Keithley’s Series 2400 Source Measure Unit (SMU) Instruments are designed specifically for test applications that demand tightly coupled sourcing and measurement. All SourceMeter models provide precision voltage and current sourcing as well as measurement capabilities. Each SourceMeter SMU instrument is both a highly stable DC power source and a true instrument-grade 6½-digit multimeter. The power source characteristics include low noise, precision, and readback. The multimeter capabilities include high repeatability and low noise. The result is a compact, single-channel, DC parametric tester. In operation, these instruments can act as a voltage source, a current source, a voltage meter, a current meter, and an ohmmeter. Manufacturers of components and modules for the communications, semiconductor, computer, automotive, and medical industries will find the SourceMeter SMU instruments invaluable for a wide range of characterization and production test applications. Series 2400 SourceMeter SMU Instruments Datasheet Key Features

  • Five instruments in one (IV Source, IVR Measure)
  • Five models: 20–100 W DC, 1000 W pulsed,

1100 V to 1 µV

  • Source and sink (4-quadrant) operation
  • 0.012% basic measure accuracy with 6½-digit resolution
  • 2-, 4-, and 6-wire remote V-source and measure sensing
  • 1700 readings/second at 4½ digits via GPIB
  • Pass/Fail comparator for fast sorting/binning
  • Programmable DIO port for automation/handler/ prober control (except 2401)
  • Standard SCPI GPIB, RS-232 and Keithley Trigger Link interfaces
  • Download free KickStart I-V Characterization Software for device characterization. Advantages of a Tightly Integrated Instrument By linking source and measurement circuitry in a single unit, these instruments offer a variety of advantages over systems configured with separate source and measurement instruments. For example, they minimize the time required for test station development, setup, and maintenance, while lowering the overall cost of system ownership. They simplify the test process itself by eliminating many of the complex synchronization and connection issues associated with using multiple instruments. And, their compact half-rack size conserves precious “real estate” in the test rack or bench. A Tektronix Company

TEK.COM2 Series 2400 SourceMeter ® SMU Instruments Local Remote Remote Local DUT IN/OUT HI SENSE HI SENSE LO IN/OUT LO Imeter Isource Vmeter/Compliance Local Remote Remote Local DUT IN/OUT HI SENSE HI SENSE LO IN/OUT LO Imeter/Compliance Vsource Vmeter Feedback to Adjust Vsource Source I–Measure V, I, or Ω configuration Source V–Measure I, V, or Ω configuration +1 A –1 A +100 mA +20 V–20 V +200 V–200 V –100 mA 2400 only 2400 only +1 A –1 A +100 mA +20 V–20 V +200 V–200 V –100 mA +1A –1 A +200 mA +20 V–20 V +200 V–200 V –100 mA +20 mA –20 mA +1000 V–1000 V +1 A +20 V–20 V –1 A +100 mA –100 mA +3 A –60 V +60 V –3 A –100 V 5 A 3 A 100 mA –10 V +10 V –100 mA –3 A –5 A –40 V +40 V 1 A –1 A Duty cycle limited Duty cycle limited Duty cycle limitedDuty cycle limited Duty cycle limited 2400 SourceMeter SMU Instrument 2410 High-Voltage SourceMeter SMU Instrument 2440 5A SourceMeter SMU Instrument 2420 3A SourceMeter SMU Instrument 2401 Low-Voltage SourceMeter SMU Instrument Series 2400 SourceMeter SMU Instruments Typical Applications Devices

  • Discrete semiconductor devices
  • Passive devices
  • Transient suppression devices
  • ICs, RFICs, MMICs
  • Laser diodes, laser diode modules, LEDs, photodetectors
  • Circuit protection devices: TVS, MOV, Fuses, etc.
  • Airbags
  • Connectors, switches, relays
  • High brightness LEDs (DC and pulse) Tests
  • Leakage
  • Low voltage/resistances
  • LIV
  • IDDQ
  • I-V characterization
  • Isolation and trace resistance
  • Temperature coefficient
  • Forward voltage, reverse breakdown, leakage current
  • DC parametric test
  • DC power source
  • HIPOT
  • Photovoltaic cell efficiency (source and sink)
  • Dielectric withstanding

