Technical content
Product Discontinuance Notice The 34401A DMM and all associated options will be discontinued December 1, 2016. The last date this product can be ordered is November 30, 2016. – For new product purchases, Keysight recommends the 34461A, 6½ digit True volt DMM. – For more information, as well as to access to detailed migration guides, please visit www.keysight.com/find/nextgenDMMs – To contact a product selection expert, visit www.keysight.com/find/contactus
02 | Keysight | 34401A Digital Multimeter - Data Sheet Measurement capability – 6½ digit resolution – 10 measurement functions: DC/AC voltage, DC/AC current, 2- and 4-wire resistance, diode, continuity, frequency, period – Basic accuracy: 0.0035% DC, 0.06% AC – 1000 V max voltage input, 3 A max current input System Capability – 1000 readings/second – 512 reading memory www.keysight.com/find/dmm Superior performance The Keysight Technologies, Inc. 34401A multimeter gives you the performance you need for fast, accurate bench and system testing. The 34401A provides a combination of resolution, accuracy and speed that rivals DMMs costing many times more. 6½ digits of resolution, 0.0015% basic 24-hr dcV accuracy and 1,000 readings/s direct to GPIB assure you of results that are accurate, fast, and repeatable. Use it on your benchtop The 34401A was designed with your bench needs in mind. Functions commonly associated with bench operation, like continuity and diode test, are built in. A Null feature allows you to remove lead resistance and other fixed offsets in your measurements. Other capabilities like min/max/avg readouts and direct dB and dBm measurements make checkout with the 34401A faster and easier. The 34401A gives you the ability to store up to 512 readings in internal memory. For troubleshooting, a reading hold feature lets you concentrate on placing your test leads without having to constantly glance at the display. Use it for systems testing For systems use, the 34401A gives you faster bus throughput than any other DMM in its class. The 34401A can send up to 1,000 readings/s directly across GPIB in user-friendly ASCII format. You also get both GPIB and RS-232 interfaces as standard features. Voltmeter Complete and External Trigger signals are provided so you can synchronize to other instruments in your test system. In addition, a TTL output indicates Pass/Fail results when limit testing is used. To ensure both forward and backward compatibility, the 34401A includes three command languages (SCPI, Keysight 3478A and Fluke8840A/42A), so you don’t have to rewrite your existing test software. An optional rack mount kit is available. Easy to use Commonly accessed attributes, such as functions, ranges, and resolution are selected with a single button press. Advanced features are available using menu functions that let you optimize the 34401A for your applications. The included Keysight IntuiLink software allows you to put your captured data to work easily, using PC applications such as Microsoft Excel or Word to analyze, interpret, display, print, and document the data you get from the 34401A. You can specify the meter setup and take a single reading or log data to the Excel spreadsheet in specified time intervals. Programmers can use ActiveX components to control the DMM using SCPI commands. To find out more about IntuiLink, visit www.keysight.com/find/intuilink 3-year warranty With your 34401A, you get full documentation, a high-quality test lead set, calibration certificate with test data, and a 3-year warranty, all for one low price.
