video output type .150 ps to 1/6 fs (fs:synchroscan frequency) Lens output type ...200 ps to 1/6 fs Synchronous frequency range .. fs ± 0.2 MHz (fs = synchroscan frequency)
- M5676 Fast Single Sweep Unit Maximum sweep repetition frequency (max.)...
- M5677 Slow Single Sweep Unit ... Magnification Effective F value F-mount 1 : 0.7 (50 mm : 35 mm) F/2.0 Signal format Coupling method Resolution CCIR or RS-170 Fiber optics 768 × 493 or 756 × 581 pixels Sweep units (Plug-in: built into main unit) D sub-connector DB-25S, 16-bit parallel output, open collector Max. horizontal blanking repetition frequency Max. MCP gate repetition frequency Max. photocathode gate repetition frequency Gate trigger delay time Gate trigger input Interface Line voltage Power consumption Factory set within a range of 75 MHz to 165 MHz Better than 2 ps at 800 nm (1.5 ps typ.) 0.15, 0.5, 1, 2, 5, 10, 20, 50 ns/full screen 0.2, 0.5, 1, 2, 5, 10, 20, 50 ns/full screen ± 5 V/50 Ω Approx. 13 ns (fastest range) Better than 20 ps 10 kHz Sweep time Maximum sweep repetition frequency (max.).. Better than 50 ps 50 ns to 1 ms/full screen ± 5 V/50 Ω Approx. 45 ns (fastest range) Better than temporal resolution
2 MHz (fastest range)
Function Expansion Units (connected to top of main unit) M5678 Synchronous Blanking Unit (designed for use in conjunction with M5675 Synchroscan) Horizontal shift width Factory set within a range of 75 MHz to 165 MHz 2.5 mm or 11 mm (at phosphor screen) M5679 Dual Time Base Extender Unit (Can be used in conjunction with all sweep units) Sweep time Trigger signal input ± 5 V/50 Ω 10 ns to 100 ms/full screen 10 HzMaximum sweep repetition frequency (max.) Gate Time 1 : 106 min. 1 : 108 min. Gate Extin- ction Ratio MCP + horizontal blanking MCP + horizontal blanking + photocathode
READOUT SYSTEM (HPD-TA) Video output (Video CCD Camera) Cooled digital camera C4742-95 Series (ORCA) Camera head Femtosecond Streak Camera Control unit Frame Grabber Streak Image Analysis Software Streak Image Analysis Systems for IBM® PC/AT Lens output Personal computer (IBM PC/AT Compatible) GP-IB board The HPD-TA (Temporal Analyzer) is a high-performance digital data acquisition and control system specifically designed to read out im- ages from the Hamamatsu streak camera’s phosphor screen. It ena- bles precise, quantitative acquisition and pre-analysis of two dimen- sional streak data that includes photon counting plus a full range of data correction and calibration possibilities. It possible to select the best camera for a given streak configuration and application. The camera is connected to an IBM-compatible PC/AT via a frame grab- ber board that can support real-time data transfer. The HPD-TA allows the remote control of the C5680 via GPIB in- terface. The entire system is controlled through a powerful but userfriendly software application that runs on a Microsoft Win- dows platform. * A read out system based on the Macintosh® computer is also available. Please consult with our sales office for more details.
- Functions & Specifications Computer Environment
1 Input optics
Resolution (pixels) Dynamic range Single frame Integration Frame rate Superpixel mode Subarray scan mode Single exposure time Analog integration Photon counting Dark correction Shading correction Curvature correction Calibration Multiple profiles Data export (images) Data export (profiles) Streak camera interface Other devices interf The HPD-TA requires an industry-standard Pentium-class (or com- patible) PC with a 32-bit Microsoft Windows version. A fast, high- resolution graphics configuration is recommended. Depending on the streak camera system configuration, a number of PCI and/or ISA slots as well as a serial interface port may be occupied. (Please consult Hamamatsu for a detailed specification for a given case.) Cooled CCD version C4742-95 Series (ORCA) Relay lens 1280 × 1024 10 or 12 bits 16 bits
9 Hz (normal) / 18 Hz (super pixel)
132 µs to 10 s on chip / into memory linear / nonlinear, both ax up to 10 Binary,TIFF, ASCII ASCII GPIB or StatusPort GPIB Video version Video CCD Fiber optics 756 × 581 8 bits 16 bits 30 Hz 40 ms or 33 ms into memory linear / nonlinear, both axes up to 10 Binary, TIFF, ASCII ASCII GPIB or StatusPort GPIB Mount Table A1471-12 Readout system (HPD-TA) Items
¶Input section
- Trigger unit §Optical trigger
- PIN Diode Head C1083-01 (for Low Repetition)
6 Optical Trigger (PIN diode head)
7 Input Section
- Spectroscopes C5094 and C5095
- PIN Diode Head C1808-03 (for High Repetition) Spectral response400 nm to 1100 nm Rise time 0.8 ns Dimensions/weight Power supply +22.5 V (battery) Head: 100 (W) × 160 to 235 (H) × 50 (D) mm/400 g Power supply unit: 100 (W) × 83(H) × 100 (D) mm/400 g <100 MHz Minimum input level Saturation output level Frequency band Power supply 1 mW (f=80 MHz, λ=800 mm, FWHM<1 ps) Approx. 1.5 Vp-p (50 Ω ) INPUT Voltage range 100 V to 240 V Input power supply frequency range 50/60 Hz C5094 C5095 Optical layout Czerny-Turner model (with toroidal mirror for aberration correction) Focal distance 250 mm 500 mm F value Incident light slit widthVariable between 10 µm to 2,000 µm Grating Up to 3 can be installed simultaneously Reciprocal dispersion 2.5 nm/mm 1.5 nm/mm (when using 1200 gr/mm) (when using 1200 gr/mm) Wavelength resolution < reciprocal dispersion × 0.06 The following are needed in order to connect these units to the C5680: A spectroscope mounting table A spectroscope adaptor A light source for wavelength axis calibration (mercury lamp, etc.)