TEK.COM 3 Series 2400 SourceMeter ® SMU InstrumentsDatasheet Power of Five Instruments in One (IV Source, IVR Measure) The tightly coupled nature of a SourceMeter SMU instrument provides many advantages over solutions configured from separate instruments, such as a precision power supply and a digital multimeter. For example, it provides faster test times by reducing GPIB traffic and simplifies the remote programming interface. It also protects the device under test from damage due to accidental overloads, thermal runaway, etc. Both the current and voltage source are programmable with readback to help maximize device measurement integrity. If the readback reaches a programmed compliance limit, then the source is clamped at the limit, providing fault protection. I-V Characteristics All SourceMeter SMU instruments provide four-quadrant operation. In the first and third quadrants they operate as a source, delivering power to a load. In the second and fourth quadrants they operate as a sink, dissipating power internally. Voltage, current, and resistance can be measured during source or sink operation. Automation for Speed A SourceMeter SMU instrument streamlines production testing. It sources voltage or current while making measurements without needing to change connections. It is designed for reliable operation in non-stop production environments. To provide the throughput demanded by production applications, the SourceMeter SMU instrument offers many built-in features that allow it to run complex test sequences without computer control or GPIB communications slowing things down. Standard and Custom Sweeps Sweep solutions greatly accelerate testing with automation hooks. Three basic sweep waveforms are provided that can be programmed for single-event or continuous operation. They are ideal for I/V, I/R, V/I, and V/R characterization.

  • Linear Staircase Sweep: Moves from the start level to the stop level in equal linear steps
  • Logarithmic Staircase Sweep: Done on a log scale with a specified number of steps per decade
  • Custom Sweep: Allows construction of special sweeps by specifying the number of measurement points and the source level at each point
  • Up to 1700 readings/second at 4½ digits to the GPIB bus
  • 5000 readings can be stored in the non-volatile buffer memory Bias Linear staircase sweep Logarithmic staircase sweep Custom sweep Bias Bias Bias Stop Stop StopStart Start User defined settings Start Bias Bias

TEK.COM4 Series 2400 SourceMeter ® SMU Instruments Built-In Test Sequencer (Source Memory List) The Source Memory list provides faster and easier testing by allowing you to setup and execute up to 100 different tests that run without PC intervention.

  • Stores up to 100 instrument configurations, each containing source settings, measurement settings, pass/fail criteria, etc.
  • Pass/fail limit test as fast as 500 µs per point
  • Onboard comparator eliminates the delay caused when sending data to the computer for analysis
  • Built-in, user definable math functions to calculate derived parameters Test 3 Test 2 V I Test 1IR VF2 VF1 Test Pass/Fail Test If Passes Test If Fails Test Test Check VF1 at 100 mA against pass/ fail limits Go to Test 2 1. Bin part to bad bin 2. Transmit data to computer while handler is placing new part 3. Return to Test 1 Test Check VF2 at

1 A against

–500 V and test against pass/fail limits 1. Bin part to good bin 2. Transmit readings to computer while handler is placing new part 3. Return to Test 1 Example test sequence Digital I/O Interface The digital I/O interface can link a SourceMeter SMU instrument to many popular component handlers, including Aetrium, Aeco, and Robotronics. Other capabilities of the interface include:

  • Tight systems integration for applications such as binning and sorting
  • Built-in component handler interface
  • Start of test and end of test signals
  • 5 V, 300 mA power supply
  • Optional expander accessory (2499-DIGIO) adds 16 digital I/O lines The digital I/O interface is available on all Series 2400 SoourceMeter instruments except the 2401. Trigger Link Interface All SourceMeter SMU instruments include Keithley’s unique Trigger Link interface, which provides high-speed, seamless communications with many of Keithley’s other instruments independent of a computer and GPIB. Unique 6-Wire Ohms Technique SourceMeter SMU instruments can make standard 4-wire, split Kelvin, and 6-wire, guarded ohms measurements and can be configured for either the constant current or constant voltage method. The 6-wire ohms technique:
  • Uses guard and guard sense leads in addition to the 4-wire sense and source leads.
  • Locks out parallel current paths when measuring resistor networks or hybrid circuits to isolate the component under test.
  • Allows users to configure and plot data easily from Series 2400 SourceMeter SMU instruments, making characterization of two, three, and four terminal devices a snap. GUARD GUARD SENSE IN/OUT HI SENSE HI SENSE LO IN/OUT LO DUT Vmeter Imeter Isource 6-Wire Ohms Circuit. All test current flows through R1 because the high current guard drives the voltage across R2 to 0 V.

