53150A KEYSIGHT | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 8

Technical content

Full-featured CW Microwave Counters for Field, Factory or Lab 53150A 20 GHz Counter 53151A 26.5 GHz Counter 53152A 46 GHz Counter Technical Overview

microwave counters: at home, in the field, bench or ATE environment – Ultrawide range, single input (from 50 MHz up to 46 GHz) – Simultaneous power and frequency measurement with analog peaking indicator – GPIB and RS-232 standard – Lightweight and rugged – Battery optional Convenience, portability and outstanding performance The innovative designs of the Keysight Technologies, Inc. 53150 Series microwave counters offer an uncluttered, feature-laden front panel. These designs present no- compromise performance and quality in a surprisingly small, light, battery operated product. The convenience of a single microwave input The Keysight 53150 Series has an advanced sampler that integrates a separate zero bias Schottky diode for the accurate measurement of input power. This allows measurement of both frequency and power with a single connection. No compromise in frequency coverage is required for this capability. The ultrawideband microwave input covers the entire RF and microwave spectrum, from intermediate frequencies IFs) of 50 MHz to millimeter waves. The power measurement accuracy and repeatability of these counters rivals power meters with diode sensors. Since frequency and power of the input signal are measured simultaneously, adjustment for the diode’s frequency response is done automatically. And like the latest in diode sensors, compensation is also made for deviation from square law. Field tough but ready for benchtop or ATE

applications

The Keysight 53150 Series is as comfortable in the field as in the laboratory. The rugged case with an integrated tilting handle can tolerate the vibration and shock expected in field use. The backlit LCD display ensures visibility in all environments, from dark to full sunlight, at distances exceeding 15 feet. If AC power is unavailable, the internal, replaceable camcorder batteries pro- vide at least 2.5 hours of continuous operation. The unit can also be powered from an external 11-18 VDC source. For benchtop and ATE applications, the Keysight 53150 Series delivers full functionality and high measurement speed. The fully programmable RS-232 interface and high speed GPIB interface are standard features. No compromise performance The Keysight 53150 Series offers exceptional sensitivity by utilizing a single board design with low phase noise PLL circuitry. Despite their simple appearance, these counters retain all the powerful functions one expects in precision instrumentation: measurement averaging, arbitrary as well as nulling offsets for both frequency and power, display of power in either dBm or Watts. Additional capabilities include full control of resolution, sampling rate, and GPIB address, plus extensive self-diagnostics, fast acquisition times and full programmability. Performance surpasses the industry standard Keysight 5350 Series, in virtually every aspect, in a package less than half the weight and size.

Measurement Specifications and Characteristics All measurement specifications are over the full signal and temperature ranges unless otherwise noted. Input characteristics 53150A 53151A 53152A Frequency range Channel 1 (Normal mode) (Low pass filter enabled) Channel 2

10 Hz - 125 MHz

10 Hz - 50 kHz

50 MHz - 20 GHz

50 MHz - 26.5 GHz

50 MHz - 46 GHz

30 Hz-125 MHz

0.3-12.4 GHz 12.4-18 GHz 18-20 GHz 20-26.5 GHz 26.5-40 GHz 40-46 GHz –20 dBm –33 dBm –33 dBm –29 dBm N/A N/A N/A –20 dBm –33 dBm –33 dBm –29 dBm –25 dBm N/A N/A –20 dBm –33 dBm –30 dBm –27 dBm –27 dBm –23 dBm –17 dBm Maximum input Channel 1 2 Vrms 2 Vrms 2 Vrms Channel 2

50 MHz - 2 GHz

+5 dBm +13 dBm +5 dBm +13 dBm +5 dBm +13 dBm Damage level Channel 1 120 V (DC + AC pk) linearly derated to 5 Vrms at 125 MHz

120 V (DC + AC pk) linearly

derated to 5 Vrms at 125 MHz derated to 5 Vrms at 125 MHz Channel 2 +27 dBm +27 dBm +27 dBm Impedance (Nominal) Channel 1 1 MΩ/60 pF 1 MΩ/60 pF 1 MΩ/60 pF Channel 2 50 Ω 50 Ω 50 Ω Connector Channel 1 Channel 2 BNC female SMA/APC-3.5 compatible female BNC female SMA/APC-3.5 compatible female BNC female 2.92 mm removable, SMA/APC-3.5 compatible female SWR (Typical) Channel 2 50-300 MHz 0.3-10 GHz 10-20 GHz 20-26.5 GHz 26.5-46 GHz 1.5:1 typical 2.0:1 typical 3.0:1 typical N/A N/A 1.5:1 typical 2.0:1 typical 3.0:1 typical 3.0:1 typical N/A 1.5:1 typical 2.0:1 typical 3.0:1 typical 2.5:1 typical 2.5:1 typical Coupling Channel 1 Channel 2 AC AC AC AC AC AC Emissions (Typical) (“kickback noise“) Channel 1 Channel 2 (measuring/no input) N/A –40 dBm/< –70 dBm N/A –40 dBm/< –70 dBm N/A –40 dBm/< –70 dBm Resolution Channel 1/Channel 2 1 Hz to 1 MHz 1 Hz to 1 MHz 1 Hz to 1 MHz

Measurement Specifications and Characteristics Continued... Input characteristics 53150A 53151A 53152A Accuracy Channel 1/Channel 2 (LSD = resolution selection) ±1 LSD ± residual stability ±(timebase error x frequency) ±1 LSD ± residual stability ±(timebase error x frequency) ±1 LSD ± residual stability ±(timebase error x frequency) Residual stability* Channel 1 Channel 2 *Counter and source tied to same timebase N/A