03 | Keysight | 34401A Digital Multimeter - Data Sheet Accuracy Specifications ± (% of reading + % of range)1 Function Range 3 Frequency, etc. 24 hour 2 23 ± 1 °C 90 day 23 ± 5 °C 1 year 23 ± 5 °C Temperature coefficient 0 – 18 °C 28 – 55 °C DC voltage 100.0000 mV 1.000000 V 10.00000 V 100.0000 V 1000.000 V 0.0030 + 0.0030 0.0020 + 0.0006 0.0015 + 0.0004 0.0020 + 0.0006 0.0020 + 0.0006 0.0040 + 0.0035 0.0030 + 0.0007 0.0020 + 0.0005 0.0035 + 0.0006 0.0035 + 0.0010 0.0050 + 0.0035 0.0040 + 0.0007 0.0035 + 0.0005 0.0045 + 0.0006 0.0045 + 0.0010 0.0005 + 0.0005 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 0.0005 + 0.0001 True rms AC voltage 4 100.0000 mV 3 – 5 Hz 5 – 10 Hz
10 Hz – 20 kHz
20 – 50 kHz 50 – 100 kHz 100 – 300 kHz 6 1.00 + 0.03 0.35 + 0.03 0.04 + 0.03 0.10 + 0.05 0.55 + 0.08 4.00 + 0.50 1.00 + 0.04 0.35 + 0.04 0.05 + 0.04 0.11 + 0.05 0.60 + 0.08 4.00 + 0.50 1.00 + 0.04 0.35 + 0.04 0.06 + 0.04 0.12 + 0.05 0.60 + 0.08 4.00 + 0.50 0.100 + 0.004 0.035 + 0.004 0.005 + 0.004 0.011 + 0.005 0.060 + 0.008 0.20 + 0.02
1.000000 V to
750.000 V 3 – 5 Hz 5 – 10 Hz 20 – 50 kHz 50 – 100 kHz 5 100 – 300 kHz 6 1.00 + 0.02 0.35 + 0.02 0.04 + 0.02 0.10 + 0.04 0.55 + 0.08 4.00 + 0.50 1.00 + 0.03 0.35 + 0.03 0.05 + 0.03 0.11 + 0.05 0.60 + 0.08 4.00 + 0.50 1.00 + 0.03 0.35 + 0.03 0.06 + 0.03 0.12 + 0.04 0.60 + 0.08 4.00 + 0.50 0.100 + 0.003 0.035 + 0.003 0.005 + 0.003 0.011 + 0.005 0.060 + 0.008 0.20 + 0.02 Resistance 7 100.0000 Ω 1.000000 kΩ 10.00000 kΩ 100.0000 kΩ 1.000000 MΩ 10.00000 MΩ 100.0000 MΩ 1 mA current source 1 mA 100 µA 10 µA 5.0 µA 500 nA 500 nA II 10 MΩ 0.0030 + 0.0030 0.0020 + 0.0005 0.0020 + 0.0005 0.0020 + 0.0005 0.002 + 0.001 0.015 + 0.001 0.300 + 0.010 0.008 + 0.004 0.008 + 0.001 0.008 + 0.001 0.008 + 0.001 0.008 + 0.001 0.020 + 0.001 0.800 + 0.001 0.010 + 0.004 0.010 + 0.001 0.010 + 0.001 0.010 + 0.001 0.010 + 0.001 0.040 + 0.001 0.800 + 0.001 0.0006 + 0.0005 0.0006 + 0.0001 0.0006 + 0.0001 0.0006 + 0.0001 0.0010 + 0.0002 0.0030 + 0.0004 0.1500 + 0.0002 DC current 10.00000 mA 100.0000 mA 1.000000 A 3.00000 A < 0.1 V burden voltage < 0.6 V < 1.0 V < 2.0 V 0.005 + 0.010 0.010 + 0.010 0.050 + 0.010 0.100 + 0.010 0.030 + 0.020 0.030 + 0.005 0.080 + 0.010 0.120 + 0.020 0.050 + 0.020 0.050 + 0.005 0.100 + 0.010 0.120 + 0.020 0.0020 + 0.0020 0.0020 + 0.0005 0.0050 + 0.0010 0.005 + 0.0020 True rms AC current 4
1.000000 A 3 – 5 Hz
5 – 10 Hz
10 Hz – 5 kHz
1.00 + 0.04 0.30 + 0.04 0.10 + 0.04 1.00 + 0.04 0.30 + 0.04 0.10 + 0.04 1.00 + 0.04 0.30 + 0.04 0.10 + 0.04 0.100 + 0.006 0.035 + 0.006 0.015 + 0.006
3.00000 A 3 – 5 Hz
5 – 10 Hz 1.10 + 0.06 0.35 + 0.06 0.15 + 0.06 1.10 + 0.06 0.35 + 0.06 0.15 + 0.06 1.10 + 0.06 0.35 + 0.06 0.15 + 0.06 0.100 + 0.006 0.035 + 0.006 0.015 + 0.006 Frequency or period 8 100 mV to 750 V 3 – 5 Hz 5 – 10 Hz 10 – 40 Hz
40 Hz – 300 kHz