TEK.COM 5 Series 2400 SourceMeter ® SMU InstrumentsDatasheet Specifications Voltage Accuracy (Local or Remote Sense) Model Range Programming Resolution Source 1 Accuracy (1 Year) 23°C ±5°C ±(% rdg. + volts) Default Measurement Resolution Measurement 2, 3, 4 Accuracy (1 Year) 23°C ±5°C ±(% rdg. + volts) Output Slew Rate (±30%) Source/Sink Limit 2400, 2401 200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV ±21 V @ ±1.05 A ±210 V @ ±105 mA* 2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV 200.000 V* 5 mV 0.02% + 24 mV 1 mV 0.015% + 10 mV 0.5 V/µs 2410 200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV ±21 V @ ±1.05 A ±1100 V @ ±21 mA 2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV 1000.00 V 50 mV 0.02% + 100 mV 10 mV 0.015% + 50 mV 0.5 V/µs 2420 200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV ±21 V @ ±3.15 A ±63 V @ ±1.05 A 2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV 2440 200.000 mV 5 µV 0.02% + 600 µV 1 µV 0.012% + 300 µV ±10.5 V @ ±5.25 A ±42 V @ ±1.05 A 2.00000 V 50 µV 0.02% + 600 µV 10 µV 0.012% + 300 µV *Not available on 2401. NOTES 1. 2400, 2401, 2410 Only: Specifications valid for continuous output currents below 105 mA. For operation above 105 mA continuous for >1 minute, derate accuracy 10%/35 mA above 105 mA. specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%. 3. Accuracies apply to 2- or 4-wire mode when properly zeroed. 4. In pulse mode, limited to 0.1 PLC measurement. Temperature Coefficient (0°–18°C and 28°–50°C) ±(0.15 × accuracy specification)/°C. Voltage Regulation Line: 0.01% of range. Load: 0.01% of range + 100 µV. Over Voltage Protection User selectable values, 5% tolerance. Factory default = none. Current Limit Bipolar current limit (compliance) set with single value. Min. 0.1% of range. Overshoot <0.1% typical (full scale step, resistive load, 10 mA range). Additional Source Specifications (All Models) Transient Response Time 30 µs minimum for the output to recover to its spec. following a step change in load. Command Processing Time Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent <nrf> command. Autorange On: 10 ms. Autorange Off: 7 ms. Output Settling Time Time required to reach 0.1% of final value after command is processed. 100 µs typical. Resistive load. 10 µA to 100 mA range. DC Floating Voltage Output can be floated up to ±250 VDC (2440 ±40 VDC) from chassis ground. Remote Sense Up to 1 V drop per load lead. Compliance Accuracy Add 0.3% of range and ±0.02% of reading to base specification. Over Temperature Protection Internally sensed temperature overload puts unit in standby mode. Range Change Overshoot Overshoot into a fully resistive 100 kΩ load, 10 Hz to 1 MHz BW, adjacent ranges: 100 mV typical, except 20 V/200 V (20 V/60 V on 2420), range boundary, and 2440. Minimum Compliance Value 0.1% of range.