0.6 LSD rms

0.8 LSD rms

1.25 LSD rms

Measurement time (Typical) Channel 1 Channel 2 1/Resolution + 20 ms 1/Resolution + Acquisition time + 20 ms 1/Resolution + 20 ms 1/Resolution + Acquisition time + 20 ms 1/Resolution + 20 ms 1/Resolution + Acquisition time + 20 ms Acquisition time (Typical) (1 MHz FM rate, power measurement off) Channel 1 Channel 2 (FM Auto/FM Off) N/A 125 ms/100 ms N/A 125 ms/100 ms N/A 140 ms/115 ms FM tolerance Channel 1 Channel 2 (FM Auto) (FM Off) N/A

20 MHz p-p max @ 10 MHz rate

1 MHz p-p @ 10 MHz rate

20 MHz p-p max to 26.5 GHz, 12 MHz p-p max above 26.5 GHz @ 10 MHz rate Channel 1, Channel 2 Any index provided minimum signal level is not less than sensitivity Any index provided minimum signal level is not less than sensitivity Any index provided minimum signal level is not less than sensitivity Amplitude discrimination Channel 1 Channel 2 below 300 MHz above 300 MHz N/A N/A Automatically measures the larg- est signal present provided signal is > 10 dB(typical) above any sig- nal separated by less than 75 MHz; > 20 dB (typical) above any signal separated by more than 75 MHz N/A N/A Automatically measures the larg- est signal present provided signal is > 10 dB(typical) above any sig- nal separated by less than 75 MHz; > 20 dB (typical) above any signal separated by more than 75 MHz N/A N/A Automatically measures the larg- est signal present provided signal is > 10 dB(typical) above any sig- nal separated by less than 75 MHz; > 20 dB (typical) above any signal separated by more than 75 MHz Power measurement Channel 1 Channel 2 N/A N/A N/A Range Accuracy at input connector (0 dBm to –20 dBm) 0.05-12.4 GHz 12.4-20 GHz 20-26.5 GHz 26.5-46 GHz Resolution Display Counter sensitivity to +7 dBm ±1.5 dB ±1.5 dB N/A N/A 0.01 dB dBm or millwatts/microwatts Counter sensitivity to +7 dBm ±1.5 dB ±1.5 dB ±2.0 dB N/A 0.01 dB dBm or millwatts/microwatts Counter sensitivity to +7 dBm ±1.0 dB ±1.5 dB ±1.5 dB ±2.0 dB 0.01 dB dBm or millwatts/microwatts see graphs for typical data

Frequency: 10 MHz Output: 10 MHz sine wave, 1 Vrms into 50 Ω External timebase input: 1, 2, 5, 10 MHz; 1 to 5 Vrms into 50 Ω Connector: BNC female located on rear panel Internal timebase stability TCXO (Standard) Oven (Option 001) Aging rate Per day — < 5 x 10–10 Per month < 1 x 10–7 < 1.5 x 10–8 Short term < 1 x 10–9 < 2 x 10–10 (1 s avg. time) Line variation < 1 x 10–7 < 1 x 10–10 (±10%) Warm-up — < 1 x 10–8 within 5 min. after turn-on at 25 °C Temperature stability General information Save and recall: Up to 9 complete instrument setups may be saved and later recalled. These setups are retained when power is removed. Sample rate: User-selectable Fast (nominally 20 ms between readings), Medium (nominally 250 ms between readings), Slow (nominally 1 s between readings) and Hold. Self test: Internal memory and count circuitry automatically tested at startup, via menu selection, or remotely. Error messages displayed to indicate failed tests. Size: 213 mm W x 88.5 mm H x 300 mm D Operating temperature: 0-55 °C With battery option: 0-40 °C Weight: 4 kg without battery option, 6.4 kg with battery option Warranty: 1 year Programming: GPIB (IEEE-488.1-1987, IEEE 488.2-1987) or RS-232C Language: SCPI-1992.0 (Standard Commands for Programmable Instruments) RS-232C rates: User-selectable 2400 to 19200 baud Power supply: AC: 90-132 Vac; 47.5-66 Hz or 360- 440 Hz 216-264 Vac; 47.5-66 Hz line selection: automatic power requirements: 75 VA max. (25 W typ.) DC: (Option 002 only): 11-18 Vdc; 2A max. Battery (Option 002): Type: VHS camcorder, lead acid (2 each) Charge time: 8 hours in unit Capacity: 2.5 hours min. at 25 °C Math functions: Offset: Last reading and/or entered offset to reading for either power or frequency Averaging: 1 to 99 measurement running average Cable loss compensation: Offsets power reading via linear interpolation of user-entered attenuations with up to 9 independent frequency points. Display: Backlit LCD. Backlight can be turned on or off via front panel control. Sleep mode (Option 002 only): Automatically activated if no input is present for 5 minutes. Safety: Designed in compliance with IEC-1010, CAN/CSA 1010.1 EMC: Designed in compliance with Accessories supplied Operating, programming, and service manuals and AC power cord.

Ordering information

53151A 26.5 GHz counter 53152A 46 GHz counter Options Option 001 Oven Timebase Option 002 Battery and DC input Option 1CM (53150-67001) Rackmount kit 53153A (Option 007) Soft carrying case for 5315xA microwave counters Option A6J ANSI 7540 Compliant Calibration Option W30 Three years of Return Repair Service Option W50 Five years of Return Repair Service

www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/frequencycounters For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 0800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-07-01-14) 08 | Keysight | Full-featured CW Microwave Counters for Field, Factory or Lab - Technical Overview This information is subject to change without notice. © Keysight Technologies, 2011 - 2014 Published in USA, August 1, 2014 5966-3118E www.keysight.com