SMFA30 SERIES 30W Open-Frame Power Supply for Medical Equipment – Wide Input Voltage 90 to 264 VAC, 47 to 63Hz – Internal EMI Filter – Single Output – Size: 2"x4" x1.16" – Class I Insulation – Input connector mates with Molex housing 09-50-3031 and Molex 2478 series crimp terminal. – Output connector mates with Molex housing 09-50-3061 and Molex 2478 series crimp terminal. – Output Voltage Available from 5VDC through 36VDC – Input Surge Current, Over Load and Over Voltage Protection

3 Year Warranty Approvals:

Product Number Output Voltage Output Current Total Regulation Maximum Output Power SMFA30-S02 5 VDC 6.00 A 5% 30W SMFA30-S03 7 VDC 4.28 A 5% 30W SMFA30-S04 9 VDC 3.33 A 5% 30W SMFA30-S05 12 VDC 2.50 A 5% 30W SMFA30-S06 15 VDC 2.00 A 5% 30W SMFA30-S07 18 VDC 1.66 A 5% 30W SMFA30-S08 24 VDC 1.25 A 5% 30W SMFA30-S09 30 VDC 1.00 A 5% 30W SMFA30-S10 36 VDC 0.83 A 3% 30W

Electrical Characteristics

Parameter Test Conditions Min. Typ. Max. Unit Input Voltage Operating Voltage 90 264 VAC Input Frequency 47 63 Hz Output Power Range Vin= 90 to 264 VAC 0 30 W Input Current (Low Line) Io=Full load, Vin=115 VAC 0.7 A Input Current (High Line) Io=Full load, Vin= 230 VAC 0.3 0.5 A Low Line Inrush Current Io=Full load, 25° C, Cool start, Vin=115VAC 15 18 A High Line Inrush Current Io=Full load, 25° C, Cool start, Vin=230 VAC 21 25 A Efficiency Io=Full Load, Vin=230VAC 70 80 85 % Line Regulation Io=Full Load 0.5 1 % Load Regulation Vin=230VAC 3 7 % Over Voltage Protection 112 132 Over Current Protection 110 150 % Transient Response Io=Full Load to Half Load, Vin=100VAC 4 mS Hold-Up Time Io=Full Load, Vin=110VAC 10 12 mS Start Up Time Io=Full Load, Vin=100VAC 0.3 1 2 S Ripple & Noise (Peak to Peak) Full Load, Vin=90VAC 0 .5 1 % Safety Ground Leakage Current Io= Full Load, Vin=240VAC 0.1 0.3 mA Temperature Coefficient All output -0.04 0.04 %/° C

Parameter Test Conditions Min. Typ. Max. Unit Operating Temperature 0 50 70 °C Storage Temperature -40 85 °C Relative Humidity 5 95 % Operation temperature at 25° C calculated per MIL-HDBK-217F 0.1M Hrs Derate linearly from 100% load at 50° C to 50% load at 70° C Approvals and Compliances Parameter Test Conditions * Min. Unit Dielectric Withstanding Voltage for Primary to secondary Primary to secondary 5656 VDC Dielectric Withstanding Voltage for Primary to Ground Primary to ground 2828 VDC Isolation Resistance Test Voltage=500VDC 50 MΩ EMI requirements for CISPR-11 Vin=220VAC B CLASS EMI requirements for FCC PART-18 Vin=110VAC B CLASS *Recommend to be used on the metal chassis. Mechanical and PIN out PIN CHART PIN MODEL 1 2 3 4 5 6 SMFA30-SXX RTN RTN RTN Vout Vout Vout Note: 1. Dimensions are shown in mm & inch 2. Weight: approx. 200g 3. Input connector mates with Molex housing 09-50-3031 and Molex 2478 series crimp terminal. 4. Output connector mates with Molex housing 09-50-3061 and Molex 2478 series crimp terminal.