TEK.COM6 Series 2400 SourceMeter ® SMU Instruments Current Accuracy (Local or Remote Sense) Model Range Programming Resolution Source 1, 3 Accuracy (1 Year) 23°C ±5°C ±(% rdg. + amps) Default Measurement Resolution Measurement 5, 6, 7 Accuracy (1 Year) 23°C ±5°C ±(% rdg. + amps) Source/Sink Limit 2400, 2401 1.00000 µA 50 pA 0.035% + 600 pA 10 pA 0.029% + 300 pA ±1.05A @ ±21 V ±105 mA @ ±210 V8 10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA 100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA 1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA 10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA 100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA 1.00000 A2 50 µA 0.27 % + 900 µA 10 µA 0.22 % + 570 µA 2410 1.00000 µA 50 pA 0.035% + 600 pA 10 pA 0.029% + 300 pA ±1.05A @ ±21 V ±21 mA @ ±1100 V 10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA 100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA 1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA 20.0000 mA 500 nA 0.045% + 4 µA 100 nA 0.035% + 1.2 µA 100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA 1.00000 A2 50 µA 0.27 % + 900 µA 10 µA 0.22 % + 570 µA 2420 10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA ±3.15A @ ±21 V ±1.05 A @ ±63 V 100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA 1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA 10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA 100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA 1.00000 A2 50 µA 0.067% + 900 µA 10 µA 0.066% + 570 µA 2440 10.0000 µA 500 pA 0.033% + 2 nA 100 pA 0.027% + 700 pA ±5.25A @ ±10.5 V ±1.05 A @ ±42 V 100.000 µA 5 nA 0.031% + 20 nA 1 nA 0.025% + 6 nA 1.00000 mA 50 nA 0.034% + 200 nA 10 nA 0.027% + 60 nA 10.0000 mA 500 nA 0.045% + 2 µA 100 nA 0.035% + 600 nA 100.000 mA 5 µA 0.066% + 20 µA 1 µA 0.055% + 6 µA 1.00000 A 50 µA 0.067% + 900 µA 10 µA 0.060% + 570 µA NOTES 1. 2400, 2401, 2410 Only: Specifications valid for continuous output currents below 105 mA. For operation above 105 mA continuous for >1 minute, derate accuracy 10%/35 mA above 105 mA. 2. Full operation (1 A) regardless of load to 30°C (50°C for 2420 and 2440). Above 30°C (50°C for 2420 and 2440) ambient, derate 35 mA/°C and prorate 35 mA/Ω load. 4-wire mode. For current sink operation on 1 A, 3 A, or 5 A ranges, maximum continuous power is limited to approximately 1/2 rated power or less, depending on current, up to 30°C ambient. See power equations in the User’s Manual to calculate allowable duty cycle for specific conditions. For 1A range, accuracy is: 2400, 2401: ±(1.5% + offset*8). 2410, 2420, 2440: ±(1.5% + offset*3). specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%. 6. Accuracies apply to 2- or 4-wire mode when properly zeroed. 7. In pulse mode, limited to 0.1 PLC measurement. 8. 2400 only. Temperature Coefficient (0°–18°C and 28°–50°C) ±(0.15 × accuracy specification)/°C. Voltage Limit Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range. Overshoot <0.1% typical (1 mA step, RL = 10 kΩ, 20 V range for 2400, 2401, 2410, 2420), (10 V range for 2440).

TEK.COM 7 Series 2400 SourceMeter ® SMU InstrumentsDatasheet Resistance Measurement Accuracy (Local or Remote Sense) 1, 2, 5 Range Default Resolution Default Test Current 2400, 2401, 2410 Default Test Current 2420, 2440 Normal Accuracy (23°C ±5°C) 1 Year, ±(% rdg. + ohms) Enhanced Accuracy (23°C ±5°C)4

