TCS MSYSTEM | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 4

Technical content

MODEL: TCS TCS SPECIFICATIONS ES-1331-E Rev.16 Page 1/4 http://www.m-system.co.jp/ Plug-in Signal Conditioners M-UNIT THERMOCOUPLE TRANSMITTER Functions & Features

  • Accepting direct input from a thermocouple and providing a standard process signal
  • 7-segment linearization
  • Burnout protection
  • High-accuracy cold junction compensation
  • Isolation up to 2000 V AC
  • Fast response type available
  • LCD meter (engineering unit display selectable)
  • Simple loop test output (0 % and 100 %)
  • High-density mounting Typical Applications
  • High-accuracy cold junction compensation benefits narrow span measurements
  • 0.1 μA burnout sensing enables long distance transmission with minimum offset drifts
  • Electric furnace (isolation)
  • No burnout type can connect to a single T/C in parallel with a recorder
  • Power plant (2000 V AC isolation, 110 V DC power supply) (1.97) 127 (5) (3.15) mm (inch) MODEL: TCS–[1][2]–[3][4]

ORDERING INFORMATION

  • Code number: TCS-[1][2]-[3][4] Specify a code from below for each of [1] through [4]. (e.g. TCS-2A-B/E2/BL/Q)
  • Temperature range (e.g. 0 – 800°C)
  • Special output range (For codes Z & 0)
  • Specify the specification for option code /Q (e.g. /C01/S01) [1] INPUT THERMOCOUPLE 1: (PR) (Usable Range 0 to 1760°C, 32 to 3200°F) 2: K (CA) (Usable range -270 to +1370°C, -454 to +2498°F) 3: E (CRC) (Usable range -270 to +1000°C, -454 to +1832°F) 4: J (IC) (Usable range -210 to +1200°C, -346 to +2192°F) 5: T (CC) (Usable range -270 to +400°C, -454 to +752°F) 6: B (RH) (Usable range 0 to 1820°C, 32 to 3308°F) 7: R (Usable range -50 to +1760°C, -58 to +3200°F) 8: S (Usable range -50 to +1760°C, -58 to +3200°F) N: N (Usable range -270 to +1300°C, -454 to +2372°F) 0: Specify [2] OUTPUT Current A: 4 – 20 mA DC (Load resistance 750 Ω max.) B: 2 – 10 mA DC (Load resistance 1500 Ω max.) C: 1 – 5 mA DC (Load resistance 3000 Ω max.) D: 0 – 20 mA DC (Load resistance 750 Ω max.) E: 0 – 16 mA DC (Load resistance 900 Ω max.) F: 0 – 10 mA DC (Load resistance 1500 Ω max.) G: 0 – 1 mA DC (Load resistance 15 kΩ max.) Z: Specify current (See OUTPUT SPECIFICATIONS) Voltage 1: 0 – 10 mV DC (Load resistance 10 kΩ min.) 2: 0 – 100 mV DC (Load resistance 100 kΩ min.) 3: 0 – 1 V DC (Load resistance 100 Ω min.) 4: 0 – 10 V DC (Load resistance 1000 Ω min.) 5: 0 – 5 V DC (Load resistance 500 Ω min.) 6: 1 – 5 V DC (Load resistance 500 Ω min.) 0: Specify voltage (See OUTPUT SPECIFICATIONS) [3] POWER INPUT AC Power B: 100 V AC C: 110 V AC D: 115 V AC F: 120 V AC G: 200 V AC (Not selectable with Option /UL) H: 220 V AC (Not selectable with Option /UL) J: 240 V AC (Not selectable with Option /UL) DC Power S: 12 V DC (Not selectable with Option /UL) R: 24 V DC (Not selectable with Option /UL) V: 48 V DC (Not selectable with Option /UL) P: 110 V DC (Not selectable with Option /E2 or /UL) [4] OPTIONS (multiple selections) Input Signal Indicator blank: Without /E: With (0.0 - 100.0 % display) /E2: With (in engineering unit with backlight and the simple loop test output) Response Time (0 – 90 %) blank: Standard (≤ 0.5 sec.) /K: Fast response (Approx. 25 msec.)

MODEL: TCS TCS SPECIFICATIONS ES-1331-E Rev.16 Page 2/4 http://www.m-system.co.jp/ (Not selectable with Option /E2) Burnout blank: Upscale burnout /BL: Downscale burnout /BN: No burnout Standards & Approvals blank: Without UL /UL: UL approval (Not selectable with /E or /E2) Other Options blank: none /Q: Option other than the above (specify the specification) SPECIFICATIONS OF OPTION: Q (multiple selections) COATING (For the detail, refer to M-System's web site.) /C01: Silicone coating /C02: Polyurethane coating /C03: Rubber coating (UL not available) TERMINAL SCREW MATERIAL /S01: Stainless steel (UL not available) GENERAL SPECIFICATIONS Construction: Plug-in Connection: M3.5 screw terminals Screw terminal: Chromated steel (standard) or stainless steel Housing material: Flame-resistant resin (black) Isolation: Input to output to power Overrange output: Approx. -10 to +120 % at 1 – 5 V Zero adjustment: -5 to +5 % (front) Span adjustment: 95 to 105 % (front) At burnout: Downscale ≤ -10 %, Upscale ≥ 110 % Linearization: Standard Cold junction compensation: CJC sensor attached to the input terminals (B thermocouple is without CJC as standard) Simple loop test output: 0 % and 100 % signal simulated by selecting the front switch positions. (Only for option code /E2) ■ DISPLAY (Input indicator)