- Fiber-optic Input Optics System (FC Connector) A6368 This fiber-optic input optics system can be connected in place of theincident light slit in the C5680.
- Objective Lens Connecting a C-mount adaptor to the incident light slit section of the C5680 enables attachment of a C-mount objective lens. F-mount ob- jective lenses can also be attached using an FC converter.
8 Trigger Units
- Delay Unit C1097-01 This unit can be used to align the operation timing of the streak camera with the target phenomenon. *The C1097-04, which has a GP-IB interface, is also available. 0 to 31.96 ns 30, 60, 120, 250, 500 ps, 1, 2 4, 8, 16 ns Approx. 12 ns 30 V AC85 V to 250 V Variable delay range Delay setting range Minimum delay time Maximum input voltage Power supply External dimensions/weight
- High-stability Delay Unit C6878 Used in combination with a synchroscan unit, this unit is used to adjust the delay times of trigger signals. In addition, the amount by which trigger signals are delayed is adjusted automatically, while monitoring the sweep signal, enabling stable acquisition of streak images over a long period of time. RF Up Converter Unit C6207 This outputs an output signal of 100 MHz synchronized to the 10 MHz input signal. Inputting reference output signals from a commercial frequency synthesizer enables stable synchroscan triggers to be obtained. Input signal frequency Input level Output frequency Output signal level Timing jitter Power supply
10 MHz ± 10 Hz
–10 dBm to 0 dBm/50 Ω
100 MHz
3 dBm/50 Ω (typ.) σ: 1 ps max. AC100/117/220/240V, 50/60 Hz
- Other Numerous other peripheral devices are also available, such as the DG535 Digital Delay Generator and the Picosecond Light Pulser PLP Series. Please feel free to contact HAMAMATSU concerning these and other devices.
DIMENSIONAL OUTLINES (Unit: mm) 300 290 160 A1976-01 C5680-1X M5676/M5677 128 360 70 160 A1976-01 C5680-0X M5676/M5677 128 360 23 160 230 320 14912 225 135 150 50 360 135 11 38 147 360 135 11 147 360 135 11 147 360 300 51.5 360 300 51.5 C5680 main unit (approx. 20 kg) C5680-1X (video output)C5680-0X (lens output) Power supply unit (approx. 10 kg) Remote control unit (approx. 1.2 kg) Synchroscan Unit M5675 (approx. 4.1 kg) Fast Single Sweep Unit M5676 (approx. 2.4 kg) Slow Single Sweep Unit M5677 (approx. 2.2 kg) Synchronous Blanking Unit M5678 (approx. 3.4 kg) Dual Time Base Extender Unit M5679 (approx. 3.4 kg) ISO 9001 Certificate: 09 105 79045 # IBM is a registered trade mark of IBM Co. # is registered trademark of Apple Computer, Inc. # Product and software package names noted in this documentation are trademarks or registered trademarks of their respective manufacturers.
- Subject to local technical requirements and regulations, availability of products included in this promotional material may vary. Please consult with our sales office.
- Information furnished by HAMAMATSU is believed to be reliable. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearance are subject to change without notice. © 2003 Hamamatsu Photonics K.K. HAMAMATSU PHOTONICS K.K., Systems Division 812 Joko-cho, Hamamatsu City, 431-3196, Japan, Telephone: (81)53-431-0124, Fax: (81)53-435-1574, E-mail:export@sys.hpk.co.jp Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49)8152-375-0, Fax: (49)8152-2658, E-mail:info@hamamatsu.de United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road Welwyn Garden City Hertfordshire AL7 1BW U.K., Telephone: (44)1707-294888, Fax: (44)1707-325777, E-mail: info@hamamatsu.co.uk North Europe: Hamamatsu Photonics Norden AB: Smidesvägen 12, SE-171-41 Solna, Sweden, Telephone: (46)8-509-031-00, Fax: (46)8-509-031-01, E-mail:info@hamamatsu.se Italy: Hamamatsu Photonics Italia S.R.L.: Strada della Moia, 1/E 20020 Arese (Milano), Italy, Telephone: (39)02-935 81 733, Fax: (39)02-935 81 741, E-mail:info@hamamatsu.it Homepage Address http://www.hamamatsu.com Cat. No. SSCS 1046E07 AUG/2003 HPK Created in Japan (PDF)