1 Year,

±(% rdg. + ohms) 2400, 2401 2410 2420, 2440 2400, 2401 <0.20000 Ω 3 – – – Source IACC + Meas. VACC Source IACC + Meas. VACC Source IACC + Meas. VACC Source IACC + Meas. VACC 2.00000 Ω3 10 µΩ – 1 A Source IACC + Meas. VACC Source IACC + Meas. VACC 0.17% + 0.0003 Ω Source IACC + Meas. VACC 200.000 MΩ3 1 kΩ 100 nA – 0.66% + 10 kΩ 0.66% + 24 kΩ Source IACC + Meas. VACC 0.35% + 5 kΩ >200.000 MΩ3 – – – Source IACC + Meas. VACC Source IACC + Meas. VACC Source IACC + Meas. VACC Source IACC + Meas. VACC Temperature Coefficient (0°–18°C and 28°–50°C) ±(0.15 × accuracy specification)/°C. Source I Mode, Manual Ohms Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense). Source V Mode, Manual Ohms Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense). 6-wire Ohms Mode Available using active ohms guard and guard sense. Max. Guard Output Current: 50 mA (except 1 A range). Accuracy is load dependent. Refer to White Paper below for calculation formula: “Obtaining More Accurate Resistance Measurements Using the 6-wire Ohms Measurement Te chnique.” Guard Output Impedance <0.1Ω in ohms mode. NOTES specifications, except 200 mV, 1 A, 10 A ranges, add 0.5%. 2. Accuracies apply to 2- or 4-wire mode when properly zeroed. 3. Manual ohms only – except 2420, 2440 for 2 Ω range and 2400, 2401, or 2410 for 200 MΩ range. 4. Source readback enabled, offset compensation ON. Also available on 2410, 2420, and 2440 with similar accuracy enhancement. 5. In pulse mode, limited to 0.1 PLC measurement. 6. Except 2440; default test current is 5 µA. 7. Except 2440; default test current is 0.5 µA.

TEK.COM8 Series 2400 SourceMeter ® SMU Instruments System Speeds Measurement 1 Maximum Range Change Rate 75/second. Maximum Measure Autorange Time 40 ms (fixed source).2 Sweep Operation 3 Reading Rates (rdg./second) for 60 Hz (50 Hz) Speed NPLC/ Trigger Origin Measure Source-Measure Source-Measure 5 Pass/Fail Test4, 5 Source-Memory 4 To Mem. To GPIB To Mem. To GPIB To Mem. To GPIB To Mem. To GPIB Fast 0.01 / internal 2081 (2030) 1754 1551 (1515) 1369 902 (900) 981 165 (162) 165 IEEE-488.1 Mode 0.01 / external 1239 (1200) 1254 1018 (990) 1035 830 (830) 886 163 (160) 163 Fast 0.01 / internal 2081 (2030) 1198 (1210) 1551 (1515) 1000 (900) 902 (900) 809 (840) 165 (162) 164 (162) IEEE-488.2 Mode 0.01 / external 1239 (1200) 1079 (1050) 1018 (990) 916 (835) 830 (830) 756 (780) 163 (160) 162 (160) Medium 0.10 / internal 510 (433) 509 (433) 470 (405) 470 (410) 389 (343) 388 (343) 133 (126) 132 (126) IEEE-488.2 Mode 0.10 / external 438 (380) 438 (380) 409 (360) 409 (365) 374 (333) 374 (333) 131 (125) 131 (125) Normal 1.00 / internal 59 (49) 59 (49) 58 (48) 58 (48) 56 (47) 56 (47) 44 (38) 44 (38) IEEE-488.2 Mode 1.00 / external 57 (48) 57 (48) 57 (48) 57 (47) 56 (47) 56 (47) 44 (38) 44 (38) Single Reading Operation Reading Rates (rdg./second) for 60 Hz (50 Hz) Speed NPLC/Trigger Origin Measure To GPIB Source-Measure 5 To GPIB Source-Measure Pass/Fail Test4, 5 To GPIB Fast (488.1) 0.01 / internal 537 140 135 Fast (488.2) 0.01 / internal 256 (256) 79 (83) 79 (83) Medium (488.2) 0.10 / internal 167 (166) 72 (70) 69 (70) Normal (488.2) 1.00 / internal 49 (42) 34 (31) 35 (30) Component for 60 Hz (50 Hz) 4, 6 Speed NPLC/Trigger Origin Measure To GPIB Source Pass/Fail Test Source-Measure Pass/Fail Test 5, 7 To GPIB NOTES 1. Reading rates applicable for voltage or current measurements. Auto zero off, autorange off, filter off, display off, trigger delay = 0, and binary reading format. 2. Purely resistive lead. 1 µA and 10 µA ranges <65 ms. 3. 1000 point sweep was characterized with the source on a fixed range. 4. Pass/Fail test performed using one high limit and one low math limit. 5. Includes time to re-program source to a new level before making measurement. 6. Time from falling edge of START OF TEST signal to falling edge of end of test signal. 7. Command processing time of :SOURce:VOLTage|CURRent:TRIGgered <nrf> command not included.