0.10 0.05 0.03 0.006 0.10 0.05 0.03 0.01 0.10 0.05 0.03 0.01 0.005 0.005 0.001 0.001 1. Specifications are for 1 hr warm-up and 6½ digits, slow ac filter. 2. Relative to calibration standards. 3. 20% over range on all ranges except 1000 Vdc and 750 Vac ranges. 4. For sinewave input > 5% of range. For inputs from 1% to 5% of range and < 50 kHz, add 0.1% of range additional error. 5. 750 V range limited to 100 kHz or 8 x 107 Volt-Hz. 6. Typically 30% of reading error at 1 MHz. 7. Specifications are for 4-wire ohms function or 2-wire ohms using Math Null. Without Math Null, add 0.2 Ω additional error in 2-wire ohms function. 8. Input > 100 mV. For 10 to 100 mV inputs multiply % of reading error x 10. 9. Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is typical. Variation in the current source will create some variation in the voltage drop across a diode junction. 254.4 mm 374.0 mm 212.6 mm 348.3 mm 103.6mm 88.5mm 254.4 mm 374.0 mm 212.6 mm 348.3 mm 103.6mm 88.5mm
04 | Keysight | 34401A Digital Multimeter - Data Sheet Measurement Characteristics DC voltage Measurement method Continuously integrating multi-slope III A-D converter A-D linearity 0.0002% of reading + 0.0001% of range Input resistance 10 MΩ or 0.1V, 1 V, 10 V ranges Selectable > 10,000 MΩ
100 V, 1000 V ranges 10 MΩ ± 1%
Input bias current < 30 pA at 25 °C Input protection 1000 V all ranges dcV:dcV ratio accuracy Vinput accuracy + Vrelevance accuracy True RMS AC voltage Measurement method AC-coupled true rms-measures the AC component of the input with up to 400 Vdc of bias on any range Crest factor Maximum of 5:1 at full scale Additional crest factor errors (non-sinewave) Crest factor 1-2: 0.05% of reading Crest factor 2-3: 0.15% of reading Crest factor 3-4: 0.30% of reading Crest factor 4-5: 0.40% of reading Input impedance 1 MΩ ± 2% in parallel with 100 pF Input protection 750 Vrms all ranges Resistance Measurement method Selectable 4-wire or 2-wire ohms Current source referenced to LO input Maximum lead resistance (4-wire) 10% of range per lead for 100 Ω, 1 kΩ ranges 1 kΩ per lead on all other ranges Input protection 1000 V all ranges DC current Shunt resistance 5 Ω for 10 mA, 100 mA 0.1 Ω for 1 A, 3 A Input protection Externally accessible 3 A 250 V fuse Internal 7 A 250 V fuse True RMS AC current Measurement method Directly coupled to the fuse and shunt. AC coupled true RMS measurement (measures the AC component only). Shunt resistance 0.1 Ω for 1 A and 3 A ranges Input protection Externally accessible 3 A 250 V fuse Internal 7 A 250 V fuse Frequency and period Measurement method Reciprocal counting technique Voltage ranges Same as AC voltage function Gate time 1 s, 100 ms, or 10 ms Continuity/diode Response time 300 samples/s with audible tone Continuity threshold Selectable from 1 Ω to 1000 Ω Measurement noise rejection 60 (50) Hz 1 dc CMRR 140 dB ac CMRR 70 dB Integration time and normal mode rejection 2 100 plc/1.67 s (2 s) 60 dB 3 0 plc/167 ms (200 ms) 60 dB 3 1 plc/16.7 ms (20 ms) 60 dB < 1 plc/3 ms or 800 µs) 0 dB Operating characteristics 4 Function Digits Reading/s dcV, dcl, and resistance 6½ 0.6 (0.5) 6½ 6 (5) 5½ 60 (50) 5½ 300 4½ 1000 acV, acl 6½ 0.15 slow (3 Hz) 6½ 1 medium (20 Hz) 6½ 10 fast (200 Hz) 5 6½ 50 Frequency or period 6½ 1 5½ 9.8 4½ 80 1. For 1 kΩ unbalanced in LO lead, ± 500 V peak maximum. 2. For power line frequency ± 0.1%. 3. For power line frequency ± 1% use 40 dB or ± 3% use 30 dB. 4. Reading speeds for 60 Hz and (50 Hz) operation. 5. Maximum useful limit with default settling delays defeated. 6. Speeds are for 4½ digits, delay 0, auto-zero and display OFF.