  • Option code: /E LCD digital display: 0.0 - 100.0 % (min. digit 0.1 %) (No scaling)
  • Option code: /E2 LCD digital display: Engineering unit Display scaling: -10000 – +10000 Decimal position: 10-1 - 10-4 or no decimal point Engineering unit: %, μV, mV, V, mA, A, °C, °F, Ω, DEG K, mHz, Hz, kHz, VAC, AAC, mg, g, kg, t, rpm or rps selectable Back light: Green at normal, red at loop test output enable Factory setting: scaling 0.00 - 100.00, unit: % INPUT SPECIFICATIONS Minimum span: 3 mV Offset: Max. 1.5 times span Input resistance: 30 kΩ min. Burnout sensing: 0.1 μA Minimum span (in °C) (PR): min. span 370°C K (CA): min. span 75°C E (CRC): min. span 50°C J (IC): min. span 60°C T (CC): min. span 75°C B (RH): min. span 780°C R: min. span 360°C S: min. span 380°C N: min. span 110°C Minimum span (in °F) (PR): min. span 670°F K (CA): min. span 140°F E (CRC): min. span 90°F J (IC): min. span 110°F T (CC): min. span 140°F B (RH): min. span 1410°F R: min. span 650°F S: min. span 690°F N: min. span 200°F Note: The described accuracy may be partially not satisfied when the temperature ranges below 0°C. Consult factory. OUTPUT SPECIFICATIONS ■ DC Current: 0 – 20 mA DC Minimum span: 1 mA Offset: Max. 1.5 times span Load resistance: Output drive 15 V max. ■ DC Voltage: -10 – +12 V DC Minimum span: 5 mV Offset: Max. 1.5 times span Load resistance: Output drive 10 mA max.; 5 mA for negative voltage output; at ≥ 0.5 V INSTALLATION Power input
  • AC: Operational voltage range: rating ±10 %, 50/60 ±2 Hz, approx. 2 VA (approx. 3 VA with Option /E2)
  • DC: Operational voltage range: rating ±10 %, or 85 – 150 V for 110 V rating (ripple 10 % p-p max.) approx. 2.6 W (110 mA at 24 V; approx. 3.6 W with Option /E2) Operating temperature: -5 to +60°C (23 to 140°F)

MODEL: TCS TCS SPECIFICATIONS ES-1331-E Rev.16 Page 3/4 http://www.m-system.co.jp/ Operating humidity: 30 to 90 %RH (non-condensing) Mounting: Surface or DIN rail Weight: 350 g (0.77 lb) PERFORMANCE in percentage of span Accuracy: ±0.3 % (at over 400°C or 750°F for R, S and PR; over 770°C or 1420°F for B) Display accuracy: ± (0.3 % of FS + 1 digit) (at over 400°C or 750°F for R, S and PR; over 770°C or 1420°F for B) Simple loop test output setting accuracy: ±0.5 % Cold junction compensation error (at 20°C ±10°C or 68°F ±18°F) K, E, J, T & N: ±0.5°C or ±0.9°F S, R & PR: ±1°C or ±1.8°F Temp. coefficient: ±0.015 %/°C (±0.008 %/°F) (at over 400°C or 750°F for R, S and PR; over 770°C or 1420°F for B) Burnout response: ≤ 10 sec. Line voltage effect: ±0.1 % over voltage range Insulation resistance: ≥ 100 MΩ with 500 V DC Dielectric strength: 2000 V AC @1 minute (input to output to power to ground) STANDARDS & APPROVALS Approval: UL/C-UL general safety requirements (UL 3111-1, CAN/CSA-C22.2 No.1010-1) EXTERNAL VIEW Span Adj. Zero Adj. LCD Meter Loop Test Selector Top Middle Bottom : 0% fixed : 100% fixed : Cancel (normal operating mode) NEXT Key ENTER Key Span Adj. Zero Adj. LCD Meter ■ OPTION /E ■ OPTION /E2

MODEL: TCS TCS SPECIFICATIONS ES-1331-E Rev.16 Page 4/4 http://www.m-system.co.jp/ EXTERNAL DIMENSIONS unit: mm (inch) 3456 2187 80 (3.15) 80 (3.15) (.79) 40 (1.57) 50 (1.97) 7.8 (.31) CLAMP (top & bottom) DIN RAIL 35mm wide2–4.5 (.18) dia. MTG HOLE 15 (.59) deep 8–M3.5 SCREW

  • When mounting, no extra space is needed between units. TERMINAL ASSIGNMENTS unit: mm (inch) 2187 7 (.28) CJC SENSOR (model: CJM) SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM Isolation Base Socket SZ Option /E, E2 Deleted with B thermocouple Option /E2 Low Drift Amplifier Output Driver 1 T / C CJC SENSOR metal leg model: CJM comp. leadwire Linearizer CJC Sensing Volt. Generator LOOP TEST SELECTOR * * OUTPUT U(+) V(–) POWER Specifications are subject to change without notice.