TEK.COM 9 Series 2400 SourceMeter ® SMU InstrumentsDatasheet General Specifications Noise Rejection NPLC NMRR CMRR Fast 0.01 — 80 dB Medium 0.1 — 80 dB Slow 1 60 dB 100 dB1 1 Except lowest 2 current ranges = 90dB. Load Impedance Stable into 20,000 pF typical. Common Mode Voltage 250 V DC (40V DC for 2440). Common Mode Isolation >109 Ω, <1000pF. Overrange 105% of range, source and measure. Max. Voltage Drop Between Input/Output and Sense Terminals 5 V. Max. Sense Lead Resistance 1 MΩ for rated accuracy. Sense Input Impedance >1010 Ω. Guard Offset Voltage <150 µV, typical (300 µV for 2440). Source Output Modes Fixed DC level Memory List (mixed function) Stair (linear and log) Memory Buffer 5,000 readings @ 5 digits (two 2,500 point buffers). Includes selected measured value(s) and time stamp. Lithium battery backup (3 yr+ battery life). Source Memory List 100 points max. Programmability IEEE-488 (SCPI-1995.0), RS-232, 5 user-definable power-up states plus factory default and *RST. Digital Interface Output Enable Active low input. Handler Interface Start of test, end of test, 3 category bits. +5 V @ 300 mA supply. Not available on 2401. Digital I/O 1 trigger input, 4 TTL/Relay Drive outputs (33 V @ 500 mA, diode clamped). Not available on 2401. Power Supply 100 V to 240 V rms, 50–60 Hz (automatically detected at power up). 2400, 2401: 190 VA. 2410: 210 VA. 2420: 220 VA. 2440: 240VA. Cooling 2401: Convection. 2410, 2420, 2440: Forced air, variable speed. EMC Conforms to European Union Directive 89/336/EEC, EN 61326-1. Safety Conforms to European Union Low Voltage Directive. 2400 and 2401 UL listed to UL 61010-1:2012 and UL 61010-2-30:2012. Vibration MIL -PRF-28800F Class 3 Random. Warm-Up 1 hour to rated accuracies. Dimensions 89 mm high × 213 mm wide × 370 mm deep (3½ in × 83⁄8 in × 149⁄16 in). Bench Configuration (with handle and feet): 104 mm high × 238 mm wide × 370 mm deep (41⁄8 in × 93⁄8 in × 149⁄16 in). Supplied Accessories

8605 Test Leads

LabVIEW Driver Software Driver (downloadable) LabTracer Software (downloadable)

TEK.COM10 Series 2400 SourceMeter ® SMU Instruments Available Accessories Test Leads and Probes 1754 2-Wire Universal 10-Piece Test Lead Kit

5804 Kelvin (4-Wire) Universal 10-Piece Test Lead Kit

5805 Kelvin (4-Wire) Spring-Loaded Probes

5808 Low Cost Single-pin Kelvin Probe Set

5809 Low Cost Kelvin Clip Lead Set

8607 2-Wire, 1000 V Banana Cables, 1 m (3.3 ft) CA-18-1 Shielded Dual Banana Cable, 1.2 m (4 ft) Cables/Adapters 7007-1 Shielded GPIB Cable, 1 m (3.3 ft) 7007-2 Shielded GPIB Cable, 2 m (6.6 ft) 7009-5 RS-232 Cable

8620 Shorting Plug

KPCI-488LPA IEEE-488 Interface/Controller for the PCI Bus KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter Triggering and Control 2499-DIGIO Digital I/O Expander Assembly (not for 2401) 8501-1 Trigger Link Cable, DIN-to-DIN, 1m (3.3 ft) 8501-2 Trigger Link Cable, DIN-to-DIN, 2m (6.6 ft)

8502 Trigger Link to BNC Breakout Box

8503 Trigger Link Cable, DIN-to-Dual BNC, 1m (3.3 ft)

8505 Male to 2-Female Y-DIN Cable for Trigger Link

4288-1 Single Fixed Rack Mount Kit 4288-2 Dual Fixed Rack Mount Kit 4288-4 Dual Fixed Rack Mount Kit 4288-5 Shelf Type Side by Side Rack Mounting Kit 4288-9 Dual Fixed Rack Mounting Kit

