Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833 07/18/2011 603-778-2300 www.wallindustries.com Fax 603-778-9797 Page 1 of 5 DATASHEET Rev. A WMH PU12 SERIES 90~264VAC Input Voltage Range

12 Watt, Single Outputs

Class II Limited Power Source US, UK, and EU Plugs Medical AC/DC Wall Mount Power Supplies

FEATURES

  • Class II Limited Power Source
  • RoHS Compliant
  • Energy Star 2.0, Efficiency Level V Compliant
  • Single Output
  • Wide Input Voltage Range: 90~264VAC, 47~63Hz
  • Optional Output Connectors
  • Over Load and Over Temperature Protection
  • US, UK, and EU Types Available
  • US Plug Medical Approvals: UL/cUL 60601-1: 2 nd Edition and TUV/T-mark EN60601-1:2 nd Edition
  • UK and EU Plug Medical Approvals: TUV/T-mark EN60601-1: 2 nd Edition
  • Meets FCC Part-18 Class B and CISPR-11 EN55011 Class B Emission Limits

DESCRIPTION

The WMHPU12 series of Class II medical AC/DC wall mount power supplies provides up to 12 W atts of continuous output power. This series consists of single output models from 5VDC to 27VDC with a 90~264VAC input voltage range. These supplies also have over load and over temperature protection. This series has UL/cUL 60601-1: 2 nd Edition (US plug type only) and TUV/T-mark EN60601-1: 2 nd Edition (US, UK, and EU types) medical approvals. All supplies are RoHS and Energy Star 2.0, Efficiency Level V compliant. These supplies also meet new CE requirements and are well suited for use in health care applications. The WMHPU12 series has three types of plugs available: US (A Type), UK (B Type), and EU (C Type). Optional output connectors are also available. Please call factory for ordering details. US Type UK Type EU Type MEDICAL

Rev. A WMHPU12 Series US, UK, and EU Plugs Class II Limited Power Source Medical AC/DC Wall Mount Power Supplies Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833 07/18/2011 603-778-2300 www.wallindustries.com Fax 603-778-9797 Page 2 of 5 SPECIFICATIONS: WMHPU12 Series All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted. We reserve the right to change specifications based on technological advances. SPECIFICATION TEST CONDITIONS Min Nom Max Unit INPUT SPECIFICATIONS Operating Input Voltage Range 90 264 Operating Voltage Range Safety Approvals Input Voltage Range 100 240 VAC Input Frequency 47 63 Hz Low Line Io = Full Load, Vin = 115VAC 0.25 0.35 A Input Current High Line Io = Full Load, Vin = 230VAC 0.17 0.22 A Low Line Io = Full Load, 25°C, Cool Start, Vin = 115VAC 14 25 A Inrush Current High Line Io = Full Load, 25°C, Cool Start, Vin = 230VAC 28 50 A OUTPUT SPECIFICATIONS Output Voltage See Table Load Regulation Vin = 230VAC 5 % Line Regulation Io = Full Load 1 % Output Power Vin = 90 to 264VAC 0 12 W Output Current Range See Table Ripple & Noise (peak to peak) Full Load, Vin = 90VAC 1 % Transient Response Time Io = Full Load to Half Load, Vin = 100VAC 4 ms Hold-Up Time Io = Full Load, Vin = 110VAC 10 ms Start-Up Time Io = Full Load, Vin = 100VAC 0.3 1 2 s Temperature Coefficient -0.04 +0.04 %/°C PROTECTION Over Current Protection 110 150 % Over Temperature Protection (See Note 2) Verified by design/characteristics of integrated controller -25 130 °C GENERAL SPECIFICATIONS Efficiency Io = Full Load, Vin = 230VAC 77.7 % Dielectric Withstanding Voltage Primary to Secondary 5656 VDC ENVIRONMENTAL SPECIFICATIONS Operating Temperature Derates linearly from 100% Load at 40°C to 50% load at 70°C 0 40 70 °C Storage Temperature -40 85 °C Operating Humidity 0 95 % Storage Humidity 0 75 % MTBF Operating Temp. at 25°C, calculated per MIL-HD BK-217F 100,000 hours PHYSICAL SPECIFICATIONS US Plug (A Type) 4.6oz (130g) UK Plug (B Type) 5.3oz (150g) Weight EU Plug (C Type) 4.9oz (140g) US Plug (A Type) 2.52 x 2.17 x 2.10 inches (64.0 x 55.0 x 53.3 mm) UK Plug (B Type) 2.52 x 2.17 x 2.37 inches (64.0 x 55.0 x 60.2 mm) Dimensions (L x W x H) EU Plug (C Type) 2.52 x 2.17 x 2.70 inches (64.0 x 55.0 x 68.6 mm) SAFETY & EMI US Plug (A Type) UL/cUL 60601-1: 2 nd Edition TUV/T-mark EN60601-1: 2 nd Edition Safety Approvals UK & EU Plugs (B & C Types) TUV/T-mark EN60601-1: 2 nd Edition EMI Requirements for CISPR-11 Vin = 220VAC B Class EMI Requirements for FCC PART-18 Vin = 110VAC B Class

