SHFA60 Vitec | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 2
Technical content
SHFA60 SERIES 60W Open-Frame Power Supply for Medical Equipment Wide Input Voltage 90 to 264 VAC, 47 to 63Hz Output Voltage Available From 12VDC Thru 40VDC Input connector mates with molex housing 09-50-3031/35977-0300 and molex 2478/ 35922 series crimp terminal. Output connector mates with molex housing 09-50-3061/35977-0600 and molex 2478/ 35922 series crimp terminal. Single Output Size: 2"x4"x1" Class I Insulation Input Surge Current, Over Load and Over Voltage Protection Solution for Health Care Equipment
2 Year Warranty Approvals:
Product Number Output Voltage Output Current Total Regulation Maximum Output Power SHFA60-S05 12 ~ 13 VDC 5.00 ~ 4.61 A 5% 60W SHFA60-S06 13 ~ 16 VDC 4.61 ~ 3.75 A 5% 60W SHFA60-S07 16 ~ 21 VDC 3.75 ~ 2.85 A 5% 60W SHFA60-S08 21 ~ 27 VDC 2.85 ~ 2.22 A 3% 60W SHFA60-S09 27 ~ 33 VDC 2.22 ~ 1.81 A 3% 60W SHFA60-S10 33 ~ 40 VDC 1.81 ~ 1.50 A 3% 60W
Electrical Characteristics
Parameter Test Conditions Min. Typ. Max. Unit Input Voltage Operating Voltage 90 264 VAC Input Frequency 47 63 Hz Output Power Range Vin= 90 to 264 VAC 0 60 W Input Current (Low Line) Io=Full load, Vin=115 VAC 1.62 A Input Current (High Line) Io=Full load, Vin= 230 VAC 0.3 0.72 A Low Line Inrush Current Io=Full load, 25° C, Cool start, Vin=115VAC 26 30 A High Line Inrush Current Io=Full load, 25° C, Cool start, Vin=230 VAC 43 47 A Efficiency Io=Full Load, Vin=230VAC 85 % Line Regulation Io=Full Load 1 % Load Regulation Vin=230VAC 5 % Over Voltage Protection 112 132 % Over Current Protection 110 150 % Transient Response Io=Full Load to Half Load, Vin=100VAC 4 mS Hold-Up Time Io=Full Load, Vin=110VAC 12 16 mS Start Up Time Io=Full Load, Vin=100VAC 0.3 1 2 S Ripple & Noise (Peak to Peak) Full Load, Vin=90VAC 1 % Safety Ground Leakage Current Io= Full Load, Vin=240VAC 0.1 0.3 mA Temperature Coefficient All output -0.04 0.04 %/° C
Parameter Test Conditions Min. Typ. Max. Unit Operating Temperature 0 40 70 °C Storage Temperature -40 85 °C Relative Humidity 5 95 % Operation temperature at 25℃, calculated per MIL-HDBK-217F 0.1M Hrs Derate linearly from 100% load at 40° C to 50% load at 70° C Approvals and Compliances Parameter Test Conditions Min. Unit Dielectric Withstanding Voltage for Primary to secondary Primary to secondary 5656 VDC Dielectric Withstanding Voltage for Primary to Ground Primary to ground 2828 VDC Isolation Resistance Test Voltage = 500VDC 50 MΩ EMI requirements for CISPR-11 Vin=220VAC B CLASS EMI requirements for FCC PART-18 Vin=110VAC B CLASS Mechanical and PIN out PIN CHART PIN MODEL 1 2 3 4 5 6 SHFA60-SXX OUT OUT OUT RTN RTN RTN Note: 1. Dimensions are shown in mm & inch 2. Weight: approx. 140g 3. Input connector mates with molex housing 09-50-3031/35977-0300 and molex 2478/ 35922 series crimp terminal. 4. Output connector mates with molex housing 09-50-3061/35977-0600 and molex 2478/ 35922 series crimp terminal.