TEK.COM 11 Series 2400 SourceMeter ® SMU InstrumentsDatasheet Available Services 2400-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 2401-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 2410-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 2420-3Y-EW 1-year factory warranty extended to 3 years from date of shipment 2440-3Y-EW 1-year factory warranty extended to 3 years from date of shipment C/2400-3Y-ISO 3 (ISO-17025 accredited) calibrations within 3 years of purchase for 2400, 2400-LV* C/2401-3Y-ISO 3 (ISO-17025 accredited) calibrations within 3 years of purchase for 2401* C/2410-3Y-ISO 3 (ISO-17025 accredited) calibrations within 3 years of purchase for 2410* C/2420-3Y-ISO 3 (ISO-17025 accredited) calibrations within 3 years of purchase for 2420* C/2440-3Y-ISO 3 (ISO-17025 accredited) calibrations within 3 years of purchase for 2440* TRN-2400-1-C Course: Unleashing the Power of Your SourceMeter SMU Instrument *Not available in all countries

Ordering Information

2400 200V, 1A, 20W SourceMeter SMU Instrument 2401 20V, 1A, 20W SourceMeter SMU Instrument 2410 1100V, 1A, 20W SourceMeter SMU Instrument 2420 60V, 3A, 60W SourceMeter SMU Instrument 2440 40V, 5A, 50W SourceMeter SMU Instrument Warranty Information Warranty Summary This section summarizes the warranties of the Series 2400. For complete warranty information, refer to the Series 2400 Reference Manual. Any portion of the product that is not manufactured by Keithley is not covered by this warranty and Keithley will have no duty to enforce any other manufacturer’s warranties. Hardware Warranty Keithley Instruments, Inc. warrants the Keithley manufactured portion of the hardware for a period of one year from defects in materials or workmanship; provided that such defect has not been caused by use of the Keithley hardware which is not in accordance with the hardware instructions. The warranty does not apply upon any modification of Keithley hardware made by the customer or operation of the hardware outside the environmental specifications. Software Warranty Keithley warrants that the Keithley produced portion of the software or firmware will conform in all material respects with the published specifications for a period of ninety (90) days; provided the software is used on the product for which it is intended in accordance with the software instructions. Keithley does not warrant that operation of the software will be uninterrupted or error-free, or that the software will be adequate for the customer’s intended application. The warranty does not apply upon any modification of the software made by the customer.

Contact Information: Australia* 1 800 709 465 Austria 00800 2255 4835 Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777 Belgium* 00800 2255 4835 Brazil +55 (11) 3759 7627 Canada 1 800 833 9200 Central East Europe / Baltics +41 52 675 3777 Central Europe / Greece +41 52 675 3777 Denmark +45 80 88 1401 Finland +41 52 675 3777 France* 00800 2255 4835 Germany* 00800 2255 4835 Hong Kong 400 820 5835 India 000 800 650 1835 Indonesia 007 803 601 5249 Italy 00800 2255 4835 Japan 81 (3) 6714 3086 Luxembourg +41 52 675 3777 Malaysia 1 800 22 55835 Mexico, Central/South America and Caribbean 52 (55) 56 04 50 90 Middle East, Asia, and North Africa +41 52 675 3777 The Netherlands* 00800 2255 4835 New Zealand 0800 800 238 Norway 800 16098 People’s Republic of China 400 820 5835 Philippines 1 800 1601 0077 Poland +41 52 675 3777 Portugal 80 08 12370 Republic of Korea +82 2 565 1455 Russia / CIS +7 (495) 6647564 Singapore 800 6011 473 South Africa +41 52 675 3777 Spain* 00800 2255 4835 Sweden* 00800 2255 4835 Switzerland* 00800 2255 4835 Taiwan 886 (2) 2656 6688 Thailand 1 800 011 931 United Kingdom / Ireland* 00800 2255 4835 USA 1 800 833 9200 Vietnam 12060128 * European toll-free number. If not accessible, call: +41 52 675 3777 Find more valuable resources at TEK.COM in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks or registered trademarks of their respective companies.

011719 SBG 1KW-2798-1