05 | Keysight | 34401A Digital Multimeter - Data Sheet
Ordering Information
Keysight 34401A multimeter accessories included: Test lead kit with probe, alligator, and grabber attachments, calibration certificate, test report, and power cord. Also includes CD with: IntuiLink software, IVI and VXI PnP drivers, Quick start tutorial, user’s guide, command quick reference, service guide, and data sheet. Options 34401A-A6J ANSI Z540 compliant calibration Accessories Probes/Leads/Clip accessories 11059A Kelvin probe set 11060A Surface mount device (SMD) test probes 11062A Kelvin clip set 34133A Precision electronic test leads 34134A DC coupled current probe 34136A High voltage probe 34138A Test lead set 34171B Input terminal connector (sold in pairs) 34172B Input calibration short (sold in pairs) 34330A 30 A current shunt E2308A 5 k thermistor probe Y1133A Low-thermal external digital multimeter scanning kit Rackmount kits 34190A Rackmount kit Designed for use with only one instrument, mounted on either the left or the right side of the rack. 34191A 2U Dual flange kit Secures the instrument to the front of the rack. This kit can be used with the 34194A dual lock link kit to mount two half-width, 2U height instruments side-by side. 34194A Dual lock link kit Recommended for side-by-side combinations and includes links for instruments of different depths. This kit can be used with the 34191A 2U dual flange kit to mount two half-width, 2U height instruments side-by-side. Other accessories 34131A Hard transit case 34161A Accessory pouch 34398A RS-232 cable, 9 pin (f) to 9 pin (f) E5810B LAN/GPIB gateway Measurement Characteristics, continued System speeds Configuration rates 26/s to 50/s Autorange rate (dc Volts) > 30/s ASCII readings to RS-232 55/s ASCII readings to GPIB 1000/s Maximum internal trig rate 1000/s Max. ext trig. rate to mem 1000/s Triggering and memory Reading HOLD sensitivity 10%, 1%, 0.1%, or 0.01% of range Samples/trigger 1 to 50,000 Trigger delay 0 to 3600 s: 10 µs step size External trigger delay < 1 ms External trigger jitter < 500 µs Memory 512 readings Math functions NULL, min/max/average, dBm, dB, limit test (with TTL output) Standard programming languages SCPI (IEEE-488.2), Keysight 3478A, Fluke 8840A/42A Accessories included Test lead kit with probe, alligator and grabber attachments Operating manual, service manual, test report and power cord General specifications Power supply 100 V/120 V/220 V/240 V ± 10% Power line frequency 45 to 66 Hz and 360 to 440 Hz, automatically sensed at power-on Power consumption 25 VA peak (10 W average) Operating environment Full accuracy for 0 to 55 °C, full accuracy to 80% R.H. at 40 °C Storage temperature –40 to 70 °C Weight 3.6 kg (8.0 lbs) Safety Designed to CSA, UL-1244, IEC-348 RFI and ESD MIL-461C, FTZ 1046, FCC Vibration and shock MIL-T-28800E, Type III, Class 5 (sine only) Warranty 3 year
For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-02-10-16) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System 06 | Keysight | M9037A PXIe Embedded Controller - Data Sheet From Hewlett-Packard through Agilent to Keysight For more than 75 years, we‘ve been helping you unlock measurement insights. Our unique combination of hardware, software and people can help you reach your next breakthrough. Unlocking measurement insights since 1939.
1939 THE FUTURE
This information is subject to change without notice. © Keysight Technologies, 2016 Published in USA, February 22, 2016 5968-0162EN www.keysight.com myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. Three-Year Warranty www.keysight.com/find/ThreeYearWarranty Keysight’s committed to superior product quality and lower total cost of ownership. Keysight is the only test and measurement company with three-year warranty standard on all instruments, worldwide. And, we provide a one-year warranty on many accessories, calibration devices, systems and custom products. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Infoline www.keysight.com/find/service Keysight’s insight to best in class information management. Free access to your Keysight equipment company reports and e-library. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. www.keysight.com/find/34401A