Rev. A WMHPU12 Series US, UK, and EU Plugs Class II Limited Power Source Medical AC/DC Wall Mount Power Supplies Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833 07/18/2011 603-778-2300 www.wallindustries.com Fax 603-778-9797 Page 3 of 5 MODEL SELECTION TABLE Model Number Input Voltage Range Output Voltage Output Current Total Regulation (3) Output Power Plug Type WMHPU12A-102 90 ~ 264 VAC 5 ~ 6 VDC 2.00 ~ 1.66 A 5% 10W WMHPU12A-103 90 ~ 264 VAC 6 ~ 8 VDC 2.00 ~ 1.50 A 5% 12W *WMHPU12A-104 90 ~ 264 VAC 8 ~ 11 VDC 1.50 ~ 1.09 A 5% 12W *WMHPU12A-105 90 ~ 264 VAC 11 ~ 13 VDC 1.09 ~ 0.92 A 5% 12W *WMHPU12A-106 90 ~ 264 VAC 13 ~ 16 VDC 0.92 ~ 0.75 A 5% 12W *WMHPU12A-107 90 ~ 264 VAC 16 ~ 21 VDC 0.75 ~ 0.57 A 3% 12W *WMHPU12A-108 90 ~ 264 VAC 21 ~ 27 VDC 0.57 ~ 0.45 A 3% 12W US WMHPU12B-102 90 ~ 264 VAC 5 ~ 6 VDC 2.00 ~ 1.66 A 5% 10W WMHPU12B-103 90 ~ 264 VAC 6 ~ 8 VDC 2.00 ~ 1.50 A 5% 12W *WMHPU12B-104 90 ~ 264 VAC 8 ~ 11 VDC 1.50 ~ 1.09 A 5% 12W *WMHPU12B-105 90 ~ 264 VAC 11 ~ 13 VDC 1.09 ~ 0.92 A 5% 12W *WMHPU12B-106 90 ~ 264 VAC 13 ~ 16 VDC 0.92 ~ 0.75 A 5% 12W *WMHPU12B-107 90 ~ 264 VAC 16 ~ 21 VDC 0.75 ~ 0.57 A 3% 12W *WMHPU12B-108 90 ~ 264 VAC 21 ~ 27 VDC 0.57 ~ 0.45 A 3% 12W UK WMHPU12C-102 90 ~ 264 VAC 5 ~ 6 VDC 2.00 ~ 1.66 A 5% 10W WMHPU12C-103 90 ~ 264 VAC 6 ~ 8 VDC 2.00 ~ 1.50 A 5% 12W *WMHPU12C-104 90 ~ 264 VAC 8 ~ 11 VDC 1.50 ~ 1.09 A 5% 12W *WMHPU12C-105 90 ~ 264 VAC 11 ~ 13 VDC 1.09 ~ 0.92 A 5% 12W *WMHPU12C-106 90 ~ 264 VAC 13 ~ 16 VDC 0.92 ~ 0.75 A 5% 12W *WMHPU12C-107 90 ~ 264 VAC 16 ~ 21 VDC 0.75 ~ 0.57 A 3% 12W *WMHPU12C-108 90 ~ 264 VAC 21 ~ 27 VDC 0.57 ~ 0.45 A 3% 12W EU NOTES 1. Models with a “*” next to the model number are approved by CEC Level V. 2. Over Temperature Protection: Thermal shutdown by junction temperature controller; the adapter will automatically recover when the fault condition is removed. 3. Models WMHPU12-102~104 needs to use AWG#18/6FT output cable in order to meet the total regulation specified. Models WMHPU12-105~108 needs to use AWG#20/6FT output cable in order to meet the total regulation specified. The regulation and efficiency will change if a different output cable is used. 4. Optional output connectors are available. Please call factory for ordering details. DERATING CURVE Ambient Temperature (°C) Output Power (%)

Rev. A WMHPU12 Series US, UK, and EU Plugs Class II Limited Power Source Medical AC/DC Wall Mount Power Supplies Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833 07/18/2011 603-778-2300 www.wallindustries.com Fax 603-778-9797 Page 4 of 5 MECHANICAL DRAWINGS POWER LED (OPTIONAL) Unit: inches (mm) 1.22±0.04 (31.1±1.0) 0.19±0.04 (4.7±1.0) 0.25 (6.3) 0.31 (8.0) POWER LED (OPTIONAL) 1.22±0.04 (31.1±1.0) 0.25±0.04 (6.4±1.0) POWER LED (OPTIONAL) 1.22±0.04 (31.1±1.0) 0.73±0.04 (18.5±1.0) Unit: inches (mm) Unit: inches (mm) US TYPE (SUFFIX “A”) UK TYPE (SUFFIX “B”) EU TYPE (SUFFIX “C”)

Rev. A WMHPU12 Series US, UK, and EU Plugs Class II Limited Power Source Medical AC/DC Wall Mount Power Supplies Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833 07/18/2011 603-778-2300 www.wallindustries.com Fax 603-778-9797 Page 5 of 5 COMPANY INFORMATION Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one example of our commitment to producing a high quality, well-documented product for our customers. Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects. Contact Wall Industries for further information: Phone : (603)778-2300 Toll Free : (888)597-9255 Fax : (603)778-9797 E-mail : sales@wallindustries.com Web : www.wallindustries.com Address : 5 Watson Brook Rd. Exeter, NH 03833